AM-640 Series Mat Table
This manual has been prepared for the owners and operators of the ARMEDICA™ AM-Series Mat Tables. It contains installation instructions, precautionary instructions and maintenance procedures for the following model numbers:
AM-640, AM-641, AM-642, AM-643, AM-644,
AM-645, AM-650, AM-652 AM-654,
AM-660, AM-662, AM-664
Armedica Treatment Tables are intended to be used as physical therapy supports for patients during clinician attended physical therapy related diagnosis, treatment, and monitoring. The able is equipped with vertical table height adjustment actuators. Patients are not to be left unattended.
INSTALLATION INSTRUCTIONS FOR ARMEDICA™ MAT PLATFORM TABLES
-
Remove the mat platform from the carton, leaving the foam covering on the upholstered top.
-
Place the platform top with the bottom side up on a clean, smooth surface and remove the wood packing frame and strapping from the motor on electric models. On adjustable models with a manual crank, remove strapping and install the unit referring to the picture of model AM-642 (pg 4).
-
Remove the screws (A) holding the legs to the underside of the table and carefully unwrap.
CAUTION:
ON ELECTRIC AND ADJUSTABLE MODELS DO NOT FOLD OR KINK HYDRAULIC LINES AT ANY TIME -
Gently rotate legs into position as shown and insert bolts (3) with the bolt head and finishing cap washers (27) on the outside of the table frame. Tighten bolts securely using the nuts (13) provided. Attach the black finishing caps (28) to bolt heads by pressing firmly. Extra cable clamps and screws are included for additional cable clamping if necessary.
-
Install leveling glides (33) by screwing them into the legs.
-
Place the mat platform upright and remove the foam covering. Adjust the leveling glides as needed. The mat platform is ready for use. For powered models plug the power cord into a properly grounded 120 volt AC outlet.
Power cord is hospital grade and reliability of unit can only be achieved when connected to an equivalent receptacle.
OPERATING INSTRUCTIONS
To adjust the height of the Armedica Power Mat Platform, a low voltage toggle
switch or optional foot switch is used to activate the motorized hydraulic
pump. The up (UP) or down (DN) motion is clearly indicated on the foot switch.
The motors are equipped with a thermal cutoff that protects them from
overheating.
Duty cycle on/off int. 1 min. / 5 min.
Adjustable Mat Platform is adjusted with a crank handle that activates
the hydraulic pump. Turning the handle clockwise will raise the table. Counter
clockwise will lower it.
The Armedica™ The handle will automatically lock into place when extended for
use. Make certain the handle is fully engaged prior to turning. To release
and lower the handle once the desired height is reached, firmly pull the
handle out until it is able to be lowered.
The moveable section on the adjustable back model is activated by grasping the
handle located at the center of the back section. Release to lock at desired
position.
This equipment has been tested and found to comply with the limits for medical
devices to the IEC 60601-1-2: Third edition 2007-03. These limits are designed
to provide reasonable protection against harmful interference in a typical
medical installation. This equipment generates, uses and can radiate radio
frequency energy and, if not installed and used in accordance with the
instructions, may cause harmful interference to other devices in the vicinity.
However, there is no guarantee that interference will not occur in a
particular installation. If this equipment does cause harmful interference to
other devices, which can be determined by turning the equipment off and on,
the user is encouraged to try to correct the interference by one or more of
the following measures:
- Reorient or relocate the receiving device.
- Increase the separation between the equipment.
- Connect the equipment into an outlet on a circuit different from that to which the other device(s) are connected.
- Consult the manufacturer or field service technician for assistance.
MAINTENANCE
The Armedica™ Electric Power Mat Platforms and Adjustable Mat Platforms are
equipped with a maintenance free hydraulic system.
On all mat platforms frequently check to make certain that all hardware (nuts,
bolts, etc.) are securely fastened.
The vinyl cover can be cleaned with mild soap and water and the frame wiped
with a dry cloth.
SYMBOLS AND INDICATORS
Attention: Consult accompanying documents.
Type B Equipment: “An adequate degree of protection against electric shock is
provided, particularly regarding leakage currents and reliability of the
protective earth connection.”
TROUBLE SHOOTING
TABLE WILL NOT OPERATE
A. Check to see if the power cord is properly plugged into the electrical
outlet.
B. Make sure the hand switch or foot switch is securely plugged in.
C. Thermal cutoff may have activated. Allow motor to cool.
MOTOR SPECIFICATIONS
Bucher Hydraulics, Inc Model 4E-D2A-10-XX028
1363 Michigan St. N.E. 115 VAC 2.4 Amps
Grand Rapids, MI 49503 Short Duty on/off 1 min. / 5 min
Transport and Storage
Ambient Temperature: -34 to 65 deg. C
Relative Humidity: 15 to 95%, including condensation
Atmospheric Pressure: 500hPa to 1030hPa
MAT PLATFORM AM-644, AM-654, AM-664
MAT PLATFORM AM-642, AM-652, AM-662
MAT PLATFORM AM-640, AM-650, AM-660
MAT PLATFORM AM-641, AM-643, AM-645
ARMEDICA™ MANUFACTURING CORP. LIMITED WARRANTY
ARMEDICA™ Manufacturing Corp. warrants that the AM-Series Mat Platforms are free from defects in material and workmanship. This warranty shall remain in effect for 18 months from the date of original consumer purchase of the product. The Hydraulic units on the Power Mat Platforms and Adjustable Mat Platforms are warranted to be free from defects in materials and workmanship for 18 months from the date of original consumer purchase of the product. The welded steel frames on all Mat Platforms are warranted to be free from defects in materials or workmanship for 5 years from the date of original consumer purchase of the product. If the table fails to function during the warranty period due to a defect in material or workmanship, ARMEDICA™ Manufacturing Corp. or the selling Dealer will repair or replace the table without charge within a 30 day period from the date on which the table is returned to ARMEDICA™ Manufacturing Corp. or the selling Dealer.
THIS WARRANTY DOES NOT COVER
- Replacement parts or labor furnished by anyone other than ARMEDICA, Selling Dealer or any approved ARMEDICA Service Agent. Parts lists are for reference only when speaking with an ARMEDICA representative. Replacements and repairs are to be done by an approved ARMEDICA Service Agent.
- Any failure of the table during the warranty period if the failure is not caused by a defect in material or workmanship or if the failure is caused by unreasonable use, including the failure to provide reasonable and necessary maintenance.
ARMEDICA™ MANUFACTURING CORP. IS NOT LIABLE FOR INCIDENTAL OR CONSEQUENTIAL
DAMAGE TO PROPERTY OR BUSINESS.
TO OBTAIN SERVICE from ARMEDICA™ or the Selling Dealer under this warranty,
please do the following:
- A written claim should be sent to ARMEDICA Manufacturing Corp., P.O. Box 880, Greenwood, AR. 72936-0880 or the selling Dealer.
- The table must be returned to ARMEDICA™ Manufacturing Corp. or the selling Dealer.
- This warranty gives you specific legal rights and you may have other rights which vary from state to state.
ARMEDICA™ does not authorize any person or dealer to create for it any other obligations or liabilities in connection with the sale of the tables. Any representation or agreement not contained in this warranty shall be void. Disposal of unused or unwanted product should be in accordance with applicable regional, national, and local laws. For information contact the local regulatory agency.
PN-03433 Rev D 12/10/13
Mat Tables Manual