TECHNICAL BULLETIN TB-3092
Full-Time Continuous Monitor
Installation, Operation and Maintenance Made in the
United States of America
19330 Full Time Continuous Monitor
Description
The Desco Full-Time Continuous Monitor is a single workstation continuous
monitor. It continuously monitors the wiring of an AC outlet, path-to-ground
integrity of one operator, and the ground plane of one ESD worksurface.
Mounting tabs make it suitable to install in workbench or equipment settings.
The embedded relay terminal activates when a pass condition is displayed on
the monitor and can be connected to machines for access control or other
devices. Connect the 19332 Light Tower accessory for increased alarm
visibility.
Connect the 19333 Power Relay accessory to control power to workbenches, hand
tools, and other electronic devices dependent on the Full-Time Continuous
Monitor’s status.
The Full-Time Continuous Monitor is designed with Wave Distortion Technology
to provide stable continuous monitoring of the path-to-ground and presence of
1 megohm resistance of an operator’s wrist strap. Audio and visual alarms
activate (in less than 500 milliseconds) when the operator unintentionally
disconnects from the monitor or the wrist strap connection points are
intermittent. The Full-Time Continuous Monitor also monitors for a worksurface
path-to-ground of less than 10 megohms and the
integrity of the electrical ground used to by the monitor.
All Full-Time Continuous Monitors are calibrated to NIST standards and include
a certificate.
Wave Distortion Technology is the Most Reliable Technology for the Continuous
Monitoring of Single-Wire (Conductor) Wrist Straps “Wave Distortion” or vector
impedance works by applying a continuous test voltage of 1.2 volts peak-to-
peak at 1 to 2 microamperes (0.000002 amperes) to the wrist strap that is
connected to the continuous or constant monitor. The test voltage creates a
square wave that the monitor circuit compares to established patterns. By
monitoring the “distortions”, or shape of the square wave, Wave Distortion
Technology determines if the monitored circuit is complete – the operator is
in the circuit and the total equivalent DC resistance is within
specifications. Wave Distortion Technology produces a very fast alarm time and
minimal false alarms.
ANSI/ESD S20.20 section 7.3 states “Compliance verification records shall be
established and maintained to provide evidence of conformity to the technical
requirements.” Per ANSI/ESD S1.1 Annex A.3 Daily (wrist strap system) testing
may be omitted if constant monitoring used.” Per ESD Handbook ESD TR 20.20
section 5.3.2.4.4 “Typical Test programs recommend that wrist straps that are
used daily should be tested daily. However, if the products that are being
produced are of such value that knowledge of a continuous, reliable ground is
needed, and then continuous monitoring should be considered or even required.”
The Full-Time Continuous Monitor and its accessories are available as the following item numbers:
Item | Description |
---|
19330 | Full-Time Continuous Monitor, with North
America Power Cord
19331 | Full-Time Continuous Monitor, with UK Power
Cord
19332 | Light Tower
19333 | Power Relay, North America
19334 | Replacement Power Relay Interface Cord
19327 | Replacement Power Cord, North America
19328 | Replacement Power Cord, UK
98221 | Wave Distortion Monitor Verification Tester
09842 | Multi-Ground Hub
Packaging
19330 Full-Time Continuous Monitor
1 Full-Time Continuous Monitor
1 Mat Monitor Cord (White)
1 Countersunk Washer
1 Flat Head Screw, 6-32 x 1/4″
2 Pan-Head Screws, #6 x 3/8″
1 Dual Lock Fastener Set, 2″ Length
1 Power Cord with North American Plug
1 Certificate of Calibration
19331 Full-Time Continuous Monitor
1 Full-Time Continuous Monitor
1 Mat Monitor Cord (White)
1 Countersunk Washer
1 Flat Head Screw, 6-32 x 1/4″
2 Pan-Head Screws, #6 x 3/8″
1 Dual Lock Fastener Set, 2″ Length
1 10mm Interchangeable Park Snap
1 10mm Operator Jack Adapter
1 Power Cord with UK Plug
1 Certificate of Calibration
19332 Light Tower
1 Light Tower with Stereo Cord
4 Pan-Head Screws, #8 x 3/4″
Features and Components
A. Monitored Ground Terminal: Provides a verified ground point. Bond to
the 09842 Multi-Ground Hub to create up to 12 ground points.
B. Light Tower Jack: Connects to the 19332 Light Tower accessory.
- +12VDC
- Green (open collector)
- Red (open collector)
C. Park Snap: Allows the operator to disconnect (and park) a wrist strap
when leaving the workstation, without creating a false alarm. Touch the park
snap or park switch, and the monitor’s audible alarm will disable for
approximately 8 seconds. Use this time to disconnect the wrist cord from the
wristband, and attach it to the park snap or remove the banana plug from the
jack.
Upon returning to the workstation and disconnecting the wrist cord from the
park snap, the operator has approximately 8 seconds to reconnect the wrist
cord to a worn wristband before activating the alarm.
D. Monitored Operator Jack: Insert the single-wire wrist cord here to
monitor the operator’s connection to ground.
E. Operator LEDs: Illuminates green when the operator is properly
grounded. Illuminates solid red and audible alarm sounds when the operator is
not properly grounded.
F. Mat LEDs: Illuminates green when the worksurface is properly grounded.
Illuminates red and audible alarm sounds when the worksurface is not properly
grounded.
G. Ground LEDs: Illuminates green when the AC outlet is properly wired
and its path to equipment ground via the equipment ground conductor is intact.
Illuminates red and audible alarm sounds when the AC outlet is not properly
wired and its path to equipment ground via the equipment ground conductor is
broken.
H. Park Switch: Allows the operator to disconnect (and park) a wrist
strap when leaving the workstation, without creating a false alarm. Touch the
park snap or park switch, and the monitor’s audible alarm will disable for
approximately 8 seconds. Use this time to disconnect the wrist cord from the
wristband, and attach it to the park snap or remove the banana plug from the
jack.
Upon returning to the workstation and disconnecting the wrist cord from the
park snap, the operator has approximately 8 seconds to reconnect the wrist
cord to a worn wristband before activating the alarm. The 8 second timer is
indicated by the growing intensity of the park switch’s blue LED. The blue LED
will pulse when the operator jack is vacant to indicate standby mode.
I. Power Inlet: Connect the included power cord here.
J. Monitored Mat Terminal: Verifies that the worksurface’s path-to-ground
is less than 10 megohms.
Connect the white mat monitor cord here.
K. Set Switch: Toggles the monitor’s settings for the audible alarm and
mat monitor circuit.
Function | Procedure |
---|---|
Audible Alarm Volume | 1. Push and release the switch to toggle the buzzer’s |
volume.
2. The volume will toggle from low, medium, and high settings.
Enable / Disable Mat Monitor Circuit| 1. Push and hold the switch until all
LEDs illmuniate orange.
2. Release the switch.
3. An illuminated MAT LED indicates that the mat monitor circuit is enabled.
A non-illuminated MAT LED
indicates that the mat monitor circuit is disabled.
L. Relay Terminal: Integrates with electronic locks, lights, buzzers,
etc. See the “Relay Terminal” section for more information.
M. Auxilary Jack: Connects to the 19333 Power Relay accessory. See
TB-3093 for more information.
Installation
Full-Time Continuous Monitor
- Remove the Full-Time Continuous Monitor from the carton, and inspect for damage.
- Determine the mounting location of the Full-Time Continuous Monitor. Its LEDs should be visible to the operator. Secure the Full-Time Continuous Monitor to a surface using its mounting tabs and the included pan head screws or dual lock fasteners.
- Secure the ring termination of the white mat monitor cord to the monitored mat terminal on the back of the Full-Time Continuous Monitor.
- Connect the mat monitor cord from the back of the Full-Time Continuous Monitor to a grounded worksurface mat. Use the included countersunk washer and flat head screw to secure the cord to the grounded mat.
- Connect the power cord to the power inlet located on the back of the Full-Time Continuous Monitor.
Connect the opposite end to an appropriate AC outlet.
Relay Terminal
The Full-Time Continuous Monitor features a relay terminal that can be
integrated with electronic locks, lights, buzzers, etc. The relay will
activate when a failure is reported by the Full-Time Continuous Monitor.
Contact Form | 2 Form C |
---|---|
Maximum Switching Power | 60 W, 125 VAC |
Maximum Switching Voltage | 220 VDC, 250 VAC |
Maximum Switching Current | 2 A |
Maximum Carrying Current | 2 A |
10 mm Snap Adapters
The 19331 Full-Time Continuous Monitor includes an interchangeable 10 mm park
snap and 10 mm banana jack adapter for operators who use wrist cords with 10
mm terminations. Use the park snap’s knurled rim to unscrew the 4 mm park snap
from the monitor and install the 10 mm park snap to the monitor. Plug the 10
mm operator jack adapter into the monitor’s operator jack.
Operation
- Fit the wristband snugly onto the wrist.
- Snap the wrist cord to the wristband.
- Plug the wrist cord into the operator jack. The green operator LED will illuminate. This indicates that the operator is properly grounded.
- If this does not happen, examine the wrist cord for continuity or damage and the wristband to ensure that it fits securely. If you have dry skin, apply an approved dissipative hand lotion such as MendaReztore® ESD Hand Lotion.
- When leaving the workstation, first touch the monitor’s park switch or park snap. The monitor’s audible alarm will disable for approximately 8 seconds. Use this time to disconnect the wrist cord from the wristband and attach it to the park snap without sounding a false alarm.
- Upon returning to the workstation and disconnecting the wrist cord from the park snap, the monitor will provide approximately 8 seconds for the operator to reconnect the wrist cord to a worn wristband before sounding the monitor’s alarm.
Calibration
Frequency of recalibration should be based on the critical nature of those ESD
sensitive items handled and the risk of failure for the ESD protective
equipment and materials. In general, Desco recommends that calibration be
performed annually.
Use the Desco 98221 Wave Distortion Monitor Verification Tester to perform
periodic verification (once every 6-12 months) of the Full-Time Continuous
Monitor. The verification tester can be used to check the test limits of the
Full-Time Continuous Monitor without removing it from the workstation.
See TB-3074 for more
information.
Specifications
Full-Time Continuous Monitor
Input Voltage and Frequency| Power Input:
100-240 VAC, 50/60 Hz Power Cord Length: 6 ft. (1.8 m)
---|---
Operating Temperature| 50 to 95° F (10 to 35° C)
Environmental Requirements| Indoor use only at altitudes less than 6500 ft. (2
km)
Maximum relative humidity of 80% up to 85° F (30° C)
decreasing linearly to 50% @ 85° F (30° C)
Dimensions| 2.85″ L x 5.13″ W x 1.00″ H (72 mm x 130 mm x 25 mm)
Weight| 0.35 lbs. (0.16 kg)
Operator Test Range*| 500 kilohms to 10 megohms impedance
Worksurface Test Limit**| 10 megohms (±20%)
Operator Test Voltage| 1.2 V peak-to-peak @ 1.2 µA, open circuit
Worksurface Test Voltage| 5 to 7.5 V, open circuit
Alarm Response Time| <500 ms
Country of Origin| United States of America
*This cannot be verified with standard DC test equipment. The continuous monitor is an impedance sensing device, and the limits are determined by the magnitude and angle of the impedance.
**The worksurface mat must have a conductive layer such as Dual Layer Rubber
or Dissipative 3-Layer Vinyl with conductive buried layers. Desco workstation
monitors are not recommended for use with homogeneous matting.
ANSI/ESD S20.20 requires the use of a surface resistance meter for periodic
verification of a worksurface’s resistance-to-ground (Rtg). Constant monitors
may not be used as a substitute for this requirement.
Light Tower
Input Voltage | 12-24 VDC |
---|---|
Dimensions | 2.8″ L x 2.8″ L x 8.9″ H |
(70 mm x 70 mm x 225 mm)
Weight| 0.5 lbs. (0.23 kg)
Cable Length| 6 ft. (1.8 m)
Country of Origin| China
Limited Warranty, Warranty Exclusions, Limit of Liability and RMA Request
Instructions
See the Desco Warranty – Desco.com/Limited-Warranty.aspx
DESCO WEST – 3651 Walnut Avenue, Chino, CA 91710 •
909-627-8178
DESCO EAST – One Colgate Way, Canton, MA 02021-1407 •
781-821-8370 • Website:
Desco.com
TB-3092 Page 7 of 7
© 2022 DESCO INDUSTRIES INC
Employee Owned
References
- Desco - Static Control Products and Supplies for ESD Protection
- Desco - Limited Warranty
- Desco - 98221 Wave Distortion Monitor Verification Tester
- Desco - Static Control Products and Supplies for ESD Protection
- Desco - 09842 Multi-Ground Hub
- Desco - 19327 Power Cord, IEC C5, North American Plug, 6'
- Desco - 19328 Power Cord, IEC C5, UK Plug, 6'
- Desco - 19330 FULL-TIME MONITOR, WITH NORTH AMERICA POWER CORD
- Desco - 19331 FULL-TIME MONITOR, WITH UK POWER CORD
- Desco - 19332 LIGHT TOWER, FOR FULL-TIME MONITOR
- Desco - 19333 POWER RELAY, NORTH AMERICA, 120VAC
- Desco - 19334 REPLACEMENT INTERFACE CORD, FOR POWER RELAY