GYENNO TC00B Hand Tremor Data Collector User Manual

June 14, 2024
GYENNO

GYENNO TC00B Hand Tremor Data Collector User Manual

User Manual

Hand Tremor Data Collector

Model TC300

Version: TC300-EN-V1.0

DEFINITIONS

WARNING: Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.

CAUTION: Indicates a potentially hazardous situation which, if not avoided, may result in invalid data collecting, damage to the equipment or invalid operation.

NOTICE: Information necessary to be known before using the equipment.

INTENDED USE

The hand tremor data collector is used to collect hand tremor data of user with Parkinson’s Disease, Parkinson’s Syndrome, Essential Tremor, or other physical disabilities. It can also be used as an assistive equipment to eat.

SAFETY INFORMATION

WARNING

  1. Do not place or store the equipment near fire or heat articles.
  2. Do not use a damaged charging cable or socket.
  3. For patients with pacemakers or other electrical stimulation devices, please do not place the product near your heart or other electrical stimulation devices due to the magnetic components of the product.
  4. Users allergic to medical rubber are not commended to use this equipment.
  5. Do not plug or unplug the charging cable when hands are wet.
  6. Do not make self-diagnose based on data collected by the equipment.
  7. Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating normally.
  8. Use of accessories, transducers and cables other than those specified or provided by the manufacturer of this equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation.

CAUTION

  1. Do not immerse the equipment except for attachments into any kind of liquid.
  2. Do not use the equipment while charging.
  3. In case the equipment falls into liquid while charging, unplug the cable immediately and wipe the equipment.
  4. The attachments should not be disinfected with high temperature or microwaved.
  5. The handle should not be washed in dishwasher, disinfected with high temperature or microwaved.
  6. Do not use this equipment in a strong magnetic environment.
  7. Do not exceed the maximum load of the equipment.
  8. Do not dismount or attempt to repair the unit or components.
  9. Do not twist the spoon head, as this may cause damage to the device.

NOTICE

  1. Clean the attachments before using the equipment.

KNOW YOUR UNIT

The equipment can collect tremor data of user by built-in sensors and transfer the data via WIFI. It can also detect and offset shaking of hand to help user eat smoothly.

Components

Handle

Handle

Spoon Attachment

Spoon Attachment

Fork Attachment

Fork Attachment

SYMBOLS USED

Table 5 Symbols Used by this Equipment

SPECIFICATIONS

Table 6 Specifications

INDICATOR LIGHT DESCRIPTION

Table 7 Indicator Lights Description

NOTICE

  1. If the indicating light alternates in yellow solid and yellow flashing, the equipment is charging and not connected to WIFI.
  2. If the indicating light alternates in green solid and yellow flashing, the equipment is fully charged and not connected to WIFI.

DOWNLOAD MOBILE APPLICATION

Please search for “GYENNO SPOON” in the appropriate mobile application store, download and install it.

You can also visit www.gyenno.com/spoon2 for downloading the mobile application “GYENNO SPOON”.

Open the app on your smartphone and follow the WIFI configuration instructions.

NOTICE:

You can record and view your tremor data on the APP. If you do not download the APP, you can also use the device.

BEFORE USING

Attachments Installation

The attachments will adhere to the handle by magnetic force when the attachments are placed near the handle as shown below.

How to Hold the Equipment

Make sure the attachment is firmly connected to the handle before using. It is recommended to hold the equipment as shown below.

For users with severe tremor or stiff, please put on the hand strap before using to avoid dropping the equipment.

Power On and Off

Turn on the equipment by touching the on/off button on the front of the handle, then it can be used when the indicator light shows green breathing status.

Conversely, if the units is switched on, touch the on/off button on the front of the handle and the indicator light goes out and the unit is switched off.

NOTICE

The equipment switches off automatically after 5 minutes of inactivity when the indicator light goes out and the equipment enters a shutdown state.

TAKING A MEASUREMENT

When the equipment is on and once it detects tremor, data collecting will start. The indicating light breaths in blue.

The data collecting process takes a few minutes (measurement period), when it is completed, the indicating light breaths in green. Each measurement period the equipment exports a measurement result including frequency and amplitude.

NOTICE

  1. Reboot the equipment before taking the next measurement.
  2. Do not power off or put down the equipment before the measurement completes.

Data Transferring

All data stored in the equipment will be transferred while charging. To transfer data, please connect the equipment with power supply by the charging cable.

The equipment will automatically retry the saved latest wireless connection in a scanning window. If successfully connected, the indicating light flashes in blue for a while. If failed, the indicating light flashes in yellow.

If WIFI connects failed, please follow the instructions in “GYENNO SPOON” to retry.

Data transfer will start once a wireless connection is successfully set up. When transferring data, the indicating light flashes in blue and when the process is completed, the indicating light stays in red (charging) or green (charged).

View the Measurement

The measurement results, including the frequency (Hz) and amplitude (mm), can be viewed in “GYENNO SPOON”.

NOTICE

If no measurement results are shown, it may be caused by:

  1. None or incomplete data transfer due to WIFI connection problems.
  2. The measurement is terminated.

ASSIST TO EAT

Shaking Offset

Once the equipment detects tremor, the internal motors actively offset the shaking to keep stable.

AFTER USING

Attachments Uninstallation

Please power off and uninstall the attachments after using. The handle should be placed horizontally. It is suggested that both the handle and attachments should be cleaned each time after using.

Charging

If the indicating light breaths in yellow, the equipment can still work for approximately 10 minutes. In order to protect the mechanic parts, please charge the equipment as soon as possible.

To start charging, plug in the charging cable and connect to power. Keep the equipment horizontal while charging. The indicating light stays in yellow while charging and it turns to green when charged.

Figure 12 Charging

NOTICE

  1. Adapter required: 5V 1A。
  2. Please dispose of disused equipment and charging cables in accordance with local regulations, otherwise they may cause contamination.
  3. Please use the charging line provided by this device for charging.

CLEANING AND DISINFECTION

Attachments

NOTICE

The attachments should be cleaned before the first use.

Please uninstall the attachments from the handle before cleaning. The attachments are suggested to wash with utensil detergent and water, they can also be cleaned in the dishwasher. Wipe the attachments with dry cloth. Do not submerge the attachments into detergent contains corrosive ingredient.

If low-level disinfection is required, use alcohol solution (ethanol (75%), isopropyl (70%), e.g.), or disinfect it with ozone, ultraviolet rays, high temperature disinfection is not permitted.

Soft Sleeve

To clean the soft sleeve, use a cotton swab with warm water and mild detergent, and dry it with a clean cotton swab. Never clean it with toothpicks, tweezers, or anything sharp.

If low-level disinfection is required, use a cotton swab with alcohol solution (ethanol (75%), isopropyl (70%), e.g.), and dry it with a clean cotton swab.

Handle and Cover

The handle should never be submerged in water or cleaned in the dishwasher. Wipe the handle or cover clean with antibacterial wipes or a cloth with alcohol solution (ethanol (75%), isopropyl (70%), e.g.) and dry it with a clean,soft cloth.

The cover can be cleaned with a sponge, dishwashing detergent, and clean water, and can also be cleaned in the dishwasher. After cleaning, dry it with a clean, soft cloth. Do not use corrosive solutions to wipe or clean.

Hand Strap

Please put off the hand strap from the handle and wash it with a mild detergent, then rinse it off.

MAINTENANCE

  1. The attachment should be uninstalled from the handle before cleaning. Use a small portion of utensil detergent and water to wash.
  2. The attachment should not be submerged into detergent contains corrosive ingredient.
  3. Any part of the handle is not waterproof, do not splash water on the handle or submerge it into water. Wipe and clean the handle with antibacterial tissue or cotton cloth soaked with alcohol.
  4. Power off the equipment before installing or uninstalling the attachment or accessories, or after using.
  5. Do not place or store the equipment near fire or heat articles.
  6. This equipment consists of motors and sophisticated mechanical parts. Performance and service life of this equipment may be significantly reduced if under Long-term magnetic interference.
  7. Please avoid dropping or impacting the equipment at any time.
  8. Do not forcefully pull, press, or rotate the unit or components. Do not press other objects with the equipment.
  9. The equipment is only used for hand tremor data collecting and an assistive tool for eating. Do not use for other purposes.
  10. The service life of this equipment is 3 years (3 meals per day) or 50000 measurements (3 minutes per time).

TROUBLESHOOTING

Troubleshooting

PRECAUTIONS

  1. It is suggested to use this equipment independently at the first time to avoid worse tremor caused by the anxiety and nervous since the users may have difficulties on daily eating and being taken care of.
  2. It is suggested to proactively adjust holding gestures and ways of sending food to mouth at the beginning of adopting the equipment. The best way of using the equipment will come after some time.
  3. A big and shallow bowl helps when using the equipment as an assistive tool to eat.
  4. It may take several days for a user to use the equipment well. Please keep using for at least a week to get used to this new way of eating.
  5. Do not apply any kind of forces to arm and hand using the equipment. Relax as much as possible, keep the original shaking to make sure high performance of the equipment.

CONTRAINDICATIONS

For patients with pacemakers or other electrical stimulation devices, please do not place the product near your heart or other electrical stimulation devices due to the magnetic components of the product. Users allergic to medical rubber are not recommended to use this equipment.

EMC INFORMATION

EMC Information

EMC Information continued 1

EMC Information continued 2

EMC Information continued 3

EMC Information continued 4

REGULATORY INFORMATION

FCC ID: 2ACGF-TC00B

This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.

NOTE: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:

  • Reorient or relocate the receiving antenna.
  • Increase the separation between the equipment and receiver.
  • Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
  • Consult the dealer or an experienced radio/TV technician for help.

NOTE: The manufacturer is not responsible for any radio or TV interference caused by unauthorized modifications to this equipment. such modifications could void the user’s authority to operate this equipment.

Changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment.

This device complies with Industry Canada RSS standard(s). Operation is subject to the following two conditions: (1) this device may not cause interference, and (2) this device must accept any interference, including interference that may cause undesired operation of the device.

Specific Absorption Rate (SAR) information

This Hand Tremor Data Collector meets the government’s requirements for exposure to radio waves. The guidelines are based on standards that were developed by independent scientific organizations through periodic and thorough evaluation of scientific studies. The standards include a substantial safety margin designed to assure the safety of all persons regardless of age or health.

FCC RF Exposure Information and Statement

The SAR limit of USA (FCC) is 1.6 W/kg averaged over one gram of tissue. Device types: DTS (FCC ID: 2ACGF-TC00B) has also been tested against this SAR limit. The highest SAR value reported under this standard during product certification for use at the body is 0.672 W/kg. This device was tested for typical body-worn operations with the back of the product kept 0 cm from the body. To maintain compliance with FCC RF exposure requirements, use accessories that maintain a 0 cm separation distance between the user’s body and the back of the handset. The use of belt clips, holsters and similar accessories should not contain metallic components in its assembly. The use of accessories that do not satisfy these requirements may not comply with FCC RF exposure requirements, and should be avoided.

LIMITED WARRANTY

GYENNO warrants the included hardware product against defects for 12 months (24 months in EU countries) from the date of original retail purchase. As for the attachments and accessories (charging cable), the warrantyperiodis6 months.

Any part of the equipment, including attachments and accessories, is not allowed to be repaired by unspecified maintenance. One of the following occurs is not eligible for warranty claim:

  • Damage caused by dismounting on customer’s own.
  • Damage caused by unspecified maintenance by manufacturer.
  • Damage caused by improper transportation.
  • Damage caused by the failure to use in normal environment or according to the user manual.
  • Manufacturing label being replaced or removed.

DISPOSAL

GYENNO has always been committed to earth care, we also encourage our users to make contribution by executing proper disposal:

  • Packaging materials should be handed over recycling companies for possibility of re-use.
  • Comply with relevant laws and regulations when disposing wasted parts and components.
  • The equipment should be disposed as electronic waste.

COPYRIGHT AND RESPONSIBILITY

The copyrights and confidentiality of this manual are owned by GYENNO.

This manual is subjected to be a reference for operating and maintaining the equipment.

MANUFACTURER INFORMATION

Manufacturer Information

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

GYENNO User Manuals

Related Manuals