Arlyn Scales MKE-5 Digital Scales User Manual
- June 13, 2024
- Arlyn Scales
Table of Contents
- Arlyn Scales MKE-5 Digital Scales User Manual
- Introduction
- RS232 Communications Port
- Configuring the Port
- Perform Test Print
- Loop Back Test
- Print Stream Mode
- Printing at Stability Using Motion Detection and Stability Control
- The Print At Stability Setup Screen
- Function List
- Tares
- Samples
- Print Frames
- Text Definitions
- The Setpoint Definition Screen
- Creating and Editing Formulas
- CMD EXP VALUE OUT
- Notes on Cycling Setpoints
- Output Connector and Pin Diagram
- Weight Averaging
- Push Button Operation
- For Multiple Platforms
- For Multiple platforms, if you have selected the Multiple Platform
- weight averaging.
- Continuous Operation
- Configuring the Weight Average Option
- Multiple Platform Operation
- Battery Pack Operation
- Description
- The Main Power Toggle Switch
- The Front Panel On/Off Switch
- Auto Shutoff Setup
- Statistical Package [N/A: For Future Expansion]
- Statistical Menu
- The Setup Screen, Configuring the Statistical Package.
- Adding a Reading to the Collection
- Key lock option
- General description:
- To set password:
- To unlock keys:
- To deactivate key lock option:
- Data Logging
- Basic Operation
- Note about Multiple Platforms
- USB Flash Drive Use
- Advanced Operation
- WARNING: The following descriptions illustrate the advanced operation of
- Bluetooth
- NOTE: Some options labeled as [N/A] are either not available or not fully
- Important Notes for Bluetooth Operation
- Outgoing Connection
- Incoming Connection
Arlyn Scales MKE-5 Digital Scales User Manual
Introduction
This manual will cover the operation and setup of all of the options available for your scale. It is important to note that some described menu items will be missing if that option was not installed in your specific scale. All of the options listed may be installed in any scale and may run concurrently. Any of the options may be installed later at the factory if it wasn’t installed at time of purchase. Call our service department for details.
RS232 Communications Port
The RS232 option is a fully capable, bi-directional communications port. The
port can be configured to operate at a variety of baud rates and the output
data frames are definable by the user. The printing of a frame can be
initiated by pressing the print button, by an external command, upon reaching
a setpoint, or continuously when the print stream mode is activated.
The communications port also contains an extensive external command interface
allowing key presses, adding and editing memory slots and even scale
calibration to be controlled from external equipment.
Configuring the Port
Baud rate and other parameters can be configured through the Port Configuration screen that can be accessed under menu SETUP MENU/OPTION SETUPS/RS232/PORT CONFIG .The options are shown below.
BAUD RATE Settable to 110, 150, 300, 600, 1200, 2400, 4800, 9600, 19200,
31250 or 38400 bits per second.
DATA BITS Seven or Eight STOP BITS One or Two
PARITY Even, odd or none
ECHO On or off. When echo is enabled the scale will echo each
character the scale receives back to the user.
Perform Test Print
As an aid to port configuration we have included a “perform test print” function to the RS232 menu. Executing this will cause the string “Your RS232 port is working fine!”, a carriage return and a line feed to be output from the port. To access this function go to the menu SETUP MENU/OPTION SETUPS/RS232/PERFORM TEST PRINT.
Loop Back Test
You may also test the operation of the RS232 port by performing a loop back test. On the output connector, TX and RX (transmitted and received data) need to be shorted together. When you perform the test, the scale sends data and reads it back through its own port, where it is compared and the results are show.
Print Stream Mode
Print stream mode will continuously print the currently active print frame at
fixed time intervals. Stream mode can be configured using menu SETUP
MENU/OPTION SETUPS/RS232/PRINT STREAM. The lower menu selection ENB will
enable or disable print stream mode, and TIME will be the number of seconds
between the beginning of each print frame. Time settings of .5 to 1638 seconds
(27.3 minutes) can be entered. Avoid using time values of less than .5 seconds
as it may slow down or lock up the scale.
When stream mode is enabled, use the print button or the external *P command
to start or stop the printing process.
Printing at Stability Using Motion Detection and Stability Control
All Arlyn digital indicators have the capability of detecting motion and/or
stabilizing the weight and using it to control printed outputs. This can be
useful in many applications. For example, if you wanted to print labels for
several items, you could simply place each item on the platform, and when the
scale gets a stable reading it will print the label automatically. The Motion
Detect, Stability and printing at stability are fully configurable through
their respective setup screens.
The Motion Detect indicator looks like an ‘M’ and it is located just to the
right of the Zero Indicator / Bar Graph located on the bottom of the display.
It is crossed out if there is no motion, and it is uncrossed if there is
motion.
The Stability Control indicator looks like a circle and it is also located
just to the right of the Zero indicator/Bar Graph located on the bottom of the
display. The circle is unfilled when the weight is not stable and filled when
the weight is stable.
Print at Stability configuration allows the user to choose what control they
want to use to print. The user can choose Motion Detect only, Stability
control only, or both Motion Detect and Stability control. Take a look at the
Instructions Manual to see how the Motion Detect and Stability Control can be
set.
Sending a PRINT command or pressing the PRINT button on the front panel will
send a PRINT request to the scale, and the scale will only print if the motion
and stability conditions are met.
The Print At Stability Setup Screen
Follow the menu tree down this path to locate this setup screen:
SETUP MENU / OPTION SETUPS / RS232 / PRINT AT STABILITY.
There are five settings that control this function:
ACTV: This will activate or deactivate this function. Select ACTV from the
horizontal menu on the bottom of the screen, and then press ENTER to change
this setting.
PCT: This is the percent of full scale capacity, which, if the scale reading
falls below, will not automatically print. For example, a 100lb full capacity
scale set to 1% will not automatically print when there is less than 1 lb on
the platform. In most applications there is no purpose to generating a print
after removing weight from the platform. This percent can be set to two
decimal places.
TIMEOUT: This is the time set by the user to timeout retrieval of the value
after a certain period of time. By default, this value is 0, which means that
the scale is never going to timeout to print a weight values. It will continue
to look for a stable point to retrieve a weight value for printing. If the
number is set other than 0, then this is the period that the scale will try to
print a weight value, and if it fails, it will print “TIMEOUT” in ASCII on the
terminal.
NEG: By setting this to “YES”, the automated print will work on both sides of
zero. When set to “NO” it will only work with positive weights. The user must
take into consideration here that if this parameter is set to “NO”, then when
the scale is reading zero, the output may be erratic even though the scale is
stable. This is because the value 0 in the scale is not a true stable zero.
The zero here may be a long decimal that goes beyond the resolution. In this
case, the value maybe negative and therefore, impeding the printing process.
It is recommend to set this to “YES” if the zero value is important.
ZERO NTFY: If this option is set to “YES”, then each time the scale is zeroed
or a zero command is sent through one of the communication ports, the
indicator will wait for the scale to stabilize and detect if there is no
motion. Once these conditions are detected, the word “ZERO” appears on the
output terminal in ASCII.
CONTROL: This option can be switched between Motion, Stability, Motion and
Stability based on your preferences. This determines how you want the Print at
Stability to operate, whether based on Motion Detect, Stability Control or
both. It is important that when you choose these options, these features need
to be activated within the system to operate properly.
Print Stream Mode with Print At Stability
Print Stream and Print at Stability can work together. If you want the scale to print weights constantly, then activate both of these features together and they will work in tandem with each other.
Print Frame Definitions
A print frame is a list of up to ten functions for the scale to perform each
time a print request is received. The list is executed in the same order as
shown is shown. A function can either be an item to be printed or a command
for the scale. Currently there is only one command and it is the CHAIN
command. All other functions are items to be printed. If more than ten
functions are needed, as when defining print frames for complex labels, the
CHAIN command can be used to chain to another print frame definition
Each print frame requires two memory slots. You can define as many frames as
you like up to the number of free memory slots available.
Each frame has a description associated with it. Descriptions are useful when
you have many print frames defined, as when you have multiple labels etc.
Only one frame can be active at any one time. Whenever the scale receives a
print request, the currently activated frame will be printed. The active frame
is showed in the list with a check mark next to it.
Defining and Editing Print Frames
Go to the menu SETUP MENU/OPTION SETUPS/RS232/PRINT FRAMES.
To activate, edit or delete a frame, use the arrow keys to line up on the
desired frame and ACTV, EDIT or DEL on the lower menu and then press ENTER.
To define a new frame, use the arrow keys to select NEW on the lower menu and
press ENTER. A new frame entry will be added to the list. Use the arrow keys
to line up on it and on EDIT in the lower menu. Pressing ENTER will bring you
to the frame editing screen. The screen will show “None defined” as there are
no functions defined yet. For this example we will make a simple data frame
that will print the weight, followed by CR and an LF. Line up on NEW on the
lower menu and then press ENTER to add a function. The function NULL will be
added to the list. We want to change this to the WEIGHT function. Press the
right arrow key to line up on FNC+ on the lower menu. Press ENTER twice to
select the WEIGHT function. To add a CR to this function, use the right arrow
key to line up on CR on the lower menu and then press ENTER. You will then see
a check mark under the column “C”. To turn on the LF, press the right arrow
once more to line up on LF and then press ENTER. This completes the
definition.
If you wish to change the description of this new frame, use the arrow keys to line up on DESC on the lower menu and then press ENTER. Its current description will be loaded into the text-editing screen. Change it as desired and then press MENU to return to the frame editing screen. To use this new frame press MENU to back out to the frame list. Use the arrow keys to line up on the new frame and also ACTV on the lower menu. Press ENTER and a check mark will show on the new frame showing that it is activated.
The lower menu items on the frame editing screens are outlined below. DESC
Allows you to change the description of this frame.
NEW Adds a new default function (NULL) to the function list.
INS Will insert a new function (NULL) before the currently selected
function line. DEL Will delete the currently selected function line.
FNC+/- Will change the function of the current line. Available functions and
their explanations are outlined below.
SRC On many of the functions you need to specify the source of the data.
For example the WEIGHT function must know what platform (on multiple platform
scales) the reading should come from. SRC will step through all available
platforms. For a TEXDEF function, it will step through all the available text
definitions. For a CHAIN function it will show a list of other frames that you
can chain to. For ASCII you can enter the decimal value of the ASCII code. For
ChFont, using SRC will enable you to step through all available choices.
CR Will include a carriage return at the end of the current function.
LF Will include a line feed at the end of the current function.
SP Allows you to print space(s) at the end of the current function.
When activated a “?” will show up in the function list
on the current line and allow you to enter a number. The number of spaces must
be between 0 and 15.
TEST This will allow you to quickly test the print frame without having to
return to the scale’s normal operating mode.
Print Frame Functions
You can add up to ten functions to each frame. The functions are executed in
the order they are defined in the list. You can add a CR (carriage return), LF
(line feed) or trailing spaces to most functions. This is useful when
designing labels.
If a CHAIN function is encountered, the scale will ignore any further
functions in the current frame and immediately chain to the next frame.
Function List
NULL Does nothing, but allows CR, LF and spaces to be printed.
WEIGHT Prints a six digit weight value. It will be net or gross
depending on mode. Use SRC to select the platform. NETWGT Prints a six digit
weight reading corrected for tare. Use SRC to select the platform.
GRSWGT Prints a six digit weight value ignoring any active tare
weight. Use SRC to select the platform.
NGTEXT Prints “NET” or “GROSS” depending on mode. This is meant to
be used with the weight function. Use SRCto select the platform.
PLATTOT Prints a gross total reading of all active platforms.
COUNT Prints the current piece count. Use SRC to select the
platform.
ACCUM Prints the value in the parts counting accumulate register.
Use SRC to select the platform.
CNTACC Prints a total of the current piece count plus the part count
in the accumulate register. Use SRC to select the platform.
TIME12 Prints the current time in 12 hour format.. (when the time
and date option is installed) TIME24 Prints the current time in 24
hour format.. (when the time and date option is installed) DAY
Prints out the current day of the week. (when the time and date option is
installed)
DATE Prints the current date in the format mm/dd/yyyy. (when the
time and date option is installed) SCADES Prints the 14 character scale
description as entered in the system menu.
SCAID# Prints the scale ID number as entered in the system menu.
WEIGH# Prints the auto-incrementing weigh number who’s starting
value was entered in the system menu. Thisnumber will increment each time a
frame is printed.
TEXDEF Prints a 14 character text string that was entered as a text
definition. Use SRC to select the text definition. UNITXT Prints a 3
character indication of the currently selected units conversion (lb, kg, oz
etc).
TARDES Prints the 14 character description of the current active
tare weight.
SMPDES Prints the 14 character description of the current active
sample weight (parts counting).
ASCII Will send a single ASCII character to the port. Among
other things this is useful to send ESC sequences to the printer to change
fonts etc. Use SRC to enter the ASCII code.
CHFONT Change Font. Works only with EPSON compatible printers. Use
SRC to select the font.. See explanation below.
CHPNT Change Point size. Use SRC to toggle through the choices
16,20,24,30,36,40,42,48,52,56,60,64 CHAIN Will end the current frame and
initiate another. Use SRC to select the next frame. Avoid chaining to the same
frame, as this will cause endless looping.
Print Frame Design Examples
The first example is a simple print frame that will print the weight followed by a carriage return and a line feed.
FUNCTN | SOURCE | C L SP |
---|---|---|
Weight | Plat#001 | 44 0 |
The next example shows a more complex definition. Here we will design a five line label. Line 1and line 2 will be blank (used for spacing the label vertically).
Line 3 will show the time in 12hr format, followed by 1 space, Line 4 shows
day of the week with 1 trailing space.
Line 5 shows the date followed by a carriage return and a line feed.
Line 6 will show the scale’s text description (as programmed through the
system menu) followed by 3 spaces.
Line 7 shows the weight followed by a carriage return and a line feed. Line #8
will be blank (again used for spacing the label vertically).
FUNCTN
Null
| SOURCE| C L SP
44 0
| Comments
Two blank lines. Note the check marks on C and L.
---|---|---|---
Null Time12| | 44| 0|
1
|
Line #3, the time and 1 space
Day
Date
| |
44
|
0
| 1| Day of week and one space
Date followed by a CR and LF
SCADES
Weight
|
Plat#001
|
44
|
0
| 3| Line #4, The scales description followed by 3 spaces
The weight followed by a CR and LF
Null| | 44| 0| | Two blank lines. Note the check marks on C and L.
The label would look like:
Remember that there are only 10 functions allowed per print frame, and we used
nearly all of them in this example. If you needed more you could define a
second print frame and use a CHAIN function at the end of first frame to chain
to the second frame.
Text Definitions
Sometimes it is desirable to include text when printing a label. Each text
definition allows you to define a 14 character text string. To access text
definitions, go to the menu SETUP MENU/OPTION
SETUPS/RS232/TEXT DEFINITIONS. A list of current text definitions is shown and
you can NEW, EDT or DEL them in the usual way. ACTV is not used here.
Changing Printer Fonts
If you have an EPSON compatible printer, you can use the ChFont print frame
function to change to one of four fonts, turn double wide on/off, double
height on/off, italics on/of, letter quality/draft, bold on/off and printer
reset. Not all point sizes or other functions are available on all fonts.
Also, not all fonts are available on all printers. Please consult your
printer’s instruction manual for details. The ChPnt print frame function
allows you to scale the font size up and down.
The scale will send the appropriate “Escape Codes” to the printer to perform
these functions. ChFont and ChPnt are only meant to be a short-cuts to reduce
the frame size. You could also perform font/point switching and many other
printer functions by using the ASCII function.
The following is a summary of the escape codes sent to the printer by the
various sub-functions of ChFont;
Bold On | ESC E | <27 69> | Bold printing ON |
---|---|---|---|
Bold Off | ESC F | <27 70> | Bold printing OFF |
Ital On | ESC 4 | <27 52> | Italics ON |
Ital Off | ESC 5 | <27 53> | Italics OFF |
Draft | ESC x 1 | <27 120 1> | Select draft quality print |
Let Qual | ESC x 0 | <27 120 0> | Select letter quality print |
Roman | ESC k 0 | <27 107 0> | Select Roman font |
Sans Ser | ESC k 1 | <27 107 1> | Select Sans Serif font |
Courier | ESC k 2 | <27 107 2> | Select Courier font |
Prestige | ESC k 3 | <27 107 3> | Select Prestige font |
Script | ESC k 4 | <27 107 4> | Select Script font |
Reset | ESC @ | <27 64> | Initialize printer |
DblW On | ESC W 1 | <27 87 1> | Double width printing ON |
DblW Off | ESC W 0 | <27 87 0> | Double width printing OFF |
DblH On | ESC w 1 | <27 119 1> | Double height printing ON |
DblH Off | ESC w 0 | <27 119 0> | Double height printing OFF |
ChPnt | ESC X 1 n 0 | <27 88 1 n 0> | Select height in points. n = # of points |
External Command Interface
The external command interface allows the user to control the scale and
manipulate memory data through the RS232 interface. There are three basic
command types, these are:
The * Command (Keyboard Emulation). This allows you to simulate key presses
remotely.
The ! Command (Remote Entry). Allows you to add, edit and delete memory
contents like tares and parts counting samples. The @ Command (Interactive
Command). Allows you to operate the scale interactively whereby you send
commands to the scale and the scale responds with text messages guiding you
through various operations.
It should be noted that the command list described below is the proposed list of commands. As our scale continues to develop, commands are added as each firmware revision comes out. The revision in which the command was added is shown just to the right of the command in parenthesis. If (X) is indicated it means that the command has not been implemented yet and is slated for future release. To check the revision number of your scale, watch the Arlyn Logo splash- screen when you first turn on your scale. Firmware upgrades can be purchased and installed in your indicator. . Some commands can be implemented based on customer request. If it is imperative for the customer to have a certain command type, contact our service department for more information and to get quote for the functionality to be implemented.
Print Request Response Time
When the scale receives a print request (*P) there is a small time delay while the scale computes and formats its output frame. For any frame function that outputs weight or parts count the response time is typically 100 milliseconds (1/10th second). For almost all other frame functions the response is almost immediate.
External Command Limitations
It is important that when sending repeated print requests (*P) to the scale that you wait for the scale to completely transmit its print frame, and then wait a small amount of time before sending the next request. If the scale receives a print request while it is still transmitting, the scale could actually lock up requiring a power off reset. Also, it is advisable to wait a small amount of time after the output frame is complete to give the scale a chance to do its own “housekeeping chores”. To immediately send another print request directly after a transmission could bog down the scale’s displayed update rate and keyboard response.
Command String Parameters
All command strings sent to the scale require a carriage return (hereafter
called CR) to be sent at the end of the command string. A CR is ASCII code 13.
Keyboard emulation commands (* commands) do not require a CR and the key press
will be initiated right after the command is received.
Many of the command strings require parameters. These are described below.
< descr> Any command that changes the contents of a memory slot will
require a fourteen character description. This description should exactly
match the description of the item in question. The description must be 14
characters in length and spaces should be used for any unused characters just
as they are used in the scale’s setup menus. For example “Tare #001 “ and
“Plat #001 “ are valid descriptions. Descriptions are case sensitive so
format your output appropriately.
If the scale can not find the description in memory, the message “Error 2:
Description not found. Aborting.” Will be sent
followed by a CR. The operation is then aborted.
< value> Some commands require numeric input. All numbers should be 9
characters in length with a decimal point at character five. Valid values, for
example, are 0123.4500 and 0001.0000. Use leading zeros as necessary to keep
the spacing correct.
If the scale receives a number in an incorrect format it will send the message
“Error 3: Illegal numeric format. Aborting.” Followed by a CR. The operation
is then aborted.
Command Abort Any command can be aborted at any time by sending a caret
(^) character which is ASCII code 94. The
scale will respond by sending the text “Command aborted” followed by a CR
(ASCII code 13).
Illegal Command If any text is received by the scale that isn’t in a
correct command format , the scale will respond with
“Error 1: Illegal command.” followed by a CR. The command is then aborted.
*The [] Command: Used for Keyboard Emulation**
A number of external commands can be sent to the scale to simulate keyboard presses. These commands are listed below. Each command must be prefixed with a
- character and no CR is required.
P or 1 – Print Key (5.000)
C or 3 – Units Key (5.000)
Z or 0 – Zero Key (5.000)
N or 9 – Net/Gross Key (5.000)
T or 7 – Tare Key (5.000)
U or 8 – Up Arrow Key (5.008)
D or 2 – Down Arrow Key (5.008)
L or 4 – Left Arrow Key (5.008)
R or 6 – Right Arrow Key (5.008)
. – Decimal Point (5.008)
O – On/Off Key, will take the scale in and out of Ready Mode (5.008)
M or 5 – Menu/Back Key (5.008)
*E – Enter Key (5.008)
Care must be taken when sending key commands via RS232 as some keys will
perform different functions depending on thescale’s operating mode.
The [!] Command: Remote Entry Commands [N/A: For Future Expansion]
Tares
NEW (X) !TN
EDIT (X) !TE
DELETE (X) !TD
ACTIVATE (X) !TA
CLEAR (X) !TC simply clears the active tare.
ACQUIRE (X) !TQ
Samples
NEW (X) !SN
EDIT (X) !SE
DELETE (X) !SD
ACTIVATE (X) !SA
CLEAR (X) !SC
ACQUIRE (X) !SQ
Print Frames
NEW (5.021) !FN
LINE EDIT (5.021) !FL<14 chr src descr> <1 chr line n# 0-9>
NEW (5.021) !XN
The [@] Command: Interactive Commands [N/A: For Future Expansion]
**Remote Span Calibration** (5.010)
Your scale can be span calibrated remotely using this function. The command
format is @CS
To do this procedure you will need a calibrated weight of a known value. For
this example we will calibrate the scale’s
primary platform whose description is “Plat #001” with a 10 lb weight, the
procedure is described below.
You send “@CSPlat #001 0010.0000
Scale responds “Remove all weight from the platform and then send
a CR.”
You remove the weight from the platform and send a CR.
Scale responds “Place the calibrated weight on the platform and
then send a CR.”
You place the calibrated weight in the center of the platform and then send a
CR.
Scale responds “Remove all weight from the platform and then send
a CR.”
You remove the weight from the platform and then send a CR. Scale responds
“Operation Complete.”
The scale will then remote calibrate itself and return to the weighing mode. You may have to press the ZERO key (or remote zero key) to get the scale to indicate properly.
Future Interactive Commands
TARE-READ VALUE (X) @TV
Tips on Interfacing RS232
RS232 transmission and reception takes place over a three wire cable, one for transmitted data, one for received data and one for ground. When wiring these lines to a standard RS232 25 pin D connector, the transmitted data line should be tied to pin 3, received data to pin 2 and ground to pin 7. For 9 pin connectors, the transmitted data line should be tied to pin 2, received data to pin 3 and ground to pin 5. In some situations, computers and other equipment require handshaking lines to work properly. Many times this can be shut off to work with a three wire system. If not, you can wire the connector to “trick” the computer etc. to work properly. This involves shorting the RS232 RTS and CTS line together and shorting the DTR and DSR lines together. For 25 pin connectors short pins 4 and 5 for RTS/CTS and pins 6 and 20 for DTR/DSR. For 9 pin connectors short pins 4 and 6 for RTS/CTS and pins 7 and 8 for DTR/DSR.
Shown below is a simple program written in BASIC that will open a com port at
1200 baud and then drop into a never ending loop that will input frames of
information and print them on the computer’s screen.
10 REM A simple example of a scale to com port interface
20 OPEN ”COM1:1200,N,8,1,CS,DS” AS #1
50 REM Never ending loop
52 REM Send a print request to the scale
55 PRINT #1,”*P”
- REM Input scale’s transmitted frame
- INPUT #1,FRAME$
69 REM display the results on the computer screen 70 PRINT FRAME$
75 REM Small time delay (at least 100ms, see External Command Limitations
above) 80 FOR X=1 TO 20000
85 NEXT X90 GOTO 50
Time and Date
When your scale is equipped with the time and date option you can get time and date and day of week readouts on the screen or use them in RS232 output frames. The time can be printed in either 12 or 24 hour format. Consult the section on Print Frame Functions for details. The screens for setting the date and time can be found under the menu SETUP MENU/OPTION SETUPS.
Analog Output
This option is a fully configurable analog output port that can be configured for current loop outputs. The factory has preset the output to 4-20 mA. This range is configurable and it is explained in the second paragraph of this section. The output follows the display and its gain is based on the platform’s programmed overload point. For example, in the case of a 100lb scale whose output is configured as 4-20mA, 0lb = 4mA, 25lb = 8mA, 50lb = 12mA and so on.
The starting and ending currents can be configured by the user. For example the output’s starting and ending points can be set 4mA to 20mA. Further, there is an option allowing negative displayed numbers to be output for cases where a reading below zero needs to be monitored. As an example, a 100lb scale is setup with its starting current at 10mA and ending current at 20mA, giving the output of -100lb= 0mA, -50lb = 5mA, 0lb = 10mA, 50lb = 15mA and 100lb = 20mA.
The Analog Output can also be configured to output voltage. The output can be configured for a voltage range between 0- 5VDC. This configuration can be requested to be done at the time of processing the order. However, this can also be done by the customer. Simply connect a 250 ohm ¼ watt resistor between the output terminals. Set the starting and ending currents to be 0mA and 20mA. Using Ohm’s law (V = I x R), you will get from 0VDC (0mA x 250 ohm = 0V) to 5VDC (20mA x 250 ohm = 5V). Just put a voltmeter across the terminals to get these values.
Using any value of the resistor, the scale can be configured to output a customized range of voltages based on customer requirements. However, the voltage cannot exceed 6VDC. The current output component cannot support a voltage drop larger than this value.
Analog Setup Menu
The analog setup menu can be found under menu SETUP MENU/OPTION SETUPS/ANALOG OUTPUT. Programmable parameters are as follows:
ACTV – Will activate or deactivate the option.
START – The starting current that will correspond to zero on the display.
END – The ending current that will correspond to the full-scale weight on the
display.
NEGRDG – Allows negative displayed readings to affect the analog output. If
disabled, the output will stop at the programmed starting current when there
is a negative number on the display. This should not be confused with a
negative current on the output. The lowest possible current is zero.
Special Note for Multiple Platform Scales
The analog output module used in the scale is a single channel module. Only a
single output from a single platform can be seen. For scales with multiple
platform, the analog output will reflect accordingly with the multiple
platform settings you have. For multiplatform scales, you can monitor the
weight from each platform on the indicator screen by using the UP/DOWN arrow
keys and SHIFT key. The analog output will only reflect the current platform
seen on the indicator screen. For example, if the screen is showing weight for
Platform 1, the analog output reading will be for Platform 1.
Indicators equipped with multiple platforms also have a special screen that
shows the weights on all the connected platforms and the total from all of
these platforms. In this screen, there is an arrow that indicates the actions
that can be taken on the indicated platform. If the arrow is on Platform 1,
then the analog output will reflect Platform 1. The same applies for platform
- While in this screen, the analog output can also reflect the total sum of weights on all the platforms.
To do this, press the SHIFT key and SIX key in sequence (not together). When this feature is activated, the analog output will only reflect the total. It won’t reflect on any of the platforms. Press the SHIFT + 6 key sequence again to remove the analog output for total weight feature.
Setpoint Controller
The setpoint controller gives your scale the capability to output a signal to
external equipment when certain conditions are met. This is particularly
useful in filling operations, either to sound alarms or to control filling
machinery.
Setpoint definitions can be created, edited and deleted. Each has a text
description, controlling parameters and a multi-line formula associated with
it. In operation, if the displayed weight satisfies the formula, you may
direct any of the outputs to switch on or off. You can create as many setpoint
definitions as you like within limits of free memory. Each setpoint uses four
memory slots.
Setpoint definitions can be activated or deactivated, and up to four setpoint
definitions can be activated at one time. When multiple setpoints are
activated, each setpoint formula is evaluated in the order it was defined. If
more than one setpoint shares the same output, then it is possible for one to
overwrite the other.
With a little finesse in creating formulas, simple setpoints, filling control
cycles, over and under check weighing, and many other functions are possible.
IMPORTANT NOTE: During Scale reboot, the scale will momentarily toggle all setpoints to check for malfunction. This might cause your valve systems to open (or close) during that initial moment. A fix is already being implemented (as of this writing) for this issue. All Software versions at 6.123b and below will have this feature. A workaround would be to add a manual switch that would be used to shut off your system if the scale is every rebooted.
Output Specifications and Notes
The setpoint controller has eight separate outputs and is expandable to sixteen. Each output is an open collector that is capable of sourcing 4.5ma or sink 10ma of current. Care must be taken not to exceed these values. This is sufficient to drive LEDs or sound buzzers. If higher output currents are needed then external relays need to be installed. We normally provide 117VAC, 10Amp optically isolated relays and mechanical relays should be avoided. We also provide 60VDC DC relays.
Optical relays have some important advantages over mechanical relays, the actuation current for optical relays is usually less than 10ma but mechanical relay coils require currents often exceeding the limit. Optical relays have no mechanical switches to wear out and more importantly they isolate the scale from the load, thus protecting it from voltage spikes. If excessive noise is coupled back to the scale it could lock up or damage its microprocessor or memory.
It should be noted that when you command an “OFF” CONDITION, the output will
be grounded. In most open collector (switched ground) wiring schemes this will
actuate the external light or relay. To help eliminate this confusion, each
setpoint formula has a optional invert function to invert the output.
Even though the output is an open collector, it differs from the standard open
collector output in that the output will assume a high condition (3.3VDC) when
it is switched on without the need for an external pull up resistor, allowing
direct connection into any TTL level input.
Creating Setpoints and Formulas
To access setpoints, go to the menu SETUPMENU/OPTION SETUPS/SETPOINT
CONTROLLER. There you will see a list of all available setpoints or “None
Defined” if none exist. The creation, editing and deletion of setpoints are
similar to the creation of tares and parts counting samples. To add a new
setpoint definition, select NEW on the lower menu and press ENTER. A new
setpoint definition will then be added to the list with a description of
“SetPt #XXX”. This new setpoint is deactivated and contains an empty formula
list.
To edit this new setpoint, or any other one, use the UP and DOWN ARROW keys to
line up on the desired setpoint, select EDIT on the lower menu and then press
ENTER. The setpoint definition screen will be shown.
The Setpoint Definition Screen
The setpoint definition screen will show the description and controlling parameters for this setpoint definition. INV=On/Off: If it is on then all outputs related to this setpoint will be inverted. This is useful when using the open collector output for the reasons stated earlier in this section.
CYCLE=On/Off: The cycle option is used to create an operation cycle as
described earlier in this section.
NEG=On/Off: The Negative option is used to determine if the scale needs
to evaluate negative weight values. If NEG = Off, then the setpoint controller
will not evaluate negative weights, as in it will treat both negative and
positive weights the same. Example, if the display shows -10 lbs or +10 lbs,
the setpoint evaluator will evaluate these values as positive values. If NEG
= On, then it will differentiate between the two values.
PLAT=PL1: This option only appears when the scale is equipped with
Multiple platforms. The user can select if the setpoint conforms to a
particular platform. If it is the case, then the setpoint will only evaluate
weights corresponding to that platform. PARTS=On/Off: This option only
appears when the scale is equipped with Parts Counting. The user is given a
choice of whether the setpoint should respond to count value instead of weight
value.
To change any of these options simply select the appropriate entry in the
lower menu and press ENTER.
Creating and Editing Formulas
The formula is a list of up to nine lines that are evaluated and acted upon in
the order they are entered.
To access the formula list select EDIT in the lower menu and then press ENTER.
All of the current formula lines will bedisplayed or “None Defined” if there
aren’t any. Using the NEW and INS in the lower menu to add or insert new
lines.
Each line consists of four parts. These are CMD (command), EXP (expression),
VALUE and OUT. CMD is an action to take. The only two commands currently
available are ON and OFF. EXP is an expression and can be < (less than), <=
(less than or equal to), > (greater than) and >= (greater than or equal to).
Value is the setpoint weight value and OUT is the output channel (1-8) that
the CMD command will work on.
For example, a simple formula to activate the output when a certain weight is exceeded is made up of two lines as shown below:
CMD| EXP| VALUE| OUT|
---|---|---|---|---
Off| <| 5.0| 1| Start with light off
On| >=| 5.0| 1| On if 5 lb or greater
This can be interpreted as:
Line 1: Output #1 is off if weight is less than 5.00 lb
Line 2: Output #1 is on if weight is greater than or equal to than 5.00 lb
You will notice that every output possibility must be dealt with when creating
a formula.
The next example represents a much more complicated definition. The scale
would have three lights connected to three outputs and operate in an UNDER /
ACCEPT / OVER check weighing scenarioThere will be a red light connected to
output #1 to indicate an “under” condition, a green light connected to output
2 for an
accept” condition and a yellow light connected to output #3 to show an “over”
condition. We can program the red light to be on whenever the weight is less
than 5lb, the green light to be on only between 5 and 5.2 lb and the yellow
light to be on whenever the weight is above 5.2lb indicating an over
condition. In practice, the user would start filling the container until the
green light comes on.
It is less confusing to make a formula of this nature if you deal with the
complete operation of one light at a time. We recommend laying the formula out
on paper first and then enter it into the system.
Here it is:
CMD| EXP| VALUE| OUT|
---|---|---|---|---
On| >| 0.00| 1| Deal with “under” (red) light first
Off| >=| 5.00| 1|
Off| <| 5.00| 2| Now the “accept” (green light), to be
On| >=| 5.00| 2| on only between 5 and 5.2 lb
Off| >| 5.20| 2|
Off| <=| 5.20| 3| Now the “over” (yellow) light to be on
On| >| 5.20| 3| anytime the weight is above 5.2
When this formula is executed, the scale evaluates each line in order. This
calculation is done in memory only, and the outputs are not affected until the
formula is complete. The entire formula is evaluated on each and every display
update, which is programmable through the system menu.
You may wonder why it is necessary to have to use an Off condition at the
beginning. This was done to increase flexibility as there may be many cases
where the user wants the light to start in the on state, turn off during a
certain weight range and on again when above that range. You have to specify
every possible condition for the output when making formulas.This last example
will show an automated filling operation using the cycle option. The goal here
would be to have output #1 connected to a filling valve and the cycle starting
with this valve in the off condition. The user places a box on the scale below
the filling apparatus, press the ZERO key to zero out the weight of the box,
press the ENTER key to start filling and have the setpoint controller switch
the filling valve off when 5lb is reached. The user then removes the box from
the scale and the output will remain off until the next cycle is initiated by
pressing the ENTER key.
Start in the setpoint definition screen by enabling the CYCLE option and
setting the INIT option to off. Here is the formula:
CMD EXP VALUE OUT
On < 5.00 1 Will actually go on as
soon as the ENTER key is pressed. Off >= 5.00 1
Switch off at 5 lb and end cycle
That’s all there is to it. The output will go off at 5 lb and remain off until
the next cycle.
Notes on Cycling Setpoints
The cycle function is used when it is desirable to “latch” the output to a
specific state when a process is complete. Consider the following examples
with and without the cycle option. In both cases the setpoint is set to fill a
box to 5lb and then shut off.
Cycle Off: The output is on as the box fills and the scale monitors the
process. At 5lb the output switches off. The user then removes the box from
the scale. When he does that, the indicated weight falls below 5 lb, the
filling valve switches back on dumping material all over the platform.
Output Connector and Pin Diagram
In most cases there will be a single cable with a 15 pin subminiature D type female connector added to the scale for interfacing to external equipment.
- Setpoint #4 Output
- Setpoint #5 Output
- Setpoint #6 Output
- Setpoint #7 Output
- Setpoint #8 Output
- Setpoint #3 Output
- Setpoint #2 Output
- Setpoint #1 Output
- Digital Ground (For Scales sold before 3/7) 10 +3.3 VDC
15 Digital Ground (For Scales sold after 3/7)
Weight Averaging
The weight averaging option allows the scale to display an average weight taken from the platform over a specific time period. Readings are sampled from the platform forty times per second so a five second average would average 200 readings. Weight averaging can be setup to operate from a front panel key (the “8” key) or to operate on a continuous basis.
Push Button Operation
Pressing the “8” key will start the average. While the average is being taken, the status bar on the bottom of the screen will display “Avg” and the graph bar will show the amount of time left. When the average is completed the result is locked on the display and the text “Loc” will be shown on the status bar. While the reading is locked the UNITS, NET/GROSS and other front panel keys will still operate as normal. To release the lock simply press the “8” key again. While the average is being taken it can be aborted at any time by pressing the “8” key.
For Multiple Platforms
For Multiple platforms, if you have selected the Multiple Platform
screen, then press the “6” button to perform
weight averaging.
Continuous Operation
When configured for continuous operation, constant averages at the selected time interval will be taken and locked on the display. The text “Avg” will be shown on the status bar and the time progress shown on the graph line. The average key (the “8” key) is used to start and stop this process.
Configuring the Weight Average Option
The configuration parameters for the weight average option can be found in the
menu SETUPMENU/OPTION SETUPS/WEIGHT AVERAGING. There are three settable
options here.
ACTV Activates or deactivates weight averaging TYPE Push button or continuous
operation
TIME Average acquire time can be set from one to sixty three seconds
Multiple Platform Operation
Your scale has the capability of running up to three input devices at any one time. Normally only one is used, a load cell, and the scale is programmed to have a single platform containing that load cell. The scale can be configured to run from multiple platforms there are only a few basic operational differences:
- An extra display screen is added showing the readings from each individual platform as well as the total weight from all
- When viewing individual platform screens, you can manually scroll through different Press SHIFT+UP in sequence or SHIFT+DOWN in sequence to scroll through the platforms. On the primary weight screens the active platform number is indicated on the bottom of the screen just to the right of the status bar (PL1, PL2 etc).
- On the extra platform you can just use UP key or DOWN key without the need to press SHIFT key. For example, to zero platform 2 use the DOWN ARROW to get the indicator on the right side of the screen to point at platform 2 and then press the ZERO key.
- An extra menu is inserted in the menu SETUP MENU/PLATFORM This is a menu that shows the description of each platform and allows you select one for editing.
- When taking or activating tares, parts counting samples, taking a quick sample or pressing the TARE, UNITS, NET/GROSS or ZERO key, the operation will be performed only on the currently selected
- An auto-scroll feature is available in the Platform Setup This feature allows the indicator to scroll through individual weight screens at specified time intervals. To setup auto-scroll, go to SETUP MENU->PLATFORM SETUP. Choose any activated platform and press EDIT. Scroll down to MULTIPLATFORM SETTINGS. Press ENTER. The horizontal options and the screen text should be clear enough to set the correct settings. It is recommended that the scroll timer should be set no less than 5 seconds. Any number lesser than that can cause some instability in scale readings.
- FOR DUAL PLATFORMS ONLY: To show the difference between the two platform scales, do the same steps as defined in step 6 above. In the MULTIPLATFORM SETTINGS, select Platform Total to be either “Sum” or “Difference”.
- If the multiple platform scale has options attached to it, such as RS232, Analog Output, Setpoints, make sure to read the special notes for that section, if any, that might apply specifically for multiple platform operation.
Battery Pack Operation
See the addendum page 24 for quick operational instructions on this feature.
Description
The Battery Pack Option will allow the scale to be operated in areas where no
power is available. The battery pack itself is installed inside the indicator
housing and is not user serviceable in any way. The only external differences
are the addition of a Main Power toggle switch and power transformer connector
located on the right side of the indicator housing.
To charge the pack, simply leave the scale plugged in for about four hours.
The charging circuitry has a peak detect and will stop charging once the
battery has reached its peak. The scale may be operating normally while
charging.
The battery pack allows a scale equipped with a single load cell platform
(models 620 and 820) to operate for about fifteen hours on a single charge.
For scales that have multiple load cell platforms, running time is slightly
shorter. A four load cell platform, like the model 320 will run for about 8
hours.
The Main Power Toggle Switch
There is a main power switch located on the right side of the indicator. Its only function is to completely remove power from the main circuit board in the event of a microprocessor program crash or any other event that causes the scale not to be able to switch itself off. This toggle switch is also responsible for charging the battery. If the toggle switch is left in the open (DOWN) position, then the battery will not charge. This toggle switch should be left in the closed (UP) position during all normal operation.
The Front Panel On/Off Switch
The front panel On/Off switch is used to switch the scale on and off during
normal operation.
To turn the scale on, simply press the button and the scale will start its
initiation sequence. To turn the scale off, press and hold this button for
approximately three seconds until the display blanks out and then release it.
Auto Shutoff Setup
The scale can (and should) be setup to automatically turn itself off when a
programmed time has been reached and/or there is no activity on the scale.
This is done through the Auto Shut-Off screen. You may also want to skip the
splash screen and ready prompt on startup. You can do this in the menu SETUP
MENU/SYSTEM/STARTUP PARAMETERS.
The scale’s auto shutoff operation can be activated and changed in the menu
SETUP MENU/ SYSTEM /AUTO SHUT-OFF. There are two options available.
ACT Activates/deactivates the auto shutoff feature.
TIME Is the time, in minutes, that the unit will shutoff due to
inactivity.
Statistical Package [N/A: For Future Expansion]
The Statistical Package allows you to save a large number of readings in memory for future use. Here are some of its features:
- You can transmit the entire data collection through the RS232 communications port to a printer or computer for later
- You can view a “result screen” which will show the minimum reading, the maximum reading, the average, and
- The data collection is held in memory even if the power is
- Readings can be saved by simply pressing a
- Shares memory with the scale’s operating system for maximum storage
- Main screen shows when data collection is active and the number of readings
Statistical Menu
This menu contains all of the functions associated with the statistical
package. It is located at SETUP MENU / OPTION SETUPS / STATISTICAL.
The items in this menu are:
SETUP: This is a simple configuration screen for the statistical package.
CLEAR MEMORY: This will clear all collected data from memory.
TRANSMIT DATA: This will transmit the entire data collection through the
RS232 communications port. Each collected reading is transmitted along with a
carriage return (ASCII 13) and a line feed (ASCII 10). All collected readings
are transmitted. Memory does not automatically clear after transmission.
VIEW RESULTS: This screen will show some basic information about the data
collected. These are the minimum reading, the maximum reading, the total and
average reading.
The Setup Screen, Configuring the Statistical Package.
This screen will allow you to specify if the statistical option is active or
not, and also how much memory to set aside for data collection. Any time you
enter this configuration screen, the data collection in memory will be erased.
There are only two configurations on this screen:
ACTIVE: Set to YES in order to enable data collection. If enabled, the
top line of the main weight display will show an inverse
“Data:”, followed by the number of readings currently collected in memory.
SIZE: The memory set aside for data collection uses the same memory and
storage techniques as the memory used by the scale’s operating system for
storing tares, counting samples, print frames and other data. If you aren’t
using the scale’s memory for those other uses, you will have roughly 470 free
slots available for data collection, depending on the scale’s configuration.
Each slot will hold 8 readings, so at maximum, 470 x 8, or 3760 readings may
be collected.
To configure the memory, enter the new size through the setup screen. Size is
the number of readings (maximum) that you would like to be able to store. This
number will automatically be rounded up to the next multiplier of 8 (8
readings per slot). The memory is then set aside for data collection, and
becomes invisible to the rest of the system. This being the case, if you
configure a large memory area for data collection, you will be restricted as
to the number of storage slots available for other uses.
Adding a Reading to the Collection
Simply press the ENTER key at any time while you are in the main weight screen. The current reading will be saved and the top line will show you how many readings have been collected. If the data collection memory is full, you will see a warning message at this time.
Key lock option
General description:
**** This key lock option adds the capability of disabling the front panel by having all of the function keys locked out, until a valid password is entered. Once a valid password is entered, the keys will remain unlocked until the scale is powered down and powered up. By default the option is not active until an initial password is set.
To set password:
- Press the menu Unlock keys by entering the current password, if necessary.
- Navigate to Options Setup and then scroll down to “Set Password”.
- “Enter new password “ will appear on the Enter the desired password (4 digits long)
- Confirm the password by retyping the intended Confirmation is used to assure that a mistyped password is not stored.
To unlock keys:
- Press menu
- “Enter Password” will appear on the top line of the Enter the password previously set.
- The function keys will remain unlocked until the scale’s power is powered down and powered
To deactivate key lock option:
- Follow the above procedure for setting the
- Instead of entering a password simply press enter
- This will clear the password and deactivate the key lock option until the password is set
Data Logging
The optional USB Data Logger module allows a large amount of data from your
scale to be recorded into a USB Flash Drive. The Flash Drive maybe removed
from the scale and inserted into a standard USB port on a PC. The data is
collected on a
CSV file (comma-separated values), which can be opened using any spreadsheet
software such as Microsoft Excel®.
Data Logger is activated by default. To activate/deactivate it, press **_Menu
-
Option Setups -> Data Logging. _**On the horizontal menu displayed in the Data Logging setup screen, select ACTV to activate/deactivate Data Logging.
Basic Operation
There are two modes of operation for using the USB Data Logger:
Triggered – The weight entry is written to the USB when the user
presses a button
Continuous – The weight entry is written to the USB at regular time
intervals selected by the user.
To switch modes of operation, on the Data Logging setup screen select TYPE on
the horizontal menu to switch between triggered and continuous type logging.
In Triggered Mode, press the TWO (2) key each time to log the current weight
and time on the USB drive. This will be noted
by a “Recording Entry” message on the status message area of the screen each
time the TWO key is pressed.
In Continuous Mode, press the TWO key to begin the data logging sequence. A
“Data Logging On” message will be displayed in the status area, followed by
the “Recording Entry” message after the time interval has passed. After that,
“Recording Entry” message will appear on regular time intervals specified by
the user. The time interval can be set in the Data Logging setup screen.
Select the TIME option on the horizontal menu to enter the time interval in
seconds.
Note about Multiple Platforms
For Multiple platform, when the scale is in the Multiple Platform screen
(where all the platforms are viewed at the same time with a TOTAL), the TWO
key is already occupied for switching between platforms. Instead, press the
SHIFT key (ON/OFF) and then the ENTER key in succession. Please make sure to
see the “Shft” notification on the top left corner of the screen before
pressing the ENTER key. The ENTER key may have to be held a half-a-second
longer to register the “Recording Entry” notification.
The SHIFT+ENTER key-sequence will have the same effect as the TWO key
described above in the Basic Operation section.
USB Flash Drive Use
Your Data Logging scale comes with a 1 Gigabyte (1 GB) USB flash drive that can hold more than 4 million readings! For those scales equipped with Data Loggers, the scale will always test the Data Logger system and check for a USB flash drive at power up. If a USB Flash drive is plugged in, the scale will show the message “USB Stick found…” and if the USB flash is not plugged in, then the message will simply say, “USB Stick not found…”. At times, the scale may show “USB Stick not found…” even when a USB flash drive is plugged in. In this case, simply recycle the power and it should initialized the Data Logger system correctly.
If the USB flash drive has never been used with the scale, the scale software will create the file DATA.CSV automatically. This is the file the scale uses to store the weight entries. When recording new data on a card that already contains the file, the scale will look for the file and append the new weight entries to the end. This file can then be opened using Microsoft Excel® or a text editor.
It is okay to remove the USB drive while the scale is on. But if the user
wants to plug in the USB drive back again, the user has to:
Shut off the scale, Plug in the USB drive, Wait for 5 seconds Turn on the
scale.
The USB drive cannot be plugged in while the scale is on since the USB Data
Logging system initialization routine happens only at power up. But there is
no harm if the USB drive is plugged in while the scale is on and an attempt is
made to write data into it. In this case, the status message will display
“Recording Entry”, followed by “Can’t Open File…” This means that the file
DATA.CSV cannot be opened at this time.
The USB drive comes pre-formatted. It is not recommended that you reformat the
USB drive.
When deleting weight entries from the USB Flash Drive, it is important not to
leave the file empty. The data on the drive can be anything – a period or a
space or carriage return – so long as there is some data in the file.
If you are using the scale in Continuous Mode , we recommend that you
temporarily put the scale into Triggered Mode while the USB drive is plugged
in and test it for operation.
Advanced Operation
WARNING: The following descriptions illustrate the advanced operation of
changing the frames being printed on the USB flash drive. Incorrect usage of the features described in this section will cause the scale to malfunction. Proceed with caution.
The USB Data Logger can be used to store more than just the time and weight.
The system uses Print Frames to determine what data can be stored on the USB
flash drive.
The Print Frames feature can be accessed by going to SETUP MENU/OPTION
SETUPS/RS232/PRINT FRAMES. Ignore the connotation RS232, as it only implicates
the RS232 transceiver not present in these scales.
In the Print Frame Definitions screen, the frame named “USB” applies to the
frame of data printed on the USB flash drive.
The factory default USB frame is:
FUNCTN | SOURCE | C L SP | Comments |
---|---|---|---|
Time24 | 1 | Line #1, the time and 1 space | |
Weight | Plat#00X | 1 | Line #2, Weight |
UniTxt | Plat#00X | √ √ 0 | Line #3, Units |
To change the frame to include more data types, you can refer to instructions
outlined in the Print Frames section defined for RS232 option in page 5.
Please note that a frame is only saved when the user moves out of the Print
Frame Definitions screen. If the scale is turned off while the user is still
on this screen, the frame will be lost permanently.
Bluetooth
NOTE: Some options labeled as [N/A] are either not available or not fully
implemented yet. Setting these parameters may result into a serious malfunction of your Bluetooth device.
Introduction
The optional Bluetooth module allows you to transfer data wirelessly for a
short distance to another Bluetooth unit. This is a simple option, which uses
the ease of use and popularity of Bluetooth devices for information transfer
while stripping the complexities and cost of higher-end wireless communication
such as WiFi.
The Bluetooth module uses a Frame Definition database built into the scale to
provide the user the ability to print custom frames of data based on various
requirements that may need to be accomplished.
Important Notes for Bluetooth Operation
Bluetooth devices have some notable limitations and some precautions need to be taken for proper Bluetooth communication.
Maximum Distance
The Bluetooth module implemented in the scale is a standard Class 2 Bluetooth device. This means that it can only communicate with another Bluetooth device within 30ft in the line of sight. If there are obstructions within this line of sight, such as concrete walls, then that distance is further diminished.
Interference
It is recommended that to get the best signal out of the Bluetooth enabled indicator, no metal should be placed near the indicator. Metals disrupt the signal coming from the Bluetooth module and might produce noise interference in the communication resulting in corrupt data being transmitted. The metal also might cause the signal range to be diminished causing serious reduction in communication range.
Overview of Menu Options
This section explains the various menu options available in the indicator to manipulate the Bluetooth module. The actual sequence of the operation will be covered in the next section (see “Basic Operation”). These menu options can be accessed by pressing MENU key on the main screen, then selecting OPTIONS SETUP->BLUETOOTH SETUP.
Bluetooth Reset
This menu option executes a soft reset on the internal Bluetooth. This reset is usually performed when the Bluetooth module is performing erratically.
Scan Remote Devices
This option allows you to scan for any Bluetooth devices present in the vicinity. All Bluetooth devices scanned will be placedin a database, which can be accessed using the “List All Devices” option .
Find Remote Devices [N/A]
This option allows you to scan for a specific device in the vicinity.
List All Devices
This option lists all the devices that have been scanned using the “Scan Remote Devices” option. In this list, you can check
the properties of a remote device, connect to a remote device and even check the connection status of a remote device.
Quick Connect Device
The Quick Connect option allows fast-connection to a particular device based on used defined parameters, without going through menu options. The user may select the name of the device to connect to or the specific address. The connection process is further explained in the section “Quick Connect Operation ”.
Bluetooth Local Settings (Advanced)
This option allows the user to set settings for the internal Bluetooth module, such as the local name, aerial virtual COM port and factory settings.
Basic Operation
Pairing and Default Pin
Please note there might be a need to pair the scale with the remote Bluetooth device. The default pin for the scale is ‘0000’. As soon as you get the prompt to enter a pin for pairing on your remote device, enter the stated pin to commence pairing. Pairing can be before or during the connection sequence. The Default Pin option is further explained in the Advance Settings section. Instructions are provided on how to change the default pin, or if so desired, to eliminate the default pin and use manual entry on the fly.
Connection
There are two types of connections that are going to be discussed in this section; Outgoing Connection and Incoming Connection.
Outgoing Connection
Outgoing connection is the connection that the user initiates from the scale. The user has full control on how this connection will be established. The following steps explain how to conduct this connection.
- As recommended, before conducting any Bluetooth operations perform a soft reset of the Bluetooth module to bring it to a known
- To establish an outgoing connection to a remote device, you must first scan the area for that device so that the scale can add it to its internal remote device database. From the Bluetooth Setup option menu, select “Scan Remote Devices” to detect any Bluetooth devices in the area. You can also do this directly in the “List All Devices” section. The following status messages should appear at the top of the
“Scanning Remote Devices” – Obtaining remote device addresses from the
vicinity
“Scanning Remote Names” – Obtaining remote names associated with the
addresses obtained.
“Done!” – Operations done
- Go to “List All Devices” to see what devices are available in the vicinity. In the “Bluetooth Devices” screen, you will see all Bluetooth devices that have been detected in the area. If you don’t see the device you are looking for, select the “SCAN” option on the horizontal menu below the list and press Since the screen can only show you ipto six devices at a time, you may have to use the UP and DOWN buttons to select a device as there maybe more than 6 devices detected. At the bottom of the screen, there are various operations that can be done on the devices detected.
- SCAN: Scan the surrounding area to find Bluetooth devices. The system will keep adding new devices to the list without erasing the previous devices found. At some point, the list may become “stale” i.e. the list is outdated and the devices in the list are no longer in the surrounding area. If this is the case, you may need to clear the To do this, go back to the main menu and select “Scan Remote Devices”. Alternatively, you can use “Bluetooth Reset” to clear the list.
- CONN [CONNECT] : Connect to the selected remote Bluetooth
- PRPT [PROPTERTIES] : Check the properties of the remote
- Name: Name of remote device
- Class: Device class of remote device, such as headset, desktop,
- Address: Address of
- STAT [STATUS] : Check the connection status of the remote This is only active if a connection has been established.
- Name: Name of remote device
- Connected: Connection status (YES or NO)
- Paired: Shows if the device is paired or [N/A]
- Port: The current local port used for connection.
- Port: The remote port of the connected device.
- Signal: Signal strength of the
- Select CONNECT on the horizontal menu and press the ENTER key to connect to a desired device. If you see a “Link Established…” message followed by a checkmark being placed besides the intended connected device, you have successfully connected to the If you see any other messages, or the checkmark was not placed, then the device was not connected.
- To check the status of the connection, select STATUS from the horizontal This will give some information about the connection as stated in step 3.
- Go back to main weight A Bluetooth indicator () shows at the bottom of the screen to indicate that a Bluetooth connection is in progress. You can now use the PRINT button on the front panel to send weight information to the remote device.
- To disconnect, go back to the Bluetooth Options menu, in “List All Devices”. Select the device that is currently connected (the device with a checkmark beside it). Select CONNECT from the horizontal menu and press ENTER The connection will be destroyed. The status message “Link Released…” will appear at the top of the screen. If this status message does not appear, then the link was not released properly.
- An easier way to disconnect is using the “Quick Disconnect” This is explained in the next section.
Incoming Connection
Incoming connection is the connection that comes from a remote device. The user of the scale can only choose whether to connect or not to the incoming device, if pairing has been enabled (no default pins defined). The following steps instruct the user how to manage an incoming connection.
- If the incoming connection from a device is a new device that has never been used before, a Pin Entry dialog box will appear if the default pin was not set (see Bluetooth Local Settings section). If the default pin was not set, and the wrong pin is entered from the remote device, then the connection will not be
- If the incoming connection from a device is a device that has been paired before, then the connection is There is no need for user interaction here. The link will be established and the device will appear in the “List All Devices” section. The Bluetooth indicator will also appear at the bottom of the screen to inform the user that a Bluetooth connection is in progress.
- You may now receive data or press the PRINT button to send the data to the remote
QuickConnect Operation
Introduction
The QuickConnect feature simplifies the connection process for everyday use so
that the user does not have to go through the menu process to establish a
simple connection. The QuickConnect feature allows the user to connect to a
predefined remote device straight from the front panel while on the main
weight screen.
The user can choose to predefine the Bluetooth remote device’s name or
address, so that when the user conducts a QuickConnect, the scale will look
for a device in the vicinity that matches the predefined name or address and
establish a connection immediately.
The QuickConnect can also start “automatically” at reboot without any user
interaction. This means that if the desired
QuickConnect set device is in range, as soon as the indicator comes out of
startup, it will connect immediately to this device.
Setup
To setup QuickConnect, select “QuickConnect Device” in the Bluetooth setup
menu. A screen shows up showing fields to
enter values for Quick Connect.
- N: Name of the remote device to perform a quick connect
- A: Address of the remote device to perform a quick connect
AutoStart:
- YES: This enables The QuickConnect process will begin at startup.
- NO: This disables The QuickConnect needs to be initiated manually. This is explained below
under “QuickConnect”.
ADVANCED
[After the above area, the rest is advanced settings that only should be played with if you have some knowledge about Bluetooth functionality]
- LPort: The default local port to use for
- Rport: The default remote port to use for
To change these fields, a horizontal menu is provided for your use.
- LIST: Opens a list of detected devices in the vicinity. This is the easiest way of entering a predefined device for quick connect purposes. Simply make sure that the scale is near the remote device you want to connect to. Select this option and press the ENTER The scale will scan the area for Bluetooth devices and present you with a list of detected devices. Select a device and press ENTER to fill the Quick Connect fields.
- AUTO: Toggle AutoStart option for Yes/No.
- NAME: Manually enter the name of the device you want to quick connect
- ADDR: Manually enter the address of the device you want to quick connect
- LPRT: Enter the local port you want to use for link [Set auto]
- RPRT: Enter the remote port want to use for link connection.
When you are satisfied with the settings, press the MENU key. This will save your settings so that the next time you power up the scale, the settings will be preloaded. There is no need to come into this menu option to redo the settings again, unless you want to change the Quick Connect parameters.
QuickConnect Process
Once the settings have been established, head over to the main weight screen. Make sure there are no connections currently in progress (check to see if the Bluetooth indicator is showing).
AutoStart
To demonstrate how the system will automatically connect to the set QuickConnect device, make sure the QuickConnect device is in the vicinity of the scale. Press the ON/OFF key until the screen goes blank. Wait for 5 seconds. Press the
ON/OFF key again and wait. After the weight appears on screen, the message “Starting QuickConnect” appears on the top part of the screen. Then you will see “Verifying Service”. After this message, you will see “Link Established. This means the QuickConnect session was successful. If you see “Service Check Failed!” and “Device not found”, this means the QuickConnect device is not in the vicinity, or in more advanced setting, the Local/Remote port is not right. If the session was successful, the Bluetooth indicator will appear. Press PRINT key to send data to the device.
Manual
For a manual operation, press ON/OFF and then press 5 (within 3 seconds). Immediately, the scale will scan the current area to see if the device you defined in the Quick Connect screen is present. If it is present, it will connect to this device and show the Bluetooth indicator. If the device is not found, then the appropriate status message will be shown. You may now send and receive data from the device. Press PRINT key to send data to the device.
Quick Disconnect
If you want to disconnect from the device, you can either use the Quick Disconnect feature or go to the Bluetooth Setup option and disconnect from there. To quick disconnect the device, use the same sequence that is described above; i.e. press ON/OFF and then press 5 (within 3 seconds). This quick disconnection sequence can also be used if you have connected to the device without going through Quick Connect.
Bluetooth Local Settings (ADVANCED)
This section introduces some of the advanced features available for the scale Bluetooth module. Most of these features have to do with the local configuration of Bluetooth module. It is recommended that you are well versed with the above described operation before you get into this section. Most of the time, you will not need to modify these settings.
Following are the local Bluetooth settings that can be modified to satisfy some specific requirements in your application. Toaccess these settings, in the Bluetooth Setup screen, scroll down “Bluetooth Local Settings.”
Bluetooth Enable
This menu option allows you to Enable or Disable the Bluetooth functionality. After toggling this option, the scale needs to be restarted.
View Device Info
This menu option allows to view the Bluetooth modules name and device address.
Change Local Name
This menu option allows you to change the local name of the scale Bluetooth module. Currently, it is defaulted to “ARLYN SCALE”. You can modify this to anything you want depending on your needs.
Virtual Port Settings
This menu option specifies the characteristics of the aerial Virtual COM port through which to send out data. This setting, if changed improperly, may disable your Bluetooth communication system.
Fixed Pin Settings
This shows the current Default Pin settings of the Bluetooth module. This
allows the Bluetooth module to send this pin to any device asking for a pin
request. You will never need to enter the pin on the scale.
Entering this option, the default pin is shown on screen. The default pin is
‘0000’. You can press ENTER to change the
default pin. You have two options here:
- Default Fixed Pin: You can put a default pin, of up to 16 alphanumeric characters in The next time a pairing is needed, this pin has to be entered on the other remote device when a dialog box pops up.
- No Fixed Pin: By pressing ENTER without entering a pin, the scale will assume that you want to enter the pin each time a pairing is The next time you pair with a device, a dialog box will pop up on the scale screen for you to enter a desired pin. That same pin has to be entered in the remote device as well.
Reset Factory Settings
Selecting this option will default the Bluetooth module to the original settings set by the Bluetooth Manufacturer. If the Bluetooth is acting erratically, and/or you might have changed settings that may have disabled Bluetooth communication, use this option to bring the settings back to original factory configuration.
Frames Settings
WARNING: The following descriptions illustrate the advanced operation of
changing the frames being printed on the USB flash drive. Incorrect usage of the features described in this section will cause the scale to malfunction. Proceed with caution.
The Bluetooth module sends out weight information by pressing the PRINT key on
the front panel. The system uses Print Frames to determine what data can be
sent out. Currently, the frame contains weight information.
The Print Frames feature can be accessed by going to SETUP MENU/OPTION
SETUPS/RS232/PRINT FRAMES. Ignore the connotation RS232, as it only implicates
the RS232 transceiver not present in these scales.
In the Print Frame Definitions screen, look for the frame named Frame#001
which is the default frame data being sent out. This is current configuration.
FUNCTN | SOURCE | C L SP | Comments |
---|---|---|---|
Weight | Plat#00X | 1 | Line #1, Weight |
UniTxt | Plat#00X | √ √ 0 | Line #2, Units |
To change the frame to include more data types, you can refer to instructions
outlined in the Print Frames section defined for RS232 option in page 5.
Please note that a frame is only saved when the user moves out of the Print
Frame Definitions screen. If the scale is turned off while the user is still
on this screen, the frame will be lost permanently.
Arlyn Scales Battery Option Addendum
- Plug in power transformer & the “Ready” sign will come
- Place the toggle switch on the right side of indicator to the “On”
- Let the batteries charge up (approx. 8hrs).
- At this point your scale should be operational & ready to
- You can leave the power transformer plugged in while using the scale
- To shut off the scale press the “On/Off” key
- To turn on the scales press the “On/Off” key and you will see the “Ready” come
- Press the “On/Off” key again & your scale will be
- Only shut off toggle switch if you are not going to use the scale for a long period of
- If the toggle switch is not on the “On” position the scale will not
_Note_
The scale will automatically shut off if the scale is not being used in 15min.
You can change the “Shut Off Time” or disable the shut off time internally in
the system menu.
Any problems please call the Service Dept. @
800-645-4301 x101
Read More About This Manual & Download PDF:
References
- Industrial Scales | Custom Weighing Solutions┃Arlyn Scales
- Industrial Scales | Custom Weighing Solutions┃Arlyn Scales
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>