AL-OPTIC 71739 Vision Screener User Manual
- June 4, 2024
- AL-OPTIC
Table of Contents
AL-OPTIC 71739 Vision Screener
Getting to KnowAI-OpticVisionScreener
This manual gives a brief introduction to the AI-Optic Vision Screener device. For more detailed information about the product, such as the product features, basic operations, technical specifications, and maintenance, service, andwarranty information, refer to the paper user manual included with the product. You can also download the electronic user manual by visiting www.ai- optic.com and finding the Support section.
Packing list
Once you receive the product, make sure that the packaging is undamagedand
that all contents are accounted for. If the outer packaging is damagedor
anything in the package is missing, contact the AI-Optic Vision Screener
distributor
- Product Instruction x 1
- Type-C power cable x 1
- Power adapter X 1
- Tripod x 1
- AI-Optic Vision Screener x 1
Charging
The AI-Optic Vision Screener supports two power modes: Wired modeandBattery mode
-
Wired mode is when the power cable is plugged in and receivingpower
from an external power source. -
Battery mode is when the device is unplugged from a power source
Be sure to charge the AI-Optic Vision Screener before using it for the first
time. To charge the battery, plug the Type-C connector into the USB port of
thedevice and the Type-A connector into the USB port of a power sourceor
computer.
When the screen displays the charging status icon, your device is beingcharged
It takes two hours to fully charge the battery, which holds a two hour charge.
When the battery is less than 20%, the system automatically displays
alowbattery icon, as shown in Figure 1-4. In this case, plug the Type
Cconnector into the USB port of the device to charge the device.
You can use the device even while it is being charged.
Power on and off
Power on: Press and hold the power button for 2 seconds to turn onthedevice.
Power off: Tap the Power off icon in the navigation bar on the Home screen; or
press and hold the power button for 4 seconds, and tap Yes intheconfirmation
dialog box that appears.
If the device does not respond, press and hold the power button for 6
secondsto force a shut down.
Screen-off protection
While the device is on, if it goes unused for 5 minutes, the
screenautomatically turns off. To wake it up, just tap anywhere on the screen.
If the device is idle for 10 minutes, the device is automatically poweredoff.
Touse the device, you need to power the device on again. To change the
screen-off protection settings, choose Settings > ScreenOff
Time on the Home screen and specify a desired value.
Product appearance
The AI-Optic Vision Screener is equipped with a 7-inch display screen, withacapacitive touch sensor at the rear. To use a specific function, just touchthecorresponding icon on the Home screen. The screen and touch sensor areprotected by a thin glass cover.
USB port
The AI-Optic Vision Screener has a USB power port for intended onlyfor
charging, not for data transmission with other devices. Do not plug any
external devices into this USB port. Otherwise, the devicemay be damaged.
Daily use
When you use the AI-Optic Vision Screener, place it in the way showninFigure
1-10.
During examination, the device takes a series of photos. The subject must look
into the camera; otherwise, the examination could fail.
Serial number
The serial number is at the bottom of the product. You will need to provideit
inthe event that you need to contact our customer service center.
Wi-Fi andSettings
Wi-Fi
Once you turn on the power, the system will show a loading screen followedby a
Wi-Fi option screen.
Select your desired Wi-Fi network, enter the password, and click Connect
tocomplete the Wi-Fi connection. To change the Wi-Fi settings, chooseSettings
Wi-Fi on the Home screen, select another network, and enter thepassword to connect to this network.
Settings
Tap Settings on the Home screen to go to the Settings screen.
You can tap the Back icon in the lower-left corner to return to the
Homescreen.
Language
On the Settings screen, you can tap Language to change the systemlanguage. The system supports two languages: English and Chinese.
Wi-Fi
On the Settings screen, tap Wi-Fi to go to the Wi-Fi configuration screen. Select your desired Wi-Fi network, enter the password, and tap Connect.
Factory Settings
On the Settings screen, you can tap Factory Settings to restore all settingsand configurations to default factory settings.
Time and Time Zone
After the device is connected to the Wi-Fi, the system time is
automaticallyupdated. The AI-Optic Vision Screener uses the international
standard date format of
YEAR-MONTH-DAY. For example, January 1 st 2020 would appear as 2020- 01-01.
The default time zone on the device is UTC+0. To change the time zone, tapTime
Zone on the Settings screen and select the country andthecorresponding time
zone. The system automatically updates the timebasedon your local time zone.
Referral Criteria
The AI-Optic Vision Screener features three examination settings: Arnold,
Matta/Silbert, and Arthur2. The default setting is Arnold. The
sensitivityandspecificity of the Arnold setting is best suited for a routine
examination.
If you are in a medically developed country or region, you can
chooseMatta/Silbert. This option is more sensitive and designed for
childrentoreceive an examination.
If you are in a medically underdeveloped country or region, you can choose
Arthur2. This option features high specificity and thus helps
reduceunnecessary referrals.
Screen Off Time
On the Screen Off Time settings screen, you can set Auto-Timeout andAuto-Power off. In the Auto-Timeout drop-down list box, tap the downarrowand select a value as required. In the Auto-Power off drop-down list box, tapthe down arrow and select a value as required.
System Update
In the system upgrade interface, if you need to check the update, click “Yes”, the machine will check whether there is the latest version in the stateof networking, if there is, you can download and install the latest software.
Machine ID
Each AI-Optic Vision Screener has a unique ID. You can click MachineIDonthe setting interface and scan the QR code on the screen to obtain the ID.
IP Address
This is set up before delivery by the manufacturer, and users do not needtomake any operation changes.
Examination Procedure
The inspection light source of the AI-Optic Vision Screener is infraredlight, which is safe for human eyes. Infrared light can avoid causing glareandmiosis.
Select a referral criterion
The AI-Optic Vision Screener features three examination settings: Arnold,
Matta/Silbert, and Arthur2. The default setting is Arnold. The
sensitivityandspecificity of the Arnold setting is best suited for a routine
examination.
If you are in a medically developed country or region, you can
chooseMatta/Silbert. This option is more sensitive and designed for
childrentoreceive an examination.
If you are in a medically underdeveloped country or region, you can
chooseArthur2.This option features high specificity and thus helps
reduceunnecessary referrals.
Enter the basic information of theSubject
Enter the name, gender, and date of birth of the subject.
Prepare the examination environment
To obtain the correct measurement values, we recommend that you performthe examination indoors and avoid direct sunlight during the examination. If the room has lots of windows, try to close as many curtains or blindsaspossible. The brightness value should be 5 or below. If the value is greater than5, theflash cannot be turned on, and you need to reduce the brightness of thesurrounding environment. If the brightness is 5 or below, turn on theflash. Then, you can start the procedure. The examination environment must be quiet and free fromdistraction. Thesubject must hold still during the procedure, so make sure that if the patient isa child that they are able to hold still.
Check the examination distance
The recommended examination distance is 3-4 feet away fromthe device.
First, place the device at eye level of the subject, about 3-4 feet away
fromtheir face.
Then, adjust the position of both the device and the subject to make surethat
their eyes are seen on the screen. Move the device back and forth until aclear
corneal reflection point is obtained. Then, you are ready to start
themeasurement.
To ensure that results are as accurate as possible, make sure that the subject
looks directly into the device during the examination. The subject shouldsit
upstraight, with the body facing the camera.
Please remove glasses or contact lenses before starting the exam.
Start examination
To start the examination, tap the camera icon on the screen. During shooting,
make sure that the camera is oriented vertically, andkeepthe camera at eye
level of the subject.
In addition, in order to improve the stability and accuracy of the
measurement, we recommend that customers use the included tripod stabilizing
device.
Remind the subject to avoid blinking during the examination,
otherwisetheexamination will fail. The entire examination process lasts about
5 seconds. After the scaniscompleted, a dialog box appears, as shown in Figure
3-5. You can tapResult to check the examination results.
View examination results
After the scan is completed, the examination results will be ready
withinapproximately 10 seconds. After you tap Result, the name of the subject,
thedate of examination, the spherical degrees and cylinder degrees of
botheyes, and the pupillary distance are displayed.
If the examination result is good, the system displays Passed in green,
asshown in Figure 3-6. If any problem is identified, the systemdisplays
Referral Required in red, as shown in Figure 3-7.
After each measurement, you can tap the Home icon to return to theHomescreen.
Operations related to subject information
Input the information of a new subject
You can input the information of a new subject directly on the Home screen
Retrieve the information of an existing subject.
Tap the Retrieval icon at the bottom of the Home screen, to go to the
retrieval screen. Enter the name of the subject and verify the date of birth
of thesubject,to avoid mistakes caused by duplicate names. After the
corresponding name is found, you can view the examination records of
thesubject or start a new examination.
Save or delete data
Every time an examination is completed, the AI-Optic Vision Screener
automatically saves the examination results by default. To delete a result,
tapthe Retrieval icon on the Home screen, enter the name of the subject,
findthe corresponding entry, and then delete the result.
Exporting Data
After the check, if you need to export the data for easy viewing, you
canclickthe export button in the menu bar below, and then enter the email
addresstoreceive the summary email about the data. You can download it as
needed.
Troubleshooting
The device cannot be turnedon
If your device cannot start, check the power supply first to make sureit isplugged in. Connect the power adapter to the power source and pressthepower button again to see if the device can start.
The device is automatically poweredoffafter power on
Connect the charging cable first. The system will show the charging statusonthe screen. Restart in about 30 minutes.
The screen suddenly turns off
The screen will automatically turn off when the device is idle for a specifiedperiod. This is to save power. You can touch the screen to wake it up. If thescreen does not wake up, the device is powered off. You just need topower on the device again. Note: You can tap Screen-off Time on the Settings screen to changetheautomatic screen-off or power-off time.
Common reasons for measurement failure
When the scan fails during examination, check the following possible causesin
sequence.
The image is not clear
Cause: If the image is not clear, it’s because the measurement distance is too far or too close.
Solution: We recommend that you place the device about 3 feet away fromthesubject. Adjust the distance between the device and the eyes of the subject to ensurethat the eyes of the subject can be seen on the screen. Then, move the devicebackand forth until a clear corneal reflection point is obtained and the eyelashes andeyebrows of the subject can be clearly seen on the screen. When the measurement distance displayed on the screen is 95-105 cm, you can start measuring.
The subject does not look at the camera
Cause: When the subject is not facing the device, or looking away
fromthecamera during the examination.
Solution: Remind the subject to turn their body towards the camera,
andlookdirectly into the lens. Sit up straight, and try not to move or blink
duringtheexamination.
Glasses are found
Cause: The subject has not taken off the glasses during the examination. Solution: Ask the subject to remove the frame glasses and restart theexamination.
No pupils are found
Cause: The subject has organic eye diseases such as corneal scars, turbidrefractive media, cataracts, keratoconus, or retinal detachment, or thepupilsof the subject are blocked by their cap, eyelashes, or hair. Solution: Proceed to the referral process in case of any eye disease. Makesure their hair or eyelashes are not blocked the eyes.
The brightness value is excessively large
Cause: When the surrounding environment is too bright, the brightness valueon
the screen is too high. Solution: Close the curtains or blinds to block out
external sunlight. Incandescent
lights tend to affect the performance of the examination, but fluorescent
andLEDlights are okay. If the brightness value on the screen is less than 5,
tap the flashiconto turn on the flash.
The measurement is not completed
Cause: If the subject closes their eyes or blinks frequently during the measurement process, the process can read as incomplete. This can also happen when the Wi-Fi network is unstable.
Solution: Remind the subject to remain still and avoid frequent blinking. If the issueiswith the Wi-Fi, connect to another network and restart the examination.
Eyes are closed or blocked
Cause: When the subject closes their eyes or their eyes are blocked by hair or other things, the examination will fail.
Solution: Remind the subject to avoid frequent blinking or remove obstructionsinfront of their eyes
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>