TEAC TD-260T Digital Transducer Indicator Instructions
- June 13, 2024
- TEAC
Table of Contents
- TEAC TD-260T Digital Transducer Indicator
- Product Usage Instructions
- Introduction
- Safety Instructions
- Controls and Functions
- Decision output LEDs
- Decision LED (JUDGE)
- Setup Mode
- of times
- of times
- of
- Error Messages
- Calibration
- Setting Functions
- Digital Zero Function
- of times
- of times V
- of times/Sec
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
TEAC TD-260T Digital Transducer Indicator
Product Information: Digital Transducer Indicator TD-260T
The TD-260T is a digital transducer indicator designed for accurate and efficient measurement in testing and manufacturing equipment. It features a directly readable, 5-digit digital display with a reading scale of -19999 to +99999. The indicator supports easy sensitivity calibration through TEDS (Transducer Electronic Data Sheet) support, allowing for auto-calibration using TEDS transducer connection. It also includes a remote sensing function to ensure measurement accuracy even with the use of long cables.
Other notable features of the TD-260T include equivalent input calibration, which eliminates the need for an actual load during sensitivity adjustment, and support for static strain measurement, enabling easy identification of defects on load cells caused by plastic deformation. The indicator also offers various comparison functions, such as High-High and Low-Low limits, as well as High and Low limits.
The TD-260T is designed with a DIN-size form factor, making it suitable for installation in testing and manufacturing equipment. It is RoHS compliant, ensuring environmentally friendly operation.
For power supply options, the TD-260T is available in both AC power and DC power models. The AC power model comes with additional accessories such as an instructions CD, AC power cord, flathead micro screwdriver, and input/output connector plugs. The DC power model includes similar accessories except for the AC power cord. Additionally, there is an option-installed model available with BCD or D/A option support.
Product Usage Instructions
-
Before using the TD-260T, carefully read the entire user manual to ensure safe and proper operation.
-
Check the package contents and verify that all accessories are included and undamaged. If anything is missing or damaged, contact the manufacturer for assistance.
-
If you have ordered an option-installed model, confirm that the received product package matches your specifications.
-
Determine the power supply type of the TD-260T (AC power or DC power) and ensure it matches your order.
-
Follow the safety instructions provided in the user manual to prevent personal injury or property damage.
-
When connecting the TD-260T to a load cell, external I/O, or option, use a shielded cable for optimal performance.
-
Allow at least five seconds between turning on and off the TD-260T to prevent any issues.
-
Avoid touching the rear panel or connectors while the product is turned on to prevent electrical hazards.
-
Before connecting to the signal I/O connector, verify the signal name and pin assignment to avoid incorrect connections.
Always turn off the power to the TD-260T before making any connections. -
For sensitivity calibration, utilize the TEDS support for easy auto-calibration or follow the equivalent input calibration method if no actual load is available.
-
If required, utilize the static strain measurement feature to identify defects caused by plastic deformation in load cells.
-
Explore the different hold functions available, including peak hold, bottom hold, peak-to-peak hold, and block setting for each hold.
-
If analog voltage output is needed, consider the D/A option for producing an output corresponding to the TD-260T’s reading.
For further assistance or inquiries, refer to the contact information provided in the user manual.
D01156700H
Z
Digital Transducer Indicator
TD-260T
Instructions for Use
Introduction
Thank you for purchasing the TD-260T digital transducer indicator (“TD-260T”).
Please read this document in its entirety before using the product to get the
best performance and ensure safe and proper operation.
Features
- Directly readable, 5-digit digital display (reading scale -19999 +99999)
- Easy sensitivity calibration facilitated by TEDS support. Auto-calibration using TEDS transducer connection TEDS: Transducer Electronic Data Sheet
- Remote sensing function, ensuring measurement accuracy even with the use of long cables
- Equivalent input calibration. No actual load required for sensitivity adjustment
- Static strain measurement supported, allowing easy identification of a defect on a load cell caused by its plastic deformation
- Comparison functions supported, including for High-High and Low-Low limits as well as High and Low limits
- Different types of hold functions available: peak hold, bottom hold, peak-to-peak hold, and block setting for each hold
- DIN-size design for installation in testing and manufacturing equipment
- Analog voltage output supported. D/A option available for producing an output that corresponds to the TD-260T’s reading
- RoHS compliant
Checking Accessories
If anything is missing or damaged, contact us. (For contact information, see
the last page.)
Accessories for AC power model Instructions for Use (CD) x 1 AC power cord x 1
Flathead micro screwdriver x 1 Input/output connector plugs B2L 3.50/18/180F
SN BK BX or its equivalent x 1 BCZ3.81/07/180F SN BK BX or its equivalent x 1
Accessories for DC power model Instructions for Use (CD) x 1 Flathead micro
screwdriver x 1 Input/output connector plugs B2L 3.50/18/180F SN BK BX or its
equivalent x 1 BCZ 3.81/07/180F SN BK BX or its equivalent x 1
Additional accessory for BCD or D/A option-installed model Ferrite core x 1
An option-installed model is also available in the TD-260T. Before operating
the product, make sure that the product package you have received is the one
you specified when you placed an order. Note that there are two types of power
supply: AC power and DC power. Check the power supply of the product received
and determine that it is the one you ordered.
NOTE
Any data, including, but not limited to information, described herein are
intended only as illustrations of such data and/or information and not as the
specifications for such data and/or information. TEAC CORPORATION disclaims
any warranty that any use of such data and/or information shall be free from
infringement of any third party’s intellectual property rights or other
proprietary rights, and further, assumes no liability of whatsoever nature in
the event of any such infringement, or arising from or connected with or
related to the use of such data and/or information.
Company names and product names in this document are the trademarks or
registered trademarks of their respective owners.
2
Safety Instructions
This document describes the safety instructions for the operation of the TD-
260T. Before operating the product, read this document carefully to
familiarize yourself with the product.
Warning Follow the instructions below to avoid risk of serious personal injury
and death.
Never use beyond the rated specifications as there is a danger of property
damage, injury, fire or electrical shock. Never use in flammable gaseous
environments including the following locations as there is a danger of
explosion. · Locations subject to corrosive or flammable gases · Locations
subject to water, oil or chemical splash If the product malfunctions (any
smoke, odor or noise is present), stop operating the product immediately and
unplug the power cord as there is a danger of fire or electrical shock. Never
attempt to disassemble the product. Carefully check connections and wiring
before applying power. Be sure to ground the product (with ground resistance
of 100 or less). The electric shock protection class of this product is Class
I. The product must be grounded to the protective ground terminal for
protection against electric shock. To allow the operator to immediately shut
off the power to the product, install a switch or circuit breaker that
complies with both IEC60947-1 and IEC60947-3 requirements near the product.
The switch or breaker installed must also indicate that it functions to shut
off the power to the TD-260T. Double insulation or reinforced insulation must
be provided between the TD-260T and an external device connected to it and
between the external device and its power source.
Caution
Follow the instructions below to avoid risk of personal injury or property damage.
Disconnect the power cord when performing the following. · Attaching or
detaching optional connectors · Wiring or connecting cables to terminal blocks
· Connecting the ground line
Wait for at least five seconds between turning on and off.
Never touch the rear panel or connectors while the product is turned on.
When connecting to the signal I/O connector, check the signal name and the pin
assignment to prevent a false connection. Be sure to shut off the power to the
product as well when making a connection.
Use a shielded cable when connecting to a load-cell, external I/O or option.
Take adequate measures when operating in the following locations. · Near a
power line · Where a strong electric or magnetic field is present · Where
static electricity or noise such as from a relay is generated
Do not install under the following environments. · Locations subject to
temperatures exceeding the specified temperature and humidity ranges ·
Locations with high salt or iron content · Locations under direct vibration or
shock · Outdoor, or locations with an altitude of higher than 2,000 m
Do not operate a damaged unit.
The TD-260T is classified as an open-type (built-in) device and must be
installed inside a control panel.
If the top cover and the panel are dirty, wipe them with a soft cloth lightly
dampened with a solution of a neutral detergent and water, and then wipe them
again with a damp, well wrung-out cloth. DO NOT use a chemically treated dust
cloth, paint thinner, or other flammable solvents. Using any of them could
damage the coating of the product.
If the product is used in a manner unintended by the manufacturer, the product
safety may be adversely affected.
3
Table of Contents
Introduction………………………………………………………………………………………… 2 Features
…………………………………………………………………………………………………………. 2 Checking
Accessories……………………………………………………………………………….. 2
Safety Instructions ……………………………………………………………………………. 3
1. Controls and Functions ……………………………………………………………. 6 1-1. Front panel
…………………………………………………………………………………………… 6 1-2. Rear
panel……………………………………………………………………………………………… 8 1-2-1. Signal I/O terminal
block………………………………………………………… 8 1-2-2. Control I/O terminal block
…………………………………………………….. 9 1-2-3. Control I/O
signals…………………………………………………………………….. 9
2. Hardware Installation …………………………………………………………………10
3. Connections…………………………………………………………………………………..11 3-1. Connecting to I/O
terminal blocks …………………………………………… 11 3-1-1. Connecting to sensor I/O terminal
block ……………………… 11 3-1-2. Connecting to control I/O terminal block……………………..
11 3-2. Connecting strain gage sensor………………………………………………….. 12 3-2-1. About
sensor connection terminals ……………………………….. 12 3-2-2. About remote sense
……………………………………………………………… 12 3-2-3. About bridge voltage (excitation voltage)
…………………… 12 3-3. Connecting to power input terminals, FG terminal, and
protective ground terminal…………………………………………………………. 13 3-3-1. AC power
supply…………………………………………………………………….. 13 3-3-2. DC power supply
……………………………………………………………………. 13 3-4. Connecting to control I/O terminals
……………………………………….. 14 3-4-1. Connecting external input………………………………………………….. 14
3-4-2. Connecting decision outputs ……………………………………………. 14 3-4-3. Connecting
voltage output (V-OUT) ………………………………. 14
4. Setup Mode……………………………………………………………………………………15 4-1. Changing set
values……………………………………………………………………….. 15 4-2. List of setup items and
settings…………………………………………………. 16 4-2-1. F1 setup
screens……………………………………………………………………… 16 4-2-2. F2 setup
screens……………………………………………………………………… 17 4-2-3. F3 setup
screens……………………………………………………………………… 18 4-2-4. F4 setup
screens……………………………………………………………………… 19
5. Error Messages ……………………………………………………………………………..20
6. Calibration………………………………………………………………………………………21 6-1. Common calibration
procedure ……………………………………………….. 22 6-1-1. Enabling and disabling CAL
lock………………………………………. 22 6-1-2. Remote Sense/TEDS ……………………………………………………………… 23
6-1-3. Bridge voltage………………………………………………………………………….. 24 6-1-4. Minimum
scale………………………………………………………………………… 24 6-1-5. Display Refresh
Rate………………………………………………………………. 25 6-2. Equivalent input
calibration………………………………………………………… 25 6-3. Actual load
calibration…………………………………………………………………… 27 6-4. TEDS
calibration……………………………………………………………………………….. 28 6-4-1. Displayed digits of
rated capacity for TEDS calibration..29 6-5. Zero-point
calibration……………………………………………………………………. 29
4
7. Setting Functions ………………………………………………………………………..30 7-1. Setting HI, LO, HH,
and LL values………………………………………………. 30 7-1-1. Statuses of decision output LEDs
…………………………………….. 30 7-1-2. HI value……………………………………………………………………………………….. 30
7-1-3. LO value……………………………………………………………………………………… 31 7-1-4. HH
value…………………………………………………………………………………….. 31 7-1-5. LL value
………………………………………………………………………………………. 32 7-1-6. Enable HH and LL
…………………………………………………………………… 32 7-2.
Hysteresis……………………………………………………………………………………………. 33 7-3. Comparative Output
Pattern………………………………………………………. 34 7-3-1. Standard
output……………………………………………………………………… 34 7-3-2. Area output
………………………………………………………………………………. 35 7-4. Comparison Mode
………………………………………………………………………….. 36 7-5. Near Zero
……………………………………………………………………………………………. 36 7-6. Motion
detection…………………………………………………………………………….. 37 7-6-1. Motion Detection
Time………………………………………………………… 37 7-6-2. Motion Detection
Bandwidth……………………………………………. 37 7-7. Zero tracking
…………………………………………………………………………………….. 38 7-7-1. Zero Tracking Time
Interval………………………………………………… 38 7-7-2. Zero Tracking Bandwidth
……………………………………………………. 38 7-8. Digital Filter
……………………………………………………………………………………….. 39 7-9. Analog
Filter………………………………………………………………………………………. 39 7-10. SET Lock
……………………………………………………………………………………………. 40 7-11. Static Strain Mode
………………………………………………………………………… 40 7-12. Decision-Output
Check………………………………………………………………. 41 7-13. Control-Input Check
……………………………………………………………………. 41 7-14. Beep
…………………………………………………………………………………………………… 42 7-15. Select Data Output
………………………………………………………………………. 42
8. Hold function ………………………………………………………………………………..43 8-1. Hold
Mode…………………………………………………………………………………………. 43 8-2. External Hold
Mode………………………………………………………………………… 43 8-3. Enable CLEAR
Signal………………………………………………………………………. 44 8-4. Enable Block Setting
………………………………………………………………………. 44 8-5. Enable JUDGE Signal
……………………………………………………………………… 45 8-6. Sample
hold………………………………………………………………………………………. 46 8-7. Peak hold
……………………………………………………………………………………………. 47 8-7-1. Operation without block setting
……………………………………… 47 8-7-2. Operation with block setting …………………………………………….. 47
8-8. Bottom hold………………………………………………………………………………………. 48 8-8-1. Operation without
block setting ……………………………………… 48 8-8-2. Operation with block setting
…………………………………………….. 48 8-9. Peak-to-Peak hold …………………………………………………………………………… 49
8-9-1. Operation without block setting ……………………………………… 49 8-9-2. Operation
with block setting …………………………………………….. 49
9. Digital Zero Function …………………………………………………………………50 9-1. Digital
Zero………………………………………………………………………………………… 50 9-2. Digital Zero Limit
…………………………………………………………………………….. 50 9-3. Clear Digital Zero
…………………………………………………………………………….. 51 9-4. Enable Digital Zero
…………………………………………………………………………. 51 9-5. Digital Offset
……………………………………………………………………………………… 52
10. BCD Data Output (TD-260T AC(BCD) / TD-260T DC(BCD))
………………………………………………………………………………………53
10-1. Panel ………………………………………………………………………………………………….. 53 10-2. Changing
logic………………………………………………………………………………. 54 10-3. Equivalent
circuit…………………………………………………………………………… 54 10-4. Signal
timing…………………………………………………………………………………… 55 10-5. BCD Data Refresh
Rate………………………………………………………………… 56 10-6. BCD Output
Check……………………………………………………………………….. 56
11. RS-232C Interface (TD-260T AC(232C) / TD-260T
DC(232C))……………………………………………………………………………………..57
11-1. Panel ………………………………………………………………………………………………….. 57 11-2. Standard
…………………………………………………………………………………………… 57 11-3. Cable wiring example
…………………………………………………………………. 58 11-4. RS-232C Interface
Setting………………………………………………………….. 58 11-5. List of
Commands…………………………………………………………………………. 59
11-5-1. Read commands ………………………………………………………………….. 59 11-5-2. Set
commands………………………………………………………………………. 59 11-5-3. Set-value retrieving set value
………………………………………….. 60 11-5-4. Control……………………………………………………………………………………… 60
11-6. Transmission Mode………………………………………………………………………. 61 11-6-1. Transmission
mode 0 …………………………………………………………. 61 11-6-2. Transmission mode 1
…………………………………………………………. 61 11-7. Format for transmission mode
0…………………………………………….. 61 11-7-1. Read displayed-reading command (RA) ……………………..
61 11-7-2. Read status command (RS)………………………………………………. 61 11-8. Set
commands……………………………………………………………………………….. 62 11-9. Set-value retrieval
command…………………………………………………… 64 11-10. Control
Commands…………………………………………………………………… 65 11-11. Format for transmission mode
1………………………………………….. 65
12. D/A Converter (TD-260T AC(D/A) / TD-260T DC(D/A))…..66 12-1. Panel
………………………………………………………………………………………………….. 66 12-2.
Connections……………………………………………………………………………………. 66 12-2-1. Connecting to
screwless terminal block…………………….. 66 12-3. D/A
Zero…………………………………………………………………………………………… 67 12-4. D/A Full Scale
…………………………………………………………………………………. 67 12-5. D/A output
error……………………………………………………………………………. 68
13. NDIS Panel (PN-260ND) ………………………………………………………….69
14. About TEDS …………………………………………………………………………………70 14-1. Save TEDS Data
……………………………………………………………………………… 70 14-2. Restore TEDS
Data………………………………………………………………………… 71
Table of Contents
15. Available functions at turn-on……………………………………………..72 15-1. Performing
self-check …………………………………………………………………. 72 15-2. Initializing
settings………………………………………………………………………… 72
16. Specifications ……………………………………………………………………………..73 17. Outline
drawings……………………………………………………………………….74 18. Block
Diagram…………………………………………………………………………….75
5
Controls and Functions
1-1. Front panel
1 Numeric display
The TD-260T shows a continuous reading or a set value on the display.
If the sensor input is erroneous or a reading cannot be displayed, the corresponding overflow message is displayed as shown below.
-LOAD: A /D conver ter over f low in the negative direction
LOAD: A/D converter overflow in the positive direction
-Full: Display overflow, lower than -19999
Full:
Display overflow, higher than 99999
Decision output LEDs
The following LEDs show statuses of the TD-260T.
HH/HI
The HH/HI LED shows the operation of the decision output for the High-High
limit (“HH”) or High limit (“HI”), meaning that the LED displays one of the
following solid colors when a displayed reading is higher than the
corresponding set value.
Red: shows that the HH decision output is ON. HH value < displayed reading
Orange: shows that the Hi decision output is ON. HI value < displayed reading
Note that the light displays solid red when the setting for Enable HH and LL
(F1-5) is OFF OK
The OK LED shows the operation of the OK decision output, meaning that the LED
displays solid green when a displayed reading is both higher than the LO value
and lower than the HI value.
Green: shows that the OK decision output is ON. LO value displayed reading HI
value
LL/LO
The LL/LO LED shows the operation of the decision output for the Low-Low limit
(“LL”) or Low limit (“LO”), meaning that the LED displays one of the following
solid colors when a displayed reading is lower than the corresponding set
value.
Red: shows that the LL decision output is ON. Displayed reading < LL value
Orange: shows that the LO decision output is ON. Displayed reading < LO value
Note that this LED displays solid red when the setting for Enable HH and LL
(F1-5) is OFF.
NOTE
The operations of decision outputs vary depending on the settings for Comparative Output Pattern (F1-15) and Hysteresis (F1-6). In the above examples, “standard” for Comparative Output Pattern and “0” for Hysteresis are selected.
Decision LED (JUDGE)
The LED lights when the decision output is enabled.
With Enable JUDGE Signal (F3-8) set to ON; the LED lights if the JUDGE signal
from A6 is ON, or the LED remains off if the JUDGE signal from A6 is OFF.
With Enable JUDGE Signal (F3-8) set to OFF, the LED lights and remains on.
4 Calibration keys
GAIN:
To perform a calibration by applying an actual load, press this key.
TEDS:
To perform a calibration when a TEDS sensor is connected, press this key.
CAL:
To perform a calibration by entering the rated output values, press this key.
5 Lock LEDs
CAL:
Lights while changing calibration values is not allowed.
SET:
Lights while changing set values is not allowed.
6 Hold-mode LEDs
Each LED displays solid green, showing that the corresponding mode is set, or it displays solid red or orange, showing that the TD-260T is operating in the corresponding mode.
PEAK:
Lights when the TD-260T is operating in the Peak-hold mode.
BOTTOM: Lights when the TD-260T is operating in the Bottom-hold mode.
P-P:
Lights when the TD-260T is operating in the Peak-to-Peak hold mode.
SAMPLE: Lights when the TD-260T is operating in the Sample hold mode.
7 HOLD key
Pressing this key starts the hold operation. To stop the hold operation, press the key again.
6
8 Set keys
FNC:
Pressing this key while a reading is displayed changes the display to the setup screen. Each time you press the key while the setup screen is displayed, the display cycles through the function numbers. Pressing the key while the setup screen is displayed with a digit of the set value blinking changes the sign setting or the decimal point position.
ESC/CLEAR: Pressing this key while the setup screen is displayed cancels the setting, and then displays a reading again. Pressing the key while the TD-260T is in a hold mode clears the hold value.
ENTER:
Pressing this key saves the item or value you have set. Pressing and holding the key while a reading is displayed disables or enables all the keys on the TD-260T. While all the keys are disabled, the key input is disabled, except the ENTER key that can enable all the keys again by being pressed and held.
5b:
Pressing these keys selects an item you want to
set or changes a set value.
gt:
Pressing these keys selects a digit you want to make changes to.
9 ZERO key
The operation of this key varies depending on the setting for CAL Lock (F3-5).
Pressing this key performs zero-point calibration while the calibration lock
setting is OFF. Pressing this key while the calibration lock setting is ON
forces a reading on the display to be zero . (Digital Zero function)
1. Controls and Functions
7
1. Controls and Functions
1-2. Rear panel
AC power model
1
2 3
7 DC power input terminals Connects DC power. The voltage range is 10 to 28 V
DC.
8 Frame ground (FG) FG terminal for DC power supply. (The protective ground
terminal and the FG terminal are electrically connected.)
1-2-1. Signal I/O terminal block 4
1
5 7
6 1 TEDS / +SENS
2 GND / SENS
DC power model
3 +EXC(A) 4 SIG(B)
1 Protective ground terminal screw To prevent an electric shock hazard and
electrostatic interference, be sure to connect the ground wire to this screw.
2 AC power input terminals Connects AC power. The voltage range is 100 to 240
V AC.
3 Frame ground (FG) FG terminal for AC power supply. (The protective ground
terminal and the FG terminal are electrically connected.)
4 Signal I/O terminal block Connects a strain gage sensor or TEDS sensor
5 Option space The optional NDIS panel (PN-260ND) can be installed in this
area. (Because the option-installed model has an option board pre-installed in
this area, the optional panel cannot be installed in the option-installed
model.) For detailed information about the option-installed model, see Chapter
16, “Specifications.”
6 Control I/O terminal block Connects control signals, decision outputs, etc.
5 EXC(C)
6 +SIG(D)
7 SHIELD
Pin assignment
1 2 3 4 5 6 7
Signal name
TEDS/+SENS GND/-SENS +EXC (A)
-SIG (B) -EXC (C) +SIG (D) Shield
Wiring color
Orange Green Red Black Blue White Yellow
< The terminals 1 and 2 on the signal I/O terminal block are used as either
remote sense terminals or TEDS sensor data terminals. Before connecting a
sensor, you need to set Remote Sense/TEDS (F2-6) to match the function you are
going to use.
The initial setting is 0.Aut: TEDS auto-calibration (disable remote sense).
< The wiring colors shown are those of TEAC strain gage sensors.
8
1-2-2. Control I/O terminal block
A1
B1
1. Controls and Functions
A9
B9
1-2-3. Control I/O signals
Pin assignment
A1
A2 A3 A4 A5
Signal name V-OUT
NC CLEAR
Description
Voltage output + An output voltage of about 2 V for an input of 1 mV/V is
produced, which is proportionate to the sensor output. V-OUT is not isolated
from the internal circuit. A negative output voltage for this output is
generated from B1. Make sure that the cable you use is as short as possible.
No connections. Do not use for relay connections.
To clear a hold mode other than the sample hold mode, set this signal to ON.
To disable this signal, set Enable CLEAR Signal (F2-8) to OFF.
Pin assignment
B1
B2 B3 B4 B5
Signal name
COM
LL LO COM HH
Description Voltage output Negative output voltage for A1
LL decision output (a contact) LO decision output (a contact) Common terminal
for B2 and B3 HH decision output (a contact)
This is the control signal for decision
outputs. Only while the control signal
A6
JUDGE is ON, decision outputs are enabled.
B6
HI
To disable the signal, set Enable JUDGE
Signal (F3-8) to OFF.
This is the start signal for a hold mode. While the signal is ON, a hold
A7
HOLD operation is performed.
B7
COM
To disable the signal, set Hold Mode
(F2-1) to 0.OFF (disable hold).
This signal digitally sets a reading to zero. The function is enabled only
A8
D/Z
while both CAL Lock (F3-5) and Enable Digital Zero (F3-1) are ON. When the
B8
OK
TD-260T is turned off, the digitally set zero is cleared.
This is the common terminal for A5,
A9
COM
A6, A7, and A8. A photocoupler is used to isolate signals from A5 to A9 and
B9
COM
the TD-260T internal circuit.
Connector used: Weidmüller S2L-SMT-3.50/1890LF3.2 SN BK BX or its equivalent Applicable plug: Weidmüller B2L 3.50/18/180F SN BK BX or its equivalent
HI decision output (a contact) Common terminal for B5 and B6 OK decision output (a contact) Common terminal for B8
9
2. Hardware Installation
Follow these steps to install the TD-260T inside a control panel. 1. Cut out
the panel according to the dimensions in the
panel cut-out drawing.
Panel cut-out dimensions Recommended panel thickness: 1.6 – 3.0 mm
4. Put the guide rails that were removed in Step 2, back in place from the
rear side, and then secure them with the two screws.
(Recommended tightening torque: 0.6 ± 0.1 N m [6 ± 1 kgf cm])
+0.8 92 0
92 +0.8 0
2. Remove the two screws, and then remove the guide rails on the right and
left sides. < DO NOT use screws other than those installed in the TD-260T.
Guide rail
screws Guide rail
3. Push the TD-260T into the panel cut-out area from the front side of the
panel.
10
3. Connections
3-1. Connecting to I/O terminal blocks
3-1-1. Connecting to sensor I/O terminal block The TD-260T uses a 2-piece
sensor input/output terminal block. To connect wires to the provided connector
BCZ 3.81/07/180F SN BK BX or its equivalent and then the connector to the
terminal block, use the provided flathead micro screwdriver. If you want to
use a screwdriver other than the provided screwdriver, be sure to use a
flathead screwdriver with teeth not exceeding either 2.5 mm in width or 0.4 mm
in thickness.
1. Strip off about 5 mm of the insulation from an end of each wire you
connect, and then gently twist each bare end. Use wires 0.2 to 1.5 mm2 in
cross-sectional area (28 to 16 AWG).
3-1-2. Connecting to control I/O terminal block
The TD-260T uses a 2-piece control input/output terminal block. To connect
wires to the provided connector B2L 3.50/18/180F SN BK BX or its equivalent
and then the connector to the terminal block, use the provided flathead micro
screwdriver. If you want to use a screwdriver other than the provided
screwdriver, be sure to use a flathead screwdriver with teeth not exceeding
either 2.5 mm in width or 0.4 mm in thickness.
1. Strip off about 7 to 8 mm of the insulation from an end of each wire you
connect, and then gently twist each bare end.
Use wires 0.13 to 1.0 mm2 in cross-sectional area (28 to 18 AWG).
Approx. 5 mm
loosetnighten
Wire insertion hole (round)
Screwdriver insertion hole
(square)
Approx. 7-8 mm
2. While keeping each bare end twisted, insert it into the corresponding
hole.
3. Secure each wire end to the hole by using the flathead screwdriver.
4. Lightly pull each wire and make sure that they are securely clamped.
5. Plug the wired connector into the terminal block in the TD-260T, and then
secure the connector plug to the terminal block with screws.
2. Insert the screwdriver into the square holes next to the wire insertion
holes to enlarge the clamp areas.
3. While keeping each bare end twisted, insert it into the corresponding
hole, and then secure it to the hole.
4. Lightly pull each wire and make sure that they are securely clamped.
5. Plug the wired connector into the terminal block in the TD-260T, and then
secure the connector plug to the terminal block with the screws.
11
3. Connections
3-2. Connecting strain gage sensor
3-2-1. About sensor connection terminals Sensor connection terminals (1, 2)
can be used as either remote sensing terminals or TEDS sensor data terminals.
Before connecting a sensor, you need to set Remote Sense/TEDS (F2-6) to match
the function you are going to use. The initial setting is 0.Aut: TEDS auto-
calibration (disable remote sense). For detailed information, see the section
6-1-2. “Remote Sense/TEDS.”
CAUTION
If you want to use a sensor with a 6-wire connection (for remote sensing), set
Remote Sense/TEDS (F2-6) to 2.rSE (disable TEDS) before connecting the sensor.
3-2-2. About remote sense The TD-260T uses a 6-wire connection (for remote
sensing) when calibrated at the factory. Be aware that, if you connect a
sensor by using 4 wires, and then perform an equivalent input calibration, an
error may occur in a reading, resulting from the bridge voltage drop that is
proportionate to the length of the cable extended. A 6-wire connection (for
remote sensing) is excellent for compensating for a voltage drop caused by the
extension of the cable length or voltage change due to changes in temperature.
For applications of the TD-260T that may be exposed to varying temperatures
including the use of the device in an outdoor installation system, or
applications of the TD-260T that require overall accuracy, we recommend the
6-wire connection that includes remote sensing.
3-2-3. About bridge voltage (excitation voltage) When you set the bridge
voltage, make sure that the output voltage from the sensor is relatively high,
but that it does not exceed the maximum excitation voltage specified on the
sensor test report at the same time. Setting a voltage that exceeds the
sensor’s maximum excitation voltage may cause the sensor to malfunction.
CAUTION
< If the sensor’s maximum excitation voltage is lower than 10 V, be sure to
set Bridge Voltage (F2-5) to 2.5 before connecting the sensor.
< For TEDS calibration, if the bridge voltage setting is higher than the
maximum excitation voltage recorded in the TEDS memory when TEDS data is
loaded, the bridge voltage setting is automatically changed to a setting equal
to or lower than the maximum excitation voltage..
NOTE
< The initial setting is 2.5 V.
TEDS sensor & 4-wire connection
Strain Gage Sensor
TD-260T
Teds Data Memory
1 (TEDS/+SENS) 2 (GND/-SENS)
A
3 (+EXC)
D
B
4 (-SIG)
C
5 (-EXC)
6 (+SIG)
7 (SHIELD)
If you do not use the TEDS function, you do not need to connect any device to the terminal 1 or 2.
6-wire connection
Strain Gage Sensor
TD-260T
1 (TEDS/+SENS) 2 (GND/-SENS)
A
3 (+EXC)
D
B
4 (-SIG)
C
5 (-EXC)
6 (+SIG)
7 (SHIELD)
12
3-3. Connecting to power input terminals, FG
terminal, and protective ground terminal
Check the power supply and determine that the type of the power supply you
have received conforms to the specifications that were provided at the time of
order placement, and then connect the AC or DC power supply.
3-3-1. AC power supply The AC power input voltage is 100240 V AC. For power
supply connection, use power wires that are 0.326 to 2.081 mm2 in cross-
sectional area (22 to 14 AWG). Connect those wires to the terminal block by
using crimp connectors (for M3, 6 mm or less in width). For grounding, use a
ground wire (green/yellow) that is 3.309 mm2 in cross-sectional area (12 AWG).
Connect the wire to the protective ground terminal by using a crimp connector
(for M4).
3. Connections
3-3-2. DC power supply The DC power input voltage is 1028 V DC. For power
supply connection, use power wires that are 0.326 to 2.081 mm2 in cross-
sectional area (22 to 14 AWG). Connect those wires to the terminal block by
using crimp connectors (for M3, 6 mm or less in width). For grounding, use a
ground wire (green/yellow) that is 3.309 mm2 in cross-sectional area (12 AWG).
Connect the wire to the protective ground terminal by using a crimp connector
(for M4).
Protective(Ggrroeuenn/dyewlliorew)
Black White Green/yellow
< Power cord wire colors (for the provided AC power cord) L: Black N: White
FG: Green/yellow
< The provided AC power cord is only intended for use within Japan, rated for
10 A at 125 V AC. If you want to use the TD-260T at a voltage exceeding the
specified voltage or overseas, be sure to use another AC power cord that is
more appropriate for your operating
< The FG terminal is internally connected to the protective ground.
< The frequency is 5060 Hz.
CAUTION
< For the TD-260T to warm up, wait for more than 5 minutes after it is turned
on.
< Be sure to install the cover over the power input terminal block.
protective(Ggrroeuenn/dyewlliorew) + FG
< The FG terminal is internally connected to the protective ground.
< The current draw at 12 V DC input is about 0.8 A. CAUTION < If you want to
use the TD-260T as a CE compliant unit, the
length of the power cord to be used for the TD-260T must be 3 m or less. < For
the TD-260T to warm up, wait for more than 5 minutes after it is turned on. <
Be sure to install the cover over the power input terminal block.
13
3. Connections
3-4. Connecting to control I/O terminals
3-4-1. Connecting external input An input signal is generated from an external
input circuit when a control terminal and COM terminal on the external input
circuit are short-circuited and then opened. Terminals can be short-circuited
through contact or noncontact technologies (transistor, TTL open collector).
TD-260T Vcc +12V
Approx. 10mA
PUSH Switch
A9 COM
When contact signal is used:
When transistor is used:
When TTL open collector is used: IN
ON when IN is high
< The current draw is about 10 mA when the external contact is in ON state.
Use a contact device with a withstand voltage of 20 V or more and about 20 mA
capacity in ON state.
< The common terminal for control input/output terminals A5 (CLEAR), A6
(JUDGE), A7 (HOLD), and A8 (D/Z) is A9 (COM).
< Do not apply external voltages.
3-4-2. Connecting decision outputs
TD-260T
Diode
N/O Load
COM
Spark Killer
DC Power Supply
< Relays are used for decision outputs. The relay contact rating is 1 A and 30
V DC. Be sure to use relays within the specified current and voltage ratings.
< Install a spark killer for noise reduction as close to the load circuit,
such as relay coil, you use as possible.
< Use a rectifying diode.
< The common terminal for control output terminals B5 (HH) and B6 (HI) is B7
(COM).
< The common terminal for control output terminals B2 (LL) and B3 (LO) is B4
(COM).
3-4-3. Connecting voltage output (V-OUT)
The output voltage is proportionate to the sensor input and approximately 2 V
for a sensor input of 1 mV/V. The above output voltage is produced after the
zero point of the sensor has been calibrated during the calibration procedure.
With the digital zero function (D/Z), the output voltage cannot be reduced to
zero.
TD-260T
Voltage Output
A1
B1
shield
External Device
+ Load resistance
2k or more GND
< This output is not isolated from the internal circuit. Be sure to install an
external device as close to this output as
possible by using 2-core shielded cables or equivalent.
14
Setup Mode
4-1. Changing set values
Each setup screen can be identified with its setup screen number consisting of
a function number and a suffix number, F1-1, for example. To change a set
value, you need to put the TD-260T into the corresponding setup mode by
displaying the appropriate setup screen number and then pressing ENTER. 1. To
Enter the setup screen number, press FNC.
The display shows “F1-1.”
2. To change the function number, press FNC. Each time you press FNC, the
display cycles through function numbers such as F2 and F3.
4. To switch to the setup mode, press ENTER.
For detailed information about operating each setup screen, see the next
chapters. 5. To change a setting in the setup mode, do the following. If the
displayed value includes a blinking digit, you can change it by pressing the
arrow buttons; if you press 5 or b, the blinking digit increases or decreases.
To move to the previous or next digit, press g or t.
To move to the previous or next digit, press:
To change the digit, press:
3. To change the suffix number, press 5 or b.
Each time you press 5, the suffix number is increased by one, or each time you
press b, the suffix number is decreased by one.
If the entire value or text on the display blinks, you can select another
setting by pressing 5 or b.
or
suffix number + 1 suffix number + 1
6. To save the selection, press ENTER, and then to exit the setup mode, press ESC. A reading is displayed.
NOTE
If you press ENTER while the setup screen number is displayed, the display
switches to the setup mode screen, shown in Step 5 of this section, before the
setup item name and set value appear. If you do not enter any key for 0.5
seconds, the setup item name appears, followed by the set value.
To save, press:
CAUTION The display also shows a reading again if you do not press any key for
one minute.
15
4. Setup Mode
4-2. List of setup items and settings
4-2-1. F1 setup screens
Setup screen number F1-1
Setup item name Displayed text
HI value
Initial setting
Setting range
150.00 -19999 to 9999
x: Applicable
Page
SET Lock
CAL Lock
Unit
30
x
LO value F1-2
050.00 -19999 to 99999
31
x
HH value F1-3
999.99 -19999 to 99999
31
x
LL value F1-4
-199.99 -19999 to 99999
32
x
Enable HH and LL F1-5
OFF OFF, ON
32
x
Hysteresis F1-6
000.00 00000 to 19999
34
x
Comparison Mode F1-7
Near Zero F1-8
0.ALL: Always comp are to gener ate decision outputs
1.Stb: Compare to generate decision output when stabilized
2.no0: Alw ays comp are to gener ate
decision outputs except near zero
0.ALL 3.n0S: Compare to generate decision 36
x
outputs when stabilized except
near zero
4.HLd: Compare to generate decision outputs when in hold mode
5.OFF: Do not compare
001.00 00000 to 09999
36
x
Motion Detection Time F1-9
1.5 0.0 to 9.9
37
x
Sec
Motion Detection Bandwidth
F1-10
000.05 00000 to 00999
37
x
Zero Tracking Time Interval F1-11
0.0 0.0 to 9.9
38
x
Sec
Zero Tracking Bandwidth F1-12
000.00 00000 to 00999
38
x
Digital Zero Limit F1-13
99999 00000 to 99999
50
x
Clear Digital Zero F1-14
0 CLr
51
x
F1-15
Comparative Output Pattern
0.nor
0.nor: Standard output 1.ArE: Area output
35
x
16
4-2-2. F2 setup screens
Setup screen number
Setup item name Displayed text
Hold Mode F2-1
Digital Offset F2-2
Digital Filter F2-3
Analog Filter F2-4
Bridge Voltage F2-5
Remote Sense/TEDS F2-6
External Hold Mode F2-7
Enable CLEAR Signal F2-8
4. Setup Mode
Initial setting
1.SPL
Setting range
0.OFF:Disable hold 1.SPL: Sample hold 2.PEk: Peak hold 3.bot: Bottom hold
4.P2P: Peak-to-Peak hold
Page
x: Applicable
SET Lock
CAL Lock
Unit
43
x
00000 -19999 to 19999
52
x
OFF
Moving average: OFF, 4, 8, 16, 32, 64, 128, 256
39
x
of times
100 3, 10, 30, 100, 300
39
x
Hz
2.5 10, 2.5
24
0.Aut
0.Aut: TEDS auto-calibration 1.tEd: Enable TEDS/disable remote sense 23 2.rSE: Enable remote sense/disable TEDS
0.LEv 0.LEv: Level, 1.PLS: Pulse
43
x
x
V
x
ON OFF, ON
44
x
17
4. Setup Mode
4-2-3. F3 setup screens
Setup screen number F3-1
Setup item name Displayed text
Enable Digital Zero
Enable Block Setting F3-2
Minimum Scale F3-3
Display Refresh Rate F3-4
CAL Lock F3-5
SET Lock F3-6
Static Strain Mode F3-7
Enable JUDGE Signal F3-8
Initial setting
Setting range
ON OFF, ON
OFF OFF, ON
1
1, 2, 5, 10
4
4, 6, 10, 20
OFF OFF, ON
OFF OFF, ON
OFF OFF, ON
OFF OFF, ON
Page
x: Applicable
SET Lock
CAL Lock
Unit
51
x
44
x
24
x
25
x
of times
22
40
40
x
45
x
18
4-2-4. F4 setup screens
Setup screen number F4-1
Setup item name Displayed text
BCD Data Refresh Rate
BCD Output Check F4-2
RS-232C Interface Setting F4-3
DA Zero F4-4
DA Full Scale F4-5
Save TEDS Data F4-6
Restore TEDS Data F4-7
Decision-Output Check F4-8
Control-Input Check F4-9
Beep F4-10
Select Data Output F4-11
4. Setup Mode
Initial setting
Setting range
100 1, 2, 5, 10, 20, 50, 100
C bcd
13100
1 3 1 0 0
2 2 2 2 2
a bc de
a Transmission mode
0: Command transmission mode
1: Continuous transmission mode
b Baud rate
0: 1200bps 1: 2400bps
2: 4800bps 3: 9600bps
4: 19200bps
c Data bits
0: 7bits
1: 8bits
d Parity bits
0: None
1: Odd
2: Even
e Stop bits
0: 1bit
1: 2bits
000.00 -19999 to 99999
Page 56
x: Applicable
SET Lock
CAL Lock
Unit
of
x
times
/sec
56
58
x
67
x
100.00 -19999 to 99999
67
x
2.000
70
x
00000
71
x
C Out C IN
Turn on the decision outputs for HH, HI, OK, LO, and LL, one by one
41
Show the ON/OFF state of CLEAR, HOLD, JUDGE, and DZ on the LED display
41
ON OFF, ON
42
x
0.dISP: Reference the display; hold value
0.dSP
is sent in hold mode 1.InP: Reference the input; input value is
42
x
sent in hold mode
19
Error Messages
ADC overflow in the positive direction
ADC overflow in the negative direction
Display overflow in the positive direction (higher than 99999) Display
overflow in the negative direction (lower than -19999) Input higher than 3.2
mV/V
Input lower than -3.2 mV/V
· Exceeding zero-point adjustable range · Zero-point adjustment not complete
within the specified time period Sensor input during calibration higher than
3.2 mV/V Rated capacity set to 0000
Sensor input lower than adjustable range (aprrox. 0.28 mV/V) Negative sensor
output during actual load calibration TEDS-supported sensor not connected
Wrong password for TEDS save operation
Invalid set value
Rated output outside setting range (lower than 0.28 mV/V or higher than 3.2
mV/V) Sensor mismatch: different from the sensor used for TEDS calibration
BV10V unsupported sensor used. BV autoset to 2.5 V Invalid operation
Digital zero limit error
Error other than those shown above
DA output higher than 10.9 V
DA output lower than -10.9 V
C-000 check error (IRAM)
C-001 check error (SRAM)
C-002 check error (EEPROM)
C-003 check error (TEDS)
Not calibrated at the factory
DA not adjusted at the factory
Adjusting zero-point now
· Accessing TEDS now · Calibrating TEDS now
20
Calibration shown in this document is the process of adjusting parameters to
determine a reading to be displayed on the TD-260T when a strain gage sensor
is connected. The following three calibration methods are available in the TD-
260T.
1. Equivalent input calibration
It is a simple calibration method that requires not the use of an actual load,
but only the key entry of the rated output (mV/V) and rated capacity (value
you want to display) of a strain gage sensor. With this method, you can easily
perform a calibration if an actual load cannot be applied. Examples:
In case of load, enter: 100 kN for the rated capacity and 2.001 mV/V for the
rated output
In case of pressure, enter: 10.00 MPa for the rated capacity and 2.002 mV/V
for the rated output
In case of torque, enter: 15.00 N m for the rated capacity and 2.502 mV/V for
the rated output.
As shown above, enter values specified on a test report so that the TD-260T
automatically determines and displays a gain.
2. Actual load calibration
It is a calibration method that requires you apply an actual load to a strain
gage sensor and then measure it. For accurate calibration, which minimizes
errors, you need to apply a load that is close to the sensor’s maximum
measurable value.
3. TEDS calibration
It is a calibration method that uses data on the rated output (mV/V) and rated
capacity of a strain gage sensor both stored in the TEDS memory. The TEDS
memory is available in 1-kilobit and 4-kilobit sizes. Note that the TD-260T
supports the 4-kilobit memory only.
Calibration
Checking a sensor before calibration Calibration cannot be done if the TD-260T
displays an unstable reading or error when you connect a sensor to the TD-260T
and then turn it on. In that case, set the TD260T to the static strain mode by
performing Static Strain Mode (F3-7), and then check a reading on the display
again. In this mode, the output from the sensor is displayed in microstrains;
the displayed output is 0 to ±6400.0 for the input of 0 to ±3.2 mV/V.
CAUTION
The terminals (1 and 2) on the signal input/output terminal block can be used
as either remote sense terminals or the TEDS sensor data terminals. Before
connecting a sensor, you need to change the setting for Remote Sense/ TEDS
(F2-6) to match the function you are going to use. The initial setting is
0.Aut: TEDS auto-calibration (disable remote sense). For detailed information,
see the section 6-12. “Remote Sense/TEDS.”
The test report included with a strain gage sensor provides the following
information. Rated Capacity: load, atomospheric pressure, etc. (Unit: kN, Mpa,
etc.) Rated output: voltage (Unit: mV/V) Linearity: %R.O Hysteresis: %R.O Safe
excitation voltage (maximum): V (bridge voltage) Input Terminal Resistance:
Output Terminal Resistance: Zero Balance: x10-6 strain (microstrain)
< Data required for equivalent input calibration are rated capacity and rated
output.
< TEDS sensor has the above data saved in its internal memory.
21
6. Calibration
6-1. Common calibration procedure
The three calibration methods available in the TD-260T, equivalent input
calibration, actual load calibration, and TEDS calibration, follow the same
steps before and after calibration. Below is the flow chart of the calibration
procedure.
Disable CAL lock (F3-5)
6-1-1. Enabling and disabling CAL lock Normally the TD-260T is used with the
calibration lock set to ON. Be sure to set the calibration lock to OFF before
calibration, and then set it to ON again after calibration. 1. Enter the setup
screen number F3-5.
Press 3 times
Press 4 times
Set remote sense/TEDS (F2-6)
2. To select ON or OFF, press 5 or b.
Set bridge voltage (F2-5) Set minimum scale (F3-3)
or
3. To save the selection, press ENTER, and then to exit the setup mode, press
ESC.
Set display refresh rate (F3-4)
To save, press:
Equivalent input calibration, actual load calibration, or TEDS calibration
Enable CAL lock (F3-5)
Shown inside the brackets ( ) are setup screen numbers. This section provides
information about the steps to be performed before and after calibration,
followed by equivalent input calibration, actual load calibration, and TEDS
calibration.
CAUTION
< The calibration lock setting changes the function of the ZERO key. If you
press ZERO with the calibration lock set to ON, the displayed reading is
digitally set to zero. With the calibration lock set to OFF, zero-point
calibration is performed.
< To prevent the calibration values from being accidentally changed, be sure
to set the calibration lock to ON.
NOTE
For detailed information about setup items whose settings cannot be changed
with the calibration lock set to ON, see the section 4-2. “List of setup items
and settings.”
22
6-1-2. Remote Sense/TEDS The signal input terminals (1 and 2) can be used as
either remote sense terminals or TEDS sensor data terminals. You need to set
the terminals to one of the available settings before connecting a sensor to
the TD-260T.
1. Enter the setup screen number F2-6.
Press twice
6. Calibration
NOTE
< The initial setting is 0.Aut: TEDS auto-calibration (disable remote sense).
< If you use neither remote sense nor TEDS sensor, be sure to select 0.Aut:
TEDS auto-calibration.
Press 3 times
2. To change the setting, press 5 or b.
or
The following three settings are available. 0.Aut: TEDS auto-calibration
(disable remote sense)
If the currently connected TEDS transducer is recognized as a transducer
different from the one used for last calibration, calibration is automatically
performed when you turn the TD-260T on. 1.tEd: Enable TEDS (disable remote
sense) You need to manually perform TEDS calibration. 2.rSE: Enable remote
sense (disable TEDS) You need to connect a 6-wire sensor. 3. To save the
selection, press ENTER, and then to exit the setup mode, press ESC.
To save, press:
CAUTION If you want to use a 6-wire connection (for remote sensing), be sure
to select 2.rSE to enable remote sense and disable TEDS before connecting a
sensor. Connecting a TEDS sensor with this setting may damage the sensor and
the TD-260T.
23
6. Calibration
6-1-3. Bridge voltage You need to set the bridge voltage, which is supplied to
a strain gage sensor. 1. Enter the setup screen number F2-5.
Press twice
6-1-4. Minimum scale You can set the minimum increment value that is used to
display a change in the digital reading. 1. Enter the setup screen number
F3-3.
Press 3 times
Press 4 times
Press twice
2. To enter 2.5 or 10, press 5 or b.
Making sure that the output voltage from the sensor is relatively high, but
that it does not exceed the maximum excitation voltage specified on the sensor
test report at the same time, choose between the two values.
2. To select 1, 2, 5, or 10, press 5 or b. or
or
3. To save the selection, press ENTER, and then to exit
the setup mode, press ESC.
3. To save the selection, press ENTER, and then to exit the setup mode, press ESC.
To save, press:
To save, press:
CAUTION < Setting a voltage that exceeds the sensor’s maximum
excitation voltage may cause the sensor to malfunction. NOTE < The initial
setting is 2.5 V. < For TEDS calibration, if the bridge voltage setting is
higher
than the maximum excitation voltage recorded in the TEDS memory when TEDS data
is loaded, the bridge voltage setting is changed to a setting equal to or
lower than the maximum excitation voltage.
NOTE The initial setting is 1.
24
6-1-5. Display Refresh Rate You can select the display refresh rate per second
for a reading. 1. Enter the setup screen number F3-4.
Press 3 times
Press 3 times
6. Calibration
6-2. Equivalent input calibration
You can determine the calibration values by setting the rated output and rated
capacity shown on the test report. Using this calibration method, you can
easily perform a calibration without applying an actual load.
Set rated output
2. All the digits of the displayed value blinks. To select 4, 6, 10, or 20, press 5 or b.
Set rated capacity
or
3. To save the selection, press ENTER, and then to exit the setup mode, press
ESC.
Zero-point calibration
1. Press CAL, and then press ENTER. The rated output (mV/V) appears on the
display with the least significant digit blinking.
NOTE The initial setting is 4.
To save, press:
2. To set the rated output; change the blinking digit of the value by
pressing 5 or b. To move to the previous or next digit, press g or t.
To move to the previous or next digit, press:
To change the digit, press:
3. To save the rated output you entered, press ENTER.
The rated capacity appears with the least significant digit blinking.
25
6. Calibration
4. To set the rated capacity; change the blinking digit of the value by
pressing 5 or b. To move to the previous or next digit, press g or t. To move
to the previous or next digit, press:
6. To perform zero-point calibration;
with no load on the sensor, press ZERO, and then press ENTER. The display
shows “0-CAL,” followed by zero for a reading, showing that the zero-point
calibration is completed.
To change the digit, press:
To change the decimal point position, blink the digit you want to the left of the decimal point by pressing g or t, and then press FNC.
If a calibration error is displayed, make the appropriate correction, and then repeat the calibration.
NOTE
The position of the decimal point you set in this step is used for readings on
the TD-260T. 5. To save the rated capacity you entered, press ENTER.
CAUTION
If you use the D/A option, the rated capacity is set as the D/ A full scale
value.
26
6-3. Actual load calibration
You can perform a calibration by applying an actual load to the sensor.
Zero-point calibration
6. Calibration
To change the decimal point position, blink the digit you want to the left of
the decimal point by pressing g or t, and then press FNC.
Load calibration
1. To perform zero-point calibration; with no load on the sensor, press ZERO,
and then press ENTER. The display shows “0-CAL,” followed by zero for a
reading, showing that the zero-point calibration is completed.
NOTE
The position of the decimal point you set in this step is used for readings on
the TD-260T. 4. To save the rated capacity you entered, press ENTER.
If a calibration error is displayed, make the appropriate correction, and then
repeat the calibration.
2. To perform a calibration by using a load;
apply an actual load to the sensor. Press GAIN, and then press ENTER. The
value for the rated capacity appears with the least significant digit
blinking.
If a calibration error is displayed, make the appropriate correction, and then
repeat the calibration.
CAUTION
If you use the D/A option, the rated capacity is set as the D/A full scale
value.
3. To set the rated capacity; Now you need to enter the value of the load
that is applied to the sensor. To do so, change the blinking digit of the
value by pressing 5 or b. To move to the previous or next digit, press g or t.
To move to the previous or next digit, press:
To change the digit, press:
27
6. Calibration
6-4. TEDS calibration
A TEDS sensor has calibration information including the rated output and
capacity stored in its internal memory. Using TEDS calibration, you can load
the calibration information and automatically register it as calibration
values.
Load TEDS data and modify loaded data
3. To set the rated capacity;
If you want to change the displayed data, you can change the blinking digit of
the value by pressing 5 or b. To move to the previous or next digit, press g
or t.
To move to the previous or next digit, press:
To change the digit, press:
Zero-point calibration
1. To load the TEDS data; press TEDS, and then press ENTER. The rated output
(mV/V) you loaded appears with the least significant digit blinking.
To change the decimal point position, blink the digit you want to the left of the decimal point by pressing g or t, and then press FNC.
If you want to change the displayed data, you can change the blinking digit of
the value by pressing 5 or b. To move to the previous or next digit, press g
or t.
CAUTION
If the bridge voltage setting is higher than the maximum excitation voltage
recorded in the TEDS memory when TEDS data is loaded, the bridge voltage
setting is automatically changed to a setting equal to or lower than the
maximum excitation voltage.
2. To save the rated output or the change you made, press ENTER. The rated
capacity appears with the least significant digit blinking.
NOTE The position of the decimal point you set in this step is used for
readings on the TD-260T.
4. To save the rated capacity or the change you made, press ENTER.
CAUTION If you use the D/A option, the rated capacity is set as the D/ A full
scale value.
28
5. To perform zero-point calibration;
with no load on the sensor, press ZERO, and then press ENTER. The display
shows “0-CAL,” followed by zero for a reading, showing that the zero-point
calibration is completed.
6. Calibration
6-5. Zero-point calibration
Using the analog circuit on the input of the strain amplifier, you can set the
A/D input voltage to zero. 1. Set CAL Lock (F3- 5) to OFF. 2. With no load on
the sensor, press ZERO, and then
press ENTER. The display shows “0-CAL,” followed by zero for a reading,
showing that the zero-point calibration is completed.
If a calibration error is displayed, make the appropriate correction, and then repeat the calibration.
NOTE
< If Remote Sense/TEDS (F2-6) is not set to 0.Aut or 1.tEd, “Err08” appears on
the display, followed by a reading.
< If Remote Sense/TEDS (F2-6) is set to 2.rSEn, “Err13” appears on the
display, followed by a reading.
< If Remote Sense/TEDS (F2-6) is set to a value other than 2.rSEn, and a TEDS
sensor is not connected to the TD-260T, “Err06” appears on the display,
followed by a reading.
6-4-1. Displayed digits of rated capacity for TEDS calibration
Rated capacity for TEDS sensor
1 2 3 4 5 10 20 30 40 50 100 200 300 400 500
Unit
N, kN N, kN N, kN N, kN N, kN N, kN N, kN N, kN N, kN N, kN N, kN N, kN N, kN
N, kN N, kN
Reading
01.000 02.000 03.000 04.000 005.00 010.00 020.00 030.00 040.00 0050.0 0100.0
0200.0 0300.0 0400.0 00500
If a calibration error is displayed, make the appropriate correction, and then
repeat the calibration. 3. Set CAL Lock (F3-5) to ON.
NOTE
You can set the sensor input of up to +2 or down to -2 mV/V to zero.
29
Setting Functions
7-1. Setting HI, LO, HH, and LL values
You can set HI, LO, HH, and LL values, compare each with a reading, and then
generate each decision output so that a comparison result is determined.
CAUTION
If you press ENTER to save settings while they do not meet the condition LL
value < LO value < HI value < HH value, “Err08” appears, and the settings
cannot be saved. In that case, enter values that meet the condition above, and
then press ENTER again.
NOTE
The operations of decision outputs vary depending on the setting for Comparative Output Pattern (F1-15).
7-1-1. Statuses of decision output LEDs
LED
Color
Description
HH/HI
Lights if HH value < reading (with the setting for Enable HH and LL (F1-5) ON) Red Lights if HI value < reading (with the setting for Enable HH and LL (F1-5) OFF)
Lights if HI value < reading < HH value Orange (with the setting for Enable HH
and LL
(F1-5) ON)
OK
Green Lights if HI value reading LO value
LL / LO
Lights if reading < LL value
(with the setting for Enable HH and LL
Red
(F1-5) OFF) Lights if reading < LO value
(with the setting Enable HH and LL (F1-5)
OFF)
Lights if LL value < reading < LO value Orange (with the setting for Enable HH
and LL
(F1-5) ON)
JUDGE
Lights if JUDGE signal is ON
Green
(with the Enable Block Setting (F3-2) ON) Always lit
(with the Enable Block Setting (F3-2) OFF)
Green Peak hold is set
PEAK Red Operating in the Peak hold mode
Orange
Operating in block setting
the
Peak
hold
mode
with
Green Bottom hold is set
BOTTOM Red Operating in the Bottom hold mode
Orange
Operating in the with block setting
Bottom
hold
mode
Green Peak-to-Peak hold is set
P-P
Red
Operating in the Peak-to-Peak hold mode
Orange
Operating in the Peak-to-Peak mode with block setting
hold
SAMPLE
Green Sample hold is set Red Operating in the Sample hold mode
CAL Orange Lights 5) ON with the setting for CAL Lock (F3- SET Orange Lights ON with the setting for SET Lock (F3-6) HOLD LOCK
7-1-2. HI value 1. Enter the setup screen number F1-1.
Press once
2. To change the blinking digit of the value, press 5 or b. To move to the
previous or next digit, press g or t. To move to the previous or next digit,
press:
To change the digit, press:
To put a minus sign in front of the set value, press FNC.
NOTE If you press FNC, regardless of the position of the blinking digit, the
sign is added or removed.
3. To save the setting, press ENTER, and then to exit the setup mode, press
ESC.
To save, press:
30
7-1-3. LO value 1. Enter the setup screen number F1-2.
Press once
Press once
7. Setting Functions 7-1-4. HH value
1. Enter the setup screen number F1-3. Press once
Press twice
2. To change the blinking digit of the value, press 5 or b. To move to the
previous or next digit, press g or t.
To move to the previous or next digit, press:
2. To change the blinking digit of the value, press 5 or b. To move to the
previous or next digit, press g or t.
To move to the previous or next digit, press:
To change the digit, press:
To change the digit, press:
To put a minus sign in front of the set value, press FNC.
To put a minus sign in front of the set value, press FNC.
NOTE
If you press FNC, regardless of the position of the blinking digit, the sign
is added or removed.
3. To save the setting, press ENTER, and then to exit the setup mode, press
ESC.
NOTE
If you press FNC, regardless of the position of the blinking digit, the sign
is added or removed.
3. To save the setting, press ENTER, and then to exit the setup mode, press
ESC.
To save, press:
To save, press:
31
7. Setting Functions
7-1-5. LL value 1. Enter the setup screen number F1-4.
Press once
Press 3 times
7-1-6. Enable HH and LL You can enable or disable HH and LL decision outputs.
If you select OFF, the HH or LL or Low-Low limit decision output is not
generated anymore, and the HH/HI or LL/LO LED display solid red instead to
show that the HI or LO decision output is generated.
1. Enter the setup screen number F1-5.
Press once
2. To change the blinking digit of the value, press 5 or b. To move to the
previous or next digit, press g or t.
To move to the previous or next digit, press:
2. To select OFF or ON, press 5 or b.
To change the digit, press:
Press 4 times
To put a minus sign in front of the set value, press FNC.
or
3. To save the selection, press ENTER, and then to exit the setup mode, press ESC.
NOTE
If you press FNC, regardless of the position of the blinking digit, the sign
is added or removed.
3. To save the setting, press ENTER, and then to exit the setup mode, press
ESC.
To save, press:
To save, press:
32
7. Setting Functions
7-2. Hysteresis
Using this function, You can buffer the set points at which the HH, HI, LO,
and LL decision outputs switch off. If fluctuations in the reading occur near
each comparative point, the corresponding decision output switches on and off,
which is a phenomenon called “chattering.” You can prevent the output from
chattering by setting a hysteresis value. Hysteresis for the HH and HI values
is enabled when the reading decreases, and hysteresis for the LL and LO values
is enabled when the reading increases, as shown below.
Reading
Reading HI/HH
Hysteresis
HI/HH
LO/LL
Hysteresis
HI/HH
Decision output
HH/HI Disable hysteresis OFF
ON
Enable hysteresis
OFF ON
Time Hysteresis
Decision output HH/HI OFF ON
LL/LO
OFF ON
Time
Conditions for generating decision outputs (with “standard output” selected for Comparative Output Pattern)
Decision output
State
Condition
OFF w ON HH value < reading HH
ON w OFF Reading (HH value hysteresis value)
OFF w ON HI value < reading HI
ON w OFF Reading (HI value hysteresis value)
OFF w ON Reading < LO value LO
ON w OFF (LO value + hysteresis value) reading
OFF w ON Reading < LL value LL
ON w OFF (LL value + hysteresis value) reading
CAUTION
Set a hysteresis value to meet the following conditions. HI value (HH value –
hysteresis value) LO value (HI value – hysteresis value) LL value (LO value –
hysteresis value)
NOTE
< The same amount of hysteresis is set for HI, LO, HH, and LL values. < If the
setting is 0, hysteresis is disabled.
33
7. Setting Functions
1. Enter the setup screen number F1-6.
Press once
Press 5 times
2. To change the blinking digit of the value, press 5 or b. To move to the
previous or next digit, press g or t. To move to the previous or next digit,
press:
To change the digit, press:
7-3. Comparative Output Pattern
The following two operating settings for decision outputs are available:
standard output and area output.
7-3-1. Standard output Each decision output switches on when the following
conditions are met. (with hysteresis set to 0)
HH: HH value < reading HI: HI value < reading LO: Reading < LO value LL:
Reading < LL value
Reading
HH HI LO LL
3. To save the value you entered, press ENTER, and then to exit the setup
mode, press ESC.
To save, press:
NOTE < The same amount of hysteresis is set for HI, LO, HH and LL
values. < If the setting is 0, hysteresis is disabled.
Decision output
HH
OFF ON
HI
OFF ON
OK
OFF ON
LO
OFF ON
LL
OFF ON
Time
NOTE
The operations of decision outputs vary depending on the setting for
Hysteresis (F1-6). For detailed information, see the section 7-2.
“Hysteresis.”
34
7-3-2. Area output Each decision output switches on when the following
conditions are met. (with hysteresis set to 0)
HH: HH value < reading HI: HI value < reading < HH value LO: LL value <
reading < LO value LL: Reading < LL value
Reading
HH
HI
LO
LL
7. Setting Functions
1. Enter the setup screen number F1-15. Press once Press once
2. To select a comparative output pattern,press 5 or b.
Decision output
HH
OFF ON
HI
OFF ON
OK
OFF ON
LO
OFF ON
LL
OFF ON
Time
or
The following two settings are available. 0.nor: Standard output 1.ArE: Area
output
3. To save the selection, press ENTER, and then to exit the setup mode, press
ESC.
NOTE
The operations of decision outputs vary depending on the setting for
Hysteresis (F1-6). For detailed information, see the section 7-2.
“Hysteresis.”
To save, press:
35
7. Setting Functions
7-4. Comparison Mode
You can set a condition for comparing a reading with HI, LO, HH, and LL values
and determining the comparison result. 1. Enter the setup screen number F1-7.
Press once
Press 6 times
7-5. Near Zero
You can set a range to be considered near zero. 1. Enter the setup screen
number F1-8.
Press once
Press 7 times
2. To change the setting, press 5 or b.
2. To change the blinking digit of the value, press 5 or b. To move to the
previous or next digit, press g or t.
To move to the previous or next digit, press:
or
To change the digit, press:
The following six settings are available. 0.ALL: Always compare to generate
decision outputs 1.Stb: Compare to generate decision outputs when stabilized
2.no0: Always compare to generate decision outputs except near zero 3.nOS:
Compare to generate decision outputs when stabilized except near zero 4.HLD:
Compare to generate decision outputs when in hold mode 5.OFF: Do not compare
NOTE
The comparison mode settings 1.Stb, 2.no0, 3.nOS are closely related to the
near-zero and motion detection settings. For detailed information about the
near-zero and motion detection settings, see the sections 7-5. “Near Zero” and
7-6. “Motion Detection.”
3. To save the selection, press ENTER, and then to exit the setup mode, press
ESC.
3. To save the value you entered, press ENTER, and then to exit the setup
mode, press ESC.
To save, press:
NOTE The operations of decision outputs near zero are closely related to the
condition you set for comparing a reading with HI, LO, HH, and LL values and
determining the comparison result. For detailed information about the
comparison mode setting, see the section 7-4. “Comparison Mode.”
To save, press:
36
7. Setting Functions
7-6. Motion detection
You can set parameters used for detecting stability. If the difference between the currently displayed reading and the reading that was displayed 100 milliseconds ago is equal to or lower than the bandwidth you set, and then the state remains unchanged for more than the length of set time, the TD-260T determines that the currently displayed reading is stable. The motion detection setting is closely related to the comparison mode setting. For detailed information, see the section 7-4. “Comparison Mode.” If the displayed reading is determined as stable, the P.C. (stable) signal from the optional BCD output is turned on and the stable output for the RS-232C read status command is set to ON.
Reading
Determine as stable Determine as unstable Determine as stable
Motion detection bandwidth
100mSec
Motion Detect time
7-6-1. Motion Detection Time 1. Enter the setup screen number F1-9.
Press once
100mSec
100mSec
Motion Detect time
Time
Stable
Unstable
7-6-2. Motion Detection Bandwidth 1. Enter the setup screen number F1-10.
Stable
Press once
Press 7 times
Press 6 times
2. To change the blinking digit of the value, press 5 or b. To move to the
previous or next digit, press g or t.
To move to the previous or next digit, press:
2. To change the blinking digit of the value, press 5 or b. To move to the
previous or next digit, press g or t.
To move to the previous or next digit, press:
To change the digit, press:
To change the digit, press:
3. To save the setting, press ENTER, and then to exit the setup mode, press ESC.
3. To save the setting, press ENTER, and then to exit the setup mode, press ESC.
To save, press:
To save, press:
37
7. Setting Functions
7-7. Zero tracking
Using zero tracking automatically tracks and corrects deviations from zero
point, such as drift. If a drift is within the bandwidth you set, the function
automatically moves a drifted zero point to zero at each set time interval.
Zero tracking is enabled only while both CAL Lock (F3-5) and Enable Digital
Zero (F3-1) are set to ON. Zero tracking is disabled while its time interval
and bandwidth are respectively set to 0.0 sec and 0.0.
Reading
Zero tracking
bandwidth
0
Time
Zero tracking time interval
zero tracking enabled
zero tracking enabled
7-7-1. Zero Tracking Time Interval 1. Enter the setup screen number F1-11.
Press once Press 5 times
7-7-2. Zero Tracking Bandwidth You can set a zero tracking bandwidth for the
least three significant digits of a reading of your device. 1. Enter the setup
screen number F1-12.
Press once
Press 4 times
2. To change the blinking digit of the value, press 5 or b. To move to the
previous or next digit, press g or t. To move to the previous or next digit,
press:
To change the digit, press:
2. To change the blinking digit of the value, press 5 or b. To move to the
previous or next digit, press g or t. To move to the previous or next digit,
press:
To change the digit, press:
3. To save the setting, press ENTER, and then to exit the setup mode, press ESC.
3. To save the setting, press ENTER, and then to exit the setup mode, press ESC.
To save, press:
To save, press:
38
7. Setting Functions
7-8. Digital Filter
You can select a number of measurements to be used for obtaining a moving
average. 1. Enter the setup screen number F2-3.
Press twice
Press twice
7-9. Analog Filter
You can select a cutoff frequency (Hz) of the low-pass filter. 1. Enter the
setup screen number F2-4.
Press twice
Press 3 times
2. To select OFF, 4, 8, 16, 32, 64, 128, or 256, press 5 or b.
2. To Select 3, 10, 30, 100, or 300 by pressing 5 or b.
or or
3. To save the selection, press ENTER, and then to exit the setup mode, press ESC.
3. To save the selection, press ENTER, and then to exit the setup mode, press ESC.
To save, press:
To save, press:
39
7. Setting Functions
7-10. SET Lock
You can prevent changes to settings you have saved. For detailed information
about the applicable setup items, see the section 4-2. “List of setup items
and settings.” 1. Enter the setup screen number F3-6.
Press 3 times
Press 3 times
7-11. Static Strain Mode
You can display the input signal in microstrains. By selecting this mode, you
can check for sensor output or fluctuations in the reading and other erroneous
phenomena caused by parts including sensor and cable. 1. Enter the setup
screen number F3-7.
Press 3 times
Press twice
2. To select OFF or ON, press 5 or b.
2. To select OFF or ON, press 5 or b.
or or
3. To save the selection, press ENTER, and then to exit the setup mode, press
ESC.
To save, press:
3. To save the selection, press ENTER, and then to exit the setup mode, press
ESC.
With ON selected, the input signal is displayed in microstrains.
To save, press:
NOTE
< To exit the static strain mode, select OFF while the above screen is
displayed, or shut off the power to the TD-260T, and then turn it on again.
< The TD-260T uses one strain gage with a gauge factor of 2.0 to display
static strain.
40
7. Setting Functions
7-12. Decision-Output Check
You can turn on decision outputs, one by one. By using this function, you can
check wiring of each decision output. Note that the TD-260T does not work as
an indicator while the function is enabled. 1. Enter the setup screen number
F4-8.
Press 4 times
7-13. Control-Input Check
Using the alphanumeric display on the TD-260T, you can check the ON or OFF
state of the following signals: D/Z, HOLD, JUDGE, and CLEAR signals. 1. Enter
the setup screen number F4-9.
Press 4 times
Press 3 times
Press 4 times
2. The OK decision output is generated.
2. The display changes, depending on the states of the input signals.
or
The TD-260T generates outputs, one by one, as shown in the image below; if you
press b while the display shows “OK,” the LO decision output turns on next, or
if you press 5, the HI decision output turns on next.
With High-High decision output ON, HH/HI LED lights solid red.
With High decision output ON, HH/HI LED lights solid oragne.
With OK decision output ON, OK LED lights solid green.
With Low decision output ON, LL/LO LED lights solid orange.
With Low-Low decision output ON, LL/LO LED lights solid red.
3. To finish checking the wiring, press ENTER or ESC.
Below is an example of the display. ON state OFF state
D/Z
JUDGE
HOLD
CLEAR
CAUTION
The display is refreshed about every 0.5 seconds. If the duration of a signal
input is 0.5 seconds or less, the display may not be refreshed.
3. To finish checking the ON/OFF states, press ESC.
or 4. To exit the setup mode, press ESC.
4. To exit the setup mode, press ESC.
41
7. Setting Functions
7-14. Beep
You can turn on or off the beep sound that you normally hear when you press a
key on the TD-260T. 1. Enter the setup screen number F4-10.
Press 4 times
Press twice
7-15. Select Data Output
You can select the data to be sent from the BCD parallel data output, the RS-
232C serial data output, or the D/A converter, which can be installed in the
option space. 1. Enter the setup screen number F4-11.
Press 4 times
Press once
2. To select ON or OFF, press 5 or b.
2. To change the data output setting, press 5 or b.
or or
3. To save the selection, press ENTER, and then to exit the setup mode, press
ESC.
To save, press:
The following two data output settings are available. 0.diSP: Reference the
display; hold value is sent in hold mode
1.InP: Reference the input; input value is sent in hold mode
3. To save the selection, press ENTER, and then to exit the setup mode, press
ESC.
To save, press:
42
8. Hold function
8-1. Hold Mode
The following hold modes are available: Sample, Peak, Bottom, and Peak-to-
Peak. 1. Enter the setup screen number F2-1.
8-2. External Hold Mode
You can select a signal type for the HOLD signal from the control input/output
terminal A7. 1. Enter the setup screen number F2-7.
Press twice
Press twice
Press twice
2. To select a hold mode setting, press 5 or b.
2. To select an external hold mode setting, press 5 or b.
or
or
The following five settings are available. 1.SPL: Sample hold 2.PEk: Peak hold
3.bot: Bottom hold 4.P2P: Peak-to-peak hold 0.OFF: Disable hold
3. To save the selection, press ENTER, and then to exit the setup mode, press
ESC.
The following two settings are available. 0.LEv: Level response 1.PLS: Pulse
response
3. To save the selection, press ENTER, and then to exit the setup mode, press
ESC.
To save, press:
To save, press:
NOTE
You can specify a block to be held for Peak, Bottom, or Peakto-Peak hold.
43
8. Hold function
8-3. Enable CLEAR Signal
You can enable or disable the CLEAR signal from the control input/output
terminal A5 and the ESC/CLEAR key. 1. Enter the setup screen number F2-8.
Press twice
Press once
8-4. Enable Block Setting
1. Enter the setup screen number F3-2.
Press 3 times Press once
2. To select ON or OFF, press 5 or b.
2. To select OFF or ON, press 5 or b.
or
3. To save the selection, press ENTER, and then to exit the setup mode, press
ESC.
or
3. To save the selection, press ENTER, and then to exit the setup mode, press
ESC.
To save, press:
To save, press:
44
8-5. Enable JUDGE Signal
You can disable or enable the signal that controls the decision outputs. 1.
Enter the setup screen number F3-8.
Press 3 times Press once
2. To select OFF or ON, press 5 or b.
or
OFF: Always enable decision outputs ON: Enable decision outputs while the
JUDGE signal is ON 3. To save the selection, press ENTER, and then to exit the
setup mode, press ESC.
To save, press:
8. Hold function
45
8. Hold function
8-6. Sample hold
You can put the TD-260T into the Sample hold mode by using the HOLD key or
setting the HOLD signal from the control input/ output terminal A7 to ON. If
the HOLD key is pressed, the TD-260T starts holding the displayed reading, and
then if the HOLD key is pressed again, the TD-260T exits the hold mode.
Reading
Reading Sensor input
HOLD
A7 HOLD
OFF
level response
ON
Time
HOLD key or
OFF
pulse response ON
Decision output
OFF
(Enable JUDGE Signal OFF) ON
Enable
OFF
A6 JUDGE
ON
Decision output
OFF
(Enable JUDGE Signal ON) ON
Enable
< You can choose between level response and pulse response for the HOLD signal
from the control input/output terminal A7 by changing the setting for External
Hold Mode (F2-7).
< The operations of decision outputs vary depending on the setting for
Comparison Mode (F1-7). The mode 4.HLd is selected in the above example.
< The operations of decision outputs vary depending on the setting for Enable
JUDGE Signal (F3-8). With the setting OFF selected, a decision output can
always be generated, depending on the input signal. With the setting ON
selected, a decision output can be generated, depending on the input signal,
only while the JUDGE signal from the A6 is ON, so that you can avoid
unnecessary switching of decision output relays.
< Block setting is unavailable for the Sample hold mode.
< You cannot use either the CLEAR key or the CLEAR signal from the control
input/output terminal A5 while the TD-260T is operating in the Sample hold
mode.
46
8. Hold function
8-7. Peak hold
8-7-1. Operation without block setting
If the HOLD key is pressed or the HOLD signal from the control input/output
terminal A7 is set to ON, the TD-260T starts holding and displaying the
maximum reading (peak value). If the HOLD key is pressed again or the HOLD
signal from the control input/ output terminal A7 is set to OFF, the TD-260T
exits the Peak hold mode, and then starts measuring the input signal again. If
the CLEAR key is pressed or the CLEAR signal from the control input/output
terminal A5 is set to ON while the TD-260T is in the hold mode, the TD-260T
resets the peak value.
Reading
Reading Sensor input
A7 HOLD
OFF
level response
ON
HOLD key or OFF pulse response ON
OFF
A5 CLEAR
ON
OFF
A6 JUDGE
ON
Decision output OFF
(with JUDGE signal ON) ON
HOLD
Time
Enable
8-7-2. Operation with block setting
The TD-260T continues to display the reading that has been displayed when a
Peak hold operation is terminated. If the CLEAR key is pressed or the CLEAR
signal from the control input/output terminal A5 is set to ON, the TD-260T
resets the display.
Reading
Reading Sensor input
A7 HOLD
OFF
level response
ON
HOLD key or OFF pulse response ON
OFF
A5 CLEAR
ON
OFF
A6 JUDGE
ON
Decision output OFF
(with JUDGE signal ON) ON
HOLD
HOLD with block setting
Enable
Time
If the CLEAR key is pressed or the CLEAR signal from the control input/output terminal A5 is set to ON while the TD-260T is in the hold mode, the TD-260T resets the peak value.
47
8. Hold function
8-8. Bottom hold
8-8-1. Operation without block setting
If the HOLD key is pressed or the HOLD signal from the control input/output
terminal A7 is set to ON, the TD-260T starts holding and displaying the
minimum reading (bottom value). If the HOLD key is pressed again or the HOLD
signal from the control input/ output terminal A7 is set to OFF, the TD-260T
exits the Bottom hold mode, and then starts measuring the input signal again.
If the CLEAR key is pressed or the CLEAR signal from the control input/output
terminal A5 is set to ON while the TD-260T is in the hold mode, the TD-260T
resets the bottom value.
Reading
Reading Sensor Input
A7 HOLD
OFF
level response
ON
HOLD key or OFF pulse response ON
OFF
A5 CLEAR
ON
OFF
A6 JUDGE
ON
Decision output OFF
(with JUDGE signal ON) ON
HOLD
Time
Enable
8-8-2. Operation with block setting
The TD-260T continues to display the reading that has been displayed when a
Bottom hold operation is terminated. If the CLEAR key is pressed or the CLEAR
signal from the control input/output terminal A5 is set to ON, the TD-260T
resets the display.
Reading
Reading Sensor Input
A7 HOLD
OFF
level response
ON
HOLD key or OFF pulse response ON
OFF
A5 CLEAR
ON
OFF
A6 JUDGE
ON
Decision output OFF
(with JUDGE signal ON) ON
HOLD
HOLD with block setting Time
Enable
If the CLEAR key is pressed or the CLEAR signal from the control input/output terminal A5 is set to ON while the TD-260T is in the hold mode, the TD-260T resets the bottom value.
48
8. Hold function
8-9. Peak-to-Peak hold
8-9-1. Operation without block setting
If the HOLD key is pressed or the HOLD signal from the control input/output
terminal A7 is set to ON, the TD-260T starts holding the maximum reading (peak
value) and the minimum reading (bottom value) for each sampling. Then, the TD-
260T starts holding and displaying the maximum difference between those two
values. If the HOLD key is pressed again or the HOLD signal from the control
input/output terminal A7 is set to OFF, the TD-260T exits the Peak-to-Peak
hold mode, and then starts measuring the input signal again. If the CLEAR key
is pressed or the CLEAR signal from the control input/output terminal A5 is
set to ON while the TD-260T is in the hold mode, the TD-260T resets the peak-
to-peak value.
Reading 0
P P
P
B
B
Reading Sensor Input P: Peak value B: Bottom value
Time
A7 HOLD
OFF
level response
ON
HOLD key or OFF pulse response ON
OFF
A5 CLEAR
ON
OFF
A6 JUDGE
ON
Decision output OFF
(with JUDGE signal ON) ON
HOLD Enable
8-9-2. Operation with block setting
The TD-260T continues to display the reading that has been displayed when a
hold operation is terminated. If the CLEAR key is pressed or the CLEAR signal
from the control input/output terminal A5 is set to ON, the TD-260T resets the
display.
Reading
Reading
P
Sensor Input
P: Peak value B: Bottom value
0
P
B
Time
B
A7 HOLD
OFF
level response
ON
HOLD
HOLD with block setting
HOLD key or OFF pulse response ON
OFF
A5 CLEAR
ON
OFF
A6 JUDGE
ON
Decision output OFF
(with JUDGE signal ON) ON
Enable
If the CLEAR key is pressed or the CLEAR signal from the control input/output terminal A5 is set to ON while the TD-260T is in the hold mode, the TD-260T resets the peak-to-peak value.
49
9. Digital Zero Function
9-1. Digital Zero
You can set the currently displayed reading to zero. 1. Press ZERO or short-
circuit the control input/output
terminals A8 for D/Z and A9 for COM.
9-2. Digital Zero Limit
You can set a range that can be digitally zeroed. (Note that a value that is
set for Digital Zero Limit does not use mV/V as the unit of measure.) 1. Enter
the setup screen number F1-13.
Press once
Press 3 times
< To set a range that can be digitally zeroed, use Digital Zero Limit (F1-13).
< To clear the digitally set zero, shut off the power to the TD260T or perform
Clear Digital Zero (F1-14).
CAUTION
< The Digital zero function is available only with both CAL Lock (F3-5) and
Enable Digital Zero (F3-5) set to ON.
< If you press ZERO with the CAL Lock (F3-5) set to OFF, zeropoint calibration
is performed instead.
2. To change the blinking digit of the value, press 5 or b. To move to the
previous or next digit, press g or t. To move to the previous or next digit,
press:
To change the digit, press:
3. To save the value you set, press ENTER, and then to exit the setup mode,
press ESC.
To save, press:
CAUTION
If the current sensor input exceeds the digital zero limit setting, the TD-
260T, which regards as zero the sensor input that was obtained when the zero-
point calibration was performed, shows Err14, meaning digital zero limit
error, and cannot set the displayed reading to zero.
NOTE
If you perform the function explained in the section 9-3. “Clear Digital
Zero,” the TD-260T displays the current sensor input value, regarding as zero
the sensor input that was observed when the zero-point calibration was
performed.
50
9. Digital Zero Function
9-3. Clear Digital Zero
You can restore the zero setting that has been used before Digital Zero was
performed. After this function is performed, the TD-260T displays the current
sensor input value, regarding as zero the sensor input that was observed when
the zero-point calibration was performed.
1. Enter the setup screen number F1-14.
Press once
9-4. Enable Digital Zero
With Enable Digital Zero set to OFF, a reading cannot be digitally set to zero
if you press ZERO or short-circuit the control input/output terminals A8 for
D/Z and A9 for COM. 1. Enter the setup screen number F3-1.
Press 3 times
Press twice
2. To select ON or OFF, press 5 or b.
2. To clear the digitally set zero, press ENTER. To exit the setup mode,
press ESC.
or
3. To save the selection, press ENTER, and then to exit the setup mode, press ESC.
To save, press:
51
Digital Zero Function
9-5. Digital Offset
You can set a value that is automatically subtracted from a measured value.
(Tare) 1. Enter the setup screen number F2-2.
Press twice Press once
2. To change the blinking digit of the value, press 5 or b. To move to the
previous or next digit, press g or t. To move to the previous or next digit,
press: To change the digit, press:
NOTE If you press FNC, regardless of the position of the blinking digit, a
minus sign is added or removed.
3. To save the vale you set, press ENTER, and then to exit the setup mode,
press ESC.
To save, press:
52
10. BCD Data Output (TD-260T AC(BCD) / TD-260T DC(BCD))
The BCD data output is an interface for sending the TD-260T displayed readings
as BCD-coded data. The interface connects the TD-260T to a computer, process
controller, sequencer or equivalent, for controlling, summarizing, recording
or any other processing purposes. The TD-260T internal circuit and external
connection circuit are electrically isolated with a photocoupler.
10-1. Panel
1
1 BCD connector
No. I/O Signal name No. I/O Signal name
1
COM
19
COM
2
1
20
NC
3
2
21
4
4
22
Near Zero
5
8
23
Negative (polarity)
6
10
24 Output
OVER
7
20
25
P.C (stable)
8
40
9
80
Output
10
100
11
200
26
STROBE
27
Hold BCD Data
Input
28
Switch Data Logic
29
12
400
30
13
800
31
NC
14
1000
32
15
2000
33
16
4000
34
80000
17
8000
35 Output 40000
18
10000
36
20000
Applicable connector: DDK 57-30360 or equivalent
NOTE
To reduce the unnecessary radio frequency emission, be sure to install the
provided ferrite core on the output cable (for voltage or current).
53
10.BCD Data Output (TD-260T AC(BCD) / TD-260T DC(BCD))
10-2. Changing logic
You can change the logic setting for all output signals except for the STROBE
signal.
If a COM terminal and the pin 28 are open, the signals are set to negative
logic. If they are short-circuited, the signals are set to positive logic.
Pin 28 Open Shorted
Logic Negative Positive
10-3. Equivalent circuit
Output equivalent circuit TD-260T
Vcc +5V Vext
COM
Output data
Negative logic
0
Output transistor (Output pin level)
OFF (H)
1
Output transistor (Output pin level)
ON (L)
Vce = 30V(max), Ic = 50mA(max)
Positive logic ON (L)
OFF (H)
Input equivalent circuit TD-260T
Vcc +12V
Open
OFF
Shorted ON
Approx. 10mA
Push Switch
COM When transistor is connected
When open collector is connected IN
ON when IN is high
CAUTION < Do not apply external voltages to the signal input circuit. < Use a
contact device with 20 mA or more capacity.
54
10.BCD Data Output (TD-260T AC(BCD) / TD-260T DC(BCD))
10-4. Signal timing
P.C (stable) The P.C signal is turned on when the input signal is stable. To
load the BCD data in synchronization with the STROBE signal, be sure to wait
25 milliseconds or more after the falling edge of the P.C signal.
BCD data
P.C 25mSec or more
OVER The OVER signal is generated when the message -LOAD or LOAD, or -Full or
Full is displayed.
STROBE The STROBE pulse is generated in synchronization with the BCD data. To
load the data, use the leading edge of the pulse. Using BCD Data Refresh Rate
(F4-1), you can change the BCD data refresh rate.
BCD data
OVER
1 STROBE 0
Strobe pulse width 7 msec, min.
CAUTION If you have set the data refresh rate to 100 times per second, be sure
to load the data within 5 milliseconds.
NOTE For detailed information about the P.C (stable) signal, see the section
7-6. “Motion Detection.”
55
10.BCD Data Output (TD-260T AC(BCD) / TD-260T DC(BCD))
10-5. BCD Data Refresh Rate
You can change the BCD data refresh rate (times per second). 1. Enter the
setup screen number F4-1.
Press 4 times
10-6. BCD Output Check
By changing the BCD output from AAAAA to 55555 in this order, you can check
the BCD data wiring to an external device for correct connection. 1. Enter the
setup screen number F4-2.
Press 4 times
Press once
2. To select 100, 50, 20, 10, 5, 2, or 1, press 5 or b.
or 3. To save the selection, press ENTER, and then to exit
the setup mode, press ESC.
2. The TD-260T goes into check mode, with AAAAA being both displayed on the
TD-260T and sent from the BCD data output. To display 55555 and send it from
the BCD output, press ENTER. Then, to put the TD-260T into setup mode, press
ENTER again. The display shows “F4-2.”
AAAAA is displayed. if you press ENTER , the display changes to 55555.
To save, press:
3. To exit the setup mode, press ESC.
NOTE
< While “F4-2” is displayed, a reading displayed on the TD260T is generated
from the BCD data output.
< Each time you press ENTER after “F4-2” is displayed, the BCD data output
cycles through the following.
AAAAA
55555
reading
56
11. RS-232C Interface (TD-260T AC(232C) / TD-260T DC(232C))
The RS-232C interface loads readings and states from and also saves set values
into the TD-260T. It connects the TD-260T to a computer, process controller,
sequencer, or equivalent, for controlling, summarizing, recording, or any
other processing purposes. The TD-260T internal circuit and external
connection circuit are electrically isolated with a photocoupler.
11-1. Panel
11-2. Standard
Signal level:
RS-232C
Transmission distance: Approx. 15 m
Transfer method:
Full duplex, asynchronous communication
Transfer rate:
1200, 2400, 4800, 9600, 19200 bps
Bit configuration: Start bit: 1
Data bits:
7 or 8 bits
Stop bits:
1 or 2 bits
Parity bits:
None, odd, or even
Flow control:
None
Code:
ASCII
1
1 RS-232C connector
No.
I/O
Signal name
No.
I/O
1
FG
14
2 Output TxD
15
3
Input RxD
16
4 Output RTS
17
5
Input CTS
18
6
19
7
SG
20 Output
8
21
9
22
10
23
11
24
12
25
13
Applicable plug: 25-pin D-SUB connector (JAE’s DB-25P-NR or equivalent)
Signal name
DTR
57
11. RS-232C Interface (TD-260T AC(232C) / TD-260T DC(232C))
11-3. Cable wiring example
TD-260T No. Signal 1 FG 2 TxD 3 RxD 4 RTS 5 CTS 8 (CD) 6 (DSR) 20 DTR 7 SG
Computer etc. No. Signal 1 FG 2 TxD 3 RxD 4 RTS 5 CTS 8 (CD) 6 (DSR) 20 DTR 7 SG
11-4. RS-232C Interface Setting
You can change the RS-232C transmission conditions: 1. Enter the setup screen
number F4-3.
Press 4 times
Press twice
2. To change the blinking digit of the value, press 5 or b. To move to the
previous or next digit, press g or t. To move to the previous or next digit,
press:
To change the digit, press:
The number for each digit represents a transmission setting as shown below.
13100 22222 abcde
a Transmission mode 0: Transmission mode 0 1: Transmission mode 1
b Baud rate 0: 1200 bps 1: 2400 bps 2: 4800 bps 3: 9600 bps 4: 19200 bps
c Data bits 0: 7 bits 1: 8 bits
d Parity bits 0: None 1: Odd
2: Even
e Stop bits
0: 1 bit
1: 2 bits
3. To save the settings, press ENTER, and then to exit the setup mode, press ESC.
To save, press:
58
11. RS-232C Interface (TD-260T AC(232C) / TD-260T DC(232C))
11-5. List of Commands
11-5-1. Read commands
Command
Function
RA Load displayed readings
RS Load statuses
RD Load statuses
Page 60 60 60
11-5-2. Set commands
Command
Setup item
W01 HI Valuet W02 LO Value W03 Comparison Mode W04 Hysteresis W05 Digital Offset W06 Near Zero W07 HH value W08 LL value W09 Enable HH and LL
W11 Digital Filter
W12 Analog Filter W13 Motion Detection Time W14 Motion Detection Bandwidth W15 Zero Tracking Time Interval W16 Zero Tracking Bandwidth W17 Hold Mode W18 Digital Zero Limit W19 Comparative Output Pattern
Page
61 61 61 61 61 61 61 61 61
61
61 61 61 61 62 62 62 62
SET Lock
CAL Lock
Unit
x
x
x
x
x
x
x
x
x
x
of times
x
Hz
x
Sec
x
x
Sec
x
x
x
x
x: Applicable
59
11. RS-232C Interface (TD-260T AC(232C) / TD-260T DC(232C))
Command
Setup item
CAL Lock W21
SET Lock W22 Minimum Scale W23 Display Refresh Rate W24 Bridge Voltage W25
External Hold Mode W26 Remote Sense/TEDS W27 Enable CLEAR Signal W28 Enable
Block Setting W29 Enable JUDGE Signal
W31 BCD Data Refresh Rate
W32 RS-232C Interface Setting W33 D/A Zero W34 D/A Full Scale W35 Beep W36 Select Data Output
Page
62 62 62 62 62 62 62 62 62 62
62
62 63 63 63 63
SET CAL Lock Lock
x x x x x x x x x x x x x x x: Applicable
Unit
of times V
of times/Sec
11-5-3. Set-value retrieving set value
Command
Function
Wxx
Set value for set command Wxx
Page 63
11-5-4. Control
Command
Function
CE Hold ON
CF Hold OFF
CG Digital zero
CH Reset digital zero
CI Reset hold
CJ JUDGE contact input ON
CK JUDGE contact input OFF
Page 64 64 64 64 64 64 64
60
11. RS-232C Interface (TD-260T AC(232C) / TD-260T DC(232C))
11-6. Transmission Mode
You can switch between the transmission mode 0 and the transmission mode 1 by
changing the transmission mode setting explained in the section 11-4. “Setting
RS-232C interface.”
11-6-1. Transmission mode 0 In the transmission mode 0, a computer or similar
device sends a command to load a reading, statuses, and set values from or to
save set values into the TD-260T.
11-6-2. Transmission mode 1 In the transmission mode 1, the TD-260T
continuously transmits readings and statuses.
11-7. Format for transmission mode 0
A host computer sends data called Request to the TD-260T, and then the TD-260T sends data called Response back to the host computer. Request and Response use ASCII characters except for CR and LF. (CR: Carriage Return (0x0D), LF: Line Field (0x0A))
11-7-1. Read displayed-reading command (RA)
Request Response
R A CR R A ± 1 0 0 . 0 0 CR LF
Sign bit Plus: + Minus: –
Reading
11-7-2. Read status command (RS)
Request Response
R S CR R S 0 0 0 0 0 0 0 0 CR LF
Hold output
0: OFF 1: ON
Stable output
0: OFF 1: ON
Near zero 0: OFF output 1: ON
HI decision 0: OFF output 1: ON
OK Output
0: OFF 1: ON
LO decision output
0: OFF 1: ON
HH decision output
0: OFF 1: ON
LL decision 0: OFF output 1: ON
TD-250T compatible command (RD)
Request Response
R D CR R D 0 0 0 0 0 0 0 CR LF
Hold output signal
0: OFF 1: ON
Stable output signal
0: OFF 1: ON
Near zero 0: OFF
output signal
1: ON
HI decision 0: OFF output 1: ON
OK output
0: OFF 1: ON
LO decision output
0: OFF 1: ON
Reserved
61
11. RS-232C Interface (TD-260T AC(232C) / TD-260T DC(232C))
11-8. Set commands
HI Value (W01)
W 0 1 0 0 0 0 0 0 CR LF
sign bit Plus: 0 Minus: –
Set value without decimal point (5 digits max)
LO Value (W02)
W 0 2 0 0 0 0 0 0 CR LF
sign bit Plus: 0 Minus: –
Set value without decimal point (5 digits max)
Compairson Mode (W03)
W 0 3 0 0 0 0 0 0 CR LF
0: A l w ay s co m p a re to g e n e r ate decision outputs
1: Compare to generate decision outputs when stabilized
2: A l w ay s co m p a re to g e n e r ate decision outputs except near zero
3: Compare to generate decision outputs when stabilized except near zero
4: Compare to generate decision outputs when in hold mode
5: Do not compare
Hysteresis (W04)
W 0 4 0 0 0 0 0 0 CR LF
Set value without decimal point (5 digits max)
Digital Offset (W05)
W 0 5 0 0 0 0 0 0 CR LF
sign bit Plus: 0 Minus: –
Set value without decimal point (5 digits max)
Near Zero (W06)
W 0 6 0 0 0 0 0 0 CR LF
Set value without decimal point (4 digits max)
HH Value (W07)
W 0 7 0 0 0 0 0 0 CR LF
sign bit Plus: 0 Minus: –
Set value without decimal point (5 digits max)
LL Value (W08)
W 0 8 0 0 0 0 0 0 CR LF
sign bit Plus: 0 Minus: –
Set value without decimal point (5 digits max)
Enable HH and LL (W09) W 0 9 0 0 0 0 0 0 CR LF
0: Disabel 1: Enable
Digital Filter (W11)
W 1 1 0 0 0 0 0 0 CR LF
0: OFF
1: 4 times 2: 8 times
3: 16 times 4: 32 times 5: 64 times
6: 128 times 7: 256 times
Analog Filter (W12)
W 1 2 0 0 0 0 0 0 CR LF
0: 3Hz 3: 100Hz
1: 10Hz 4: 300Hz
2: 30Hz
Motion Detection Time (W13) W 1 3 0 0 0 0 0 0 CR LF
Time
Motion Detection Bandwidth (W14) W 1 4 0 0 0 0 0 0 CR LF
Set value without decimal point (3 digits max) Zero Tracking Time Interval
(W15)
W 1 5 0 0 0 0 0 0 CR LF
Time
62
11. RS-232C Interface (TD-260T AC(232C) / TD-260T DC(232C))
Zero Tracking Bandwidth (W16) W 1 6 0 0 0 0 0 0 CR LF
External Hold Mode (W25) W 2 5 0 0 0 0 0 0 CR LF
Set value without decimal point (3 digits max)
Hold Mode (W17)
W 1 7 0 0 0 0 0 0 CR LF
0: Peak hold 1: Sample hold 2: Bottom hold 3: Peak-to-peak hold 4: Disable
hold
Digital Zero Limit (W18)
W 1 8 0 0 0 0 0 0 CR LF
Set value without decimal point (5 digits max)
Comparative Output Pattern (W19)
W 1 9 0 0 0 0 0 0 CR LF
0.nor: Standard output 1.ArE: Area output
Lock (W21)
W 2 1 0 0 0 0 0 0 CR LF
SET Lock
CAL Lock
Enable Digital Zero Enable Hold
0: OFF 1: ON 0: OFF 1: ON 0: OFF
1: ON
0: OFF 1: ON
Minimum Scale (W22)
W 2 2 0 0 0 0 0 0 CR LF
0: Level 1: Pulse
Remote Sense/TEDS (W26) (Write-protected, read only)
W 2 6 0 0 0 0 0 0 CR LF
0: TEDS auto-calibration 1: Enable TEDS /disable remote sense 2: Enable remote
sense/disable TEDS
Enable CLEAR Signal (W27)
W 2 7 0 0 0 0 0 0 CR LF
0: Disable 1: Enable
Enable Block Setting (W28) W 2 8 0 0 0 0 0 0 CR LF
0: Disable 1: Enable
Enable JUDGE Signal (W29) W 2 9 0 0 0 0 0 0 CR LF
0: Disable 1: Enable
BCD Data Refresh Rate (W31) W 3 1 0 0 0 0 0 0 CR LF
0: 1 times/Sec 2: 5 times/Sec 4: 20 times/Sec 6: 100 times/Sec
1: 2 times/Sec 3: 10 times/Sec 5: 50 times/Sec
RS-232C Interface Setting (W32) W 3 2 0 0 0 0 0 0 CR LF
0: 1
1: 2
2: 5
3: 10
Display Refresh Rate (W23) W 2 3 0 0 0 0 0 0 CR LF
0: 4 times/Sec 2: 10 times/Sec
1: 6 times/Sec 3: 20 times/Sec
Bridge Voltage (W24) (Write-protected, read only)
W 2 4 0 0 0 0 0 0 CR LF
0: 10V 1: 2.5V
0: Command
Tr ansmissio n transmission
mode
1: Continuous
transmission
0: 1200bps
1: 2400bps
Baud rate 2: 4800bps
3: 9600bps
4: 19200bps
Data bits
0: 7bits 1: 8bits
0: None
Parity bit
1: Odd
2: Even
Stop bit
0: 1bit 1: 2bits
63
11. RS-232C Interface (TD-260T AC(232C) / TD-260T DC(232C))
D/A Zero Setting (W33)
W 3 3 0 0 0 0 0 0 CR LF
sign bit Plus: 0 Minus: –
Set value without decimal point (5 digits max)
D/A Full-Scale Settings (W34) (Write-protected, read only)
W 3 4 0 0 0 0 0 0 CR LF
11-9. Set-value retrieval command
You can retrieve a set value by using the following command.
Request
W
Response
CR W
CR LF
Setup value No.
Setup value No. Same as data in set command
sign bit Plus: 0 Minus: –
Set value without decimal point (5 digits max)
BEEP (W35)
W 3 5 0 0 0 0 0 0 CR LF
0: OFF 1: ON
Select Data Output (W36) W 3 6 0 0 0 0 0 0 CR LF
0: Reference the display; hold value is sent in hold mode
1: Reference the input; input value is sent in hold mode
64
11. RS-232C Interface (TD-260T AC(232C) / TD-260T DC(232C))
11-10. Control Commands
Hold ON (CE)
C E CR (Only with a hold mode setting other than 0.OFF enabled)
Hold OFF (CF)
C F CR (Only with a hold mode setting other than 0.OFF enabled)
Digital Zero (CG)
C G CR (Only with both Enable Digital Zero and CAL Lock ON)
Reset digital zero (CH)
C H CR (With Enable Digital Zero ON)
Reset Hold (CI)
(Only in hold mode)
C I CR
JUDGE contact input ON (CJ)
C J CR (With Enable JUDGE Signal ON)
JUDGE contact input OFF (CK)
C K CR (With Enable JUDGE Signal ON)
11-11. Format for transmission mode 1
The TD-260T continuously transmits 20-byte data in the following format.
G S , 1 , 2 , 3 , 4 , ± 0 0 0 0 0 0 CR LF
sign bit Plus: 0 Minus: –
Reading
1 Displayed status O: Overflow (±LOAD,±Full) S: Stable M: Unstable H: Hold
Status notification priority if more than one status observed H > O > (S or M)
2 Zero tracking A: OFF T: ON
3 Decision output H: Hi decision output ON L: LO decision output ON O: HH
decision output ON U: LL decision output ON G: OK decision output ON (HI, LO,
HH, and LL decision outputs OFF) F: All decision outputs OFF Output
notification priority if more than one output observed (O or U) > (H or L or
G) > F
4 Near zero N: OFF Z: ON
65
12. D/A Converter (TD-260T AC(D/A) / TD-260T DC(D/A) )
The D/A output circuit is isolated from the TD-260T internal circuit. The
analog output ranges are 0 to ± 10 V in voltage and 4 to 20 mA in current.
Using the D/A zero and D/A full-scale functions, you can respectively obtain
analog output zero (0 V, 4 mA) and full scale (±10 V, 20 mA) for any digital
value you set. Notice that the output current for the output voltage 0 V is 4
mA and the output current for the output voltage 10 V is 20 mA. You cannot set
a value for the output current and output voltage separately when you change
the zero point or full scale setting. Resolution is ±1 per 10000 for 0 to ±10
V, and the conversion rate is 100 times per second. The D/A converter has an
overrange capability of about ±10 % on the output.
12-1. Panel
12-2. Connections
Connect an external device with a load resistance of 2 k or more to the
voltage outputs V + and V -. Connect an external device with a load resistance
of 350 or less, including a cable resistance, to the current outputs I + and I
-.
TD-260T
+ V
+ I
shield shield
External + device
2k or more of load
GND resistance
External device
350 or less of load GND resistance
12-2-1. Connecting to screwless terminal block
1
To connect wires to the terminal block, use the provided
2
flathead micro screwdriver or equivalent.
3
1. Strip off about 10-12 mm of the insulation from an
4
end of each wire you connect, and then gently twist
each bare end.
Use wires 0.4-1.2 mm² in cross-sectional area.
2. Press the button next to the corresponding wire insertion hole by using the screwdriver.
3. While keeping each wire end twisted, insert it into the corresponding wire insertion hole.
4. Release each button.
1 Voltage output V +
5. Lightly pull each wire and make sure that they are securely clamped.
2 Voltage output V 3 Current output I + 4 Current output I –
NOTE
To reduce the unnecessary radio frequency emission, be sure to install the
provided ferrite core on the output cable (voltage or current).
66
12. D/A Converter (TD-260T AC(D/A) / TD-260T DC(D/A) )
12-3. D/A Zero
You can set a value so that the TD-260T generates a D/A zero output (voltage 0
V and current 4 mA) when it displays the same reading as the value you set. 1.
Enter the setup screen number F4-4.
Press 4 times
12-4. D/A Full Scale
You can set a value so that the TD-260T generates a D/A full scale output
(voltage 10 V and current 20 mA) when it displays the same reading as the
value you set. 1. Enter the setup screen number F4-5.
Press 4 times
Press 3 times
Press 4 times
2. To change the blinking digit of the value, press 5 or b. To move to the
previous or next digit, press g or t.
To move to the previous or next digit, press:
2. To change the blinking digit of the value, press 5 or b. To move to the
previous or next digit, press g or t.
To move to the previous or next digit, press:
To change the digit, press:
To change the digit, press:
3. To save the setting, press ENTER, and then to exit the setup mode, press ESC.
3. To save the setting, press ENTER, and then to exit the setup mode, press ESC.
To save, press:
To save, press:
CAUTION
The rated capacity is set as the D/A full-scale value when you save the rated
capacity during calibration.
67
12. D/A Converter (TD-260T AC(D/A) / TD-260T DC(D/A) )
Setting example 1 D/A Zero D/A Full Scale Reading 100.00
000.00
000.00 100.00 D/A Output 10V(20mA)
0V(4mA)
12-5. D/A output error
Below are the messages that are displayed when errors are caused by the D/A
option.
dErr: the D/A output is higher than +10.9 V. -dErr: the D/A output is lower
than -10.9 V.
100.00
Setting example 2 D/A Zero D/A Full Scale Reading 120.00
020.00
10V(–mA)
020.00 100.00 D/A Output 10V(20mA) 0V(4mA)
80.00
Setting example 3 D/A Zero D/A Full Scale Reading 120.00
020.00
10V(–mA)
020.00 -100.00 D/A Output 10V(20mA) 0V(4mA)
80.00
Setting example 4 D/A Zero D/A Full Scale Reading 100.00
000.00
10V(–mA)
-10.00 20.00 D/A Output 10V(20mA) 5V(4mA)
10.00
0V(–mA)
68
The NDIS panel is an input connector panel that conforms to the Non-
Destructive Inspection Standard (NDIS) (established by the Japanese Society
for Non-Destructive Inspection). You can connect a sensor with an NDIS
connector to the panel.
1
13. NDIS Panel (PN-260ND)
CAUTION
< The terminals F and G are used as either remote sense terminals or TEDS
sensor data terminals. Before connecting a sensor, you need to set Remote
Sense/TEDS (F2-6) to match the function you are going to use. The initial
setting is 0.Aut: TEDS auto-calibration (disable remote sense).
< If you use a 4-wire connection, but you do not connect a TEDS sensor, do not
connect any device to the terminal F or G.
< If you connect a sensor to the NDIS connector on the panel, do not connect
any sensor to the signal input/output terminal block on the TD-260T.
1 NDIS connector
Pin assignmentment A B C D E F G
Signal name +EXC -SIG -EXC +SIG
SHIELD(FG) TEDS/+SENS GND/-SENS
Applicable plug: PRC03-12A10-7M10.5 manufactured by Tajimi Electronics Co., Ltd.
69
14. About TEDS
Connecting a sensor that conforms to IEEE1451.4 (Transducer Electronic Data Sheet (TEDS)) to the TD-260T, you can load the rated output values that are saved on the sensor into the TD260T and use the values for calibration with the TD-260T. You can also use the TD-260T to save on your TEDS sensor values that are used for calibration with the TD-260T or to restore the TEDS sensor’s factory settings. TEDS memory is available in 1-kilobit and 4-kilobit sizes. Note that the TD-260T supports the 4-kilobit memory only.
TEDS standard IEEE1451.4(V1.0)
IEEE 1451.4 (V0.9)
Bridge Sensors
(33)
Template ID
Strain gage (35)
Others
: TEDS calibration, Save TEDS Data (F4-6), and Restore TEDS Data (F4-7) all
supported
: TEDS calibration supported : None of the above functions supported
14-1. Save TEDS Data
You can save the current calibration values into the TEDS memory. 1. Enter the
setup screen number F4-6.
Press 4 times
Press 5 times
3. Press ENTER. The TD-260T displays “tEdS” and starts verifying data in the
TEDS memory,
4. To enter a calibration date; enter a year by pressing g, t, 5, or b, and
then press ENTER. For month and day settings, too, press ENTER after you enter
values. When the date setting is completed, the TD-260T starts saving the
calibration values. Enter year and press
Enter month and press
Enter day and press
NOTE If you press ESC while “tEdS” is not displayed, the setting is
interrupted, and then the TD-260T exits the setup mode.
5. The display shows “tEdS” during the save process.
2. To enter 00015, change the blinking digit of the value, press 5 or b. To
move to the previous or next digit, press g or t.
To move to the previous or next digit, press:
6. The TD-260T displays “F4-6,” showing that the save process is completed, and then waits for a key to be pressed. To exit the setup mode, press ESC.
To change the digit, press:
NOTE
< Be sure to enter this value to prevent the original TEDS data from being
overwritten by mistake.
< If you press ESC while “tEdS” is not displayed, the setting is interrupted,
and then the TD-260T exits the setup mode.
70
14-2. Restore TEDS Data
You can restore the factory calibration setting (“original TEDS data”) into
the TEDS memory after saving data by using the function explained in the
previous section 14-1. “Save TEDS Data.” 1. Enter the setup screen number
F4-7.
Press 4 times
14. About TEDS
4. The TD-260T displays the rated output (mV/V), showing that the load
operation is completed. Check the values, and then press ENTER.
Press 6 times
2. To enter 00015, change the blinking digit of the value, by pressing 5 or
b. To move to the previous or next digit, press g or t. To move to the
previous or next digit, press:
NOTE
If you press ESC while “tEdS” is not displayed, the setting is interrupted,
and then the TD-260T exits the setup mode.
5. The rated capacity is displayed. Check the value, and then press ENTER.
To change the digit, press:
NOTE
< Be sure to enter this value to prevent the original TEDS data from being
overwritten by mistake.
< If you press ESC while “tEdS” is not displayed, the setting is interrupted,
and then the TD-260T exits the setup mode.
3. Press ENTER. The TD-260T displays “tEdS” and starts loading the oritinal
TEDS data from the TEDS memory.
NOTE
If you press ESC while “tEdS” is not displayed, the setting is interrupted,
and then the TD-260T exits the setup mode.
6. Press ENTER. The TD-260T displays “tEdS” and starts saving the data into
another area of the TEDS memory.
7. The TD-260T displays “F4-7,” showing that the save process is completed,
and then waits for a key to be pressed.
To exit the setup mode, press ESC.
71
15. Available functions at turn-on
You can check operations of the TD-260T when turning it on.
15-1. Performing self-check
This function allows the TD-260T to perform a self-diagnostic check of LEDs,
RAM, SRAM, EEPROM, and TEDS, in this order. 1. While pressing and holding ESC,
turn the TD-260T on.
6. The TD-260T displays the following message during the TEDS check process.
Turn the TD-260T on while holding:
7. The TD-260T displays a reading after all the check processes are completed.
2. The TD-260T displays sets of the same numbers, such as 88888, to check LEDs, one after another.
3. The TD-260T displays the following message during the internal RAM check
process.
4. The TD-260T displays the following message during the external SRAM check
process.
NOTE
The TD-260T displays one of the following messages when an error is detected:
cErr0, cErr1, cErr2, and cErr3. If you encounter any one of the errors, shut
off the power to the TD-260T, and then, while pressing and holding ESC, turn
it on again to perform a self-check. If the error continues to occur, see the
last page of this document and contact us.
15-2. Initializing settings
You can initialize settings all at once to restore the factory preset values.
- While pressing and holding ESC and ENTER, turn the
TD-260T on.
Turn the TD-260T on while holding
&
5. The TD-260T displays the following message during the EEPROM check process.
After the settings have been initialized, a self-check starts. NOTE
Calibration values are not initialized.
72
16. Specifications
Bridge voltage (BV)
10 V DC, 2.5 V ±10 % (Current: 60mA max. Remote sensing available)
Singal input range
±3.2 mV/V
Equivalent input/ TEDS
Calibration range
Calibration accuracy
0.3 mV/V to 3.2 mV/V Within 0.1% FS (at 10V for BV and 0.5mV/V or more for sensor sensitivity)
Zero adjustment range
±2.0 mV/V
Linearity
Within 0.01% FS + and 1 digit (at an input of 1m V/V or more)
Accuracy
Zero drift
0.5 V/°C or less (RTI)
Gain drift
Within ±0.005% FS/°C
A/D conversion rate
100 times/sec
Analog filter
Selectable from 3, 10, 30, 100, 300 Hz (-12 dB/oct)
Analog voltage output
Voltage Load register
Approx. 2 V for 1 mV/V 2 k or more
TEDS function
IEEE1451.4 Class 2 Mixed-mode interface Note that the TEDS function cannot be used when remote sensing of excitation power supply is used.
Characters
14.6mm high alphanumeric, shown by 7-segment green LED display
Display
Range Decimal point
-19999 to 99999 Position to be selectable
Refresh rate
4, 6, 10, 20 times/sec
Calibration setting Zero CAL/Span CAL (TEDS CAL, Actual load CAL, Equiv. input CAL)
Displayable items
Function settings
HI value, LO value, HH value, LL value, Comparison mode, Hysteresis, Near- zero, Digital filter, Analog filter, Motion detection, Zero tracking, Static strain, Digital zero, Zero offset, Hold mode, Key lock, Minimum scale, Refresh rate, Bridge voltage, B.C.D data refresh rate, RS232C, D/A converter
Hold functions
Sample hold, Peak hold, Bottom hold, Peak-to-Peak hold, block-specified hold (Peak, Bottom, Peak-to-Peak)
External input/ output signal
Input: Output:
Hold, Decision, Clear, Digital zero 1a contact for each relay: HH, HI, OK, LO, LL, analog voltage
Power supply
AC power rating: DC power rating:
100 to 240 V ±10% 5060 Hz AC, 13 W 10 to 28 V DC, 8 W
Operating temperature range
-10°C to 40°C
Storing temperature range
-20°C to 60°C
Operating humidity range
85% RH or less (no condensation)
Standard
Safety standard EMC
Installation class I, EN61010-1 Pollution degree 2 Overvoltage category II, CE marking VCCI (Class A), EN61326 (Class A), EN61000-3-2, EN61000-3-3
External dimensions (W x H x D)
Approx. 96 mm x 96 mm x 146 mm (protrusions not included)
Weight
Approx. 950 g
Option (Not used for option-installed models) NDIS panel (PN-260ND)
Option-installed models TD-260T AC(BCD) / TD-260T DC(BCD) TD-260T AC(232C) /
TD-260T DC(232C) TD-260T AC(D/A) / TD-260T DC(D/A)
73
12
17. Outline drawings
96
Dimensional drawing of panel cut-out (recommended panel thickness: 1.6-3.0 mm) Panel cut-out
121
92 +0.8 0
Standard: DIN IEC61554, published in 2002
13
91 +0.8 92
0
96
96
91
< Specifications and appearance are subject to change for improvement without notice. < Some of the images shown in this document may differ from the actual product for product improvement.
74
18. Block Diagram
A
D
B
C
Strain gage amplifier Gain regulator circuit Buffer Analog output Filter circuit Voltage shift circuit TEDS memory Bridge power supply Zero regulator circuit Reference voltage line
F.G
AC power type
+12V
12V
+5V +3.3V
TEDS I/F
LED indicator
EEPROM (4kbit)
Power supply unit
FRAME
DC power type +12V 12V
+5V +3.3V
L AC IN
N
F.G
SRAM (128kByte)
I/O signal control connector
Power supply unit FRAME
+ DC IN
F.G
A/D conversion
CPU
RS232C I/F
BCD I/F
D/A I/F Option card
75
Z
TEAC CORPORATION
TEAC AMERICA, INC.
TEAC EUROPE GmbH. (EU Importer)
TEAC SALES & TRADING (ShenZhen) CO., LTD.
1-47 Ochiai, Tama-shi, Tokyo 206-8530, Japan 10410 Pioneer Blvd. Suite #1,
Santa Fe Springs, California 90670, U.S.A. Bahnstrasse 12, 65205 Wiesbaden-
Erbenheim, Germany
Room 817, Xinian Center A, Tairan Nine Road West, Shennan Road, Futian
District, Shenzhen, Guangdong Province 518040, China
Phone: +81-42-356-9154
Phone: +1-323-726-0303
Phone: +49-(0)611-7158-349
Phone: +86-755-88311561-2
0222.MA-1699H
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>