TEAC TU-NR-C-G Strain Gauge Load Cell Instruction Manual

June 12, 2024
TEAC

TEAC TU-NR-C-G Strain Gauge Load Cell

Product Information

  • Product Name: TU-NR-C-G Tension/Compression Load Cell
  • Model Number: D01396001A
  • Product Description: The TU-NR-C-G load cell is a strain gauge load cell designed for both tension and compression applications. It is used to measure the load or force applied to an object. This load cell provides accurate and reliable measurements, ensuring safe and proper operation.
  • Included Accessories: Please check the package contents upon purchase. If anything is missing or damaged, kindly contact the retailer where you purchased the product.

Product Usage Instructions

Important Safety Instructions:

  • WARNING: If any abnormality occurs, do not attempt to open or alter the unit. Contact the retailer for repair. Avoid placing foreign objects or liquids into the unit as it may cause malfunction.
  • CAUTION: Ensure proper installation in suitable locations. Do not expose the unit to smoke, steam, excessive humidity, direct sunlight, or unstable surfaces. Cut off power supply when not in use for a long time. Do not operate a damaged unit.

Introduction

Thank you for purchasing the TU-NR-C-G load cell. Please read this document completely before using this load cell to achieve its best performance and ensure safe and proper operation.

  • Company names and product names in this document are the trademarks or registered trademarks of their respective owners.

Included accessories

If anything is missing or damaged, contact the retailer where you purchased the product.

  • Test report × 1
    • Instructions for Use (this document) × 1

IMPORTANT SAFETY INSTRUCTIONS

WARNING

  • If something abnormal occurs
    Request repair from the retailer where you purchased the product.

  • Do not open the cover.
    Never remove the cover from this unit. Doing so could cause malfunction. Request inspection and repair from the retailer where you purchased the product. Do not alter this unit. Doing so could cause malfunction

  • Do not put foreign objects or water, for example, into the unit.
    Do not place a container that holds water, for example, on top of this unit. If liquid is spilled, for example, and enters the unit, this could cause malfunction.

  • Do not use the unit with any power supply voltage other than that specified.
    Do not use the unit with any power supply voltage other than that specified. Doing so could cause malfunction.

CAUTION Unsuitable installation locations
Do not place the unit in the following types of locations. Doing so could cause malfunction.

  • Locations where it might be exposed to smoke or steam, such as near a kitchen table or humidifier
  • Unstable locations, including unsteady stands and tilted places
  • Locations that are very humid or dusty
  • Locations that are exposed to direct sunlight

When not using the unit for a long time
For safety, cut the power supply when not using this unit for a long time. Do not operate a damaged unit.

Precautions for use

  • This unit is not built to be water or splash resistant, and it cannot be used in conditions when the relative humidity is high. Moreover, use in atmospheres with corrosive gases should be avoided.
  • Be careful to prevent water, oil and other substances from getting on the unit.
  • Avoid use in conditions where condensation could occur.
  • Connect cores to the load cell after discharging (eliminating) static electricity from your body.
  • If the surrounding temperature changes suddenly, the values output by this device could become unstable, making accurate measurement impossible. (This could occur, for example, in a location blown by warm or cold air.)
  • Conduct load calibrations periodically.

Installation procedures

  • Install this unit in a place where the structure is level and can sufficiently bear the load being used. Output precision could be affected if the attachment surface does not meet parallelism and flatness requirements.
  • Refer to Figure 1 for use with only compression loads. Refer to Figure 2 for use with tension/compression loads.
  • Attach a load button when using with compression. When using with tension, attach a load rod, for example.

Figure 1. Compression Figure 2. Compression/ tension

  • Insert screws (steel hexagon socket head cap screws) into the screw holes near the edge of the load cell, and fix it to the flat attachment surface. (Output precision could be affected if the attachment surface is not flat.) The size of the screws used and their number depends on the rated capacity (model name) of the load cell as shown below.

Screw types and tightening torques

Rated capacity Center screw section Flange bolt section
Tightening torque Dimension of load cell

screw hole

| Tightening torque| Dimension of load cell

screw hole

1 kN| 96 N·m| M12 × 1| 10 N·m| M6
2 kN| 96 N·m| M12 × 1| 10 N·m| M6
5 kN| 96 N·m| M12 × 1| 10 N·m| M6
10 kN| 96 N·m| M12 × 1| 10 N·m| M6
20 kN| 96 N·m| M12 × 1| 10 N·m| M6
50 kN| 400 N·m| M20 × 1.5| 25 N·m| M8
100 kN| 1.3 kN·m| M32 × 2| 43 N·m| M10
200 kN| 3.2 kN·m| M40 × 2| 190 N·m| M16

Use screws with a strength grade of 12.9.
In addition, the length of thread engagement for the screw and the mounting part (internal threads) should be about twice the nominal diameter of the screw. Please consult with us if you are uncertain about screw selection, for example.

  • Tighten with the torque shown in the “Screw types and tightening torques” table to prevent the load button or load rod from becoming loose. Be sure not to apply torque that exceeds the tightening torque shown in the table even when removing an attached load button or load rod.
  • A thread-locking fluid can be used on the screw, but be aware that this could prevent removal of the load button or load rod.
  • When connecting a load button or load rod to this unit (or disconnecting one), check the values in the Values Table of the Specifications.
  • When using with tension, consider the detent of all the screws, and consider implementing inspections before starting work, for example.
  • In addition, also consider examinations to prevent accidents caused by falling objects due to damaged load cells resulting from overloads and unbalanced loads, for example.

Precautions when placing loads on the unit

  • Make sure the load is perpendicular to the surface to which this unit is attached.
  • Apply the load so that it is centered on the center of the unit. If the load is not centered (eccentric load), twisting, for example, and measurement errors could occur. This could even result in damage.
  • If the load receiving area is contacted by something that is at a different temperature and the load is increased, the values output by this device could become unstable, making accurate measurement impossible. In such a case, wait until the temperature difference ceases to exist before measuring.
  • Do not apply any load to the area shown by shading with diagonal lines in the illustration below. Applying aload to the shaded area will break the load cell.

Electrical connection

Electrical connection of load cell
Connect as shown in the illustration below. Incorrect connections could result in inability to balance and in errors occurring in the output voltage when loads are applied.
Receptacle in device

  • This unit does not support remote sense. See the operation manuals of indicators and strain amps that support remote sense for how to connect sensors with those units.
  • The shield is connected to the main body of this product. For this reason, if grounding is necessary because of external noise or another issue, arrange to ground  the shield to a part other than the body of this unit,for example.
  • Use a specialized cable when replacing the cable or to increase the length. (Please consult with us.)

When included cable connected

Handling after use

  • When moving this unit while it is attached, take protective measures to prevent it from being shaken or subjected to excessive external forces.
  • When storing it, keep it in a dry place where it will not be exposed to water or oil, for example.

Specifications

  • TU-NR-C 1KN-G to TU-NR-C 200KN-G
  • Rated capacity: 1 kN, 2 kN, 5 kN, 10 kN, 20 kN, 50 kN, 100 kN, 200 kN
  • Safe overload rating: 150% R.C.
  • Rated output: 0.75 mV/V ±1% (1 kN)
    • 1.0 mV/V ±1% (2 kN)`
    • 1.5 mV/V ±1% (5 kN to 200 kN)
  • Linearity: 0.15% R.O.
  • Hysteresis: 0.15% R.O.
  • Repeatability: 0.1% R.O.
  • Zero balance: 5% R.O.
  • Safe excitation voltage: 20 V (12 V recommended)
  • Input terminal resistance: 350 Ω ±1%
  • Output terminal resistance: 350 Ω ±1%
  • Insulation resistance: 1000 MΩ or more (DC 50 V)
  • Compensated temperature range: −10 to 60°C
  • Permissible temperature range: −30 to 80°C
  • Temperature effect on zero balance:
    • 0.1% R.O./10°C (1 kN to 2 kN)
    • 0.05% R.O./10°C (5 kN to 200 kN)
  • Temperature effect on output: 0.1% R.C./10°C
  • Connector: PRC03-21A10-7F
  • Cable: Ø8, 4-core shielded, 5m, bare lead wires for one end, PRC03-12A10-7M for other end
  • Body material: alloy tool steel
  • Environmental compliance: RoHS (10 substances)
  • TEDS: not supported

Dimensional

Dimensional drawings

Rated capacity| A| ØB| C| ØD| ØE| ØF| ØG

(H7)

| H| J| K| L| M| N| ØP| ØQ
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---
1 kN| 102 kgf| 25| 65| 22| 70| 41| 18| 14| 0.5| 2| 55| 22| 50| M12 × 1| 52| 6.5
2 kN| 204 kgf| 25| 65| 22| 70| 41| 18| 14| 0.5| 2| 55| 22| 50| M12 × 1| 52| 6.5
5 kN| 510 kgf| 25| 65| 22| 70| 41| 18| 14| 0.5| 2| 55| 22| 50| M12 × 1| 52| 6.5
10 kN| 1.02 tf| 25| 65| 22| 70| 41| 18| 14| 0.5| 2| 55| 22| 50| M12 × 1| 52| 6.5
20 kN| 2.04 tf| 30| 65| 22| 70| 41| 18| 14| 0.5| 2| 55| 22| 50| M12 × 1| 52| 6.5
50 kN| 5.1 tf| 30| 88| 27| 92| 60| 30| 22| 1| 2| 67| 22| 50| M20 × 1.5| 74| 9
100 kN| 10.2 tf| 34| 117| 31| 121| 82| 46| 34| 1| 2| 81| 22| 50| M32 × 2| 100| 11
200 kN| 20.4 tf| 50| −| −| 166| 116| 60| 44| 1| 2| 117| 40| 70| M40 × 2| 142| 17

Values Table

Model

| Allowable bending moment| Allowable lateral load| Allowable torque| Natural frequency

(kHz)

| Weight (kg)
---|---|---|---|---|---
Extent of precision

compensation

| Extent that will not cause

mechanical damage

| Extent of precision

compensation

| Extent that will not cause

mechanical damage

TU-NR-C 1KN-G| 0.6 N·m| 8.4 N·m| 58.8 N·m| 2.7 kN·m| 2.1 N·m| 6.5| 0.6
TU-NR-C 2KN-G| 1.1 N·m| 16.3 N·m| 98.0 N·m| 4.1 kN·m| 3.9 N·m| 8| 0.6
TU-NR-C 5KN-G| 1.7 N·m| 25.3 N·m| 166.7 N·m| 6.8 kN·m| 11.4 N·m| 11| 0.6
TU-NR-C 10KN-G| 3.7 N·m| 53.8 N·m| 323.6 N·m| 14.4 kN·m| 51.5 N·m| 16| 0.6
TU-NR-C 20KN-G| 10.4 N·m| 148.9 N·m| 657.0 N·m| 29.9 kN·m| 164.6 N·m| 21| 0.7
TU-NR-C 50KN-G| 13.6 N·m| 195.2 N·m| 1.6 kN·m| 30.1 kN·m| 392.0 N·m| 18| 1.1
TU-NR-C 100KN-G| 48.1 N·m| 687.6 N·m| 3.2 kN·m| 63.9 kN·m| 1.9 kN·m| 16| 2.2
TU-NR-C 200KN-G| 190.7 kN·m| 2.7 kN·m| 6.5 kN·m| 175.2 kN·m| 7.9 kN·m| 12| 6.0

Contact

Contact information

  • TEAC CORPORATION (Manufacturer)
    • 147 Ochiai, Tama-shi, Tokyo 206-8530 Japan
    • Phone: +81-042-356-9154
  • TEAC AMERICA, INC.
    • 10410 Pioneer Blvd. Unit #1, Santa Fe Springs, California 90670, U.S.A.
    • Phone: +1-323-726-0303
  • TEAC EUROPE GmbH. (EU Importer)
    • Bahnstrasse 12, 65205 Wiesbaden-Erbenheim, Germany
    • Phone: +49-611-7158-349
  • TEAC UK Limited (UK Importer)
    • Luminous House, 300 South Row, Milton Keynes, Buckinghamshire, MK9 2FR, UK
    • Phone: +44-1923-797205

Warranty explanation

The warranty period for this device is one year from the date of purchase. Be aware that repairs will require payment in the following cases even during the warranty period.

  1. Malfunction or damage due to misuse
  2. Malfunction or damage caused by modifications or repairs conducted by any party other than our company or a service person designated by our company
  3. Malfunction or damage caused by dropping, transportation or similar handling after product delivery
  4. Malfunction or damage caused by fire, earthquake, water, lightning or other natural disaster
  5. Malfunction or damage caused by external factors, including power supplies and equipment environmental conditions, that deviate from the operation requirements of this product
  6. Malfunction or damage if the product was not purchased from our company or an agent designated by our company
    • We offer paid service after the conclusion of the warranty period. For details, please contact the retailer where you purchased the unit.
    • Be aware that our company will bear no responsibility for any secondary damages resulting from the operation of this device or related to data.
    • Information is given about products in this manual only for the purpose of example and does not indicate any guarantees against infringements of third-party intellectual property rights and other rights related to them.
    • TEAC Corporation will bear no responsibility for infringements on third-party intellectual property rights or their occurrence because of the use of these products.

Strain gauge load cell Instructions for Use D01396001A

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals