Inside Installation
Whenever possible,
EOGas Sterilizers should be located near an outside wall, and exhaust directly
to the outside via the included exhaust ventilator. AN 333
Sterilizers (shown) and AN306 and AN310 Sterilizers on base cabinets have
mounted casters providing convenient mobility
Installation with Emissions Abato
Alternative Installations to the Outside
- Please note: AAMI guidelines recommend that all EtO sterilizers should be vented (exhausted) through a dedicated vent line to the outside
atmosphere directly or through an emission control system. The vent should not terminate within 25 feet (7.6m) of any building air intake source or near pedestrian or animal traffic, nor should it allow EtO release into or to reenter the building