UT345A Ultrasonic Thickness Gauge Operating Manual
Overview
This product is an intelligent ultrasonic thickness gauge that adopts the latest high performance and low power consumption microprocessor technology. Based on the principle of ultrasonic measurement, it can measure the thickness and sound speed of metal and other materials. It can detect the thinning of pipes and pressure vessels after corrosion, and can make accurate measurements of plates and machined parts. The gauge can be widely used in petroleum, chemical, metallurgy, shipbuilding, aviation, aerospace and other fields.
1.1 Operating Principle
When measuring the thickness, the ultrasonic pulse is generated by the probe
to reach the measured body through the coupling agent. Part of the ultrasonic
signal is reflected from the bottom of the object, and the probe receives the
reflected echo. The gauge calculates the round trip time of the ultrasonic
wave and the thickness value according to the following formula, and then
displays the calculated results. In the formula: H-Measured thickness;
V-Material sound velocity; t-Round trip time of ultrasonic wave in specimen
1.2 Specifications
Display | 2.4″ monochrome dot matrix screen |
---|---|
Language | English/Chinese |
Measurement range | 1.00-300.00mm |
Sound velocity range | (1000^-9999) m/s |
Unit | 0.1mm/0.01mm/0.01in |
Accuracy | H<10mm, ±0.1mm, H is the actual |
thickness;
H ..– 10mm, ±(1%H+0.1)mm
Lower limit of pipe measurement| 020x3mm (steel)
Data storage| 300 groups, including material, sound
velocity, unit, measured value, MAX,
MIN and AVG of each group
Main functions| Single measurement, continuous
measurement, alarm, sound velocity
measurement, auto calibration
Battery| Built-in 3.7V 2000mAh lithium battery
Charging specification| DC5V 1A, Type-C interface
Operating time (full battery)| About 16h
Operating temperature and humidity| 0°C-40°C 10%RH-80%RH
Storage temperature and humidity| -10°C-50°C 10%RH-70%RH
Product size| 140x66x28.5mm
Note:
-
Oil coupling agent cannot be used for the standard probe, otherwise the probe will be damaged.
After using the probe, please wipe the residual coupling agent on it to extend its service life. -
Please avoid using the product in oily environment as far as possible. If unavoidable, minimize the contact time with oily substances, and wipe the probe surface with oil-absorbing paper after use.
Structure and Appearance
Product features
A. 2.4″ monochrome dot matrix screen
B. English/Chinese interface
C. Built-in 2000mAh lithium battery
D. DC5V/1A, Type-C charging interface
E. 300 groups of data storage
- Outer shell
- Buttons
- Display
- Transmitting socket 5. Receiving socket
- Standard thickness block
- Probe
2.1 Main Interface
After the gauge is turned on, it will automatically enter the main interface,
as shown below:
- Coupling state: The coupling state of the probe and the measured workpiece
- Mode: Display the current measurement mode
- Unit: mm (metric), or in (imperial)
- Battery status: Display the remaining battery power
- Sound: Alarm sound on/off
- Information display: Display measured thickness, material, sound velocity, MAX, MIN and AVG
Battery Charging and Maintenance
The product is equipped with a non-removable 3.7V 2000mAh lithium battery.
When the product cannot be turned on or the battery is empty, please charge it
in time.
Charge the product with a DC5V adapter larger than 1A and the charging
interface is Type-C.
When charging, the battery icon scrolls . When fully charged, the battery icon
becomes full .
Note: When not in use for a long time, the product should be fully charged and recharged every six months to avoid battery damage.
General Measurement Process
A. Prepare the object to be measured, refer to 5.1 Treatment of Measured
Workpiece Surface.
B. Insert the probe and turn on the gauge.
C. Gauge calibration, refer to 6.5 Calibration.
D. Sound velocity measurement. The sound velocity of the workpiece is
uncertain, refer to 6.6 Sound Velocity Measurement.
E. Measure the thickness. Measure the thickness of the workpiece of the same
material with the sound velocity.
Preparation
5.1 Treatment of Measured Workpiece Surface If the measured object surface is
rough or seriously rusted, please use the following methods:
●Use coupling agent on the measured object surface
●Use rust remover, wire brush or sandpaper to treat the surface
●Multiple measurements near the same point
Operation
6.1 Power On/Off
Insert the probe into the gauge and short press to power on.Long press to
power off.
Auto power off: The product will shut down automatically when there is no
operation. The default time is 5 minutes.
Users can select the auto power off time, refer to 6.18 Power Off.
Force shutdown: Long press the Power button for more than 10s.
6.2 Single Measurement
Evenly smear the coupling agent on the area to be measured. Tightly couple the
probe to the material surface, and the thickness of the measured area will
show on the screen. When the probe is well coupled to the material, the screen
will display the coupling symbol . If the coupling symbol flickers or there is
no coupling symbol, it indicates that the coupling condition is bad and the
coupling agent needs to be added. After the probe is removed, the coupling
symbol disappears and the thickness value remains.
Figure 1 Add the coupling agentFigure 2 Evenly smear the coupling
agent and tightly couple the probe to the material surface. 6.3 Continuous Measurement
Evenly smear the coupling agent on the area to be measured and couple the
probe to the material surface.
Move the probe along the surface. The gauge will display the current measured
value, MAX, MIN and AVG in real time.
6.4 Alarm
Users can monitor the unqualified materials with the alarm function. Long
press to turn on/off the alarm.
If is displayed on the screen, the alarm is turned on.
If is displayed on the screen, the alarm is turned off.
For example:
Standard value: 4.00mm, limit: ±0.1mm. When the measured thickness is less
than 3.90mm or greater than 4.10mm, the gauge will give an alarm sound,
indicating that the thickness is not qualified. Users can press any button to
exit the alarm, or continue measuring.
6.5 Calibration
Note: Calibration should be carried out when the probe is replaced, the
ambient temperature changes greatly or the measurement deviation occurs.
Calibrate the gauge if it is used for the first time or not used for a long
time. Calibration must be done using the standard thickness block. After
calibration, the material automatically changes to steel, and the sound
velocity changes to 5920m/s (0.233in/us).
Calibration steps:
-
Long press to enter the calibration mode.
“Please Calibrate” will show on the screen. -
Refer to Figure 1 and Figure 2 in 6.2 Single
Measurement to evenly smear the coupling agent on the surface of the standard thickness block. Press the probe on the coupling agent to make it tightly contact with the block surface. (Note: the coupling agent should not be too thick. Otherwise the gauge will count the thickness of the coupling agent, resulting in calibration errors). -
Press to save the calibration data. (Note: Before pressing , the value displayed might not be 4mm.
When the value is stable, press . After calibration, the displayed value should be within the range of 4±0.04mm, otherwise recalibrate).
6.6 Sound Velocity Measurement
Different materials have different sound velocities. The materials and
velocities listed in Appendix A are for r eference only.This function is used
to measure the sound velocities of materials. The steps are as follows:
- First calibrate the gauge.
- Use a caliper to measure the thickness of the workpiece.
- Use the gauge to measure the thickness of the workpiece.
- Press or to make the thickness measured by the gauge the same as the value measured by the caliper.
- Press to calculate the workpiece sound velocity.
- At this point, the correct sound velocity can be obtained, and the correct thickness can be obtained by measuring the same material with the sound velocity.
Note: The thickness adjustment range is ±20mm (±0.79in) (for customized
materials, ±250mm (±9.84in)).
The upper limit is 300mm, and the lower limit is 1mm.
In case of misoperation, users can refer to 6.7 Sound Velocity Selection.
6.7 Sound Velocity Selection
Users can select the sound velocity according to the material. When the
current unit is mm, the unit of sound velocity is m/s. When the unit is in,
the unit of sound velocity is in/us. In the main interface, short press to
select the material. The gauge has three types of custom materials. If users
do not know which material to select, they can select the custom materials and
then measure the sound velocity of the material, refer to 6.6 Sound Velocity
Measurement.
Material selection interface
- Press to step through the options upwards.
- Press to step through the options downwards.
- Press to confirm the selected material and enter the sound velocity adjustment interface.
- Press to return to the main interface.
Sound velocity adjustment interface
-
Press to add the value. Long press it to scroll up.
-
Press to subtract the value.
Long press it to scroll down. -
Press to confirm the sound velocity and return to the main interface.
-
Press to return to the material selection interface.
6.8 Clear Measured Results
In the single and continuous measurement interfaces, short press to clear the
current measured results (including MAX, MIN and AVG). In the continuous
measurement mode, the results can only be cleared after the probe is removed.
6.9 Data Storage
In the single and continuous measurement interfaces, long press to save
records. Each record includes the current measured value, MAX, MIN, AVG and
material sound velocity. A maximum of 300 groups of data can be stored. To
view and delete records, refer to 6.14 Records. In the continuous measurement
mode, the data can only be saved after the probe is removed to avoid storing
unstable data.
6.10 Menu
Long press to enter the menu. The options include Material, Speed, Unit,
Records, Setting, About and Reset.
- Press to step through the options upwards.
- Press to step through the options downwards.
- Press to enter the interface of the selected option.
- Press to exit the menu and return to the main interface.
6.11 Material
In the material selection interface:
- Press to step through the materials upwards
- Press to step through the materials downwards.
- Press to confirm the selected material and return to the menu.
- Press to return to the menu.
6.12 Speed
The default value is ±200m/s (0.008in/us). The adjustment range of custom
materials is 1000m/s to 9999m/s (0.039in/us to 0.394in/us).
-
Press to add the value. Long press it to scroll up.
-
Press to subtract the value.
Long press it to scroll down. -
Press to save and return to the menu.
-
Press to return to the menu.
6.13 Unit
In the unit selection interface:
- Press to step through the units upwards.
- Press to step through the units downwards.
- Press to save and return to the menu.
- Press to return to the menu.
6.14 Records
In the menu, select “Records” and press to enter the records interface. When
there is no record, the screen will display “No records”, and the gauge cannot
enter the records interface.
- Press to step through the options upwards.
- Press to step through the options downwards.
- Press to enter the interface of the selected option.
- Press to return to the menu.
6.14.1 First Page
Select “First page” and the screen will display the records from the first
page. The number and value of each record will also be displayed.
- Press to page up.
- Press to page down
- Press to select a record.
- Press to return to the records interface.
When a record is selected, press or to step through the records upwards or
downwards.
Short press to view the detailed information of the selected record (NO.,
measured value, MAX, MIN and AVG). Short press to return to the record list.
When viewing a record, users can long press to delete this record. Press or to
view the previous or next record.
6.14.2 Last Page
Select “Last page” and the screen will display the records from the last page.
The number and value of each record will also be displayed, refer to 6.14.1
First Page.
6.14.3 Selected Item
In the records interface, select “Selected item” to choose a record to be
viewed. The screen will display the record list starting from the page where
the selected record is located. The number and value of each record will also
be displayed.
-
Press or to adjust the record number.
-
Press to confirm the adjusted digit (tens digit/tens digit/ units digit).
After the units digit is confirmed, it will enter the page where the selected record is located. -
Press to return to the records interface.
6.14.4 Delete the Item
In the records interface, select “Delete the item” to choose the record to be
deleted.
- Press or to adjust the record number.
- Press to confirm the adjusted digit (tens digit/tens digit /units digit). After the units digit is confirmed, a prompt will pop up.
- Press to return to the records interface.
When the prompt pops up, users can press or to select Yes/No, and then press
to delete or not.
Press to return to the previous interface. Deleting records takes a while.
6.14.5 Delete All
In the records interface, select “Delete all” to delete all records. Press or
to select Yes/No, and then press to delete or not. Press to return to the
records interface. 6.15 Setting
In the setting interface:
- Press to step through the options upwards.
- Press to step through the options downwards.
- Press to enter the interface of the selected option.
- Press to return to the menu.
6.16 Sound
In the sound interface:
- Press or to select ON/OFF.
- Press to save.
- Press to return to the setting interface.
6.17 Backlight
In the backlight interface:
- Press to add the backlight time.
- Press to subtract the backlight time.
- Press to confirm.
- Press to return to the setting interface.
6.18 Power Off
In the power off interface:
- Press to add the power off time.
- Press to subtract the power off time.
- Press to confirm.
- Press to return to the setting interface.
6.19 Alarm
In the alarm interface, first adjust the standard value, and then adjust the
limit. Long press or to quickly adjust the value.
- Press to add the standard/limit.
- Press to subtract the standard/limit.
- Press to confirm.
- Press to return to the setting interface.
6.20 Language
In the language interface:
- Press to step through the options upwards.
- Press to step through the options downwards.
- Press to save.
- Press to return to the setting interface.
6.21 About
Press to return to the setting interface in the about interface.
The content will be updated later. The actual situation prevails.
6.22 Reset
In the reset interface:
- Press or to select Yes/No.
- Press to confirm.
- Press to return to the setting interface.
Measurement Applications
7.1 Measurement Modes
● Single point measurement: Use the probe to measure any point of the measured
object. The displayed value is the thickness.
● Two-point measurement: Use the probe to measure the same point of the
measured object twice. The probe parting plane is 90°, and the smaller value
is the thickness.
● Multi-point measurement: Multiple measurements are made in a circle with a
diameter of about 30mm, and the minimum value is the thickness.
● Continuous measurement: Continuously measure along a specified route with
the single point method.
The distance between each point is not less than 5mm, and the minimum value is
the thickness.
7.2 Tube Wall Measurement
The probe split surface can measure along the axis of the pipe or the axis of
the vertical pipe. If the pipe diameter is large, the measurement should be in
the direction of the vertical axis. When the pipe diameter is small, it should
be measured in both directions, and take the minimum value as the thickness.
Maintenance and Cautions
8.1 General Cautions
● The gauge and probe should be avoided from strong vibration.
●Do not place the gauge in a damp environment.
●Do not rotate the probe when inserting or removing it to avoid damaging the
cable core.
●The attachment of oil and dust will make the probe cable gradually aging and
fracture. After use, the dirt on the cable should be removed.
8.2 Cautions during Measurement
● Good measurements are made only when stable coupling symbol appears;
● When there is a large amount of coupling agent on the object surface, it
will cause mismeasurement.
Therefore, at the end of the measurement, the probe should be quickly removed
from the surface.
● The probe surface is acrylic resin and should be gently pressed when used.
When measuring rough surface, try to reduce the probe scratching on it.
● When measuring at room temperature, the surface of the measured object
should not exceed 60 ℃ , otherwise the probe can no longer be used.
● If the probe wears, the indicating value will be unstable. Please replace
it.
8.3 Clean the Standard Thickness Block
The standard thickness block should avoid rust.
Wipe the block after use, and don’t come into contact with sweat. If not used
for a long time, a little grease should be coated on the surface of the block,
and the grease can be wiped after the next use.
8.4 Clean the Outer Shell
Alcohol and diluent have corrosive effect on the casing, so wipe it gently
with a wet cloth.
8.5 Repair
When there are problems with the gauge (such as inability to measure, abnormal
LCD display, large error and button operation failure or confusion), please do
not disassemble or adjust any parts, and contact after-sales.
Appendix A Material Sound Velocity
Note: The sound velocities listed are approximate and are for reference
only.
| Sound velocity
---|---
Material| Mips| m/s
User define 1| 0.233| 5920
User define 2| 0.233| 5920
User define 3| 0.233| 5920
Aluminum| 0.250| 6340-6400
Steel, common| 0.233| 5920
Steel, stainless| 0.226| 5740
Brass| 0.173| 4399
Copper| 0.186| 4720
Iron| 0.233| 5930
Cast Iron| 0.173-0.229| 4400-5820
Lead| 0.094| 2400
Nylon| 0.105| 2680
Silver| 0.142| 3607
Gold| 0.128| 3251
Zinc| 0.164| 4170
Titanium| 0.236| 5990
Tin| 0.117| 2960
Acrylic resin| 0.109| 2760
Epoxy resin| 0.100| 2540
Ice| 0.157| 3988
Nickel| 0.222| 5639
Plexiglass| 0.106| 2692
Porcelain| 0.230| 5842
PVC| 0.094| 2388
Quartz glass| 0.222| 5639
Rubber, vulcanized| 0.091| 2311
Appendix B FAQ and Solutions
B.1 Influence of Surface Conditions on Measurement Results
B.1.1 Surface Coverings
Before measurement, remove dust, dirt, rust, paint and other coverings on the
object surface.
B.1.2 Rough Surfaces
Too rough surface can cause error, or even no reading.
The material surface should be as smooth as possible.
Grinding, polishing, filing and other methods are helpful, and a high-
viscosity coupling agent can also be used.
B.1.3 Roughwrought Surfaces
Regular grooves created by roughwrought surfaces (such as lathes or machines)
can also cause errors, as described above. In addition, adjust the angle
between the probe partition board (through the bottom center of the probe
metal thin layer) and the measured material groove, so that the partition
board and the groove are vertical or parallel to each other. Take the minimum
value as the thickness.
B.1.4 Cylindrical Surfaces
When measuring cylindrical materials, such as pipes, oil drums, etc., it is
very important to correctly choose the included angle between the probe
partition board and the axis of the measured material. The probe is coupled
with the measured material, and the probe partition board is parallel or
perpendicular to the material axis. Slowly shake the probe vertically along
the axis, and the reading on the screen will change regularly. Take the
minimum value as the thickness.
Select the angle between the probe partition board and the axis according to
the curvature of the material. For pipe with large diameter, the probe
partition board is perpendicular to the pipe axis. For pipe with small
diameter, the probe partition board is perpendicular or parallel to the pipe
axis. Take the minimum value as the thickness.
B.1.5 Composite Shapes
The method described above can be used when measuring materials with composite
shapes (such as pipe elbows). The difference is that two measurements are
made, reading the probe partition and axis perpendicular and parallel values
respectively. Take the smaller value as the thickness.
B.1.6 Unparallel Surfaces
To obtain a stable and reliable value, the other surface of the material must
be parallel or coaxial to the measured surface.
B.2 Influence of Temperature on Measurement Results
The thickness and the velocity of ultrasonic wave in the material are affected
by temperature. When higher accuracy is needed, the blocks with the same
material and approximate thickness can be measured under the same temperature.
The temperature compensation coefficient can be obtained, and the measured
value of the workpiece can be corrected with this coefficient.
B.3 Influence of Material Attenuation on Measurement Results
Fiber, porous, coarse crystal materials will cause a lot of ultrasonic
scattering and energy attenuation, so that the gauge may have abnormal reading
or no reading (abnormal reading is usually less than the actual thickness). In
this case, the material is not suitable for measurement with this gauge.
B.4 Reference Blocks
The closer the material of the calibrated block is to the measured material,
the more accurate the measurement will be. An ideal reference block would be a
set of blocks of different thicknesses of the measured material, which would
provide instrumental compensatory correction factors (such as material
microstructure, heat treatment conditions, particle orientation, surface
roughness, etc.).
In most cases, satisfactory measurement accuracy can be obtained by using only
one reference block. The block shall be of the same material and of a similar
thickness as the measured material. The uniform material can be used as a
block after measuring with a micrometer.
For thin materials, a block may be used to determine the exact lower limit
when its thickness approaches the lower limit measured by the probe. Do not
measure material below the lower limit of thickness. If the thickness range is
estimable, the thickness of the block shall be the upper limit. When the
material to be tested is thick, especially the alloy with a complex internal
structure, one of the blocks close to the measured material should be selected
to facilitate calibration.
The internal structure of most forgings and castings is directional. The sound
velocity will vary in different directions. To solve this problem, the test
block should have an internal structure in the same direction as the measured
material, and sound waves should travel in the same direction in the test
block as in the material.
In certain cases, the reference block can be replaced by the sound velocity
table of the known material. But this is only approximating the substitution
of some reference blocks. In some cases, the values in the sound velocity
table differ from the actual measurements because of differences in the
physical and chemical conditions of the material. This method is often used to
measure mild steel, but only as a rough measurement. This gauge has the
function of measuring the sound velocity, so it can measure the sound
velocity first, and then measure the workpiece with the sound velocity.
B.5 Casting Measurement
The grain size of casting material is coarse and the structure is not dense
enough, which will cause great attenuation of sound energy. In addition, the
measurement is often carried out in the state of wool, so it is difficult to
measure. The degree of attenuation is closely related to grain size and
ultrasonic frequency. At the same frequency, the attenuation increases with
the increase of grain diameter, but when it exceeds the highest point, the
attenuation tends to a fixed value. For probes with different frequency, the
attenuation increases with the increase of frequency. When the coarse grain
and casting in the presence of coarse heterogeneous structure, the ultrasonic
signal will produce abnormal reflection, grass echo or tree echo, resulting in
the reading error.
Coarse grain size and inconsistent densification of tissues at different
positions in the workpiece will cause differences in sound velocity and make
the measurement results inaccurate.
Note:
● When measuring the casting with rough surface, a coupling agent with high
viscosity must be used.
● It is recommended to calibrate the sound velocity of the material with a
block of the same material and direction as the measured object.
B.6.4 Probe Wear
The probe surface is acrylic resin. Long-term use will increase its roughness,
resulting in a decrease in sensitivity. If the probe is badly worn, sandpaper
or whetstone can be used to smooth the surface. If the measured value is still
unstable, replace the probe.
B.6.5 Multilayer and Composite Materials
It is impossible to measure multilayer materials with incompact bonding
surfaces because ultrasonic waves cannot penetrate uncoupled surfaces.
Ultrasonic waves cannot spread at uniform speed in composite materials, so the
gauges measuring thickness by ultrasonic reflection principle are not
suitable for measuring multilayer and composite materials.
B.6.6 Influence of Oxide Layer on Metal Surface
Some metals, such as aluminum, have a dense oxide layer on their surfaces.
This oxide layer is closely bonded with the matrix, but the spread speeds of
ultrasonic wave in these two substances are different, so it will cause error.
Please pay attention to this condition when using. Users can select a sample
block from the same batch of measured materials, measure its thickness with
micrometer or caliper, and calibrate the gauge with this sample block.
B.6.7 Abnormal Thickness Reading
Operators should be able to identify abnormal readings, which are usually
caused by rust spots, corrosion pits, and internal defects. Solutions can be
found in the relevant sections of this manual.
B.6.8 Selection and Use of Coupling Agents
The coupling agent is used as the carrier of ultrasonic signal spread between
the probe and the measured material. If the type or use method of coupling
agent is improper, it may cause error, or the coupling symbol flickers, and
the measured value cannot be stable. The coupling agent should be applied in
appropriate amount and coated evenly. Low viscosity coupling agents can be
used on smooth surfaces. High viscosity coupling agents can be used on rough,
vertical and top surfaces.
B.7 Methods for Reducing Errors
B.7.1 Ultra-Thin Materials
With any ultrasonic thickness gauge, errors will occur when the measured
material thickness falls below the lower limit of the probe.
When measuring ultra-thin materials, a false result called “double refraction”
sometimes occurs: the reading is twice as thick as it actually is. Another
type of false result is called “pulse envelope, cycle jump”: the measured
value is greater than the actual thickness. In order to prevent such errors,
the ultra-thin material measurements should be checked repeatedly.
B.7.2 Rust Spots and Corrosion Pits
Rust pits on another surface of the measured material (small rust spots are
sometimes difficult to detect) will cause irregular changes in the reading and
in extreme cases no reading at all. When pits are found, this area should be
measured with great care. Multiple measurements can be performed by selecting
the orientation of the probe partition board at different angles.
B.7.3 Material Identification Error
When the gauge is calibrated with one material and then measures another, the
wrong result will be obtained.
Please select the correct sound velocity.
Packing List
Standard accessories | NO. | Item | Quantity | Remarks |
---|---|---|---|---|
1 | Ultrasonic thickness | |||
gauge | 1 | |||
2 | Probe (5MHz c1)10) | 1 | ||
3 | Coupling agent | 1 | ||
4 | Carrying bag | 1 | ||
5 | Data cable | 1 | ||
6 | User manual | 1 |
UNI-
TREND TECHNOLOGY (CHINA) CO., LTD.
No.6, Gong Ye Bei 1st Road,
Songshan Lake National High-Tech Industrial
Development Zone, Dongguan City,
Guangdong Province, China
Tel: (86-769) 8572 3888
http://www.uni-trend.com
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>