KaWe Dermatoscope Piccolight User Manual
- June 12, 2024
- KaWe
Table of Contents
KaWe Dermatoscope Piccolight
Product Information
- Product Name: Dermatoskope
- Model: QM-1-101H
- Manufacturer: KaWe
- Application: The dermatoscope is designed for skin surface examination and can only be used by qualified medical professionals.
- Indications for Use: The device is suitable for people of any age, male and female, who require an examination of skin surface structures.
- Power Source: Batteries (not specified)
- Operating Time: The devices are designed for periodic use with a maximum operation time of 1 minute. There should be a 10-minute pause between applications.
- Warranty: The device has a two-year warranty from the date of purchase, except for lamps/rechargeable batteries. LED lamps have a five-year guarantee from the date of purchase.
Product Usage Instructions
- Before using the dermatoscope, carefully read and understand the user manual.
- Ensure that the device is properly maintained and follows the maintenance recommendations.
- Only qualified medical professionals should use the dermatoscope.
- To use the dermatoscope, remove the investigation head by lowering the throw-on insert.
- Do not exceed the maximum operation time of 1 minute and allow at least a 10-minute pause between applications.
- Do not put the dermatoscope in fluid and keep fluid off the handle. Clean or disinfect the dermatoscope with surface alcohol-containing disinfectants. Do not use abrasive cleaning agents.
- Do not sterilize the dermatoscope.
- If the device heats excessively, immediately turn off the handle to prevent burn injuries.
- Follow additional maintenance recommendations to extend the service life of the rechargeable batteries.
- If any issues arise or repair is necessary, please contact your dealer. The warranty covers the device for two years from the date of purchase (excluding lamps/rechargeable batteries).
Dear buyer! Thank you for choosing a KaWe product. Our products are known for
high quality and durability. This KaWe device complies with the provisions of
Directive EU 2017/745 (European Directive on Medical Devices), whereby it is
classified as a medical product of class 1. Please carefully consider this
instruction and read all the provisions prior to use. Follow the maintenance
recommendations. Prior to use, please carefully consider the maintenance
procedure.
Application: Only qualified medical professionals can use the
dermatoscope. Group of patients with indications for use: People of any age,
male and female, requiring an examination of skin surface structures.
Purpose: KaWe dermatoscope is designed for visual examination of the body
surface. It zooms skin surface structures almost ten-fold, and a halogen lamp
lights the area in question. The mechanism is analogous to microscopic
examination in the reflected right (chemiluminescence microscopy).
Dermatoscopy is aimed at the precise examination of skin pigment changes in
everyday practice, specifically, within the early detection of malignant
melanoma. Skin reflexes are suppressed by a special dermatoscopic moisturizing
gel (KaWe Derma-Gel), and skin color changes are localized and detected more
efficiently. Contact lens scale allows for establishing the sizes of changes.
The device is designed only for skin surface examination.
OVERVIEW
Operation of KaWe EUROLIGHT D30 2.5 V and 3.5 V dermatoscopes:
Dermatoskope
- KaWe EUROLIGHT D30 2.5 V dermatoscope
- KaWe PICCOLIGHT D 2.5 V dermatoscope
- KaWe EUROLIGHT D30 LED 3.5 V dermatoscope
Maintenance of KaWe EUROLIGHT D30 2.5 V and 3.5 V dermatoscopes: For
proper functioning, use standard KaWe 2.5 V handles with two KaWe 2.5 V or 3.5
V “C” or Baby 1.5 V rechargeable batteries. Put them on and fix the device’s
head. To switch the device on, press and turn to the left a white plastic
button on the rotary switch. Variable resistor allows quick brightness
regulation (0=off, 1=max. brightness). To switch the device off, turn the
button the other way back to the reference position (switching off movement).
To replace the battery or rechargeable battery, open the lower cover. To
charge handles with rechargeable batteries, take off the plastic cover and
connect with KaWe Med- Charge 4000 charging device. To charge KaWe EUROLIGHT
D30 LED 3.5 V, unscrew two connecting parts of the handle and put the lower
part of the handle with the mains plug into the socket.
(ATTENTION: 3.5 V version is compatible with KaWe 3.5 V rechargeable
batteries only). Remove the investigation head by lowering the thrown insert.
ATTENTION: Follow additional maintenance recommendations to make the
service life of the rechargeable batteries longer.
Maintenance of KaWe PICCOLIGHT: Proper functioning is provided by KaWe
PICCOLIGHT handles with 2 batteries (“АА”-Mignon 1.5 V). To switch the device
on/off, use a sliding switch. To switch the device on using a sliding switch,
move it towards the dermatoscopy head. To switch the device off, move the
sliding switch in reverse direction, to reference position. To replace the
battery and lamp, unscrew the head of KaWe PICCOLIGHT.
Time of application: The devices are designed for periodic use, maximum
operation time totals 1 minute. There shall be 10-minute pauses between the
applications.
Lamp replacement of KaWe PICCOLIGHT: For proper light functioning and
quality, use only original KaWe lamps. Unscrew the head of the dermatoscope
from the handle and remove the lamp. Clean the glass bulb of the new lamp with
an alcohol-containing cloth. The glass bulb shall be clean and free from
fingerprints (grease spots). Put the lamp all the way into its groove in the
side pin. Then screw the investigation head back to the handle.
Lamp replacement of KaWe EUROLIGHT: For proper light functioning and
quality, use only original KaWe lamps. Unlatch the head of the dermatoscope
from the handle and remove the lamp. Now repeat the steps above in the section
on replacing the KaWe PICCOLIGHT lamp. The glass bulb shall be clean and free
from fingerprints (grease spots). Put the lamp all the way in.
Handling: Put special moisturizing KaWe Derma-Gel dermatoscope gel on the
skin area of interest. Turn the light on and set the contact lens on the skin
without excessive pressing. Even lighting will cover the skin area with a
large examination diameter (25 mm). Place the examined area in the center of
the contact lens, if possible. Bring the eyepiece as close to the eye as
possible. With a free hand, turn the focus ring until the image of the area of
interest becomes sharp. Try looking through the eyepiece without pressure on
eye muscles (not accommodating them for a sharper view). Scales contact lens
is designed to control pigmented skin changes. To examine raised or palpable
skin areas (damages), a dermatoscope can be used without the contact lens. For
a sharper view, you can raise the dermatoscope over the area of interest but
not further than 5 mm.
Technical characteristics
- Magnifier glass with 10x magnification
- Contact lens diameter: 25 mm
- Focusing / sharpness adjustment: -6…+3.5 diopter
- Halogen lamp aperture: 2.5 V (KaWe EUROLIGHT D30)
- Vacuum lamp aperture: 2.5 V (KaWe PICCOLIGHT D)
- LED lamp aperture: 3.5 V (KaWe EUROLIGHT D30 LED)
Additional directions, maintenance, and storage: The device will serve
for years with proper handling and storage.
Warranty: The warranty for the device totals two years from the date of
purchase with proper handling and following the operation manual (the warranty
does not cover lamps / rechargeable batteries). LED lamps are subject to five
years of guarantee from the date of purchase. If any issues arise or repair is
necessary, please contact your dealer.
Warnings: An error may cause extensive heat posing a threat of burn
injury. If the devices heats excessively, immediately turn the handle off.
Warnings: Do not put the dermatoscope in fluid and keep fluid off the
handle! The dermatoscope is not designed for surgery, so it is sufficient to
clean or disinfect it with surface alcohol-containing disinfectants. Do not
use abrasive cleaning agents! The dermatoscope shall not be sterilized.
Secondary treatment limitations: Secondary treatment has a minor effect on
these devices. The service life of the product is usually defined based on the
degree of wear and use-related damage. Directives
Working area: Clean the surface with a disposable/ paper cloth.
Environment conditions
Transportation and storage: Store and transport dermatoscopes protecting
them from dust, moisture and dirt.
Preparation for cleaning: No special requirements.
Manual cleaning: Clean the dermatoscope external surface with a damp,
soft and lent-free cloth. Do not put the dermatoscope in fluid and keep body
off the handle! The dermatoscope is not designed for surgery, so it is
sufficient to clean it with surface alcohol-containing detergent. Do not use
abrasive cleaning agents! The device shall not be sterilized. Clean the
magnifying glass with a glass detergent only.
Automatic cleaning: Automatic cleaning is impossible/ unnecessary.
Disinfection: Clean the dermatoscope with surface alcohol-containing
disinfectants. Use a slightly damp cloth.
Sterilization: Scaled lens shall be sterilized in autoclave at 134 °C for
at least 3 minutes.
Additional information: When sterilizing several devices in autoclave, do
not exceed the maximum load indicated by the sterilizer manufacturer.
Maintenance
Remove or replace the damaged parts. If the device does not work, do the
following: 1. Replace the batteries or rechargeable batteries and check
whether they are serviceable; 2. Replace the lamp and check whether it is
serviceable; 3. Send the device for repair.
Control and functional check: Check the smooth travel of moveable parts,
such as magnifying glass frame and plugs. Switch the device on. All the
instruments: Visually check for damages and wear. Keep all the electric
contacts clean.
Package: Separate zipped bag / fabric bag.
Disposal: Electric/electronic devices shall not be disposed together with
household waste! Dispose the device in accordance with the local regulations
at a special collection point.
General accessories: To find out more about these products, visit our
homepage: www.kawemed.com.
Manufacturer: KaWe
Manufacturer’s contact details: Dealer’s address or phone number:
+49-7141-68188-0.
Basic UDI-DI 4030155KaWe0103YV
Legend
Instructions for service personnel and patients: In case of problems during the device use, it is necessary to immediately contact the manufacturer and the competent authorities of the relevant member state of the Union, in which the service personnel or the patient are located.
CONTACT
- KIRCHNER & WILHELM GmbH + Co. KG
- Eberhardstr. 56
- 71679 Asperg
- Germany
- Zentrale / Central offi ce
- Fon: +49 -7141-68188-0
- Fax: +49 -7141-68188-11
- Email: info@kawemed.de.
- Internet: www.kawemed.com.
QM-1-101H / B-26790 / 2021-04