JUNCTEK KG-F Series DC Voltage and Current Meter
Product Information
- Product: Hangzhou Junce Instruments Co., Ltd.
- Trademark: JUNCTEK
- Manufacturer: Hangzhou Junce Instruments Co., Ltd.
Guaranty and Declaration: Hangzhou Junce Instruments Co., Ltd. guarantees
the quality and functionality of the product. This document replaces all
previously published documentation.
Safety Requirement: The meter should be operated with caution to avoid
personal injury or damage to the meter and connected products. Please follow
the safety precautions specified in this manual.
Product Usage Instructions
Observe All Terminal Ratings:
Before connecting the meter, check the ratings and markers on the meter and
refer to the manual for more information.
Use Proper Over-voltage Protection:
Ensure that the product is protected from over-voltage, such as lightning
strikes, to prevent electric shock hazards.
Do Not Operate Without Covers:
Always operate the meter with covers or panels in place to prevent any
damage or malfunction.
Do Not Insert Anything Into the Air Outlet:
Avoid inserting any objects into the air outlet to avoid damaging the
meter.
Avoid Circuit or Wire Exposure:
Do not touch exposed junctions and components when the unit is powered on
to prevent electric shock.
Do Not Operate With Suspected Failures:
If you suspect any damage to the meter, have it inspected by JUNCTEK
authorized personnel before further operations. Only authorized personnel
should perform maintenance, adjustment, or replacement.
Provide Adequate Ventilation:
Ensure that the meter is well-ventilated to prevent overheating and damage.
Regularly inspect the air outlet and fan for any blockages.
Do Not Operate in Wet Conditions:
Avoid using the meter in humid environments to prevent short circuits and
electric shock hazards.
Do Not Operate in an Explosive Atmosphere:
Never operate the meter in an explosive atmosphere to avoid personal
injuries and damage to the meter.
Keep Meter Surfaces Clean and Dry:
Regularly clean and keep the surfaces of the meter dry to maintain optimal
performance and prevent dust or moisture from affecting its functionality.
Prevent Electrostatic Impact:
Operate the meter in an electrostatic discharge protective environment.
Ground both internal and external conductors of cables before making
connections to avoid damage from static discharges.
Handle with Caution:
During transportation, handle the meter with care to prevent damage to
keys, knobs, interfaces, and other parts on the panels.
Note: For any problems or requirements when using the product or this manual, please contact JUNCTEK by email at junce@junteks.com or visit their website at www.junteks.com.
Hangzhou Junce Instruments Co., Ltd.
I
Hangzhou Junce Instruments Co., Ltd.
Guaranty and Declaration
Copyright Hangzhou Junce Instruments Co., Ltd. all right reserved. Trademark
Information JUNCTEK is a registered trademark of Hangzhou Junce Instruments
Co., Ltd. Notices JUNCTEK products are covered by P.R.C. patents, issued and
pending. This document replaces all previously published documentation.
Contact Us If you have any problem or requirement when using our products or
this manual, please contact JUNCTEK. E-mail: junce@junteks.com Website:
www.junteks.com
I
Hangzhou Junce Instruments Co., Ltd.
Safety Requirement
General Safety Summary
Please review the following safety precautions carefully before putting the
meter into operation so as to avoid any personal injury or damage to the meter
and any product connected to it. To prevent potential hazards, please follow
the instructions specified in this manual to use the meter properly.
Observe All Terminal Ratings To avoid fire or shock hazard, observe all
ratings and markers on the meter and check your manual for more information
about ratings before connecting the meter.
Use Proper Over-voltage Protection Ensure that no over-voltage (such as that
caused by a bolt of lightning) can reach the product. Otherwise, the operator
might be exposed to the danger of an electric shock.
Do Not Operate Without Covers Do not operate the meter with covers or panels
removed.
Do Not Insert Anything Into the Air Outlet Do not insert anything into the air
outlet to avoid damage to the meter.
Avoid Circuit or Wire Exposure Do not touch exposed junctions and components
when the unit is powered on.
Do Not Operate With Suspected Failures If you suspect that any damage may
occur to the meter, have it inspected by JUNCTEK authorized personnel before
further operations. Any maintenance, adjustment or replacement especially to
circuits or accessories must be performed by JUNCTEK authorized personnel.
Provide Adequate Ventilation Inadequate ventilation may cause an increase of
temperature in the meter, which would cause damage to the meter. So please
keep the meter well ventilated and inspect the air outlet and the fan
regularly.
Do Not Operate in Wet Conditions To avoid short circuit inside the meter or
electric shock, never operate the meter in a humid environment.
II
Hangzhou Junce Instruments Co., Ltd.
Do Not Operate in an Explosive Atmosphere To avoid personal injuries or damage
to the meter, never operate the meter in an explosive atmosphere.
Keep Meter Surfaces Clean and Dry To avoid dust or moisture from affecting the
performance of the meter, keep the surfaces of the meter clean and dry.
Prevent Electrostatic Impact Operate the meter in an electrostatic discharge
protective environment to avoid damage induced by static discharges. Always
ground both the internal and external conductors of cables to release static
before making connections.
Handle with Caution Please handle with care during transportation to avoid
damage to keys, knobs, interfaces, and other parts on the panels.
Notices
1. Do not exceed the voltage and current range of the meter, otherwise the
meter will be damaged. 2. The positive and negative poles cannot be reversed,
and the reverse connection cannot be handled correctly. 3. The shell of the
meter is fragile and easy to corrode. Please don’t hit or close to chemicals
to avoid corrosion. 4. Storage temperature: -25~50, and keep the meter in a
dry environment. 5. Do not attempt to disassemble the meterit will void the
warranty. There are no user-serviceable parts inside the meter. Repairs can
only be made through designated repair outlets or sent back to the factory. 6.
The display screen is a fragile device, please do not touch or bump it .
Please avoid children playing with the meter. When there is dirt on the LCD
surface, wipe it carefully with a soft cloth. 7. Please do not move the meter
violently to avoid causing irreparable damage to the internal circuit. If the
meter does not work properly, please contact the supplier!
III
Inspection
When you get a new KG-F series coulometer, it is recommended that you inspect
the meter according to the following steps.
Inspect the Packaging If the packaging has been damaged, do not dispose the
damaged
packaging or cushioning materials until the shipment has been checked for
completeness and has passed both electrical and mechanical tests. The
consigner or carrier shall be liable for the damage to the meter resulting
from shipment. We would not be responsible for free maintenance/rework or
replacement of the meter.
Check the Contents Please check the contents according to the packing lists.
If the meter is
damaged or incomplete, please contact your JUNCTEK sales representative.
KG-F series of Measuring module
1pc
Host
KG-F series of display module
1pc
KG-F series of the Sampler
1pc
3m communication cable
1pc
Temperature Sensor
1pc
4P connection cable
1pc
Accessory 3P connection cable (external relay cable)
1pc
Terminal
1pc
Quick guide
1pc
Inspect the Meter In case of any mechanical damage, missing parts, or failure
in passing the
electrical and mechanical tests, contact your JUNCTEK sales representative.
2
Hangzhou Junce Instruments Co., Ltd.
Chapter Overview
Brief Introduction
The KG-F series voltage and current meter is a new type of coulometer that can
measure various parameters such as voltage, current, power, charge and
discharge AH value, watt-hours, time, etc. It can also set parameters to
achieve multi protection functions such as over-voltage protection, under-
voltage protection, over-current protection, over-power protection, over
temperature protection and time limit protection.The meter can automatically
identify the direction of the current, and can monitor the battery capacity in
real time and display the measured data in color LCD screen. The KG-F series
voltage and current meters add voltage and current curve display and export
functions on the basis of the original functions. At the same time, you can
use the mobile APP and the computer to control the measuring module, and the
firmware can also be updated in real time.
Dimensions
Dimension of display module Note: display module inlay hole size: 85.1 * 58.1mm
Display module hole bitmap
Dimension of measuring module
Hole location of measuring module
3
Hangzhou Junce Instruments Co., Ltd.
Dimension of the 100A sampler
Dimension of the 100A sampler
Dimension of the 400A sampler
Dimension of the 600A sampler
Specification
Model Sampling method Voltage measurement range (External power supply) Voltage measurement range (Self-powered) Voltage resolution Current measurement range Current resolution Relay
KG105F The Shunt
0-120V
10-120V 0.01V 0~50A 0.01A
KG110F The Shunt
KG140F The Shunt
0-120V
0-120V
10-120V
10-120V
0.01V
0.01V
0~100A
0~400A
0.01A
0.1A Optional
4
KG160F The Shunt
0-120V
10-120V 0.01V 0~600A 0.1A
KG610F The Shunt
0-600V
10-120V 0.01A 0~100A 0.01A
Temperature measurement range Capacity display range Power measurement range Power resolution AH measureme nt range Capacity resolu tion Watt-hour measurement range Watt-hour accuracy Time measurement range Time resolution Bluetooth communication distance Voltage accura cy Current accurac y Sampling rate Measuring module power consumption Display module power consumption Over power protection Negative over current protection Forward over
Hangzhou Junce Instruments Co., Ltd.
-20-120 0%~100% 0~180KW
0.01W 0~9999.99AH
0.001AH 0~9999.99kWH
0.001WH 0~999:59:59S
1 second 10m
±2%+3 digits ±5%+10 digits 1 time / second
About 0.4W
About 0.5W 0-99999.99W 0~50A/100A/400A/600A 0~50A/100A/400A/600A
5
current protection Over voltage protection Under voltage protection External over temperature protection Protection recovery time Enter standby time Delay time setting Communication address
Hangzhou Junce Instruments Co., Ltd.
0~120V/600V 0~120V/600V
0-120 0-99s 0-60s 0-99S
There are 99 P01-P99, P00 is the broadcast address
Chapter Meter Introduction 1.Introduction of display module
Item 1 2 3
Figure 2-1-1 KG-F series display module diagram Table 2-1-1 KG-F series display module diagram instructions
Description
Item
Description
LCD screen
4
button
SETbutton
5
OKbutton
button
6
Wiring port
(1).LCD screen The 2.4-inch TFT color LCD screen displays the current function
menu
and parameter settings.
(2).SETbutton
6
Hangzhou Junce Instruments Co., Ltd.
Shortly pressing the SET button can quickly switch to the setting interface.
In the main interface, press and hold the SET button can select the address
and use to quickly adjust the address. (3).button
When setting parameters, button is used to change the parameters. When setting
the system, button is used to select the corresponding system setting.
(4).button When setting parameters, button is used to change the parameters.
When setting the system,button is used to select the corresponding system
setting. In the main interface, press and hold the button will jump out the
“Clear data” pop-up window to confirm or cancel clear data. (5).OKbutton In
the main interface, shortly pressing the OK button can control on or off of
output status and data record. Pressing and hold the OK button can quickly
lock or unlock the buttons. (6).Wiring port The wiring port is used to connect
to the measuring module.
2.Introduction of the measuring module
7
Hangzhou Junce Instruments Co., Ltd.
Figure 2-2-1 KG-F series measuring module diagram Table 2-2-1 Instructions diagram of the KG-F series measuring module
Item 1
2 3 4
Description Indicator light
2wires&3wires switch Power supply interface
485 communication connection interface
Item 5
6 7 8
Description
485 communication display interface Relay output control interface
Sampler interface
External temperature measurement interface
(1). Indicator light The blinking of the indicator light indicates the working
state, and the slow
flashing indicates that the power supply is normal and the measurement is
under normal operation.
(2). 2-wire 3-wire power supply switch selection The switch can be selected
from external power supply or self-powered,
when the switch is moved up to 2W, it is suitable for self-power supply, and
the voltage measurement range is 10-120V, when the switch is moved down to 3W,
it is suitable for external power supply, and the external power supply
voltage range is 10- 80V, the voltage measurement range is 0-120V.
(3). Power supply interface There are three power supply interfaces, the
battery positive interface:
Vsns, the external power supply positive interface: VEXT, and the battery and
external power supply negative interface: GND.
(4). 485 communication connection interface The interface used to connect to
the computer, and it can also be
connected to another measuring module to realize multi-computer communication.
The internal sequence from left to right is: B, A, GND, NC.
8
Hangzhou Junce Instruments Co., Ltd.
(5). 485 communication display interface Used to connect with the display
module, when the measuring module has
power supply, the communication interface is live. The internal sequence from
left to right is: B, A, GND, +5V.
(6). Relay output control interface The relay output control interface can be
used with relays. The internal
sequence from left to right is: GND, OUT, VEXT.
(7). The Sampler interface Used to connect with the sampler for current
detection. The internal
sequence from left to right is: VCC, GND, ISNS+, ISNS-.
(8). External temperature measurement interface Used to connect with a
temperature sensor to measure the external
temperature. The internal sequence from left to right is: TSNS, GND.
3.Introduction of the Sampler
Figure 2-4-1 KG-F series of the Sampler diagram Table 2-4-1 KG-F series of the Sampler diagram instructions
Item 1
Description The Sampler interface
Item
Description
The Sampler interface Connect to the Sampler interface of the measuring module for current measurement.
9
Hangzhou Junce Instruments Co., Ltd.
4.Introduction of the display interface
Item 1 2 3 4 5 6 7 8
Figure 2-5-1 KG-F series display interface diagram Table 2-5-1 KG-F series display interface diagram instructions
Description
Item
Description
Key lock status
Output status and data record
Communication address
9
Current direction
10
Battery remaining capacity
11
Battery internal resistance
External temperature
12
Battery life
Running time
Communication signal indication
Power
13
Elapsed AH value
14
Remaining AH value
15
Measured current value
Elapsed electric energy
16
Measured voltage value
(1). Key lock status The state in the figure indicates that the button can be
operated. After
pressing and holding the OK button, is displayed ,which means the
10
Hangzhou Junce Instruments Co., Ltd.
button is locked.
(2). Output status and data record Shortly press the OKbutton to control the
output status and data record
on or off.
(3). Communication address The communication address range is P00-P99. P00 is
the broadcast
address, used for multi-machine communication, P01 represents that the current
communication address is P01, and view the data of P01 address.
(4). External temperature It displays the ambient temperature. The current
display is 19°C, which
means that the temperature of the environment where the external sensor is
located is 19°C.
(5). Running time It represents the cumulative value of a single running time.
(6). Communication signal indication The current graph means that the display
module and the measuring
module are connected properly. When there is a red × in the graph, it means
that the communication is interrupted, and the address needs to be switched or
further checked.
(7). Power It measures the current power value, the unit is W.
(8). Elapsed electric energy It represents the actual accumulated electric
energy during the running
time, the unit is: KW.h.
(9). Current direction It represents the direction of current passing through
the Sampler. The
current icon can also indicate that the battery is in a discharged state.
(10). Battery remaining capacity The remaining capacity of the battery is
highlighted more intuitively with
graphics, where the 50% value means that the remaining AH. value of the
battery accounts for 50% of the preset battery AH. value. In the main
interface, when the remaining capacity percentage is less than or equal to
20%, the graph of remaining AH value is red, when remaining AH value
percentage is greater than 20%, the graph of remaining AH value is green.
11
Hangzhou Junce Instruments Co., Ltd.
(11).Battery internal resistance (estimated) It indicates the resistance value
of the current battery internal resistance.
(12). Battery left According to the charge and discharge current and capacity,
calculate the
battery left and charging time.
(13). Elapsed AH value It represents the elapsed AH value of charge and
discharge during this
period running time.
(14). Remaining AH value It represents the remaining capacity of the battery
after charging and
discharging, remaining AH. value = preset battery AH. Value – elapsed AH
value.
(15). Measured current value It represents the actual current value through
the Sampler.
(16). Measured voltage value It represents the voltage value of the voltage
measurement interface in the
power supply interface.
ChapterBasic Operation of the Meter
1.System settings
System settings Video: http://68.168.132.244/KG-F_xtsz_EN.mp4 In the main
interface, short press the SET button to enter the system setting interface.
As shown in Figures 3-1-1 to 3-1-3 below.
Figure 3-1-1 System Setting 01
Figure 3-1-2 System Setting 02
12
Hangzhou Junce Instruments Co., Ltd.
Figure 3-1-3 System Setting 03
(1). Language selection When the “language” selection is green, you can select
the language,
short press theOKbutton, the cursor selects Chinese, press the button to
switch, there are two language modes, Chinese and English, then short press
the OK button after confirming.
(2). Set current to zero In the system setting interface, press the button to
switch to
“Set current to zero”, short press theOKbutton and the “Set current to zero”
window will pop up, and you can choose to confirm or cancel “set current to
zero” by pressing the buttons , When the cursor is selected to confirm, short
press theOKbutton, if it displays “Completed” , that means setting current to
zero is completed, and the current will becomes 0A.
Precautions for the first startup and / or after restoring factory settings:
When the customer starts the meter for the first time or restores the factory
settings, the display will have a small current deviation under normal
circumstances when the meter is not connected to the load. At this time, the
charging state arrow and the discharging state arrow will jump back and forth,
which means that the meter is being calibrated automatically. This phenomenon
will last about 30 seconds, and then the current will return to zero
automatically. At this time, we need to press the key for a long time to
confirm that the current returns to zero, so that the meter can remember the
zero state, and the zero current will not jump again no matter whether the
machine is switched on or off or restarted.
(3). Clear accumulated data In the system setting interface, press the buttons
can switch
to the “Clear Data”. Short press theOKbutton will jump out the “Clear data”
pop-up window, and then you can pressbuttons to confirm or cancel clear data.
When the cursor is selected to “OK”, short press theOKbutton, if it displays
“Completed”, that meas the data is cleared. Time, watt-hour, and
13
Hangzhou Junce Instruments Co., Ltd.
elapsed AH value are cleared at this time, but the measuring is not affected.
(4). Preset battery AH. value In the system setting interface, pressing the
buttons can
switch to the “AH. preset”. Short press the OK button, the cursor will select
the value, and it can be changed by pressing thebuttons. Pressing the OK
button can switch the cursor position, from the smallest digit to the highest
digit. At the highest position, press the OK button again to confirm and save
the preset battery AH. value.
Before using the battery for the first time, you need to set the effective
capacity of the battery and the current remaining power percentage correctly.
If the effective capacity of the battery is unknown or the current remaining
capacity is unknown, the following operation steps should be followed:
a. Empty the battery pack or put it until you don’t want to discharge, at this
time, the battery power is the default 0 power.
b. Enter the preset battery capacity setting in the system settings, and set
the capacity value as large as possible (for example, set the estimated value
20Ah to 30Ah).
c. Set the remaining capacity percentage to 0%, and then charge the battery
pack.
d. After the battery is fully charged, check the remaining capacity value
displayed on the screen, and set this value to the preset battery capacity.
(5). Percentage of remaining AH value In the system setting interface,
pressing thebuttons can switch
to the “AH. remaining”. Short press theOKbutton, the cursor will select the
value, and it can be changed by pressing the buttons. Press the OKbutton again
to confirm that the set remaining AH value accounts for the preset battery AH
value. The remaining AH value can be set according to actual usage. And the
percentage of remaining AH value is continuously adjustable from 0% to 100%.
In the main interface, when the percentage of remaining AH value is less than
or equal to 20%, the graph of remaining AH value displays red, when the
percentage of remaining AH value is greater than 20%, the graph of remaining
AH value displays green.
After using or replacing the battery for the first time, the capacity value
displayed by the cooulometer is not the actual value of the battery, and
capacity setting is required.
Suggestion: Set 100% for full charge and 0% for discharge, which is more
accurate.
14
Hangzhou Junce Instruments Co., Ltd.
(6). Screen brightness In the system setting interface, press the buttons can
switch
to the “Brightness”. Short press the OK button, the cursor will select the
value, and it can be changed by pressing the buttons. Press the OKbutton again
to confirm the screen brightness. The screen brightness is continuously
adjustable from 1% to 100%.
(7). Under-voltage protection In the system setting interface, press the
buttons can switch
to the “LVP”. Short press the OK button, the cursor will select the value, and
it can be changed by pressing the buttons. Press the OK button can switch the
cursor position, from the smallest digit to the highest digit. At the highest
position, press the OK button again to confirm the set under-voltage
protection value.
When the value is 000.00V, the protection function does not start, when the
value is greater than 000.00V, the protection will start. If the current
voltage is less than the set under-voltage protection value, the upper left of
the screen will display LVP, which is in the protection state. Time stops and
no more capacity is calculated.
The protection function needs to purchase the relay (If there is no relay,
please do not set this value.) and install a relay in the circuit before it
can play a protective role in breaking the circuit.
(8). Over-voltage protection In the system setting interface, pressing the
buttons can switch
to the “OVP”. Short press the OK button, the cursor will select the value, and
it can be changed by pressing thebuttons. Pressing theOK button can switch the
cursor position, from the smallest digit to the highest digit. At the highest
position, press the OK button again to confirm the set over-voltage protection
value.
When the value is 000.00V, the protection function does not start, when the
value is greater than 000.00V, the protection will start. If the current
voltage is greater than the set over-voltage protection value, the upper left
of the screen will display OVP, which is in the protection state. Time stops
and no more capacity is calculated.
The protection function needs to purchase the relay (If there is no relay,
please do not set this value.) and install a relay in the circuit before it
can play a protective role in breaking the circuit.
(9). Positive over-current protection In the system setting interface,
pressing the button to switch
to the discharge over current protection setting. Short press the OKbutton,
the cursor will select the value, and it can be changed by pressing the
buttons. Pressing theOKbutton can switch the cursor position, from
15
Hangzhou Junce Instruments Co., Ltd.
the smallest digit to the highest digit, at the highest position, press the OK
button again to confirm the discharge over current protection.
When the value is 000.00V, the protection function does not start, when the
value is greater than 000.00V, the protection will start. If the current
discharge current is greater than the set discharge over-current protection
value, OCP will be displayed on the upper left of the screen, which is in the
protection state. Time stops and no more capacity is calculated.
The protection function needs to purchase the relay (If there is no relay,
please do not set this value.) and install a relay in the circuit before it
can play a protective role in breaking the circuit.
(10). Negative over-current protection In the system setting interface, press
thekey to switch to the
setting of discharge over-current protection. After briefly pressing the OK
key, the cursor selects the value,and it can be changed by pressing the
buttons. Pressing theOKbutton can switch the cursor position, from the
smallest digit to the highest digit, at the highest position, press the OK
button again to confirm the set charge over-current protection.
When the value is 000.00A, the protection function will not be started. If the
value is greater than 000.00A, the protection function will be started. If the
charging current is greater than the set charging over-current protection
value, NCP will be displayed at the top left of the screen, which is in the
protection state. Time stops and no more capacity is calculated.
The protection function needs to purchase the relay (If there is no relay,
please do not set this value.) and install a relay in the circuit before it
can play a protective role in breaking the circuit.
(11). Over power protection In the system setting interface, pressing the
buttons can switch
to the “OPP”. Short press the OK button, the cursor will select the value, and
it can be changed by pressing thebuttons. Pressing theOK button can switch the
cursor position from the smallest digit to the highest digit. At the highest
position, press the OK button again to confirm the set over power protection
value.
When the value is 00000.00W, the protection function does not start, when the
value is greater than 00000.00W, the protection will start. If the current
power is greater than the set over power protection value, the upper left of
the screen will display OPP, which is in the protection state.Time stops and
no more capacity is calculated.
The protection function needs to purchase the relay (If there is no relay,
please do not set this value.) and install a relay in the circuit before it
can play a protective role in breaking the circuit.
(12). Over temperature protection
16
Hangzhou Junce Instruments Co., Ltd.
In the system setting interface, pressing the buttons can switch to the “OTP”.
Short press the OK button, the cursor will select the value, and it can be
changed by pressing the buttons. Press the OK button again to confirm the set
over temperature protection.
When the value is 00, the protection function does not start, when the value
is greater than 00 , the protection will start. If the current ambient
temperature exceeds the set over-temperature protection value, the upper left
of the screen will display OTP, which is in the protection state. Time stops
and no more capacity is calculated.
The protection function needs to purchase the relay (If there is no relay,
please do not set this value.) and install a relay in the circuit before it
can play a protective role in breaking the circuit.
(13). Reverting time protection In the system setting interface, pressing
thebuttons can switch
to the “Revert time”. Short press the OK button, the cursor will select the
value, and it can be changed by pressing the buttons. Press the OK button
again to confirm the reverting time protection.
When the value is 00s, the protection status will continue until the OK button
is pressed in the main interface to implement the protection status. If the
value is greater than 00s, for example, the protection recovery time is set to
5s, which means 5 seconds after the value returns to the normal value, the
protection state is automatically canceled, the circuit is closed and work
begins.
(14). Enter standby time In the system setting interface, pressing thebuttons
can switch
to the “Standby Time”. Short press theOK button, the cursor will select the
value, and it can be changed by pressing the buttons. Press the OK button
again to confirm the set standby time.
When the value is 00s, the meter will never stand by and the display will
always be on. If the value is greater than 00s, for example, the standby time
is set to 5s, it means that the meter will enter the standby state after not
operating and the current is 0A for 5s, and the display will go out; if there
is operation or current during the period, the display will be on, and no
operation is required. And after the current is 0A for 5s, the display screen
turns off again.
(15). Voltage calibration In the system setting interface, press the buttons
can switch
to the “Voltage Cal.”. Short press theOK button, the cursor will select the
value and it can be changed by pressing the buttons. Press the OK button again
to confirm the set voltage calibration. The voltage fine-tuning calibration
can be performed online. The default voltage calibration is 0, and the maximum
voltage calibration range is ±10%.
17
Hangzhou Junce Instruments Co., Ltd.
(16). Current calibration In the system setting interface, pressing thebuttons
can switch
to the “Current Cal.”. Short press theOK button, the cursor will select the
value and it can be changed by pressing the buttons. Press the OK button again
to confirm the set current calibration. The current fine-tuning calibration
can be performed online. The default current calibration is 0, and the maximum
current calibration range is ± 25%.
(17). Temperature calibration In the system setting interface, pressing the
buttons can switch
to the “Temp. Cal.”. Short press theOK button, the cursor will select the
value and it can be changed by pressing the buttons. Press the OK button again
to confirm the set temperature calibration. The temperature fine-tuning
calibration can be performed online. The default temperature calibration is 0.
(18). Set address In the system setting interface, press the button to switch
to
the address setting, short press theOK button and select the value with the
cursor, press the button to change the communication address, Press
theOKbutton to switch from the preset communication address to the current
communication address, and press theOK button again to confirm the set
address.
Example: As shown in Figure 3-1-3, the set address is 03-01, 01 represents the
preset communication address, the preset communication address range is 01-99,
03 represents the current measuring module communication address, and the
current communication address range is 00-99 , 00 represents the broadcast
address. If we don’t know the current communication address of the measuring
module, we can set the current communication address to 00, and set the
default communication address to the desired address 03 (for example: 00-03).
After setting, press OK to complete the setting. At this time, the current
communication address of the measuring module is 03, and the meter can
communicate normally. (Use the broadcast address with caution in multi-machine
communication)
When multiple measuring modules are connected, the current communication
address can be switched to control the measuring module corresponding to the
address. At 00, multiple measuring modules can be controlled at the same time.
(19). Set delay time In the system setting interface, pressing the buttons can
switch
to the “Delay Time”. Short press the OK button, the cursor will select the
value and it can be changed by pressing the buttons. Press the
18
Hangzhou Junce Instruments Co., Ltd.
OK button again to confirm the set delay time. When the delay time is set to
00s and the measured value exceeds the
set protection value, the meter will immediately enter the protection state,
when the delay time is greater than 00s, for example, it is set to 5s, when
the measured value exceeds the set protection value, the meter enters the
protection state after 5s.
(20). Current ratio (this function is only available for KG-H series Hall
sensor version) In the system setting interface, press thebutton to switch to
the current ratio setting, short press theOKbutton and the cursor selects the
value, press the button to change the value, and then press the OKbutton again
to confirm the set current ratio. The default is X01. Example: When the wire
of the Hall sensor is wound 10 times, the current multiplier is set to X01,
the measured current value is displayed as the total current passing through
the Hall sensor 10 times, the current multiplier is set to X10, and the
measured current value is displayed as passing The current value of one turn
of the Hall sensor wire. Can be used to measure small currents.
(21). Monitor function In the system setting interface, pressing thebuttons
can switch
to the “Monitor function”. Short press the OK button and then press the
buttons to switch on or off. The monitor function must be turned on when using
the computer software control. After the monitor function is turned on, the
setting function in the display module can only set the system language,
working screen brightness and enter the standby time (After the monitoring
function is turned on, the display module will not automatically update the
data.).
(22).Relay mode In the system setting interface, pressing thebuttons can
switch
to the “Relay Mode”. Short press the OK button and then press the buttons to
switch between “NO” and “NC”.
After the selection is completed, press the OKbutton again to confirm the type
of relay. When using this function, you need to select normally open or
normally closed according to the relay type connected during actual wiring.
(23). Restore factory settings In the system setting interface, press the key
to switch to
restore the factory settings. After a short press of theOK key, the restore
factory settings box will pop up. Through the key, you can select to confirm
or cancel the restore factory settings. When the cursor is selected to
confirm, press the OK key briefly. if the display shows success, the restore
19
Hangzhou Junce Instruments Co., Ltd.
factory settings will be completed.
2.Wiring method
Wiring instructions video: http://68.168.132.244/KG-F_jxfs_EN.mp4 (1). Self-
powered wiring method
If the voltage range of the tested battery is between 10-120V during normal
operation,self-powered wiring can be used. First, turn the power supply
selection interface switch to “2W”, and then connect the positive terminal of
the battery to the power supply interface “Vsns” , This line does not need to
be particularly thick,13-16AWG wire is fine. Pay attention to the positive and
negative poles of the battery, do not connect the wrong or reverse connection.
The negative pole of the battery is connected to the screw with the battery
GND on the sampler, and the negative pole of the charger and the negative pole
of the load are connected to another screw of the sampler. It is better to use
a copper nose to connect it firmly. When charging, the current direction
symbol color is red, and the remaining capacity value increases. When
discharging, the color of the current direction symbol is sky blue, and the
remaining capacity value decreases.
Figure 3-3-1 Self-powered wiring diagram
20
Hangzhou Junce Instruments Co., Ltd.
(2). External power supply wiring method The measuring voltage range of the
external power supply is 0-120V.
First, set the power selection interface switch to “3W”, connect the positive
pole of the external power supply to the power interface “VEXT”, and connect
the negative pole of the external power supply to “GND”. When wiring, connect
the positive pole of the battery to the measurement interface “Vsns” in the
power interface. Note that the positive and negative poles of the battery and
the external power supply should not be connected incorrectly or reversed.
The negative pole of the battery is connected to the screw with the battery
GND on the sampler, and the negative pole of the charger and the negative pole
of the load are connected to the other screw of the sampler together. It is
better to use copper nose to connect firmly. When charging, the current color
is green and the residual capacity value increases. When discharging, the
current color is sky blue and the residual capacity value decreases.
Figure 3-3-2 External power supply wiring diagram (3). External power supply
wiring methodRelay control
The working power of the relay is provided by an external power supply. If the
relay is connected, an external power supply with the same working voltage as
the relay should be provided. Connect the control ports of the relay to the
“OUT” and “VEXT” of the measurement module Output Control, connect the
positive pole of the external power supply to the “VEXT” of the Output
21
Hangzhou Junce Instruments Co., Ltd.
Control, and the negative pole of the external power supply to the “GND”. Note
that the positive and negative poles of the battery and the external power
supply should not be connected wrongly or reversed.
If you want to control the charging or discharging, you should connect the
wiring according to the wiring diagram of external power supply relay in
figure 3-3-3. When the relay is closed, the indicator light will be on, and
when it is disconnected, it will be off as a prompt.
The negative pole of the battery is connected to the screw with the battery
GND on the sampler. The negative pole of the charger and the negative pole of
the load are connected to the other screw of the sampler. It is better to use
a copper nose to connect it firmly. When charging, the current direction
symbol color is red, the remaining capacity value increases. When discharging,
the current direction symbol color is sky blue, and the remaining capacity
value decreases.
Figure 3-3-3 External power supply wiring diagram (Relay control)
3.Communication protocol control
We only provide communication protocol, customers can carry out secondary
development according to communication protocol. 1.Overview
The command line is used as the control command, and the
22
Hangzhou Junce Instruments Co., Ltd.
communication rate is 115200. The PC sends out the command, the meter parses
and executes it, and then returns the result to the PC. The different commands
are described below. The sending data format is as follows
Start bit
Checksums 1-255
Function code W,R
Spacer ,
Function No. 00~99
Data field See instructions
Connector =
Spacer ,
Address symbol 1-99
Terminator
Spacer ,
Explain: (1) Function code: “W” is the write instruction used to set various
parameters, and “R” is the parameter used by the read instruction to return to
the machine. (2) Function No.: different values represent different parameter
settings. (3) Address symbol: address range is 1-99, 0 is broadcast address.
(4) Checksum: the value of the sum is obtained by adding 1 to the remainder of
255 after the sum of all the numbers after the check sum. If the sum of the
check is taken as 0, it means that it is not verified. (5) Data field: the
data field is equivalent to the number of operations of the command. For
example: W20=12162000, the instruction operands are 2000, which means that the
over voltage protection is set to 20.00V. (6) Terminator: Each instruction
ends with a carriage return + line feed, and
1. W instructions
Function code
Set up
Send command
01
Set address :W01=1,3,2,
10
Turn on the :W10=1,2,1,
output
:W10=1,0,0,
Illustrate
The communication address is set to 2 (this command should be used with
caution, if you want to use it, you must know the current communication
address and the communication address to be set, it is recommended to use the
mobile APP to modify) Turn on output status and data recording Turn off output
status and data recording
23
Hangzhou Junce Instruments Co., Ltd.
20
Set over-voltage protection
:W20=1,216,2000,
The over-voltage protection value is set to 20.00V.
21
Set under-voltage protection
:W21=1,216,2000,
The under-voltage protection value is set to 20.00V.
22
Set positive over-current protection
:W22=1,216,2000,
The positive over-current protection value is set to 20A.
23
Set negative over-current protection
:W23=1,216,2000,
The
negative
over-current
protection value is set to 20A.
24
Set over power protection
:W24=1,216,2000,
The over power protection value is set to 20W.
25
Set
over :W25=1,151,150,
temperature
protection
:W25=1,211,210,
The over temperature protection value is set at 50 .
The over temperature protection value is set to 110 .
26
Set protection reverting time
:W26=1,31,30,
The protection reverting time is set to 30s.
27
Set delay time :W27=1,31,30,
The delay time is set to 30s.
28
Set battery capacity
:W28=1,216,2000,
The battery 200.0Ah.
capacity
is
set
at
Set the voltage calibration value to
:W29=1,121,120,
20, and fine-tune to increase the voltage value.
Set the voltage calibration to a value
Voltage
of -20, and fine-tune to reduce the
29
calibration
voltage value.
settings
(The value only represents the
fine-tuning factor, the greater the
:W29=1,81,80,
absolute value of the value, the greater the amplitude of the
fine-tuning)
:W30=1,121,120,
Set the current calibration value to 20, and fine-tune to increase the
current value.
Current
30
calibration
settings
Set the current calibration to a value of -20, and fine-tune to reduce the current value. (The value only represents the
fine-tuning factor, the greater the
:W30=1,81,80,
absolute value of the value, the greater the amplitude of the
24
Hangzhou Junce Instruments Co., Ltd.
fine-tuning)
Set the temperature calibration
31
Temperature calibration settings
:W31=1,104,103, :W31=1,99,98,
value to 3°C, and the temperature increases by 3°C. Set the temperature calibration value to -2°C, and the temperature
decreases by 2°C.
33
Reserved settings (coming soon)
Reserved settings (coming soon)
Reserved settings (coming soon)
:W34=1,2,1,
Set the relay type to normally
34
Set relay type
closed.
:W34=1,0,0,
Set the relay type to normally open.
35
Restore factory settings
:W35=1,2,1,
Perform factory reset
36
Set ratio
current :W36=1,4,3,
The current ratio is set to 3.
37
Setting voltage curve scale
:W37=1,4,3,
The voltage curve scale is set to 03v / Div.
38
Setting current curve scale
:W38=1,4,3,
The current curve scale is set to 03v / Div.
Set
the
percentage of
60
battery
:W60=1,51,50,
50% of battery capacity remaining
remaining
capacity
61
Current to zero
clear :W61=1,2,1,
Perform current clear to zero
Clear
62
accumulated :W62=1,2,1,
Perform clear accumulated data
data
2. R instructions The R command is a read command, and its command format is
basically
the same as the write command format. The description will not be repeated
here. The following is the data returned by the machine, which is just an
example.
Read
Read basic machi ne
PC sending
:R00=1,2, 1,
Machine return
:r00=1,47,1120,100,10 1,
Read instructions
1 represents the communication address; 47 stands for check sum; In 1120, the
first one represents Hall
25
Hangzhou Junce Instruments Co., Ltd.
inform ation
Read all meas ured values
:R50=2,2, 1,
Read all setting s
:R51=1,2, 1,
:r50=2,215,2056,200, 5408,4592,9437,14353 , 134,4112,0,0,162,3068 2,
:r51=1,211,3000,100,2 000, 2000,10000,151,10,7,2 00, 120,90,101,0,0,2,12,13 ,
sensor (1-hall sensor, 2-shunt sampler), the second one represents 100V, and 20 represents 200A; 100 represents version 1.00; 101 represents the machine serial number; 2 represents the communication address; 215 represents the checksum; 2056 represents the voltage of 20.56V; 200 represents current 2.00A; 5408 represents the remaining battery capacity is 5.408Ah; 4593 means the cumulative capacity is 4.593Ah; 9437 represents the watt-hour is 0.09437kw.h; 14353 represents the running time of 14353s; 134 represents the ambient temperature is 34; 4112 represents the power of 41.12W; 0 means the output status is ON; (0-ON, 1-OVP, 2-OCP, 3-LVP, 4-NCP, 5-OPP, 6-OTP, 255-OFF) 0 represents the direction of current, and the current is forward current; (0-forward, 1-reverse) 162 means battery life is 162 minutes; 30682 represents the internal resistance of the battery is 306.82m . 1 represents the communication address; 211 stands for checksum; 3000 means that the over- voltage protection is 30.00V; 100 means that the under-voltage protection is 1.00V; 2000 means that the positive over-current protection is 20.00A;
26
Hangzhou Junce Instruments Co., Ltd.
2000 represents the negative over-current protection is -20.00A; 10000 means
that the over power protection is 100.00W; 151 means that the over-temperature
protection is 51; 10 means the protection recovery time is 10s; 7 means the
delay time is set to 7s; 200 means that the preset battery capacity is 20.0Ah;
120 represents the value of the voltage calibration is 20, and the voltage
fine-tuning increases; (100 means no fine-tuning) 90 represents the value of
the current calibration is 10, and the current fine-tuning is reduced; (100
means no fine-tuning) 101 represents a temperature calibration increase of 1;
(100 means no fine-tuning) 0 stands for reserved settings (coming soon); 0
represents the relay type is normally open; (0-normally open relay, 1-normally
closed relay) 2 represents the current ratio is 2; 12 represents the voltage
curve scale is 12V/div; 13 represents the current curve scale is 13A/div.
4.Mobile APP control
1. App Instructions (For Android) Android mobile app installation and
operation demonstration video: http://68.168.132.244/KG-F_app_Android_EN.mp4
(1) App download
Google download link:
https://play.google.com/store/apps/details?id=com.juntek.vat Server download
link: http://68.168.132.244/app/KG.apk
27
Hangzhou Junce Instruments Co., Ltd.
(If you cannot download, you can ask customer service to obtain the software.)
(2) App Software Installation This software only supports Android 5.0 and more
advanced systems. It
will apply for location during the installation. Please agree and turn on
location services. The Bluetooth module cannot be plugged or unplugged in a
powered state, which may cause damage. This guide corresponds to software
version 1.3.24, and different versions may be slightly different. It is
recommended to upgrade to the latest version for a better experience. The
installation steps are shown in figure 3-5-1 to 3-5-3 below.
Figure 3-5-1 Installation step 1
Figure 3-5-2 Installation step 2
Figure 3-5-3 Installation step 3
(3) Software update Click the App icon, after the App is started, the system
will automatically
detect whether there is a new version of the App in the background, and the
new version will pop up to remind you to update. Apps downloaded from Google
play need to manually detect new versions.
28
(4) App interface
Hangzhou Junce Instruments Co., Ltd.
Figure 3-5-4 Main interface
Figure 3-5-5 Curve
Figure 3-5-6 System Settings
Figure 3-5-7 About us
(5) App operation instructions Connection
Open the APP on the homepage and click the Search in the upper right corner to
search for the corresponding Bluetooth. After clicking the corresponding
Bluetooth, the interface will pop up “Connected”, and the Search in the upper
right becomes disconnected, and the meter model appears, indicating that the
connection is complete. As shown in Figure 3-5-8,
29
Hangzhou Junce Instruments Co., Ltd.
Figure 3-5-9, Figure 3-5-10.
Figure 3-5-8 Connect 1 Figure 3-5-9 Connect 2 Figure 3-5-10 Connect 3 Main
interface operation Output state control switch: Click to control the output
state, and different states will be displayed under different protection
states, as shown in Figure 3-5-11 and 3-5-12. Voltage calibration: Click on
the voltage value, “input voltage calibration” will pop up, enter the
corresponding value for fine-tuning, as shown in Figure 3-5-13 below. Current
calibration: Click on the current value, “input current calibration” will pop
up, enter the corresponding value for fine-tuning, as shown in Figure 3-5-14
below. Voltage range: Click on the voltage dial, the pop-up box “Input voltage
range” will pop out, enter the corresponding value, you can adjust the scale
range of the voltage dial, as shown in Figure 3-5-15 Voltage scale range.
Current scale maximum value: Click on the current scale dial, the pop-up box
“Input the maximum current value” pops out, enter the correct value to adjust
the maximum scale value of the current dial, as shown in Figure 3-5-16 Current
scale maximum value. Zero current: When the load is 0, click the zero current
button, the pop-up box “Zero current” pops out, click OK to memorize the
current in this state, as shown in Figure 3-5-17 Zero current.
30
Hangzhou Junce Instruments Co., Ltd.
Clear data: Click the clear data button, the pop-up box “Clear cumulative data
” will pop up, click OK to reset the time, watt-hour, and cumulative capacity,
as shown in Figure 3-5-18 Clear data.
Remaining capacity percentage: Click the battery icon, the pop-up box “Input
percentage of remaining capacity” will pop up, enter the corresponding value
and click OK to complete the modification, as shown in Figure 3-5-19 Remaining
capacity percentage.
Preset battery capacity: Click the setting button to the right of the preset
battery capacity, the pop-up box “Input preset battery AH value” will pop up,
enter the corresponding value and click OK to complete the modification, as
shown in Figure 3-5-20 Preset battery capacity.
Ambient temperature: Click on the ambient temperature, the “Input temperature
calibration” pop-up box will pop up, enter the correct value and click OK to
complete the modification, as shown in Figure 3-5-21 ambient temperature.
Figure 3-5-11 Output status :Off
Figure 3-5-12 Over-voltage protection
status
Figure 3-5-13 Input Voltage
calibration
31
Hangzhou Junce Instruments Co., Ltd.
Figure 3-5-14 Input Current calibration
Figure 3-5-15 Voltage Range
Figure 3-5-16 Input Max Current
Figure 3-5-17 Zero current
Figure 3-5-18 Clear Accumulated
Data
Figure3-5-19 Remaining Capacity Percentage
32
Hangzhou Junce Instruments Co., Ltd.
Figure 3-5-20 Input Preset battery AH
value
Figure 3-5-21 Input Temperature
calibration
Curve interface operation Voltage and current real-time curve: When the green
circle icon appears
before the text of the voltage and current real-time curve, and the text turns
green, it means that the voltage and current real-time curve is selected,
click the value displayed by the voltage and current, and the voltage or
current real-time curve can be displayed separately. And the opening and
closing of the voltage and current real-time curve is associated with the
output state and the data recording switch of the main interface. Only when
the output state and the data recording switch are turned on (ON), the voltage
and current real-time curve recording will be turned on. The real-time curve
of voltage and current is shown in Figure 3-5-22 below.
Voltage and current history curve: Click the “History: OFF” button to turn on the voltage and current history curve. At this time, a green circular icon appears before the text of the voltage and current history curve, and the text turns green, and the button changes to “History: ON”; click The value displayed by voltage and current can display the voltage or current history curve separately; in the voltage and current history curve graph, you can slide left and right to view the voltage and current curve in the historical time, as shown in Figure 3-5-23 Voltage and current history curve .
Export curve: When viewing the voltage and current history curve, the export curve button will appear. After clicking it, an EXECL file will be generated, and the voltage and current values of the history curve can be viewed at a specific time, as shown in Figure 3-5-24.
33
Hangzhou Junce Instruments Co., Ltd.
Custom export curve: When viewing the voltage and current history curve, the
export curve button will appear. After clicking, select the time range of the
export curve. After the export is confirmed, the export progress will be
displayed on the screen. The completion of the progress bar indicates the
success of the export curve data. Note: Do not close the KG application during
the process of exporting the curve. The maximum time range for exporting is 12
hours, and the longer the exporting time is, the longer the exporting time
will be. When exporting curve data, please wait patiently. The following
figure 3-5-25 is the export curve, As shown in the figure below3-5-26 is the
export curve progress bar.
Custom viewing curve: When viewing the voltage and current history curve, the
View Curve button will appear. After clicking, select the time range for
viewing the curve. After confirming the viewing, the export progress will be
displayed on the screen. After the progress bar is finished, it will enter the
horizontal screen interface. You can view the voltage and current trends
during this period of time, click on the curve, you can easily view the
voltage and current values at that point in time. Note: Do not close the KG
application while viewing the curve. The maximum time range you can select to
view is 12 hours, and the longer you select to view, the longer you can view
the curve. Please be patient when entering to view the curve. As shown in the
figure below, 2-5-27 is the export curve, and the figure 2-5-28 is the
progress bar of the export curve.
Figure 3-5-22 Voltage/current real-time curve
Figure 3-5-23 Voltage/current history
curve
34
Figure 3-5-24 Export curves
Hangzhou Junce Instruments Co., Ltd.
Figure 3-5-25 Export curve
Figure 3-5-26
Figure 3-5-27 Select
Export curve progress bar
Time Range
Figure 3-5-28 view curve
System setting interface operation Over-voltage protection: Click the OVP
button to jump out the “Input over-voltage protection” pop-up window, enter
the corresponding value and click OK to complete the modification, as shown in
Figure 3-5-29 below.
Under-voltage protection: Click the LVP button to jump out the “Input under-
voltage protection” pop-up window, enter the corresponding value and click OK
to complete the modification, as shown in Figure 3-5-30 below.
Input Positive Over-current Protection: Click the OCP discharge over current
protection button, and the “Input positive over current protection” pop-up box
will pop up, enter the corresponding value and click OK to complete the
modification, as shown in Figure 3-5-31 below.
Over-temperature protection: Click the OTP over temperature protection button,
the “Input external over temperature protection” pop-up box will pop
35
Hangzhou Junce Instruments Co., Ltd.
up, enter the corresponding value and click OK to complete the modification,
as shown in Figure 3-5-32 below.
Over power protection: Click the OPP over-power protection button, the “Input
over power protection” pop-up box will pop up, enter the corresponding value
and click OK to complete the modification, as shown in Figure 3-5-33 below.
Input Negative Over-current Protection: Click the NCP charging over-current
protection button, the “Input negative over current protection” pop-up box
will pop up, enter the corresponding value, click OK to complete the
modification, as shown in Figure 3-5-34 below.
Low-capacity reminder: Click on the low-capacity reminder, and the “Input low
capacity reminder” pop-up box will pop up, enter the corresponding capacity
percentage, click OK to complete the modification, and of course, the battery
capacity percentage is lower than When the value is set, a low-capacity
reminder will be displayed in the status bar of the mobile phone, as shown in
Figure 3-5-35 below.
Protection-revert-time: Click the Revert time button, the “Protection-revert-
time” pop-up box will pop up, enter the corresponding value and click OK to
complete the modification, as shown in Figure 3-5-36 below.
Relay mode: Click the relay type button,the “Relay mode” pop-up box will pop
up, and select the relay type that corresponds to the actual one, as shown in
Figure 3-5-37 below.
Set delay time: Click the Delay time button, the “Input delay time” pop-up box
will pop up, enter the corresponding value and click OK to complete the
modification, as shown in Figure 3-5-38 below.
Set address: Click the Set address button, the “Input address” pop-up box will
pop up, enter the corresponding value and click OK to complete the
modification, as shown in Figure 3–5-39 below.
Firmware upgrade: When the measurement module is not powered, short-circuit
the external temperature measurement interface with a wire, and then supply
power to the measurement module. Now turn on the KG-F mobile phone APP,
connect the instrument via Bluetooth, and click the firmware upgrade button on
the mobile phone, “Please input the firmware code. ” pop-up box will pop up,
enter the firmware code “0 0 0”, click OK, the instrument will start the
firmware upgrade, and the indicator light of the measurement module flashes
rapidly at this moment, which means the
36
Hangzhou Junce Instruments Co., Ltd.
firmware upgrade is in progress. Please wait patiently during the upgrade
process, and do not perform other operations. , “Upgrade success” will be
displayed after the upgrade is complete. The firmware upgrade sequence is as
follows: Figure 3-5-40 Firmware upgrade 01, Figure 3-5-41 Firmware upgrade 02,
Figure 3-5-42 Firmware upgrade 03. Firmware update log: http://68.168.132.244
/KG-F_DFU_log_EN.pdf
Low temperature protection: Click on the low temperature protection, the “low
temperature protection ” pop-up box will pop up, you can choose to turn on or
off the low temperature protection, click on, the “input low temperature
protection (-20-20) ” pop-up box will pop up, enter the corresponding value,
Click OK to complete the modification. The low temperature protection function
is not activated by default. When we set the low temperature protection value
to 0°C, if the value is less than 0°C, the protection will be activated. The
upper right corner of the APP main interface will display LTP, which is in the
protection state. As shown in Figure 3-5-43 and Figure 3-5-44 below.
The protection function needs to purchase additional relays and install the
relays in the circuit to protect the circuit.
Figure 3-5-29 Input Over-voltage Protection
Figure 3-5-30 Input Under-voltage Protection
Figure 3-5-31 Input Positive Over-current Protection
37
Hangzhou Junce Instruments Co., Ltd.
Figure 3-5-32 Input External Over
Temperature Protection
Figure 3-5-33 Input Over Power Protection
Figure 3-5-34 Input Negative Over-current
Protection
Figure 3-5-35 Low-capacity reminder
Figure 3-5-36 Protection-revert-time
Figure 3-5-37 Relay Mode
38
Hangzhou Junce Instruments Co., Ltd.
Figure 3-5-38 Input Set delay time
Figure 3-5-39 Input Setting address
Figure 3-5-40 firmware upgrade 01
Figure 3-5-41firmware upgrade 02
Figure 3-5-42 firmware upgrade 03
Figure 3-5-43 Low temperature protection
01
39
Hangzhou Junce Instruments Co., Ltd.
Figure 3-5-44 Low temperature protection
02
About…interface operation Version: you can view the software version
information, as shown in figure 3-5-45.
Official website: after clicking the official website, you can jump to our
company’s official website, as shown in figure 3-5-46 below.
Figure 3-5-45 version
Figure 3-5-46 official website
40
Hangzhou Junce Instruments Co., Ltd.
2. App Instructions (For IOS) Apple Mobile App installation and operation
demonstration video: http://68.168.132.244/KG-F_app_IOS_EN.mp4 (1) APP
download Search for “KG-F series” in the Apple Store to download.
(2) App software installation The software only supports IOS 9.0 or above.
Bluetooth will be accessed
when the software is first linked. Please agree to visit. Bluetooth module can
not be plugged in and out of power state, which will
cause damage. This manual corresponds to software version 1.1.10. Different
versions may be slightly different. It is recommended to upgrade to the latest
software to obtain better use experience.
(3) Software update You can get the latest version from the Apple Store. The
current manual
corresponds to the IOS software version is 1.1.10.
(4) APP interface display
Figure 3-5-47 Main interface
Figure 3-5-48 Curve diagram
Figure 3-5-49 System settings
41
Hangzhou Junce Instruments Co., Ltd.
Figure 3-5-50 About us
(5) App operating instructions The operation of IOS is the same as the
Android.
Figure 3-5-51 Connection 01
Figure 3-5-52 Connection 02
Figure 3-5-53 Connection 03
42
Hangzhou Junce Instruments Co., Ltd.
Figure 3-5-54 Output status is off
Figure 3-5-55 Over-voltage protection status
Figure 3-5-56 Input Voltage
calibration
Figure 3-5-57 Input Current calibration
Figure 3-5-58 Voltage scale range
Figure 3-5-59 Current maximum
43
Hangzhou Junce Instruments Co., Ltd.
Figure 3-5-60 Zero current
Figure 3-5-61 Clear Accumulated
Data
Figure 3-5-62 Input Percentage Of
Remaining Capacity
Figure 3-5-63 Input Preset battery
AH value
Figure 3-5-64 Input Temperature
calibration
Figure 3-5-65 Voltage and current
real time curve
44
Hangzhou Junce Instruments Co., Ltd.
Figure 3-5-66 Voltage and current
history curve
Figure 3-5-67 Export curve data
Figure 3-5-68 Custom export curves
Figure 3-5-69 Customize export data
progress
Figure 3-5-70 Custom view
Figure 3-5-71 Customize Viewing
Waveforms
45
Hangzhou Junce Instruments Co., Ltd.
Figure 3-5-72 Input Over-voltage Protection
Figure 3-5-73 Input Under-voltage Protection
Figure 3-5-74 Input Positive Over current
Protection
Figure 3-5-75 Input External Over Temperature Protection
Figure 3-5-76 Input Over Power
Protection
Figure 3-5-77 Input Negative Over current
Protection
46
Hangzhou Junce Instruments Co., Ltd.
Figure 3-5-78 Low-capacity
reminder
Figure 3-5-79 Input
Protection-revert-time
Figure 3-5-80 Input Set delay time
Figure 3-5-81 Input setting address
Figure 3-5-82 Relay Mode
Figure 3-5-83 firmware upgrade 01
47
Hangzhou Junce Instruments Co., Ltd.
Figure 3-5-84 firmware upgrade 02
Figure 3-5-85 firmware upgrade 03
Figure 3-5-86 Low temperature protection
01
Figure 3-5-87 Low temperature protection 02
Figure 3-5-88 version
Figure 3-5-89 official website
48
Hangzhou Junce Instruments Co., Ltd.
Chapter Troubleshooting
The following lists the possible problems and troubleshooting methods of the
KG-F series during use. When you encounter these problems, please follow the
corresponding steps to deal with it. If you cannot deal with it, please
contact JUNCTEK Hangzhou Junce Instruments Co., Ltd., and please provide your
equipment information.
If the screen is still black after power on, and there is nothing display:
Check whether the power wiring and the wiring between the display board
and the main board are well connected. After finishing the above checks,
restart the meter. If the product still cannot be used normally, please
contact JUNCTEK.
If the screen display is too dark to see clearly: Check whether the brightness
setting value of the LCD screen is too small.
Press the SET button to enter the system setting interface, then press buttons
to make the cursor on the working screen brightness. Press the OKbutton and
then use the buttons to adjust the brightness of the LCD screen to a suitable
state.
Chapter For More product information
For more information about this product, please refer to the relevant manual
(you can log in to the JUNCTEK official website (www.junteks.com) to
download).
“KG-F series operation demonstration video” provides the operation video of
this product.
“KG-F series PC software and communication protocol” provides the
communication protocol.
“KG-F Series User Manual” provides the function introduction and operation of
this product, possible problems and treatment methods during use.
49
Hangzhou Junce Instruments Co., Ltd.
Appendix 1: English and abbreviations of the interface
English Output State!
English abbreviation Output State!
Data Record
Data Record
Series number Display module version
SN Disp Mod Ver.
Measuring module version
Meas Mod Ver.
Remaining AH. value Elapsed AH value
AH.Rmn. AH.Elp.
Battery left
BatLeft
Battery internal resistance Language selection
IntRes Language
Set current to zero
Setcurrentto0
Clear accumulated data
Preset battery AH. value Percentage of remaining capacity Screen brightness
Under-voltage protection
Clear data AH.Preset: AH.Remaining: Brightness LVP
Over-voltage protection
OVP
Forward over-current protection OCP Negative over-current protection NCP
Over power protection
OPP
Over temperature protection Protection-revert-time
OTP Revert Time
Enter standby time
Standby Time
Voltage calibration Current calibration
Voltage Cal. Current Cal.
Temperature calibration
Temp. Cal
Set address
Address
50
Hangzhou Junce Instruments Co., Ltd.
Set delay time Current ratio Monitor function Relay Mode The scale of the voltage curve The scale of the current curve Restore to factory settings About Enter the curve interface OK Cancel Voltage real-time curve Current real-time curve Voltage/current real-time curve Voltage history curve Current history curve Voltage/current history curve Voltage/current Curve Voltage curve Current curve Normally open Normally close Completed Failed NO OFF
Delay Time Current Ratio Monitor Relay Mode Voltage Scale Current Scale Restore all About Curve Interface OK Cancel Voltage Curve Current Curve V/A Real-Time Curve V History Curve A History Curve V/A History Curve Voltage/current Curve V-Curve A-Curve NO NC Completed Failed NO OFF
51