JUNCTEK KG-F Series DC Voltage and Current Meter User Manual

June 12, 2024
JUNCTEK

JUNCTEK KG-F Series DC Voltage and Current Meter

Product Information

  • Product: Hangzhou Junce Instruments Co., Ltd.
  • Trademark: JUNCTEK
  • Manufacturer: Hangzhou Junce Instruments Co., Ltd.

Guaranty and Declaration: Hangzhou Junce Instruments Co., Ltd. guarantees the quality and functionality of the product. This document replaces all previously published documentation.
Safety Requirement: The meter should be operated with caution to avoid personal injury or damage to the meter and connected products. Please follow the safety precautions specified in this manual.

Product Usage Instructions

Observe All Terminal Ratings:
Before connecting the meter, check the ratings and markers on the meter and refer to the manual for more information.

Use Proper Over-voltage Protection:
Ensure that the product is protected from over-voltage, such as lightning strikes, to prevent electric shock hazards.

Do Not Operate Without Covers:
Always operate the meter with covers or panels in place to prevent any damage or malfunction.

Do Not Insert Anything Into the Air Outlet:
Avoid inserting any objects into the air outlet to avoid damaging the meter.

Avoid Circuit or Wire Exposure:
Do not touch exposed junctions and components when the unit is powered on to prevent electric shock.

Do Not Operate With Suspected Failures:
If you suspect any damage to the meter, have it inspected by JUNCTEK authorized personnel before further operations. Only authorized personnel should perform maintenance, adjustment, or replacement.

Provide Adequate Ventilation:
Ensure that the meter is well-ventilated to prevent overheating and damage. Regularly inspect the air outlet and fan for any blockages.

Do Not Operate in Wet Conditions:
Avoid using the meter in humid environments to prevent short circuits and electric shock hazards.

Do Not Operate in an Explosive Atmosphere:
Never operate the meter in an explosive atmosphere to avoid personal injuries and damage to the meter.

Keep Meter Surfaces Clean and Dry:
Regularly clean and keep the surfaces of the meter dry to maintain optimal performance and prevent dust or moisture from affecting its functionality.

Prevent Electrostatic Impact:
Operate the meter in an electrostatic discharge protective environment. Ground both internal and external conductors of cables before making connections to avoid damage from static discharges.

Handle with Caution:
During transportation, handle the meter with care to prevent damage to keys, knobs, interfaces, and other parts on the panels.

Note: For any problems or requirements when using the product or this manual, please contact JUNCTEK by email at junce@junteks.com or visit their website at www.junteks.com.

Hangzhou Junce Instruments Co., Ltd.
I

Hangzhou Junce Instruments Co., Ltd.

Guaranty and Declaration

Copyright Hangzhou Junce Instruments Co., Ltd. all right reserved. Trademark Information JUNCTEK is a registered trademark of Hangzhou Junce Instruments Co., Ltd. Notices JUNCTEK products are covered by P.R.C. patents, issued and pending. This document replaces all previously published documentation. Contact Us If you have any problem or requirement when using our products or this manual, please contact JUNCTEK. E-mail: junce@junteks.com Website: www.junteks.com
I

Hangzhou Junce Instruments Co., Ltd.

Safety Requirement

General Safety Summary
Please review the following safety precautions carefully before putting the meter into operation so as to avoid any personal injury or damage to the meter and any product connected to it. To prevent potential hazards, please follow the instructions specified in this manual to use the meter properly.
Observe All Terminal Ratings To avoid fire or shock hazard, observe all ratings and markers on the meter and check your manual for more information about ratings before connecting the meter.
Use Proper Over-voltage Protection Ensure that no over-voltage (such as that caused by a bolt of lightning) can reach the product. Otherwise, the operator might be exposed to the danger of an electric shock.
Do Not Operate Without Covers Do not operate the meter with covers or panels removed.
Do Not Insert Anything Into the Air Outlet Do not insert anything into the air outlet to avoid damage to the meter.
Avoid Circuit or Wire Exposure Do not touch exposed junctions and components when the unit is powered on.
Do Not Operate With Suspected Failures If you suspect that any damage may occur to the meter, have it inspected by JUNCTEK authorized personnel before further operations. Any maintenance, adjustment or replacement especially to circuits or accessories must be performed by JUNCTEK authorized personnel.
Provide Adequate Ventilation Inadequate ventilation may cause an increase of temperature in the meter, which would cause damage to the meter. So please keep the meter well ventilated and inspect the air outlet and the fan regularly.
Do Not Operate in Wet Conditions To avoid short circuit inside the meter or electric shock, never operate the meter in a humid environment.
II

Hangzhou Junce Instruments Co., Ltd.
Do Not Operate in an Explosive Atmosphere To avoid personal injuries or damage to the meter, never operate the meter in an explosive atmosphere.
Keep Meter Surfaces Clean and Dry To avoid dust or moisture from affecting the performance of the meter, keep the surfaces of the meter clean and dry.
Prevent Electrostatic Impact Operate the meter in an electrostatic discharge protective environment to avoid damage induced by static discharges. Always ground both the internal and external conductors of cables to release static before making connections.
Handle with Caution Please handle with care during transportation to avoid damage to keys, knobs, interfaces, and other parts on the panels.
Notices
1. Do not exceed the voltage and current range of the meter, otherwise the meter will be damaged. 2. The positive and negative poles cannot be reversed, and the reverse connection cannot be handled correctly. 3. The shell of the meter is fragile and easy to corrode. Please don’t hit or close to chemicals to avoid corrosion. 4. Storage temperature: -25~50, and keep the meter in a dry environment. 5. Do not attempt to disassemble the meterit will void the warranty. There are no user-serviceable parts inside the meter. Repairs can only be made through designated repair outlets or sent back to the factory. 6. The display screen is a fragile device, please do not touch or bump it . Please avoid children playing with the meter. When there is dirt on the LCD surface, wipe it carefully with a soft cloth. 7. Please do not move the meter violently to avoid causing irreparable damage to the internal circuit. If the meter does not work properly, please contact the supplier!
III

Inspection

When you get a new KG-F series coulometer, it is recommended that you inspect the meter according to the following steps.
Inspect the Packaging If the packaging has been damaged, do not dispose the damaged
packaging or cushioning materials until the shipment has been checked for completeness and has passed both electrical and mechanical tests. The consigner or carrier shall be liable for the damage to the meter resulting from shipment. We would not be responsible for free maintenance/rework or replacement of the meter.
Check the Contents Please check the contents according to the packing lists. If the meter is
damaged or incomplete, please contact your JUNCTEK sales representative.

KG-F series of Measuring module

1pc

Host

KG-F series of display module

1pc

KG-F series of the Sampler

1pc

3m communication cable

1pc

Temperature Sensor

1pc

4P connection cable

1pc

Accessory 3P connection cable (external relay cable)

1pc

Terminal

1pc

Quick guide

1pc

Inspect the Meter In case of any mechanical damage, missing parts, or failure in passing the
electrical and mechanical tests, contact your JUNCTEK sales representative.

2

Hangzhou Junce Instruments Co., Ltd.

Chapter Overview

Brief Introduction
The KG-F series voltage and current meter is a new type of coulometer that can measure various parameters such as voltage, current, power, charge and discharge AH value, watt-hours, time, etc. It can also set parameters to achieve multi protection functions such as over-voltage protection, under- voltage protection, over-current protection, over-power protection, over temperature protection and time limit protection.The meter can automatically identify the direction of the current, and can monitor the battery capacity in real time and display the measured data in color LCD screen. The KG-F series voltage and current meters add voltage and current curve display and export functions on the basis of the original functions. At the same time, you can use the mobile APP and the computer to control the measuring module, and the firmware can also be updated in real time.

Dimensions

Dimension of display module Note: display module inlay hole size: 85.1 * 58.1mm

Display module hole bitmap

Dimension of measuring module

Hole location of measuring module

3

Hangzhou Junce Instruments Co., Ltd.

Dimension of the 100A sampler

Dimension of the 100A sampler

Dimension of the 400A sampler

Dimension of the 600A sampler

Specification

Model Sampling method Voltage measurement range (External power supply) Voltage measurement range (Self-powered) Voltage resolution Current measurement range Current resolution Relay

KG105F The Shunt
0-120V
10-120V 0.01V 0~50A 0.01A

KG110F The Shunt

KG140F The Shunt

0-120V

0-120V

10-120V

10-120V

0.01V

0.01V

0~100A

0~400A

0.01A

0.1A Optional
4

KG160F The Shunt
0-120V
10-120V 0.01V 0~600A 0.1A

KG610F The Shunt
0-600V
10-120V 0.01A 0~100A 0.01A

Temperature measurement range Capacity display range Power measurement range Power resolution AH measureme nt range Capacity resolu tion Watt-hour measurement range Watt-hour accuracy Time measurement range Time resolution Bluetooth communication distance Voltage accura cy Current accurac y Sampling rate Measuring module power consumption Display module power consumption Over power protection Negative over current protection Forward over

Hangzhou Junce Instruments Co., Ltd.
-20-120 0%~100% 0~180KW
0.01W 0~9999.99AH
0.001AH 0~9999.99kWH
0.001WH 0~999:59:59S
1 second 10m
±2%+3 digits ±5%+10 digits 1 time / second
About 0.4W
About 0.5W 0-99999.99W 0~50A/100A/400A/600A 0~50A/100A/400A/600A
5

current protection Over voltage protection Under voltage protection External over temperature protection Protection recovery time Enter standby time Delay time setting Communication address

Hangzhou Junce Instruments Co., Ltd.
0~120V/600V 0~120V/600V
0-120 0-99s 0-60s 0-99S
There are 99 P01-P99, P00 is the broadcast address

Chapter Meter Introduction 1.Introduction of display module

Item 1 2 3

Figure 2-1-1 KG-F series display module diagram Table 2-1-1 KG-F series display module diagram instructions

Description

Item

Description

LCD screen

4

button

SETbutton

5

OKbutton

button

6

Wiring port

(1).LCD screen The 2.4-inch TFT color LCD screen displays the current function menu
and parameter settings.
(2).SETbutton

6

Hangzhou Junce Instruments Co., Ltd.
Shortly pressing the SET button can quickly switch to the setting interface.
In the main interface, press and hold the SET button can select the address and use to quickly adjust the address. (3).button
When setting parameters, button is used to change the parameters. When setting the system, button is used to select the corresponding system setting. (4).button When setting parameters, button is used to change the parameters. When setting the system,button is used to select the corresponding system setting. In the main interface, press and hold the button will jump out the “Clear data” pop-up window to confirm or cancel clear data. (5).OKbutton In the main interface, shortly pressing the OK button can control on or off of output status and data record. Pressing and hold the OK button can quickly lock or unlock the buttons. (6).Wiring port The wiring port is used to connect to the measuring module.
2.Introduction of the measuring module
7

Hangzhou Junce Instruments Co., Ltd.

Figure 2-2-1 KG-F series measuring module diagram Table 2-2-1 Instructions diagram of the KG-F series measuring module

Item 1
2 3 4

Description Indicator light
2wires&3wires switch Power supply interface
485 communication connection interface

Item 5
6 7 8

Description
485 communication display interface Relay output control interface
Sampler interface
External temperature measurement interface

(1). Indicator light The blinking of the indicator light indicates the working state, and the slow
flashing indicates that the power supply is normal and the measurement is under normal operation.
(2). 2-wire 3-wire power supply switch selection The switch can be selected from external power supply or self-powered,
when the switch is moved up to 2W, it is suitable for self-power supply, and the voltage measurement range is 10-120V, when the switch is moved down to 3W, it is suitable for external power supply, and the external power supply voltage range is 10- 80V, the voltage measurement range is 0-120V.
(3). Power supply interface There are three power supply interfaces, the battery positive interface:
Vsns, the external power supply positive interface: VEXT, and the battery and external power supply negative interface: GND.
(4). 485 communication connection interface The interface used to connect to the computer, and it can also be
connected to another measuring module to realize multi-computer communication. The internal sequence from left to right is: B, A, GND, NC.
8

Hangzhou Junce Instruments Co., Ltd.
(5). 485 communication display interface Used to connect with the display module, when the measuring module has
power supply, the communication interface is live. The internal sequence from left to right is: B, A, GND, +5V.
(6). Relay output control interface The relay output control interface can be used with relays. The internal
sequence from left to right is: GND, OUT, VEXT.
(7). The Sampler interface Used to connect with the sampler for current detection. The internal
sequence from left to right is: VCC, GND, ISNS+, ISNS-.
(8). External temperature measurement interface Used to connect with a temperature sensor to measure the external
temperature. The internal sequence from left to right is: TSNS, GND.
3.Introduction of the Sampler

Figure 2-4-1 KG-F series of the Sampler diagram Table 2-4-1 KG-F series of the Sampler diagram instructions

Item 1

Description The Sampler interface

Item

Description

The Sampler interface Connect to the Sampler interface of the measuring module for current measurement.

9

Hangzhou Junce Instruments Co., Ltd.
4.Introduction of the display interface

Item 1 2 3 4 5 6 7 8

Figure 2-5-1 KG-F series display interface diagram Table 2-5-1 KG-F series display interface diagram instructions

Description

Item

Description

Key lock status
Output status and data record
Communication address

9

Current direction

10

Battery remaining capacity

11

Battery internal resistance

External temperature

12

Battery life

Running time
Communication signal indication
Power

13

Elapsed AH value

14

Remaining AH value

15

Measured current value

Elapsed electric energy

16

Measured voltage value

(1). Key lock status The state in the figure indicates that the button can be operated. After
pressing and holding the OK button, is displayed ,which means the
10

Hangzhou Junce Instruments Co., Ltd.
button is locked.
(2). Output status and data record Shortly press the OKbutton to control the output status and data record
on or off.
(3). Communication address The communication address range is P00-P99. P00 is the broadcast
address, used for multi-machine communication, P01 represents that the current communication address is P01, and view the data of P01 address.
(4). External temperature It displays the ambient temperature. The current display is 19°C, which
means that the temperature of the environment where the external sensor is located is 19°C.
(5). Running time It represents the cumulative value of a single running time.
(6). Communication signal indication The current graph means that the display module and the measuring
module are connected properly. When there is a red × in the graph, it means that the communication is interrupted, and the address needs to be switched or further checked.
(7). Power It measures the current power value, the unit is W.
(8). Elapsed electric energy It represents the actual accumulated electric energy during the running
time, the unit is: KW.h.
(9). Current direction It represents the direction of current passing through the Sampler. The
current icon can also indicate that the battery is in a discharged state.
(10). Battery remaining capacity The remaining capacity of the battery is highlighted more intuitively with
graphics, where the 50% value means that the remaining AH. value of the battery accounts for 50% of the preset battery AH. value. In the main interface, when the remaining capacity percentage is less than or equal to 20%, the graph of remaining AH value is red, when remaining AH value percentage is greater than 20%, the graph of remaining AH value is green.
11

Hangzhou Junce Instruments Co., Ltd.
(11).Battery internal resistance (estimated) It indicates the resistance value of the current battery internal resistance.
(12). Battery left According to the charge and discharge current and capacity, calculate the
battery left and charging time.
(13). Elapsed AH value It represents the elapsed AH value of charge and discharge during this
period running time.
(14). Remaining AH value It represents the remaining capacity of the battery after charging and
discharging, remaining AH. value = preset battery AH. Value – elapsed AH value.
(15). Measured current value It represents the actual current value through the Sampler.
(16). Measured voltage value It represents the voltage value of the voltage measurement interface in the
power supply interface.
ChapterBasic Operation of the Meter
1.System settings
System settings Video: http://68.168.132.244/KG-F_xtsz_EN.mp4 In the main interface, short press the SET button to enter the system setting interface. As shown in Figures 3-1-1 to 3-1-3 below.

Figure 3-1-1 System Setting 01

Figure 3-1-2 System Setting 02

12

Hangzhou Junce Instruments Co., Ltd.
Figure 3-1-3 System Setting 03
(1). Language selection When the “language” selection is green, you can select the language,
short press theOKbutton, the cursor selects Chinese, press the button to switch, there are two language modes, Chinese and English, then short press the OK button after confirming.
(2). Set current to zero In the system setting interface, press the button to switch to
“Set current to zero”, short press theOKbutton and the “Set current to zero” window will pop up, and you can choose to confirm or cancel “set current to zero” by pressing the buttons , When the cursor is selected to confirm, short press theOKbutton, if it displays “Completed” , that means setting current to zero is completed, and the current will becomes 0A.
Precautions for the first startup and / or after restoring factory settings:
When the customer starts the meter for the first time or restores the factory settings, the display will have a small current deviation under normal circumstances when the meter is not connected to the load. At this time, the charging state arrow and the discharging state arrow will jump back and forth, which means that the meter is being calibrated automatically. This phenomenon will last about 30 seconds, and then the current will return to zero automatically. At this time, we need to press the key for a long time to confirm that the current returns to zero, so that the meter can remember the zero state, and the zero current will not jump again no matter whether the machine is switched on or off or restarted.
(3). Clear accumulated data In the system setting interface, press the buttons can switch
to the “Clear Data”. Short press theOKbutton will jump out the “Clear data” pop-up window, and then you can pressbuttons to confirm or cancel clear data. When the cursor is selected to “OK”, short press theOKbutton, if it displays “Completed”, that meas the data is cleared. Time, watt-hour, and
13

Hangzhou Junce Instruments Co., Ltd.
elapsed AH value are cleared at this time, but the measuring is not affected.
(4). Preset battery AH. value In the system setting interface, pressing the buttons can
switch to the “AH. preset”. Short press the OK button, the cursor will select the value, and it can be changed by pressing thebuttons. Pressing the OK button can switch the cursor position, from the smallest digit to the highest digit. At the highest position, press the OK button again to confirm and save the preset battery AH. value.
Before using the battery for the first time, you need to set the effective capacity of the battery and the current remaining power percentage correctly. If the effective capacity of the battery is unknown or the current remaining capacity is unknown, the following operation steps should be followed:
a. Empty the battery pack or put it until you don’t want to discharge, at this time, the battery power is the default 0 power.
b. Enter the preset battery capacity setting in the system settings, and set the capacity value as large as possible (for example, set the estimated value 20Ah to 30Ah).
c. Set the remaining capacity percentage to 0%, and then charge the battery pack.
d. After the battery is fully charged, check the remaining capacity value displayed on the screen, and set this value to the preset battery capacity.
(5). Percentage of remaining AH value In the system setting interface, pressing thebuttons can switch
to the “AH. remaining”. Short press theOKbutton, the cursor will select the value, and it can be changed by pressing the buttons. Press the OKbutton again to confirm that the set remaining AH value accounts for the preset battery AH value. The remaining AH value can be set according to actual usage. And the percentage of remaining AH value is continuously adjustable from 0% to 100%.
In the main interface, when the percentage of remaining AH value is less than or equal to 20%, the graph of remaining AH value displays red, when the percentage of remaining AH value is greater than 20%, the graph of remaining AH value displays green.
After using or replacing the battery for the first time, the capacity value displayed by the cooulometer is not the actual value of the battery, and capacity setting is required.
Suggestion: Set 100% for full charge and 0% for discharge, which is more accurate.
14

Hangzhou Junce Instruments Co., Ltd.
(6). Screen brightness In the system setting interface, press the buttons can switch
to the “Brightness”. Short press the OK button, the cursor will select the value, and it can be changed by pressing the buttons. Press the OKbutton again to confirm the screen brightness. The screen brightness is continuously adjustable from 1% to 100%.
(7). Under-voltage protection In the system setting interface, press the buttons can switch
to the “LVP”. Short press the OK button, the cursor will select the value, and it can be changed by pressing the buttons. Press the OK button can switch the cursor position, from the smallest digit to the highest digit. At the highest position, press the OK button again to confirm the set under-voltage protection value.
When the value is 000.00V, the protection function does not start, when the value is greater than 000.00V, the protection will start. If the current voltage is less than the set under-voltage protection value, the upper left of the screen will display LVP, which is in the protection state. Time stops and no more capacity is calculated.
The protection function needs to purchase the relay (If there is no relay, please do not set this value.) and install a relay in the circuit before it can play a protective role in breaking the circuit.
(8). Over-voltage protection In the system setting interface, pressing the buttons can switch
to the “OVP”. Short press the OK button, the cursor will select the value, and it can be changed by pressing thebuttons. Pressing theOK button can switch the cursor position, from the smallest digit to the highest digit. At the highest position, press the OK button again to confirm the set over-voltage protection value.
When the value is 000.00V, the protection function does not start, when the value is greater than 000.00V, the protection will start. If the current voltage is greater than the set over-voltage protection value, the upper left of the screen will display OVP, which is in the protection state. Time stops and no more capacity is calculated.
The protection function needs to purchase the relay (If there is no relay, please do not set this value.) and install a relay in the circuit before it can play a protective role in breaking the circuit.
(9). Positive over-current protection In the system setting interface, pressing the button to switch
to the discharge over current protection setting. Short press the OKbutton, the cursor will select the value, and it can be changed by pressing the buttons. Pressing theOKbutton can switch the cursor position, from
15

Hangzhou Junce Instruments Co., Ltd.
the smallest digit to the highest digit, at the highest position, press the OK button again to confirm the discharge over current protection.
When the value is 000.00V, the protection function does not start, when the value is greater than 000.00V, the protection will start. If the current discharge current is greater than the set discharge over-current protection value, OCP will be displayed on the upper left of the screen, which is in the protection state. Time stops and no more capacity is calculated.
The protection function needs to purchase the relay (If there is no relay, please do not set this value.) and install a relay in the circuit before it can play a protective role in breaking the circuit.
(10). Negative over-current protection In the system setting interface, press thekey to switch to the
setting of discharge over-current protection. After briefly pressing the OK key, the cursor selects the value,and it can be changed by pressing the buttons. Pressing theOKbutton can switch the cursor position, from the smallest digit to the highest digit, at the highest position, press the OK button again to confirm the set charge over-current protection.
When the value is 000.00A, the protection function will not be started. If the value is greater than 000.00A, the protection function will be started. If the charging current is greater than the set charging over-current protection value, NCP will be displayed at the top left of the screen, which is in the protection state. Time stops and no more capacity is calculated.
The protection function needs to purchase the relay (If there is no relay, please do not set this value.) and install a relay in the circuit before it can play a protective role in breaking the circuit.
(11). Over power protection In the system setting interface, pressing the buttons can switch
to the “OPP”. Short press the OK button, the cursor will select the value, and it can be changed by pressing thebuttons. Pressing theOK button can switch the cursor position from the smallest digit to the highest digit. At the highest position, press the OK button again to confirm the set over power protection value.
When the value is 00000.00W, the protection function does not start, when the value is greater than 00000.00W, the protection will start. If the current power is greater than the set over power protection value, the upper left of the screen will display OPP, which is in the protection state.Time stops and no more capacity is calculated.
The protection function needs to purchase the relay (If there is no relay, please do not set this value.) and install a relay in the circuit before it can play a protective role in breaking the circuit.
(12). Over temperature protection
16

Hangzhou Junce Instruments Co., Ltd.
In the system setting interface, pressing the buttons can switch to the “OTP”. Short press the OK button, the cursor will select the value, and it can be changed by pressing the buttons. Press the OK button again to confirm the set over temperature protection.
When the value is 00, the protection function does not start, when the value is greater than 00 , the protection will start. If the current ambient temperature exceeds the set over-temperature protection value, the upper left of the screen will display OTP, which is in the protection state. Time stops and no more capacity is calculated.
The protection function needs to purchase the relay (If there is no relay, please do not set this value.) and install a relay in the circuit before it can play a protective role in breaking the circuit.
(13). Reverting time protection In the system setting interface, pressing thebuttons can switch
to the “Revert time”. Short press the OK button, the cursor will select the value, and it can be changed by pressing the buttons. Press the OK button again to confirm the reverting time protection.
When the value is 00s, the protection status will continue until the OK button is pressed in the main interface to implement the protection status. If the value is greater than 00s, for example, the protection recovery time is set to 5s, which means 5 seconds after the value returns to the normal value, the protection state is automatically canceled, the circuit is closed and work begins.
(14). Enter standby time In the system setting interface, pressing thebuttons can switch
to the “Standby Time”. Short press theOK button, the cursor will select the value, and it can be changed by pressing the buttons. Press the OK button again to confirm the set standby time.
When the value is 00s, the meter will never stand by and the display will always be on. If the value is greater than 00s, for example, the standby time is set to 5s, it means that the meter will enter the standby state after not operating and the current is 0A for 5s, and the display will go out; if there is operation or current during the period, the display will be on, and no operation is required. And after the current is 0A for 5s, the display screen turns off again.
(15). Voltage calibration In the system setting interface, press the buttons can switch
to the “Voltage Cal.”. Short press theOK button, the cursor will select the value and it can be changed by pressing the buttons. Press the OK button again to confirm the set voltage calibration. The voltage fine-tuning calibration can be performed online. The default voltage calibration is 0, and the maximum voltage calibration range is ±10%.
17

Hangzhou Junce Instruments Co., Ltd.
(16). Current calibration In the system setting interface, pressing thebuttons can switch
to the “Current Cal.”. Short press theOK button, the cursor will select the value and it can be changed by pressing the buttons. Press the OK button again to confirm the set current calibration. The current fine-tuning calibration can be performed online. The default current calibration is 0, and the maximum current calibration range is ± 25%.
(17). Temperature calibration In the system setting interface, pressing the buttons can switch
to the “Temp. Cal.”. Short press theOK button, the cursor will select the value and it can be changed by pressing the buttons. Press the OK button again to confirm the set temperature calibration. The temperature fine-tuning calibration can be performed online. The default temperature calibration is 0.
(18). Set address In the system setting interface, press the button to switch to
the address setting, short press theOK button and select the value with the cursor, press the button to change the communication address, Press theOKbutton to switch from the preset communication address to the current communication address, and press theOK button again to confirm the set address.
Example: As shown in Figure 3-1-3, the set address is 03-01, 01 represents the preset communication address, the preset communication address range is 01-99, 03 represents the current measuring module communication address, and the current communication address range is 00-99 , 00 represents the broadcast address. If we don’t know the current communication address of the measuring module, we can set the current communication address to 00, and set the default communication address to the desired address 03 (for example: 00-03). After setting, press OK to complete the setting. At this time, the current communication address of the measuring module is 03, and the meter can communicate normally. (Use the broadcast address with caution in multi-machine communication)
When multiple measuring modules are connected, the current communication address can be switched to control the measuring module corresponding to the address. At 00, multiple measuring modules can be controlled at the same time.
(19). Set delay time In the system setting interface, pressing the buttons can switch
to the “Delay Time”. Short press the OK button, the cursor will select the value and it can be changed by pressing the buttons. Press the
18

Hangzhou Junce Instruments Co., Ltd.
OK button again to confirm the set delay time. When the delay time is set to 00s and the measured value exceeds the
set protection value, the meter will immediately enter the protection state, when the delay time is greater than 00s, for example, it is set to 5s, when the measured value exceeds the set protection value, the meter enters the protection state after 5s.
(20). Current ratio (this function is only available for KG-H series Hall sensor version) In the system setting interface, press thebutton to switch to the current ratio setting, short press theOKbutton and the cursor selects the value, press the button to change the value, and then press the OKbutton again to confirm the set current ratio. The default is X01. Example: When the wire of the Hall sensor is wound 10 times, the current multiplier is set to X01, the measured current value is displayed as the total current passing through the Hall sensor 10 times, the current multiplier is set to X10, and the measured current value is displayed as passing The current value of one turn of the Hall sensor wire. Can be used to measure small currents.
(21). Monitor function In the system setting interface, pressing thebuttons can switch
to the “Monitor function”. Short press the OK button and then press the buttons to switch on or off. The monitor function must be turned on when using the computer software control. After the monitor function is turned on, the setting function in the display module can only set the system language, working screen brightness and enter the standby time (After the monitoring function is turned on, the display module will not automatically update the data.).
(22).Relay mode In the system setting interface, pressing thebuttons can switch
to the “Relay Mode”. Short press the OK button and then press the buttons to switch between “NO” and “NC”.
After the selection is completed, press the OKbutton again to confirm the type of relay. When using this function, you need to select normally open or normally closed according to the relay type connected during actual wiring.
(23). Restore factory settings In the system setting interface, press the key to switch to
restore the factory settings. After a short press of theOK key, the restore factory settings box will pop up. Through the key, you can select to confirm or cancel the restore factory settings. When the cursor is selected to confirm, press the OK key briefly. if the display shows success, the restore
19

Hangzhou Junce Instruments Co., Ltd.
factory settings will be completed.
2.Wiring method
Wiring instructions video: http://68.168.132.244/KG-F_jxfs_EN.mp4 (1). Self- powered wiring method
If the voltage range of the tested battery is between 10-120V during normal operation,self-powered wiring can be used. First, turn the power supply selection interface switch to “2W”, and then connect the positive terminal of the battery to the power supply interface “Vsns” , This line does not need to be particularly thick,13-16AWG wire is fine. Pay attention to the positive and negative poles of the battery, do not connect the wrong or reverse connection.
The negative pole of the battery is connected to the screw with the battery GND on the sampler, and the negative pole of the charger and the negative pole of the load are connected to another screw of the sampler. It is better to use a copper nose to connect it firmly. When charging, the current direction symbol color is red, and the remaining capacity value increases. When discharging, the color of the current direction symbol is sky blue, and the remaining capacity value decreases.
Figure 3-3-1 Self-powered wiring diagram
20

Hangzhou Junce Instruments Co., Ltd.
(2). External power supply wiring method The measuring voltage range of the external power supply is 0-120V.
First, set the power selection interface switch to “3W”, connect the positive pole of the external power supply to the power interface “VEXT”, and connect the negative pole of the external power supply to “GND”. When wiring, connect the positive pole of the battery to the measurement interface “Vsns” in the power interface. Note that the positive and negative poles of the battery and the external power supply should not be connected incorrectly or reversed.
The negative pole of the battery is connected to the screw with the battery GND on the sampler, and the negative pole of the charger and the negative pole of the load are connected to the other screw of the sampler together. It is better to use copper nose to connect firmly. When charging, the current color is green and the residual capacity value increases. When discharging, the current color is sky blue and the residual capacity value decreases.
Figure 3-3-2 External power supply wiring diagram (3). External power supply wiring methodRelay control
The working power of the relay is provided by an external power supply. If the relay is connected, an external power supply with the same working voltage as the relay should be provided. Connect the control ports of the relay to the “OUT” and “VEXT” of the measurement module Output Control, connect the positive pole of the external power supply to the “VEXT” of the Output
21

Hangzhou Junce Instruments Co., Ltd.
Control, and the negative pole of the external power supply to the “GND”. Note that the positive and negative poles of the battery and the external power supply should not be connected wrongly or reversed.
If you want to control the charging or discharging, you should connect the wiring according to the wiring diagram of external power supply relay in figure 3-3-3. When the relay is closed, the indicator light will be on, and when it is disconnected, it will be off as a prompt.
The negative pole of the battery is connected to the screw with the battery GND on the sampler. The negative pole of the charger and the negative pole of the load are connected to the other screw of the sampler. It is better to use a copper nose to connect it firmly. When charging, the current direction symbol color is red, the remaining capacity value increases. When discharging, the current direction symbol color is sky blue, and the remaining capacity value decreases.
Figure 3-3-3 External power supply wiring diagram (Relay control)
3.Communication protocol control
We only provide communication protocol, customers can carry out secondary development according to communication protocol. 1.Overview
The command line is used as the control command, and the
22

Hangzhou Junce Instruments Co., Ltd.
communication rate is 115200. The PC sends out the command, the meter parses and executes it, and then returns the result to the PC. The different commands are described below. The sending data format is as follows

Start bit
Checksums 1-255

Function code W,R
Spacer ,

Function No. 00~99
Data field See instructions

Connector =
Spacer ,

Address symbol 1-99
Terminator

Spacer ,

Explain: (1) Function code: “W” is the write instruction used to set various parameters, and “R” is the parameter used by the read instruction to return to the machine. (2) Function No.: different values represent different parameter settings. (3) Address symbol: address range is 1-99, 0 is broadcast address. (4) Checksum: the value of the sum is obtained by adding 1 to the remainder of 255 after the sum of all the numbers after the check sum. If the sum of the check is taken as 0, it means that it is not verified. (5) Data field: the data field is equivalent to the number of operations of the command. For example: W20=12162000, the instruction operands are 2000, which means that the over voltage protection is set to 20.00V. (6) Terminator: Each instruction ends with a carriage return + line feed, and represents the carriage return in the ASCII character table (hexadecimal representation is 0x0d). is the newline character in the ASCII character table (in hexadecimal notation is 0x0a). The above two methods indicate carriage return and line feed.
1. W instructions

Function code

Set up

Send command

01

Set address :W01=1,3,2,

10

Turn on the :W10=1,2,1,

output

:W10=1,0,0,

Illustrate
The communication address is set to 2 (this command should be used with caution, if you want to use it, you must know the current communication address and the communication address to be set, it is recommended to use the mobile APP to modify) Turn on output status and data recording Turn off output status and data recording

23

Hangzhou Junce Instruments Co., Ltd.

20

Set over-voltage protection

:W20=1,216,2000,

The over-voltage protection value is set to 20.00V.

21

Set under-voltage protection

:W21=1,216,2000,

The under-voltage protection value is set to 20.00V.

22

Set positive over-current protection

:W22=1,216,2000,

The positive over-current protection value is set to 20A.

23

Set negative over-current protection

:W23=1,216,2000,

The

negative

over-current

protection value is set to 20A.

24

Set over power protection

:W24=1,216,2000,

The over power protection value is set to 20W.

25

Set

over :W25=1,151,150,

temperature

protection

:W25=1,211,210,

The over temperature protection value is set at 50 .
The over temperature protection value is set to 110 .

26

Set protection reverting time

:W26=1,31,30,

The protection reverting time is set to 30s.

27

Set delay time :W27=1,31,30,

The delay time is set to 30s.

28

Set battery capacity

:W28=1,216,2000,

The battery 200.0Ah.

capacity

is

set

at

Set the voltage calibration value to

:W29=1,121,120,

20, and fine-tune to increase the voltage value.

Set the voltage calibration to a value

Voltage

of -20, and fine-tune to reduce the

29

calibration

voltage value.

settings

(The value only represents the

fine-tuning factor, the greater the

:W29=1,81,80,

absolute value of the value, the greater the amplitude of the

fine-tuning)

:W30=1,121,120,

Set the current calibration value to 20, and fine-tune to increase the

current value.

Current

30

calibration

settings

Set the current calibration to a value of -20, and fine-tune to reduce the current value. (The value only represents the

fine-tuning factor, the greater the

:W30=1,81,80,

absolute value of the value, the greater the amplitude of the

24

Hangzhou Junce Instruments Co., Ltd.

fine-tuning)

Set the temperature calibration

31

Temperature calibration settings

:W31=1,104,103, :W31=1,99,98,

value to 3°C, and the temperature increases by 3°C. Set the temperature calibration value to -2°C, and the temperature

decreases by 2°C.

33

Reserved settings (coming soon)

Reserved settings (coming soon)

Reserved settings (coming soon)

:W34=1,2,1,

Set the relay type to normally

34

Set relay type

closed.

:W34=1,0,0,

Set the relay type to normally open.

35

Restore factory settings

:W35=1,2,1,

Perform factory reset

36

Set ratio

current :W36=1,4,3,

The current ratio is set to 3.

37

Setting voltage curve scale

:W37=1,4,3,

The voltage curve scale is set to 03v / Div.

38

Setting current curve scale

:W38=1,4,3,

The current curve scale is set to 03v / Div.

Set

the

percentage of

60

battery

:W60=1,51,50,

50% of battery capacity remaining

remaining

capacity

61

Current to zero

clear :W61=1,2,1,

Perform current clear to zero

Clear

62

accumulated :W62=1,2,1,

Perform clear accumulated data

data

2. R instructions The R command is a read command, and its command format is basically
the same as the write command format. The description will not be repeated here. The following is the data returned by the machine, which is just an example.

Read
Read basic machi ne

PC sending
:R00=1,2, 1,

Machine return
:r00=1,47,1120,100,10 1,

Read instructions
1 represents the communication address; 47 stands for check sum; In 1120, the first one represents Hall

25

Hangzhou Junce Instruments Co., Ltd.

inform ation

Read all meas ured values

:R50=2,2, 1,

Read all setting s

:R51=1,2, 1,

:r50=2,215,2056,200, 5408,4592,9437,14353 , 134,4112,0,0,162,3068 2,
:r51=1,211,3000,100,2 000, 2000,10000,151,10,7,2 00, 120,90,101,0,0,2,12,13 ,

sensor (1-hall sensor, 2-shunt sampler), the second one represents 100V, and 20 represents 200A; 100 represents version 1.00; 101 represents the machine serial number; 2 represents the communication address; 215 represents the checksum; 2056 represents the voltage of 20.56V; 200 represents current 2.00A; 5408 represents the remaining battery capacity is 5.408Ah; 4593 means the cumulative capacity is 4.593Ah; 9437 represents the watt-hour is 0.09437kw.h; 14353 represents the running time of 14353s; 134 represents the ambient temperature is 34; 4112 represents the power of 41.12W; 0 means the output status is ON; (0-ON, 1-OVP, 2-OCP, 3-LVP, 4-NCP, 5-OPP, 6-OTP, 255-OFF) 0 represents the direction of current, and the current is forward current; (0-forward, 1-reverse) 162 means battery life is 162 minutes; 30682 represents the internal resistance of the battery is 306.82m . 1 represents the communication address; 211 stands for checksum; 3000 means that the over- voltage protection is 30.00V; 100 means that the under-voltage protection is 1.00V; 2000 means that the positive over-current protection is 20.00A;

26

Hangzhou Junce Instruments Co., Ltd.
2000 represents the negative over-current protection is -20.00A; 10000 means that the over power protection is 100.00W; 151 means that the over-temperature protection is 51; 10 means the protection recovery time is 10s; 7 means the delay time is set to 7s; 200 means that the preset battery capacity is 20.0Ah; 120 represents the value of the voltage calibration is 20, and the voltage fine-tuning increases; (100 means no fine-tuning) 90 represents the value of the current calibration is 10, and the current fine-tuning is reduced; (100 means no fine-tuning) 101 represents a temperature calibration increase of 1; (100 means no fine-tuning) 0 stands for reserved settings (coming soon); 0 represents the relay type is normally open; (0-normally open relay, 1-normally closed relay) 2 represents the current ratio is 2; 12 represents the voltage curve scale is 12V/div; 13 represents the current curve scale is 13A/div.
4.Mobile APP control
1. App Instructions (For Android) Android mobile app installation and operation demonstration video: http://68.168.132.244/KG-F_app_Android_EN.mp4 (1) App download
Google download link: https://play.google.com/store/apps/details?id=com.juntek.vat Server download link: http://68.168.132.244/app/KG.apk
27

Hangzhou Junce Instruments Co., Ltd.
(If you cannot download, you can ask customer service to obtain the software.)
(2) App Software Installation This software only supports Android 5.0 and more advanced systems. It
will apply for location during the installation. Please agree and turn on location services. The Bluetooth module cannot be plugged or unplugged in a powered state, which may cause damage. This guide corresponds to software version 1.3.24, and different versions may be slightly different. It is recommended to upgrade to the latest version for a better experience. The installation steps are shown in figure 3-5-1 to 3-5-3 below.

Figure 3-5-1 Installation step 1

Figure 3-5-2 Installation step 2

Figure 3-5-3 Installation step 3

(3) Software update Click the App icon, after the App is started, the system will automatically
detect whether there is a new version of the App in the background, and the new version will pop up to remind you to update. Apps downloaded from Google play need to manually detect new versions.

28

(4) App interface

Hangzhou Junce Instruments Co., Ltd.

Figure 3-5-4 Main interface

Figure 3-5-5 Curve

Figure 3-5-6 System Settings

Figure 3-5-7 About us
(5) App operation instructions Connection
Open the APP on the homepage and click the Search in the upper right corner to search for the corresponding Bluetooth. After clicking the corresponding Bluetooth, the interface will pop up “Connected”, and the Search in the upper right becomes disconnected, and the meter model appears, indicating that the connection is complete. As shown in Figure 3-5-8,
29

Hangzhou Junce Instruments Co., Ltd.
Figure 3-5-9, Figure 3-5-10.
Figure 3-5-8 Connect 1 Figure 3-5-9 Connect 2 Figure 3-5-10 Connect 3 Main interface operation Output state control switch: Click to control the output state, and different states will be displayed under different protection states, as shown in Figure 3-5-11 and 3-5-12. Voltage calibration: Click on the voltage value, “input voltage calibration” will pop up, enter the corresponding value for fine-tuning, as shown in Figure 3-5-13 below. Current calibration: Click on the current value, “input current calibration” will pop up, enter the corresponding value for fine-tuning, as shown in Figure 3-5-14 below. Voltage range: Click on the voltage dial, the pop-up box “Input voltage range” will pop out, enter the corresponding value, you can adjust the scale range of the voltage dial, as shown in Figure 3-5-15 Voltage scale range. Current scale maximum value: Click on the current scale dial, the pop-up box “Input the maximum current value” pops out, enter the correct value to adjust the maximum scale value of the current dial, as shown in Figure 3-5-16 Current scale maximum value. Zero current: When the load is 0, click the zero current button, the pop-up box “Zero current” pops out, click OK to memorize the current in this state, as shown in Figure 3-5-17 Zero current.
30

Hangzhou Junce Instruments Co., Ltd.
Clear data: Click the clear data button, the pop-up box “Clear cumulative data ” will pop up, click OK to reset the time, watt-hour, and cumulative capacity, as shown in Figure 3-5-18 Clear data.
Remaining capacity percentage: Click the battery icon, the pop-up box “Input percentage of remaining capacity” will pop up, enter the corresponding value and click OK to complete the modification, as shown in Figure 3-5-19 Remaining capacity percentage.
Preset battery capacity: Click the setting button to the right of the preset battery capacity, the pop-up box “Input preset battery AH value” will pop up, enter the corresponding value and click OK to complete the modification, as shown in Figure 3-5-20 Preset battery capacity.
Ambient temperature: Click on the ambient temperature, the “Input temperature calibration” pop-up box will pop up, enter the correct value and click OK to complete the modification, as shown in Figure 3-5-21 ambient temperature.

Figure 3-5-11 Output status :Off

Figure 3-5-12 Over-voltage protection
status

Figure 3-5-13 Input Voltage
calibration

31

Hangzhou Junce Instruments Co., Ltd.

Figure 3-5-14 Input Current calibration

Figure 3-5-15 Voltage Range

Figure 3-5-16 Input Max Current

Figure 3-5-17 Zero current

Figure 3-5-18 Clear Accumulated
Data

Figure3-5-19 Remaining Capacity Percentage

32

Hangzhou Junce Instruments Co., Ltd.

Figure 3-5-20 Input Preset battery AH
value

Figure 3-5-21 Input Temperature
calibration

Curve interface operation Voltage and current real-time curve: When the green circle icon appears
before the text of the voltage and current real-time curve, and the text turns green, it means that the voltage and current real-time curve is selected, click the value displayed by the voltage and current, and the voltage or current real-time curve can be displayed separately. And the opening and closing of the voltage and current real-time curve is associated with the output state and the data recording switch of the main interface. Only when the output state and the data recording switch are turned on (ON), the voltage and current real-time curve recording will be turned on. The real-time curve of voltage and current is shown in Figure 3-5-22 below.

Voltage and current history curve: Click the “History: OFF” button to turn on the voltage and current history curve. At this time, a green circular icon appears before the text of the voltage and current history curve, and the text turns green, and the button changes to “History: ON”; click The value displayed by voltage and current can display the voltage or current history curve separately; in the voltage and current history curve graph, you can slide left and right to view the voltage and current curve in the historical time, as shown in Figure 3-5-23 Voltage and current history curve .

Export curve: When viewing the voltage and current history curve, the export curve button will appear. After clicking it, an EXECL file will be generated, and the voltage and current values of the history curve can be viewed at a specific time, as shown in Figure 3-5-24.

33

Hangzhou Junce Instruments Co., Ltd.
Custom export curve: When viewing the voltage and current history curve, the export curve button will appear. After clicking, select the time range of the export curve. After the export is confirmed, the export progress will be displayed on the screen. The completion of the progress bar indicates the success of the export curve data. Note: Do not close the KG application during the process of exporting the curve. The maximum time range for exporting is 12 hours, and the longer the exporting time is, the longer the exporting time will be. When exporting curve data, please wait patiently. The following figure 3-5-25 is the export curve, As shown in the figure below3-5-26 is the export curve progress bar.
Custom viewing curve: When viewing the voltage and current history curve, the View Curve button will appear. After clicking, select the time range for viewing the curve. After confirming the viewing, the export progress will be displayed on the screen. After the progress bar is finished, it will enter the horizontal screen interface. You can view the voltage and current trends during this period of time, click on the curve, you can easily view the voltage and current values at that point in time. Note: Do not close the KG application while viewing the curve. The maximum time range you can select to view is 12 hours, and the longer you select to view, the longer you can view the curve. Please be patient when entering to view the curve. As shown in the figure below, 2-5-27 is the export curve, and the figure 2-5-28 is the progress bar of the export curve.

Figure 3-5-22 Voltage/current real-time curve

Figure 3-5-23 Voltage/current history
curve
34

Figure 3-5-24 Export curves

Hangzhou Junce Instruments Co., Ltd.

Figure 3-5-25 Export curve

Figure 3-5-26

Figure 3-5-27 Select

Export curve progress bar

Time Range

Figure 3-5-28 view curve
System setting interface operation Over-voltage protection: Click the OVP button to jump out the “Input over-voltage protection” pop-up window, enter the corresponding value and click OK to complete the modification, as shown in Figure 3-5-29 below.
Under-voltage protection: Click the LVP button to jump out the “Input under- voltage protection” pop-up window, enter the corresponding value and click OK to complete the modification, as shown in Figure 3-5-30 below.
Input Positive Over-current Protection: Click the OCP discharge over current protection button, and the “Input positive over current protection” pop-up box will pop up, enter the corresponding value and click OK to complete the modification, as shown in Figure 3-5-31 below.
Over-temperature protection: Click the OTP over temperature protection button, the “Input external over temperature protection” pop-up box will pop
35

Hangzhou Junce Instruments Co., Ltd.
up, enter the corresponding value and click OK to complete the modification, as shown in Figure 3-5-32 below.
Over power protection: Click the OPP over-power protection button, the “Input over power protection” pop-up box will pop up, enter the corresponding value and click OK to complete the modification, as shown in Figure 3-5-33 below.
Input Negative Over-current Protection: Click the NCP charging over-current protection button, the “Input negative over current protection” pop-up box will pop up, enter the corresponding value, click OK to complete the modification, as shown in Figure 3-5-34 below.
Low-capacity reminder: Click on the low-capacity reminder, and the “Input low capacity reminder” pop-up box will pop up, enter the corresponding capacity percentage, click OK to complete the modification, and of course, the battery capacity percentage is lower than When the value is set, a low-capacity reminder will be displayed in the status bar of the mobile phone, as shown in Figure 3-5-35 below.
Protection-revert-time: Click the Revert time button, the “Protection-revert- time” pop-up box will pop up, enter the corresponding value and click OK to complete the modification, as shown in Figure 3-5-36 below.
Relay mode: Click the relay type button,the “Relay mode” pop-up box will pop up, and select the relay type that corresponds to the actual one, as shown in Figure 3-5-37 below.
Set delay time: Click the Delay time button, the “Input delay time” pop-up box will pop up, enter the corresponding value and click OK to complete the modification, as shown in Figure 3-5-38 below.
Set address: Click the Set address button, the “Input address” pop-up box will pop up, enter the corresponding value and click OK to complete the modification, as shown in Figure 3–5-39 below.
Firmware upgrade: When the measurement module is not powered, short-circuit the external temperature measurement interface with a wire, and then supply power to the measurement module. Now turn on the KG-F mobile phone APP, connect the instrument via Bluetooth, and click the firmware upgrade button on the mobile phone, “Please input the firmware code. ” pop-up box will pop up, enter the firmware code “0 0 0”, click OK, the instrument will start the firmware upgrade, and the indicator light of the measurement module flashes rapidly at this moment, which means the
36

Hangzhou Junce Instruments Co., Ltd.
firmware upgrade is in progress. Please wait patiently during the upgrade process, and do not perform other operations. , “Upgrade success” will be displayed after the upgrade is complete. The firmware upgrade sequence is as follows: Figure 3-5-40 Firmware upgrade 01, Figure 3-5-41 Firmware upgrade 02, Figure 3-5-42 Firmware upgrade 03. Firmware update log: http://68.168.132.244 /KG-F_DFU_log_EN.pdf
Low temperature protection: Click on the low temperature protection, the “low temperature protection ” pop-up box will pop up, you can choose to turn on or off the low temperature protection, click on, the “input low temperature protection (-20-20) ” pop-up box will pop up, enter the corresponding value, Click OK to complete the modification. The low temperature protection function is not activated by default. When we set the low temperature protection value to 0°C, if the value is less than 0°C, the protection will be activated. The upper right corner of the APP main interface will display LTP, which is in the protection state. As shown in Figure 3-5-43 and Figure 3-5-44 below.
The protection function needs to purchase additional relays and install the relays in the circuit to protect the circuit.

Figure 3-5-29 Input Over-voltage Protection

Figure 3-5-30 Input Under-voltage Protection

Figure 3-5-31 Input Positive Over-current Protection

37

Hangzhou Junce Instruments Co., Ltd.

Figure 3-5-32 Input External Over
Temperature Protection

Figure 3-5-33 Input Over Power Protection

Figure 3-5-34 Input Negative Over-current
Protection

Figure 3-5-35 Low-capacity reminder

Figure 3-5-36 Protection-revert-time

Figure 3-5-37 Relay Mode

38

Hangzhou Junce Instruments Co., Ltd.

Figure 3-5-38 Input Set delay time

Figure 3-5-39 Input Setting address

Figure 3-5-40 firmware upgrade 01

Figure 3-5-41firmware upgrade 02

Figure 3-5-42 firmware upgrade 03

Figure 3-5-43 Low temperature protection
01

39

Hangzhou Junce Instruments Co., Ltd.

Figure 3-5-44 Low temperature protection
02
About…interface operation Version: you can view the software version information, as shown in figure 3-5-45.
Official website: after clicking the official website, you can jump to our company’s official website, as shown in figure 3-5-46 below.

Figure 3-5-45 version

Figure 3-5-46 official website
40

Hangzhou Junce Instruments Co., Ltd.
2. App Instructions (For IOS) Apple Mobile App installation and operation demonstration video: http://68.168.132.244/KG-F_app_IOS_EN.mp4 (1) APP download Search for “KG-F series” in the Apple Store to download.
(2) App software installation The software only supports IOS 9.0 or above. Bluetooth will be accessed
when the software is first linked. Please agree to visit. Bluetooth module can not be plugged in and out of power state, which will
cause damage. This manual corresponds to software version 1.1.10. Different versions may be slightly different. It is recommended to upgrade to the latest software to obtain better use experience.
(3) Software update You can get the latest version from the Apple Store. The current manual
corresponds to the IOS software version is 1.1.10.
(4) APP interface display

Figure 3-5-47 Main interface

Figure 3-5-48 Curve diagram

Figure 3-5-49 System settings

41

Hangzhou Junce Instruments Co., Ltd.

Figure 3-5-50 About us
(5) App operating instructions The operation of IOS is the same as the Android.

Figure 3-5-51 Connection 01

Figure 3-5-52 Connection 02

Figure 3-5-53 Connection 03

42

Hangzhou Junce Instruments Co., Ltd.

Figure 3-5-54 Output status is off

Figure 3-5-55 Over-voltage protection status

Figure 3-5-56 Input Voltage
calibration

Figure 3-5-57 Input Current calibration

Figure 3-5-58 Voltage scale range

Figure 3-5-59 Current maximum

43

Hangzhou Junce Instruments Co., Ltd.

Figure 3-5-60 Zero current

Figure 3-5-61 Clear Accumulated
Data

Figure 3-5-62 Input Percentage Of
Remaining Capacity

Figure 3-5-63 Input Preset battery
AH value

Figure 3-5-64 Input Temperature
calibration

Figure 3-5-65 Voltage and current
real time curve

44

Hangzhou Junce Instruments Co., Ltd.

Figure 3-5-66 Voltage and current
history curve

Figure 3-5-67 Export curve data

Figure 3-5-68 Custom export curves

Figure 3-5-69 Customize export data
progress

Figure 3-5-70 Custom view

Figure 3-5-71 Customize Viewing
Waveforms

45

Hangzhou Junce Instruments Co., Ltd.

Figure 3-5-72 Input Over-voltage Protection

Figure 3-5-73 Input Under-voltage Protection

Figure 3-5-74 Input Positive Over current
Protection

Figure 3-5-75 Input External Over Temperature Protection

Figure 3-5-76 Input Over Power
Protection

Figure 3-5-77 Input Negative Over current
Protection

46

Hangzhou Junce Instruments Co., Ltd.

Figure 3-5-78 Low-capacity
reminder

Figure 3-5-79 Input
Protection-revert-time

Figure 3-5-80 Input Set delay time

Figure 3-5-81 Input setting address

Figure 3-5-82 Relay Mode

Figure 3-5-83 firmware upgrade 01

47

Hangzhou Junce Instruments Co., Ltd.

Figure 3-5-84 firmware upgrade 02

Figure 3-5-85 firmware upgrade 03

Figure 3-5-86 Low temperature protection
01

Figure 3-5-87 Low temperature protection 02

Figure 3-5-88 version

Figure 3-5-89 official website

48

Hangzhou Junce Instruments Co., Ltd.
Chapter Troubleshooting
The following lists the possible problems and troubleshooting methods of the KG-F series during use. When you encounter these problems, please follow the corresponding steps to deal with it. If you cannot deal with it, please contact JUNCTEK Hangzhou Junce Instruments Co., Ltd., and please provide your equipment information.
If the screen is still black after power on, and there is nothing display: Check whether the power wiring and the wiring between the display board
and the main board are well connected. After finishing the above checks, restart the meter. If the product still cannot be used normally, please contact JUNCTEK.
If the screen display is too dark to see clearly: Check whether the brightness setting value of the LCD screen is too small.
Press the SET button to enter the system setting interface, then press buttons to make the cursor on the working screen brightness. Press the OKbutton and then use the buttons to adjust the brightness of the LCD screen to a suitable state.
Chapter For More product information
For more information about this product, please refer to the relevant manual (you can log in to the JUNCTEK official website (www.junteks.com) to download).
“KG-F series operation demonstration video” provides the operation video of this product.
“KG-F series PC software and communication protocol” provides the communication protocol.
“KG-F Series User Manual” provides the function introduction and operation of this product, possible problems and treatment methods during use.
49

Hangzhou Junce Instruments Co., Ltd.

Appendix 1: English and abbreviations of the interface

English Output State!

English abbreviation Output State!

Data Record

Data Record

Series number Display module version

SN Disp Mod Ver.

Measuring module version

Meas Mod Ver.

Remaining AH. value Elapsed AH value

AH.Rmn. AH.Elp.

Battery left

BatLeft

Battery internal resistance Language selection

IntRes Language

Set current to zero

Setcurrentto0

Clear accumulated data
Preset battery AH. value Percentage of remaining capacity Screen brightness
Under-voltage protection

Clear data AH.Preset: AH.Remaining: Brightness LVP

Over-voltage protection

OVP

Forward over-current protection OCP Negative over-current protection NCP

Over power protection

OPP

Over temperature protection Protection-revert-time

OTP Revert Time

Enter standby time

Standby Time

Voltage calibration Current calibration

Voltage Cal. Current Cal.

Temperature calibration

Temp. Cal

Set address

Address

50

Hangzhou Junce Instruments Co., Ltd.

Set delay time Current ratio Monitor function Relay Mode The scale of the voltage curve The scale of the current curve Restore to factory settings About Enter the curve interface OK Cancel Voltage real-time curve Current real-time curve Voltage/current real-time curve Voltage history curve Current history curve Voltage/current history curve Voltage/current Curve Voltage curve Current curve Normally open Normally close Completed Failed NO OFF

Delay Time Current Ratio Monitor Relay Mode Voltage Scale Current Scale Restore all About Curve Interface OK Cancel Voltage Curve Current Curve V/A Real-Time Curve V History Curve A History Curve V/A History Curve Voltage/current Curve V-Curve A-Curve NO NC Completed Failed NO OFF

51

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)