AirtraQ A-390 WiFi Camera Instruction Manual
- June 12, 2024
- AirtraQ
Table of Contents
- Instructions For Use
- A-390 INTENDED USE AND BASIC INFORMATION
- A-390 COMPONENTS
- A-390 SWITCHES, INDICATORS AND CONNECTORS
- A-390 BATTERY CHARGE AND CHARGE STATUS CHECKING
- MOUNTING THE A-390 ONTO THE AIRTRAQ
- TURN ON/OFF.
- TOUCH PANEL & MENUS
- AIRTRAQ CAM APP FOR PC & MOVILE DEVICES
- Downloading app:
- INITIAL SET UP. CAMERA SETTINGS
- PAEDIATRIC MODE
- IMAGE ROTATION
- RECORD , AUTO-RECORD & SNAPSHOT
- PLAY RECORDED VIDEOS AND VIEW SNAPSHOTS ON A-390
- DOWNLOAD VIDEOS AND SNAPSHOTS TO A PC /MAC
- CONNECTING WI-FI LIVE VIDEO & RECORDING ON TABLET / PHONE /PC
- VIEW SYSTEM INFORMATION
- UPDATING A-390 INTERNAL SOFTWARE
- TROUBLESHOOTING
- CLEANING INSTRUCTIONS
- OPERATING, STORAGE AND TRANSPORT CONDITIONS
- TECHNICAL DATA
- REGULATORY
- DISPOSAL INSTRUCTIONS
- MANUFACTURER’S WARRANTY
- IMPORTANT SAFETY INSTRUCTIONS
- USER ASSISTANCE INFORMATION
- GLOSSARY OF SYMBOLS
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
PL-C390-921 Rev. 2.2 2022-04
A-390 WI-FI CAMERA For Software version
2.00 and posteriors
For Serial Numbers higher than M-20000uctio
Instructions For Use
The Instructions for Use may be updated without further notice. The latest
version is available on www.airtraq.com.
To check for latest software versions available connect A-390 to “Airtraq Cam”
application on PC or visit www.airtraq.com
A-390 INTENDED USE AND BASIC INFORMATION
Code: A-390
Description: Airtraq Wi-Fi Camera
The A-390 is an accessory for the Airtraq Video Laryngoscope. It is compatible
with all Airtraq Video laryngoscopes models.
Intended Use: The A-390 shall be attached to an Airtraq Video
Laryngoscope and its intended purpose is to provide full colour, real-time
images on its 2.8 inch integrated touch screen.
A-390 is powered by an internal rechargeable 3.7 V Li-Po battery. A-390 does
not work while connected to a battery charger. A-390 battery shall be charged
at least 1.8 m. away from the patient. In case A390 is not operative by any
cause (battery not charged, etc) the user can still intubate looking directly
through the eyecup of the Airtraq Video Laryngoscope.
Auxiliary Functions: The A390 provides the following additional functions that
are not included in its intended purpose: sending real time images via Wi-Fi
to a mobile device; video playback on A390´s screen; wire connection to a PC
for video downloading or software update. A temporary failure of any of these
functions, due to an abnormal circumstance, will not increase the patient´s
risk and will not compromise A390´s intended use, therefore it would be
acceptable.
A-390 COMPONENTS
2.1. A-390 Camera
CMOS video camera with integrated touch screen and rechargeable Li-Po Battery.
2.2. USB-C cable
To be used to connect the A-390 to a PC for video downloading or for camera’s
settings administration.
May also be used to charge the A390 directly from a USB power source.
2.3. Silicon protective case
Provides additional protection against shocks.
It can be easily installed onto or removed from the camera.
2.4. Battery charger
Must be connected to mains supply to charge the 3.7 volts, Li-Po Battery.
2.5. Docking station
Accessory to facilitate charge of A-390.
Docking Station can also be used to charge the A-360 Airtraq Wi-Fi Camera.
The mains power supply is connected to the Docking Station through its micro
USB-C connector.
2.6. Neoprene pouch
Can be used to store some of the A390 accessories.
2.7. Storage case (two sizes)
Can be used to store all or some of the above items. It is recommended that
the A-390 be stored and transported inside the storage case to protect the
camera while not in use.
A-390 can be charged when it is inside the storage case by connecting the
USB-C cable directly to
the camera and zipping the case partially closed allowing room for the cable
to exit the case and
plugged into power.
A-390 SWITCHES, INDICATORS AND CONNECTORS
A. Automatic power button: This button is automatically pushed when the A-390
is inserted onto any Airtraq. It can also be pushed manually to turn on the
A-390. To turn off the A-390 press the corresponding icon on the menu.
B. Battery charge connector: Brass surface connectors that connect to docking
station for battery charge.
C. USB-C female connector connector: to connect to PC or to battery charger.
D. Battery charge connector: Brass surface connectors that connect to Camera
for battery charge.
E. USB-C female connector: to connect to mains power supply.
A-390 BATTERY CHARGE AND CHARGE STATUS CHECKING
The A-390 DOES NOT WORK WHEN IT IS BEING CHARGED.
IT SHOULD BE CHARGED AT LEAST 1.8 M AWAY FROM THE PATIENT
It is recommended to perform a full charge/discharge cycle of the battery
before starting to use the A-390 for clinical purposes.
After a full charge, the A-390 can operate for around 120 minutes when Wi-Fi
is deactivated and for around 90 minutes when Wi-Fi is activated. Recharging
takes approximately 110 minutes.
Without use, the battery will discharge at a rate of around 1% per day.
If the A-390 can be turned on, then a minimum of 10 minutes of operating time
is guaranteed.
When the A-390 is turned on, a battery charge status icon and available
minutes of operating time appear on the screen.
To charge battery:
- Connect the battery charger to Docking station and to mains electrical supply.
- Place A-390 onto docking station.
- A-390 can also be charged connecting directly the battery charger provided to the USB-C connector of the A390.
Upon starting to charge battery the A-390 will automatically turn Off.
Also, touching the screen while charging or when fully charged will show
available minutes of operating time and a battery icon.
When in use, the battery charge indicator on the header of the screen will
show available minutes of operating time and a battery icon in different
colours.
Service life for the rechargeable battery depends upon the number of charge-
discharge cycles performed. Battery can be expected to lose around 25% of its
capacity after 500 charge cycles.
At around 75%, (after 500 cycles) the camera should continue working well; it
just won’t provide as many minutes of service per charge.
The A-390´s battery can only be replaced by the manufacturer. The A-390 should
never be opened by unauthorized personnel. It could result in a HAZARDOUS
SITUATION and will void the warranty.
MOUNTING THE A-390 ONTO THE AIRTRAQ
To mount the A-390 onto the Airtraq:
- Remove eyecup from Airtraq.
- Place A-390 onto Airtraq’s proximal end. Make sure the A-390 is fully inserted.
- A-390 will automatically turn on and start in Live Video mode.
- If during intubation, A-390 accidentally becomes partially dislodged from the Airtraq blade, a warning message will appear on the screen indicating ”Blade detached”. This is to notify the user that the camera is not fully seated on the Airtraq blade. If A-390 accidentally becomes partially dislodged and A-390 is recording video, and within five seconds the user re-attaches the A-390 securely to the blade, then recording will restart automatically maintaining the same video file. If the user does not correct the ”Blade detached” condition within 5 seconds then the recording will be stopped. To remove A-390 from the Airtraq once intubation is finished lift up and pull A-390 away from Airtraq.
TURN ON/OFF.
A-390 can be started in two different ways, in both methods it will start in “Live Video” mode:
- Automatically by sliding A-390 onto an Airtraq blade.
- Pressing the automatic power button.
- Boot up will take approximately 7 seconds.
To turn off the A-390 press the “Power off” icon on the main menu. The unit
will show a message to the user and a “cancel” icon will appear and it will
automatically shut down in 3 seconds unless the user cancels the action.
A-390 automatically shuts off when there is no action from the user for 30
minutes when inserted onto an Airtraq blade or 5 minutes if not. The A-390
will show a shutdown message and the user has 5 seconds to cancel the process
if wished.
TOUCH PANEL & MENUS
Press the action icons on the screen to select an action (touch panel also
works when icons are pressed using gloves).
When in Live Video, pressing the “Back” (1/2) Icon will make A-390 to return
to the main menu. Navigation is shown below:
AIRTRAQ CAM APP FOR PC & MOVILE DEVICES
8.1 AIRTRAQ CAM APP FOR PC
Functions:
- Initial set up (A-390 connected to PC through USB).
- Download videos recorded in A-390 (A-390 connected to PC through USB).
- Live Video on PC connecting via Wi-Fi to A-390.
- Recording Video in PC while A-390 and PC are connected via Wi-Fi.
Environments & versions:
- Windows and MacOS
Downloading app:
Download to PC The “Airtraq Cam” app from the downloads section on the Airtraq website ( www.airtraq.com ) and install it on the PC.
8.2 AIRTRAQ CAM APP FOR MOBILE DEVICES
Functions:
- Live Video on Mobile Device connecting via Wi-Fi to A-390.
- Recording Video in Mobile Device when A-390 is connected via Wi-Fi.
Environments & versions:
- Android
- iOS (iPhone & iPad)
Downloading app:
Before starting to use a tablet/ smartphone with A-390, download from “App Store” SM or “Google Play” TM the “Airtraq Cam” app and install it. Make sure that you have the latest available version installed on your device before attempting to connect to it.
INITIAL SET UP. CAMERA SETTINGS
Initial set up must be performed connecting the A-390 to a PC through the USB
cable provided and using the “Airtraq Cam” App. Only one A-390 at a time can
be connected to the PC.
Some options can also be selected directly on the Settings screen of the
A-390.
When the App is launched a screen will show up with the A-390 connected to the
PC. App will prompt the user to register the warranty. Manufacturer recommends
registering the warranty before starting to use A-390.
- Select Language: Scroll and select the desired language.
- Date and Time: Date and time are used to name video files recorded. They are automatically updated when connecting A390 to a PC and opening the Airtraq Cam App.
- Set Cam ID: Cam ID is used to name the Wi-Fi Network generated by the A-390. Any combination of a maximum of 10 letters/numbers.
- Set optional Wi-Fi Password: Optionally, to better protect patient’s information activate Wi-Fi password. Wi-Fi password is used to block non-authorized devices from connecting to A-390. User can set it to any combination of five letters/numbers. Password protection can be set on or off. It is off by default. This option can also be set in the A-390 Settings screen. When Password is activated the Wi-Fi icon includes a padlock .
- Set limitation to Wi-Fi Power (Canada only): Users can limit the maximum RF power in order to comply with Canadian regulations Auto-recording options can be set in the A-390 camera. Press “Settings” icon.
- Set Auto Recording Mode: The auto-recording function will automatically start a new video recording every time the A-390 goes into Live Video. Set it on or off as desired. This option can also be set in the A-390 Settings screen.
- Set Auto-recorded video maximum size: select the maximum video file size: (5 min [15 mb], 6min [18 mb], 10 min [30 mb], 15 min [45 mb], 20 min [60 mb], 30 min [90 mb]). This selection can also be set in the A-390 Settings screen. When the maximum size is reached, video recording is automatically stopped. Videos can be downloaded using “Airtraq Cam” in order to compress them and reduce its size..
PAEDIATRIC MODE
When using a pediatric Airtraq Blade the A-390 must be manually set to
Pediatric mode, other wthe image will be too bright.
Upon pressing the “Start Pediatric” icon, a message saying “Pediatrics Mode
On” will appear on screen and the icon will change to “Stop Pediatric” .
The message will blink for the first 3 seconds.
If the A-390 is set to pediatric mode when using an adult Airtraq blade the
image will becomedark.
IMAGE ROTATION
Depending on intubation position, the user can rotate the image 180 degrees to correct the glottis orientation by pressing the rotate icon .
RECORD , AUTO-RECORD & SNAPSHOT
RECORDING AND SNAPSHOT ARE DEACTIVATED WHEN A-390 CAMERA IS NOT INSERTED ONTO
AN AIRTRAQ BLADE.
RECORDING AND SNAPSHOT ARE DEACTIVATED WHEN WiFi IS ACTIVATED. The user has
the option to record or take snapshots in the mobile device or PC that is Wi-
Fi connected to the A-390 through the corresponding recording or snapshot
button of the “Airtraq Cam” App.
To start recording when in Live Video press icon . While recording, the icon
will change to and a timer will appear on the header of the screen. To stop
recording press the “Stop Recording” icon . The screen will show a message
informing
that recording will stop in 5 seconds unless user presses “Cancel”. When Auto-
Recording function is activated, it will automatically start a new video
recording every time the A-390 enters into Live Video. Video files are
recorded in avi format and stored in A-390’s internal memory. A-390 is able to
store 200 videos or around 6 hours of video. A Snapshot can be taken by
pressing in the centre of the image of the live video screen. The Snapshot
feature also works while recording video. Multiple Snapshots can be taken at a
maximum speed of 1 picture every 2 seconds. Snapshots are stored on the
A-390´s internal memory as a jpeg file and can be viewed in the snapshot
gallery Free memory available is displayed in the Setting/ info Screen. If
when the A390 is turned on memory is full or there are more than 199 videos or
snapshots, then A-390 automatically deletes the oldest videos/snapshots stored
until there is enough free storage to record a 30 minutes video or to record
10 snapshots. To free up memory space the user should delete videos/snapshots
using “Airtraq Cam” App for PC/MAC. A-390 can store up to 200 snapshots.
PLAY RECORDED VIDEOS AND VIEW SNAPSHOTS ON A-390
Select “Videos ” from Main Menu.
- The screen will display a list of the recorded videos showing starting date/time and duration. Last recorded video will be first on the list.
- Navigate through different pages using Up and Down navigation icons and select the desired video.
- Action icons provide the following functions: play/pause and fast forward/rewind. Select “Snapshots” at the bottom of the Videos Menu.
- The screen will display a list of Snapshots taken showing date/time with the most recent Snapshots on top of the list.
- Navigate through different pages using Up and Down navigation icons and select the desired Snapshot.
DOWNLOAD VIDEOS AND SNAPSHOTS TO A PC /MAC
Videos are recorded into the A-390 in avi format and named automatically with the device ID and date and time at which they were recorded “deviceid_yy_mm_dd_xxh_xxm_xxs.avi”. Snapshots are stored in jpeg format and named automatically with the device ID, date and time at which they were taken “deviceid_yy_mm_dd_xxh_xxm_xxs.jpeg”. Videos and snapshots are downloaded connecting the A-390 to a PC (Windows or Mac) through the USB cable provided and using the “Airtraq Cam” App for Windows or Mac. The user can select to download videos in its original resolution or to download compressing the video file to reduce its size. When the App is launched, a screen will show up with the A-390 connected to the PC. Only one A390 at a time can be connected to the PC. Click on the download icon selected and the App will display the list of videos and snapshots stored into the A-390. Select the videos and snapshots to be downloaded. Alternatively, click on “Download All” icon. Select video or snapshots to be deleted and press the delete icon. Alternatively, click on “Delete All” if you want to remove all stored videos and snapshots for the connected A-390.
CONNECTING WI-FI LIVE VIDEO & RECORDING ON TABLET / PHONE /PC
To enable Wi-Fi go to Main Menu and press Wi-Fi icon . Wi-Fi icon will turn to
and an icon will appear on the header of the screen.
In case Wi-Fi was activated the last time that A-390 was turned off, then
during booting process A390 will automatically activate WiFi. A-390 CAN ONLY
BE WI-FI CONNECTED TO ONE DEVICE AT A TIME. Before connecting to another
device, the user should exit the “live Video” screen on “Airtraq Cam” App be.
When the A-390 is connected to an external device, and thus it is not possible
to connect another external device, the A-390´s header displays a “connected”
icon Each A-390 creates an ad-hoc Wi-Fi network named AIQ+DeviceID+5 numbers
automatically generated. To connect a Tablet /smartphone to an A390, launch
“Airtraq Cam” App, add on the external device the A-390 that you want to
connect with by entering its Cam Id, either manually or by scanning its QR. QR
can be found on the Settings screen of A-390. After adding the A-390 to the
Tablet/Smartphone, select the A390 from the list and click on “Connect”.
Repeat this procedure for every new camera that needs to be connected to the
external device. For the best image, increase screen brightness in your mobile
device.
To avoid incoming calls, turn airplane mode on and then turn Wi-Fi on.
“Airtraq Cam” App for mobile devices allows the user to record videos and take
snapshots directly on the mobile device. To connect a PC download and install
“Airtraq Cam” App for PC. Upon launching the App, it will search automatically
for A390 that are emitting Wi-Fi signal near the PC. Select the camera that
you want to connect to. If Wi-Fi password required has been set to ”on”, enter
it now. Click on Live Video and the PC will display real time video captured
by A-390. The app will allow user to record Video directly on the PC.
VIEW SYSTEM INFORMATION
Select “Settings” in main menu to see system information on the A-390 Screen. A-390 settings Screen allows to modify: Activation of Wi-Fi Password, Activation of Auto Recording Mode and Selection of Auto-recorded video maximum size.
UPDATING A-390 INTERNAL SOFTWARE
A-390 internal software can be upgraded using the “Airtraq Cam” app for
PC/Mac.
The PC should have Internet connection in order to automatically retrieve the
latest version of A-390 internal software from Airtraq website (
www.airtraq.com ).
Connect A-390 to a PC (Windows or Mac) through the USB cable provided. When
“Airtraq Cam” app is launched a screen will appear with the A-390 connected to
the PC. If there is a newer version available a message will prompt the user
to update.
It is strongly recommended to check periodically if a new software version for
A-390 is available.
TROUBLESHOOTING
Make sure that A-390 battery is properly charged. Connect A-390 to a PC, open
“Airtraq Cam” and make sure that the latest version is installed.
To reset A390 connect A-390 to a PC, open the “Airtraq Cam” app and click on
“RESET A390” in the app toolbar and carefully follow the prompts. The reset
process can be performed even if the A-390 is not recognized by the PC as a
mass storage device and its screen is totally black.
Reset process allows the user to perform only step 1 to keep all the videos
and snapshots in internal memory or to perform also step 2 which deletes all
files stored in internal memory.
CLEANING INSTRUCTIONS
The A-390 is a non-critical device. Recommended cleaning is low-level disinfection between each patient use. Be sure to follow your institution’s specific cleaning procedures in consultation with this manual.
-
Remove A-390 from the Airtraq and turn it off.
-
Cleaning: Use clean cotton gauze pads that are saturated with the cleaning solution to wipe down the exterior surfaces of the A-390. Use soft brushes with the cleaning solution to remove any residues from areas that cannot be reached with the gauze pads. Be sure to remove all visible contaminant or foreign matter. Be careful to keep running liquid off the surfaces. The following cleaning solutions may be used:
a. Enzymatic Cleaning Solutions (e.g.: ENZOL TM ).
b. Neutral pH soap and water.
c. Sodium bicarbonate solution (8- 10 %). -
The following disinfection agents may be used:
a. Peracetic Acid Solution (0.08%).
b. Isopropyl Alcohol (70%).
c. Solution containing 70% Isopropyl Alcohol and 2% chlorhexidine (e.g. Clinell Wipe).
d. Solution containing chlorine dioxide (e.g. Tristel Wipe).
e. PDI Sani-Cloth® Germicidal Wipes (AF3, Bleach, Plus or Super Sani-Cloth®). -
Blot dry the A-390 using an individual sterile surgical towel.
-
Silicone case can be cleaned using the following disinfection agents:
a. Isopropyl Alcohol (70%).
b. Solution containing 70% Isopropyl Alcohol and 2% chlorhexidine (e.g. Clinell Wipe).
Isopropyl alcohol can cause slight temporary swelling of the silicone, but as long as it is allowed to evaporate completely it will not deteriorate silicone case.
Caution:
• Do not autoclave.
• Do not rinse under running water.
• Do not soak in liquids.
• Avoid liquid or moisture from going into the A-390.
• Avoid touching the lens of the A-390.
OPERATING, STORAGE AND TRANSPORT CONDITIONS
- Storage and Transport Conditions : Temperature: between -5ºC/23ºF and 55ºC/131ºF; Humidity: between 10% and 95% ; Pressure: between 500 to 1060 hPa
- Operating Conditions: Temperature: between -5ºC/23ºF and 40ºC/104ºF; Humidity: between 10% and 95% ; Pressure: between 500 to 1060 hPa
- Battery Charging Conditions: Temperature: between -5ºC/23ºF and 40ºC/104ºF; Humidity: between 10% and 95% ; Pressure: between 500 to 1060 hPa
TECHNICAL DATA
21.1. A-390 Camera
Electrical supply input ratings: 5 V DC, 2.0 A
Supply Voltage: Rechargeable Li-Po 3.7 V, 1400mAh.
Baseband Modulation – Wi-Fi – 802.11b: DSSS
Output RF Power: + 10 dBm Typ.
Transmission range: 5 m (approx.).
Screen: 2.8” TFT-LCD, 640 x 480 RGB.
Touch Panel: OGS type 2.8” Projected Capacitive Touch Panel.
Video Sensor: Digital CMOS video sensor, VGA 640 x 480.
Video resolution: 480 x 480.
Dimensions: 61 mm x 91 mm x 64 mm.
Weight: 110 grams.
21.2 Stand-alone Battery Charger
Electrical supply input ratings: 100-240 V ~, 50-60 Hz, 0.7 A.
Electrical supply Output ratings: 5.0 V = , 2.0 A, 10.0 W
EMC: EN55024, EN55022/B, EN 60601-1-2.
UL recognized component. Valid for USA and Canada.
Micro USB-C male connector.
Cable length: 1.9 m.
REGULATORY
This device, when working as an accessory for Airtraq Video Laryngoscopes,
meet the requirements of ISO 8600, ISO 14971, ISO 62304, ISO 62366, ISO 10993,
European Medical Devices Regulation MDR (EU) 2017/745 of 5 April 2017. The
Declaration of Conformity (DoC) is available from the manufacturer.
This device complies with IEC 60601-1, IEC 60601-1-2 and IEC 60601-2-18
International Standards for Medical Electrical Equipment.
This device is in compliance with the RoHS European Directive.
IEC 60601-1 and IEC 60601-1-2 NOTICES AND WARNINGS
This device is specified as Internally Powered equipment, accessory, type BF
Applied Part, IPXO, ordinary equipment, not intended for use with flammable
agents, continuous operation.
According to IEC 60601 Standard A-390 together with Airtraq Laryngoscope
comprise a Medical Electrical System. It complies with requirements regarding
Electromagnetic Compatibility. Detailed results of Emissions and Immunity
Tests are available from the manufacturer.
A-390 is a portable RF communications equipment and can affect Medical
Electrical Equipment.
This equipment is intended for use by healthcare professionals only. This
equipment may cause radio interference or may disrupt the operation of nearby
equipment. It may be necessary to take mitigation measures, such as,
reorienting or relocating the A-390.
This device shall only be used with cables and accessories provided by the
manufacturer. Using other cables or accessories may result in increased
emissions or decreased immunity of the ME System.
This device is classified as Internally Powered equipment powered by 3.7 V
rechargeable battery for protection against electric shock.
The A-390 is classified as BF Applied Part.
This device is classified as IPX0, ordinary equipment, for degree of
protection against ingress of water, as per IEC 60529 International Standard
(Degrees of Protection provided by enclosures).
No sterile parts. Mode of operation continuous. This device has non-ionizing
radiation.
FCC AND INDUSTRY CANADA NOTICES AND WARNINGS
Caution: The user is cautioned that changes or modifications not expressly
approved by the party responsible for compliance could void the user’s
authority to operate the equipment.
This device complies with Part 15 of the FCC Rules and Industry Canada
licence-exempt RSS standard(s). Operation is subject to the following two
conditions: (1) this device may not cause harmful interference, and (2) this
device must accept any interference received, including interference that may
cause undesired operation.
NOTE: This equipment has been tested and found to comply with the limits
for a Class B digital device, pursuant to Part 15 of the FCC Rules. These
limits are designed to provide reasonable protection against harmful
interference in a residential installation. This equipment generates, uses and
can radiate radio frequency energy and, if not installed and used in
accordance with the instructions, may cause harmful interference to radio
communications. However, there is no guarantee that interference will not
occur in a particular installation.
If this equipment does cause harmful interference to radio or television
reception, which can be determined by turning the equipment off and on, the
user is encouraged to try to correct the interference by one or more of the
following measures:
— Reorient or relocate the receiving antenna.
— Increase the separation between the equipment and receiver.
— Connect the equipment into an outlet on a circuit different from that to
which the receiver is connected.
— Consult the dealer or an experienced radio/TV technician for help.
FCC & IC Radiation Exposure Statement:
This equipment complies with FCC and Canada radiation exposure limits set
forth for an uncontrolled environment. This equipment should be installed and
operated with a minimum distance of 15mm between the radiator and your body.
This transmitter must not be co-located or operating in conjunction with any
other antenna or transmitter.
DISPOSAL INSTRUCTIONS
The A-390 and related accessories contain batteries, that cannot be removed, and other environmentally hazardous materials. When A-390 has reached the end of its useful service life, it must be disposed of in accordance with WEEE or other national and local regulations on disposal of electronic and electrical equipment.
MANUFACTURER’S WARRANTY
Warranty & Service
The manufacturer warrants this device against faulty materials or
manufacturing defects for a period of one (1) year from the date of purchase,
provided that it is used in accordance with the procedures set up in the
instructions for use. This warranty is only applicable if the device has been
bought from an authorized distributor and its warranty has been registered
using the “Airtraq Cam” App for PC. The manufacturer disclaims all other
warranties, whether expressed or implied, including, without limitation the
warranties of merchantability or fitness for a particular purpose. Handling
Instructions for Defective Devices
- Return units must be packed in original and secured packaging materials. All accessories of the device have to be sent back.
- Customer must have registered the device warranty prior to the date of failure and provide serial number of the A-390, description of the problem and proof of purchase.
- In all cases when the device shows defects which are not covered by the warranty, the customer will remain responsible for all handling costs and transportation fees.
Exclusions from Warranty
- Product, where the serial number has been defaced, modified or removed.
- Damage, deterioration or malfunction resulting from:
A. Accidental drop of the device.
B. Misuse, neglect, fire, water, lightning or other acts of nature.
C. Repair or attempted repair by anyone not authorized by Supplier.
D. Modification by anyone not authorized by Supplier.
E. Any damage of the product due to shipment.
F. Causes external to the product, such as: over powering the unit, electric power fluctuations or failure, pouring liquids over the device, etc.
G. Normal wear and tear.
H. Any other causes which does not relate to a product defect.
IMPORTANT SAFETY INSTRUCTIONS
- Keep, read and follow all of these instructions.
- To minimize risk of contamination, ensure that A-390 is cleaned and low-level disinfected before and after each use in accordance with chapter 19.
- Heed all warnings in these instructions.
- The A-390 does not work when it is being charged. It should be charged at least 1.8 m away from the patient.
- Use only the A-390´s battery charger supplied by the manufacturer.
- A-390 is portable RF communications equipment. It may cause radio interference or may disrupt the operation of nearby equipment. It may be necessary to take mitigation measures, such as, reorienting or relocating the A-390. ‘
- To reduce the risk of fire or shock hazard, do not expose this equipment to rain or moisture.
- Do not charge near any heat source such as radiators, heat registers, stoves or other apparatus that produces heat.
- Protect the charging cord from being damaged.
- Do not use this equipment in presence of flammable anaesthetics.
- ESD (Electrostatic discharge) may cause malfunction of these devices.
- Unplug this apparatus during lightning storms or when unused for long periods of time.
- Do not place vessels containing liquids on top of the apparatus. Do not splash liquids on top of the device.
- Do not attempt to service these devices yourself as opening or removing covers may expose you to hazards. Refer all servicing to qualified service personnel from your Airtraq distributor.
- A-390 shall never be serviced or maintained while in use with the patient.
- A-390 shall only be connected to Airtraq video laryngoscopes. Connection to other ME Equipment or ME System may cause unexpected harm.
- That before each use or after a change of viewing modes / settings, the operator should check to ensure the view observed through the device provides a live image (rather than a stored one) and has the correct image orientation.
USER ASSISTANCE INFORMATION
Instructions for use are available online at https://www.airtraq.com/IFU
Visit www.airtraq.com for further advice on
using the A-390
Any serious incident that occurs in relation to this device should be reported
to the manufacturer and the competent Health Authority in which the user
and/or patient is established. For communication with manufacturer, please
email user. assistance@airtraq.com
-
Contact directly with Manufacturer at its address or by phone or email:
USA & Canada: +1877-624-7929
EU & Other: +34944804690 -
Contact Manufacturer´s Representative for your area (details below) or
-
Contact your Local Distributor
Airtraq is a registered trademark.
Apple and the Apple logo, iPad and iPhone are trademarks of Apple Inc.,
registered in the U.S. and other countries. App Store SM is a service mark of
Apple Inc. Wi-Fi and the Wi-Fi logo are registered trademarks of the Wi-Fi
Alliance. All other trademarks and copyrights are the property of their
respective owners.
STORAGE AND TRANSPORT CONDITIONS: Temperature: between -5ºC/23ºF and 55ºC/131ºF; Humidity: between 10% and 95% ; Pressure: between 500 to 1060 hPa OPERATING CONDITIONS : Temperature: between -5ºC/23ºF and 40ºC/104ºF; Humidity: between 10% and 95% ; Pressure: between 500 to 1060 hPa
GLOSSARY OF SYMBOLS
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>