** UT3550 Battery Tester
User Manual**
Preface
Thank you for purchasing this product. In order to use this product safely and
correctly, please read this manual thoroughly, especially the safety notes.
Copyright Information
UNI-T products are protected by patent rights in China and other countries,
including issued and pending patents. Uni-Trend reserves the rights to any
product specification and pricing changes.
Uni-Trend Technology (China) Co., Ltd. all rights reserved. Trend reserves all
rights. Information in this manual supersedes all previously published
versions. No part of this manual may be copied, extracted, or translated by
any means without the prior permission of Uni-Trend.
UNI-T is the registered trademark of Uni-Trend Technology (China) Co., Ltd.
Warranty Service
The instrument has a warranty period of one year from the date of purchase. If
the instrument is damaged due to improper operation by the user during the
warranty period, the maintenance fee and the costs caused by the maintenance
shall be borne by the user, and the instrument shall be maintained by the
company for life.
If the original purchaser sells or transfers the product to a third party
within one year from the date of purchase of the product, the warranty period
of one year shall be from the date of the original purchase from UNI-T or an
authorized UNl-T distributor. Power cords, accessories, fuses, etc. are not
included in this warranty.
If the product is proved to be defective within the warranty period, UNI-T
reserves the right to either repair the defective product without charge for
parts and labor or exchange the defective product for a working equivalent
product (determined by UNI-T). Replacement parts, modules, and products may be
brand new or perform at the same specifications as brand-new products. All
original parts, modules, or products that were defective become the property
of UNI-T.
The”customer” refers to the individual or entity that is declared in the
guarantee. In order to obtain the warranty service, the “customer” must inform
the defects within the applicable warranty period to UNI-T, and perform
appropriate arrangements for the warranty service.
The customer shall be responsible for packing and shipping the defective
products to the individual or entity that is declared in the guarantee. In
order to obtain the warranty service, the customer must inform the defects
within the applicable warranty period to UNI-T, and perform appropriate
arrangements for the warranty service. The customer shall be responsible for
packing and shipping the defective products to the designated maintenance
center of UNI-T, pay the shipping cost, and provide a copy of the purchase
receipt of the original purchaser. If the product is shipped domestically to
the purchase receipt of the original purchaser. If the product is shipped to
the location of the UNI-T service center, UNI-T shall pay the return shipping
fee. If the product is sent to any other location, the customer shall be
responsible for all shipping, duties, taxes, and any other expenses.
Guarantee Limit
This warranty shall not apply to any defects, malfunction, or damages caused
by accidental, machine parts’ wear and tear, using outside the product’s
specifications, improper use, and improper or lacking maintenance. UNIT under
the provisions of this warranty has no obligation to provide the following
services:
a) Any repair damage caused by the installation, repair, or maintenance of the
product by non-UNI-T service representatives;
b) Any damage caused by improper use or connection to an incompatible device;
c) Any damage or malfunction caused by the use of a power source not provided
by UNI-T;
d) Any maintenance on altered or integrated products (if such alteration or
integration leads to an increase in time or difficulty of product
maintenance).
This warranty is written by UNI-T for this product and it is used to
substitute any other express or implied warranties. UNI-T and its distributors
do not offer any implied warranties for merchantability or applicability
purposes. For violation of this guarantee, UNI-T is responsible for the repair
or replacement of defective products as the only and complete remedy available
to customers. Regardless of whether UNI-T and its distributors are informed
that any indirect, special, incidental, or consequential damage may occur,
UNI-T and its distributors shall not be responsible for any of these damages.
Safety Instructions
Warning Danger:
To avoid possible electric shock and personal safety problems, please follow
the instructions below.
Disclaimer| Please read the following safety information carefully before
starting to use the instrument. Uni-Trend will not be responsible for the
personal safety and property damage caused by the user’s failure to comply
with the following terms.
---|---
Instrument Grounding| To prevent the risk of electric shock, please connect
the power ground wire.
DO NOT use the instrument in an explosive atmosphere| Do not use the
instrument in flammable and explosive gas, steam or dusty environment.
The use of any electronic equipment in such an environment is a risk to
personal safety.
DO NOT the outer shell of the instrument| Non-professional maintenance
personnel should not open the outer shell of the instrument to try to repair
the instrument. The undischarged charge still exists for a period of time
after the instrument is turned off, which may cause electric shock.
DO NOT use instruments that work
abnormally| If the instrument does not work properly and its danger is
unpredictable, please disconnect the power cord, do not use it, and do not try
to repair it yourself.
DO NOT use the instrument beyond
the way specified in this user manual| If it exceeds the range, the protective
measures provided by the instrument will become invalid.
| Warning:
For the instrument model UT3550, do not apply AC voltage and DC voltage
exceeding100V to the test terminal, otherwise, the instrument will be damaged.
Waste Electrical and Electronic Equipment (WEEE) Directive 2002/96/EC
| Must not be discarded in the trash can.
Introduction
This chapter includes:
Product Introduction
Main function
Front panel & Interface panel
1.1 Product Introduction
UT3550 battery internal tester, measured voltage range is 100V. See the below
table, a more specific technical index refers to the last chapter.
Model | Accuracy | Measurement Range |
---|---|---|
UT3550 | Resistance: ±0.5%rdg±5dgt | |
Voltage: ±0.05%rdg±5dgt | Resistance: 0.001mΩ~30.00Ω |
Voltage: 0.0001~100.00V
Product Characteristics UT3550 is an automated and real-time detection
handheld instrument made by a high-performance 32-bit ARM microprocessor. This
instrument can test resistance 0.001mΩ~30.00Ω and DC voltage 0.0001V~100.00V.
It has a built-in USB-Type-C interface, through data collection software to
remote control, collect and analyze data.
UT3550 uses a remote control instruction set and is compatible with SCPI
(Standard Command for Programmable Instrument) to fulfill remote control and
data collection.
The AC resistance testing principle can be used for the internal resistance
testing of almost all batteries, including lithium batteries, Hydrogen fuel
batteries, lead-acid batteries, button batteries, and other batteries. The
design allows direct online measurement of the UPS without loss of accuracy
and stability. IP65 is dustproof and waterproof, even in the dust pollution,
splashing industrial site, can also help you to accurately measure.
1.2 Main Function
1.2.1 Calibration Function
Full-range short circuit zeroed function.
1.2.2 Comparator Function(sorting)
Built-in sorting data, estimate measured object by OK/NG.
-
Compare Method:
ABS tolerance ± TOL sorting: compared with the upper/lower value of measured value and absolute deviation value
PER tolerance TOL sorting: compared with the upper/lower value of measured value and percentage deviation of nominal value
SEQ comparison sorting: measured value compared with the upper/lower value -
Beep
The beep can be set OFF / OK / NG beep by user-defined.
1.3 System
- Keypad lock function
- Switch Chinese/English language
- Date and time setup
- Administrator and user account, administrator account can set code
- Backlight setup
- Auto power off time setup
1.3.1 Remote Control
Support maximum 115200bps baud rate, compatible with SCPI protocol, ASCII
transmission.
1.4 Front Panel
1.4.1 Introduction of Front Panel
Front Panel | Number | Function |
---|
| 1| 3.5 inches
LCD
2| Function keys
3| Power switch
4| Test terminals
5| Function short cuts
6| Numeric keypad
Introduction of Symbol
Symbol | Description |
---|---|
**** | U disk inserts into the instrument, it can save data and screenshot |
**** | Key lock function enabled |
**** | There is no measured value, or measurement line and the object does not |
connect
| It presents measured value is qualified within comparator range
| It presents measured value is not qualified with comparator range
| After setting up the comparator, the measured result is failed
(unqualified)
| After setting up the comparator, the measured result is passed
(qualified)
1.5 Interface Panel
Top View | Number | Function |
---|
| 1| Type-C
interface
2| USB interface
3| Power Socket
4| 0.96 inches sub-display
5| Replaceable clip
Inspection and Installation
2.1 Packing List
Before use:
- Check whether product appearance is damaged, scratched, or has other defects;
- Check with the packing list to confirm that the accessories have no loss.
If there have any problems, please contact with Uni-Trend Instrument Sale Department or the distributor.
Components | Quantity | Remarks |
---|---|---|
Battery internal resistance tester | 1 | |
Power adapter | 1 | |
Soft carrying pouch | 1 | |
Shoulder belt | 1 | |
USB communication cable-Type-C | 1 | |
Magnetic strap | 1 | |
UT3550 kelvin clips wire | 1 | special for UT3550 battery tester |
UT3550 test pen-crown probe | optional | optional, special for UT3550 battery |
tester
User Manual| 1| electronic file, download from the official website
2.2 Requirements of Power Supply
It has a standard power adapter for the UT3550 battery internal resistance
tester.
When connecting with an external power supply, the instrument will use an
external power supply, and the cell stop supplying power automatically. The
external power supply also can charge up a built-in lithium battery. It is
suggested that use our company’s specialized power adapter.
Insert the power adapter into the AC input terminal on the instrument.
AC Power Adapter:
Input: 100V-240V, ~50Hz/60Hz, 0.6A Max
Output: 9V,2000mA
Warning: It can not use another standard power adapter. This instrument
can only match our company’s power adapter and recharge lithium battery!
If the battery capacity is not full, then inserting the power adapter into the
instrument enables to charge up the built-in lithium battery. When the
backlight of the power switch light on yellow, it presents in-charing. The
indicator light still lights on after powering off, it will light on the green
when fully charged.
2.3 Operating Environment
UT3550 series must be used under the following environmental conditions:
Temperature: 0℃~40℃
Humidity: 10~80%RH
Altitude: 0~2000 meter
2.4 Cleaning
To prevent the risk of electric shock, please pull out the power line before
cleaning.
Please use a clean cloth dipped in clean water to clean the cover and panel.
Do not clean the inside of the instrument.
Notes: Do not use solvents (alcohol or gasoline etc.) to clean the
instrument.
2.5 Replace Battery
The instrument has a built-in rechargeable lithium battery, and the battery is
installed in the cell barn.
If it needs to replace the battery, please do it as the figure:
- Take down screws
- Open cell cover
- Pull out the built-in lithium battery
2.6 Replace Strap
The strap can replace or take down, please the following steps,
- Take down screw
- Push the strap
- Push down and then pull out the strap
2.7 Tilt Stand
Preparation before Measurement
3.1 Power on and Run
3.1.1 Power on and off
Turn on the power switch , when the indicator light display green(no power
supply), it presents the power has been turned on.
3.1.2 Introduction of Test Lead
UT3550 series battery internal resistance tester is equipped with UT-L84
Kalvin clips and UTL86 test pen-crown probe, which is convenient for users to
measure more professionally.
3.1.3 Test Lead
From the perspective of safety and test accuracy, it is recommended to use the
standard test lead for measurement. The contact surface of the test lead is
gold plating, in the long-term work (1~2 years), it may cause plating
abrasion, it is suggested that the user replace the new test lead if the test
result appears deviation.
3.1.4 Connection of Test Lead
Before measurement, follow the steps below to connect the test leads to the
test terminals of the instrument:
- Before connecting the test line, please make sure that the power switch is OFF;
- Make sure there is no connection at the top of the test leads of the four terminals;
- Connect the test leads of the four terminals to the input terminals of the instrument according to the method shown in Figure 3-1-4, the steps are as follows,
Connect the black test lead to the black terminal hole of the instrument and
the red one to the red terminal hole.
The▲symbol of the black test lead must match the black SENSE hole on the front
panel, and the▲ symbol of the red test lead must match the red SENSE hole.
The test lead must connect with the terminal correctly, otherwise, it may
affect the output of the test result.
Figure 3-1-4
Schematic Diagram of the Connection between the Test Lead and the Terminal
Hole of the Instrument
Note: In order to ensure the accuracy of the instrument, please use the
attached test lead for testing.
Warning: It is prohibited to connect the AC current and voltage source
directly to the test terminals.
3.2 Measurement Method of DUT
After correctly connecting the test leads to the Sense and Source test
terminals of the instrument, please connect the test leads to the DUT as
follows. Note that the red test pen should be connected to the positive
electrode of the battery, and the black test pen should be connected to the
negative electrode. As shown in Figure 3-2-a.
If the positive and negative poles of the terminal of the DUT are connected
reversely, and the test voltage value is negative, please exchange the
positions of the red and black test pen.
- When testing with an alligator clip, the user can use the test lead to directly champ the terminal of the DUT for four-terminal testing.
- If testing with a crown probe test pen, the user can put the red and black test lead probes against the corresponding positive and negative terminals of the DUT for four-terminal testing, as shown in Figure 3-2-b.
Note: When a negative voltage appears, it is caused by the reverse
connection of the test leads, and the positions of the red and black test pens
need to be changed.
3.3 Set Zero and Calibration
Before testing, please perform the short circuit zeroing step to remove the
stray resistance and bias voltage caused by the test lead or external
environment factors.
The measured resistance value may be very small, such as 3mΩ and 30mΩ range.
When the test current flows through the resistance, the voltage signal
generated will be very weak, and the maximum is only a few mV, so the
position, length, and shape of the test lead may all have effects on the
measurement. Therefore, the positions and conditions during setting zero
should be the same as those during subsequent measurements.
3.3.1 Short-Circuit Test
- Method of Short Circuit Zeroed****
The first step| push【Utility】key to enter the setup page
---|---
The second step| the short contact test pen
The third step| use the function key to select【F4 SET ZERO】
The fourth step| push【F4 OK】, the instrument start to zeroing
The fifth step| When zeroing is finished, go back to the setup page
Correct Short Circuit Method of Crown Probe Test Pen
The center probe of the crown probe test pen is the Sense end, and the outer
cylindrical probe is the Source end.
First, place the probes of the crown probe test pen as shown in Figure
3-3-1-a. Make sure that the surface of the object at the red dot is in
contact, that is, the center probe is in contact with the center probe (Sense
and Sense). And the center probe is in contact with the peripheral probe
(Sense and Source). Second, apply force in the direction of the arrow shown in
Figure 3-3-1-b to make the surface of the object at the third red point
contact, that is, the periphery probe is in contact with the peripheral probe
(Source and Source). Finally, only when the three red dots as shown above are
actually touched, the next setting zero steps can be performed.
[Test] Measurement Page
4.1
- TRIGGER – The Trigger Mode of the Measurement
- FUNC – Parameters of Measurement
- RANGE – Setting and Display of Resistance and Voltage Range
4.1.1 【Trigger】
The instrument has 2 trigger modes: internal trigger and external trigger
(including manual /remote)
Function | Description |
---|---|
INT | Also called continuous test, the trigger signal is continuously tested by |
the internal of the instrument according to the inherent cycle. Generally,
choose this method for measurement
EXT| Manual: Each time press the【Trigger】key, the instrument will execute a
measurement cycle, and the instrument is in a waiting state at other times.
Remote: Send the command TRIGGER to measure once and return the measured
value.
Setting Steps
- Press【Test】key to enter the main page of
; - Use the cursor key▼to move the cursor to the 【TRIGGER】 field;
- According to the user’s needs, use the function keys at the bottom of the screen to select the desired trigger mode.
4.1.2 【Function】
The instrument has 3 measurement functions: RV, R, and V.
Function | Description |
---|---|
RV | Measure and display resistance and voltage values at the same time |
R | Only measure and display the resistance value |
V | Only measure and display the voltage value |
Setting Steps
- Press【Test】key to enter the main page of
; - Use the cursor key▼to move the cursor to the 【FUNC】field;
- According to the user’s needs, use the function keys at the bottom of the screen to select the desired measurement items.
4.1.3 【R-Range】
UT3550 series battery internal resistance tester has 5 ranges, divided into
manual and automatic mode, as shown in the following Table 4-1-2,
Table 4-1-2 Range Selection Method and Variation Rane
Range Mode | Description |
---|---|
AUTO | The instrument automatically selects the corresponding best test range. |
MANUAL | The instrument will always use the user-specified range for testing. |
Manual Range | Automatic Range |
<4>30Ω | <4> AUTO |
<3>3Ω | <3> AUTO |
<2>300mΩ | <2> AUTO |
<1>30mW | <1> AUTO |
<0>3mΩ | <0> AUTO |
Setting Steps
- Press【Test】key to enter the
page or press【Utility】key to enter the page; - Use the cursor key▼to move the cursor to the the【R-RANGE】field;
- According to the user’s needs, use the function keys at the bottom of the screen to select the desired range method and range. If the user selects AUTO, the instrument will automatically select the corresponding best test range and range; if the user selects MANUAL, the instrument will always use the user-specified range for testing; INC+ means to increase the range, while the range is changed to lock; DEC- means decrease the range, and the range is changed to lock at the same time.
When the range is automatic, the instrument will predict the range in each measurement period, so the test speed will be slightly slower than the locked range. Moreover, in the case of automatic measurement, frequently changing the range will slow down the response. Usually, when the instrument is used as a sorting measurement, the automatic range mode is not suitable. For sorting measurement, it is recommended to select manual range mode.
4.1.4 【V-Range】
Voltage range setting: UT3550 series battery internal resistance tester has 3
ranges and two ranges selection modes: automatic and manual
Table 4-1-4 Range Selection Mode and Variation Range
Range Mode | Description |
---|---|
AUTO | The instrument automatically selects the corresponding best test range. |
MANUAL | The instrument will always use the user-specified range for testing. |
Manual Range | Automatic Range |
<1>100V | <1> AUTO |
<0>10V | <0> AUTO |
Setting Steps
- Press【Test】key to enter the
page or press【Utility】key to enter the page; - Use the cursor key▼to move the cursor to the the【V-RANGE】field;
- According to the user’s needs, use the function keys at the bottom of the screen to select AUTO, MANUAL, INC+, or DEC-.
When the measured value exceeds the setting range, OL is displayed on the test
interface. When the range is automatic, the instrument will make range
predictions at each measurement period, so the test speed may be lower than
the locking range; In addition, in the automatic measurement, changing the
range frequently will lead to a slow response.
When the instrument at sorting measurement, automatic mode is not suitable.
For sorting, it is suggested to select manual mode.
If the measured value exceeds the setting range, the test interface display
OL; if the test circuit is open fully, the test interface display ——-
4.2 Save and View Data
The measured display value can be manually saved (up to 500 sets of data can
be stored) and quickly browsed on the instrument screen. Press the function
key [SAVE DATA] at the bottom of the screen to manually save the data. Each
time user presses the key, the data will be saved manually. Via the function
key [VIEW DATA], the user can view the manually saved measurement data on the
instrument screen, as shown in Figure 4-2 On the
- Save to USB disk: after inserting the U disk, this function key can save the data in the U disk.
- Clear: this function key can clear all data.
- Page Up/ Page down/ Jump to: this function key can realize the page up and down or jump of multiple pages of data.
If the user wants to delete a row of data, you can press the arrow keys to
stop the cursor on a specific row, and select the [DELETE] function key at the
bottom of the screen to delete this row of data.
4.3 Screenshot Function
The instrument provides a screenshot function, inserts a USB storage device
into the USB interface on the front panel, and presses the [ENTER] key to save
the current screen capture to the USB disk for subsequent reference.
If the user needs to save data, it is recommended to use a branded U disk to
insert the instrument interface. The format and capacity can refer to FAT,
FAT32, or EXFAT.
When the data is being collected, it may cause the screen capture to fail,
which can be executed after the collection is completed. If the user needs to
save data, it is recommended to use a branded U disk to insert the instrument
interface. The format and capacity can refer to FAT, FAT32, or EXFAT.
When the data is being collected, it may cause the screen capture to fail,
which can be executed after the collection is completed.
[Utility] Setting
5.1 Parameters of Measurement
All the settings related to the measurement are operated on the
The rest of the settings include the following parameters,
- Timer – delayed before measurement of external trigger
- Beep – beep alarm
5.1.1 Timer
In order to conveniently collect data into internal storage, users can use the
Timer collect function. When the test is ready, the instrument will start the
timer, when it counts to 0, it will keep the current data in internal storage.
Users can view the collected data on the
Setting Steps of Timer
- press【Utility】key to enter the setup page;
- Use the cursor to select【Timer】filed;
- Use the numeric keypad to enter time, the unit is second.
5.1.2 Beep
The beep function is effective only when the comparator function is turned on.
The beep function has three settings: OK (qualified beep), NG (unqualified
beep), and OFF (turn off beep).
OK: the buzzer will sound when the sorting result is qualified
NG: the buzzer will sound when the sorting result is unqualified
Setting steps
- Press【Test】or【Utility】key, and select【COMP SET】with the function key at the bottom of the screen to enter the 【COMP SET】page;
- Use the cursor key to move the cursor to the【Beep】filed;
- According to the user’s needs, use the function keys at the bottom of the screen to select OK, NG and OFF.
Comparator Sorting
6.1 Comparator Setting
6.1.1 Comparison Mode Selection
The instrument can compare resistance and voltage simultaneously or
separately. There are 3 comparison modes to choose from: a) Absolute Value
Comparison ABS
(a): Absolute Value Δ= Measured Value – Nominal Value
b) Percentage Comparison PER(Δ%): Percentage Δ% = (Measured Value – Nominal
Value) / Nominal Value × 100%
c) Direct-Reading Comparison(SEQ): The direct-reading value SEQ comparison
uses the direct reading measurement value to compare with the upper and lower
limit ranges, so the nominal value is not required to participate in the
calculation.
6.1.2 【nominal value】set up
The absolute value and percentage comparison mode must enter the nominal
value. As for the direct reading comparison mode, the nominal value does not
participate in the calculation.
Setting Steps for Absolute Value Comparison ABS(Δ)or Percentage Comparison
PER(Δ%)
- Press【Test】or【Utility】key, and select【COMP SET】with the function key at the bottom of the screen to enter the 【COMP SET】page;
- Use the cursor to the【R-COMP】or【V-COMP】field to select ON;
- Move cursor move to【Mode】field, and select comparison mode absolute value or percentage;
- After operating the third step, move the cursor to the corresponding【nominal value】,【Upper limit】 or lower limit】field, and use the numeric keyboard to enter data, unit set by functional key.
6.1.3 【Upper limit】and【Lower limit】set up
Setting Steps for Direct-Reading Comparison SEQ
- Press【Test】or【Utility】key, and select【COMP SET】with the function key at the bottom of the screen to enter the 【COMP SET】page;
- Use the cursor to the【R-COMP】or【V-COMP】field to select ON;
- Move cursor move to【Mode】field, and select absolute value;
- After operating the third step, move the cursor to the corresponding【Upper limit】or【Lower limit】field, and use the numeric keyboard to enter data, unit set by functional key.
Notice: Percentage mode Δ% does not need to select unit ratio, enter
percentage value; absolute value Δ and direct reading SEQ mode need to use the
function key to select unit.
The instrument shares the same storage space for the three comparison modes,
so the comparator data needs to be reset after switching the comparison mode.
6.2 Display and Determination
6.2.1 Display after Opening the Comparator
When the comparator setting is completed, press【Test】to enter the main page of
- After the comparator is turned on, the “COMP” logo will appear at the top of the screen, which means that the comparator function has been turned on at this time. The determinating symbol is displayed on the screen, which is a normal interface.
- “ ” means that the test clip or test pen is not properly connected to the object under test. Please connect the test clip or test pen to the object under test correctly to enter the state to be measured.
6.2.2 Example of Qualified Determination
- When the voltage or resistance measurement value is within the setting range of the comparator, the judgment symbol of the test value will be displayed on the screen, indicating that the measured value is within the specification range. Only when the resistance and voltage are within the range, the instrument will judge the total result of this group of test data as qualified, and a qualified mark will appear on the screen.
- If the buzzer is set to “beep when qualified” at this time, the instrument will emit a qualified buzzer for the qualified situation.
- If the buzzer setting is OFF at this time, there will be no buzzer sound.
6.2.3 Example of Unqualified Determination
- When the voltage or resistance measurement value exceeds or falls below the comparator setting value, in any of these situations, the instrument determines that the test result is unqualified, the symbol is .
- If the buzzer is set to “beep when unqualified” at this time, the instrument will emit an unqualified buzzer for this situation.
- If the buzzer setting is OFF at this time, there will be no buzzer sound.
System Configuration
7.1 System Configuration Settings
- Press【Test】or【Utility】key, and select【SYSTEM】with the function key at the bottom of the screen to enter the < SYSTEM CONFIG> page;
- Use the cursor key to move the cursor to【LANGUAGE】;
- According to the user’s needs, press the function key at the bottom of the screen to select Chinese [CHN] or ENGLISH (English).
7.1.2 DATE and TIME Setting
The instrument uses a 24-hour clock, and it can modify the date and time.
Setting Steps
- Press【Test】or【Utility】key, and select【SYSTEM】with the function key at the bottom of the screen to enter the
page; - Use the cursor key to move the cursor to the [DATE] or [TIME] field;
- According to the user’s needs, press the function key at the bottom of the screen to select the year, month, day, hour, minute, and second that you want to set.
+ means the value plus 1 and – means the value minus 1. For example, year+, month+, day+, hour+, minute+, and second+ represent +1 year, +1 month, +1 hour, +1 minute, and +1 second respectively and vice versa.
7.1.3 AUX SCREEN Setting
Sub-screen has three settings: OFF/ UP/ DOWN
- Press【Test】or【Utility】key,and select【SYSTEM】to enter < SYSTEM CONFIG > page;
- Use the cursor key to move the cursor to【AUX SCREEN】field;
- According to the user’s needs, press the function key at the bottom of the screen to select OFF, UP, and DOWN.
7.1.4 KEY BEEP Setting
The instrument key can be set with or without the key prompt sound.
Setting Steps
- Press【Test】or【Utility】key, and select【SYSTEM】to enter < SYSTEM CONFIG > page;
- Use the cursor key to move the cursor to【KEY BEEP】field;
- According to the user’s needs, press the function key at the bottom of the screen to select ON or OFF. ON means that the key beep is on; OFF means that the key beep is off.
7.1.5 BAUD Setting
The instrument has a built-in Type-C interface, after the instrument senses
the signal conversion the Type-C interface immediately communicates with the
host at the set baud rate, and the keyboard is locked.
In order to communicate correctly, it is necessary to confirm whether the baud
is set correctly, otherwise, the upper computer cannot communicate correctly.
The instrument with Type-C use SCPI language for programming.
Setting Steps
- Press【Test】or【Utility】key, and select【SYSTEM】to enter < SYSTEM CONFIG > page;
- Use the cursor key to move the cursor to【BAUD】field;
- According to the user’s needs, press the function key at the bottom of the screen to select different baud rates,9600, 19200, 38400, 57600, 115200.
- To communicate with the host computer, it is recommended to use 115200 high-speed baud rates.
7.1.6 BRIGHTNESS Setting
The more the brightness is darker, the more power consumption is lesser, and
the instrument operating time will be longer. The instrument can set 5
brightness levels to meet different light needs.
Setting Steps
- Press【Test】or【Utility】key to enter the main page;
- Select【SYSTEM】key to enter
; - Use the cursor key to move the cursor to【BRIGHTNESS】field;
- Use the function key to select the brightness levels.
7.1.7 APO Setting
When the test measurement values exceed the range, no operating, serial port
pause to respond, and reaches auto power off time, the instrument will save
the data and turn off automatically.
Setting Steps
- Press【Test】or【Utility】key to enter the main page;
- Select【SYSTEM】key to enter
; - Use the cursor key to move the cursor to
filed; - Use the function key to select auto power off time.
Function key | Description |
---|---|
5 minute | |
15 minute | The default value to save power |
30 minute | |
60 minute | |
OFF |
7.2 System Information
Press the [ Test] or [ Utility ] key, and select [ SYSTEM ] with the function
key at the bottom of the screen to enter the
This page does not require the user to set up.
7.3 System Service
Press the [ Test] or [ Utility ] key, and select [ SYSTEM ] with the function
key at the bottom of the screen to enter the
Warning: This page is not open to users. It is used to calibrate data when
leaving the factory. Non-professionals cannot enter by force. Otherwise, the
calibration data may be lost, resulting in a large deviation of the
measurement data.
Technical Index
8.1 Technical Index of Product
Function| Range No.| Range| Resolution| Accuracy|
Measuring current| Description
---|---|---|---|---|---|---
Resistance Measurement| 0| 3mΩ| 1μΩ| ±0.5%rdg.
±5dgt| 150mA| Measuring current frequency: 1kHz±30Hz
1| 30mΩ| 10μΩ| 150mA
2| 300mΩ| 100μΩ| 15mA
3| 3Ω| 1mΩ| 1.5mA
4| 30Ω| 10mΩ| 150uA
Voltage Measurement| Range name| Range| Resolution|
Accuracy| Description
0| 10V| 100μV| ±0.05%rdg.
±5dgt| /
1| 100V| 1mV
Display| 3.5 inch LCD display +0.96 inch OLED on the top
Range Mode| Auto and Manual(resistance and voltage)
Calibration| Short circuit full range to zeroed
Testing Speed| 1/s(manual range)
Sorting Function| It has three output situations: exceed resistance and
voltage upper limit/ exceed resistance and voltage low limit and conform The
sorting result has two output situations: PASS and FAIL
Beep| OFF, OK, NG
Comparison Mode| Absolute deviation(ABS), relative deviation(PER)
Direct-reading(SEQ)
Other Function| Lock keypad, U disk storage, keypad tone, Chines/English
language, screenshot
Test Terminal| Four-terminal test method
Trigger| Internal, manual, and remote
Storge| The maximum storage is 500 groups, supports USB to copy data
Interface| Type-C interface(communication), USB interface(storage), Charging
interface
Upper-computer Software| SCI
Charging Time| <5 hour
Operating Time| About 5 hours (20% brightness display)
Power| The maximum 5VA
Power Supply| 7.4V,2000mAh rechargeable lithium battery or 9VDC 2A power
adapter
Preheating Time| At least 15 minute
Temperature/Humidity Index| Temperature:18℃~28℃, Humidity:<65%RH
Operating Temperature/Humidity| Temperature:0℃~40℃, Humidity:10~80%RH
Storage Temperature/Humidity| Temperature:-20℃~60℃, Humidity:10~90%RH
Temperature Index| Over 28℃, less 18℃, each degree plus 0.1* the specified
precision
---|---
Accuracy guarantee time| 12 months
Operating Altitude| ≤2000m
UNI-TREND TECHNOLOGY (CHINA) CO., LTD.
No.6, Gong Ye Bei 1st Road,
Songshan Lake National High-Tech Industrial
Development Zone, Dongguan City,
Guangdong Province, China