Suzhou Nargon Electronic Equipment Co LTD NMC161 Moisture-Proof Cabinet Instruction Manual

October 30, 2023
Suzhou Nargon Electronic Equipment Co LTD

Suzhou Nargon Electronic Equipment Co LTD NMC161 Moisture-Proof Cabinet

Operation instruction for moisture-proof cabinet

Welcome to choose NMC series electronic moisture-proof cabinet produced by SuzhouNargon Electronic Equipment Co., LTD. Please read this manual carefully before usingsoastogive the best performance and avoid unnecessary loss.

  • First, we would like to thank you for purchasing Nargon NMC series electronic moisture-proofcabinet. We are convinced that this product will improve your working environment andplayagood role in improving product quality. This manual describes all the characteristics of NMC series electronic moisture-proof cabinets. For special non-standard products, please read and use the manual together with the additional
    instructions. For non-standard products with only ruler marking or material change, noadditional instructions will be given. In order to ensure the most effective operation of the product, please read this manual carefullyand get familiar with the product. And keep the manual in a convenient place for referenceat any time in the use of the product. In order to ensure the safe and reliable operation of the product, please designate a special person as the operation manager of the product. In case of failure of the product, please immediately report to the manager or contact the company. Contact number: 0512-52179330

specification table of Nargon NMC series electronic moisture-proof cabinet

nominal volume

**** L

| Medium humidity model 20%-60%RH

**** adjustable

| Ultra-low humidity model 1%-40%RH

f ully automatic

| External dimensions

**WDH(mm)**

| **** Internal dimensions

**WDH (mm)**

| Layers
---|---|---|---|---|---
160L| NMC161| NMC161B| W450D450H1010| W405D420H840| 3
240L| NMC242| NMC242B| W600D450H1250| W555D420H1080| 3
320L| NMC322| NMC322B| W900D450H1010| W855D420H840| 3
450L| NMC452| NMC452B| W900D600H1010| W855D570H840| 3
540L| NMC542| NMC542B| W600D690H1485| W555D640H1315| 3
730L| NMC732| NMC733B| W600D690H1830| W555D640H1660| 5
730| NMC733| NMC733B| W600D690H1830| W555D640H1660| 5
880L| NMC884| NMC884B| W900D600H1860| W855D550H1680| 5
1430L| NMC1434| NMC1434B| W1200D690H1830| W1155D640H1660| 5
1430L| NMC1436| NMC1436B| W1200D690H1830| W1155D640H1660| 5

Functions and features

Humidity control:
Touch CNC humidity control; humidity range is adjustable. Rated voltage (Rated voltage) : AC220V / 60 hz, extremely low power consumption The dehumidifier core whose performance is highly reliable and durable adopts the dehumidifier technology with Chinese patent, making it work without noise or water. Keep at room temperature and humidity, no heating effect; which will not affect the quality of store items. The shelf position can be adjusted to facilitate the storage of articles of different heights. Use special hygroscopic material which needs no change. No vibration, no water; clean storage For detailed information, please refer to the “Nargon NMC series electronic moisture-proof cabinet user manual”.

General product installation steps (when the moisture-proof box is used for the first time or hasbeen out of use for a period of time and needs to be reused), please follow the followingstepstoinstall:
Note: Turn on the power after confirming that the four legs of the goods are adjusted andstabilized.And adjust humidity when the power is on. After 24 hours, check and confirmthehumidity reaches the set humidity,then put the product in.

Preparation before starting:
Cabinet placement:
1. Moisture-proof cabinet must be placed in the horizontal solid surface;Turn the balancingfeetto solid
2. Reserve a certain space on the back to facilitate moisture removal (no less than 20cmawayfrom the movement)
3. Please do not put the moisture-proof cabinet in the following places:

  1. air flow direction, such as air outlet of air conditioner. 2) high temperature
    3). Direct sunlight
  2. smoke, water vapor places
    4. Adjust the height of each layer of the adjustable laminates in the cabinet accordingtotheheight of

Preparation before starting:

Cabinet placement:

  1. Moisture-proof cabinet must be placed in the horizontal solid surface;Turn the balancingfeetto solid

  2. Reserve a certain space on the back to facilitate moisture removal (no less than 20cmawayfrom the movement)

  3. Please do not put the moisture-proof cabinet in the following places:
    1) air flow direction, such as air outlet of air conditioner.
    2) high temperature
    3). Direct sunlight
    4) smoke, water vapor places

  4. Adjust the height of each layer of the adjustable laminates in the cabinet accordingtotheheight of the placeme

How to use

  1. It is recommended to use wall fixed independent socket, 220V/ 10a socket
  2. Press the “ON/OFF” button ON the far right of the numerical control panel. Humiditysettingcan be started when the display screen lights up .
  3. Press the “+” and “-” button on the numerical control panel and adjust it to the humidityyouneed. After 5 seconds, the computer will record it.Set humidity values
  4. At this time, the red light of the dehumidifier movement is on. And lighting up of green”RUN” light on the left side of the display indicates that the valve has been opened outwardlynormally. After a few minutes, the green light of the dehumidifier movement is on. Thenthered”OUT” light is on. Movement heating indicates that the movement starts to drainmoistureoutwards. 5. After the dehumidifying of the dehumidifier movement is finished, dehumidifyingshall bestarted. When lights “RUN” “OUT”are off, it begins to turn into the moisture insidetheabsorption box, and at the same time, the red light and green light on the dehumidifyingmovement are off. The dehumidifying work starts. 6. Leave the moisture-proof box empty and the warm-up machine running for one night or morethan 10 hours (do not place any storage items)
  5. Confirm the relative humidity before opening: it is normal that it is belowthe set humidity;
  6. When the articles are put in, the relative humidity will rise; and the host machinewilldehumidify slowly and stably. For items with high moisture content, it will take a longtimeforthem to fall back. This is normal. Please feel free to use. 9. The goods stored in the cabinet should not be too close to the housing of the movement, soasnot to cover the moisture absorption port of the dehumidifier.

Matters needing attention

This is not a failure in this case

  1. The temperature and humidity control panel on the cabinet has been calibrated andcalibratedby the temperature and humidity instrument calibrated by the national standard measurementdepartment for 2 hours when the product leaves the factory. To avoid misleading,it is necessaryto seek the local national standard measurement department for calibration.
  2. In the comparison process, the humidity value of the two instruments fluctuates normallyon5%RH, and the humidity error value is 5%RH.
  3. It is a normal phenomenon that the humidity in the cabinet increases when the movement valve plate is opened externally, because the humidity in the cabinet is far lower thantheambient humidity before the valve plate is opened. The valve blade switch is not completedinan instant. It takes about 3 minutes. Therefore, the environmental humidity has some influenceon the cabinet during this process.
  4. When dehumidifying, that is, when the valve plate is opened externally, the surfacetemperature of the dehumidifying movement is higher than the normal temperature, whichisnormal. Please do not mistake it for a failure of the dehumidifying movement.
  5. Please try to shorten the opening time when you open the door to pick up and release. 6
  6. Do not cover the back dehumidifier core cooling hole
  7. When the object is put into the moisture-proof cabinet, the humidity in the moisture-proofbox will rise due to the moisture contained in the object itself. And it may take 1 to 2daysforthe humidity to fall. For the article containing wrapping paper or cloth, the humidity will fall fora longer time.

Failure and elimination

Please confirm before repair services

Abnormal conditions| Items to be inspected| Reasons| Processing methods
---|---|---|---
Humidity can’t go down|

  1. power on or not
  2. Whether the power pl ug is firmly inserted in to the power socket
  3. For those equipped wi th sub-units, please c heck whether the po wer cables of the main and sub-units are con nected

|

  1. When power is off,
    h umidity will not rise; But if the door is often   op en or for days without po wer, moisture will gradu ally enter the box

|

  1. please confirm the power
  2. Please secure the po wer socket
  3. The humidity canno t drop below the set v alue because the auxiliary   machine i s not working

| 4. Switching time of   gla ss door is too long

5. Glass doors’ opening and closing are too often

| 2. causing a large number of moisture into the box and the relative humidity rise.| 4. Reduce opening ti me and frequency.
| 6. Newly added items wi th high moisture content: such as: paper products, wood products, leather pr oducts, etc| 3. As water  molecules ar e constantly released due t o the high moisture cont ent of the articles, the rate of humidity decline will b e affected.| 5. It takes a long time to observe, sometim es as long as several weeks, to reach the l ow wet range.
The indicator light “RU N” and “OUT” on the te mperature and humidity display screen is not on| Whether the relative hu midity is in stable range or in continuous declin e.| The dehumidifying heart i s absorbing the humidity i n the chamber.| This is a normal phe nomenon. When desi ccant heart hygrosco pic saturation, it w ill automatically light up to discharge the h umidity of the cabin et.
The indicator light in th e dehumidifier is not on| Whether the relative hu midity is in stable range or in continuous declin e.| The dehumidifying heart is absorbing the humidity in the chamber.| This is a normal phe nomenon. When desi ccant heart hygroscopic saturation, it w ill automatically light up to discharge the humidity of the cabin et.

applications and storing items for reference

Relative humidity Suitable storage of items
45%RH-60%RH Books, antiques, paper money, old books, fax paper, copy paper
45%RH-55%RH Camera, video camera, camera, microscope, magnifying glass,

binoculars, magnetic tape, magnetic disk, tape, film (light), the positive and negative film, Musical Instruments, stamps, fur,

Chinese traditional medicine, tea, coffee, cigarettes, etc

35%RH-45%RH| Precised hardware molds, measuring instruments, electronic parts, motherboard, metal powder, semiconductor, pharmaceutical raw

materials,etc

35%RH 或以下| Samples, standard gauges, seeds, pollen, bulbous roots, etc.
10% RH -40%RH| Patch elements,PCB boards
10%RH 以下| IC. Silicon wafer, wafer.TFT-LCD, LTPS, etc