Shanghai Yinxi Instrument Co Ltd 250SD Precision Incubators of Constant Temperature and Humidity Instruction Manual

October 30, 2023
Shanghai Yinxi Instrument Co Ltd

Shanghai Yinxi Instrument Co Ltd logo

250SD Precision Incubators of Constant Temperature and Humidity

Shanghai Yinxi Instrument Co Ltd 250SD Precision Incubators of Constant
Temperature and Humidity product

Instruction Manual for Precision Incubators of Constant Temperature and

Humidity

Product Information

The Precision Incubator of Constant Temperature and Humidity is a product manufactured by Shanghai Yinxi Instrument Co., Ltd, a professional manufacturer of temperature and humidity controllers.
The incubator has the following features:

  • Large color display screen: The liquid crystal screen displays humidity and temperature. The dynamic icons visually indicate the running condition of the incubator.
  • High precision: The temperature is accurately controlled by microcomputer, which ensures a highly precise constant temperature and humidity. The touch-button facilitates the temperature regulation.
  • Safe and reliable: Over-voltage protection, under-voltage protection, and double temperature limits (Alarms are given in case of a superhigh temperature (over 69.9) or an ultralow temperature (under 0). The upper-limit protection can be provided when the temperature reaches 70.)
  • Environment-friendly and energy-saving: The insulating layer is integrally injected with hard PU, thus ensuring long durability and high effectiveness in energy saving.
  • Self-diagnosis: The fault can be diagnosed automatically and displayed visually. The supply voltage can be detected automatically, and alarms can be given.
  • Automatic printing and recording: The data printer or curve recorder prints or records the historical data or curve of temperature automatically for easy reference and filing. (The configuration is optional. Please mark it when placing an order.)
  • RS485 computer communication interfaces: (The configuration is optional. Please mark it when placing an order.)

The Precision Incubator of Constant Temperature and Humidity has a variety of specifications and can be used under the following conditions:

  • Ambient temperature: 5-30℃
  • Ambient humidity: 80%RH
  • Range of atmospheric pressure: 86kPa-106kPa

Product Usage Instructions

Please follow the below instructions to use the Precision Incubator of Constant Temperature and Humidity:

  1. Read the Instruction Manual carefully before using the incubator.
  2. Ensure that the incubator is used under the specified environmental conditions.
  3. Press the touch-button to regulate the temperature as required.
  4. Observe the running condition of the incubator through the dynamic icons displayed on the liquid crystal screen.
  5. In case of a superhigh temperature (over 69.9) or an ultralow temperature (under 0), alarms will be given. The upper-limit protection can be provided when the temperature reaches 70.
  6. In case of a fault, the fault will be diagnosed automatically and displayed visually. The supply voltage can be detected automatically, and alarms can be given.
  7. If automatic printing and recording is required, please mark it when placing an order. The data printer or curve recorder prints or records the historical data or curve of temperature automatically for easy reference and filing.
  8. If RS485 computer communication interfaces are required, please mark it when placing an order.

Ensure that the incubator is properly maintained and kept clean for optimal performance.

Summary

Product Features

  • Large color display screen: The liquid crystal screen displays humidity and temperature. The dynamic icons visually indicate the running condition of the incubator.
  • High precision: The temperature is accurately controlled by microcomputer, which ensures a highly precise constant temperature and humidity. The touch-button facilitates the temperature regulation.
  • Safe and reliable: Over-voltage protection, under-voltage protection, and double temperature limits (Alarms are given in case of a superhigh temperature (over 69.9) or an ultralow temperature (under 0). The upper-limit protection can be provided when the temperature reaches 70.)
  • Environment-friendly and energy-saving: The insulating layer is integrally injected with hard PU, thus ensuring a long durability and a high effectiveness in energy saving.
  • Self-diagnosis: The fault can be diagnosed automatically, and displayed visually. The supply voltage can be detected automatically, and alarms can be given.
  • Automatic printing and recording: The data printer or curve recorder prints or records the historical data or curve of temperature automatically for easy reference and filing. (The configuration is optional. Please mark it when placing an order. )
  • RS485 computer communication interfaces: (The configuration is optional. Please mark it when placing an order. )

Application Scope and Main Purposes

The Precision Incubator of Constant Temperature and Humidity (referred to as the incubator, the same below)is a kind of high-precision device for laboratory providing constant temperature and humidity, thus making itself ideal for colleges and universities, scientific research institutions, medical field, animal husbandry, chemical, food, agriculture, aquaculture, biogenetic engineering and so on. It can find wide application in testing effectiveness and the aging of the medical field, chemical material, wood and building materials, as well as in the research on the culture of tissues, cells, mould and microorganism and on other tests in the condition of constant temperature and humidity.

Type and Specification

The incubator has a variety of specifications.

Composition and Meaning of Model

Shanghai Yinxi Instrument Co Ltd 250SD Precision Incubators of Constant
Temperature and Humidity fig 1

Service Environment

The incubator should be used under the following conditions:

  1. Ambient temperature: 530
  2. Ambient humidity: 80%RH
  3. Range of atmospheric pressure: 86kPa106kPa.

Operating Conditions

It should be operated on an AC (220 ± 22) V (50 ± 1) Hz power source. Operating mode: Continuous operation.

Structural Features

  • Body: The exterior housings are made of superior cold-roll steel sheetsand the inner housing is made of brushed stainless steel.The insulating layer is made of hard PU foam.
  • Door: The door is mounted with hollow glass in its middle so as to facilitate observing the experiment effect. The insulating layer is made of hard PU foam. The door gasket is sealed together with the housing with magnetic adhesive.
  • Refrigerating system: It is made up by the compressor, condenser, capillary tube, evaporator and so on.
  • Control system: It is controlled by microcomputer.
  • Supporting net: The round steel is made into the strip shelf which is then sprayed with plastics. The height is adjustable.

Technical Features

The incubator is a kind of thermostatic and constant-humidity device that can achieve humidifying, heating and refrigerating. It is designed for the required purposes (No.2 in the summary).
The main technical parameters are shown as follows( chart 1 and 2):

Table 1: Dimension of incubator Body

Product Model Boundary Dimension Dimension of Inner Body
SHH-250SD 570mm×570mm×1600mm 500mm×480mm×1050mm

Table 2: Major Performance Parameters

Model Item SHH-250SD
Nominal volume 250L
Temperature control range 5℃~65℃(Note1)

Allowance of temperature fluctuation

| ****

±0.5℃



Allowance of temperature uniformity

| ****

1.2℃

Humidity control range

(Note 2)

| 35%RH ~ 95%RH
Humidity      fluctuation| ±5%RH
Lighting power| 6VA


Compressor power

| ****

145VA

Heating power| 300VA
Humidifying power| 340VA
Refrigerant| R134a

Note

  • The above parameters are measured in the 30±2 thermostatic incubator.
  • In case a low temperature is used, the following principle shall be followed: operating temperature ambient temperature-20.
  • Within the temperature and humidity ranges as follows, it is fully controlled, (shown in Figure)Shanghai Yinxi Instrument Co Ltd 250SD Precision Incubators of Constant Temperature and Humidity fig 2

The humidity controlling area (Figure 1)

Directions for Use and Precautions

The indication panel of the incubator

Shanghai Yinxi Instrument Co Ltd 250SD Precision Incubators of Constant
Temperature and Humidity fig 3

Key and identifier description

The definition of Identifier

  • T1Reservation: when entering the reservation timing state, T1 will flash and (2) the display window will display the appointment countdown
  • T2 Heating: when heating has output, T2 lights up
  • T3Compressor: when the compressor starts, T3 lights up; when the compressor is waiting for start-up , T3 flashes
  • T4 Humidification: when humidifying has output, T4 lights up
  • T5 Daytime: when entering day mode, T5 lights up
  • T7 Timing: when entering the operation timing state, T7 will flash, and (2) display window will display the timed countdown
  • T8 Water shortage: when there is a signal of water shortage, T8 lights up; when there is an alarm of water shortage, T8 flashes
  • T9 Open door: when the door is opened, T9 lights up
  • T10 Alarm: when there are alarms of temperature and humidity , T10 lights up; when entering the low temperature or high temperature protection state, T10 flashes
  • T11 Lock the screen: when the screen is locked, T11 lights up
  • T12 Adding water: when the pump has output, T12 lights up
  • T13 Defrosting: when defrosting has output, T13 lights up
  • T14 Solenoid valve: when the solenoid valve has output, T14 lights up

Display window

  1. Display window shows values of period or segment
  2. Display window shows: The values of timing or time setting
  3. Display window shows the measured temperature
  4. Display window shows the setting temperature
  5. Display window shows the measured humidity
  6. Display window shows the setting temperature
  7. Display window shows the setting illumination or the output power of heating .

Definition of Keys

Shanghai Yinxi Instrument Co Ltd 250SD Precision Incubators of Constant
Temperature and Humidity fig 4

Instructions of operation

  1. When the controller is powered on, 3Display window shows “TH” 5 Display window shows”V01″, the buzzer gives a short beep. After about 2 seconds, it will enter the normal display state.

  2. The modification of setting values
    In normal display state, click setting key the two identifiers, “TIME” and “SV” flash at the same time. At this time, the value of the number flashed by the cursor can be modified by shift, increase, decrease keys, and click setting key again to switch to the next group of values. After modification, long press the setting key for 1 second to exit the setting state, or in the fixed mode, continuously click the setting key and modify a group of data and exit. The buzzer will give a short beep to enter the normal display interface, and the parameters will be saved automatically.
    When the controller is running in the program mode, click the setting key, and the value of segments starts to flash. At this time, you can view the setting values of time, temperature, humidity and illuminance under the corresponding number of segments by modifying the value of segments with the increase, decrease keys. And then click the setting key again to modify the setting parameters one by one in the present segment. When the cursor returns to the value of segments and flashes, modify the value of segments again to modify the next segment. When the controller is running in the DAY /NIGHT mode, click the setting key, and the “DAY” identifier will flash. Through the increase, decrease keys, it can be switched, and the “NIGHT” identifier flashes, to view the setting parameters under the day and night. Click the setting key to modify the setting parameters in the day or night mode.

  3. To set the number of segments or cycle
    In the non-fixed value mode (see parameter table-1-u1 for details), when the controller stops running, long press the setting key for 3 seconds, the (1) display window displays the “LC” prompt, and the (2) display window displays the password. Modify the password to 3 through the increase, decrease keys to enter the cycle and segment number settings. In the program mode, if the “PRO” prompt flashes, you can set the total operation sections; when you click the setting key, and the “CYC” prompt flashes, you can set the total operation cycles (when the number of cycles is set to 0, the controller always runs). After setting, long press the setting key for 3 seconds to exit the setting and the parameters will be saved automatically. In the day and night mode, there is no total operation sections, only the cycle can be set.

  4. Start and stop
    Long press the R / S key U7 (see user parameter TABLE-1 for details) seconds to start the the controller, and the remaining running time will be displayed in (2) display window . When the operation time is over, the controller will stop working, and the buzzer will beep u9 (see user parameter TABLE-1 for details) seconds.(2) display windowwill display “End”, or long press R / S key U7 seconds to stop the controller working and (2) display window will display “oFF”. When the controller is running in program mode and the total number of segments is greater than 1, if the constant temperature and humidity condition (see parameter table-1-u5, U6 for details) is not on, it is necessary to set each segment of time before starting the controller. When the time is all set to 0, the start will fail.

  5. Reservation function
    When the value of the reservation function AP (see the user parameter table-7) is set to 1, click the setting key again to set the reservation time, and the unit is minutes. Otherwise, it will directly exit the normal display interface. After the reservation time setting, it is needed to click the R / S key to start the machine. In the reservation timing, you can enter the parameter table again to modify the reservation time, or click the R / S key to stop the controller, and the setting time of running the machine is automatically cleared, which is only valid once.

  6. The memory function for power-down By modifying the parameter of power down memory (see “U2” parameter: internal parameter TABLE-1), you can select whether there is memory function for power-down.

Functions of Intrinsic Parameters

Press Setting to enter into the parameter setup menu after the instrument is powered on.

Internal Operation Parameters of the Control Panel


Press setting key: LC=000| |
Prompt| Name| Setting Range| Instruction
LCK=9 the parameter table of users

U1

|

Operation mode

|

(0~2)0

| 0:Fixed value operation mode;

1: Day and night mode, 0 ~ 99 cycles;

2: Program mode, which is programmable 1 ~ 30 segments, 0 ~ 99

cycles.

U2

| Operation mode of

power-down

|

(0~2)0

| 0: No operation;

1:     Operate from the first segment (daytime); 2:     Operate from power- down time.

U3

|

Timing correction

| (-999~999) 0| Correct the total timing error,

The correction value =  【 operating time (seconds) – actual time

(seconds)】 * 10 ÷ actual time (minutes).

U4| Time unit| (1~2) 1| 1:minute 0~9999; 2:hour 0~9999。

U5

|

Timing deviation in the condition of constant temperature

|

(0~10.0℃) 0

| If  the  deviation  between  the  measured  temperature  and  the

setting      value is less than U5, start the timing. Note: 0 means that there is no need to judge the temperature when timing.

U6

|

Timing deviation in the condition of constant humidity

|

(0~50.0%) 0

| If the deviation between the measured humidity and the setting

value is less than U6, start timing. Note: 0 means that there is no need to judge the humidity when timing.

U7| Setting Key Effective time| (0~10s)0| After long press U7(the time),Setting function key is valid

UA

|

illumination time

|

(0~9999min)      5

| After the lighting is turned on, the illumination time is automatically turned off.

Note: 0 means that the lighting must be turned off manually.

Ub

|

sterilization time

|

(0~9999min) 30

| After the sterilization is turned on, the sterilization time is automatically turned off.

Note: 0 means that sterilization must be turned off manually .

Uc| Address| (1~16) 1| Address of the machine.
LCK=103 **parameter table of temperature**

TH

|

the upper deviation is over temperature and an alarm appears;

|

(0~20.0℃) 4.0

| If “the measured value > the setting value + TH “, the upper deviation alarms and the temperature and humidity output will be turned off.

The relay has output when there is a temperature alarm; the buzzer

beeps; the alarm indicator is on, and the temperature unit flashes rapidly. Click any key to cancel the buzzer.

TL

| the lower deviation is over temperature and| (-50.0~0℃)

0

| If “the measured value < the setting value + TL “, the lower deviation alarms.

The relay has output when there is a temperature alarm; the

| an alarm appears;| | buzzer beeps, and the temperature unit flashes slowly. Click any key to cancel the buzzer.

Note: when “TL = 0”, this function is invalid.

---|---|---|---
Tb|

Deviation correction

| (-99.9~99.9℃)

0

| Correct measurement errors of the sensor in low temperature;

Tb = the actual temperature – the measured value.

TA

|

Slope correction

|

(-999~999) 0

| Correct measurement errors of the sensor in high temperature;

TA = **1000 *** (the actual temperature – the measured value) / the

measured value.

TP| Heating ratio| (0.1~50.0)      8.0| Adjust the function of the proportion of time
TI| Heating integral| (1~2000s)     500| Adjust the function of the integral
TD| Heating differential| (0~2000s)     200| Adjust the function of the differential
TT| Heating cycle| (1~60s)     5| The cycle of controlling Heating

Tc

|

Heating off

|

(-5.0~5.0℃)0

| Restrict the heating overshoot; when “the measured temperature

the setting temperature + Tc ”, heating stops output.

To| heating power| (0~100%) 100| The percentage of the maximum power of the heating output.

Tu

|

Turn on refrigeration

|

(-10.0~10.0℃)

0.6

| When the compressor is in manual start-stop mode and the compressor is under disconnected controlling mode, if “the measured temperature the setting temperature + Tu “, start the

compressor.

Tn

|

Turn off cooling

|

(-10.0~ uP )

0.6

| When the compressor is in manual start-stop mode and the compressor is under disconnected controlling mode, if “the measured temperature  ≤ the setting temperature + Tu Tn “, shut

down the compressor .

TE

|

High-temperature control, heating off

|

0

| Under controlling of high temperature, the temperature point where heating stops is effective only when the setting temperature is

higher than the ambient temperature.

LCK=203 **parameter table of humidity**

HH

| The upper deviation is

over humidity and an alarm appears

|

(0~50.0%)

20.0

| If “the measured humidity > the setting value + HH “, the upper deviation alarms; the humidity alarm relay is disconnected and the humidity output is turned off.

The humidity alarm relay has output when there is an alarm; the alarm indicator is on, and the humidity unit flashes rapidly.

HL

| The lower deviation is over humidity and an alarm appears|

(-50.0~0%)

0

| If “the measured humidity < the setting value+ HL “, the lower deviation alarms

The humidity alarm relay has output when there is an alarm; the alarm indicator is on, and the humidity unit flashes rapidly.

Note: this function is invalid when “HL = 0”

Hb

|

Deviation correction

| (-99.9~99.9%)

0

| Correct measurement errors of the sensor in low humidity;

Hb = the actual humidity – the measured value.

HA

|

Slope correction

|

(-999~999) 0

| Correct measurement errors of the sensor in high humidity;

HA = 1000 *** ( the actual humidity –** the measured value ) / the

measured value

HP| Humidifying ratio| (0.0~90.0)  10.0| Adjust the function of the proportion of time
---|---|---|---
HI| Humidifying integral| (1~999s)    200| Adjust the function of the integral

Hd

| Humidifying

differential

|

(0~999s)    30

|

Adjust the function of the differential

HT| Humidifying cycle| (0~60s)    5| The cycle of controlling Humidifying

Hc

| Low-temperature control, Humidifying off| (-50.0~50.0%)

0.0

| Under  controlling  of  low  humidity,  the  humidity  point  where

humidifying stops.

Ho| Humidification power| (0~100%) 100| The percentage of the maximum power of the humidifying output.

Hu

|

Turn on dehumidification

|

( Hn ~20.0%)3.0

| When the compressor is in manual start-stop mode and the compressor is under disconnected controlling mode, if “the measured  humidity the  setting humidity  +  Hu “,  start  the

compressor.

Hn

|

Turn off dehumidification

|

(-20.0~ Hu )3.0

| When the compressor is in manual start-stop mode and the compressor is under disconnected controlling mode, if “the measured  humidity the  setting humidity  +  Hn “,  stop  the

compressor.

HE

|

The point where there is no humidification

|

(0.0~10.0)2.0

| When the compressor is working in intermittent mode, if the setting value meets the conditions, the value can be modified and stop the

humidification in advance.

HF

| The point where there is no humidification in low humidity| (-10.0~10.0%)

-5.0

| When controlling low humidity, if “the measured humidity  ≥ the

setting humidity + HF “, the controller will prohibit humidification.

Instructions for humidifier

Connect the water inlet hole of the humidifier and the water tank with the hose of the machine, and add water to the water tank of the humidifier at the same time.

Precautions for Use

  1. This indicator has pre-set the following functions and protective measures:
  2. When the unit is energized for the first time or after the power failure, the compressor can enter into the operating state 1 to 3 minutes later.
  3. During the time: from starting the indicator or resetting the temperature to the temperature of the indicator firstly reaches the setting temperature:
  * No dehumidification when heating.
  * No humidification when refrigerating.
3. When the temperature in the indicator is 4 over or under the setting value, the buzzer will beep. The power of the internal heating tube will cut off when the temperature is over 70.
  1. In case of the circuit protection act caused by a too long door-opening time, the main power can be cut off.
  2. When turning high humidity to low humidity or using the low humidity function from low-temperature area to high-temperature area, please check whether there is any water in the incubator. If there is some, clear it out before use.
  3. There shall not be too much culture in the incubator, otherwise the air circulation inside will be affected.
  4. Do not impact the unit, or pull the probe in the unit, so as to avoid that it goes out of control.

Fault Analysis and Removal

The simple faults and removal methods are shown in Chart 4:

Fault Symptom

|

Analysis of Causes

|

Troubleshooting

|

Remarks

---|---|---|---
The power light isn’t working.| 1.  There is no power supply in the socket.

2.  The fuse protector blows out.

| 1.Check the socket 2.Replace the fuse protector|

Behind the box

The unit does not heat (or temperature does not rise.).

| 1.  The heating light is off, indicating the control board failure.

2.  The heating light is on, indicating the

blowing out of heating tube.

|

1.  Replace the control board.

2.  Replace the heating tube.

|

The unit can not refrigerate (or cool).

| 1.  The refrigeration light is off, indicating the control board failure.

2. The refrigeration light is on, indicating the failure in the refrigerating system.

3.  The time delay for compressor startup or re-operation is less than three minutes. Please

wait

|

1.  Replace the control board

2.  Repair the refrigerating system

3.  Please wait for a moment.

|
The heating or refrigeration light is on, but there is no sign of temperature rise or drop.|

Control board failure

|

Replace the control board

|

Located      in     the instrument

The temperature display stays at“ —” always| There is a short circuit in the temperature probe or an open circuit in the leading wire.|

Replace the probe Pt100

|
The buzzer sounds, and the temperature exceeds the set| The incubator failure. The possible causes include:| 1. Securely shut the door, and restart the unit after|
values ±4℃.| 1.  The door of the incubator is not securely shut.

2.  The refrigerant leaks.

3.  The circulating fan does not work.

4.  Control board failure

| having the power off for two minutes.

2.  Find out the leaking points, carry out repairing welding and recharge refrigerant.

3.  Replace the circulating fan

4.  Control board failure

Replace the control board

|
---|---|---|---
No humidification (The humidification light on

the display screen is on)

|

Humidifier failure

|

Replace humidifier

|
| ① The     humidity     in     the     incubator     is| ① Dehumidify the incubator|
The humidity always displays “99.9”| supersaturated

② There is an short circuit in the humidity sensor or the leading wire.

| or  open  the  door  of  the

incubator

| | ②Change the humidity sensor

Notes: In case the failures are not included in the above table or the above failures can not be removed, please contact the technical service department of our company or our maintenance points.

Maintenance

  1. The incubator must be kept in a well-ventilated and dry place away from sunshine. The distance between the unit and the wall shall not be less than 10cm.
  2. The incubator shall be equipped with a three-pin plug which shall be reliably grounded, in order to ensure the normal operation and personal safety.
  3. It is not allowed to clean the inner and exterior surfaces of the incubator with acid, alkali and other corrosive substances for the sake of beauty. It can be cleaned with neutral detergent and then wiped with dry cloth.
  4. When replacing the fuse protector, the one with the same blowing current and the same dimension as marked in the fuse protector shall be selected.
  5. During use, the incubator should be checked often to ensure that there is water in the humidifier tank.
  6. When humidifying is not used, the water in the humidifier should be removed and dried.
  7. When the incubator is not used, unplug it, and keep it dry and clean within.
  8. The dirt on the condenser (at the back of the incubator) shall be cleaned every six months, so as to ensure a favorable heat dissipation during the operation.

Transport and Storage

  1. Transport

    • Rough transport is not allowed.
    • During the transportation, the incubator shall be inclined at an angle less than 45 degrees. It is not allowed to place it flatwise or upside-down.
    • During the transportation of unpacked incubator, pay attention not to impact the refrigerating pipeline at the back of the incubator.
  2. Storage
    In case the incubator is not used for a long time before or after unpacking, it shall be stored in a place with the following conditions.

    • Ambient humidity: 040
    • Relative humidity: 80%RH
    • Range of atmospheric pressure: 86kPa106kPa

Out of box audit

  1. After unpacking the incubator, you shall take out the Instruction Manual to check whether all the attachments are complete and whether the product specification conforms to the Manual according to the “List of Attachments”.
  2. Please inspect and operate the incubator according to the Instruction Book.

List of Attachments

Model      Name SHH-250SD
1. Instruction Manual 1
2.Quality tracking card 1
3. Supporting net 4
4. Hook 18
5. Drip tray 1

Instruction of the printer and USB controller (optional when ordering)

Accessories

  1. Switch Power SupplyAC100-264VDC5V 4A
  2. The size of holes of the controller28*70millimeter

Indication Panel of the incubator

  1. Time window; display the current time (hours. minutes) under normal state
  2. OUT indicator light: when the light is on, it indicates that there is print output or data written into USB flash disk;
  3. COM indicator light: flashing indicates that the current communication is normal;
  4. USB indicator: when the light is on, it indicates that the current USB flash disk has been inserted correctly, and the data can be written in;
  5. key to increase: In normal state, click or long press this key to increase the setting interval time of USB flash disk data writing . In parameter setting state, click or long press this key to increase the setting parameter. When the setting value is the maximum, pressing this key again will automatically flip to the minimum value;
  6. key to decrease and confirm: In normal state, click or long press this key to decrease the setting interval time of USB flash disk data writing . In parameter setting state, click this key to switch setting parameters; long press this key to exit the setting state and save the setting values;
  7. composite key: In normal state, long press two keys at the same time; 3 seconds later, enter the parameter setting state;

Operation

After power on, the digital tube and indicator light are all on for 3 seconds, and enter the running state. The PRT indicator light on, the time window displays the current time (hours. minutes). The printer prints “print test”, “current date” and “current data”, and the OUT indicator light is on when printing output. After that, the data are printed according to the printing interval time, and when the date changes, the date will be printed.
In normal stateclick or to set the printing interval. The setting range is 1-9999 minutes.In normal state, long press at the same time for 3 seconds to enter the password-input state, and input the corresponding password to adjust the time and parameters.

Prompts for abnormal conditions

  1. The time window shows “TErr”, which indicates communication error.
  2. Check whether the communication line with the instrument is connected correctly or not.
  3. Check whether the power is on.
  4. The time window shows UErr, which indicates there is an error in the reading and writing of the USB flash disk.
  5. Check whether the USB flash disk is inserted correctly
  6. Format the USB flash disk or replace it
  7. If USB storage is not required, you can enter the internal parameters to turn off USB flash disk function

Internal parameters

In normal state, long press at the same time for 3 seconds; ” Lc00 ” displays; type the correct password 3 or 9, and click again to enter the internal parameter. Then long press for 3 seconds to exit that situation. The details are as follows:

Passw

ord

| Parameter

indication

| Parameter

name

| Instructions of

function of parameter

|

Range

---|---|---|---|---

3

| Y-| Year| Set the current date  Year| 00~99
N-| Month| Set the current date     Month| 00~12
d-| Day| Set the current date     Day| 00~31
H-| Hour| Set the current date     Hour| 00~59
n-| Minute| Set the current date     Minute| 00~59
S-| Second| Set the current date     Second| 00~59

Wiring connection

  1. As shown in the figure, connect pins 1 and 2 of the printer controller and power Cords of the printer (the red are positive) to 5V output terminals of power supply of switch(4 and 5 terminals); 1 and 2 pins of power supply of switch are connected with 220 V AC.
  2. Connect the communication ports (A and B) of the matching temperature-control instrument with the communication ports (5 and 6) of the printer controller. Connect PC communication ports (3, 4) with PC. (Note: the communication connecting wires must be placed separately from the strong wires. It is better to use shielded wires to avoid interference.)Shanghai Yinxi Instrument Co Ltd 250SD Precision Incubators of Constant Temperature and Humidity fig 6

Special reminder

  1. In order to reduce the interference to the instrument during wiring, please pay attention to the separation of strong current lines (such as power lines and load lines) and weak current lines (communication data lines, etc.).
  2. The cables between the printer and the controller must be compressed, otherwise there will be printing errors.
  3. The power supply of both the printer and its controller is DC5V, only the power supply at the switch needs to be connected to AC220V. Do not connect the terminals of high voltage power supply and low voltage power supply incorrectly.
  4. If the computer monitoring is not needed, communication ports of the PC does not need wiring. Please refer to the printer manual for maintenance and operation of the printer.

Special announcement

  1. According to related regulations, the product, from the date of purchase, will be guaranteed for one year, except the man-made faults.
  2. Our company reserves the right to change the design of the product. The design is subject to change without prior notice.
  3. The Instruction Manual covers the contents of technical instruction.

Wiring diagram

Shanghai Yinxi Instrument Co Ltd 250SD Precision Incubators of Constant
Temperature and Humidity fig 7

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals