METERK 9999 Intelligent Digital Counting Multimeter User Manual

June 11, 2024
Meterk

9999 Intelligent Digital Counting Multimeter

Product Information: Intelligent Digital 9999 Counting

Multimeter

The Intelligent Digital 9999 Counting Multimeter is an
electronic measuring instrument that meets the safety requirements
of IEC-61010-1, IEC-61010-2-030, IEC-61010-2-032. It has a maximum
voltage of 1000V DC or 750V AC and belongs to secondary pollution,
with an overvoltage standard of CAT III 600V. The meter has a
non-contact voltage detection sensing area, flashlight function
switch button, LCD display monitor, input socket, and power button.
The LCD touch display shows AC and DC connection/disconnection
indication, automatic range mode, automatic shutdown indication,
low battery reading hold state, volt (voltage), ampere (current),
ohm kilohm and megohm (resistance), hertz, diode, microfarad,
centigrade, and Fahrenheit. The meter has a working height of up to
2000m and a maximum display value of 9999 digits.

Product Usage Instructions:

  1. Safety Information: To ensure safe operation, please
    read the user manual carefully and follow the safety regulations
    and effective safety precautions.

  2. Part Name: The meter has a non-contact voltage detection
    sensing area, flashlight function switch button, LCD display
    monitor, input socket, and power button.

  3. LCD Touch Display: The LCD touch display shows AC and DC
    connection/disconnection indication, automatic range mode,
    automatic shutdown indication, low battery reading hold state, volt
    (voltage), ampere (current), ohm kilohm and megohm (resistance),
    hertz, diode, microfarad, centigrade, and Fahrenheit.

  4. Technical Indicators: The meter has a maximum voltage of
    1000V DC or 750V AC and belongs to secondary pollution, with an
    overvoltage standard of CAT III 600V. It has a working height of up
    to 2000m and a maximum display value of 9999 digits. The meter has
    a non-contact voltage detection sensing area, flashlight function
    switch button, LCD display monitor, input socket, and power
    button.

    • Alternating Current: The range is 999.9mA to 10A, with a
      minimum input current of 20mA AC and a maximum input current of 10A
      AC. The frequency range is 40HZ-1000Hz.

    • Direct Current: The range and resolution are not
      specified in the user manual.

Intelligent digital 9999 counting multimeter

AUTO

SEL

9999 Counts TRUE RMS NCV

SEL

A

com

INPUT

directory
directory
1. Safety information……………………………………………………… 1 1.1 safety instructions……………………………………………………….1 1.2 electrical symbols………………………………………………………..1 1.3 maintenance………………………………………………………………..2 2. The description…………………………………………………………….2 2.1 part name…………………………………………………………………… 2 2.1 LCD touch monitor……………………………………………………….3 3. The technical indexes…………………………………………………..4 3.1 comprehensive index…………………………………………………..4 3.2 technical indicators……………………………………………………..5 4. Operating instructions……………………………………………….. 8 4.1 keep reading………………………………………………………………. 8 4.2 the flashlight……………………………………………………………….8 4.3 automatic shutdown…………………………………………………… 8 4.4 measurement……………………………………………………………… 8 4.5 non-contact voltage detection (NCV)…………………………..9 4.6 dc voltage measurement…………………………………………….. 9 4.7 ac voltage measurement…………………………………………….. 10 4.8 ac/dc current measurement………………………………………… 11 4.9 resistance measurement………………………………………………11 4.10 wire test……………………………………………………………………. 12 4.11 capacitance measurement………………………………………… 12 4.12 temperature measurement……………………………………….. 12 4.13 diode test…………………………………………………………………..13 4.14 circuit on-off test……………………………………………………….13 5. Maintain……………………………………………………………………….14 5.1 Replace Battery…………………………………………………………… 14 5.2 replacement pens and………………………………………………….14 6. The attachment…………………………………………………………….14

1.Safety information warning

Safety information

To fully understand the function of the meter and ensure safe operation, please read and follow it carefully Follow the use method of this manual.
Special care should be taken when using this meter, improper use may cause electric shock or damage to the meter. In use, it should follow the usual safety regulations and take effective safety precautions.

The instrument meets the safety requirements of IEC-61010-1, IEC-61010-2-030, IEC-61010-2-032) electronic measuring instrument, belongs to secondary pollution, and the overvoltage standard is CAT III 600V
Please follow the safety operation guide and use the meter safely. 1.1Safety instructions 1.1.1 When using the meter, users must follow standard safety rules ‘ -Universal protection against electric shock ‘ -Prevent misuse of meter 1.1.2 After receiving the meter, check for damage in transit. 1.1.3 After storage and shipment in poor condition, check and
confirm whether the meter is damaged. 1.1.4 The pen must be in good condition. Before use, check whether
the insulation of the pen is damaged and whether the metal wire of the wire is exposed 1.2Electrical symbols Note (For important safety information, please refer to the instruction manual) Can be used on dangerous charged objects Double insulation protection (Class II)
In accordance with iec-61010-1 overvoltage (installation) Class II, contamination level 2 refers to the level of voltage protection provided to withstand pulses.
1

describe
Comply with European Community (EU) standards. Grounding 1.3maintenance 1.3.1Do not attempt to open the bottom case to adjust or repair the instrument. Such operations should only be performed by technicians who are fully aware of the instrument and the risks of electric shock. 1.3.2The pen should be removed from the line under test before opening the instrument backing. 1.3.3To avoid electric shocks that may be caused by incorrect readings, charge the meter immediately when it shows the ” ” symbol. 1.3.4Clean the meter with a damp cloth and mild detergent. Do not use abrasives or solvents. 1.3.5Turn off the power when the meter is not in use.
2describe 2.1 Part name

Non-contact voltage detection sensing area Flashlight Function switch button LCD display monitor Input socket The power button
2

AUTO

SEL

9999 Counts TRUE RMS NCV

SEL

A

com

INPUT

2.2 LCD touch display

describe

AUTO SEL
AC & DC Connection/Disconnection indication Automatic range mode Automatic shutdown indication Low battery Reading hold state Volt (voltage), ampere (current) Ohm kilohm and megohm (resistance) Hertz DIODE Microfarad Centigrade and Fahrenheit
The line of fire test
3

3. Technical indicators

Technical indicators

Gauges shall be specified for a period of one year, between

18°C and 28°C, with relative humidity less than Recalibration in

75% condition.

3.1 Comprehensive Indicators

1.Automatic selection of measuring function and range.

2.Full range overload protection.

3.The maximum voltage between the measuring end and the

ground is 1000V DC or 750V AC

4.Working height: maximum 2000m

5.Display: LCD

6.The maximum display value is 9999 digits.

7.Polarity indication: automatic indication, ‘-‘ indicates negative

polarity.

8.Superview display: ‘OL’ or ‘-OL’.

9.Sampling time: about 3 times/second

10.Unit display: with function, electric quantity unit display.

11.Automatic shutdown time: 15 minutes

12.Power supply: Two 1.5V AAA 7 batteries

13.Battery undervoltage indicator: LCD

symbol.

14.Temperature coefficient: less than 0.1X accuracy /°C

15.Operating temperature: 0° C-40 ° C

16.Storage temperature: -10°C- 50°C

4

3.2Technical indicators 3.2.1Alternating current

Technical indicators

Range

Resolution

Accuracy

999.9mA 10A

± (1.0% reading + 8 digits)

Minimum input current 20mA AC current

Maximum input current 10A AC current

Frequency range: 40HZ-1000Hz;

3.2.2Dc current

Range

Resolution

Accuracy

999.9mA 10A

± (1.0% reading +5 digits)

Minimum input current 20mA DC current Maximum input current 10A DC current 3.2.3Dc voltage

Range

Resolution

Accuracy

600.0mV

9.999V 99.99V

± (0.5% reading + 3 digits)

999.9V
In AUTO mode, the input voltage ranges from 0.5V to 1000V DC The input voltage ranges from 0.0mV to 600.0mV DC in mV mode

Note:
In the mV range, with the input open, the meter may have a pulsating reading, This is normal because of the high sensitivity of the meter when connected to the pen When you measure the circuit, you get a true measurement.

5

3.2.4Ac voltage

Technical indicators

Range

Resolution

Accuracy

600.0mV

9.999V 99.99V

±( 0.8% reading + 5 digits)

999.9V
In AUTO mode, the input voltage ranges from 0.5V to 1000V AC In mV mode, the input voltage ranges from 0.0mV to 600.0mV AC Frequency rate pattern: 40HZ- 1000Hz

3.2.5Line connection test

Range

Resolution

explain

If resistance of the line be ing measured is less than 50 , buzzer in instrument will make continuous alarming sounds.

Overload protection: 600V DC or AC (valid value) 3.2.6 Resistance

Range

Resolution

Accuracy

999.9

9.999K

99.99K 999.9K

± (0.8% reading + 3 digits)

9.999M

99.99M

Input protection: maximum 600V DC or AC valid value.

6

3.2.7Diode Range

Resolution Accuracy

Technical indicators

Input protection: the maximum value is 600V DC or AC. 3.2.8Capacitance

Range

Resolution

Accuracy

9.999nF

reading digits

999.9nF

9.999uF

99.99uF 999.9uF

±( 2.5%reading+20 digits)

9.999mF

99.99mF

Input protection: the maximum value is 600V DC or AC. 3.2.9The temperature

Range Resolution

Accuracy

± (1% reading +3)

± (1% reading +3)
Accuracy does not include error of thermocouple probe. Overload protection 250V DC or AC (valid value)

7

3.2.10Frequency 3.2 .10.1 Pass grade A:

Range

Resolution

Operational guidelines
Accuracy reading digits

Measuring range: 40HZ-1000Hz Input signal pattern: > 3A AC current (Effective Value) 3.2.10.2 Through V gear:

Range

Resolution

Accuracy

reading digits

Measuring range: 40HZ-1000Hz Input signal pattern: > 0.8V AC current (Effective value) 4.Operational guidelines 4.1Reading hold
In the process of measurement, if you need to hold the reading, touch the key, the display value will be locked, touch the key again, can remove the reading hold. 4.2Flashlight
Long press the button to turn on the lighting function of the flashlight barrel, which will be automatically turned off after about 5 minute. 4.3Automatic shutdown 4.3.1If there is no operation within 15 minutes after startup, the
meter will enter hibernation state and automatically shut down to save electric energy. The buzzer has one sound one minute before the shutdown. 4.3.2After automatic shutdown, press the power button to wake up the instrument to work.
8

Operational guidelines
4.4Preparation before measurement 4.4.1 Tap the power source button to power on. If the battery
voltage is low (about 2.8V), the monitor will display ” ” symbol, then should charge. Press the power button again to turn it off. 4.4.2 When the instrument is not measured, the instrument enters the automatic scanning state and the instrument displays 4. 5 Non-contact Voltage Detection (NCV) 4.5.1Touch the NCV key to enable the NCV function, and the instrument displays NCV 4.5.2The NCV sensor is placed close to the measured wire, and the instrument can detect whether the measured wire is gt. 90V ac voltage. When the meter detects the ac power, the meter buzzer alarms and the LCD analog bar displays the induction intensity.

AUTO

SEL

9999 Counts TRUE RMS NCV

SEL

Note:

A

com

INPUT

  1. Voltage may still exist even if there is no alarm indication. Do not

rely on contactless voltage detectors To see if there’s a voltage

on the wire. Probe operation may depend on socket design,

insulation thickness Different types of factors such as the impact.

  1. In NCV mode, the meter does not measure voltage, resistance,

and current at the same tim

9

Operational guidelines

4.6Ac-dc current measurement

4.6.1Insert the stylus into the 10A current input hole, and the

current input terminal is inserted into the stylus automatic

identification function on on Automatically enters the default

DC current gear of the current gear system. Remove the 10A

current input hole and the stylus automatically jumps out of the

current gear and enters the AUTO measurement mode

Note:

When the meter is in diode

gear, the pen is inserted into the

10A input hole and the meter will

not enter current gear.

4.6.2Connect the pen into the

measured signal, when the

measured signal > The meter will display the current measured value at 20mA. When measured

AUTO SEL

9999 Counts TRUE RMS NCV

SEL

ac signal > At 3000rnA, the

instrument pair displays the AC

A

com

INPUT

current frequency.

4.7Dc voltage measurement Insert the black stylus into the COM jack and the red stylus into
the V jack. Connect the markers together The voltage source or load ends are measured and read on the LCD. AUTO automatic measurement mode:
when the measured signal > At 0.5V, the meter will display the current measured DC voltage value. When the signal being measured. At 0.5V, the meter will default to the resistance value and display the internal resistance value of the measured signal.
Press ” SEL ” key to switch to DC mV The voltage profile. Connect the pen to both ends of the voltage source or load for measurement, reading on the LCD.

10

DC mV measurement mode: when the measured signal < At 600.0mV, the meter will display the current measured DC voltage value. When the measured signal > The meter will display “OL” at 600.0mV.

Operational guidelines

AUTO

SEL

9999 Counts TRUE RMS NCV

SEL

A

com

INPUT

4.8Ac voltage measurement

Insert the black stylus into the COM jack and the red stylus into

the V jack. Connect the pen to both ends of the voltage source or

load for measurement, reading on the LCD.

AUTO automatic measurement mode:

when the measured signal > At 0.5V, the meter will display the

current measured AC voltage value. When the signal being

measured. 0.5 VInstrument will

default to the resistance value,

display the measured signal internal

resistance value.

Press “SEL” to switch to AC mV

voltage.Connect the pen to both

ends of the voltage source or load

for measurement, reading on the LCD. AC mV measurement mode:
when the measured signal < At

AUTO

SEL

9999 Counts TRUE RMS NCV

SEL

600.0mV, the meter will display the current measured AC voltage value.

A

com

INPUT

When the signal is measured>

The meter will display “OL” at 600.0mV.”

11

4.9 Resistance measurement warning

Operational guidelines

When measuring the impedance on the line, make sure that the circuit is disconnected and the capacitor on the circuit discharges completely.

4.9.1 Insert the black stylus into the COM jack and the red stylus
into the ” ” jack.
4.9.2 AUTO AUTO range tap. At this time, the meter is in automatic scanning state.
4.9.3 Connect the pen to the measured resistance or both ends of the line for measurement, reading in the LCD display.
Note: 1) When the input is open, the LCD will display the “OL” overrange status. 2) If the resistance being measured is higher than 10M, the meter may take several seconds to stabilize the reading, which is normal for high resistance readings.

AUTO

SEL

9999 Counts TRUE RMS NCV

SEL

A

com

INPUT

12

Operational guidelines
4 .10 The line of fire test 4.10.1Press “SEL” key to switch to LIVE gear, the instrument
shows LIVE 4.10.2Connect the red test pen to the Live input socket. Insert the
single pen into the power socket L jack or close to the live wire. If the instrument detects ac voltage, it will display voltage signal strength according to the detected signal strength, and the buzzer will give alarm sound of different intensity. 4 .11Capacitance measurement 4.11.1Press the “SEL” key to switch to capacitor gear. 4.11.2Use the other end of the test pen to measure the capacitance value of the capacitor to be measured and read the measured value from the LCD. Note: 1)When measuring large capacitors, a fixed time is
required to stabilize readings. 2)When measuring polar capacitors, pay attention to
the corresponding polarity to avoid damaging the instrument. 4 .12Temperature measurement Press “SEL” to switch TEMP files, thermocouple red plug into the end of °C /°F, black plug into the COM jack. When the reading is stable, the temperature can be read directly from the display screen. Note: The maximum measurement temperature of type K thermocouple is 250°, instantaneous measurement up to 300°C. 4.13Diode test 4.13.1 Insert the black stylus into the COM jack and the red stylus into the ” ” jack. 4.13.2 Touch SEL key to switch to diode position. 4.13.3 Connect the red stylus to the diode anode and the black stylus to the diode cathode for testing.
13

Operational guidelines

4.13.4 Read on the LCD. Note: The meter displays an approximation of the forward voltage drop of the diode. If the pen is connected inversely or the pen is open, the LCD displays “OL”.
4 .14Line connection test
warning

AUTO

SEL

9999 Counts TRUE RMS NCV

SEL

A

com

INPUT

Danger of electric shock. When testing the circuit on and off, make sure the circuit is disconnected and the capacitor on the circuit is completely discharged

4.14.1Insert the black marker into

the COM jack and the red

marker into the ” ” jack. 4.14.2Touch the “SEL” key to switch
to the line connection test state.

AUTO

SEL

9999 Counts TRUE RMS NCV

SEL

4.14.3Connect the pen to both ends of the line for measurement.

A

com

INPUT

If the resistance of the measured

line is less than 50 , the buzzer inside the meter will sound.

Note:

If the meter is open or the resistance of the measured

line is greater than 999.9, “OL” is displayed.

14

5.Maintenance 5.1 Replace Battery

Maintenance

warning

When replacing the battery, remove the pen from the measuring circuit to avoid electric shock.

5.1.1 the ” ” symbol indicates that charging is needed.

5.2Replace the test pen warning

When replacing the stylus, you must replace the same or the same grade of the table. Pen meter must be in good condition, pen grade: 1000V 10A.

Note: If the insulation layer of the pen is damaged, such as
exposed wire of the wire, the pen must be replaced.

6.The attachment 1)Test the pen 2)Directions for use 3)thermocouple

Grade: 1000V 10A

A pair of A copy of the
a

15

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Meterk User Manuals

Related Manuals