METERK 9999 Intelligent Digital Counting Multimeter User Manual
- June 11, 2024
- Meterk
Table of Contents
9999 Intelligent Digital Counting Multimeter
Product Information: Intelligent Digital 9999 Counting
Multimeter
The Intelligent Digital 9999 Counting Multimeter is an
electronic measuring instrument that meets the safety requirements
of IEC-61010-1, IEC-61010-2-030, IEC-61010-2-032. It has a maximum
voltage of 1000V DC or 750V AC and belongs to secondary pollution,
with an overvoltage standard of CAT III 600V. The meter has a
non-contact voltage detection sensing area, flashlight function
switch button, LCD display monitor, input socket, and power button.
The LCD touch display shows AC and DC connection/disconnection
indication, automatic range mode, automatic shutdown indication,
low battery reading hold state, volt (voltage), ampere (current),
ohm kilohm and megohm (resistance), hertz, diode, microfarad,
centigrade, and Fahrenheit. The meter has a working height of up to
2000m and a maximum display value of 9999 digits.
Product Usage Instructions:
-
Safety Information: To ensure safe operation, please
read the user manual carefully and follow the safety regulations
and effective safety precautions. -
Part Name: The meter has a non-contact voltage detection
sensing area, flashlight function switch button, LCD display
monitor, input socket, and power button. -
LCD Touch Display: The LCD touch display shows AC and DC
connection/disconnection indication, automatic range mode,
automatic shutdown indication, low battery reading hold state, volt
(voltage), ampere (current), ohm kilohm and megohm (resistance),
hertz, diode, microfarad, centigrade, and Fahrenheit. -
Technical Indicators: The meter has a maximum voltage of
1000V DC or 750V AC and belongs to secondary pollution, with an
overvoltage standard of CAT III 600V. It has a working height of up
to 2000m and a maximum display value of 9999 digits. The meter has
a non-contact voltage detection sensing area, flashlight function
switch button, LCD display monitor, input socket, and power
button.-
Alternating Current: The range is 999.9mA to 10A, with a
minimum input current of 20mA AC and a maximum input current of 10A
AC. The frequency range is 40HZ-1000Hz. -
Direct Current: The range and resolution are not
specified in the user manual.
-
Intelligent digital 9999 counting multimeter
AUTO
SEL
9999 Counts TRUE RMS NCV
SEL
A
com
INPUT
directory
directory
1. Safety information……………………………………………………… 1 1.1 safety
instructions……………………………………………………….1 1.2 electrical
symbols………………………………………………………..1 1.3 maintenance………………………………………………………………..2 2.
The description…………………………………………………………….2 2.1 part
name…………………………………………………………………… 2 2.1 LCD touch monitor……………………………………………………….3
3. The technical indexes…………………………………………………..4 3.1 comprehensive
index…………………………………………………..4 3.2 technical indicators……………………………………………………..5 4.
Operating instructions……………………………………………….. 8 4.1 keep
reading………………………………………………………………. 8 4.2 the
flashlight……………………………………………………………….8 4.3 automatic
shutdown…………………………………………………… 8 4.4 measurement……………………………………………………………… 8 4.5
non-contact voltage detection (NCV)…………………………..9 4.6 dc voltage
measurement…………………………………………….. 9 4.7 ac voltage measurement……………………………………………..
10 4.8 ac/dc current measurement………………………………………… 11 4.9 resistance
measurement………………………………………………11 4.10 wire test……………………………………………………………………. 12
4.11 capacitance measurement………………………………………… 12 4.12 temperature
measurement……………………………………….. 12 4.13 diode test…………………………………………………………………..13
4.14 circuit on-off test……………………………………………………….13 5.
Maintain……………………………………………………………………….14 5.1 Replace
Battery…………………………………………………………… 14 5.2 replacement pens
and………………………………………………….14 6. The attachment…………………………………………………………….14
1.Safety information warning
Safety information
To fully understand the function of the meter and ensure safe operation,
please read and follow it carefully Follow the use method of this manual.
Special care should be taken when using this meter, improper use may cause
electric shock or damage to the meter. In use, it should follow the usual
safety regulations and take effective safety precautions.
The instrument meets the safety requirements of IEC-61010-1, IEC-61010-2-030,
IEC-61010-2-032) electronic measuring instrument, belongs to secondary
pollution, and the overvoltage standard is CAT III 600V
Please follow the safety operation guide and use the meter safely. 1.1Safety
instructions 1.1.1 When using the meter, users must follow standard safety
rules ‘ -Universal protection against electric shock ‘ -Prevent misuse of
meter 1.1.2 After receiving the meter, check for damage in transit. 1.1.3
After storage and shipment in poor condition, check and
confirm whether the meter is damaged. 1.1.4 The pen must be in good condition.
Before use, check whether
the insulation of the pen is damaged and whether the metal wire of the wire is
exposed 1.2Electrical symbols Note (For important safety information, please
refer to the instruction manual) Can be used on dangerous charged objects
Double insulation protection (Class II)
In accordance with iec-61010-1 overvoltage (installation) Class II,
contamination level 2 refers to the level of voltage protection provided to
withstand pulses.
1
describe
Comply with European Community (EU) standards. Grounding 1.3maintenance
1.3.1Do not attempt to open the bottom case to adjust or repair the
instrument. Such operations should only be performed by technicians who are
fully aware of the instrument and the risks of electric shock. 1.3.2The pen
should be removed from the line under test before opening the instrument
backing. 1.3.3To avoid electric shocks that may be caused by incorrect
readings, charge the meter immediately when it shows the ” ” symbol.
1.3.4Clean the meter with a damp cloth and mild detergent. Do not use
abrasives or solvents. 1.3.5Turn off the power when the meter is not in use.
2describe 2.1 Part name
Non-contact voltage detection sensing area Flashlight Function switch button
LCD display monitor Input socket The power button
2
AUTO
SEL
9999 Counts TRUE RMS NCV
SEL
A
com
INPUT
2.2 LCD touch display
describe
AUTO SEL
AC & DC Connection/Disconnection indication Automatic range mode Automatic
shutdown indication Low battery Reading hold state Volt (voltage), ampere
(current) Ohm kilohm and megohm (resistance) Hertz DIODE Microfarad Centigrade
and Fahrenheit
The line of fire test
3
3. Technical indicators
Technical indicators
Gauges shall be specified for a period of one year, between
18°C and 28°C, with relative humidity less than Recalibration in
75% condition.
3.1 Comprehensive Indicators
1.Automatic selection of measuring function and range.
2.Full range overload protection.
3.The maximum voltage between the measuring end and the
ground is 1000V DC or 750V AC
4.Working height: maximum 2000m
5.Display: LCD
6.The maximum display value is 9999 digits.
7.Polarity indication: automatic indication, ‘-‘ indicates negative
polarity.
8.Superview display: ‘OL’ or ‘-OL’.
9.Sampling time: about 3 times/second
10.Unit display: with function, electric quantity unit display.
11.Automatic shutdown time: 15 minutes
12.Power supply: Two 1.5V AAA 7 batteries
13.Battery undervoltage indicator: LCD
symbol.
14.Temperature coefficient: less than 0.1X accuracy /°C
15.Operating temperature: 0° C-40 ° C
16.Storage temperature: -10°C- 50°C
4
3.2Technical indicators 3.2.1Alternating current
Technical indicators
Range
Resolution
Accuracy
999.9mA 10A
± (1.0% reading + 8 digits)
Minimum input current 20mA AC current
Maximum input current 10A AC current
Frequency range: 40HZ-1000Hz;
3.2.2Dc current
Range
Resolution
Accuracy
999.9mA 10A
± (1.0% reading +5 digits)
Minimum input current 20mA DC current Maximum input current 10A DC current 3.2.3Dc voltage
Range
Resolution
Accuracy
600.0mV
9.999V 99.99V
± (0.5% reading + 3 digits)
999.9V
In AUTO mode, the input voltage ranges from 0.5V to 1000V DC The input voltage
ranges from 0.0mV to 600.0mV DC in mV mode
Note:
In the mV range, with the input open, the meter may have a pulsating reading,
This is normal because of the high sensitivity of the meter when connected to
the pen When you measure the circuit, you get a true measurement.
5
3.2.4Ac voltage
Technical indicators
Range
Resolution
Accuracy
600.0mV
9.999V 99.99V
±( 0.8% reading + 5 digits)
999.9V
In AUTO mode, the input voltage ranges from 0.5V to 1000V AC In mV mode, the
input voltage ranges from 0.0mV to 600.0mV AC Frequency rate pattern: 40HZ-
1000Hz
3.2.5Line connection test
Range
Resolution
explain
If resistance of the line be ing measured is less than 50 , buzzer in instrument will make continuous alarming sounds.
Overload protection: 600V DC or AC (valid value) 3.2.6 Resistance
Range
Resolution
Accuracy
999.9
9.999K
99.99K 999.9K
± (0.8% reading + 3 digits)
9.999M
99.99M
Input protection: maximum 600V DC or AC valid value.
6
3.2.7Diode Range
Resolution Accuracy
Technical indicators
Input protection: the maximum value is 600V DC or AC. 3.2.8Capacitance
Range
Resolution
Accuracy
9.999nF
reading digits
999.9nF
9.999uF
99.99uF 999.9uF
±( 2.5%reading+20 digits)
9.999mF
99.99mF
Input protection: the maximum value is 600V DC or AC. 3.2.9The temperature
Range Resolution
Accuracy
± (1% reading +3)
± (1% reading +3)
Accuracy does not include error of thermocouple probe. Overload protection
250V DC or AC (valid value)
7
3.2.10Frequency 3.2 .10.1 Pass grade A:
Range
Resolution
Operational guidelines
Accuracy reading digits
Measuring range: 40HZ-1000Hz Input signal pattern: > 3A AC current (Effective Value) 3.2.10.2 Through V gear:
Range
Resolution
Accuracy
reading digits
Measuring range: 40HZ-1000Hz Input signal pattern: > 0.8V AC current
(Effective value) 4.Operational guidelines 4.1Reading hold
In the process of measurement, if you need to hold the reading, touch the key,
the display value will be locked, touch the key again, can remove the reading
hold. 4.2Flashlight
Long press the button to turn on the lighting function of the flashlight
barrel, which will be automatically turned off after about 5 minute.
4.3Automatic shutdown 4.3.1If there is no operation within 15 minutes after
startup, the
meter will enter hibernation state and automatically shut down to save
electric energy. The buzzer has one sound one minute before the shutdown.
4.3.2After automatic shutdown, press the power button to wake up the
instrument to work.
8
Operational guidelines
4.4Preparation before measurement 4.4.1 Tap the power source button to power
on. If the battery
voltage is low (about 2.8V), the monitor will display ” ” symbol, then should
charge. Press the power button again to turn it off. 4.4.2 When the instrument
is not measured, the instrument enters the automatic scanning state and the
instrument displays 4. 5 Non-contact Voltage Detection (NCV) 4.5.1Touch the
NCV key to enable the NCV function, and the instrument displays NCV 4.5.2The
NCV sensor is placed close to the measured wire, and the instrument can detect
whether the measured wire is gt. 90V ac voltage. When the meter detects the ac
power, the meter buzzer alarms and the LCD analog bar displays the induction
intensity.
AUTO
SEL
9999 Counts TRUE RMS NCV
SEL
Note:
A
com
INPUT
- Voltage may still exist even if there is no alarm indication. Do not
rely on contactless voltage detectors To see if there’s a voltage
on the wire. Probe operation may depend on socket design,
insulation thickness Different types of factors such as the impact.
- In NCV mode, the meter does not measure voltage, resistance,
and current at the same tim
9
Operational guidelines
4.6Ac-dc current measurement
4.6.1Insert the stylus into the 10A current input hole, and the
current input terminal is inserted into the stylus automatic
identification function on on Automatically enters the default
DC current gear of the current gear system. Remove the 10A
current input hole and the stylus automatically jumps out of the
current gear and enters the AUTO measurement mode
Note:
When the meter is in diode
gear, the pen is inserted into the
10A input hole and the meter will
not enter current gear.
4.6.2Connect the pen into the
measured signal, when the
measured signal > The meter will display the current measured value at 20mA. When measured
AUTO SEL
9999 Counts TRUE RMS NCV
SEL
ac signal > At 3000rnA, the
instrument pair displays the AC
A
com
INPUT
current frequency.
4.7Dc voltage measurement Insert the black stylus into the COM jack and the
red stylus into
the V jack. Connect the markers together The voltage source or load ends are
measured and read on the LCD. AUTO automatic measurement mode:
when the measured signal > At 0.5V, the meter will display the current
measured DC voltage value. When the signal being measured. At 0.5V, the meter
will default to the resistance value and display the internal resistance value
of the measured signal.
Press ” SEL ” key to switch to DC mV The voltage profile. Connect the pen to
both ends of the voltage source or load for measurement, reading on the LCD.
10
DC mV measurement mode: when the measured signal < At 600.0mV, the meter will display the current measured DC voltage value. When the measured signal > The meter will display “OL” at 600.0mV.
Operational guidelines
AUTO
SEL
9999 Counts TRUE RMS NCV
SEL
A
com
INPUT
4.8Ac voltage measurement
Insert the black stylus into the COM jack and the red stylus into
the V jack. Connect the pen to both ends of the voltage source or
load for measurement, reading on the LCD.
AUTO automatic measurement mode:
when the measured signal > At 0.5V, the meter will display the
current measured AC voltage value. When the signal being
measured. 0.5 VInstrument will
default to the resistance value,
display the measured signal internal
resistance value.
Press “SEL” to switch to AC mV
voltage.Connect the pen to both
ends of the voltage source or load
for measurement, reading on the LCD. AC mV measurement mode:
when the measured signal < At
AUTO
SEL
9999 Counts TRUE RMS NCV
SEL
600.0mV, the meter will display the current measured AC voltage value.
A
com
INPUT
When the signal is measured>
The meter will display “OL” at 600.0mV.”
11
4.9 Resistance measurement warning
Operational guidelines
When measuring the impedance on the line, make sure that the circuit is disconnected and the capacitor on the circuit discharges completely.
4.9.1 Insert the black stylus into the COM jack and the red stylus
into the ” ” jack.
4.9.2 AUTO AUTO range tap. At this time, the meter is in automatic scanning
state.
4.9.3 Connect the pen to the measured resistance or both ends of the line for
measurement, reading in the LCD display.
Note: 1) When the input is open, the LCD will display the “OL” overrange
status. 2) If the resistance being measured is higher than 10M, the meter may
take several seconds to stabilize the reading, which is normal for high
resistance readings.
AUTO
SEL
9999 Counts TRUE RMS NCV
SEL
A
com
INPUT
12
Operational guidelines
4 .10 The line of fire test 4.10.1Press “SEL” key to switch to LIVE gear, the
instrument
shows LIVE 4.10.2Connect the red test pen to the Live input socket. Insert the
single pen into the power socket L jack or close to the live wire. If the
instrument detects ac voltage, it will display voltage signal strength
according to the detected signal strength, and the buzzer will give alarm
sound of different intensity. 4 .11Capacitance measurement 4.11.1Press the
“SEL” key to switch to capacitor gear. 4.11.2Use the other end of the test pen
to measure the capacitance value of the capacitor to be measured and read the
measured value from the LCD. Note: 1)When measuring large capacitors, a fixed
time is
required to stabilize readings. 2)When measuring polar capacitors, pay
attention to
the corresponding polarity to avoid damaging the instrument. 4 .12Temperature
measurement Press “SEL” to switch TEMP files, thermocouple red plug into the
end of °C /°F, black plug into the COM jack. When the reading is stable, the
temperature can be read directly from the display screen. Note: The maximum
measurement temperature of type K thermocouple is 250°, instantaneous
measurement up to 300°C. 4.13Diode test 4.13.1 Insert the black stylus into
the COM jack and the red stylus into the ” ” jack. 4.13.2 Touch SEL key to
switch to diode position. 4.13.3 Connect the red stylus to the diode anode and
the black stylus to the diode cathode for testing.
13
Operational guidelines
4.13.4 Read on the LCD. Note: The meter displays an approximation of the
forward voltage drop of the diode. If the pen is connected inversely or the
pen is open, the LCD displays “OL”.
4 .14Line connection test
warning
AUTO
SEL
9999 Counts TRUE RMS NCV
SEL
A
com
INPUT
Danger of electric shock. When testing the circuit on and off, make sure the circuit is disconnected and the capacitor on the circuit is completely discharged
4.14.1Insert the black marker into
the COM jack and the red
marker into the ” ” jack. 4.14.2Touch the “SEL” key to switch
to the line connection test state.
AUTO
SEL
9999 Counts TRUE RMS NCV
SEL
4.14.3Connect the pen to both ends of the line for measurement.
A
com
INPUT
If the resistance of the measured
line is less than 50 , the buzzer inside the meter will sound.
Note:
If the meter is open or the resistance of the measured
line is greater than 999.9, “OL” is displayed.
14
5.Maintenance 5.1 Replace Battery
Maintenance
warning
When replacing the battery, remove the pen from the measuring circuit to avoid electric shock.
5.1.1 the ” ” symbol indicates that charging is needed.
5.2Replace the test pen warning
When replacing the stylus, you must replace the same or the same grade of the table. Pen meter must be in good condition, pen grade: 1000V 10A.
Note: If the insulation layer of the pen is damaged, such as
exposed wire of the wire, the pen must be replaced.
6.The attachment 1)Test the pen 2)Directions for use 3)thermocouple
Grade: 1000V 10A
A pair of A copy of the
a
15