NULINE BW10 Blanket Warming Cabinet Instruction Manual
- June 10, 2024
- NULINE
Table of Contents
NULINE BW10 Blanket Warming Cabinet
To ensure that the equipment operates at its optimal condition. Please read the manual prior to connecting to power.
Quality Policy
Every care has been taken to ensure that the cabinet has been manufactured to
Australian Standards and carries the T.G.A. approval number 257613
Installation
- Unpack and place cabinet in position
- Ensure that it is level
- Remove any plastic coating on the stainless panels, both on the interior and exterior of cabinet
- Place shelves in desired position
- Connect to power supply
- Allow unit to operate for 2 hours prior to placing products on shelves
- Control setting have already been set for Fluid 35℃, Blanket 60℃
- Unit can be left on permanently as power consumption is minimal.
Electrical Rating – 240 volts 50 hz
3.2 amps FW 5 – 2 amp
FW10 – 3 amp
FW21- 6 amp
BW5 – 2 amp
BW10- 3 amp
BW21 – 6 amp
- For Fluid Warming cabinets check with Fluid Supplier at what temperature they recommend product be stored at.
- Temperature settings can be changed to suit products, however do take care not to exceed manufacturer recommendation and this should only be carried out by an authorized person.
- Loading of Stock – ensure that you have left sufficient space between bottles or bags so that you can obtain even temperature throughout. A full cabinet of Fluid either in bottles or bags may take 8 – 10 hours to reach desired temperature.
- Loading of Stock into Blanket Warmer – ensure that blankets DO NOT touch or block air vents.
Control Panel
Once you have connected equipment to your power supply and you have turned on power switch, located next to the controller, control panel will illuminate and start heating.
Controller – is a microprocessor which will indicate both the cabinet temperature on the top of screen, the one below is set temperature.
When cabinet has reached heating temperature, cold indicator light will stop
glowing.
To change temperature settings and this should only be carried out by a person
with authority to do so – Push the set button, go up or down with key pad,
once you had reached new temperature settings, push set button and that will
lock in new set point.
Temperature alarm
Should the temperature exceed the set points by 2 ℃, above the alarm will
activate. Plus, it will cut out power to heating element.
If temperature falls below set point, indicator light will glow.
Warranty
Subject to the following conditions from date of purchase, the following Warranty:
- Functional parts found within the unit to be defection in material will be replaced free of charge within Warranty period of 24 months from Date of Purchase
- A decision of where parts are repaired or replaced is at the discretion of Nuline Refrigeration.
- Structural components are covered by 24 months Warranty except where equipment has not being used for normal operation.
- Glass is not covered under warranty
- Under no circumstances will Nuline Refrigeration be responsible for any loss of stock
- To the extent of the law, this Warranty shall not cover damaged, malfunction or failure resulting from accident, misuse, improper installation or modification, unauthorised replacement of parts or cleaned by harsh chemicals.
The warranty shall be void.
The TEMP parameter table
Symbol| Name| Setting Range| Description| Factory set
value
---|---|---|---|---
AL/ AL| Alarming setting| 0-Full Range 0.0-Full Range| When temperature
is beyond SP+AL, the ALM indicator turn on. The buzzer sounds and the
heating power turn off.
|
CL/ CL| Refrigeration control
setting
| 0-Full Range
0.0 Full Range
| When the temperature less than SP+COL, the refrigeration point is turn on to drive the
compressor
| Without this function
CT/ CT| Refrigeration control delay| (0 –3600)s| When measured value
reaches alarm value, the alarm replay will not output before it passes CT
time
| Without this function
P/ P| Proportional Band| 1-Full Range
1.0 Full Range
| Proportional control action, if P is larger, then
the gain of system is lower. It only use on the heating side
|
I/ I| Integral Time| 0-3600 seconds| Integrated time constant. 1 larger,
then the
integrated action is smaller
|
d /d| Derivative Time| 0-3600 seconds| Derivative time constant. D
larger, then the derivative action is also larger D can overcome
overshoot. 1=0 and D =0 is called half
proportional control
|
AR/ AR| Overshoot control (resetting)| ****
1-100%
| At on-off PID control, the value of Ar is a constant which equals 1.5-2 times of the ratio of on time to the on*off period in the equilibrium state. In the half proportional control, the Ar
equals coefficient/P
|
T /T| Heating period| 1-300 seconds| For the thyristor output it is about
1-3 seconds.
For those equipment that the superfluous power provided is comparatively larger, select larger T would decrease the stable error caused
by PID control
| ****
3
Pd/ Pb| Zero point adjust (intersection)| -100-100-100.0
-100.0
| When the zero error comparatively smaller and the full point error comparatively larger, to update this value should be needed. Ordinary
for pt100, updating this value is rarely needed.
|
Pk/ PK| Full point adjust (intercept)| -1000-1000
seconds
| When zero error comparatively larger and the full point error also comparatively larger, to update this value should be needed. PK=4000 x (setting value-actual value)/actual value. For
pt100 adjusting this value is needed at first time.
|
/FN| Fan control| 1–100| When PV is at the range of SV+2.0C, the fan motor
will run at lower speed, otherwise the fan
will run at a high speed
|
L/LK| Password
key
| 0-255| When LK=18 the parameters listed above then
the above parameters can be updated
|
Operating method to the intelligent temperature controller
2. Energization to stun up
-
PVwindow(red)
- measured value displayed
- all of prompting symbols are displayed in line with the instrument status.
-
SVwindow(green)
- set value is displayed.
- all types of prompting symbols are displayed in line with the instrument status.
-
Pilot lamp
- OUT heating output lamp (green) It shines when working for output.
- ALM deviation alarm output lamp (red) It shines when working for output.
( Note: COLD cooling function)
-
SET functional key revision of set value
- call out of.parameter symbols and confirmation of revised parameters.
-
shift key
- for shifting the set value or observing ruming fixed time
-
7. digital plus and minus keys
for adjusting parameters or entry into self-adjustment state
CAPACITY
| ****
BW5 FW5
| ****
BW FW10
| ****
BW21 FW21
---|---|---|---
Electrical Ratings
| ****
240V
50ZH
| ****
240V
50HZ
| ` 240V
50HZ
Elements
| ****
350 watts
| ****
550watts
| ****
2 x 600 watts
Control
| ****
Micro processor
| ****
Micro processor
| ****
Micro processor
Electrical connection
| ****
10 amp Power cord
| ****
10 amp Power cord
| ****
10 amp Power cord
Over Temperature Protects
| ****
Auto Reset
| ****
Auto Reset
| ****
Auto Reset
Cabinet Material
| ****
430 Stainless
| ****
430 Stainless
| ****
430 Stainless
Doors
| ****
Double glazed in aluminium frame
| ****
Double glazed in aluminium frame
| ****
Double glazed in aluminium frame
Preventative Maintenance
- Wipe out inside of unit and doors with warm water and detergent on weekly basis.
- Exterior of cabinet should be cleaned on regular basis with a stainless steel cleaner
- Electrical cord should be checked every 6 months and have regular Tag and Test undertaken.
Trouble Shooting Assistance
PROBLEM
| ****
CAUSE
| ****
READY
---|---|---
Display not on
| · Power point not on
· On/off switch on unit is on
Display board not on Faulty on/off switch
| Check to ensure if supply lead is plugged in
Faulty Board, call for assistance
Call for assistance
Display panel on but not heating to correct temperature
| ****
Fan Failure
| ****
Turn power off and call for assistance
Display panel on but not heating
| ****
Faulty element
| ****
Turn power off and call for assistance
Since 1985, Nuline Refrigeration has been manufacturing refrigerators and
freezer cabinets for the Australian market. Our specialised range of
biomedical cabinets are used extensively across the hospitals, medical
centres, pharmacies and laboratories.
Focusing on customer needs, we manufacture, import and customise the
biomedical cabinets from a range of selected global suppliers. Our factory is
based in Melbourne, Australia and we are proudly 100% Australian owned
business.
Each of our manufactured cabinets are custom built based on customer orders by
our craftsmen right here in Melbourne.
We have extensive capability to design, engineer and build a range of
refrigerated cabinets for you. Our standard product range includes, fridges,
freezers (up to—40 degrees), blanket warmer, mortuary refrigerators and
refrigerated incubators.
Please contact us for your medical equipment needs.
Nuline Refrigeration, 7 Longford Court, Springvale, Victoria, 3171
Ph: 03 9562 3355
Nuline Refrigeration reserves the right to change specifications without any
prior notice
Nuline Refrigeration, 7 Longford Court, Springvale, Victoria, 3171
Ph: 03 9562 3355
Email:
info@nulinerefrigeration.com.au
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>