HANNA instruments HI9810432 Wireless Refillable pH Tester for Cosmetic Creams Instruction Manual
- June 10, 2024
- HANNA instruments
Table of Contents
- HANNA instruments HI9810432 Wireless Refillable pH Tester for Cosmetic
- Product Information: Wireless Refillable pH Tester for Cosmetic Creams
- Product Usage Instructions
- PRELIMINARY EXAMINATION
- SPECIFICATIONS
- GENERAL DESCRIPTION & INTENDED USE
- FUNCTIONAL DESCRIPTION & LCD DISPLAY
- GENERAL OPERATIONS
- SETUP
- BLUETOOTH
- HANNA LAB APPLICATION
- CALIBRATION
- CARE & MAINTENANCE
- WARNING & ERROR MESSAGES
- ABBREVIATIONS
- ACCESSORIES
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
HANNA instruments HI9810432 Wireless Refillable pH Tester for Cosmetic
Creams
Product Information: Wireless Refillable pH Tester for Cosmetic Creams
The HI9810432 is a wireless refillable pH tester designed specifically for testing the pH of cosmetic creams. It comes with a built-in specialized electrode that provides accurate and reliable measurements. The device is manufactured by Hanna Instruments.
Specifications
- Measurement range: 0.00 to 14.00 pH
- Resolution: 0.01 pH
- Accuracy: ±0.02 pH
- Operating temperature: 0 to 50°C (32 to 122°F)
- Power supply: 1.5V AAA batteries (2)
- Dimensions: 163 x 40 x 26 mm (6.4 x 1.6 x 1 in.)
General Description & Intended Use
The HI9810432 has two operating modes: continuous measurement and single measurement. It features a specialized electrode that is designed to test the pH of cosmetic creams accurately. The device comes with a built-in Bluetooth module that allows for wireless communication with the Hanna Lab application.
Probe Features
- Refillable junction
- Polycarbonate body
- Specialized glass bulb
Functional Description & LCD Display
The device has a large LCD display that shows the pH reading, temperature, and battery level. It also displays the operating mode and Bluetooth status.
Product Usage Instructions
Preliminary Examination
Before using the HI9810432, make sure to examine the device and check for any damage or defects. If you notice any issues, do not use the device and contact the manufacturer for assistance.
General Operations
To turn on the tester, press and hold the power button for two seconds. To turn it off, press and hold the power button again for two seconds. To take a single measurement, press the measurement button once. To switch between continuous measurement and single measurement modes, press and hold the measurement button for two seconds.
Battery Replacement
If the battery level is low, replace the batteries with two new 1.5V AAA batteries. Make sure to follow the polarity markings on the battery compartment.
Bluetooth
The HI9810432 comes with a built-in Bluetooth module that allows for wireless communication with the Hanna Lab application. To use this feature, make sure that Bluetooth is enabled on your device and pair it with the HI9810432. Follow the instructions in the Hanna Lab application to tag a measurement and view the data.
Care & Maintenance
Refill the electrode as needed to maintain accuracy. Store the device in a cool, dry place when not in use. Do not immerse the device in water or any other liquid. Clean the electrode with a soft cloth or tissue after each use.
Warning & Error Messages
If the device displays a warning or error message, refer to the user manual for troubleshooting instructions.
Abbreviations
The user manual includes a list of abbreviations used throughout the document for reference.
Dear Customer,
Thank you for choosing a Hanna Instruments product.
Please read this instruction manual carefully before using this instrument.
This manual will provide you with the necessary information for correct use of
this instrument, as well as a precise idea of its versatility.
If you need additional technical information, do not hesitate to e-mail us at
tech@hannainst.com or view our contact list at
www.hannainst.com.
All rights are reserved. Reproduction in whole or in part is prohibited
without the written consent of the copyright owner, Hanna Instruments Inc.,
Woonsocket, Rhode Island, 02895, USA.
PRELIMINARY EXAMINATION
Remove the tester and accessories from packaging and examine them carefully.
For further assistance, please contact your local Hanna Instruments office or
email us at tech@hannainst.com.
Each HI9810432 is supplied with a starter kit consisting of:
- pH 4.01 buffer solution, 20 mL sachet (2 pcs.)
- pH 7.01 buffer solution, 20 mL sachet (2 pcs.)
- Electrode cleaning solution, 20 mL sachet (2 pcs.)
- Electrode storage solution, 13 mL dropper bottle (1 pc.)
- Electrolyte refill solution, 30 mL
- Pipette
- 3V Lithium battery – CR2032
- Instrument quality certificate and Instruction manual
Note: Save all packing material until you are sure that the tester works correctly. Any damaged or defective item must be returned in its original packing material with the supplied accessories.
SPECIFICATIONS
- Available with Hanna Lab App
- Measuring outside the recommended operating temperature range may damage the gel electrolyte and void product warranty.
Note: The tester can display measurements from –2.00 to 16.00 pH. Measurements outside of the pH range will flash. In this case assess the integrity of the tester and the type of measured sample.
GENERAL DESCRIPTION & INTENDED USE
HI9810432 is a professional wireless pH tester, part of Hanna Instruments HALO2 family.
- The integrated Bluetooth module allows the tester to be connected to a compatible smart device with the Hanna Lab App.
- The tester has a compact, waterproof casing, and automatic pH calibration at up to three points, or four points when used with the Hanna Lab App.
- Automatically compensated temperature readings are displayed on a large LCD.
- Accurate and easy to use, the tester is ideal for measurements in samples that would be a challenge for standard design pH electrodes.
Operating Modes
The HI9810432 can be used as a stand-alone pH tester or connected to the Hanna
Lab App.
The Hanna Lab App turns a compatible smart device into a full-featured pH
meter. Features include: electrode condition, GLP with time-stamp, live-
readings, mV resolution, manual temperature compensation, stability criteria,
calibration reminder, pH (mV) and temperature alarms, tester ID, and data
sharing.
Probe Features
The HI9810432 features a non-porous, glass body that is easy to clean and
withstands harsh chemicals.
The conical glass tip allows for penetration into semisolids and emulsions.
The triple ceramic junction allows a higher flow rate of electrolyte from the
reference cell into the solution. This high flow rate provides faster
electrode response and a more stable measurement in viscous solutions or
samples of low conductivity.
Built-in temperature sensor at the tip of the pH electrode allows for rapid
determination of the sample temperature and a high-accuracy temperature
reading.
FUNCTIONAL DESCRIPTION & LCD DISPLAY
Front & Back view
LCD display
- Calibration indicator
- Stability indicator
- Battery percentage indicator
- Battery icon
- Bluetooth icon
- Temperature compensation
- Second LCD line
- Temperature unit
- First LCD line
- Measurement unit
GENERAL OPERATIONS
Turning the Tester On & Off
- Turn the tester face down on a clean, dry surface. Rotate the battery cover on the back of the tester counterclockwise and set it aside. Remove the battery insulation film.
- Press the ON/OFF button to turn the tester on or off. Initialization screen displays all LCD segments, followed by the battery percentage. The tester enters measurement mode. Before the tester turns off, “OFF PWR” is displayed briefly.
- Press and hold the ON/OFF button to turn the tester off when connected to Bluetooth.
Battery Replacement
- Turn off the tester. Turn the tester face down and rotate the battery cover counterclockwise.
- Set the battery cover aside. Press the metallic pin to push old battery out.
- Place the new battery with positive (+) sign facing out.
- Reset the date and time in setup, or connect to the Hanna Lab App to update it automatically.
- Align the mark on the cover with the open lock ico on the case. Rotate the cover clockwise until the mark on the cover aligns with the closed lock icon .
Note: Only use specified battery type. Dispose of old battery in accordance with local regulations.
SETUP
The Setup button is located inside the battery compartment. After Setup configuration, replace the cover. Setup Menu Navigation
- Press the Setup button to enter Setup mode and navigate menu items.
- To exit Setup mode, press the Setup button after “SEt TIME” option is displayed.
- Press the ON/OFF button to configure menu item options.
Temperature Unit
Option: °C or °F
Press ON/OFF button to select desired temperature unit.
Auto-Off Interval
Option: 8, 60 min., or “—-” (disabled)
Press ON/OFF button to select desired interval.
To save battery life, after the selected auto-off interval has elapsed, the
tester will automatically turn off.
Calibration Points
Option: 2P or 3P
Press ON/OFF button to select between up to two- or up to three-point
calibration.
Buffer Set
Option: 7.01 pH (Hanna) or 6.86 pH (NIST) Press ON/OFF button to select the
calibration buffer set (Hanna or NIST).
pH Resolution
Option: 0.01 pH or 0.1 pH
Press the ON/OFF button to select resolution.
Bluetooth Wireless Mode
Option: On, PAIr, or OFF
Press ON/OFF button to select Bluetooth option at start-up.
Bluetooth Pairing
Option: dEL PAIr
Press ON/OFF button to delete saved paired device.
Date & Time
Option: SEt TIME
Press ON/OFF button to set the date and time.
Option: YEAR, MO, DAY, HOUR, and MIN
Use the Setup button to select the option and press the ON/OFF button to
change the selected option.
BLUETOOTH
With “PAIr BLU” or “On BLU” selected in Setup, the Bluetooth icon will blink for up to 45 seconds, indicating the tester is in discoverable mode. Once connected, the icon will stop blinking. If not connected, the icon is not displayed.
- Select “On BLU”, to enable Bluetooth without bonding.
- Select “PAIr BLU”, to enable Bluetooth with bonding. A 6-digit bonding pin is displayed the first time the tester and smart device are paired. Once the devices are paired, the bonding pin is not required when reconnected.
- Select “OFF BLU”, to disable Bluetooth.
- Select “dEL PAIR”, to delete all paired devices. If PAIr BLU is enabled, a bonding pin will need to be re-entered.
HANNA LAB APPLICATION
- The Hanna Lab App is available on the App Store® and on Google Play.1
- Consult the Help section of the application for information on calibration, measurement, data logging, and sharing.
- When the tester is in discoverable mode, it will appear in the list of “Available Devices”.
- Within the application, tap “Connect” to pair the tester with the device. All readings are transmitted directly to the application. Tag a
Measurement
Once connected to the application, the ON/OFF button can be pressed to tag the
current reading.
- Press the ON/OFF button from measurement mode. The display shows “SEt TAG”, followed by “- TAG”.
- The reading on the application will flash green and the note icon will be displayed. Tap the note icon to add an annotation.
CALIBRATION
Preparation & Guidelines
- Unscrew the top collar to vent the cap before removing the probe from the plastic storage tube. Save the cap for probe storage.
- Rinse off any storage solution or salts that may be on the body.
- Verify there is solution inside the pH bulb by shaking the probe down to restore continuity as the solution may have moved up the stem during shipping.
- For best results, use a rinse beaker and a separate calibration beaker for each buffer. Discard rinsing buffers after use. Procedure
For most applications it is recommended to start with pH 7.01 buffer (or pH 6.86).
To restore factory defaults, press and hold the ON/OFF button. “CAL CLR” is displayed.
Note: It is recommended to calibrate the electrode with buffers at the
temperature it will be used at.
App Store is a trademark of Apple, Inc.
Google Play and the Google Play logo are trademarks of Google LLC.
Stand-Alone Tester (Up to Three-Point Calibration)
-
Rinse the electrode tip with purified water and blot dry. Then rinse with the buffer being used for calibration.
-
Press and hold ON/OFF button until “CAL MODE” is displayed.
-
When “7.01 USE” or “6.86 USE” is displayed with “CAL” tag blinking, place the tip of the electrode in the correct buffer.
-
When the buffer has been recognized, “REC“ is displayed.
“WAIT” is displayed with the stability indicator blinking until the reading is stable. Wait until the measurement is stored and the stability indicator disappears. -
To save a one-point calibration and return to measurement mode, press the ON/OFF button.
“CAL SAVE” is displayed briefly. With 2P option selected -
Place the tip of the electrode in the second buffer rinse beaker, then in the second calibration buffer. Wait until the measurement is stored and the stability indicator disappears.
-
After the second point has been stored, “CAL SAVE” is displayed briefly and the tester will return to measurement mode automatically.
With 3P option selected -
Place the tip of the electrode in the second buffer rinse beaker, then in the second calibration buffer. Wait until the measurement is stored and the stability indicator disappears.
-
Place the tip of the electrode in the third buffer rinse beaker, then in the third calibration buffer. Wait until the measurement is stored and the stability indicator disappears.
-
After the third point has been stored, “CAL SAVE” is displayed briefly and the tester will return to measurement mode automatically.
With Hanna Lab App (Up to Four-Point Calibration)
Connect the tester to the Hanna Lab App and follow the calibration procedure.
See the App Help section for the calibration procedure.
CARE & MAINTENANCE
- Never immerse the tester over the maximum immersion level.
- Fresh buffers should be used for each calibration. Once the sachets are opened the buffer value can change over time.
- If the electrode is slow or sluggish, soak it in cleaning solution for 20 minutes. Rinse with water and hydrate the electrode in storage solution for a minimum of 30 minutes before calibrating.
- If measurements are taken successively, rinse the probe thoroughly in distilled or deionized water to eliminate cross-contamination between measurements.
Refilling the Electrode
- If the filling solution (electrolyte) is more than 1 cm (½”) below the fill hole, add HI7082 3.5M KCl electrolyte fill solution for double junction electrodes.
- Unscrew the fill hole cover during measurements so the liquid reference junction maintains an outward flow of electrolyte.
Storage
To ensure a quick response, the glass tip and junction should be kept
hydrated.
-
When not in use, add a few drops of storage solution to the storage cap. If storage solution is not available, pH 4.01 or pH 7.01 buffer can be used.
Note: Do not store the electrode in distilled or deionized water. -
Place the probe tip into the cap and gently tighten the collar.
WARNING & ERROR MESSAGES
- “—- WRNG” displayed during calibration. Invalid buffer. Check the buffer value and use fresh buffer.
- Measured value displayed blinking. Measured value is out of electrode range. Clean the electrode to improve condition.
- Battery icon displayed blinking. Battery is below 10 %, replace the battery.
- “BAtt DEAD” and tester turns off. Depleted battery, replace the battery.
ABBREVIATIONS
- ATC Automatic Temperature Compensation
- GLP Good Laboratory Practice
- MTC Manual Temperature Compensation
- NIST National Institute of Standards and Technology
ACCESSORIES
Ordering Information : Product Description
pH Buffer Solution
- HI50016-02 pH 1.68 buffer solution, 20 mL sachet (25 pcs.)
- HI70004P pH 4.01 buffer solution, 20 mL sachet (25 pcs.)
- HI70006P pH 6.86 buffer solution, 20 mL sachet (25 pcs.)
- HI70007P pH 7.01 buffer solution, 20 mL sachet (25 pcs.)
- HI70009P pH 9.18 buffer solution, 20 mL sachet (25 pcs.)
- HI70010P pH 10.01 buffer solution, 20 mL sachet (25 pcs.)
- HI77400P pH 4.01 & 7.01 buffer solution, 20 mL sachet (10 pcs., 5 each)
- HI770710P pH 10.01 & 7.01 buffer solution, 20 mL sachet (10 pcs., 5 each)
Electrode Cleaning Solution
- HI700601P Electrode cleaning solution for general use, 20 mL sachet (25 pcs.)
- HI7073L Protein cleaning solution, 500 mL
- HI7074L Inorganic cleaning solution, 500 mL
- HI7077L Oil & Fat cleaning solution, 500 mL
Electrode Storage Solution
- HI70300L Electrode storage solution, 500 mL
- HI70300M Electrode storage solution, 230 mL
- HI70300S Electrode storage solution, 30 mL dropper
- HI9072 Electrode storage solution, 13 mL dropper
- Electrode Fill Solution & Other Accessories
- HI7082 3.5M KCl electrolyte fill solution for double junction electrodes (4 x 30 mL)
- HI740155P Electrode refilling pipette (20 pcs.)
RECOMMENDATIONS FOR USERS
Before using this product, make sure it is entirely suitable for your specific
application and for the environment in which it is used. Any variation
introduced by the user to the supplied equipment may degrade the tester’s
performance. For your and the tester’s safety do not use or store it in
hazardous environments.
WARRANTY
HI9810432 is warranted for a period of one year against defects in workmanship
and materials when used for its intended purpose and maintained according to
instructions. This warranty is limited to repair or replacement free of
charge. Damage due to accidents, misuse, tampering, or lack of prescribed
maintenance is not covered. If service is required, contact your local Hanna
Instruments office. If under warranty, report the model number, date of
purchase, serial number, and the nature of the problem. If the repair is not
covered by the warranty, you will be notified of the charges incurred. If the
tester is to be returned to Hanna Instruments, first obtain a Returned Goods
Authorization (RGA) number from the Technical Service department and then send
it with shipping costs prepaid. When shipping any product, make sure it is
shipped back in compliance with shipping regulations, thoroughly cleaned and
free of any chemicals, and is properly packaged for complete protection.
MAN9810432 10/21
Hanna Instruments reserves the right to modify the design, construction or
appearance of its products without advance notice.
CERTIFICATION
All Hanna Instruments conform to the CE European Directives.
Disposal of Electrical and Electronic Equipment. The product should not be
treated as household waste.
Instead, hand it over to the appropriate collection point for the recycling of
electrical and electronic equipment, which will conserve natural resources.
Disposal of waste batteries. This product contains batteries, do not dispose
of them with other household waste. Hand them over to the appropriate
collection point for recycling. Ensuring proper product and battery disposal
prevents potential negative consequences for the environment and human health.
For more information, contact your city, your local household waste disposal
service, or the place of purchase.
Regulatory Notices for stand-alone, Bluetooth, low-energy modules
All modules have identical operation. All references to US FCC Rules and
Canadian RSS standards on device classification and operation, listed here
under BMD-300 Module, apply to all models noted here. Remove the battery cover
to check the installed module.
BMD-300 Module
United States (FCC) FCC ID: 2AA9B04
This device complies with FCC Rules, Part 15, Subpart C “Intentional
Radiators” and Subpart B, Chapter §15.105. This equipment has been tested and
found to comply with the limits for a Class A digital device, pursuant to part
15 of the FCC Rules. These limits are designed to provide reasonable
protection against harmful interference when the equipment is operated in a
commercial environment. This equipment generates, uses, and can radiate radio
frequency energy and, if not installed and used in accordance with the
instruction manual, may cause harmful interference to radio communications.
Operation of this equipment in a residential area is likely to cause harmful
interference in which case, users are required to correct the interference at
their own expense.
Canada (ISED) IC: 12208A-04
This device complies with Industry Canada license exempt RSS standard(s).
Operation is subject to the following two conditions:
- This device may not cause interference, anTd
- his device must accept any interference, including interference that may cause undesired operation of the device.
Australia / New Zealand (RCM) BMD-300 complies with the AS/NZS 4268:2017.
| Japan (MIC)| South Korea (KCC) R-CRM-Rgd-BMD-300
---|---|---
| Brazil (ANATEL): Contains ANATEL approved module # 00820-21-05903.| Mexico
(IFETEL): Este equipo contiene el módulo con IFT #: NYCE/CT/0146/17/TS.
| BMD-350 Module
| United States (FCC) FCC ID: 2AA9B05| Canada (ISED) IC: 12208A-05
| Japan (MIC)| Australia / New Zealand (RCM) BMD-350 complies with the AS/NZS
4268:2017
| Eurasia (EAC) EAЭC N RU Д-US.HA27.B.00650/18| Brazil (ANATEL)
Contains ANATEL approved module # 00857-21-05903
| China (SRRC) CMIIT ID: 2018DJ7255| Mexico (IFETEL) Este equipo contiene el
módulo con IFT #: RCPRIBM18-1491
ANNA-B112 Module|
United States (FCC) FCC ID: XPYANNAB1| Canada (ISED) IC: 8595A-ANNAB1|
Contains Transmitter Module
Taiwan (NCC) CCAI18LP2200T2| South Korea (KCC) R-C-ULX-ANNA-B112|
South Africa (ICASA) ICASA TA-2019/1203 Approved| China (SRRC) CMIIT ID:
2021DJ6698|
Australia / New Zealand (ACMA) ANNA-B1 complies with AS/NZS 4268:2012
standard|
The module complies with the Japanese Technical Regulation Conformity
Certification of Specified Japan (MIC) 5 Radio Equipment (ordinance of MPT
N°. 37, 1981), Article 2, Paragraph 1, Item 19 “2.4 GHz band wide band low
power data communication system”.|
Brazil (ANATEL) This equipment operates on a secondary basis and,
consequently, must accept harmful interference, including from stations of the
same kind, and may not cause harmful interference to systems operating on a
primary basis.|
World
Headquarters
Hanna Instruments Inc.
Highland Industrial Park
584 Park East Drive
Woonsocket, RI 02895 USA
www.hannainst.com
MAN9810432
Printed in ROMANIA
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>