EXCELITAS TECHNOLOGIES X-Cite 110LED Microscope LED Fluorescence Light Source User Guide

June 9, 2024
EXCELITAS TECHNOLOGIES

EXCELITAS-TECHNOLOGIES-X-Cite-110LED-Microscope-LED-Fluorescence-Light-
Source-logo

EXCELITAS TECHNOLOGIES X-Cite 110LED Microscope LED Fluorescence Light

Source

EXCELITAS-TECHNOLOGIES-X-Cite-110LED-Microscope-LED-Fluorescence-Light-
Source-PRODUCT-IMAGE

X-Cite® 110LED: Quick Start Guide
Warning – Refer to Safety Precautions Booklet 035-00636R for all safety precautions prior to use. User Guide 035-00575R download: http://www.excelitas.com

SETUP

Unpack

  1. Carefully unpack the unit and accessories from shipping carton.
  2. When unpacking the light guide take care not to bend sharply.

Install the Microscope Adaptor (if applicable)

  1. Remove protective wrap from adaptor.
  2. Insert the flange portion of the Microscope Adaptor into the light port on the microscope, and secure it using the hardware provided on the microscope.

Connections and Positioning

  1. Remove red caps from light guide. Insert large end into light guide port on unit (up to “B” line). Insert small end into microscope light guide port (or adaptor), and secure it with hardware provided.
  2. Position the equipment. Ensure adequate clearance for ventilation around unit, and slack in the light guide.
  3. Connect speed DIAL cable to “remote” port on main unit.
  4. Connect USB cable to ports on main unit and computer. (For Windows 7/8.1, USB driver will install via internet after unit powers up, see user guide for additional information.)
  5. Connect external power supply to main unit.
  6. Connect AC power to properly grounded outlet.

MANUAL OPERATION

  1. Flip rocker switch to turn unit on.
  2. When display shows “x%”, it is ready to use.
  3. Click speed DIAL to illuminate a specimen.
  4. Turn speed DIAL to adjust intensity.

REGULATORY (refer to user guide for complete details)

Product Safety and Electromagnetic Compatibility
The X-Cite 110LED has been tested and found to comply with product safety and electromagnetic compatibility requirements. For a complete list of tests and for certification details, please contact your X-Cite representative or visit
http://www.excelitas.com

CE Marking

  • Council Directive 2014/35/EU
  • Council Directive 2014/30/EU
  • Council Directive 2012/19/EU
  • Council Directive 2011/65/EU as amended by (EU) 2015/863

Low Voltage Directive
EMC Directive
WEEE Directive
RoHS

EXCELITAS-TECHNOLOGIES-X-Cite-110LED-Microscope-LED-Fluorescence-Light-
Source-02

BASIC TROUBLESHOOTING

(refer to user guide for complete details)

Category Symptom Action
Error messages Error 1 (High Temp.) Turn off, let cool; check

clearance around vents
Error 3 or 5 (Internal)| Turn off, restart
Error 6 (Low Temp.)| Allow unit to warm to room temperature and restart
Intensity| Low output intensity| Check speed DIAL settings Check light guide position Check adaptor type
Check microscope beam path, filters, apertures, shutters, objectives
speed DIAL| Double-click turns off LED| Increase delay in the speed (Spd) setting


New settings forgotten when unit is shut down

| After changing settings, wait at least five minutes before shutting unit down
Dial does not scroll through the menu| Turn dial in the opposite direction
“Lock” icon on display, speedDIAL not responding to manual control| Send “unlock” command via computer OR power down unit and restart
TTL| **** TTL response time is longer than usual| Enable TTL mode
Verify that timeout settings are appropriately set

EXCELITAS-TECHNOLOGIES-X-Cite-110LED-Microscope-LED-Fluorescence-Light-
Source-03WARNING: Never look at operating LED! The light could severely damage the cornea and retina of the eye if the light is observed directly. Eye shielding must be used at all times as well as clothing to protect exposed skin Refer to user guide for more details.

SpeedDIAL Home Screen

EXCELITAS-TECHNOLOGIES-X-Cite-110LED-Microscope-LED-Fluorescence-Light-
Source-04

SpeedDIAL Menus and Settings

  • To access the main menu, press and hold the dial (shutter button) for one (1) second.
  • To navigate the menus, turn the dial to scroll through the options. An arrowhead will indicate the currently selected menu option, click the dial to make a selection.
  • To adjust settings, turn the dial. To select a setting and return to sub-menu, click the dial.
  • To exit menu system at any time, press and hold the dial for one (1) second.

SpeedDIAL Menu
(refer to user guide for complete instructions)

Menu Item Description
LCD Display Screen Brightness and Color Settings
Dim Backlight timeout (off = backlight always on)
Brig Screen brightness (0% = backlight off)
Color Screen color
Favo Favorite Intensity Setting

– Double-click jumps to preset intensity level
Int| Select favorite intensity level
Enbl| Enable or disable this feature
Spd| Adjust delay time between clicks
(defines “double-click”)
TTL| TTL Mode Control
Required for TTL triggering
Enbl| Enable or disable this mode
Time| Define a timeout setting (recommended)
Hand| Display Screen Orientation
– Rotate display screen 180°
Srvc| Service Data
– Embedded software versions, serial number, LED hours, system temperatures

Excelitas Canada Inc.
2260 Argentia Road Mississauga
ON L5N 6H7
Tel: 1-905-821-2600
Toll Free: 1-800-668-8752
www.excelitas.com/x-cite

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

EXCELITAS TECHNOLOGIES User Manuals

Related Manuals