the best grow LGBQM2 Advanced Quantum PAR Meter User Manual

October 29, 2023
the best grow

Advanced Quantum PAR Meterthe best grow LGBQM2 Advanced Quantum PAR
Meter

OVERVIEW

The PHOTOBIO by Phantom Quantum PAR meter is designed to measure PAR (Photosynthetically Active Radiation) flux in wavelengths ranging from 400 to 700nm.  There is a proportional relationship between the number of photons absorbed in 400 to 700nm band and the rate of photosynthesis in plants, which is important for  horticultural studies and monitoring plant physiology.
IMPORTANT: To ensure safety, please read this manual carefully before installation and follow the instructions herein. Store this manual in a secure place for future  reference.

PARTS LIST (What’s in the Box)

the best grow LGBQM2 Advanced Quantum PAR Meter -
Meter

GETTING STARTED

  1. Connect the sensor unit to PAR meter main unit with coiled cable or straight cable.
  2. Open the battery cover.
  3. Insert two AAA batteries into the battery compartment.
  4. Close the battery cover.

the best grow LGBQM2 Advanced Quantum PAR Meter -
battery

LCD screen will be on and the Quantum PAR Meter will begin operating.
There is no need for initial setup or calibration.

OPERATING INSTRUCTIONS

  1. Power On/Off
    Press the Power button to turn on the Quantum PAR meter.
    Press and hold the Power button for 3 seconds to turn off the Quantum PAR meter.

  2. Log function
    Press▼ to record the measurement data.

  3. Hold function
    Press ENTER to hold the measurement value.
    Press ENTER again to exit hold function.

  4. Mode Functions
    Press the MODE button to cycle through the Mode functions (see table below).
    Press ENTER to enter each option menu.

Mode Directions
CALI To calibrate the meter, enter CALI mode (CALI icon will be flashing),

then press ENTER.
Completely cover the photosensitive portion of the probe, and hold ENTER for 3 seconds to reset the zero point.
LOG| To display your data history, enter LOG mode (LOG icon will be flashing). Press ENTER.
Use p and q to cycle through data records. Press ENTER to exit this mode.
RCFS| To reset the device to factory settings, enter RcFS mode (RcFS icon will be flashing).
Press ENTER  ( “NO” will be flashing on LCD screen). Use ▲or ▼to select “YES,” and then hold  ENTER for 3 seconds. An audible beep will indicate when the device has been reset.   If you hold  ENTER for 3 seconds while “NO” is flashing, the device will not reset to factory settings.  Important:  This option erases all stored log data.

SPECIFICATIONS

SPECIFICATIONS
Typical test conditions, unless otherwise specified: Ambient Temp =23+/-3°C, RH=50%–70%, Altitude=0~100 meter

Measurement Specification
Operating Temperature 32°F to 122°F (0°C to 50°C)
Storage Temperature -4°F to 140°F (-20°C to 60°C)
Operating & Storage RH 0-95%, non-condensing

PPFD Measurement I
Repeatability| ±1 pmoVm2/s
Measurement Range| 0- 3,999 pmoVm2/s
Display Resolution| 0.1 pmoVm2/s (0-999);1 pmol/m2/s (1,000-3,999)
Cut-On Wavelength| 400±10nm
Cut-Off Wavelength| 700±10nm
Power Requirements| 2 x AAA batteries
Dimension| Main Unit: 115 x 60 x 24mm Sensor Unit 80 x 45.6 x 26mm
Weight| 115g (without batteries and cable)

TROUBLESHOOTING

Symptom Possible Cause Solution
No display on LCD screen No power Press the Power button to turn on PAR

meter.
Bad Batteries| Replace battery.
Cable error| Check the cable. Maybe there is an electrical short. Replace the sensor cable if needed.
LCD main display shows “–“| Sensor unit is not working| Check sensor cable to make sure it is securely connected to the PAR meter main body and sensor unit.
Sensor cable is not connected
Some foreign materials are in the R.111 port| Please check the RJ11 port and clean any foreign material that could cause interference.
Battery icon is flashing on the LCD screen| Battery is running low| Replace the batteries.
Reset to zero point failed| The light sensitive
portion of the probe is not completely covered during calibration| Completely cover the photosensitive portion of the probe before calibrating.

WARRANTY

LIMITED WARRANTY: Hydrofarm warrants the LGBQM2 to be free from defects in materials and workmanship. The warranty term is for 1 year beginning on the date of  purchase. Misuse, abuse, or failure to follow instructions is not covered under this warranty. Hydrofarm’s warranty liability extends only to the replacement cost of the  product. Hydrofarm will not be liable for any consequential, indirect, or incidental damages of any kind, including lost revenues, lost profits, or other losses in connection  with the product. Some states do not allow limitation on how long an implied warranty lasts or the exclusion of incidental or consequential damages, so the above limitations  or exclusions may not apply to you. Hydrofarm will, at our discretion, repair or replace the LGBQM2 covered under this warranty if it is returned to the original place of  purchase. To request warranty service, please return the LGBQM2, with original sales receipt and original packaging, to your place of purchase. The purchase date is based on your original sales receipt.

WARNING – POSSIBLE RISK OF INJURY TO EYES AND SKIN: Hazardous optical UV, HEV, and IR radiation may be emitted from the light source. Always wear personal protective equipment ensuring complete shielding of skin and eyes. Avoid prolonged exposure and looking directly at light source.