Ralston Instruments QTVC Volume Controller
Volume Controller (QTVC) Operation
For all models of QTVC Volume Controllers
Specifications
-
Pressure Range: 0 to 3,000 psi (0 to 210 bar)
-
Vacuum Range: 0 to 10 inHg (0 to 260 mmHg)
-
Temperature Range: 0 to 130 °F (-18 to 54 °C)
-
Construction: Anodized Aluminum, Brass, Plated Steel, Stainless Steel
-
Seal Materials: Buna-N, Delrin, Teflon
-
Pressure Media: Fine Adjust Resolution ±0.0005 PSI (0.03 mbar)
-
Inlet Port: Male Ralston Quick-test™, brass
-
Outlet Port A
Male Ralston Quick-test™ with cap and chain, brass -
Outlet Port B: Male Ralston Quick-test™, brass
-
Outlet Port C: Male Ralston Quick-test™ with cap and chain, brass
-
Weight: 5.38 lb (2.4 kg)
-
Dimensions
W: 8.5 in (21.59 cm)
H: 6.16 in (15.65 cm)
D: 7.38 in (18.75 cm) -
Fill and Vent Valves: Soft seated construction
-
Mechanical Rotation: 42 turns (pressure balanced)
Requirements
What you need to use your Volume Controller:
- Wrenches
- Thread Tape
- Ralston Quick-test™ Adapters
- Device Under Test
- Ralston Quick-test™ Hoses
- Pressure Reference
- Pressure Source
Important Safety Notices
WARNING: Do not attempt to operate this product until you have read and fully understand the instructions and hazards of the product.
- Any modifications to this product with custom parts can result in hazardous operation of the product.
- Use eye protection while using this product. Leaking gas, parts or hoses can be ejected at high speed and may cause injury.
Volume Controller Overview
- A. Outlet Port A
- B. Outlet Port B
- C. Outlet Port C
- 1. Fill Valve
- 2. Fine Adjust Valve
- 3. Balance Valve
- 4. Vent Valve
- 5. Removable Front Panel
- 6. Carry Handle
- 7. Inlet Port
- 8. Stand
Setting Up
Connecting Reference Gauge
Male NPT Reference Gauge
-
Reference Gauge with
NPT male connection -
NPT Female Ralston
Quick-test™ Gauge Adapter -
Ralston Quick-test™ Hose
-
NPT Female Ralston
Quick-test™ Adapter
Male BSPP Reference Gauge
-
Reference Gauge with
BSPP male connection -
BSPP Washer
-
BSPP Female Ralston
Quick-test™ Adapter -
Ralston Quick-test™ Hose
-
BSPP Female (RG)
Ralston Quick-test™
Adapter
Female NPT Pressure Reference Gauge
-
Reference Gauge with
NPT female port -
NPT Male Ralston Quicktest
™ Gauge Adapter -
Ralston Quick-test™ Hose
-
NPT Male Ralston
Quick-test™ Adapter
Connecting Device Under Test (DUT) and Pressure Source
- Device under test (DUT)
- Ralston Quick-test™ Adapters
- Ralston Quick-test™ Hoses
- Pressure source
Calibration
Prepare the Volume Controller
Close Fill Valve.
Close Vent Valve
Set Fine Adjust Valve to 50% of travel.
Pull Balance Valve out.
Increase Pressure
Slowly open Fill Valve to just below first test point.
Close Fill Valve.
Push Balance Valve in to close.
Use Fine Adjust Valve to put reference gauge on exact test point.
To Continue Moving Up-scale in Pressure
Pull Balance Valve out to open.
Slowly open Fill Valve to just below next test point.
Close Fill Valve.
Push Balance Valve in to close.
Fine-adjust to exact test point.
Repeat for each test point up-scale until range is complete.
To Move Down-scale in Pressure
Pull Balance Valve out to open.
Slowly open Vent Valve to just above next test point.
Close Vent Valve.
Push Balance Valve in to close.
Fine-adjust to exact test point.
Venting System
Pull Balance Valve out.
Open the Vent Valve.
Storage and Transport
Disconnect hoses and pressure reference, and store everything.
Maintenance
Maintenance Interval
Every 300 uses or 3 months
Maintenance Procedure
- Lubricate the Ralston Quick-test™ fittings by squirting 2 ml of oil inside the connection.
- Lubricate the balance valve O-rings with silicone lubricant.
Troubleshooting
There is a drop in system pressure when the Volume Controller has been
pressurized and the Fill Valve is closed
If there is a drop in system pressure when the Volume Controller has been
pressurized and the Fill Valve is closed, then there is a leak.
Follow these instructions to locate and repair the leak:
- Connect the Volume Controller to a Device Under Test (DUT) and connect a Ralston Quick-test™ hose to the Inlet Port.
- Make sure the process connections are assembled wrench-tight.
- Close Vent Valve.
- Open Balance and Fill Valves.
- Apply pressure to unit.
- Close Fill Valve.
- Spray soapy water or leak detection fluid where leaks are suspected or immerse the Volume Controller in water. Be careful not to immerse the pressure gauge or calibrator.
- Observe where the bubbles are coming from to determine where there is a leak.
- Remove the leaking part and remove the O-ring.
- Clean and lubricate the O-ring, and backup ring if applicable.
- Replace the O-ring, and backup ring if applicable.
- Reassemble.
Fine Adjust Valve is difficult to operate
If the Fine Adjust Valve is difficult to operate over years of service, then
the inside walls of the piston need grease.
- Remove the Fine Adjust Valve.
- Apply a thin coat of graphite grease, such as Dow Corning® Moly-kote G-n Metal Assembly Paste (or equivalent) to the inside walls of the piston.
- Reassemble.
The Volume Controller does not adjust pressure
If the Volume Controller does not adjust pressure, then the O-rings in the
Balance Valve and/or Fine Adjust Valve need to be cleaned and lubricated.
- Remove the Balance Valve assembly from the front of the panel.
- Clean and lubricate the O-ring.
- Replace the O-ring.
- Reassemble.
- If the Volume Controller still does not adjust pressure, then remove the Fine Adjust Piston.
- Clean and lubricate the O-ring and backup ring.
- Reassemble.
The Balance Valve gets stuck in the closed position and cannot be opened
If the Balance Valve gets stuck in the closed position and cannot be opened,
then there is gas trapped in the top of the fine adjust piston, because the
Volume Controller was vented with the Balance Valve in the closed position.
- Open the Vent Valve 4-5 turns until you hear gas escaping from the top of the fine adjust piston. It will take several turns as there is a secondary seal in the Vent Valve that must be opened.
If the issue was not resolved by these troubleshooting instructions, then please contact support listed on page 38.
Volume Controller (QTVC) Operation Manual
For all models of QTVC Volume Controllers