UNI-T UT219P AC Digital Power Clamp Meter User Manual
- June 9, 2024
- UNI-T
Table of Contents
UT219P AC Digital Power Clamp Meter
P/N:110401111507X
UT219P
AC Digital Power Clamp Meter User Manual
EN
Preface
Thank you for purchasing this brand new product. In order to use this product
safely and correctly, please read this manual thoroughly, especially the
safety notes. After reading this manual, it is recommended to keep the manual
at an easily accessible place, preferably close to the device, for future
reference.
Limited Warranty and Liability
Uni-Trend guarantees that the product is free from any defect in material and
workmanship within one year from the purchase date. This warranty does not
apply to damages caused by accident, negligence, misuse, modification,
contamination or improper handling. The dealer shall not be entitled to give
any other warranty on behalf of Uni-Trend. If you need warranty service within
the warranty period, please contact your seller directly. Uni-Trend will not
be responsible for any special, indirect, incidental or subsequent damage or
loss caused by using this device.
2
UT219P User Manual
Contents
I. Overview ————————————————————————————- 4 II. Features
———————————————————————————— 4 III. Accessories —————————————————————————— 5
IV. Safety Information ——————————————————————— 5 V. Electrical Symbols
———————————————————————- 6 VI. General Specifications —————————————————————- 6
VII. External Structure and Test Leads ————————————————- 7 VIII. LCD Display
——————————————————————————- 8 IX. Rotary Switch —————————————————————————- 9
X. Button Descriptions ——————————————————————— 9 XI. Operating Instructions
—————————————————————– 9 XII. Technical Specifications ———————————————————— 27
XIII. Bluetooth Software —————————————————————— 31 XIV. Maintenance
————————————————————————– 32
3
UT219P User Manual
I. Overview
UT219P is a digital power clamp with fundamental frequency up to 1 kHz. This
clamp meter can be used to measure AC voltage, AC current, active power,
apparent power, reactive power, power factor, phase angle, power quality,
frequency, phase sequence, total harmonic distortion, harmonic component, etc.
It has different measurement modes including single phase, threephase three-
wire, and three-phase four-wire. Designed with Bluetooth function, measurement
data can be monitored and recorded via mobile phone and tablet, and
professional report can be generated. Featuring easy operation, stable
performance, accurate measurement and more, UT219P can be applied in multiple
fields such as: 1) Measure power quality of UPS to evaluate if operation is
normal. 2) Measure power factor of power system to analyze capacitive or
inductive load. 3) Measure harmonic of power grid of power system, to find out
harmonic issue that may
damage the load of nonlinear equipment of power grid. 4) Troubleshoot 400Hz AC
power supply system of aviation and 25Hz AC power supply system
of railway. 5) Test and maintain high-power motor, centrifugal fan,
centrifugal water pump, wind power
generation, solar power generation. 6) Test new energy vehicles and hybrid
electric vehicles, evaluate motor efficiency, invertor
efficiency and energy consumption of driving.
II. Features
- Completely sealed (Ip54). 2) Wide jaw opening (80mm) enables measuring current of thick conductor (Max. 75mm). 3) Category rating: CAT IV 600V AC, CAT III 1000V AC. 4) 1000V AC RMS voltage measurement, 15Hz~1000Hz frequency response range. 5) 1000A AC RMS current measurement, 15Hz~1000Hz frequency response range. 6) Ability to measure and display RMS and peak simultaneously.
- Multiple wire connecting modes including single-phase two-wire, single-
phase three-phase,
balanced three-phase wire-wire, unbalanced three-phase three-wire, balanced three-phase four-wire, unbalanced three-phase four-wire, and more. Simple guidelines for wire connecting places can be shown on the upper and lower displays when performing three-phase connection. 8) Wide fundamental frequency range for power measurement: 15Hz~1000Hz. 9) Ability to analyze 40th voltage and current harmonics. THD %f (total harmonic to fundamental wave distortion), THD %r (total harmonic to full wave) and RMS of harmonic component of each order. 10) Identify if electricity theft occurs by comparing measurement of single-phase active electric
energy of electric energy meter. 11) Detect phase sequence of motor to prevent inversed phase sequence of power supply
from causing motor to rotate inversely. 12) Via Bluetooth APP, the product can record measurement data and generate graph, table,
and report. The product has multiple functions including real-time monitoring for rapid troubleshooting, data recording and processing, integrated management, and more, so as to achieve integrated monitoring, maintenance and management.
4
UT219P User Manual
III. Accessories
Check the accessories in the package, if any of them is missing or damaged,
please contact your supplier immediately. User manual ————————————————– 1 pc
Test leads (red, black) ————————————- 1 pair Alligator clips (red, black)
——————————— 1 pair 1.5V AAA battery ——————————————– 3 pcs Tool box
——————————————————- 1 pc Brown box —————————————————- 1 pc AAA Ni-MH
rechargeable battery ———————– 3 pcs (optional) Test probes (red, black)
———————————– 1 pair
IV. Safety Information
Note the “Warning labels and sentences”. A Warning identifies conditions and
procedures that are dangerous to the user and that can cause damage to the
Product or the equipment under test. The product is designed in accordance
with IEC/EN61010-1, 61010-2-032, Electromagnetic Radiation EN61326-1 Standard,
and conforms to Double Insulation, Overvoltage CAT III 1000V, CAT IV 600V and
pollution degree 2. Failure to follow operating instructions can impair the
protection provided by the product. 1. Check the product and test leads before
use. Pay attention to any damage or abnormal
situation. Please stop use if test lead and casing insulation are damaged, or
the LCD displays nothing, or the product cannot work normally. 2. Before each
use verify tester operation by measuring a known voltage. 3. It is forbidden
to use without rear cover or battery cover set in place. Otherwise it may
cause electric shock. 4. Keep fingers behind the finger guard and never make
contact with exposed wire, connecto,r input terminal not in use, or circuit
being measured during measurement. 5. Do not exert AC voltage over 1000V
between terminal and grounding to prevent electric shock and product damage.
6. Use caution when working with voltages over 60V DC or 30Vrms AC. 7. Never
use the tester on a circuit with frequency that exceed the rating of this
tester. 8. To meet the requirements of safety standard, please use the product
together with equipped test leads and alligator clips. If test leads or
alligator clips are damaged, replace a new one which should meet IEC61010-031
standard, rated follow parameters of the product or better. 9. When the symbol
” ” appears on the LCD, please replace the battery in time to ensure
measurement accuracy. Remove the battery if the product is not used for a long
time. 10. Do not alter the internal wiring to avoid product damage and safety
hazard. 11. Do not keep or use the product in environments with high
temperature, high humidity, strong
electromagnetic field, or inflammable and explosive environments. 12. Please
wipe the casing with soft cloth and neutral cleaning agent, do not use
abrasives
or solvents, so as to avoid casing corrosion, product damage, and safety
hazard.
5
UT219P User Manual
V. Electrical Symbols
System Description
Application around and removal from UNINSULATED HAZARDOUS LIVE conductors is permitted
AC (Alternating Current)
Bluetooth communication
Double insulated Grounding
Warning
Conform to European Union standards
Conform to UL STD 61010-1, 61010-2-032 Certified to CSA STD C22.2 NO. 61010-1, 61010-2-032
CAT III
UKCA certification mark
Applicable to test and measure the circuit connected with the power
distribution part of building’s low-voltage MAINS installation.
CAT IV
Applicable to test and measure the circuit connected with the power supply of building’s low-voltage MAINS installation.
The product conforms to the identification requirement of WEEE directive. Do not place equipment and its accessories in the trash.
VI. General Specifications
- Overload protection voltage between input terminal and grounding: 1000V 2)
Ingress protection: Ip54 3) Polarity indication: Auto 4) Overload indication:
“OL” or “-OL” 5) Error caused by testing area: An additional error of ± 1.0%
of reading can occur when the
source to be measured is not placed at the center of the clamp jaws for current measurement. 6) Drop proof: 1 m 7) Maximum jaw opening: 80mm in diameter 8) Maximum diameter of measured conductor: 75 mm 9) Power supply:3 X 1.5V AAA batteries (type: LR03) or 3 X 1.2V rechargeable Ni-MH batteries 10) Auto power off: The product powers off automatically after 15 minutes of inactivity. 11) Dimensions: 295 mm118mm50mm 12) Weight: About 630g (including battery) 13) Operating altitude: 2000 m 14) Intended use: Indoor user 15) Operating temperature and humidity: 0°C~30°C (80%RH); 30°C~40°C (75%RH);
40°C~50°C (45%RH) 16) Storage temperature and humidity: -20°C~+60°C (80%RH) - EMC: Under radio frequency field of 1V/m, overall accuracy = specified
accuracy + 5%
of range. Under radio frequency field over 1V/m, there is no specified specification.
6
VII. External Structure and Test Leads (Fig. 1)
UT219P User Manual
1
2
3
4
5
6
8
5 9
7
10
11
12
CAT1000V CAT600V
CAT1000V CAT600V
Fig. 1
1. Clamp jaw: A sensing device, used to measure AC current, can convert
current to voltage. 2. Finger guard: Designed to protect user’s finger from
touching danger zone. 3. Trigger: Press to open jaws; release to close. 4.
Rotary switch: Select function positon 5. Functional buttons: Select basic
functions 6. LCD display: Display measurement data and function symbols 7.
Input terminals: Input voltage signal 8. Battery cover 9. Screws 10. Alligator
clips: Conform to CAT III 1000V, CAT IV 600V 11. Test leads: Conform to CAT
III 1000V, CAT IV 600V 12. Test probes
7
UT219P User Manual
VIII. LCD Display (Fig. 2)
Fig. 2
-
: Bluetooth communication symbol
-
RANGE: MANUAL: Manual current measurement mode
-
PF: Power factor
-
3P3W: 3-phase 3-wire measurement mode
-
3P4W: 3-phase 4-wire measurement mode
-
UNBALANCED: Unbalanced measurement mode
-
MAX: Maximum value measurement
-
MIN: Minimum value measurement
-
AVG: Average value measurement
-
THD %f: Total harmonic to fundamental distortion
-
THD %r: Total harmonic to full wave distortion
-
: Flash symbol to indicate pressing HOLD button
-
HOLD: Data hold, or enter next operation with finger symbol
-
MEASURED: Actual measured accumulative electric energy
-
EXPECTED: Theoretical accumulative electric energy
-
: Dangerous voltage
-
h : m: hour/minute
-
m : s: minute/second
-
cyc: Cyclic accumulative electric energy
-
PEAK: Peak measurement
-
V/A: Unit of voltage/current
-
Hz: Unit of frequency
-
kW: Unit of active power
-
kVA: Unit of apparent power
-
kVAr: Unit of reactive power
-
Wh, kWh: Unit of electric energy
-
: Auto power off
-
Top right corner:
Auxiliary display 1
- Bottom left corner:
A uxiliary display 2
- Bottom right corner:
Auxiliary display 3
: Battery voltage
8
UT219P User Manual
IX. Rotary Switch (Fig. 3)
Wh
Phase Detect
HARM
3P
W W
A OFF
Fig. 3
Position OFF A/V W/VA/var 3PW HARM Wh/Setting Phase Detect
Description Turn off the internal power supply of the product AC current/voltage measurement Single-phase power measurement Three-phase power measurement Harmonic measurement Electric energy/electric energy accumulation mode setting Phase sequence detection
X. Button Descriptions
1. SELECT
(1) A~/V~ position: Short press this button to cyclically select ACA ACV on
main display,
the default position is ACA. (2) W/VA/var position: Short press this button to
cyclically select active power apparent
power reactive power power factor phase angel on main display, the default
position is active power. (3) 3PW position: Under 3-phase 3-wire (3P3W,
balanced) or 3-phase 4-wire (3P4W) mode, short press this button to cyclically
select active power apparent power reactive power power factor phase angel on
main display, the default position is active power. Under 3-phase 3-wire
(3P3W) mode, short press this button to cyclically select phase-combined
active power P1 active power P2 active power on main display, the default
display is phase-combined active power. Under 3-phase 4-wire (3P4W) mode,
short press this button to cyclically select phase-combined active power
phase-combined apparent power phase-combined reactive power phase-combined
phase angle P1 active power P2 active power PC active power on main display,
the default display is phase-combined active power. (4) HARM position: Short
press this button to cyclically select voltage harmonic analysis current
harmonic analysis, the default position is current harmonic analysis. (5)
Wh/Setting position: Short press this button to cyclically select single-
phases electric energy electric energy accumulation mode setting, the default
position is single-phase electric energy. (6) Long press this button to turn
on/off Bluetooth.
9
UT219P User Manual
2. MAX/MIN
Short press this button to cyclically select real-time value maximum value
minimum value average value on the main display, and to record the time of
capturing maximum
and minimum values. Long press this button to exit maximum/minimum value
measurement. The measurement is valid only under such modes including AC
voltage, AC current, singlephase power, balanced three-phase power, etc.
3. RANGE When this button is short pressed in auto range mode, the product
enters manual mode, the LCD shows “RANGE: MANUAL”, and the display is in
current range. Press this button continuously to select AC current ranges
cyclically. When this button is long pressed, the product exits manual range
and enters auto range, “RANGE: MANUAL” does not appear on the LCD.
4. HOLD/ Under measurement interface (except unbalanced three phase, electric
energy and phase sequence), short press this button to enter or exit data
hold. When entering data hold mode, the LCD shows “HOLD”; when exiting data
hold mode, the LCD does not show “HOLD”. Under the interface of switching wire
connection, three-phase power enters its corresponding wire-connection
measurement interface when this button is short pressed. Under electric energy
mode, short press this button to start or stop accumulative counting. Long
press this button to turn on/off backlight. The backlight turns off
automatically after it is turned on for one minute.
5. /3P3W For 3PW position, short press this button to select measurement mode
of 3P3W balanced load and unbalanced load. Under harmonic measurement mode,
short press to adjust down the order of harmonic.
6. /3P4W For 3PW position, short press this button to select measurement mode
of 3P4W balanced load and unbalanced load. Under harmonic measurement mode,
short press to adjust up the order of harmonic. Note: For the operating
instructions of button of setting electric energy accumulation mode, please
refer to the operating instructions of measurement in electric energy
position.
XI. Operating Instructions
Please check the three pieces of AAA batteries before measurement, if the
battery voltage is low, the symbol ” ” will appears on the LCD, then please
replace the battery in time. Note the symbol ” ” around the terminal, this
symbol warns that the measured voltage cannot exceed the specified voltage.
10
1. AC current measurement (Fig. 4)
L/R/A
UT219P User Manual
Power supply Load
N
AC current measurement
Fig. 4
- Set the rotary switch to “A~/V~”, press SELECT to select AC current measurement. Press the trigger to clamp the measured conductor and then release the trigger slowly until the clamp jaws closes completely. Make sure that the measured conductor centers at the clamp jaws, error will be produced if the measured conductor does not centers at the clamp jaws. The clamp can only measure a current conductor at a time, if it measures two or more current conductors at the same time, the reading will be incorrect.
- Read from the LCD the True RMS, peak and frequency of AC current, as shown below:
Power supply Load
Note: Do not release the trigger abruptly when pressing and holding it. Hall
sensor is sensitive to magnet, heat and mechanic stress, impact can cause
reading variation in a short time. To ensure measurement accuracy, please
place the measured conductor at the center of the clamp jaws. Error of ±1.0%
of reading will be produced if the measured conductor is not placed at the
center of the clamp jaws.
2. AC voltage measurement (Fig. 5)
Red
L/R/A
Black
N
AC voltage measurement
Fig. 5
11
UT219P User Manual
- Connect red test lead (connected with red alligator clip already) with “V” terminal; and black (connected with black alligator clip already) with “COM”.
- Set the rotary switch to “A~/V~”, press SELECT to select AC voltage measurement, and connect the alligator clips with power source or load to be measured in parallel.
- Read from the LCD the True RMS, peak and frequency of AC voltage, as shown below:
Note: Do not input voltage over 1000Vrms. It is possible to measure higher
voltage, but it may damage the clamp. Avoid electric shock when working with
high voltage. Disconnect test leads with the measured circuit after all
measurement operations are completed. The high voltage warning symbol ” ” will
appear on the LCD if the measured voltage is over 30V AC; the red backlight
flashes if the measured voltage is over 1000VAC.
3. W/VA/var (Fig. 6 & Fig. 7)
Red
L/R/A
Load
Power supply
Black
N
AC single-phase power measurement (1P2W) Fig. 6
Load
Power supply
Red
L1
Black
N
L2
Black Red
AC single-phase power measurement (1P3W) Fig. 7
- Connect red test lead (connected with red alligator clip already) with “V” terminal; and black (connected with black alligator clip already) with “COM”.
- Set the rotary switch to “W/VA/Var”, connect alligator clips with power
source and load to be measured, and clamp the conductor to be measured, the
current flows from up to down (up: front; down: bottom cover), then release
the trigger slowly until the clamp jaws close completely. Please make sure
that the measured conductor centers at the clamp jaws, error will be produced
12
Power supply Load
Power supply Load
UT219P User Manual
if the measured conductor is not placed at the center of the clamp jaws. The clamp can only measure a current conductor at a time, if it measures two or more current conductors at the same time, the reading will be incorrect. 3) Press SELECT to cyclically read active power, apparent power, reactive power, power factor, and phase angle, as shown below:
Short press
Short press
Short press
Main display: Active power Auxiliary display 1: Power factor Auxiliary display 2: AC voltage Auxiliary display 3: AC current
Main display: Apparent power Auxiliary display 1: Power factor Auxiliary display 2: AC voltage Auxiliary display 3: AC current
Main display: Reactive power Auxiliary display 1: Power factor Auxiliary display 2: AC voltage Auxiliary display 3: AC current
Main display: Power factor Auxiliary display 1: Active power Auxiliary display 2: AC voltage Auxiliary display 3: AC current
Short press
Short press
Main display: Phase angle Auxiliary display 1: Active power Auxiliary display
2: AC voltage Auxiliary display 3: AC current
Note: Phase angle is in reference to voltage. Do not input voltage over
1000Vrms. It is possible to measure higher voltage, but it may damage the
clamp. Avoid electric shock when working with high voltage. Disconnect test
leads with the measured circuit after all measurement operations are
completed. The high voltage warning symbol ” ” will appear on the LCD if the
measured voltage is over 30V AC; the red backlight flashes if the measured
voltage is over 1000VAC. Do not release the trigger abruptly when pressing and
holding it. Hall sensor is sensitive to magnet, heat and mechanic stress,
impact can cause reading variation in a short time. To ensure measurement
accuracy, please place the measured conductor at the center of the clamp jaws.
Error of ±1.0% of reading will be produced if the measured conductor is not
placed at the center of the clamp jaws. Incorrect wire connection will cause
negative active power and cause the red backlight to flash, please check if
the connections of voltage input terminal and clamp jaws are correct.
4. 3-phase power measurement
Wire connection interface: The places of clamp jaws, black test lead and red
test lead are shown in the figure below.
Clamp jaws
Black test lead
Red test lead
4.1 3-phase 3-wire (3P3W) balanced mode (Fig. 8)
L1/R/A
Red
Black
L2/S/B
L3/T/C
AC 3-phase 3-wire power measurement (3P3W balance) Fig. 8
13
UT219P User Manual
- Connect red test lead (connected with red alligator clip already) with “V”
terminal; and black 2)
(Sceotntnheectreodtawryithswbiltacchktoall”i3gPatWor”,cplipreaslsre”ad/y3)Pw3iWth””CtoOsMel”e.ct
3P3W balance mode, then the
clamp enters wire connection interface, as shown below:
“L1” on top right corner indicates the jaws clamp L1 live wire, “L2” on bottom
left corner indicates test lead from COM terminal connects with L2 live wire,
“L1” on bottom right corner indicates test lead from V terminal connects with
L1 live wire. The finger symbol on “HOLD” segment flashes to indicate pressing
“HOLD” button to enter measurement interface.
3) According to the wire connection interface (Fig. 8), clamp L1 live wire,
connect L1 live wire with test lead from V terminal, and connect L2 live wire
with test lead from COM terminal, then press “HOLD” button to enter
measurement interface.
4) In measurement interface, press SELECT button to cyclically read active
power, apparent power, reactive power, power factor and phase angle, as shown
below:
Short press
Short press
Short press
Main display: Active power Auxiliary display 1: Power factor Auxiliary display 2: AC voltage Auxiliary display 3: AC current
Main display: Apparent power Auxiliary display 1: Power factor Auxiliary display 2: AC voltage Auxiliary display 3: AC current
Main display: Reactive power Auxiliary display 1: Power factor Auxiliary display 2: AC voltage Auxiliary display 3: AC current
Main display: Power factor Auxiliary display 1: Active power Auxiliary display 2: AC voltage Auxiliary display 3: AC current
Short press
Short press
Main display: Phase angle Auxiliary display 1: Active power Auxiliary display
2: AC voltage Auxiliary display 3: AC current
4.2 3-phase 4-wire (3P4W) balance mode (Fig. 9)
L1/R/A
Red
L2/S/B
L3/T/C
Black
N
AC 3-phase 4-wire power measurement (3P4W balance) Fig. 9
14
UT219P User Manual
- Connect red test lead (connected with red alligator clip already) with “V” terminal; and black (connected with black alligator clip already) with “COM”.
- Set the rotary switch to “3PW”, press “/3P4W” to select 3P4W balance mode, then the clamp enters wire connection interface, as shown below:
UT219P User Manual
- Connect red test lead (connected with red alligator clip already) with “V”
terminal; and
black (connected with black alligator clip already) with “COM”. - Set the rotary switch to “3PW”, press ” /3P3W” to select 3P3W unbalance mode, then the clamp enters P1 wire connection interface, as shown below:
“L1” on top right corner indicates the jaws clamp L1 live wire, “n” on bottom
left corner indicates test lead from COM terminal connects with neutral wire,
“L1” on bottom right corner indicates test lead from V terminal connects with
L1 live wire. The finger symbol on “HOLD” segment flashes to indicate pressing
“HOLD” button to enter measurement interface.
3) According to the wire connection interface (Fig. 9), clamp L1 live wire,
connect L1 live wire with test lead from V terminal, and connect neutral wire
with test lead from COM terminal, then press “HOLD” button to enter
measurement interface.
4) In measurement interface, press SELECT button to cyclically read active
power, apparent power, reactive power, power factor and phase angle, as shown
below:
Short press
Short press
Short press
Main display: Active power Auxiliary display 1: Power factor Auxiliary display 2: AC voltage Auxiliary display 3: AC current
Main display: Apparent power Auxiliary display 1: Power factor Auxiliary display 2: AC voltage Auxiliary display 3: AC current
Main display: Reactive power Auxiliary display 1: Power factor Auxiliary display 2: AC voltage Auxiliary display 3: AC current
Main display: Power factor Auxiliary display 1: Active power Auxiliary display 2: AC voltage Auxiliary display 3: AC current
Short press
Short press
Main display: Phase angle Auxiliary display 1: Active power Auxiliary display
2: AC voltage Auxiliary display 3: AC current
4.3 3-phase 3-wire (3P3W) unbalance mode (Fig. 10)
Power supply Load
Red
L1/R/A
Black
L2/S/B
L3/T/C
Black Red
AC 3-phase 3-wire power measurement (3P3W unbalance) Fig. 10
15
“L1” on top right corner indicates the jaws clamp L1 live wire, “L2” on bottom
left corner indicates test lead from COM terminal connects with L2 live wire,
“L1” on bottom right corner indicates test lead from V terminal connects with
L1 live wire. The finger symbol on “HOLD” segment flashes to indicate pressing
“HOLD” button to enter P1 measurement interface. 3) According to the wire
connection interface (Fig. 10), clamp L1 live wire, connect L1 live wire
with test lead from V terminal, and connect L2 live wire with test lead from
COM terminal, then press “HOLD” button to enter P1 measurement interface, as
shown below:
The finger symbol on “HOLD” segment flashes to indicate pressing “HOLD” button
to enter P2 wire connection interface. 4) In P1 measurement interface, after
the displayed value stabilizes, press “HOLD” button to
enter P2 wire connection interface, as shown below:
“L3” on top right corner indicates the jaws clamp L3 live wire, “L2” on bottom
left corner indicates test lead from COM terminal connects with L2 live wire,
“L3” on bottom right corner indicates test lead from V terminal connects with
L3 live wire. The finger symbol on “HOLD” segment flashes to indicate pressing
“HOLD” button to enter P2 measurement interface.
16
UT219P User Manual
5) According to the wire connection interface (Fig. 10), clamp L3 live wire,
connect L3 live wire with test lead from V terminal, and connect L2 live wire
with test lead from COM terminal, then press “HOLD” button to enter P2
measurement interface, as shown below:
UT219P User Manual
- Connect red test lead (connected with red alligator clip already) with “V” terminal; and black (connected with black alligator clip already) with “COM”.
- Set the rotary switch to “3PW”, press “/3P4W” to select 3P4W unbalance mode, then the clamp enters P1 wire connection interface, as shown below:
The finger symbol on “HOLD” segment flashes to indicate pressing “HOLD” button
to enter phase-combined active power interface. 6) In P2 measurement
interface, press “HOLD” button to read phase-combined active power,
press SELECT button to cyclically read phase-combined active power, P1 active
power and P2 active power, as shown below:
Short press
“L1” on top right corner indicates the jaws clamp L1 live wire, “n” on bottom
left corner indicates test lead from COM terminal connects with neutral wire,
“L1” on bottom right corner indicates test lead from V terminal connects with
L1 live wire. The finger symbol on “HOLD” segment flashes to indicate pressing
“HOLD” button to enter P1 measurement interface.
3) According to the wire connection interface (Fig. 11), clamp L1 live wire,
connect L1 live wire with test lead from V terminal, and connect neutral wire
with test lead from COM terminal, then press “HOLD” button to enter P1
measurement interface, as shown below:
Second wire connection interface Main display: Active power P2 Auxiliary display 1: P2 prompt Auxiliary display 2: Voltage Auxiliary display 3: Current
Second wire connection interface Main display: Active power P1 Auxiliary display 1: P1 prompt Auxiliary display 2: Voltage Auxiliary display 3: Current
Phase-combined interface Main display: Phase-combined active power P1+P2 Auxiliary display 1: Phase-combined prompt
4.4 3-phase 4-wire (3P4W) unbalance mode (Fig. 11)
Power supply Load
L1/R/A
Red
L2/S/B
L3/T/C
Black
N
AC 3-phase 4-wire power measurement (3P4W unbalance) Fig. 11
17
The finger symbol on “HOLD” segment flashes to indicate pressing “HOLD” button
to enter P2 wire connection interface. 4) In P1 measurement interface, after
the displayed value stabilizes, press “HOLD” button to
enter P2 wire connection interface, as shown below:
“L2” on top right corner indicates the jaws clamp L2 live wire, “n” on bottom
left corner indicates test lead from COM terminal connects with neutral wire,
“L2” on bottom right corner indicates test lead from V terminal connects with
L2 live wire. The finger symbol on “HOLD” segment flashes to indicate pressing
“HOLD” button to enter P2 measurement interface.
18
UT219P User Manual 5) According to the wire connection interface (Fig. 11),
clamp L2 live wire, connect L2 live
wire with test lead from V terminal, and connect neutral wire with test lead
from COM terminal, then press “HOLD” button to enter P2 measurement interface,
as shown below:
The finger symbol on “HOLD” segment flashes to indicate pressing “HOLD” button
to enter P3 wire connection interface. 6) In P2 measurement interface, after
the displayed value stabilizes, press “HOLD” button to
enter P3 wire connection interface, as shown below:
“L3” on top right corner indicates the jaws clamp L3 live wire, “n” on bottom
left corner indicates test lead from COM terminal connects with neutral wire,
“L3” on bottom right corner indicates test lead from V terminal connects with
L3 live wire. The finger symbol on “HOLD” segment flashes to indicate pressing
“HOLD” button to enter P3 measurement interface. 7) According to the wire
connection interface (Fig. 11), clamp L3 live wire, connect L3 live wire
with test lead from V terminal, and connect neutral wire with test lead from
COM terminal, then press “HOLD” button to enter P3 measurement interface, as
shown below:
The finger symbol on “HOLD” segment flashes to indicate pressing “HOLD” button
to enter phase-combined active power interface.
19
UT219P User Manual
8) In P2 measurement interface, press “HOLD” button to read phase-combined
active power, press SELECT button to cyclically read phase-combined active
power, phase-combined apparent power, phase-combined reactive power, phase-
combined phase angle, P1 active power, P2 active power and P3 active power, as
shown below:
Short press
Short press
Main display: Phase-combined reactive power Auxiliary display 1: Phase- combined power factor
Main display: Phase-combined apparent power Auxiliary display 1: Phase- combined power factor
Main display: Phase-combined active power Auxiliary display 1: Phase-combined power factor
Short press
Short press
Short press
Short press
Short press
Main display: Phase-combined phase angle Auxiliary display 1: Phase-combined power factor
Main display: First active power Auxiliary display 1: P1 prompt Auxiliary display 2: Voltage Auxiliary display 3: Current
Main display: Second active power Auxiliary display 1: P2 prompt Auxiliary display 2: Voltage Auxiliary display 3: Current
Main display: Third active power Auxiliary display 1: P3 prompt Auxiliary display 2: Voltage Auxiliary display 3: Current
Note:
Phase angle is in reference to voltage. Do not input voltage over 1000V. It is
possible to measure higher voltage, but it may damage the clamp. Avoid
electric shock when working with high voltage. Disconnect test leads with the
measured circuit after all measurement operations are completed. The high
voltage warning symbol ” ” will appear on the LCD if the measured voltage is
over 30V AC; the red backlight flashes if the measured voltage is over
1000VAC. Do not release the trigger abruptly when pressing and holding it.
Hall sensor is sensitive to magnet, heat and mechanic stress, impact can cause
reading variation in a short time. To ensure measurement accuracy, please
place the measured conductor at the center of the clamp jaws. Error of ±1.0%
of reading will be produced if the measured conductor is not placed at the
center of the clamp jaws. Incorrect wire connection will cause negative active
power and cause the red backlight to flash, please check if the connections of
voltage input terminal and clamp jaws are correct.
5. Harmonic analysis
5.1 Current harmonic analysis (Fig. 4)
- Set the rotary switch to “HARM”, press SELECT button to select AC current harmonic analysis. Press the trigger to clamp the measured conductor and then release the trigger slowly until the clamp jaws closes completely. Make sure that the measured conductor centers at the clamp jaws, error will be produced if the measured conductor does not centers at the clamp jaws. The clamp can only measure a current conductor at a time, if it measures two or more current conductors at the same time, the reading will be incorrect.
- Press ” ” or ” ‘ to read THD %f, THD %r, RMS of harmonic component of each
order,
and harmonic ratio of harmonic component of each order to fundamental wave, for examples:
20
THD %f
THD %r
UT219P User Manual Analysis of component
of 30th harmonic
Main display: THD %f Auxiliary display 1: Frequency Auxiliary display 3: True RMS of THD current
Main display: THD %r Auxiliary display 1: Frequency Auxiliary display 3: True RMS of THD current
5.2 Voltage harmonic analysis (Fig. 5)
Main display: Harmonic ratio of component of 30th harmonic to fundamental wave Auxiliary display 1: Frequency Auxiliary display 2: Prompt of analyzing 30th harmonic Auxiliary display 3: True RMS of component of 30th harmonic
- Connect red test lead (connected with red alligator clip already) with “V” terminal; and black
(connected with black alligator clip already) with “COM”.
-
Set the rotary switch to “HARM”, press SELECT button to select AC voltage harmonic
analyzing, Press ”
then ” or ”
con”nteocrteaaldligTaHtoDr c%lipf,sTwHitDh
power source or load %r, RMS of harmonic
to be measured in parallel. component of each order ,
and harmonic ratio of harmonic component of each order to fundamental wave, for examples:
THD %f
THD %r
Analysis of component of 30th harmonic
Main display: THD %f Auxiliary display 1: Frequency Auxiliary display 3: RMS
of THD voltage
Note:
Main display: THD %r Auxiliary display 1: Frequency Auxiliary display 3: RMS of THD voltage
Main display: Harmonic ratio of component of 30th harmonic to fundamental wave Auxiliary display 1: Frequency Auxiliary display 2: Prompt of analyzing 30th harmonic Auxiliary display 3: RMS of component of 30th harmonic
Do not input voltage over 1000V. It is possible to measure higher voltage, but it may damage the clamp. Avoid electric shock when working with high voltage. Disconnect test leads with the measured circuit after all measurement operations are completed. The high voltage warning symbol ” ” will appear on the LCD if the measured voltage is over 30V AC; the red backlight flashes if the measured voltage is over 1000VAC. Do not release the trigger abruptly when pressing and holding it. Hall sensor is sensitive to magnet, heat and mechanic stress, impact can cause reading variation in a short time. To ensure measurement accuracy, please place the measured conductor at the center of the clamp jaws. Error of ±1.0% of reading will be produced if the measured conductor is not placed at the center of the clamp jaws. If zero crossing of harmonic does not meet the zero-crossing condition of fundamental wave signal for multiple times, harmonic analysis can be performed by using the mode of setting fixed frequency.
21
UT219P User Manual
6. Set the mode of electric energy accumulation (Fig. 12)
Red
L/R/A
Power supply Load
Black
N
Single-phase active electric energy measurement Fig. 12
- Set the rotary switch to “Wh/Setting”, press SELECT button to select the mode of electric energy accumulation, the setting parameters are shown as below:
Serial No. Setting value
oFF
No value (time accumulation)
01 3200 cyc./1kWh
02 1600 cyc./1kWh
03 1200 cyc./1kWh
04 1000 cyc./1kWh
05 600 cyc./1kWh
06 500 cyc./1kWh
Change the setting value Yes: No: —
—
Serial No.
07 08 09 10 0.10kWh 0.05kWh 0.01kWh
Setting value
300 cyc./1kWh 250 cyc./1kWh 150 cyc./1kWh 125 cyc./1kWh 0.10kWh 0.05kWh
0.01kWh
Change the setting value Yes: No: —
— — —
Press ” ” or ” ” to change the serial No. of main display. When the setting value can be changed by selecting serial No., press “HOLD” to select the part to be changed (auxiliary 2, auxiliary 3), then press “MAX MIN” to select ones place, tens place, hundreds place and thousands place of the value in auxiliary display 2. Press ” ” or ” ” to change partial values. After the setting is changed, press “SELECT” to enter the mode of measuring electric energy accumulation.
Note: The changed setting will be saved in the clamp.
-
Connect alligator clips with power source or load to be measured in parallel, and clamp the conductor to be measured, the current flow from up to down (up: front; down: bottom cover), then release the trigger slowly until the clamp jaws close completely. Please make sure that the measured conductor centers at the clamp jaws, error will be produced if the measured conductor is not placed at the center of the clamp jaws. The clamp can only measure a current conductor at a time, if it measures two or more current conductors at the same time, the reading will be incorrect.
-
In measurement interface of electric energy accumulation, press “HOLD” to start or stop accumulating electric energy. Measurement interface of time accumulation of electric energy (Serial No.: OFF)
22
UT219P User Manual
Start accumulating Short press
Start accumulating Short press
Main display: Standby mode prompt Auxiliary display 1: Active power Auxiliary display 2: Time
Main display: Measured electric energy Auxiliary display 1: Active power
Auxiliary display 2: Elapsed time
Restart Short press
Main display: Measured electric energy Auxiliary display 1: Active power Auxiliary display 2: Elapsed time
Measurement interface of compare function of electric meter
Start accumulating Short press
Start accumulating Short press
Main display: Standby mode prompt Auxiliary display 1: Active power Auxiliary display 2: Time
Main display: Operation prompt Auxiliary display 1: Active power Auxiliary
display 2: Time Auxiliary display 3: Theoretical value
Restart
Short press
Main display: Actual measured value–theoretical value Auxiliary display 1: Actual measured value, theoretical value Auxiliary display 2: Actual measured value Auxiliary display 3: Theoretical value
Operate the compare function of electric meter For comparing mechanical meter:
Press “Hold” to start accumulation when the round disk rotates for one round;
press “HOLD” again to stop accumulation when the round disk rotates for
another round, as shown below:
Mechanical Meter Round disc rotates for one round
Start accumulating
Round disc
23
Stop accumulating
UT219P User Manual
For comparing electronic meter: Press “Hold” to start accumulation when the LED flashes for
one time, press “HOLD” again to stop accumulation when the LED flashes again, as shown
below:
Electronic Meter LED flashes for one time
Start accumulating
Stop accumulating
For comparing fixed electric energy meter: Take 0.1kWh as an example, press
“HOLD” to start accumulation when the value changes by 0.1kWh; press “HOLD”
again to stop accumulation when the value changes by 0.1kWh again, as shown
below:
Start accumulating
Decimal place: 0.1 kWh
Stop accumulating
Note:
Phase angle is in reference to voltage. Do not input voltage over 1000V. It is
possible to measure higher voltage, but it may damage the clamp. Avoid
electric shock when working with high voltage. Disconnect test leads with the
measured circuit after all measurement operations are completed. The high
voltage warning symbol ” ” will appear on the LCD if the measured voltage is
over 30V AC; the red backlight flashes if the measured voltage is over
1000VAC. Do not release the trigger abruptly when pressing and holding it.
Hall sensor is sensitive to magnet, heat and mechanic stress, impact can cause
reading variation in a short time. To ensure measurement accuracy, please
place the measured conductor at the center of the clamp jaws. Error of ±1.0%
of reading will be produced if the measured conductor is not placed at the
center of the clamp jaws. Incorrect wire connection will cause negative active
power and cause the red backlight to flash, please check if the connections of
voltage input terminal and clamp jaws are correct.
24
Power supply Load
7. Phase sequence detection (Fig. 13)
L1/R/A
Red
Black
L2/S/B
L3/T/C
Black Red
UT219P User Manual
Phase detection
Fig. 13
- Connect red test lead (connected with red alligator clip already) with “V” terminal; and black (connected with black alligator clip already) with “COM”.
- Set the rotary switch to “Phase Detect” to enter the first prompt interface of wire connection, as shown below:
“L2” on bottom left corner indicates test lead from COM terminal connects with
L2 live wire, “L1” bottom right corner indicates test lead from V terminal
connects with L1 live wire.
3) According to the first prompt interface of wire connection (Fig. 13),
connect test lead from V terminal with L1 live wire, then connect test lead
from COM terminal with L2 live wire. When the voltage is locked, the clamp
enters the second prompt interface of wire connection, as shown below:
“L2” on bottom left corner indicates test lead from COM terminal connects with
L2 live wire, “L3” bottom right corner indicates test lead from V terminal
connects with L3 live wire, then the prompt time “10s” on top right corner
starts to decrease.
25
UT219P User Manual
4) According to the second prompt interface of wire connection (Fig. 13),
within 10 seconds, connect test lead from V terminal with L3 live wire, and
connect test lead from COM terminal with L2 live wire, then three kinds of
testing results will be displayed. For positive sequence, “L1”, “L2” and “L3”
light up clockwise, as shown below:
For inverse sequence, “L1”, “L2” and “L3″ light up counterclockwise, and red
backlight appears, as shown below:
Note: Do not input voltage over 1000Vrms. It is possible to measure higher
voltage, but it may damage the clamp. Avoid electric shock when working with
high voltage. Disconnect test leads with the measured circuit after all
measurement operations are completed. The high voltage warning symbol ” ” will
appear on the LCD if the measured voltage is over 30V AC; the red backlight
flashes if the measured voltage is over 1000VAC.
5) Press “HOLD” to return to initial interface to start phase sequence
detection.
8. Other functions
Auto power off: The clamp powers off automatically after 15 minutes of
inactivity. To awaken the clamp, please press SELECT button in Auto-off state.
To disable the Auto-off function, please hold down SELECT button and power on
the clamp. Restart the clamp to enable the Auto-off function.
Battery voltage detection: When the battery voltage is: >3.6V, ” ” appears.
3.4V~3.6V, ” ” appears. 3.2V~3.4V, ” ” appears. 3.0V~3.2V, ” ” appears.
2.8V~3.0V, ” ” appears. <2.8V, the clamp powers off automatically.
26
UT219P User Manual Red backlight as warning indication: Incorrect wire
connection will cause negative active power and cause the red backlight to
flash. The red backlight flashes when the measured voltage is >1000VAC and
current >1000A. The red backlight flashes if the phase sequence is inverse.
Analyzing fixed-frequency harmonic: Long press RANGE button to power on the
clamp to HARM position, then the clamp enters the fixed-frequency harmonic
setting interface, as shown below:
Press “MAX MIN” to select ones place, tens place, hundreds place and thousands
place of the value in auxiliary display. Press ” ” or ” ” to change partial
values. When “HOLD” is pressed after the setting is completed, the clamp
enters fixed-frequency harmonic analysis mode, “SET” in auxiliary display and
fixed frequency are displayed alternately for one second, as shown below,
other displayed values are same with harmonic sweeping analysis mode.
To reset fixed-frequency mode of other frequencies, please switch to other
positions and then return to HARM position. Note: Other positions are also in
fixed-frequency mode.
XII. Technical Specifications
Accuracy: ± (a% of reading + b digits), one-year warranty period Ambient
temperature: 23°C±5°C (73.4°F±9°F); relative humidity: 75% Note: The
temperature condition of accuracy is 18°C~28°C, the fluctuation range of
ambient temperature keeps within ±1°C. If the temperature is 18°C or >28°C,
the additional error of temperature coefficient is “0.1 × (specified
accuracy)/°C”. To ensure measurement accuracy, please place the measured
conductor at the center of the clamp jaws. Error of ±1.0% of reading will be
produced if the measured conductor is not placed at the center of the clamp
jaws.
27
UT219P User Manual
1. AC current
Range
60.00A 600.0A 1000A
Resolution 15Hz~40Hz
0.01A
0.1A
±(2.5%+5)
1A
Accuracy 40Hz~70Hz ±(2.0%+3)
±(1.5%+3)
70Hz~1000Hz ±(2.5%+5)
Overload protection
1000A AC
ACA frequency response: 15~1kHz ACA displayed true RMS, measurement range:
60.00A range: 0.30A~62.00A 600.0A range: 3A~620.00A 1000A range: 30A~1100A
Range to ensure accuracy: 5~100% of Range PEAK measurement range: 60.00A
range: 0.3A~120.0A 600.0A range: 3A~1200A 1000A range: 30A~1500A PEAK
accuracy: 60A position:
40Hz~70Hz: ±(5.0%+15) 70Hz~1kHz: ±(6.5%+15) 600A/1000A position: 40Hz~70Hz:
±(5.0%+5) 70Hz~1kHz: ±(6.5%+5) PEAK coefficient: <2.0
2. AC voltage
Range Resolution 15Hz~40Hz
Accuracy 40Hz~70Hz
70Hz~1000Hz
999.9
0.1V
±(2.0%+5)
±(0.7%+3)
±(2.0%+5)
ACV input impedance: 2M ACV frequency response: 15~1kHz ACV displayed true RMS, measurement range: 30.0V~999.9V PEAK measurement range: ± (30.0V~1500V) PEAK accuracy: 40Hz~70Hz: ±(2.5%+5) 15Hz~40Hz; 70~1000Hz: ±(4.0%+5) PEAK coefficient: <1.5
Overload protection
1000Vrms
3. Power 3.1 Active power
Range
60.00kW 600.0kW 1000kW
Accuracy
Resolution 15Hz~40Hz 40Hz~70Hz Power factor: 1 Power factor: 1
0.01kW 0.1kW 1kW
±(2.5%+5)
±(2.0%+5) ±(1.7%+5)
70Hz~1000Hz Power factor: 1
±(2.5%+5)
Overload protection
1000A AC 1000Vrms
28
UT219P User Manual
Note: 1) If the displayed power factor is not 1, please calculate the
specification of power according
to the phase angle error. 2) [1P] 0.09kW~1000kW 3) [3P3W] 0.15kW~1732kW 4)
[3P4W] 0.27kW~3000kW
3.2 Apparent power
Accuracy
Range
Resolution 15Hz~40Hz 40Hz~70Hz Power factor: 1 Power factor: 1
70Hz~1000Hz Power factor: 1
Overload protection
60.00kVA 600.0kVA 1000kVA
0.01kVA 0.1kVA 1kVA
±(2.5%+5)
±(2.0%+5) ±(1.7%+5)
±(2.5%+5)
1000A AC 1000Vrms
Note: 1) If the displayed power factor is not 1, please calculate the
specification of power according
to the phase angle error. 2) [1P] 0.09kVA~1000kVA 3) [3P3W] 0.15kVA~1732kVA 4)
[3P4W] 0.27kVA~3000kVA
3.3 Reactive power
Accuracy
Range
Resolution 15Hz~40Hz 40Hz~70Hz Power factor: 1 Power factor: 1
70Hz~1000Hz Power factor: 1
Overload protection
60.00kVAr 0.01kVAr 600.0kVAr 0.1kVAr 1000kVAr 1kVAr
±(3.0%+5)
±(2.5%+5) ±(2.0%+5)
±(3.0%+5)
1000A AC 1000Vrms
Note: 1) If the displayed power factor is not 1, please calculate the
specification of power according
to the phase angle error. 2) [1P] 0.09kVAr~1000kVAr 3) [3P3W]
0.15kVAr~1732kVAr 4) [3P4W] 0.27kVAr~3000kVAr
3.4 Power factor
Range Resolution
Accuracy
15Hz~1000Hz
-1~ 1
0.001
Calculate the specification of power factor according to the phase angle error
Overload protection
1000A AC 1000Vrms
Remark Min. measured voltage: 30V Max. measured current: 10A
29
UT219P User Manual
3.5 Phase angle
Range
Resolution
-180° (advance) ~179.9° (delay)
0.1°
15Hz 40~Hz
Accuracy 40Hz 70~Hz
70Hz 100~0Hz
±5° ±3° ±5°
Overload protection
Remark
1000A AC 1000Vrms
Min. measured voltage: 30V Max. measured current: 10A
Note: 1. The null-position cross phase difference between current and voltage
waveforms is positive
if current lags behind voltage; negative if current is ahead of voltage. 2. An
error of 2° is added if the measured conductor is not placed at the center of
the clamp jaws.
4. Harmonic analysis
Function
Harmonic order
Accuracy
Overload protection
15Hz~400Hz
RMS of harmonic of each order
1 2 6 7 8 9 10 11 15 16 30 30 40
±(3.0%+10) ±(3.5%+10) ±(4.5%+10) ±(5.0%+10) ±(7.0%+10) ±(10.0%+10) ±(20.0%+10)
1000A AC 1000Vrms
Note: 1) Minimum measured voltage is 30V, minimum measured current: >10% of
current range. 2) THD-F and THD-R are displayed on the LCD. 3) Harmonic
analysis is recorded in the clamp, and can be displayed and viewed on mobile
APP via Bluetooth module. 4) If fundamental frequency is 100Hz, the harmonic
order reaches 40.
If fundamental frequency is >100Hz, the harmonic order reaches 15.
5. Frequency
Range 15Hz 1000Hz
Resolution 0.1Hz
Accuracy 15Hz~1000Hz ±(0.3%+3)
Overload protection
1000A AC 1000Vrms
Note: Minimum measured voltage is 30V, minimum measured current: >5% of current range.
6. Phase detection
Function
Range
Frequency
Overload protection
Phase Detect
80V~1000V
40Hz~80Hz
1000Vrms
Note: The LCD shows “1 2 3” to indicate positive phase sequence or “3 2 1” to indicate negative one. The LCD shows “– — — –” to indicate phase loss or unmeasurable.
30
UT219P User Manual
7. Single-phase active electric energy
Range
Calculating method
0.00Wh~999.9k Wh
Every 0.3 second, add the step value of electric energy of last positive active power; add zero for negative active power.
Overload protection
1000A AC 1000Vrms
XIII. Bluetooth software
1. Introduction The Bluetooth software is a mobile APP, supporting mobile
phones running iOS 10.0 or newer and Android 5.0 or newer. Mobile phones
running other operating system depends on the actual issued application
software.
2. Installation For iOS: Search “UNI-T Smart Measurement” in “App Store”. For
Android: Search “UNI-T Smart Measurement” on UNI-T official website or scan
the QR code below.
(QR code for IOS) (QR code for Android)
3. Usage 3.1. Long press “SELECT” to turn on Bluetooth, if the clamp fails to
connect with the mobile APP after Bluetooth is turned on, the Bluetooth symbol
on the LCD will flash. Click “UNI-T Smart Measurement” APP icon, select
“UT219P” , then click connection. User can also scan the QR code at the clamp
for direct connection. After successful connection, the Bluetooth symbol is
displayed on the LCD for a long time. Data communication, measurement result
viewing, button control and other operations can be achieved through “UNI-T
Smart Measurement” APP.
3.2. “UNI-T Smart Measurement” APP has multiple functions including Bluetooth
communication, data recording, meter management, report generation, data
share, data synchronization, and more. Please refer to “UNI-T Smart
Measurement” User Manual for how to use the functions mentioned.
4. Uninstallation To uninstall the APP, please use the uninstaller of mobile
phone.
UT219P User Manual
XIV. Maintenance (Fig. 14)
Warning: 1. General maintenance a. The product must be maintained or serviced
by qualified professional repair personnel or
designated repair department. b. Clean the casing with dry cloth periodically,
do not use cleaning agent containing abrasive
or solvent.
2. Battery installation or replacement The product is powered by 3 AAA
batteries, please install or replace the battery according to the steps below:
a. Power off the product and remove the test leads from the input terminal. b.
Loosen the screw, take off the battery cover, remove the battery, and then
install new
battery according to correct polarity. c. Please install appropriate battery
with same model as original battery. d. Rejoin the battery cover and tighten
the screw.
Fig. 14
31
32