BIOWRAPS Knee BioWrap Blue User Guide
- June 9, 2024
- BIOWRAPS
Table of Contents
BIOWRAPS Knee BioWrap Blue
This Package Contains
- One Knee BioWrap Compression Electrode Garment – for use on LEFT or RIGHT knee.
- One 4 oz. Bottle of BioWrap Conductive Cream
Quick Set Up and Use
See video at biowave.com/support/biowrap
- The 4” Snap Cables should already be snapped onto each of the 2 snaps on the outside of the BioWrap.
- Plug the blue connectors at the other end of each BioWrap cable into your BioWave Leadwire Cable.
- Clean your skin with soap and water using a washcloth in the locations that the silver fabric will touch your skin.
- Squeeze a pea size amount of BioWrap Conductive Cream onto your finger tip and rub into your skin thoroughly in the locations that the silver fabric will touch your skin.
- Keep knee and leg fairly straight. See diagrams on opposite side to put on the BioWrap and velcro it closed firmly so there is strong but comfortable compression over the silver fabric locations.
- Plug the single end of the leadwire cable into the BioWave device.
- Press the power button to turn on the BioWave device.
- Start the treatment by pressing the PLUS (+) button to increase intensity so the sensation felt is strong but still comfortable. As your body adapts to the electrical field and the sensation fades, keep pressing the PLUS (+) button to increase the intensity to maintain a strong but comfortable sensation at the treatment location.
- When the 30-minute treatment is over, disconnect the BioWave Leadwire cable from the 4” Snap Cables, remove the BioWrap from your body and store in a cool dry place.
Directions for Use
These instructions are for attaching the Knee BioWrap to the RIGHT or LEFT knee.
Squeeze a pea size amount of BioWrap Conductive Cream onto your finger tip and rub into your skin thoroughly in the two locations that the silver fabric will touch the skin.
Both 4” Snap Cables should already be snapped onto the 2 snaps on the back
side of the BioWrap. Attach your BioWave Leadwire cable to the two connectors
on the back of the BioWrap.
Line up hole over center of kneecap.
Insert top elastic strap through top D-ring.
Pull elastic strap so BioWrap is tight.
Make sure hole stays centered over kneecap.
Pull top strap across front of knee so BioWrap is tight.
Secure Velcro strap to close.
Make sure hole stays centered over kneecap.
Insert lower elastic strap through lower D-ring.
Pull elastic strap so BioWrap is tight.
Make sure hole stays centered over kneecap.
Pull lower strap across front of knee so BioWrap is tight.
Secure Velcro strap to close.
TROUBLESHOOTING
My BioWave shows a connection error but my BioWrap is plugged in
If your BioWave device shows a connection error when you are fully connected to the BioWrap, remove the BioWrap and follow these steps:
If you forgot to apply BioWave Conductive Cream:
- CLEAN YOUR SKIN. Remove BioWrap and clean your skin with soap and water in the region that the silver fabric electrodes touch your skin. Shave any excessive hair in the locations that the silver fabric electrodes touch your skin.
- APPLY A PEA-SIZE AMOUNT OF BIOWAVE CONDUCTIVE CREAM directly to skin areas that will be against the silver fabric inside the BioWrap. Gently massage BioWave Conductive Cream into the skin allowing it to absorb, before putting on the BioWrap.
If you applied too much BioWave Conductive Cream (more than a pea-size amount):
- CLEAN YOUR SKIN. Remove BioWrap and use a dry towel to remove all excess Conductive Cream from your skin and any excess cream from the silver fabric in the BioWrap. Next, reattach the BioWrap and begin the treatment.
The treatment doesn’t feel strong enough
- If the intensity is turned up to the maximum setting and the treatment doesn’t feel strong under the electrodes, pause the treatment, remove the BioWrap and apply more conductive cream to the skin in the areas the silver fabric touches the skin.
- Tighter Compression and/or More Conductive Cream = Stronger Treatment Sensation.
- Looser Compression and/or Less Conductive Cream = Less Treatment Sensation.
Questions? Call us!
Call technical support:
1-877-BIOWAVE (1-877-246-9283)
biowave.com/support/biowrap
support@biowave.com
IMPORTANT
Contraindications When Not to Use BioWrap
- DO NOT use if you have an implanted cardiac pacemaker.
- DO NOT place the silver fabric electrodes in the BioWrap over open wounds, broken skin, rashy skin or sensitive skin areas like sunburned skin.
NOTE:
- BioWrap Wearables MAY BE PLACED over or in the proximity of implanted orthopedic hardware, including total joint replacements, anchors, plates, rods, screws, pins and clips.
- Any serious incident that has occurred in relation to the device should be reported to the manufacturer and the competent authority of the Member State in which the user and/or patient is established.
Cleaning
- The Silver Fabric Electrodes may be cleaned using Purell® brand Hand Sanitizing Wipes.
- BioWraps may be HAND WASHED using mild soap. Disconnect the two 4” snap cables from their snaps before hand washing the BioWrap.
- AIR DRY.
- DO NOT wash in a washing machine.
- DO NOT place in a dryer.
Need help?
Contact us at:
1-877-BIOWAVE • (1-877-246-9283)
support@biowave.com
The information, including but not limited to text, graphics, images and other material contained herein are for informational purposes only. Learn more at BioWave.com/terms-of-use
Agenoria Medical Ltd. Blair House Upper O’Connell St. Ennis, Co Clare Ireland
BioWave Corporation 8 Knight St., Suite 201 Norwalk, CT 06851
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>