Unity Lab Services Heraguard ECO Clean Bench User Guide

June 8, 2024
Unity Lab Services

Heraguard ECO Clean Bench
User Guide

Heraguard ECO Clean Bench

Troubleshooting guide
How do I activate the UV light on my Heraguard ECO?
Heraguard ECO
Revision Date: November 10, 2014
How do I activate the UV light on my Heraguard ECO?

The Heraguard Eco must be turned off for the UV light to function.

  1. Turn the cabinet blowers off ( hold the “ON” key for 10 seconds)
  2. Install the UV light resistant cover.
  3. Press and hold the UV light button for 10 secondsUnity Lab Services Heraguard ECO Clean Bench - Figure 2
  4. If the UV light still isn’t functioning verify the UV bulb cover has been removed.
  5. Please note the UV light resistant covers along with the UV lamp are recommended to be replaced every 1500 hours. See page 21, 37 in the operator’s manual.
    UV bulb replacement
    ▶ Align the pins of bulb with the sockets
    ▶ Insert the bulb into sockets
    ▶ Twist’ turn until the bulb locks
    ▶ Test Install light coverUnity Lab Services Heraguard ECO Clean Bench -
Figure 4 UV light resistant cover
    ▶ No interlock as with BSC
    ▶ Held in place with elastic cordsUnity Lab Services Heraguard ECO Clean
Bench - Figure 5

www.unitylabservices.com/contactus

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Unity Lab Services User Manuals

Related Manuals