OFITE 170-92 Safety Clamp for HTHP Fluid Loss Cells Instruction Manual

July 30, 2024
OFITE

OFITE 170-92 Safety Clamp for HTHP Fluid Loss Cells

Specifications

  • Product: Safety Clamp for HTHP Fluid Loss Cells
  • Part No.: #170-92
  • Instruction Manual Updated: 3/1/2024 Ver. 2

Product Usage Instructions

Operation
The Test Cell Safety Clamp is designed to hold an HTHP test cell securely in place and prevent an uncontrolled pressure release. The two plates clamp the cell in place while the locking screws are removed. Pressure can then be released gradually and safely.

Steps to Use the Safety Clamp

  1. After a filter press test is complete, allow the cell to cool completely before removing it from the heating jacket.
  2. Adjust the upper nuts on the Safety Clamp to be high enough to raise the bar above the top of the cell. Turn them counterclockwise up.
  3. Place the cell in the clamp with the inlet side pointing up. Rest the plate on top of the cell.
  4. Adjust the lower nuts to be level with the top of the cell. Turn them counterclockwise to raise them up and clockwise to lower them.
  5. Adjust the upper nuts to be tight against the top of the plate. Turn them clockwise to lower them.
  6. Once the cell is secure in the clamp, carefully unscrew and remove the six locking screws around the top (inlet) of th cell.
  7. Slowly move the upper nuts upward by turning them counterclockwise at the same time.
  8. If pressure is trapped in the cell, the cap will be lifted as the plate is raised. Continue adjusting the upper nuts upward until the pressure has been released.
  9. Once the pressure has been released, remove the cell from the clamp and thoroughly clean it as instructed in the filter press instruction manual.

FAQ
Warranty and Return Policy:

Warranty:
In the event that the buyer has purchased installation and commissioning services on applicable products, the above warranty shall extend for an additional period of twelve (12) months from the date of the original warranty expiration for such products.

Safety Clamp for HTHP Fluid Loss Cells
Part No. #170-92

Instruction Manual
Updated 3/1/2024

  • Ver. 2
  • OFI Testing Equipment, Inc.
  • 11302 Steeplecrest Dr. · Houston, Texas · 77065 · U.S.A.
  • Tele: 832.320.7300 · Fax: 713.880.9886 · www.ofite.com
  • ©Copyright OFITE 2018

Operation

The Test Cell Safety Clamp is designed to hold an HTHP test cell securely in place and prevent an uncontrolled pressure release. The two plates clamp the cell in place while the locking screws are removed. Pressure can then be released gradually and safely.

  1. After a filter press test is complete, allow the cell to cool completely before removing it from the heating jacket.
  2.  Adjust the upper nuts on the Safety Clamp to be high enough to raise the bar above the top of the cell. Turn them counterclockwise up.
  3. Place the cell in the clamp with the inlet side pointing up. Rest the plate on top of the cell.
  4. Adjust the lower nuts to be level with the top of the cell. Turn them counterclockwise to raise them up and clockwise to lower them.
  5. Adjust the upper nuts to be tight against the top of the plate. Turn them clockwise to lower them.
  6. Once the cell is secure in the clamp, carefully unscrew and remove the six locking screws around the top (inlet) of the cell.
  7. Slowly move the upper nuts upward by turning them counterclockwise at the same time.
  8. If pressure is trapped in the cell, the cap will be lifted as the plate is raised. Continue adjusting the upper nuts upward until the pressure has been released.
  9. Once the pressure has been released, remove the cell from the clamp and thoroughly clean it as instructed in the filter press instruction manual.

Warranty and Return Policy

Warranty

OFI Testing Equipment, Inc. (OFITE) warrants that the products shall be free from liens and defects in title, and shall conform in all respects to the terms of the sales order and the specifications applicable to the products. All products shall be furnished subject to OFITE’s standard manufacturing variations and practices. Unless the warranty period is otherwise extended in writing, the following warranty shall apply: if, at any time prior to twelve (12) months from the date of invoice, the products, or any part thereof, do not conform to these warranties or to the specifications applicable thereto, and OFITE is so notified in writing upon discovery, OFITE shall promptly repair or replace the defective products. Notwithstanding the foregoing, OFITE’s warranty obligations shall not extend to any use by the buyer of the products in conditions more severe than OFITE’s recommendations, nor to any defects which were visually observable by the buyer but which are not promptly brought to OFITE’s attention.

The event that the buyer has purchased installation and commissioning services on applicable products, the above warranty shall extend for an additional period of twelve (12) months from the date of the original warranty expiration for such products.
In the event that OFITE is requested to provide customized research and development for the buyer, OFITE shall use its best efforts but makes no guarantees to the buyer that any products will be provided.
OFITE makes no other warranties or guarantees to the buyer, either express or implied, and the warranties provided in this clause shall be exclusive of any other warranties including ANY IMPLIED OR STATUTORY WARRANTIES OF FITNESS FOR PURPOSE, MERCHANTABILITY, AND OTHER STATUTORY REM-EDIES WHICH ARE WAIVED.

This limited warranty does not cover any losses or damages that occur as a result of:

  • Improper installation or maintenance of the products
  • Misuse
  • Neglect
  • Adjustment by non-authorized sources
  • Improper environment
  • Excessive or inadequate heating or air conditioning or electrical power failures, surges, or other irregularities
  • Equipment, products, or material not manufactured by OFITE
  • Firmware or hardware that has been modified or altered by a third party
  • Consumable parts (bearings, accessories, etc.)

Returns and Repairs:

Items being returned must be carefully packaged to prevent damage in shipment and insured against possible damage or loss. OFITE will not be responsible for equipment damaged due to insufficient packaging.
Any non-defective items returned to OFITE within ninety (90) days of the invoice are subject to a 15% restocking fee. Items returned must be received by OFITE in original condition for it to be accepted. Reagents and special order items will not be accepted for return or refund.
OFITE employs experienced personnel to service and repair equipment manufactured by us, as well as other companies. To help expedite the repair process, please include a repair form with all equipment sent to OFITE for repair. Be sure to include your name, company name, phone number, email address, detailed description of work to be done, purchase order number, and a shipping address for returning the equipment. All repairs performed as “repair as needed” are subject to the ninety (90) day limited warranty. All “Certified Repairs” are subject to the twelve (12) month limited warranty.
Returns and potential warranty repairs require a Return Material Authorization (RMA) number. An RMA form is available from your sales or service representative.

Please ship all equipment (with the RMA number for returns or warranty repairs) to the following address:

  • OFI Testing Equipment, Inc.
  • Attn: Repair Department
  • 11302 Steeplecrest Dr.
  • Houston, TX 77065

USA
OFITE also offers competitive service contracts for repairing and/or maintaining your lab equipment, including equipment from other manufacturers. For more information about our technical support and repair services, please contact techservice@ofite.com

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals