Product Usage Instructions
Operation
The Test Cell Safety Clamp is designed to hold an HTHP test cell securely in
place and prevent an uncontrolled pressure release. The two plates clamp the
cell in place while the locking screws are removed. Pressure can then be
released gradually and safely.
Steps to Use the Safety Clamp
- After a filter press test is complete, allow the cell to cool completely before removing it from the heating jacket.
- Adjust the upper nuts on the Safety Clamp to be high enough to raise the bar above the top of the cell. Turn them counterclockwise up.
- Place the cell in the clamp with the inlet side pointing up. Rest the plate on top of the cell.
- Adjust the lower nuts to be level with the top of the cell. Turn them counterclockwise to raise them up and clockwise to lower them.
- Adjust the upper nuts to be tight against the top of the plate. Turn them clockwise to lower them.
- Once the cell is secure in the clamp, carefully unscrew and remove the six locking screws around the top (inlet) of th cell.
- Slowly move the upper nuts upward by turning them counterclockwise at the same time.
- If pressure is trapped in the cell, the cap will be lifted as the plate is raised. Continue adjusting the upper nuts upward until the pressure has been released.
- Once the pressure has been released, remove the cell from the clamp and thoroughly clean it as instructed in the filter press instruction manual.
FAQ
Warranty and Return Policy:
Warranty:
In the event that the buyer has purchased installation and commissioning
services on applicable products, the above warranty shall extend for an
additional period of twelve (12) months from the date of the original warranty
expiration for such products.
Safety Clamp for HTHP Fluid Loss Cells
Part No. #170-92
Instruction Manual
Updated 3/1/2024
- Ver. 2
- OFI Testing Equipment, Inc.
- 11302 Steeplecrest Dr. · Houston, Texas · 77065 · U.S.A.
- Tele: 832.320.7300 · Fax: 713.880.9886 · www.ofite.com
- ©Copyright OFITE 2018
Operation
The Test Cell Safety Clamp is designed to hold an HTHP test cell securely in place and prevent an uncontrolled pressure release. The two plates clamp the cell in place while the locking screws are removed. Pressure can then be released gradually and safely.
- After a filter press test is complete, allow the cell to cool completely before removing it from the heating jacket.
- Adjust the upper nuts on the Safety Clamp to be high enough to raise the bar above the top of the cell. Turn them counterclockwise up.
- Place the cell in the clamp with the inlet side pointing up. Rest the plate on top of the cell.
- Adjust the lower nuts to be level with the top of the cell. Turn them counterclockwise to raise them up and clockwise to lower them.
- Adjust the upper nuts to be tight against the top of the plate. Turn them clockwise to lower them.
- Once the cell is secure in the clamp, carefully unscrew and remove the six locking screws around the top (inlet) of the cell.
- Slowly move the upper nuts upward by turning them counterclockwise at the same time.
- If pressure is trapped in the cell, the cap will be lifted as the plate is raised. Continue adjusting the upper nuts upward until the pressure has been released.
- Once the pressure has been released, remove the cell from the clamp and thoroughly clean it as instructed in the filter press instruction manual.
Warranty and Return Policy
Warranty
OFI Testing Equipment, Inc. (OFITE) warrants that the products shall be free from liens and defects in title, and shall conform in all respects to the terms of the sales order and the specifications applicable to the products. All products shall be furnished subject to OFITE’s standard manufacturing variations and practices. Unless the warranty period is otherwise extended in writing, the following warranty shall apply: if, at any time prior to twelve (12) months from the date of invoice, the products, or any part thereof, do not conform to these warranties or to the specifications applicable thereto, and OFITE is so notified in writing upon discovery, OFITE shall promptly repair or replace the defective products. Notwithstanding the foregoing, OFITE’s warranty obligations shall not extend to any use by the buyer of the products in conditions more severe than OFITE’s recommendations, nor to any defects which were visually observable by the buyer but which are not promptly brought to OFITE’s attention.
The event that the buyer has purchased installation and commissioning services
on applicable products, the above warranty shall extend for an additional
period of twelve (12) months from the date of the original warranty expiration
for such products.
In the event that OFITE is requested to provide customized research and
development for the buyer, OFITE shall use its best efforts but makes no
guarantees to the buyer that any products will be provided.
OFITE makes no other warranties or guarantees to the buyer, either express or
implied, and the warranties provided in this clause shall be exclusive of any
other warranties including ANY IMPLIED OR STATUTORY WARRANTIES OF FITNESS FOR
PURPOSE, MERCHANTABILITY, AND OTHER STATUTORY REM-EDIES WHICH ARE WAIVED.
This limited warranty does not cover any losses or damages that occur as a result of:
- Improper installation or maintenance of the products
- Misuse
- Neglect
- Adjustment by non-authorized sources
- Improper environment
- Excessive or inadequate heating or air conditioning or electrical power failures, surges, or other irregularities
- Equipment, products, or material not manufactured by OFITE
- Firmware or hardware that has been modified or altered by a third party
- Consumable parts (bearings, accessories, etc.)
Returns and Repairs:
Items being returned must be carefully packaged to prevent damage in shipment
and insured against possible damage or loss. OFITE will not be responsible for
equipment damaged due to insufficient packaging.
Any non-defective items returned to OFITE within ninety (90) days of the
invoice are subject to a 15% restocking fee. Items returned must be received
by OFITE in original condition for it to be accepted. Reagents and special
order items will not be accepted for return or refund.
OFITE employs experienced personnel to service and repair equipment
manufactured by us, as well as other companies. To help expedite the repair
process, please include a repair form with all equipment sent to OFITE for
repair. Be sure to include your name, company name, phone number, email
address, detailed description of work to be done, purchase order number, and a
shipping address for returning the equipment. All repairs performed as “repair
as needed” are subject to the ninety (90) day limited warranty. All “Certified
Repairs” are subject to the twelve (12) month limited warranty.
Returns and potential warranty repairs require a Return Material Authorization
(RMA) number. An RMA form is available from your sales or service
representative.
Please ship all equipment (with the RMA number for returns or warranty repairs) to the following address:
- OFI Testing Equipment, Inc.
- Attn: Repair Department
- 11302 Steeplecrest Dr.
- Houston, TX 77065
USA
OFITE also offers competitive service contracts for repairing and/or
maintaining your lab equipment, including equipment from other manufacturers.
For more information about our technical support and repair services, please
contact techservice@ofite.com
OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>