Amptek XR-100CR Si Pin X Ray Detector and Preamplifier User Manual
- July 26, 2024
- AMPTEK
Table of Contents
- Warnings and Precautions
- Description
- Amptek SiPIN Specifications
- XR-100 Preamplifier Specifications
- SiPIN Mechanical Specifications
- Electrical Interface
- Operating Notes with Amptek power supply
- Troubleshooting
- Warranty and Technical Questions
- X-Ray Fluorescence Process Example: Titanium Atom (Ti = 22)
- Amptek K and L Emission Line Lookup Chart
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
XR100CR SiPIN User Manual -B3
Warnings and Precautions
CAUTION: READ MANUAL BEFORE USING THE XR-100CR
DOUBLE INSULATED. FOR INDOOR USE ONLY.
DO NOT DROP THE DETECTOR, CAUSE MECHANICAL SHOCK TO THE DETECTOR, OR CAUSE
DAMAGE TO THE DETECTOR
-
Mechanical shock can damage components inside the TO-8 package.
-
There is vacuum inside the TO-8 for cooling. Damage to the package can cause a vacuum leak, preventing good cooling. DAMAGE TO THE PACKAGE IS NOT COVERED UNDER WARRANTY.
DO NOT TOUCH THE THIN WINDOW ON THE END OF THE DETECTOR -
BROKEN WINDOWS DAMAGED BY IMPROPER HANDLING WILL NOT BE COVERED BY WARRANTY.
-
The detector window is made from either thin beryllium (13 µm or less) or from thin Si3 (as thin as 40 nm). The windows are extremely brittle and shatter easily.
-
Do not permit any object to come into contact with the window.
-
The window cannot be repaired or replaced. If the window breaks, the detector must be replaced.
-
Keep the red protective cover installed when not in use.
AVOID RADIATION DAMAGE TO THE DETECTOR -
A RADIATION DAMAGED DETECTOR WILL NOT BE COVERED UNDER WARRANTY.
-
The detector will experience radiation damage if it is exposed to a high flux environment, e.g. directly from a synchrotron.
-
If the flux is low enough for spectroscopic operation, e.g. a count rate of a few hundred kcps, there will be no radiation damage in many years of continuous operation. But there are beams that produce a flux many orders of magnitude higher than this, and these will cause damage.
-
Also, avoid radiation exposure to the electronics, the preamplifier and signal processor. High voltage is present inside the preamplifier. This is typically +180 V. The current is limited to <100
-
A so is not a personnel hazard.
For best performance the detector and preamplifier should be mounted to a heat sink. They should be kept away from incandescent lamps and not held in the hand. The thermoelectric cooler dissipates up to 2 W. A low thermal resistance path to a heat sink is needed to keep the detector cool, which is needed for the lowest electronic noise and for spectrum stability.
For best performance pay attention to possible sources of electromagnetic interference. Use a single point electrical ground, use the shortest length cables possible, and keep the system far from sources of electromagnetic interference, such as computer monitors, high power high voltage power supplies, etc. The signals from the detector are very small so performance can be degraded by EMI.
WARNING
This product contains the following chemicals, which are known to the State of
California to cause cancer, birth defects or other reproductive harm if
exposed to them through improper use, storage, or disposal of the product:
Prop 65 Chemical| Type of Toxicity| CAS No.| Product part containing the
chemical
---|---|---|---
Beryllium| Cancer| —| Detector window
Please consult this owner’s manual for proper use, storage, care and disposal of the product. For more information, go to: www.p65warnings.ca.gov
Description
2.1 Core detector technology
Amptek provides a family of high performance, compact X-ray detectors and
associated signal processing electronics. The radiation detectors are custom
photodiodes, including the traditional Si-PIN diodes, Silicon Drift Detectors
(SDDs), and CdTe Schottky diodes. The detector is mounted on a two-stage
thermoelectric cooler along with the key preamplifier components. The cooler
keeps the detector and key components at -25 C or below, reducing electronic
noise without cryogenic liquid nitrogen and drawing <1W. This cooling permits
high performance in a compact, convenient package, and has been critical to
the development of portable XRF analyzers and of high performance, benchtop
XRF and EDS systems. Amptek’s detectors represent the state-of-the-art in
X-ray spectroscopy, delivering the best energy resolution, best efficiency at
low energies, highest count rates, highest peak to background ratios, all at
low cost and suitable for portable systems, vacuum systems, etc.
They are used by OEMs and by laboratory researchers. The core enabling
technologies include the detectors themselves (which are designed and
manufactured by Amptek), the low noise JFET and CMOS technology, and the
packaging which enables good cooling in a robust system. The sketch above
illustrates a detector mounted on a thermoelectric cooler, on a TO-8 header.
The input FET and other components are also mounted on the cooler. A nickel
cover (also shown) is welded to the TO-8 header with vacuum inside the
enclosure for optimum cooling. In the cover is a window (shown green above) to
enable soft X-ray detection. This is typically beryllium for energies > 2 keV,
with Si3 N4 available for lower energies. The entire assembly shown above is
sometimes called the “detector”, though strictly speaking it is the photodiode
which detects the X-rays.
2.2 Preamplifiers and signal processors
The detector assembly shown above must be connected to a preamplifier (a
circuit board containing those portions of the preamp not in the TO-8). Amptek
uses reset-style charge sensitive preamplifiers for the lowest noise and
highest count rates. Each type of Amptek detector (FAST SDD®, SDD, Si-PIN, and
CdTe) requires its own preamp circuit. These are available in several
different standard package options: as an XR-100 box, as a standard OEM preamp
(PA210 or PA230) or in a custom board. The output of the preamplifier must be
connected to signal processing electronics (which includes pulse shaping and a
multichannel analyzer) and power supplies. Amptek has several different
options for these, including the X123 (where all are integrated in a single,
small box), the DP5/PC5 board stack (bare boards, for integrating into
customer systems), and the PC5 module (usually used in the laboratory). Please
refer to Amptek’s website for more information.
Amptek SiPIN Specifications
3.1 Specification Table
General
Detector Type| SiPIN
Detector Size (mm2)| 6| 13| 25
Collimated Area mm2| 4.4| 11.1| 21.5
Thickness| 500 µm
Collimator Type| Internal Multilayer
Preamplifier Type| JFET with Amptek custom reset through HV
Energy resolution @ 5.9 keV ( 55Fe)
6 mm2:139 to 159 eV FWHM at Tpk=32 µs
13 mm2:150 to 205 eV FWHM at Tpk=32 µs
25 mm2:190 to 225 eV FWHM at Tpk=32 µs
Other Performance
Peak to background (typical)| > 16,000:1 for 6 mm2
> 15,000:1 for 13 mm2
> 12,000:1 for 25 mm2
Maximum input count rate| > 50 kcps
Throughput count rate stability| Determined by signal processor &
its settings
Window Options
Be| 1 mil (25 µm)
0.5 mil (12.5 µm)
1 µm Parylene on each
Signal Output|
Sensitivity| 1 mV/keV
Gain stability| <20 ppm/oC
Polarity| Negative signal
Reset range| -5 to +5 V (typical)
Cooling|
Cooling performance| T > 85oC
Cooler type| Two stage thermoelectric
Temp monitor| Diode
Power|
HV Bias| + 180V (typical)
Max cooling power| 3.5 V / 0.45A
---|---
Total power| < 2 W (full cooling)
Other|
Operating range| -35 oC to +80 oC Performance degrades at elevated detector
temperatures.
Storage & shipping| -40 oC to +85 oC, 10% to 90% RH noncondensing
RoHS| Compliant
Lifetime| Typical 5 to 10 years, depending on use
Warranty Period| 1 year
Preamplifier Options
Amptek’s Si-PINs are available with the XR-100 preamplifier, with the PA210 or
PA230 OEM preamplifiers, or as part of the X-123 and X-55.
Signal Processing Options
Amptek’s Si-PINs may be used with Amptek’s DP5 or DP5-X, or the PX5, or as
part of the X-123 or X-55.
Specification Notes
-
Performance listed here is measured at full cooling (<220K).
-
The “peak to background ratio” is the ratio of counts in the Mn K
-
peak channel to the counts at 1 keV.
-
“Gain stability” refers to the change in sensitivity when the preamp temperature changes but the detector temperature is stable.
Operating Notes -
Detector must be operated with an appropriate heat sink.
-
The SiPIN is suitable for vacuum applications. Amptek provides several system configurations.
3.2 Efficiency
The intrinsic efficiency is the probability that an X-ray incident on the
detector window (in the active region) will interact in the detector and thus
deposit its energy, creating a signal pulse. It is the product of (a) the
probability of transmission through the window and (b) the probability of
interacting in the 500 m silicon depth (the detector is fully depleted). The
low energy portion of the curves is dominated by the stopping in the window,
while the high energy portion is dominated by transmission through the active
depth of the detector. The plot below shows the computed intrinsic efficiency
for a 500 m thick silicon detector with Amptek’s standard window options. See
the Amptek website for more information.
3.3 Energy resolution and electronic noise
The energy resolution, E, is a function of the energy E of the X-ray and of
the intrinsic electronic noise, ENC, which for a given detector depends on the
peaking time (Tpeak ) and the detector temperature:
where KF E is due to “Fano broadening” and equals 119 eV at the 5.895 keV K
line of Mn. Fano broadening is the theoretical limit for a noiseless detector.
The plots below illustrate the resolution for a typical SiPIN.
The first plot shows the resolution at the 5.895 keV Kline of Mn, measured
with 55 Fe for 6, 13 and 25mm. The second plot shows the electronic noise and
the third plot shows how the resolution depends on X-ray energy, for various
values of the noise.
3.4 Spectra
The plots below show typical spectra measured with a 6 mm 2 SiPIN.
XR-100 Preamplifier Specifications
General|
---|---
Preamp Type: Reset style charge sensitive preamplifier.
Power|
HV Bias| +180V
Max cooling power| 3.5 V / 0.45 A
Low voltages| +/- 8 to 9 V @15 mA. No more than 50 mV p-p
Mechanical|
Case size (without extender)| 3.00” x 1.75” x 1.13” 7.6 cm x 4.4 cm x 2.9 cm
Weight| 4.4 oz (125 g)
XR-100 Power Connector
Type| 6 pin LEMO
Pin 1| Temperature monitor diode
Pin 2| Detector Bias.
Pin 3| -9 V power
Pin 4| +9 V power
Pin 5| Cooler power return
Pin 6| Cooler power
Case| Ground and shield
---|---
XR-100 Signal Connector
Type| BNC Coax
Other|
Operating range| -35 oC to +80 oC – @components
Storage & shipping| -40 oC to +85 oC, 10% to 90% RH noncondensing
TUV Certification| Certificate #: CU 72072412 01 Tested to: UL 61010-1: 2004
R7 .05 CAN/CSA-C22.2 61010-1: 2004
RoHS| Compliant
Warranty Period| 1 year
XR-100 Options
- The XR-100 preamplifiers are available with extender lengths of none, 1.5”, 5”, and 9”.
- The XR-100 is available with a circuit to regulate the detector temperature when not used with Amptek power supplies.
- The XR-100 may be used inside a vacuum chamber or with a vacuum feedthrough. Contact Amptek for more information.
4.1 XR-100 Mechanicals
Mechanical design of standard XR-100 (1.5” extender). Visit www.amptek.com or contact Amptek for dimensions of the other configurations (5” extender, 9” extender, etc.).
SiPIN Mechanical Specifications
- All dimensions shown are in inches unless otherwise specified.
- All dimensions are “typical” unless a tolerance is listed.
5.1 Dimensions
Compatibility
- The 6 mm 2 , 13 mm 2 , and 25 mm 2 Si-PIN detectors are electrically compatible with one another.
- Amptek’s SiPIN detectors are NOT electrically compatible with the FAST SDD®, SDD, CdTe, or other Amptek detectors.
5.2 Construction
The detector element (the SiPIN itself) is mounted on a substrate, which is
mounted on a thermoelectric cooler, which is attached to the TO-8 header. A
multilayer collimator is mounted on the detector to mask the edges of the
active volume. Interactions outside the active volume can cause spectral
anomalies. Over this assembly is a cover (typically nickel) containing a
window (typically beryllium) through which low energy X-rays pass.
- Window: The Be windows are attached to the cover with a lead-free solder.
- Cover: Standard units are 0.25 mm (10 mils) thick nickel, with an O.D. of 0.55”.
- Header: Made of Kovar. O.D. is 0.60”.
- Multilayer collimator: Base metal is 100 µm of tungsten (W), the first layer is 35 µm of chromium (Cr), the second layer is 15 µm of titanium (Ti), and the last layer is 75 µm of aluminum (Al).
- Substrate: Alumina (Al2 O3 )
Electrical Interface
6.1 Detector Output Connection Diagram
| 6 pin LEMO on XR-100 LEMO P/N ERA.1S.306.CLL| | 9 pin D Connector on
ACH-403 cable Standard 9-pin D
---|---|---|---
1| Temperature Monitor| 1| +9 V Preamp Power
2| HV Bias (180 V)| 2| -9 V Preamp Power
3| -9 V Preamp Power| 3| Cooler power (3.5 V / 0.45 A MAX)
4| +9 V Preamp Power| 4| Temperature Monitor
5| Cooler power return| 5| HV Bias (180 V)
6| Cooler power (3.5 V / 0.45 A MAX)| 6| Ground (signal & chassis)
Shield| Ground (signal and chassis)| 7| Cooler power return
| | 8, 9| Ground (signal & chassis)
6.3 Preamplifier Output Signal
The plots below illustrate the typical signals seen from a SiPIN preamplifier
and DPP.
Preamplifier outputs. These traces illustrate typical output pulses with a
SiPIN detector, measured using 5.89 and 22.1 keV X-rays. Each X-ray
interaction results in a negative going step of ~1 mV/keV.
6.4 Preamplifier Reset Signal
Preamplifier output. This trace also shows the preamp output but on a
different voltage and time scale to illustrate the reset. The small steps from
each signal integrate towards the negative rail, where a reset signal is
generated. This results in a sawtooth of several volt amplitude. The period
depends on the total current through the detector (signal current plus leakage
current).
This was measured using an XR-100, which has a reset range of about +/-5V. The
PA210 and PA230 have reset ranges of about +/- 2.5V.
6.5 Shaped peak and input to the ADC Signals
Signals. These traces show the preamplifier output (dark blue), the input
to the ADC (light blue) and the shaped output (magenta) for typical pulses.
The trace on the left shows three pulses which are well separated in time. The
plot on the right shows several different traces.
Note that the ADC input has an offset of approximately 0.2 V, with positive
going pulses exhibiting an exponential 3.2 us tail. A 1V step into the ADC
corresponds to a full scale event in the histogram.
6.6 Cooling
Recommended temperature
What is the recommended operating temperature for the detector (i.e. the
photodiode itself)? The electronic noise is improved as cooling is reduced. If
the goal is the best possible energy resolution, the lowest possible noise,
then operate the detector as cold as possible, with the maximum voltage across
the cooler. But this has a couple of disadvantages: (1) power dissipation (and
heat sink requirements) increase rapidly as the detector is cooled, and (2)
the detector temperature will vary as the ambient temperature varies, and gain
is a function of detector temperature, so the stability of the spectrum
suffers. For most users, stability and power are more important than achieving
the absolute minimum noise. For the best stability run the detector at the
maximum ambient temperature that is expected (e.g. 35 C) at full cooling. Note
the temperature the detector reaches (perhaps 236 K in this example). Then use
a set point that is slightly higher (240 K in this example). That way, the
detector temperature is as cold as feasible, for stability, in this particular
instrument.
Temperature controller
The XR-100CR includes a closed loop temperature controller that regulates the
cooler temperature. By default it is enabled and set for maximum cooling and
therefore does not regulate on its own. The thermoelectric cooler can achieve
85 o C temperature differential, so at room temperature the detector can reach
220 K to 230 K.
The PX5 includes a power supply for the thermoelectric cooler which provides
its own regulation. When the XR-100 SI-PIN is to be used with a PX5, set the
XR-100CR control to maximum cooling and let the PX5 regulate. If the PX5 and
XR-100CR are set to different temperature set points, whichever set point is
set the warmest will control the detector temperature.
The figure above shows how to adjust the controller on the XR-100CR. Remove
the top cover and adjust the R39 pot to achieve the desired TMP SET value.
This is the set point temperature. It reads in mV with the calibration curve
shown in the next section. TMP MON shows the measured temperature in mV with
the same calibration curve.
Temperature sensor
The temperature sensor in Amptek’s detectors is a forward connected 1N914
diode. At a fixed current the forward voltage is a function only of
temperature as shown below. Amptek’s standard processors use a 730 A current
yielding the calibration curve shown. The calibration will change if a
different current is used.
Thermoelectric cooler
The plot below shows the typical voltage across the cooler and the current
through the cooler, as a function of the temperature across the cooler at a
base temperature of 30 o C. Cooling is improved at a higher base temperature.
Note that T is difference between the temperature of the detector and that of
the base, the TO-8 header and stud. This is typically warmer than the ambient
environment. Good performance requires the base to be kept as cold as
feasible. Note that the power drawn by the cooler decreases rapidly with T. If
maximum cooling (T = 85 o C) occurs at 1 W, at 0.5 W T is about 78 o C, and at
0.2 W, T is 55 C.
6.7 Supply Voltages
High Voltage
The recommended bias voltage for Amptek’s Si-PIN is typically +180 V at a low
current (dark current is a fraction of a pA). This is the voltage at which
Amptek tests Si-PIN detectors and at which performance is guaranteed. In some
units, the bias voltage may be increased and a performance improvement may be
found, but this is not guaranteed. The typical maximum is >250V but, again,
performance above 180V is not guaranteed.
There must be very low noise and ripple on the HVPS. Amptek’s power supplies
include several low pass filter stages.
Low Voltage Supplies
The preamp power should be between +/- 8 and 9 VDC at 15 mA. There must be <
50 mV p-p noise (lower noise is recommended).
Operating Notes with Amptek power supply
If the XR-100 Si-PIN is to be used with a power supply from Amptek, e.g. a
PX5, PC5, or X-123, then follow the instructions in the “Quick Start” guide
provided to you and in the “User Manual” for your power supply.
Operating Notes without Amptek power supply
If the XR-100 SI-PIN is to be used with your own power supplies, then follow
these instructions.
CAUTION: SUPPLY VOLTAGES MUST BE PROPERLY CURRENT LIMITED when using the
XR-100CR without a PX5 power supply. Current should be limited to values
listed in section 7.2
7.1 Equipment Required
-
Power Supplies
➢ A dual tracking +/-9 VDC @ 35 mA with voltage meter & current limit
➢A zero to +3.5 VDC @ 0.7 A adjustable with voltage and current meter
➢A zero to 180 VDC adjustable @ 25 µA (see HV requirements)
➢Equivalent supplies, such as Amptek’s PX5 or PC5, may be used. -
Multimeter with high input impedance (>1000 M).
-
Signal processing electronics. For a Si-PIN, a digital pulse processor such as Amptek’s PX5 or DP5 is strongly recommended. The signal processing electronics should include pulse shaping and multichannel analysis functions.
-
Oscilloscope
-
Low energy radioactive x-ray source (preferably 55 Fe)
-
AC power outlet strip (preferably with surge suppression & EMI/RFI filtering).
7.2 Absolute Maximum Ratings
o Cooler power. ……….. +0.7 AMPS
o Preamp power. ……… +/- 9 VOLTS
o Detector Bias (HV). … +200 VOLTS
7.3 Connections and Turn-On Procedure (without PX5 or other Amptek DPP/supply)
- Turn all power supplies OFF. Plug all equipment to be used into one common AC power outlet strip. This will help prevent ground loops, which is crucial in getting good performance.
- Set voltages and current limits on all power supplies as noted above. Turn supplies off.
- Connect the LEMO CONNECTOR cable to the XR-100CR, according to the pin assignments given in section 6.1 of this manual.
- Attach the OUTPUT of the XR-100CR to the INPUT of the signal processor. A BNC connector is provided on the rear panel of the XR-100CR. The output pulses of the XR-100CR are NEGATIVE.
- Attach the OUTPUT of the XR-100CR to one oscilloscope input and the OUTPUT of the shaping amplifier to a second oscilloscope input.
- Turn ON the +/-9 VDC power supplies to power the charge sensitive preamplifier. Verify that both the + and – Volt outputs are between 8 and 9 Volts. NEVER EXCEED 9 VOLTS.
- ALWAYS INCREASE THE HV POWER SLOWLY TO PROTECT THE INPUT FET. WHEN TURNING OFF THE XR-100CR, DECREASE THE HV SLOWLY TO ZERO VOLTS BEFORE TURNING OFF THE XR-100CR. IF THE FET IS DAMAGED DUE TO HIGH VOLTAGE TRANSIENTS, THE WARRANTY WILL BE VOID.
- Increase the HV supply to + 180 V. The output of the preamplifier should exhibit a sawtooth waveform as in section 6.3 at a fairly high frequency.
- Power the temperature sensor using a current source, as shown in section 0, and monitor Vtemperature (mV) with a meter.
- While observing the meter slightly increase the cooler supply current until the temperature reading starts to change on the meter. Observe that the voltage is increasing, indicating that the temperature is decreasing. The reset frequency will decrease as the detector cools.
- Once the temperature gets below -40 °C the performance of the XR-100CR system will not change with a temperature variation of a few degrees. Now the XR-100CR is fully operational.
- THE COOLER IS FRAGILE AND WILL BE PERMANENTLY DAMAGED IF EXCESSIVE CURRENT OR IF REVERSE POLARITY IS APPLIED. THE WARRANTY WILL BE VOID IF THE COOLER IS DAMAGED DUE TO EXCESSIVE CURRENT OR REVERSE POLARITY, OR IF THE THIN Be WINDOW IS DAMAGED.
- Remove the red protective cover from the detector of the XR-100CR. Place the X-ray source in front of the detector.
- The output of the signal processor should show the X-ray pulses as shown in section 6.3.
- Once the temperature has stabilized (about one minute), start taking data on the MCA. For normal operation there is no need to monitor the temperature.
Troubleshooting
The XR-100CR has undergone extensive testing and burn-in before leaving the factory. If the performance of the system is not similar to the one recorded at the factory before shipping please perform the following tests: If no spectrum is observed and no counts are observed:
-
Double check all the power supply voltages and the signal connection.
-
Make sure the XR-100CR output is connected to a high impedance (not 50 ).
-
Using an oscilloscope verify the presence of the periodic reset signal shown in section 6.3. If there is no reset sawtooth the detector is not functioning.
-
Verify that the reset period decreases when you place a source in front of the detector.
-
Check your signal processor. There are many settings in a modern signal processor (gain, input offset, thresholds, etc.) which can prevent the processor from observing the X-rays. If a spectrum is observed but the resolution is worse than expected or the spectrum is otherwise distorted:
-
Check the detector temperature and the heat sink.
First, make sure that cooling is properly enabled. Observe the temperature the detector reaches.
Second, check the heat sink. The thermoelectric cooler draws up to 1.5 W from the detector to the XR-100 box. If this box is thermally isolated, it will heat up and the detector will heat up.
Third, inspect the Be window on the detector. There is vacuum inside the detector. If the seal on the window is damaged and air enters the detector will not cool fully. -
Look for interference noise. Connect the XR-100CR output to its signal processor and this to an oscilloscope. Remove all X-ray sources.
Look for periodic noise pick-up on the scope by changing the time-base dial on the scope back and forth. If you find any periodic signal on the scope (other than the Reset Waveforms), try to eliminate its source or place the XR-100SDD away from the pick-up area. Any periodic signal detected on the scope will degrade the resolution of the XR-100SDD.
The XR-100 produces very small signals which are susceptible to electromagnetic interference. PLACE THE XR-100CR AWAY FROM ANY COMPUTER TERMINAL OR CRT MONITOR. KEEP THE XR- 100CR DETECTOR AWAY FROM MAGNETIC FIELDS. -
If you are using an Amptek signal processor and/or power supply refer to its User Manual and to Amptek’s “Troubleshooting guide” and to Amptek’s “Grounding and shielding” application note.
-
IF ANY QUESTIONS REMAIN CONTACT THE FACTORY FOR FURTHER ASSISTANCE AND RETURN PROCEDURES. If you are using an Amptek signal processor please follow the instructions for “Saving a .MCA file and obtaining diagnostic data,” which can be found on your installation CD.
Warranty and Technical Questions
9.1 Warranty
AMPTEK, INC. warrants to the original purchaser this instrument to be free
from defects in materials and workmanship for a period of one year from
shipment. AMPTEK, INC. will, without charge, repair or replace (at its option)
a defective instrument upon return to the factory. This warranty does not
apply in the event of misuse or abuse of the instrument or unauthorized
alterations or repair. AMPTEK, INC. shall not be liable for any consequential
damages, including without limitation, damages resulting from the loss of use
due to failure of this instrument. All products returned under the warranty
must be shipped prepaid to the factory with documentation describing the
problem and the circumstances under which it was observed. The factory MUST be
notified prior to return shipment. The instrument will be evaluated, repaired,
or replaced, and promptly returned if the warranty claims are substantiated. A
nominal fee will be charged for unsubstantiated claims. Please include the
model and serial number in all correspondence with the factory.
9.2 Technical Questions
- Please refer to http://amptek.com/technical-support/
- Please have the model and serial numbers of your Amptek device(s) available. Please have available a description of the signal processing and other electronics used with the device.
- Contact Amptek at amptek.sales@ametek.com or +1 781-275-2242.
X-Ray Fluorescence Process Example: Titanium Atom (Ti = 22)
Amptek K and L Emission Line Lookup Chart
Amptek, Inc
14 DeAngelo Dr, Bedford MA 01730 USA
Phone: +1 781 275 2242
Fax: +1 781 275 3470
Email: Amptek.Sales@ametek.com
Web: www.amptek.com
Certificate #: CU 72072412 01
Tested to: UL 61010-1: 2004 R7 .05
CAN/CSA-C22.2 61010-1: 2004
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>