BRETHALYZER AC-020 Portable Breath Alcohol Tester User Manual

June 24, 2024
BRETHALYZER

BRETHALYZER Logo User’s Manual

AC-020 Portable Breath Alcohol Tester

AC-020 is a sleek and compact designed breathalyzer which can connectable with smartphone app. via Bluetooth BLE function, so possible to manage a data and share by SMS, SNS and e-mail.
Furthermore, it is usable without connecting with smartphone as “Standalone” mode so handy and useful for personal checking.
The enhanced quality check function “Self Diagnosis System (SDS)” on an app. will verify the quality of breathalyzer’s sensor and will make sure the accurate and reliable test result.

Components

Contents
Breathalyzer, Zipper case, Mouthpieces(x5), AA Batteries(x2), Outer box, and User’s manual
How to use : WITHOUT paring iSOBER application

  1. Press the power button.
    If you keep pressing the power button, a number of testing after calibration will be shown as toggling displaying
    e.g. : 6 times tested after calibration
  2. Warming up until you can see ‘bLo(blow)’ sign.
  3. Blowing through the mouthpiece into the unit for 4 seconds.
  4. The result will be displayed for 10 seconds and shut off automatically.

*Passive test operation:
Press the power button once at ‘bLo(blow)’ status, then result will be displayed as P(PASS) or F(FAIL) according to preset limit value
How to use : WITH paring iSOBER application

  1. Download iSOBER application from the App store or Google play and log in.
  2. Turn on Bluetooth mode of your mobile device to be pairing with AC-020 unit.
  3. Power on your AC-020 unit then confirm “APP” showing on the display.
  4. Press the test button on an app then unit will be ready to blow status with showing ‘bLo(blow)’
  5. Blowing through the mouthpiece into the unit for 4 seconds following guides of App.
  6. If camera mode is on, camera of your mobile device will take a picture while blowing it. *if you want to verify your actual alcohol testing, camera mode will be a useful.
  7. App. will give you information like test result, test date & time, model of unit, sensor status, GPS location.
  8. Also user can leave a comment about testing as well as sharing via SMS,e-Mail and social networking service.
  9. Until the unit displaying “APP”, user can perform the test in a row using iSOBER application.

Caution
a. Please Wait at least 20 minutes after eating or drinking before testing. Alcohol remaining in the mouth, or even excess saliva, may interfere with testing. This includes alcohol-based mouthwash or similar products.
b. Do not blow smoke, saliva or other contaminants into the mouthpiece, as the sensor may get damaged.
c. Do NOT use AC-020 as a tool to drink and drive.
d. Keep out of extreme temperatures and out of the reach of children.
e. Please do not use manganese battery. Use alkaline battery ONLY.

Specification

Sensor Type Fuel Cell sensor
Battery AA Size Alkaline Battery 2EA
Weight 118g (Including batteries)
Size (mm) 113(Height) X 59(Width) X 22(Thickness)
Indication of Test result Display

0.00 ~ 5.00     ‰
0.00 ~ 0.50     %BAC
0.00 ~ 2.50     mg/l,
.000 ~ .500     %BAC (3 decimal point)
000 ~ 500       mg/100mL
Warming up time| 5 sec ~ 120 sec
Time for analysis| Within 5 seconds (at 0.050%BAC, 24°C)
Accuracy| ±10% (at 0.050%BAC)
Working / Storage temperature| 5°C ~ 40°C / 0°C ~ 50°C
Mouthpiece| MP1000
Calibration interval| Every 1,000 tests or 12 months (whichever comes first)
Certificate| CE RED, FCC, Bluetooth SIG
Relative Humidity| 40 % R.H. to 80 % R.H.
Used Place| Indoor
Altitude| less than 2 000 m
Additional Feature| BLE 4.2
Reminder for calibration
Compatible with mobile application iSOBER

Warning Display

| The breath sample was not strong enough for measurement.
After warming automatically, please blow strong and steady during sampling.
---|---
| The battery power is low. Please replace the batteries.
| If you test over 1,000 tests or 12 months, AC-020 will display “CAL” sign.
Press the power button once then unit goes to warm up for the test.
| Sensor error – Please contact an authorized service center
| AC-020 is Bluetooth paired with iSOBER application. It will be last 5 minutes.

Warranty

The manufacturer warrants the product to be free from defects in workmanship or material (excluding recalibration: recalibration is not covered by warranty) under normal use for TWELVE (12) months from the date of purchase. The manufacturer’s obligation under this warranty is limited to replacing, adjusting or repairing the unit if returned along with proof of purchase. This warranty is void if the unit has been tampered with or maliciously damaged.
Recalibration & service’
Please contact local distributor

FCC Instructions

FCC Compliance Statement This device complies with part 15 of the FCC rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
FCC Interference Statement This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to correct the interference by one of the following measures:

  • Reorient or relocate the receiving antenna.
  • Increase the separation between the equipment and receiver.
  • Connect the equipment into an outlet on a circuit different from which the receiver is connected.
  • Consult the dealer or an experienced radio/TV technician for help.

FCC Caution
Any changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate this equipment. This transmitter must not be co-located or operating in conjunction with any other antenna or transmitter.BRETHALYZER Logo

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals