SPEED WOLF Optical Power Meter User Manual
- June 7, 2024
- SPEED WOLF
Table of Contents
SPEED WOLF Optical Power Meter
Model Introductions
PM630-A:
Measurement range: -70~+10dBm,calibrated
wavelength:850nm、1300nm、1310nm、1490nm、1550nm、 1625nm
PM630-C:
Measurement range: -50~+26dBm,calibrated
wavelength:850nm、1300nm、1310nm、1490nm、1550nm、 1625nm
Features
-
High measurement accuracy and display resolution
-
Quick response and measurement: It can display fast-tracking and real-time measurement of optical power when output power changes
-
Wide measurement range: it can provide 80dB of optical power measurement range
-
Six calibrated wavelength: 850nm、1300nm、1310nm、1490nm、1550nm and 1625nm(other wavelength can be provided according to customer’s request)
-
Provide absolute optical power measurement and relative power measurement.
-
Real-time monitoring and displaying the battery level
Warning notice is available when the battery is low -
Auto-off function
-
Backlight LCD display
-
AAA batteries which is more convenient to carry
-
User self-calibration function
Technical Specifications
Item | PM630-A | PM630-C |
---|---|---|
Measurement range dBm | -70~+10 dBm | -50~+26 dBm |
Wavelength range(nm) | 800~1700 | |
Calibrated wavelength | 850nm、 1300nm、 1310nm、 1490nm、 1550nm、 1625nm | |
Detector | InGaAs | |
Accuracy | <±3% (-10dBm、22℃) | |
Resolution | Linearity:0.1%,Non-linearity:0.01dBm | |
Working temperature | -10℃~+50℃ | |
Storage temperature | -20℃~+70℃ | |
Relative humidity | 90%(+30℃) | |
connector | FC/PC SC/PC, (2.5mm universal interface)( ST as Optional) | |
Power supply | 3pcs AAA batteries | |
DC power supply | Optional accessory which require special order | |
Working hours | >120 hours(without backlight) | |
Weight(g) | 160g | |
Dimension(mm) | 130×66×27 |
Layout
-
Wavelength : Display the wavelength, such as 850nm、1300nm、1310nm、1490nm、1550nm and 1625nm.
-
Power value: Displays the current output power under absolute test mode
-
Unit: mW、uW、nW、dBm、dB.
-
Battery level : Indicate the current battery level.
When the icon is blank, it means the power is too low, please change the batteries. -
Auto-off :Activate the 10-minute auto off function
-
ON/OFF
-
Wavelength select
-
VFL:VFL status
-
DBm/w: Absolute power
-
Optical detector
-
Visual Laser(650nm)
-
VFL display
Function Keys
-
ON/OFF: press ON/OFF to turn it on. And press it for 3 seconds to turn it off. Under power-on
-
mode, press this key shortly to activate or deactivate the 10-minute auto off function. The default setting is auto-off function ON when start the meter. Operators can press ON/OFF shortly to close the auto-off function.
dBm/W: press dBm/W to enter absolute measurement mode. Press it repeatly to switch the display power unit between W and dBm -
λ:Switch the wavelength from 6 calibrated wavelengths, the current choice of wavelength will appears on the LCD
-
VFL :press VFL to switch the VFL status, ON Glint or OFF. At the same time, the VFL status will be simultaneously displayed in the upper right corner of the LCD screen.
-
Backlit is on when turning on the unit. It will turn off one minute later. Operators can press any key to make backlit for another one minute.
Operation
Turn on/off
Press ON/OFF shortly to turn on the unit. Press ON/OFF for a few seconds to turn it off and press this key slightly to activate or deactivate the auto-off function.
Absolute power measurement
- Insert optical signal (Insert the pigtail properly), turn on the meter
- Choose the wavelength via λ key. If the tested wavelength is not exactly same to the one in the power meter, then choose a close one from the power meter. For example, 1300nm wavelength is near to 1310nm, we can choose 1310nm to substitute 1300nm.
- After the above steps, the LCD will show the test power value.
User self-calibration function
During the operation, when operators need to calibrate the meter to be accordance with other meters. So our product offers operators self-calibration function, the following steps will tell you how to do:
- Press power key & dBm key at the same time, the instrument goes into manual calibration mode (at this time the four row units are all displayed).
- Press λ key to choose the wavelength for calibration.
- Now connect it with a pigtail which we already know its power and press REF or dBm/W to increase or decrease the power value by 0.05d every time. Until the LCD display the pigtail’s power (Or the difference gap less than 0.3db).
- Short press power key to save calibration data (backlight will flash once).
- Repeat above operation for other wavelength.
- After recalibrate each wavelength, long press power key to exit calibration mode.
Auto off function
The auto off is on when turning on the meter. Press ON/OFF to activate or deactivate this function.
Visual Fault Locator function
- Press VFL key, the VFL indicator is on, the locator output 650nm red laser.
- Press the VFL key again, the locator and indicator are blinking at the same time, the frequency is about 2Hz, which is convenient to check line faults.
- Press the VFL key the third time, then the VFL indicator is off, and the locator laser is off.
Replace battery
- If the battery level is very low, please turn off the unit immediately and replace the batteries
- Take out the batteries if not in use for a long period of time (like one month)
- NOTE: do not charge the non-rechargeable battery, dangerous!
Maintenance
- Keep all optical connectors and surfaces free from oil, dirt or other contamination to ensure proper operation.
- Keep using the same type of connector
- Please cover the dust cap when not in use to keep the connector clean.
- Carefully plug in or out the adapter
- Regularly clean the connector.
- Take out the battery when not in use for a period of time.
Trouble-Shootings
Problem | Reason | Solution |
---|---|---|
Faint LCD screen | Lower power | Replace battery |
No backlight | Lower power | Replace battery |
No display when turning on | Lower power/other | Turn on again or replace |
battery
Fail to turn on| Lower power| Replace battery
No changing on LCD screen| Lower power| Replace battery
In sensitive LCD screen| Dirty or polluted connector| Use the correct
connector and clean it.
Warranty Period:
12 months from the purchasing date
Warranty clause:
- Under the warranty, we would repair the problems caused under the normal operation free of charge. Note: remember to show us the warranty card when repair
- For the following situation, we have to charge certain cost:
- Out of the warranty;
- Fail to provide the warranty card
- Alter and omit the warranty card
- Wrong operation including the problem caused by human actors, abnormal working environment and so on.
- Problems or damages are not caused by the products quality
- Do not operate it according to this manual.
- For the following situation, we do not repair it
- Damaged seal label.
- Take apart it without permission
- Products from other manufacture
Packing List
- PM630 Optical Power Meter ……1pc 2)
- Soft case …… 1pc
- User Manual ……1pc
- Battery(AAA, transportation permit) ………3pcs
Read User Manual Online (PDF format)