Dwyer HTDL-20/30 Series High Temperature Data Logger Instruction Manual
- June 3, 2024
- Dwyer
Table of Contents
Dwyer HTDL-20/30 Series High Temperature Data Logger
Series HTDL-20/30 High Temperature Data Logger
NOTICE
Before the device is submerged, screw on the cap tightly to make sure that
the logger is properly sealed.
SOFTWARE INSTALLATION
-
Download at www.dwyer-inst.com from the software tab on the product page.
-
Follow onscreen instructions.
-
Click Install.
NOTICE A message may appear stating that the software has not passed Windows Logo Testing. This software has been tested and works properly. Click Continue anyway if this window appears.
CONNECT THE LOGGER
- Place Data Logger into the Model HTDL-DS docking station.
- Insert one end of the cable into the docking station and the other end into the PC.
SPECIFICATIONS
- Range: -328 to 500°F (-200 to 260°C).
- Memory Size: 65,536 readings.
- Accuracy: 0.18°F (0.1°C) @ 68 to 284°F (20 to 140°C); 0.54°F (0.3°C) @ -4 to 67.98°F (-20 to 19.99°C); 0.72°F (0.4°C) @ -40 to -4°F (-40 to –20°C).
- Resolution: 0.02°F (0.01°C).
- Temperature Limits: -40 to 284°F (-40 to 140°C).
- Sampling Method: Stop on memory full or continuous recording.
- Sampling Rate: Selectable from 1 s to 24 hrs.
- Computer Requirements: Windows XP SP3 or later.
- Power Requirements: 3.6 V 1/2 AA ER14250SM lithium metal battery, installed functional, user replaceable.
- Battery Life: 1 year (approx.).
- Interface: Docking station and USB cable.
- Housing Material: 316 SS.
- Weight: 4.2 oz (120 g).
START THE LOGGER
- To start a configured run, select custom start in the device tab.
- From the menu choose the desired Reading Rate, the preferred method for starting the logger and the alarm settings.
- Click the start button.
- The logger will stop recording when the device is full or when Stop Device is selected.
DOWNLOAD DATA
- To download data, connect the logger to the PC.
- From the Device tab, select download.
- Data is presented graphically and can be exported to Excel for further analysis under report options.
NOTICE
Starting the device erases all data currently stored in the logger.
Note: For more information on using the HTDL-DS software, please consult the Software Manual.
BATTERY
The battery is user replaceable, and the typical battery life is 1 year. To
replace the battery, unscrew the bottom of the data logger and remove the old
battery. Insert the new battery and screw the cap back on. Make sure the cap
in on tight before submersing the data logger. Fast logging will considerably
shorten the battery life. To preserve battery life, it is recommended to use
the longest practical sampling rate, and when the logger is not in use, select
Stop Device from the Device menu.
MAINTENANCE/REPAIR
Upon final installation of the Series HTDL-20/30, no routine maintenance is
required. The Series HTDL-20/30 is not field serviceable and should be
returned if repair is
needed. Field repair should not be attempted and may void warranty.
WARRANTY/RETURN
Refer to “Terms and Conditions of Sales” in our catalog and on our website.
Contact customer service to receive a Return Goods Authorization number before
shipping the product back for repair. Be sure to include a brief description
of the problem plus any additional application notes.
Windows® is a registered trademarks of Microsoft Corporation.
Phone: 219-879-8000
Fax: 219-872-9057
www.dwyer-inst.com
e-mail: info@dwyermail.com