Ez Life Infrared Thermometer BSX906 Instruction Manual

June 6, 2024
Ez Life

Ez Life Infrared Thermometer BSX906

‘Please thoroughly read the Instruction Manual before using the unit
‘Please save this instruction Manual for further reference
Version: A/O
NO REF No.:8187M90601
Revision Date:2020.03.25

Foreword

Thank you for purchasing the Infrared Thermometer.
The Infrared Thermometer is a high-quality product incorporating the latest technology and tested in accordance with international standards. With its unique technology, the Infrared Thermometer can provide a stable, heat- Interference-free reading with each measurement. The infrared thermometer shows the temperature of the subject by measuring the heat radiation from the forehead.

Some Helpful Guidelines For Best Results When Using the Infrared Thermometer

  • Call your doctor when your child has diarrhea, vomiting, or changes in appetite. lethargy (sluggishness) or is unusually sleepy.
  • It is important to know each individual’s normal temperature when they are well. This Is the only way to accurately diagnose a fever. Take multiple readings when healthy to determine normal temperature.
  • A child’s normal temperature can be as high as 37.7°C or as low as 36.1*C. Re-measure with a standard digital thermometer for conformation, especially on infants (It is recommended to use a digital thermometer rectally on Infants and on sleeping Infants if there is a question about the measurement. Be sure to note this unit reads 0.5’C lower than a rectal digital measurement).
  • This product Is a Class II medical device and belongs to the internal power supply equipment, type BF application part, the protection grade is ordinary equipment, non
  • AP/APG. intermittent operation equipment. the EU is classified as Class Ile.

Please read these instructions carefully before using this Instrument and keep the Instructions and the thermometer In a safe place.

Advantages of the Infrared Thermometer Quick measurement

The inflated technology allows moo securement of bOdytereporature in intron 3 seconds.

Appropriate Medical Application
The infrared thermometer provides either forehead or Object temperature mode. The range of the forehead temperature is from (32.0-43.0)°C 1(93.2-109.4)°F and the Object temperature is from (5.0-95.0)°C 1(41.0-203.0)°F which can meet the application demands normally.

Accurate and Reliable Measurement
Due to the unique probe assembly construction, the advanced infrared sensor, and the complete calibration process this unit can offer a very accurate and reliable temperature measurement comparable to one taken from the oral site with a regular digital thermometer.

Gentle and Easy to Use

  • The Infrared Thermometer is non-invasive. A measurement can be taken even while a child is sleeping.
  • Design is easy to hold and use.
  • The Infrared Thermometer is less threatening to a child than a rectal thermometer and easier to use than other methods.

Multiple Reading Recordings
Users can record the last 20 readings when entering the Memory'” Tracker recall mode, enabling efficient tracking of temperature history and variations.

Multiple Reading Recordings
Users can record the last 20 readings when entering the Memory'” Tracker recall mode, enabling efficient tracking of temperature history and variations.

Safe and Hygienic

  • No risk of broken glass or mercury ingestion.
  • Completely safe for use on children.

Fever Indications
When the measurement reading exceeds 38.0°C (100.4°F), please go to the hospital for a medical examination as soon as possible.

How the Infrared Thermometer Measures Body Temperature
Objects in nature radiate infrared rays if it, s temperature above absolute zero (-270 °C). The energy of the infrared rays and temperature are proportional. With this relationship, The infrared thermometer calculates the temperature by measuring the infrared radiation intensity of the object.

Intended Use/Indications for Use
The infrared thermometer shows the temperature of the subject by measuring the heat radiation from the forehead. It is suitable for people of all ages.
The Infrared thermometer is intended for both professionals and consumers, the patient is the intended operator.

Important Safety Instructions

  • Never use the thermometer for purposes other than those it has been intended for. Please follow the general safety precautions when using on children.
  • Never immerse the Infrared Thermometer into water or other liquids (not waterproof).
  • For cleaning and disinfecting please follow the instructions in the ‘Cleaning and Storage” section.
  • Keep the instrument away from direct exposure to the sun and keep it in a dust-free, dry area at a temperature between (15-40)°C.
  • Do not use the thermometer if there are signs of damage on the measuring sensor or on the instrument itself. If damaged, do not attempt to repair the instrument! Please contact the dealer.
  • This Infrared Thermometer consists of high-quality precision parts. Do not drop the instrument. Protect it from severe impact and shock. Do not twist the instrument or the measuring sensor, otherwise, it will not be able to measure body temperature accurately or unable to measure body temperature.

WARNING:

  • Use of this Infrared Thermometer is not intended as a substitute for consultation with your physician or pediatrician.
  • The thermometer is not waterproof! NEVER immerse into liquids!
  • Please consult your doctor if you see symptoms such as unexplained irritability, vomiting, diarrhea, dehydration, changes in appetite or activity, seizure, muscle pain, shivering, stiff neck, pain when urinating, etc., even in the absence of fever or who exhibits a normal temperature, may still need to receive medical attention.
  • People who are on antibiotics, analgesics, or antipyretics should not be assessed solely on temperature readings to determine the severity of their illness.
  • Temperature elevation may signal a serious illness, especially in adults who are old, frail, have a weakened immune system, or neonates and infants. Please seek professional advice immediately when there is a temperature elevation and if you are taking temperature for whom are:
  1. Neonates and infants under 3 months;
  2. Over 60 years of age;
  3. Having diabetes mellitus or a weakened immune system (e.g. HIV positive, cancer chemotherapy, chronic steroid treatment, splenectomy).
  4. Bedridden (e.g. nursing home patient, stroke, chronic illness, recovering from surgery)
    5)A transplant patient (e.g. liver, heart, lung, kidney)
  •  Fever may be blunted or even absent in elderly patients.
  • This thermometer is not intended for pre-term babies or small-for-gestational-age babies. This thermometer is not Intended to interpret hypothermic temperatures.
  • Do not allow children to take their temperatures unattended.
  • Do not modify this equipment without the authorization of the manufacturer.
  • Please do not service and maintain the thermometer when it is in use, for example replacing the battery, placing something on the forehead, etc.
  • This product contains small parts such as(batteries, battery cover, and screws)that may cause suffocation if swallowed by children.

PRECAUTIONS:

To avoid an Inaccurate measurement, note the following:

  • 1 hour is required for the infrared thermometer to warm from the minimum storage temperature between uses until the infrared thermometer is ready for its INTENDED USE when the ambient temperature is 20 °C;
  • 1 hour is required for the 1 hour is to cool from the maximum storage temperature between uses until the 1 hour is ready for its INTENDED USE when the ambient temperature is 20 °C;
  • Before and after use, please keep the sensor and probe cavity clean:
  • Please use the thermometer in a stable temperature environment. When the ambient temperature fluctuates greatly (from indoor to outdoor), please leave it for about half an hour to measure;
  • Do not start measuring body temperature immediately after measuring extremely high or low temperature, please leave it for ten minutes to measure;
  • Patients should stay in steady-state room condition for at least 30 minutes.
  • Don’t remove the measurement device from the measuring area before hearing the final beep.
  • Don’t take a measurement while or immediately after nursing a baby.
  • Don’t use the Infrared thermometer in high humidity environments
  • Patients should not drink, eat, or be physically active before/while taking the measurement.
  • Before placing the thermometer sensor onto the measurement area, remove dirt, hair, or sweat.
  • Always take the temperature in the same location, since the temperature readings may vary from different locations.
  • In the following situations, it is recommended that three temperatures in the same location be taken and the highest one
    taken as the reading:
  1. Newborn infants in the first 100 days.
  2. Children under three years of age with a compromised immune system and for whom the presence or absence of fever is critical.
  3. When the user is learning how to use the thermometer for the first time until he/she has familiarized himself/herself with the instrument and obtains consistent readings.

Product Description

Ez Life Infrared Thermometer BSX906 - FIG

The above picture is only for reference. Please use the material object as the standard.

  1. Start button: Measuring the forehead temperature and objects temperature.
  2. Setting: This product supports a total of four settings options, which are temperature unit, operating sound, fever indication point, temperature offsets.

Installation and Use

6.1 Initial settings
6.1.1 Battery installation

Install new battery

  1. Pull out the battery cover on the bottom of the thermometer
  2. Install two “AAA’ batteries correctly, and pay attention to the direction of the battery’s positive and negative electrodes.
  3. Close the battery cover(as the picture below).

Warning:
a) Please do not use this thermometer inflammable and explosive anesthetic gas, oxygen, nitrous oxide, etc.
b) To ensure the performance accuracy of this product and prevent equipment failure. please do not use it in a humid place.
c) Caution: Do not install waste batteries on the thermometer and do not input external power.
d) When not in use for a long time, please remove the battery to prevent leakage.
e) Never dispose of waste batteries in a fire.
f) Please properly dispose of used batteries in accordance with local regulations to avoid contamination.

6.2 Measurement Instructions
6.2.1 Body temperature
6.2.1.1 Short press the start button to turn on, all the icons are displayed on the screen, the buzzer two beeps to complete the self-test, and enter the temperature measurement state. 6.2.1.2 Point the thermometer to the center of the human forehead at a straight angle, at a distance of 0-3 cm. In about half, a second user may hear the sound-beep” and see the temperature value. The measurement is done.

After the measurement is done, if there is no operation within 30 seconds, it will automatically shut down.
6.2.1.3 Read the recorded temperature from the LCD display.

NOTE
a)Before and after use, please keep the sensor and probe cavity clean;
b) Please use the thermometer in a stable temperature environment. When the ambient temperature fluctuates greatly (from indoor to outdoor), please leave it for about half an hour to measure;
c) Do not start measuring body temperature immediately after measuring extremely high or low temperature, please leave it for ten minutes to measure;
d) When the measurement object comes from a place with a large temperature difference from the measurement environment, it should stay in the test environment for at least half an hour; e) Try not to measure with forehead blown, dripping, sweating, or applying cosmetics.
f) Do not take body temperature within 30 minutes after exercise, bathing, or eating.

6.2.2 Objects temperature
6.2.2.1 Short press the start button to turn on, all the icons are displayed on the screen, the buzzer two beeps to complete the self-test, and enter the temperature measurement state. 6.2.2.2 Point the thermometer straight to the object surface under test at a distance of 0-3 cm. Press the Start button, and in about half a second you will hear the sound ‘beep’ with the reading displayed and the measurement is done.

NOTE:
a) The temperature here refers to only the surface temperature of the object under test, not its core temperature;
b) The infrared ray radiation rate of different materials varies, which may affect the measurement results. The defaulted radiation rate of this device is 0.95. For example, the measured temperature of stainless steel may be much lower than the actual temperature. Please lake care to prevent from getting burned in such case.

6.3 Function settings
This product supports a total of five settings options, such as temperature unit, operating sound, fever indication point, temperature offsets, and measurement mode.

Icon

| Function| ‘Start button|

Defaulted value

---|---|---|---
IMF| Read Memory| View records| Latest 20 records
F1| Temp. offsets| Revise offsets| 0.0°C/°F
F2| Warning point| Revise warning point| 38.0°C
F3| Temp. unit| Switch temp. unit| °C/°F
F4| Operating sound| On/ Off| On

6.3.1 Read the Memory
When the thermometer turns off, short press the Set button, then you will see on the screen, then press the Set button to view the latest 20 records. Press the Start button to exit from the memory view.

6.3.2 Temperature offset setting: F1
When the thermometer turns on,long-press the Set button until appears F1 on the screen, press the Start button consecutively to set the range of temperature offsets/deviation between (-2-2)°C.

6.3.3 indication temperature point setting: F2
Press once more the Set button to view the current indication temperature point after F2 appears on the screen, and then press the Start button to modify the temperature point.

6.3.4 Temperature unit setting: F3
Press once more the Set button to view the current temperature unit after F3 appears on the screen, and then press the Measure button to switch the temperature unit(°C/°F).
6.3.5 Operating sound: F4
Press once more the Set button, you will see F4 and the operating sound icon (ta)) at the current status on the screen. Press the Start button to switch on/off the operating sound which indicates “OPEN” and “CLOS” to alternately.
6.3.6 Save and exit:
Press once more the Set button, you will see the word SAVE on the screen, and then save the settings and exit. Once you enter the settings menu, if you want to change some options, you can press consecutively. Set the options in order and save them. If you turn off the unit during operation, new settings won’t be saved. When the Settings menu is open, the thermometer will not take any measurement.
NOTES: a) Body temperature mode is used to measure the temperature of the human body with dynamic compensation of ambient and forehead temperature; b) Object temperature mode takes the temperature of the object surface, which does not mean the temperature of the human body under test; c) The temperature offsets/deviations can be corrected according to the difference between the target measurements and actual temperature due to different ambient temperature, measuring distance, conditions of individual skin. The temperature correction range is (-2.0-2.0)°C/°F, and the defaulted value is 0.0 °C/°F. For instance, if the measured temperature is 36.2°C, while the actual temperature of the object under test is 37.0°C.you can enter the interface of F1 to increase the offsets up to 0.8°C. After setting, the user can get the final result came as the actual temperature.
6.4 The temperature of the measured object is out of range
6.4.1Body temperature measurement: a) When the measured value is lower than 32°C, “Lo” will be displayed and there will be four beeps; b) When the measurement temperature is higher than 43.0°C, “Hi” is displayed and there will be four beeps.
6.4.2 Objects temperature measurement: a) When the measured value is lower than 5.0 °C, ‘to” will be displayed and there will be four beeps; b)When the measured temperature value is higher than 95.0°C, “Hi” is displayed and there will be four beeps.
NOTES: 1) When the ambient temperature is lower than 15 °C or higher than 40 °C, the thermometer will not be able to measure. It will display to” or “Hi” and shut down. 2) The patient is the intended operator. 3) The functions (measuring forehead temperature, measuring objects temperature, the buzzer, etc.) that patients can safely use.

Error messages and troubleshooting

Troubles

|

Solutions

---|---
The display indicates o” of”| 1. Check the target you are going to measure. In the case he is facing a fan, sweating, wet from rain, or making up with cosmetics, the accuracy cannot be ensured;
2. Check the temperature offset settings. The factory default is 0.0°C/°F;
3. Check the working environment. The environmental change imposes great impacts on the measured results. If the change varies too much, or the device has just been used to measure objects of extra-high or low temperature, a large deviation may occur. Measurement should be performed in a relatively stable and constant environment where the device stays for a least half an hour.
4. Check the measuring distance(53cm).
Buttons failure| 1. Re-install the batteries;
2. Check whether the device stays in the setting menu.
No display or abnorm at the display| Re-install the batteries.
No operating sound| Check whether the operating sound is turned off.
Shutdown soon after turned on| Re-install the batteries.
The display indicates play| Excessive ambient temperature changes, please leave it for at least half an hour before using it

Maintenance, Repair, and Calibration
Please check carefully before use to ensure that the temperature sensor and the probe cover are clean, make sure the batteries are properly installed and the power is sufficient, otherwise:
1) Care and Cleaning
Wipe the sensor mirror with a cotton ball with a small amount of 75% alcohol. Make sure that the sensor mirror is neat and tidy, otherwise, the measurement accuracy may be affected.

NOTES:
• Do not use a wet cotton ball to wipe the sensor mirror to prevent liquid from entering the thermometer.
• Never use abrasive cleaning agents, thinners, or benzene for cleaning and never immerse the instrument in water or other cleaning liquids.
2) Replacing the Battery
If the Low Voltage indicator appears on the display, it is time to replace your batteries. The thermometer takes two AAA batteries. Slide open the battery cover as shown and remove batteries. Replace the batteries being sure to align properly ( as indicated inside the battery compartment.
NOTES: If you use the thermometer every day, it is recommended to clean the shell, temperature sensor at least before the first use and after the last use of the day, and ensure that the battery is properly installed and fully charged; If the interval is longer, it is recommended to clean the shell, temperature sensor, and probe cover before and after each use, and remove the battery after use.
3) When not in use, please keep the Infrared thermometer and its accessories in the packing box. If you do not use the thermometer for a long time, please remove the batteries.
4) In order to prevent the Infrared thermometer from damage, please comply with the following items, and improper use will result in the loss of warranty services:

  • Please do not place the Infrared thermometer in an environment of high temperature, dampness, full of water vapor, or direct sunlight.
  • Please refer to the specification table for details. If stored and used outside the specified range of temperature and humidity, the system may not be able to achieve the claimed specifications.
  • Please do not disassemble the Infrared thermometer.
  • Please do not allow the Infrared thermometer to be subjected to strong impact or vibration (for example, drop the Infrared thermometer on the ground).
  • Please do not use volatile liquid to clean the Infrared thermometer. Please use a soft and dry cloth to clean the outer shell and battery.
  • Clean the thermometer with a soft dry cloth. Do not use any abrasive or volatile.
  • Please do not use water to clean or moisten the temperature, and when necessary, absorbent cotton dipped with alcohol (75% alcohol) can be used to wipe it gently, so as to conduct disinfection.
  • Drop, collision and bending, etc. may reduce the performance of the battery; if the battery is found abnormal, such as packaging damage and battery pack deformation, etc., do not use these batteries.
  • The manufacturer will make available on request circuit diagrams, component part lists, descriptions.
  • The patient can perform changing batteries.
  1. calibration instructions, or other information that will assist service personnel to repair those parts of ME equipment that are designated by the manufacturer as repairable by service personnel.
    6) For customer service:
    Warranty service, please refer to “warranty card”.
    7) Calibration
    When using, if the measurement results are biased and the accuracy is reduced, Please contact the service center for calibration or repair. Please do not disassemble the instrument. Or it is recommended the performance should be checked every 1 year or after repair. Please contact the service center.

IX. Icon meaning

Definition

| Icon|

Status Description

---|---|---
Power level| | With display| Replace battery
No display| Power level is high enough
Sound| | With display| Buzzer is on with sound
No display| Buzzer is off, no sound
Temperature unit|

°C

| Celsius

°F

| Fahrenheit
Readings|

8888

| Indicating measured results

Icon

| Status Description
---|---
| WEEE Symbol
| Type BF applied part

| Refer to instruction manual/booklet.

LOT

| Batch code

SN

| Serial number

| Manufacturer
EC REP| Authorized Representative in the European Community
| Date of manufacture
| Use-by date
| Distribution packages shall not be exposed to sunlight
| Contents of the distribution packages are fragile therefore they shall be handled with care
---|---
| Distribution packages shall be kept away from rain and be kept in dry conditions
|  Distribution packages shall be stored, transported, and handled within temperature limits indicated
| This is the correct upright position of the distribution packages for transport and/or storage.
| A maximum number of identical transport packages/items may be stacked on the bottom package, where “n” is the limiting number.

AP/APG

| It cannot be used in a mixture of flammable anesthetic and air, oxygen, or nitrous oxide.

Disposal treatment statement

The product manufacturing date can be seen in the qualification certificate, and the product use period is 5 years. If electronic products and batteries are directly thrown into the trash can, they can cause harm.
NOTES:

  • Dispose of the device, components, and optional accessories according to applicable local regulations. Unlawful disposal may cause environmental pollution.
  • If electronic products and batteries are directly thrown into the trash can, they can cause harm to the environment. At the end of the infrared thermometer, please do not throw it into the trash can, and please dispose of it according to local law.

Warranty

This product has been produced with the greatest care for normal, household use. Following directions carefully will ensure dependable operation. If for any reason (other than misuse or normal battery operation) you are dissatisfied with your thermometer, we will repair or replace it, at our option, at no charge during the 1 year warranty period. Modification to the product by the consumer is not authorized and voids the warranty. If the thermometer does not function properly, first check the battery. Replace if necessary. Batteries and packaging are excluded from the warranty. If something goes wrong, please contact BSX. If repair is required under the warranty period, the product should be returned with proof of purchase to the BSX service center. The purchaser assumes responsibility for the proper care and use of the product in accordance with the printed operating manual. The purchaser or user must make his or her judgment as to when to use the product and the length of use. Keep the operation manual available as it is your guide to safe, efficient operation. The product comes with guarantees the compliance with European Union Law. You are entitled to a replacement or refund for a major failure and for compensation for any other reasonably foreseeable loss or damage. You are also entitled to have the goods repaired or replaced if the goods fail to be of acceptable quality and the failure does not amount to a major failure.

Technical Specifications

Items

|

Parameter

---|---
Product name| Infrared Thermometer
Model name| BSX906
Measuring Range| Forehead: 32.0°C-43.0°C/89.6°F-109.4°F objects :
5.0°C 95.0°C/41.0°F-203.0°F
Display Resolution| 0.1°C/0.1°F
Measurement Accuracy| ±0.3°C(±0.6°F)
Safety classification| Class Ila, Internally powered, Type BF applied part
Fever indication| Normal body temperature: 32.0-38.0°C, the screen normally displays the body temperature value, a buzzer sounding a “beep”.
fever:>38.0°C, the screen normally displays the body temperature value, the buzzer sends out four beeps.
Waterproof grade level| IP22
Display screen| LCD
Sound setting| Available
Measurement Time| Within 3s
The service life of Battery| 10000 times
Memory capability| 20 reading sets
Operating environment range| Temperature:(15-40)°C/(59.0-104.0°F);
Humidity: (20-85)%RH,Atmosphere: (70-106)kPa
Storage/Transport environment range| Temperature: (-20-55)°C/(-4-131°F); Humidity: RH593%,Atmosphere: (70-106)kPa
Automatic Switch-off| Approx 30 seconds after the fast measurement has been taken.
Battery| 2″ 1.5 V AAA Batteries
Weight| 87g (with battery), 73g (without battery)
---|---
Lifetime| 5 years
Remarks:
1. The measurement site of the infrared thermometer is different from the contact thermometers. When measuring, please ensure that the forehead is not blocked by objects, and the measuring distance is 53 cm.
2. Clinical accuracy characteristics and procedures are available from the manufacturer on request.

If there is any performance improvement, the above specifications are subject to change without notice.
Electronic compatibility warning and statement Notes:
The equipment complies with the relevant electromagnetic compatibility requirements of EN 60601-1-2:2015; Users shall conduct installation and use according to the electromagnetic compatibility information provided by random files; Portable and mobile RF communication equipment may affect the performance of the device, so strong electromagnetic interference shall be avoided when in use, such as close to mobile phone and microwave oven, etc.;
Warning: The equipment shall not be close to or stacked with other equipment when in use, and if it is necessary to close to or stack with other equipment when in use, it shall be observed and verified that whether it can run normally under its configuration; Except for the cable sold by the equipment manufacturer as the spare parts of internal components, the use of unspecified accessories and cables may lead to emission increase or anti- interference reduction of the equipment.
— potential allergic reactions to accessible materials used in my equipment.

Guidance and manufacturer’s declaration – electromagnetic emissions

The model BSX906 is intended for use in the electromagnetic environment specified below. The customer or the user of the model BSX906 should assure that they are used in such an environment.

Emissions test

| Compliance|

Electromagnetic environment – guidance

RF emissions CISPR 11| Group 1| The model BSX906 uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions CISPR11| Class B| The model BSX906 is suitable for use in the domestic establishment and in the establishment directly connected to a low voltage power supply network that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2| Not applicable
Voltage fluctuations/
flicker emissions
IEC 61000-3-3| Not applicable
Guidance S Declaration — electromagnetic Immunity

The model SMOG Intended for use in the electromagnetic environment is specified below. The customer or the user of model 85)(906 should assure that they are used in such an environment.

Immunity test

| IEC 60601 test level| Compliance level|

Electromagnetic environment – guidance

Electrostatic discharge (ESL)) IEC 61000 4-2| se kV. ill kV contact 12 kV. /1 kV. ill kV.

I n kV ale

| Rafi kV. ta kV contact *2 kV. to kV. t8 kV.* 151O/ air| Floors should be wood. concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least: 30%.
Electrical fast transient/burst IEC 61000-4-4| 12kV for power supply Mak : I kV for Input/| output lines Not applicable| Mains power quality should be
that of a typical commerc ia l or
environment
Surge IEC 61000.4 5| ill kV line toline 12 kV line to earth| Nol ePPIloafik 7| Mains power quality should be %hal of 8 IOU! Commercial or hospital environment.
V011age dIpS. Short Interruptions and voltage variations On power supply input lines IEC 61000-4-11.| < 5% U. 1 ,95% dip in U,) for 0.5 cycle <5% %4**

ran% dip in U,) for 1 cycle

70% U, 130% dip in U.) for 25730 Cycles .5% kJ, (.95 %dip In U.) for 516 sec

| Not applicable| Mains power quality should be that of a typical commercial or hospital environment. If the user or the m00183%908 requires continued operation during power mains interruptions. Ills recommended that the model 135)(906 be powered from an uninterruptible power supply or a battery.
Power frequency (50/60 Hz) magnetic field
IEC 61000 4 8| 3A/m. 30 Arm| 3 Arra 30Nm| Power frequency magnetic fields should be at levels characteristic era typical location in et)the commercial or hospital environment.
NOTE: U, is the a.c. mains vol age prior to application of the test level.
Guidance 84 Declaration – Electromagnetic immunity

The model BSX906 is intended for use in the electromagnetic environment specified below. The customer or the user of the model 6SX906 should assure that it is used in such an environment.

Immunity test

| IEC 60601 test level| Compliance level|

Electromagnetic environment – guidance

Conducted RF
IEC 81000
4 6
Radiated RE IEC 61000-4-3| 3 Vrms
150 kHz to 80 MHZ
8 Vrms In ISM bands
10 V/m
80 MHz to 2.7 Ghz
385MHz-5785MHz Tests specifications for ENCLOSURE PORT IMMUNITY to RF wireless communication equipment (Refer to table 9 of IEC 80801.1-2:2014)| Not applicable
Not applicable
10 V/m
80 MHz to 2.7GHz
385MHz-5785MHz Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communication equipment (Refer to table 9 of IEC 60601-1-2:2014)| Portable and mobile RF communications

equipment should be used no closer to any pan of the model BSX908. including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter
Recommended separation distance
d•43,5P/JSP” d41.2xla ni 80 MHz to 800 MHZ d•2.3xP” 800 MHz to 2.7 Ghz where P is the maximum output power rating of the transmitter In watts (W) according to the transmitter manufacturer and d Is the recommended separation distance in meters (m).

Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey. a should be less than the compliance level in each frequency range.b

Interference may occur In the vicinity of equipment marked with the following.

NOTE 1 At 80 MHz and 800 MHz. the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.
Field strengths from fixed transmitters, such as base stations for radio (cellular/ cordless) telephones and land mobile radios, amateur radio, AM, and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy.

To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the model BSX906 Is used exceeds the applicable RE compliance level above, the model BSX906 should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the model BSX906.
° Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.

Recommended separation distances between portable and mobile RF communications equipment and the model BSX906.

The model BSX906 is intended for use in an electromagnetic environment in which radiated RF disturbances is controlled. The customer or the user of the model BSX906 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the model BSX906 as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter W
| Separation distance according to the frequency of transmitter m
150kHz to 80MHz
d=1.242“| 80MHz to 800MHz
d=1.2k p•”| 800MHz to 2,5GHz
d=2.3kP”
0.01| 0.12| 0.12| 0.23
0.1| 0.38| 0.38| 0.73
1| 1.2| 1.2| 2.3
10| 3.8| 3.8| 7.3
100| 12| 12| 23
For transmitters rated at a maximum output power not listed above, he recommended separation distance d in me ers (m) can be estimated using the equation applicable to the frequency of the transm tier, where P is the maximum output power rating of the transmitter in watts (W) accordable to the transmitter manufacturer. NOTE 1 At 80 MHz and 800 MHz. the separation distance for the higher frequency range applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

Recommended separation distances between portable and mobile RF communications equipment and the model BSX906.

The model BSX906 is intended for use in an electromagnetic environment in which radiated RF disturbances is controlled. The customer or the user of the model BSX906 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the model BSX906 as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter W
| Separation distance according to the frequency of transmitter m
150kHz to 80MHz
d=1.242“| 80MHz to 800MHz
d=1.2k p•”| 800MHz to 2,5GHz
d=2.3kP”
0.01| 0.12| 0.12| 0.23
0.1| 0.38| 0.38| 0.73
1| 1.2| 1.2| 2.3
10| 3.8| 3.8| 7.3
100| 12| 12| 23
For transmitters rated at a maximum output power not listed above, he recommended separation distance d in me ers (m) can be estimated using the equation applicable to the frequency of the transmit tier, where P is the maximum output power rating of the transmitter in watts (W) accordable to the transmitter manufacturer.

NOTE 1 At 80 MHz and 800 MHz. the separation distance for the higher frequency range applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals