AQUAREAD AP-700 Multiparameter Water Quality Probe Instruction Manual
- June 1, 2024
- AQUAREAD
Table of Contents
- AQUAREAD AP-700 Multiparameter Water Quality Probe
- Specifications
- Product Usage Instructions
- FAQs
- Introduction
- What’s in the Box
- Battery Installation and Care
- Overview of the Operating System
- Connecting an AP-2000
- Taking Measurements
- Depth Measurement
- Memory Mode
- Setup & Install
- RapidCal Calibration Method
- After Use
- General Probe Maintenance
- pH/ORP Electrode Calibration and Maintenance
- DO Electrode Calibration and Maintenance
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
AQUAREAD AP-700 Multiparameter Water Quality Probe
Specifications
- Product Name: Multiparameter Water Quality Probe
- Document No.: 10401-00870
- Revision: W
- Date: March 7th, 2024
- Manufacturer: Aquaread
- Address: Bridge House Northdown Industrial Park, Broadstairs Kent, CT10 3JP ENGLAND
- Phone: +44 1843 600 030
- Website: www.aquaread.com
Product Usage Instructions
1. Introduction:
Before using the Multiparameter Water Quality Probe, read through this instruction manual carefully to ensure proper usage and understanding of the device.
2. What’s in the Box?
The package should include the Multiparameter Water Quality Probe, necessary cables, connectors, and any additional accessories specified in the product description.
3. Battery Installation and Care:
Follow the guidelines provided in the manual for installing and handling batteries to ensure optimal performance of the device.
4. Overview of the Operating System:
Learn about the initial setup process, language selection, and clock setup to operate the device efficiently.
5. Connecting an AP-2000:
Follow the instructions to properly connect the AP-2000 device for accurate measurements.
6. Taking Measurements:
Instructions on how to take water quality measurements using the Multiparameter Water Quality Probe.
7. Depth Measurement:
For Aquaprobes with a -D suffix only, follow the detailed guidelines for depth measurements, differential depth measurements, and depth calibration.
8. Memory Mode:
Understand how to utilize the memory mode feature for storing and accessing measurement data.
9. Setup & Install:
Guidance on setting up and installing the Multiparameter Water Quality Probe for efficient use.
FAQs
- Q: Can this probe detect harmful bacteria or toxins in water?
- A: No, this equipment cannot detect the presence of harmful bacteria or toxins. Always treat suspect water before consumption.
- Q: Is the water safe for consumption based on readings from this probe?
- A: Favorable indications from this equipment do not guarantee that water is safe for human or animal consumption. It is recommended to treat suspect water before consumption.
“`
Instruction Manual
for the
Aquaprobe® AP-700, AP-800 & AP-2000
Multiparameter Water Quality Probe
and associated
Aquameter®, Utilities & Accessories
Aquaprobe® firmware Revision 4.07 and Above Aquameter® firmware Revision 6.20
and Above
Document No. 10401-00870
Revision: W
Date: March 7th, 2024
Aquaread® Ltd
Bridge House Northdown Industrial Park
Broadstairs Kent
CT10 3JP ENGLAND
Phone: +44 1843 600 030 www.aquaread.com
Aquaprobe® AP-700, AP-800 & AP-2000 Instruction Manual
10401-00870 Rev W
PROPRIETARY NOTICE
The information disclosed within this document or separately supplied in
furtherance of this document includes proprietary rights of Aquaread® Ltd.
Neither this document nor the information disclosed herein or of a
supplementary nature shall be reproduced or transferred to other documents or
used for any other purposes, except as specifically authorised in writing by
Aquaread® Ltd. Copyright 2019 Aquaread® Ltd. All rights reserved.
SOFTWARE LICENCE AGREEMENT
By using the Aquameter®, its associated probes and PC based software
(AquaLink), you agree to be bound by the terms and conditions of the following
software licence agreement. Please read this agreement carefully.
Aquaread® Ltd grants you a limited license to use the software embedded in the
devices (the “firmware”) in binary executable form in the normal operation of
the products. Title, ownership rights and intellectual property rights in and
to the Software remain in Aquaread® Ltd.
You acknowledge that the Software is the property of Aquaread® Ltd and is
protected under British copyright laws and international copyright treaties.
You further acknowledge that the structure, organisation and code of the
Software are valuable trade secrets of Aquaread® Ltd.
You agree not to decompile, dissemble, modify, reverse assemble, reverse
engineer or reduce to human readable form the Software or any part thereof or
create any derivative works based on the Software.
RECYCLING
Any product bearing this symbol must be handed over to an applicable
collection point for the recycling of electrical and electronic equipment. By
ensuring this product is disposed of properly, you will prevent possible
adverse consequences for the environment which could be caused by
inappropriate waste handling.
Note: For return for recycling, please contact Aquaread® Ltd for instructions
on how to return end-of-life equipment, accessories, and all auxiliary items
for proper disposal.
WARNING
Favourable indications made by this equipment do not guarantee water is safe
for human or animal consumption. This equipment can not detect the presence of
harmful bacteria or toxins. If in doubt, ALWAYS TREAT SUSPECT WATER PRIOR TO
CONSUMPTION.
Limitations of Liability
Aquaread® Ltd and its affiliates specifically disclaim liability for any and
all direct, indirect, special, general, incidental, consequential, punitive or
exemplary damages including, but not limited to, loss of profits or revenue or
anticipated profits or revenue or expenses incurred arising out of the use or
inability to use any Aquaread® product, miscalibration, loss of data or
product failure, even if Aquaread® Ltd and/or its affiliates has been advised
of the possibility of such damages or they are foreseeable or for claims by
any third party. Notwithstanding the foregoing, in no event shall Aquaread®
Ltd and/or its affiliates aggregate liability arising under or in connection
with Aquaread® products, regardless of the number of events, occurrences, or
claims giving rise to liability, be in excess of the price paid by the
purchaser for the Aquaread® product.
TRADEMARKS
Aquaread®, AquaPlusTM, Aquameter®, Aquaprobe®, AquaLinkTM and RapidCalTM are
all trademarks of Aquaread® Ltd.
Microsoft®, Windows® and Excel® are trademarks of Microsoft Corporation
GOOGLETM is a trademark of Google, Inc.
Aquaread® Ltd has no affiliation with Microsoft Corporation or Google, Inc.
© 2024 Aquaread® Ltd.
Page 2 of 141
Aquaprobe® AP-700, AP-800 & AP-2000 Instruction Manual
© 2024 Aquaread® Ltd.
Page 7 of 141
Aquaprobe® AP-700, AP-800 & AP-2000 Instruction Manual
10401-00870 Rev W
Introduction
This manual specifically covers the setup, operation, calibration and maintenance of Aquaprobe® models AP-700, AP-800 & AP-2000 running V4.07+ firmware, Aquameter® models running V6.20+ firmware, AquaLink V5.00+ PC software and associated Aquaprobe® accessories. If your Aquaprobe® or Aquameter® are running earlier firmware, the functionality may differ from that shown in this manual. In this case, contact Aquaread for an earlier version of this manual or return your equipment for firmware upgrade.
What’s in the Box
The Aquameter® is supplied with the following:
The Aquameter® unit. Quick release lanyard. Set of 5 AA Alkaline batteries.
USB Cable for downloading logged data to a PC. Cross-head screwdriver for
fitting the batteries and Probe maintenance. Getting started card for quick
reference.
The Aquaprobe® is supplied with the following:
Protective Sleeve End Cap. Calibration bottle filled with RapidCal Solution.
Spare calibration / rinse bottle. One mounting nut (pre-fitted). Getting
started card for quick reference. 25mL bottle of pH storage solution. Pot of
silicone grease. Spare Galvanic DO Membrane Cap (AP-700 & AP-800 only) 25mL
bottle of Galvanic DO filling solution (AP-700 & AP-800 only)
If using an AP-2000, you will also need an AP-2000 Extension Cable, which
should be purchased separately.
2.1. The Aquameter® and the Environment The Aquameter® is designed to be used
outdoors and is rated to IP67, that is to say it is waterproof but it is not
designed for submersion. In order to prevent accidental dunking or loss, a
lanyard is supplied.
Please note that the socket on the Aquameter® is only waterproof when the
associated plug is fitted. Without the plug fitted, water can enter the
socket. Damage caused by water ingress through the socket is not covered by
your warranty.
You may notice a small hole on the rear of the unit near the top. This is a
waterproof vent for the internal barometric sensor. Do not poke anything in
this hole! Doing so will cause major damage to the vent’s waterproof membrane
and invalidate your warranty.
2.2. The Aquaprobe® and the Environment The Aquaprobe® AP-700 and AP-800
models are designed to be fully submerged in water and are rated to IP68, that
is to say, they are rated for continual immersion to a depth of 10 meters, and
short term immersion (less than 12 hours) to a depth of 50 meters.
© 2024 Aquaread® Ltd.
www.aquaread.com
Page 8 of 141
Aquaprobe® AP-700, AP-800 & AP-2000 Instruction Manual
10401-00870 Rev W
The Aquaprobe® AP-2000 and AP-2000-D models are designed to be fully submerged in water and are rated to IP68, that is to say, they are rated for continual immersion to a depth of 30 meters, and short term immersion (less than 12 hours) to 100 meters.
2.2.1. Important Notes Regarding Galvanic Corrosion
Galvanic corrosion, sometimes also known as bimetallic corrosion, is an
electrochemical process in which one metal corrodes preferentially when it is
in contact with a dissimilar metal in the presence of an electrolyte (such as
water). A similar galvanic reaction is exploited in batteries to generate an
electrical voltage. When installing an Aquaprobe, it is important to recognise
and avoid the possibility of creating a situation where galvanic corrosion can
occur.
All Aquaprobes are made primarily from hard-galvanised marine grade aluminium. All the time that the aluminium body of the Aquaprobe is insulated from any other type of metal, there should be no problem with corrosion (unless the Probe is placed in strong acids or alkalis, for which it is not designed).
Corrosion problems can occur if the Aquaprobe is mounted near to, and connect to a dissimilar metal, such as stainless steel. Typical examples of this are mounting the Probe inside a steel pipe or suspending it from a steel dock on a steel wire. In these situations, a steel-aluminium battery is created that will generate a voltage of around 0.5V and lead to the corrosion of the metal with the lower potential, in this case the aluminium of the Aquaprobe.
It is important to avoid creating a situation where galvanic corrosion can occur in your installation. The easiest way to do this is by electrically insulating the Aquaprobe from the supporting structure. This can be done by suspending the Aquaprobe on a Mylar or Nylon rope rather than a conductive steel wire.
If the Aquaprobe is to be mounted inside a steel pipe, wrap a good thickness of electrical insulation tape around the top and bottom of the Probe to form `fenders’ in order to prevent the Probe making physical contact with the inside of the pipe. When doing this, ensure that you do not block any of the holes in the Probe’s sleeve, which are essential for water flow and correct operation of the Probe.
If the above techniques are not possible and a metal-to-metal connection is unavoidable, a sacrificial anode must be attached to the Aquaprobe. Sacrificial anodes are manufactured from zinc and are designed to slide onto the threaded connector section of the Aquaprobe and to be held on by the mounting nut. Zinc is much more active than aluminium and will therefore corrode first leaving the Aquaprobe undamaged.
Sacrificial anodes are, as the name suggest, sacrificial, so will need replacing periodically.
Sacrificial anodes are available for all models of Aquaprobes. Please contact Aquaread Ltd for more details.
2.3. Important Information about the Probe Sleeve & Sleeve End Cap
The Aquaprobe® is constructed with an aluminium sleeve surrounding the
delicate sensing electrodes. The Sleeve can be easily removed by unscrewing to
allow cleaning of the individual electrodes, however, the Probe sleeve forms
an integral, working part of the Probe’s measurement system, and MUST be
fitted for correct operation.
© 2024 Aquaread® Ltd.
www.aquaread.com
Page 9 of 141
Aquaprobe® AP-700, AP-800 & AP-2000 Instruction Manual
10401-00870 Rev W
Probe sleeves are also specifically matched to the Probe with which they were supplied. If you have more than one Aquaprobe®, be sure not to mix the sleeves up between the Probes. Doing so may seriously affect the performance of the EC electrode.
All Aquaread® Optical Electrodes are incredibly sensitive. For example, the Turbidity electrode is capable of measuring between 0 and 3000NTU with an internal resolution of greater than 0.1NTU. This means that the electrode is able to detect changes in turbidity that are less than 0.003% of the full range! The other optical electrodes have a similar level of sensitivity. It follows, therefore, that in order to provide stable, repeatable readings, the environment in which the measurements are made must be completely stable and repeatable.
For this reason, the Aquaprobe® is constructed with a matt black aluminium sleeve and end cap that enclose the sensing electrodes and provide a closed, constant condition, non reflective measurement chamber.
This is essential for the correct calibration and operation of all types of optical electrodes.
A diagram of the Aquaprobe’s measurement chamber is shown here. Please note, the design of the End Cap may vary depending upon the age and model of your Aquaprobe.
In order to obtain consistent results, the measurement chamber created within
the Aquaprobe® must remain physically constant during both calibration and
measurement.
If the optical electrode is calibrated under one set of conditions then used to measure under another set of conditions, the readings will naturally be erroneous, especially at low concentrations.
A perfect example of this is calibrating with the end cap removed then measuring with the end cap fitted (or vice-versa). By changing the physical characteristics of the measurement chamber, you also change the calibration and response of the electrode.
Another particular problem when trying to measure very low concentrations is air in the form of both visible and microscopic bubbles. These act like tiny prisms and can refract and reflect both the excitation light and the return signal being measured.
The photograph to the right was taken in a calibration tube after fresh water was poured in. The bubbles are clearly visible in the light beam.
2.4. Top Tips for successful measurements using optical electrodes Always keep the measurement chamber and electrode lenses clean. Always fit the sleeve and end cap during both calibration and measurement.
© 2024 Aquaread® Ltd.
www.aquaread.com
Page 10 of 141
Aquaprobe® AP-700, AP-800 & AP-2000 Instruction Manual
10401-00870 Rev W
Always allow the readings to settle completely during both calibration and measurement.
Always try to eliminate air bubbles by agitating the Probe after insertion both during calibration and measurement.
Always calibrate and zero the electrode as close to your sample temperature as possible. This is especially important with the Ref-Oil electrode.
Always zero the optical electrodes just prior to use in clean water (bottled still mineral water is ideal) then deploy without disturbing the measurement chamber. This is especially important when using the Turbidity and Ref-Oil electrodes.
2.5. About the Lanyard The lanyard supplied with the Aquameter® may, at first,
appear to be a little long. This is intentional. In order to keep the Meter
out of the way whilst your hands are full, the lanyard has been made long
enough to wear round your neck and over your shoulder so the Meter sits on
your hip.
The extra length also allows the meter to be held in a comfortable position in
front of you during normal use. In order to prevent you being dragged into the
water in the event of the Probe cable becoming snagged, the lanyard includes a
quick-release clip.
© 2024 Aquaread® Ltd.
www.aquaread.com
Page 11 of 141
Aquaprobe® AP-700, AP-800 & AP-2000 Instruction Manual
10401-00870 Rev W
Battery Installation and Care
The Aquameter® requires five AA size batteries. To install the batteries, loosen the two screws on the centreline of the rear of the meter and remove the battery compartment lid. Following the battery polarity markings inside the battery compartment, insert five AA cells then replace the compartment lid and tighten the screws.
3.1. Choice of Battery Type
Alkaline or rechargeable batteries may be used, but never mix battery types in
the meter. If you choose to use rechargeable batteries, we recommend Energizer
2500mAh (or greater) Nickel-Metal Hydride cells, which are widely available.
If the Meter is to be out of use for a long period, remove the batteries to
prevent damage due to possible leakage.
3.2. Battery Life
A set of fresh alkaline cells will give over 20 hours use in the AM-200 GPS
Aquameter®. A fully charged set of 2500mAh NiMH cells will give up to 40 hours
use in the AM-200 GPS Aquameter®. Please be aware however that alkaline
battery capacities are extremely temperature dependant. The figures quoted
throughout this manual for battery life assume a temperature of 21°C. Battery
life can be significantly shorter (by up to 50%) at lower temperatures.
3.3. Battery Charging
During the charging process, batteries generate heat and vent gasses, and must
never be charged inside a sealed unit. Because the Aquameter® is a sealed
unit, we do not allow charging in-situ. Batteries must be removed and charged
with a suitable battery charger outside the Meter. We recommend the use of one
of the Energizer range of NiMH chargers.
3.4. Battery Condition Icon
On all the main Aquameter® screens, a battery condition icon is displayed in
the top left corner. The icon shows full when the batteries are fresh, and
gradually empties as the batteries are used. When the batteries need
replacing, the empty battery icon will flash on and off. If you ignore this,
the Meter will automatically switch itself off when the battery voltage
becomes too low for reliable operation.
When using rechargeable batteries, the battery icon will not show completely full, even with freshly charged cells. This is due to the fact that rechargeable batteries are only rated at 1.2V per cell compared to 1.5V per cell for alkaline batteries. This indication does not affect battery life. The icon will simply sit at the ¾ full mark for a longer period of time.
3.5. Battery Saver Functions
The Aquameter® is designed to switch off automatically if you do not touch any
of the keys for 30 minutes. The only exception to this is if you have
activated the Automatic Data Logging feature. In this case, the Meter will
continue to operate until either the memory is full or the batteries go flat.
The display on the Aquameter® incorporates a white backlight to improve visibility in lowlight conditions. As on a mobile phone, the backlight switches on each time a key is pressed, and stays on at full brightness for 15 seconds. After 15 seconds, the backlight will fade to half brightness. After a further 15 seconds the backlight will switch off.
During normal operation, if you want to activate the backlight without changing the Meter function, simply press the ESC key.
© 2024 Aquaread® Ltd.
www.aquaread.com
Page 12 of 141
Aquaprobe® AP-700, AP-800 & AP-2000 Instruction Manual
10401-00870 Rev W
Overview of the Operating System
The operating firmware in the Aquameter® has been designed for simple, intuitive use. Similarly, a great deal of development work has been put into simplifying and automating the calibration procedures in the Aquameter® in order to allow normal field operatives (as opposed to trained lab technicians) to achieve quick and accurate results.
If you are used to operating a mobile phone or programming audio/visual equipment using a remote control, you should feel at home with the familiar up/down left/right arrow shaped navigation keys and central OK key.
The tree structure behind the MENU key should also be very familiar. Each item on the menu leads to a sub menu and then either onto further menus or final choices. Each branch of the menu system is navigated using the arrow keys. At each point, selections can be made by either pressing the OK key or the right arrow key.
To reverse along a branch of the menu system, use the ESC (escape) key or left arrow key. After a short time, you should be able to navigate around the entire menu system at speed using just the four arrow keys. If, at any time, you leave the Meter in one of the sub-menu screens, it will automatically back out to the main operating screen after 15 seconds.
4.1. Initial Switch On, Language and Clock Setup
To switch the meter on or off, briefly press the red key. Do not hold it down.
The meter contains a clock and is capable of operating in several different
languages. When switching on for the first time, you must select an operating
language and set the clock. The first screen you will see is the Language
Selection Screen.
English Italiano Francais Portugues Deutsch Malaysia Espanol Indonesia
To select a language, move the cursor around the list using the arrow keys. To
enter your selection, press the OK key or the right arrow key.
The next screen to be displayed is the Time & Date Setting Screen.
Time & Date Time:15:46:37
Date:15/Jun/17
To set the time and date, use the arrow keys to move the cursor around the screen. Use the up and down arrow keys to adjust values. When the time and date are correct, press the OK key. Don’t worry if you make a mistake first time round. You can easily get back to these screens later through the MENU key.
© 2024 Aquaread® Ltd.
www.aquaread.com
Page 13 of 141
Aquaprobe® AP-700, AP-800 & AP-2000 Instruction Manual
10401-00870 Rev W
Connecting an AP-2000
The AP-2000 is designed to connect to the Aquameter® using an AP-2000
Extension Cable. The cable is built-in on the AP-700 & AP-800 models. The
AP-2000 Extension Cable features high-pressure metal connectors, which
incorporate several O-ring seals at the Probe end. Prior to first connection,
the seals must be lubricated using the silicone grease supplied. Apply a
generous smear of grease to the O-rings where indicated above. Be careful not
to get any grease inside the connector near the gold contacts. A small smear
of grease should also be applied to the thread on the Probe to allow easy
tightening of the collar.
To connect the Extension Cable to the AP-2000, align the coloured dot on the
AP-2000 with the Aquaread® logo on the plug body, then press the plug into the
socket and tighten the retaining collar fully. DO NOT TWIST THE CONNECTOR BODY
WITH RESPECT TO THE PROBE. Once the AP-2000 has been connected to the
Extension Cable, the Aquameter® can be connected.
5.1. Connecting the Aquameter® Always ensure the Aquameter® is switched off
prior to connecting or disconnecting an Aquaprobe®. Align the Aquaread® logo
on the plug body with the red on/off switch on the Aquameter®, then press the
plug into the socket and tighten the retaining collar.
Once the Aquaprobe® is connected to the Aquameter®, switch the Meter on by
briefly pressing the red on/off switch. The Aquameter® should detect the Probe
and automatically start displaying readings.
Taking Measurements
The Aquaprobe® includes a pH/ORP electrode, which is kept moist by a storage
cap. Remove the storage cap by pulling the red lanyard marked `Remove Before
Use / Replace After Use’ straight down. Do not use a twisting motion to remove
or replace the cap as this can unscrew the electrode from the Probe body.
Rinse any salty deposits from the pH/ORP electrode with fresh water.
Fit the protective Sleeve End Cap into the end of the Probe sleeve. Switch the
Aquameter ® on and immerse the Aquaprobe® in the sample water, making sure
that the water level covers the minimum immersion depth groove halfway up the
Probe sleeve.
TIP: Occasional application of a smear of silicone grease or similar lubricant
to the protective Sleeve End Cap O ring and the inside rim of the Probe sleeve
will make fitting and removal of the Cap easier.
If the Aquaprobe® is connected correctly, the meter will read the Probe’s
serial number and model number, then will automatically configure itself to
display only those readings the current Aquaprobe® is capable of taking.
Initial Probe readings will be displayed on the meter’s screen along with the
current GPS status. The initial data screen for the GPS Aquameter® in
conjunction with the Aquaprobe® is shown below.
© 2024 Aquaread® Ltd.
TEMP:018.5°C
ORP:0415.2 mV
pH:06.48
GPS:Acquiring
www.aquaread.com
Page 14 of 141
Aquaprobe® AP-700, AP-800 & AP-2000 Instruction Manual
10401-00870 Rev W
Left/right arrows at the bottom corners of the screen indicate further data screens are available. To access these screens, simply press either the left or right arrow keys. Any value that is out of range or unavailable will be displayed as dashes. The other four screens available with the standard AM-200/AP-2000-D combination are shown below.
DO:098.7%
EC:6541µS/cm
TDS:3271mg/L
GPS:3D Pos
SAL:03.57 PSU
SSG:01.3t
GPS:3D Pos
BARO:1013mb
DEPTH:01.75m
Hit [OK] to zero
GPS:3D Pos
Lat:N51 °21.498 Long:E001°24.323 Alt:00050M 1013mb
Sats in use:09
6.1. What Does It All Mean?
The screens above show the full default range of readings for the
AM-200/AP-2000-D combination. If you are using a different Meter/Probe
combination, you may have fewer screens to choose from and the readings may
appear in a different order to facilitate logical screen layouts. If an
asterisk (*) character is flashing just below the battery symbol, this
indicates that Auto Data Logging is switched on. See Automatic Data Logging in
section 8.
The table below explains the readings and indicates which to expect with each Meter/Probe combination.
Prefix TEMP
pH ORP GPS DO EC TDS SAL SSG BARO DEPTH Lat Long Alt
Meaning Probe Temperature
pH (Acidity/Alkalinity) Oxidation Reduction Potential GPS Status Dissolved
Oxygen Electrical Conductivity Total Dissolved Solids Salinity Sea Water
Specific Gravity Barometric Pressure Depth above / below zero datum Latitude
Longitude Altitude above Sea Level
Units
C or F pH or pHmV mV See section 6.5 %Sat & mg/L µS/cm or mS/cm mg/L or g/l
PSU or ppt t mb or mmHg Meters / Feet Degrees & Mins Degrees & Mins Meters
or Feet
Available On All Probe Types
All Probe Types All Probe Types AM-200 + All Probe Types All Probe Types All
Probe Types All Probe Types All Probe Types All Probe Types AM-200 + All Probe
Types – D models only AM-200 + All Probe Types AM-200 + All Probe Types AM-200
- All Probe Types
Note: the Depth page is not displayed with the standard Aquaprobe®. In this case,
© 2024 Aquaread® Ltd.
www.aquaread.com
Page 15 of 141
Aquaprobe® AP-700, AP-800 & AP-2000 Instruction Manual
10401-00870 Rev W
barometric pressure is displayed after the altitude (Alt) at the end of line 3 on the position and altitude screen, with no prefix (as shown [1013mb] on the above screen example).
Items in the Units column marked with an asterisk (*) can be selected as alternative units of measurement in the Settings Menu (see section 9 Setting Units of Measurement). Items in the Units column marked with a dagger () are auto-ranging, i.e. when the values become too large to display, the units of measurement automatically re-scale.
The EC field can be replaced by its reciprocal value, RES (Resistivity), if selected in the Settings Menu. If selected, readings will be displayed in either cm or Kcm, depending on the value. See section 9 Setting Units of Measurement for more details.
6.2. Trend Indication
To the right of each reading, (except position, BARO and Depth), a trend
indication is given. This consists of either an upwards facing arrow (which
indicates the numeric value of the reading is rising), a downwards facing
arrow (which indicates the numeric value of the reading is falling) or a two-
headed arrow, which indicates a stable reading. Readings are judged to be
stable when the variation over a ten second period drops below 1%.
6.3. Global Stability Indication
In addition to the individual trend indications, there is a global stability
indication, which is displayed when all readings are stable. This takes the
form of a flashing double headed arrow which is displayed at the start of the
third line of the display.
When taking a set of readings, gently stir the Probe, or raise and lower it in the sample (if there is no natural water flow) until the global stability icon appears. The initial display of the global stability icon will be accompanied by a double beep. When this occurs, all values are stable and ready for reading or saving.
6.4. Temperature Compensation
The electrochemical properties of all solutions change with the solution’s
temperature. In addition, the response of electrochemical measuring electrodes
change with temperature. It is a fundamental, practical requirement in the
field of water quality monitoring that test measurements taken at different
temperatures can be compared.
In order to facilitate this, the Aquaprobe® automatically applies corrections for temperature wherever required.
During three point calibration of the ISE electrodes, the variation in response of the electrodes due to temperature is automatically calculated. During measurement, the variation in response of the electrodes due to temperature is automatically compensated for.
During calibration of the EC electrode, the variation in the calibration
buffer solution due to temperature is automatically corrected for. During
measurement of EC, the readings can be
displayed without any temperature correction, corrected to 20C, or corrected
to 25C. See section 9 Setting Units of Measurement for more details.
During calibration of the DO electrode, variations due to temperature and air pressure are automatically compensated for. During the measurement of DO, temperature, air pressure and salinity are automatically compensated for.
© 2024 Aquaread® Ltd.
www.aquaread.com
Page 16 of 141
Aquaprobe® AP-700, AP-800 & AP-2000 Instruction Manual
10401-00870 Rev W
During calibration of the ORP electrode, the variation in the calibration
buffer solution due to temperature is automatically corrected for. During
measurement of ORP however, temperature corrections are not applied as the
correction factors are system and chemical dependent and are not easily
determined.
ORP potential measurements are mostly made to follow reactions rather than for
their own sake. The completion of an ORP reaction is normally accompanied by a
sharp change in the ORP millivolts reading. This change is usually much larger
than the errors induced by temperature side effects.
During calibration of the optical electrodes, variations in the calibration
solutions due to temperature are automatically compensated for. During the
measurement, temperature is automatically compensated for.
During calibration of the pH electrode, the small variation in the calibration
buffer solutions due to temperature is not compensated for due to the
differences in thermal coefficient between various buffer manufacturers. For
this reason, the three pH points should be calibrated as close to the buffer
manufacturer’s specified temperature as possible (usually 20ºC or 25ºC)
although a variation of up to +/-10ºC makes very little difference in reality.
During pH measurement, temperature variation is automatically compensated for.
6.5. GPS Reception The GPS version of the Aquameter® (AM-200) contains a
built-in GPS/GLONAS receiver and antenna. The antenna is situated at the top
of the case, just behind the Aquaread® Logo. For optimum signal reception, the
antenna must be able to `see’ a reasonably large amount of the sky. The GPS
receiver will not work indoors or when shielded from the sky by any solid
structure.
After switch-on, the GPS receiver will automatically start to search for
satellites. During this phase, the message GPS:Acquiring will be shown on the
bottom line of all the screens. As soon as three satellites are acquired, two
dimensional position (no altitude) will be calculated and the message GPS:2D
POS will be shown on the bottom line of the screens.
Once a fourth satellite is acquired, altitude will be calculated and GPS:3D
POS will be shown on the bottom line of the screens. With a good view of the
sky, position should be calculated within ninety seconds of switch-on. To see
your geographic position and the number of satellites in use, use the left or
right arrow keys to scroll to the Position page.
If you switch the meter on indoors, then carry it outside after several
minutes, there may be a considerable delay in acquiring satellites. In this
case, switch the meter off, then back on again to reset the acquisition
process.
© 2024 Aquaread® Ltd.
www.aquaread.com
Page 17 of 141
Aquaprobe® AP-700, AP-800 & AP-2000 Instruction Manual
10401-00870 Rev W
Depth Measurement
(Aquaprobes with a – D suffix only) Depth is measured in the Aquaprobe® by a pressure sensor mounted inside the body of the probe.
Depth is calculated by subtracting the barometric pressure being measured in
the Aquameter® from the water pressure being measured in the Aquaprobe®. The
pressure differential, once corrected for temperature and salinity (water
density), is directly proportional to depth.
The depth measurement system uses the EC sensor to detect when the probe has
been placed in water. All the time the probe is measuring an EC of zero, the
depth will read zero. As soon as an EC value is detected, the meter will start
to calculate depth. For this reason, it is important to ensure the Probe is
connected to the Meter and switched on prior to submerging the probe in water.
7.1. Taking Depth Measurements Connect the Probe to the Meter and switch on
prior to submerging the probe in water. Select the Baro/Depth screen as
illustrated below. The depth should be reading zero.
BARO:1013mb
DEPTH:00.00m
Hit [OK] to zero
GPS:3D Pos
If the depth is not reading zero (this is possible if the probe is wet and a low EC reading is registering), press the OK key. You will be asked to confirm by pressing OK again.
Slowly lower the probe into the water. As soon as the depth value starts to register, you can lower the probe more quickly.
7.2. Differential Depth Measurement
If you want to measure changes in depth, it may be more convenient to zero the
depth measurement once the probe has been submerged.
To do this, press the OK key whilst displaying depth, then confirm. The unit will now read positive or negative changes in depth from the current depth (zero datum).
If the values are positive, the water level has increased from the zero datum. If the values are negative, the water level has decreased.
Using the Automatic Data Logging feature detailed in the following section, it is possible to monitor water levels over a period of time for later recall.
7.3. Depth Calibration
The depth sensor is automatically re-zeroed each time the DO 100% point is
calibrated in free air. The depth sensor is factory calibrated at two
temperatures and two pressures and vary rarely needs to be recalibrated during
the lifetime of the Probe. If two point depth calibration is required, it must
be done using an AP-PC KIT and associated AquaCal PC Software. Full depth
calibration can not be performed using an Aquameter®.
© 2024 Aquaread® Ltd.
www.aquaread.com
Page 18 of 141
Aquaprobe® AP-700, AP-800 & AP-2000 Instruction Manual
10401-00870 Rev W
Memory Mode
8.1. Manually Saving Readings When you are happy that the readings are stable
(see section 6.3: Global Stability Indication), press the M+ key to snapshot
the readings along with the time, date, GLP (calibration) data and position
(GPS models only).
As each reading is saved, a numeric memory location `Tag’ will be briefly
displayed which you can note down. This Tag can be used to identify readings
at a later date, both on the Aquameter® and when using AquaLink software.
8.2. Recalling and Viewing Saved Readings To recall your readings, press the
MR key. On entering Memory Recall mode, the most recent Tag and set of
readings are displayed first along with the date and time the readings were
taken shown on the bottom line of the screen.
M TEMP:012.5°C M ORP:0415.2mV pH:08.21
02/Apr/17 15:04:01
During Memory Recall, an `M’ is flashed in the top left and right corners of
the screen alternatively with an up/down arrow and a left/right arrow. This is
to indicate that the Meter is in Memory Recall mode and that other screens can
be accessed using the arrow keys.
To see earlier readings, press the up arrow key. Just before each set of
readings is displayed, the Tag will be briefly displayed. To view all the
parameters within one set of readings, use the left/right arrow keys as
described earlier. To exit Memory Recall mode, press the ESC key. If no key is
pressed for 30 seconds, Memory Recall mode will be automatically cancelled.
8.3. Recalling GLP Data Each time a set of readings is added to memory, the
date of the last successful calibration of each electrode is also appended.
This is called GLP (Good Laboratory Practice) Data. In addition to the date of
the last successful EC calibration, the Calibration Standard value at which
the EC was calibrated is also displayed (see section 15.: EC / Temperature
Electrode Calibration and Maintenance for further details).
To view the last successful calibration date for each electrode for any
particular stored reading, enter Memory Recall mode, scroll to the reading you
are interested in using the up/down keys, then press the MENU key. The screen
below will be displayed.
GLP DATA >pH/REDOX(ORP)
DO/EC Aux Electrodes
Using the up/down keys, select the electrode you are interested in, then press either the OK key or the right arrow key. If, for instance, you selected pH/REDOX(ORP), the screen below would be displayed.
© 2024 Aquaread® Ltd.
www.aquaread.com
Page 19 of 141
Aquaprobe® AP-700, AP-800 & AP-2000 Instruction Manual
10401-00870 Rev W
PH 7.00 [31/Jan/17] PH 4.01 [07/Feb/17] PH 10.0 [07/Feb/17] ORP+250
[09/Feb/17] This tells you that the last successful calibration, prior to the
recorded reading being taken, was January 31st for the pH 7.00 point, February
7th for the pH 4.01 & pH 10 points and February 9th for ORP. If the date field
is dashed (==/===/==), this means the electrode was either not fitted or had
never been calibrated.
Pushing the left or right arrow keys will toggle this screen with the
calibration report screen.
PH7 Offset:+01.2mV PH4 Slope:56.8mV/pH PH10 Slope:56.3mV/pH ORP Offset:-02.6mV
The calibration report screen displays the calibration report values from the
last calibration.
To exit this screen press the ESC key.
8.4. Clearing the Memory The memory within the Aquameter® is capable of
storing 10,000 full sets of readings.
To clear the entire memory, switch the Meter off, hold down the M+ key, then
switch the Meter back on. A screen will be displayed asking you to confirm
your request. Press OK to clear the memory or ESC to cancel and return to
normal operation.
8.5. Automatic Data Logging If you want to save readings on a regular basis,
in order, say, to check water quality at a certain location over a period of
time, you can set the Meter to record readings automatically.
Readings can be logged for short periods with the Meter permanently displaying
readings, or for much longer periods in a Low Power Mode, where the Meter
switches itself off between readings in order to extent the battery life.
To activate Automatic Logging, press the MENU key. The Main Menu screen will
be displayed. Please note, the first item on the menu, ‘Clean Probe’, will
only be active if an Aquaprobe® AP-7000 (which has an automatic cleaning
system) is connected.
Clean Probe Auto Data Logging Calibration Setup & Install
Select Auto Data Logging by pressing the down arrow key then the right arrow key or the OK key. The Auto Data Logging screen will be displayed.
© 2024 Aquaread® Ltd.
www.aquaread.com
Page 20 of 141
Aquaprobe® AP-700, AP-800 & AP-2000 Instruction Manual
10401-00870 Rev W
Auto Data Logging
Interval:10 Mins Status:OFF
Using the arrow keys to navigate, set the desired logging interval anywhere
between 1 and 90 minutes.
To select Sub-Minute Logging Mode, move the cursor right again to the word
‘Mins’, then use the up/down arrow keys to toggle the setting to ‘Secs’. Now
the logging interval can be set anywhere between 2 and 90 seconds.
To select permanent display logging mode, set the Status to ON. To select Low
Power logging mode, set the Status to LOW POWER. Low Power mode is not
available in SubMinute Logging Mode.
To activate the selected logging mode, press the OK key then revert back to
the normal operation screen from the Main Menu by pressing the left arrow key.
To indicate that Auto Data Logging is switched on, an asterisk (*) character
will flash on and off just below the battery symbol on all the main reading
screens.
If permanent display logging mode was selected (Status set to ON), the Meter
will record a full set of data automatically at the set rate until either the
memory is full or the batteries go flat.
If Low Power Logging Mode was selected (Status set to LOW POWER), the Meter
will switch itself off 30 seconds after your last key-press. Thereafter it
will switch back on at the set rate, stay on for 30 seconds, log the data,
then switch back off again. This will be repeated until either the memory is
full or the batteries go flat.
If you press any key while the Meter is off between readings in low power
mode, the Meter will switch back on. If no further key is pressed, the Meter
will switch back off again after 30 seconds and resume Low Power Mode.
You can cancel Auto Data Logging at any time by going back into the screen
above and setting the Status to OFF. Auto Data Logging will also be cancelled
if you switch the Meter off manually.
© 2024 Aquaread® Ltd.
www.aquaread.com
Page 21 of 141
Aquaprobe® AP-700, AP-800 & AP-2000 Instruction Manual
10401-00870 Rev W
8.6. Battery and Memory Duration in Low Power Logging Mode Low Power Logging
Mode is specifically designed for long term data logging. In order to estimate
battery life and memory usage, the following table can be used.
The battery life figures quoted below are based on fresh, good quality
alkaline batteries at a Meter temperature of 21°C or over. Colder Meter
temperatures will drastically reduce the battery life. For example, at 5°C,
the battery life will be approximately half that quoted.
Logging Rate 90 mins 60 mins 45 mins 30 mins 15 mins 5 mins 1 min
Battery Life (at 21°C) 38 Days 36 Days 34 Days 30 Days 20 Days 10 Days 2 Days
Memory Duration* 625 Days 416 Days 312 Days 208 Days 104 Days 34 Days 6.9 Days
So, it can be seen that although the Meter has a maximum data capacity of 625 days, for logging rates above 15 minutes, fresh batteries would need to be fitted every 30 days or so in order to make use of the Meter’s full memory capacity.
Conversely, a logging rate of 2 seconds will fill the Meter’s memory on a single set of batteries (at 21°C or greater).
Useful Tip: If you want GPS data logged in association with your other data, ensure the Meter is positioned face up with a clear view of the sky.
8.7. Important Information About Memory Mode
When data is saved in the Meter, it is compressed in raw Probe format. In
other words, the same way that it came up from the Probe. When you recall the
data in Memory Recall mode, the data is decompressed, then processed for
display.
The advantage of this is that the readings will always appear in the current Meter configuration. For example, if you spent a day taking readings with the Meter set to read EC corrected to 25°C, then when you got back you really want to see EC corrected to 20°C or even raw EC, you can do this by simply changing the Meter settings (see section 9 Setting Units of Measurement).
The stored data can be displayed any way you want on recall. You are not limited to viewing the data in the same way it was logged. This is a major advantage and allows you to actually store and recall far more parameters than can be displayed at any one time.
The same rules apply when data is output to a PC running AquaLink Software via the USB cable. The data that is output is always as per the Meter’s current configuration. You can output the data as many times as you like in various Meter configurations.
© 2024 Aquaread® Ltd.
www.aquaread.com
Page 22 of 141
Aquaprobe® AP-700, AP-800 & AP-2000 Instruction Manual
10401-00870 Rev W
Setup & Install
To alter the way the Aquameter® displays readings, press the MENU key to get to the Main Menu, then choose Setup & Install. The Settings Menu will be displayed. Please note, the ‘Socket Assignment’ option on this screen is only accessible when a Probe is connected.
Time & Date Units Language Socket Assignment
9.1. Setting Units of Measurement
From this screen choose Units. The Units Menu will be displayed. Remember, you
can use just the arrow keys to navigate through the branches of the menus. You
don’t need to press OK or ESC at each level.
Units Menu EC/TDS
TEMP/pH/SAL BARO/Alt & Depth
At the Units Menu, you have a choice of which units you want to adjust. Choose
the first line if you want to adjust Electrical Conductivity or TDS. Choose
line 2 if you want to adjust Temperature, pH or Salinity. Finally, line 3 will
give access to Barometric Pressure, Altitude and Depth settings.
Moving the cursor right onto the first line will display the following screen.
Units EC:Ref 25°C
TDS Fact:0.65
The first option on this screen allows you to choose how the Meter displays
Electrical Conductivity. There are four options. EC can be displayed as
Absolute EC’ without any temperature correction [ABS EC], as
Specific EC’
referenced to 20C [Ref 20C], as `Specific EC’ referenced to 25C [Ref 25C] or
as a reciprocal of Absolute EC, which is Absolute Resistivity [ABS RES].
Finally, this screen allows you to set the factor that the Meter uses to
calculate Total Dissolved Solids from Specific EC. This is the TDS Fact: (TDS
= EC x TDS Fact) and can be set anywhere between 0.00 and 1.00. Default value
is 0.65.
© 2024 Aquaread® Ltd.
www.aquaread.com
Page 23 of 141
Aquaprobe® AP-700, AP-800 & AP-2000 Instruction Manual
10401-00870 Rev W
Selecting the second line of the Units Menu will display the following screen.
Units TEMP:°C
pH:pH SAL:PSU
The first option on this screen allows you to change the temperature display
between C and F.
The second option allows you to change the pH display between plain pH and
pHmV. Plain pH displays normal, temperature compensated pH values in the range
0 – 14.
pHmV displays the actual voltage being generated by the pH electrode in +/-
millivolts (mV) over a range of +/- 625mV. This is not temperature
compensated.
The last option on this screen allows you to choose between displaying
salinity in Practical Salinity Units (PSU), or parts per thousand (ppt), which
is the same as grams per litre.
Selecting the third line of the Units Menu will display the following screen.
Units BARO:mb
ALT:Metres
The first line allows you to choose between displaying Barometric pressure in millibars (mb) or in mm of mercury (mmHg).
The second line allows you to choose between displaying altitude and depth in
metres (M) or feet (F). Whatever units ALT is set to, DEPTH (-D models only)
will follow. Altitude is displayed with respect to mean sea level.
Depth is displayed with respect to the depth zero datum, which can be the
water surface or any point at which the depth has been zeroed. See section 7:
Differential Depth Measurement for further details.
9.2. AUX Socket Assignment (not applicable to the AP-700) The AP-800 features
one, and the AP-2000 features two AUX (axillary) sockets into which additional
electrodes may be fitted. AUX socket 1 can be fitted with either AP-2000
Optical electrodes or AP-5000/7000 type ISE (Ion Specific) electrodes. AUX
socket 2 can be fitted with AP-2000 type ISE electrodes only.
When an electrode has been fitted to an AUX socket (see appendix 3 for fitting
instructions), the socket must be assigned to the specific electrode type.
The Socket Assignment option is only available if the Aquameter® is connected
to a Probe. This is because the assignment data is held in the Aquaprobe®, not
in the Aquameter®.
When the Socket Assignment option has been selected, the following screen will be displayed.
© 2024 Aquaread® Ltd.
www.aquaread.com
Page 24 of 141
Aquaprobe® AP-700, AP-800 & AP-2000 Instruction Manual
10401-00870 Rev W
The numbers 1 6 represent the AUX socket numbers. Only socket 1 is available on the AP-800 and sockets 1 and 2 on the AP-2000. Unavailable sockets are shown as N/A. The additional sockets are available on larger Probes.
SOCKET ASSIGNMENTS 1:EMPTY l 4:N/A
2:EMPTY l 5:N/A 3:N/A l 6:N/A
Using the up and down arrow keys, select the AUX socket you wish to assign then move the cursor to the right by pressing the right arrow key. When the cursor has moved to the right of the AUX socket number, use the up and down arrow keys to select the appropriate electrode type.
The tables below show the available electrode options and the selection that should be made on this screen:
AP-2000/5000 type Optical Electrodes (AUX1 only)
Electrode Part No.
Function
2000-TURB
Turbidity
2000-CPHYLL
Chlorophyll
2000-BGA-PC
Phycocyanin (Blue-Green Algae PC)
2000-BGA-PE
Phycoerythrin (Blue-Green Algae PE)
2000-RHOD
Rhodamine WT Dye
2000-FSCEIN
Fluorescein Dye
2000-REFOIL
Refined Oil
2000-CDOM
CDOM/FDOM
Aquameter® Selection TURB Cphl
BGA-PC BGA-PE
Rhod Fcein R-OIL CDOM
AP-5000/7000 type ISE Electrodes (AUX1 only)
Electrode Part No.
Function
7000-AMM
Ammonium/Ammonia
7000-CHL
Chloride
7000-FLU
Fluoride
7000-NIT
Nitrate
7000-CAL
Calcium
Aquameter® Selection NH4 Cl F NO3 Ca2
AP-2000 type ISE Electrodes (AUX2 only)
Electrode Part No.
Function
2000-AMM
Ammonium/Ammonia
2000-CHL
Chloride
2000-FLU
Fluoride
2000-NIT
Nitrate
2000-CAL
Calcium
Aquameter® Selection NH4 Cl F NO3 Ca2
When the desired electrode type is showing, move the cursor back to the left
of the socket number then press OK to send the selection to the Aquaprobe®.
The socket assignments are stored in the Aquaprobe®. If you press the ESC key
whilst in this screen, any changes you have made will not be transferred to
the Aquaprobe®. Please note: changing an AUX
Socket assignment will clear all the calibration data for that socket.
If you subsequently remove an electrode, be sure to set the socket assignment back to EMPTY.
© 2024 Aquaread® Ltd.
www.aquaread.com
Page 25 of 141
Aquaprobe® AP-700, AP-800 & AP-2000 Instruction Manual
10401-00870 Rev W
RapidCal Calibration Method
10.1. About Calibration Calibration is a very important part of successful
water quality measurement and should be carried out regularly as detailed in
each separate section of this manual. A great deal of development work has
been put into simplifying and automating the calibration procedures in the
Aquameter® in order to allow normal field operatives (as opposed to trained
lab technicians) to achieve quick and accurate results.
As a general rule, pH and EC should be calibrated as close to 25ºC as
possible. Optical electrodes should be calibrated as close to their deployment
temperature as possible.
In order to standardise calibration techniques, Aquaread® provide plastic
calibration bottles into which the Aquaprobe® can be directly inserted. The
Aquaprobe® is designed to be calibrated in these calibration bottles with the
Probe Sleeve and Sleeve End Cap fitted.
The Probe Sleeve and Sleeve End Cap form an integral, working part of the
Probe’s measurement system, and MUST be fitted during calibration and
measurement for correct operation. See section 2.3. Important Information
about the Probe Sleeve & Sleeve End Cap for further details.
10.2. Special Notes Concerning ISE Electrodes The high ionic concentration of
pH calibration solutions (buffers), including RapidCal, can cause significant
offsets in ISE electrodes. ISE calibration solutions other than those for that
specific ISE can contain interfering ions, again causing offsets.
These offsets are temporary, but best avoided because they can cause
significant errors during both calibration and normal operation.
For this reason all ISE electrodes are supplied with a red rubber sealing cap.
The caps should be fitted to all ISE Electrodes when using pH calibration
solutions or other ISE calibration solutions other then that specific for the
ISE being calibrated in order to protect the ISE electrodes from the effects
of the buffer solution and interfering ions.
10.3. Using RapidCal
RapidCal calibrates EC at 2570µS/cm, the pH7.00 point and the zero point of
all optical electrodes (except Turbidity) simultaneously. Ideally, this
procedure should be carried out at the beginning of each day the Probe is to
be used. In addition, you should check the DO 100% calibration point and zero
the Turbidity electrode if fitted. To use RapidCal:
1. Remove the lid from a fresh bottle of RapidCal solution, remove the storage cap from the pH electrode if fitted, wash the Probe in distilled water, ensure the Probe Sleeve and End Cap are fitted, then drop the Probe into the RapidCal solution. Bang the Probe against the bottom of the bottle several times in order to remove any air
© 2024 Aquaread® Ltd.
www.aquaread.com
Page 26 of 141
Aquaprobe® AP-700, AP-800 & AP-2000 Instruction Manual
bubbles that may be clinging to the electrodes.
10401-00870 Rev W
2. When the Probe is inserted, ensure the level of the solution is at least up to the minimum insertion line scribed around the Probe sleeve. If the level is low, the EC electrodes will not be covered and EC will not calibrate properly. If the level is low, top up with fresh RapidCal solution.
3. Switch the Aquameter® on and wait until all readings are completely stable. The longer you can leave the probe to achieve thermal equilibrium before proceeding, the better.
4. Ensure the temperature of the solution is between 5ºC and 40ºC (41ºF 104ºF). The closer to 25ºC the better.
5. Press the MENU key then select Calibration. The following screen will be displayed.
Calibration RapidCal
DO 100% Full Cal
6. Select RapidCal. The screen will change to:
PLEASE WAIT Stabilising
000%
The Meter will wait until all readings are stable, then it will send the RapidCal command to the Probe, where the calibration takes place. During calibration, the Calibrating screen is displayed and the progress counter counts up. If the calibration is successful, the counter will reach 100% and the following screen will be displayed.
Calibrating 100%
Press [OK] When calibration is complete, press OK then ESC to return to normal
reading mode.
© 2024 Aquaread® Ltd.
www.aquaread.com
Page 27 of 141
Aquaprobe® AP-700, AP-800 & AP-2000 Instruction Manual
10401-00870 Rev W
Now the DO 100% saturation point should be checked and if necessary, calibrated in damp air.
To Check / Calibrate the 100% DO Saturation Point in Damp Air
1. After calibrating with RapidCaL, remove the Probe from the bottle, wash in
fresh water, then shake off ensuring there are no droplets adhering to the DO
membrane.
2. Moisten a clean cloth or piece of tissue paper with fresh water and wrap it around the open end of the probe ensuring all the holes are covered. Place the probe on a flat surface. Do not hold the probe, the heat from your hands will warm the probe up and interfere with calibration.
3. Wait until the temperature and DO measurements are completely stable. This is very important. If the DO measurement is 100% +/- 1%, there is no need to recalibrate.
4. If recalibration is needed, refer back to the screen shown in item 5 above and select DO 100%.
5. Wait while the Aquameter® carries out the calibration procedure.
6. When the `Calibrating 100%’ screen (shown above) is displayed, press OK then ESC repeatedly to return to normal reading mode.
Finally, if you have a Turbidity electrode fitted, you should calibrate the zero point now using fresh, still mineral water. Refer to section 16.4.7. Turbidity Zero Point Calibration .
© 2024 Aquaread® Ltd.
www.aquaread.com
Page 28 of 141
Aquaprobe® AP-700, AP-800 & AP-2000 Instruction Manual
10401-00870 Rev W
10.4. Calibration Error Messages
If the Aquameter® detects a problem with either the Aquaprobe® or the
calibration solution during the calibration procedure, an error will be
indicated. The chart below shows the possible errors and how to correct them.
Error Message
Problem
Action
REPLACE DO Full re-calibration required or Optical DO
CAP
Cap needs replacing
See note below.
BATTERIES Battery Voltage is too low for reliable TOO LOW calibration
Replace the batteries
NO PROBE RESPONSE
The Probe is not responding
Check connections / cycle power
READINGS Readings did not stabilise within the
Top up / replace the RapidCal
UNSTABLE expected period
Allow longer for stabilisation.
OUT OF CAL RANGE
Readings are outside calibration limits (can Top up / check calibration be caused by low level / incorrect calibration solution is correct type. Ensure solution). Or the Probe Sleeve is not fitted the Probe Sleeve is fitted
OUT OF TEMP RANGE
Temperature is outside 5C 40C limit
Warm / cool the RapidCal
You are trying to calibrate an upper
Calibrate the zero point first,
CAL ZERO calibration point on an optical electrode
then without switching the
FIRST
without first calibrating the zero point in the Aquameter off, calibrate the
current calibration session.
upper point.
If the ‘REPLACE DO CAP’ error occurs during Optical DO Zero calibration, this usually
indicates that the DO Cap needs replacing. Perform a full DO calibration first at DO Zero
then at 100% DO. If that does not cure the problem, replace the DO Cap ( see Replacing
the Optical DO Cap in section 14).
If the corrective actions shown above for ‘READINGS UNSTABLE’ or ‘OUT OF CAL RANGE’ errors do not work, thoroughly clean the Probe and try again. If the ‘OUT OF CAL RANGE’ error persists, reset the calibration values to Factory Defaults then try again.
If the ‘OUT OF CAL RANGE’ error persists when calibrating EC, check you are using the correct EC Calibration Standard and that the Probe Sleeve is fitted and tight.
If the ‘OUT OF CAL RANGE’ error persists when calibrating pH, check you are using the correct pH Calibration Standard for the calibration point selected.
If the ‘OUT OF TEMP RANGE’ error persists when carrying out a three point ISE calibration, check your solution temperatures are within the specified limits with respect to each other.
Remember: The Probe sleeve forms an integral, working part of the Probe’s measurement system, and MUST be fitted during calibration and measurement for correct operation. If you try to calibrate the Probe without the sleeve fitted, you may get an error message.
10.5. Resetting to Factory Calibration Defaults
In some cases, if there has been a serious calibration error, the easiest way to rectify the situation is to reset the Probe to its factory defaults. To do this, first bring up the Calibration screen:
© 2024 Aquaread® Ltd.
www.aquaread.com
Page 29 of 141
Aquaprobe® AP-700, AP-800 & AP-2000 Instruction Manual
10401-00870 Rev W
Calibration RapidCal
DO 100% Full Cal
Select Full Cal. This will give you a choice of three electrodes:
Calibration pH/REDOX(ORP)
DO/EC Aux Electrodes
Move the cursor arrow to the electrode you want to reset, then press the MR
key. If you select Aux Electrodes, you must press OK first to enter the Aux
Electrode selection screen. Once in that screen, select the Aux electrode you
want to reset then press MR.
A confirmation screen will be displayed.
Are you sure you want to restore the factory calibration values? [ESC]=NO
If you are sure, press the OK key. If you want to change your mind, press the ESC key. If you press OK, you will see a message that says CAL RESTORED.
Once factory calibration defaults have been restored, you must carry out a full calibration of the electrode in question.
10.6. Calibration Reports
At the conclusion of each successful individual electrode calibration, a
single line Calibration Report is displayed. This report contains the raw
output of the electrode under calibration, uncorrected for temperature.
These values can be recorded and used to track the performance and ageing of the individual electrodes. Please note however, in order to maximise the value of this feature, all calibrations must be performed at the same temperature otherwise the recorded values will not be comparable over time.
10.7. Calibration Data Storage and Retrieval
The Aquaprobe® contains its own microprocessor and memory. All calibration
data, including the GLP data, is stored within the Probe’s memory. When a
Probe is connected to a Meter, this data is transferred for display and
logging.
This is a major advantage and allows you to use a variety of different Probes with a single Meter, without the need for re-calibration.
In order to recall the calibration data for a certain electrode on the Meter, first select the calibration screen for that electrode. If, for instance, you selected pH/REDOX(ORP), the screen below would be displayed.
© 2024 Aquaread® Ltd.
www.aquaread.com
Page 30 of 141
Aquaprobe® AP-700, AP-800 & AP-2000 Instruction Manual
10401-00870 Rev W
PH 7.00 [31/Jan/17] PH 4.01 [07/Feb/17] PH 10.0 [07/Feb/17] ORP+250
[09/Feb/17] This tells you that the last successful calibration was January
31st for the pH 7.00 point, February 7th for the pH 4.01 & pH 10 points and
February 9th for ORP. If the date field is dashed (==/===/==), this means the
electrode is either not fitted or has never been calibrated.
Pushing the left or right arrow keys will toggle this screen with the
calibration report screen.
PH7 Offset:+01.2mV PH4 Slope:56.8mV/pH PH10 Slope:56.3mV/pH ORP Offset:-02.6mV
The calibration report screen displays the calibration report values from the
last calibration.
To exit this screen press the ESC key.
© 2024 Aquaread® Ltd.
www.aquaread.com
Page 31 of 141
Aquaprobe® AP-700, AP-800 & AP-2000 Instruction Manual
10401-00870 Rev W
After Use
The Aquaprobe® should always be cleaned after every use.
It is advisable to clean the Probe after use with the cable attached. This
will prevent any water entering the Probe’s socket and will allow any deposits
to be removed from the connector collar and shell.
The Sleeve on the Aquaprobe® can be removed by unscrewing to allow cleaning of
the individual electrodes. After every use, remove the protective Sleeve End
Cap then unscrew the sleeve. With the Sleeve removed, the individual
electrodes are very vulnerable, so please handle the Probe with extreme care.
If you drop it, it’s going to break!
Rinse the exposed electrodes, the inside of the Sleeve and the Sleeve End Cap
with fresh, clean water. Shake the water from inside the Sleeve, then
reattach. Dry the outside of the Probe using a soft cloth.
Remember to replace the pH/ORP storage cap after use. Failure to do so will
damage the electrode. For more details, see Keeping the Electrodes Moist in
section 13.
Never clean the Probe with solvents, alcohol or concentrated acid/alkaline
based cleaning products such as Decon 90. These products can strip the
anodised finish from the Probe and damage the plastic and rubber components.
Damage caused by the use of aggressive cleaning agents or solvents is not
covered by your warranty.
Store the Probe without the protective Sleeve End Cap fitted in order to allow
free air circulation around the individual electrodes.
TIP: Occasional application of a smear of silicone grease or similar lubricant
to the connector O-rings and thread, Sleeve thread, the protective Sleeve End
Cap O-ring and the inside rim of the lower Probe Sleeve will make fitting and
removal of these parts easier.
© 2024 Aquaread® Ltd.
www.aquaread.com
Page 32 of 141
Aquaprobe® AP-700, AP-800 & AP-2000 Instruction Manual
10401-00870 Rev W
General Probe Maintenance
Other than regular cleaning and calibration, very little in the way of maintenance is needed.
12.1. Identifying The Individual Electrodes The photograph below shows the
standard AP-2000 electrodes. On the AP-2000, the DO and EC sensors are
incorporated into one electrode.
On the AP-700 and AP-800, the DO and EC sensors are separate electrodes.
Please note: The photograph shows the AP-2000 with Optical DO. The AP-700 and AP-800 models may have either Optical DO or Galvanic DO sensors.
Galvanic DO caps and later model Optical DO caps do not have the red warning label fitted to the DO cap.
© 2024 Aquaread® Ltd.
www.aquaread.com
Page 33 of 141
Aquaprobe® AP-700, AP-800 & AP-2000 Instruction Manual
10401-00870 Rev W
pH/ORP Electrode Calibration and Maintenance
13.1. Recognising the pH/ORP Electrode The combined pH/ORP electrode is easy
to recognise because it is the only electrode that is not black. This
electrode has a clear, gel filled body.
13.2. Electrode Removal and Replacement The pH/ORP electrode can be unscrewed
from the Probe body by rotating it anti-clockwise. When replacing an
electrode, apply a little silicone grease or similar lubricant to the thread
and O ring, then screw fully in.
Gripping the black collar at the top of the electrode, tighten until the O
ring is fully compressed. Do not twist the clear section of the electrode
whilst tightening.
Useful Tip: The red lanyard that is attached to the pH/ORP storage cap makes a
very useful belt wrench for tightening and loosening the pH/ORP and AUX
electrodes.
Slide the lanyard over the electrode and use it to grip the knurled body.
Never immerse an Aquaprobe® with the pH/ORP electrode removed. This will cause
serious damage to the electrode socket. This is not covered by your warranty.
13.3. Keeping the Electrodes Moist It is very important that the pH/ORP
electrode is kept moist when not in use. This is achieved by always fitting
the storage cap, which incorporates a sponge that should be soaked in a
special storage solution.
The sponge within the storage cap should be moistened with a few drops of pH
Electrode Storage Solution each time it is removed and replaced. If a pH/ORP
electrode is inadvertently allowed to dry out, it must be re-hydrated by
soaking in storage solution for at least one hour prior to use.
13.4. Calibrating pH pH electrodes should be calibrated fully at least once a
week to ensure optimum accuracy. Full calibration involves calibrating at pH
7.00 first, then at pH 4.01 and/or pH 10.00. The Aquaprobe® allows for both
two and three point pH calibration. Should you decide to carry out just a two
point calibration, the probe will automatically calculate and save a
calibration value for the uncalibrated third point in order to maintain the
electrode’s linearity over the full range of 0 14.
For best results, calibrate all three points as close to 25C as possible.
© 2024 Aquaread® Ltd.
www.aquaread.com
Page 34 of 141
Aquaprobe® AP-700, AP-800 & AP-2000 Instruction Manual
10401-00870 Rev W
13.5. Special Notes Concerning ISE Electrodes
The high ionic concentration of pH calibration solutions (buffers), including
RapidCal, can cause significant offsets in ISE electrodes. ISE calibration
solutions other than those for that specific ISE can contain interfering ions,
again causing offsets.
These offsets are temporary, but best avoided because they can cause significant errors during both calibration and normal operation.
For this reason all ISE electrodes are supplied with a red rubber sealing cap.
The caps should be fitted to all ISE Electrodes when using pH calibration solutions or other ISE calibration solutions other then that specific for the ISE being calibrated in order to protect the ISE electrodes from the effects of the buffer solution and interfering ions.
13.5.1.
Calibrating the First Point (pH 7.00)
Due to the way in which pH calibration works, the Probe must be calibrated at pH7.00 before calibrating at pH 4.01 or pH 10.00. Never calibrate at pH 4.01 or pH 10.00 before first calibrating at pH7.00.
To calibrate the pH electrode follow these steps:
1. Fill a calibration bottle with fresh pH 7.00 solution or RapidCal, remove
the storage cap from the pH electrode, wash the Probe in distilled water, then
drop the Probe in all the way.
2. Switch the Aquameter® on and wait until the temperature and pH
measurements are completely stable.
3. Ensure the temperature of the solution is between 5ºC and 40ºC (41ºF –
104ºF). 4. Press the MENU key then select Calibration. The following screen
will be displayed.
© 2024 Aquaread® Ltd.
www.aquaread.com
Page 35 of 141
Aquaprobe® AP-700, AP-800 & AP-2000 Instruction Manual
10401-00870 Rev W
Calibration RapidCal
DO 100% Full Cal
5. Select Full Cal. The screen will change to:
Calibration pH/REDOX(ORP)
DO/EC Aux Electrodes
6. Select pH/REDOX(ORP). The screen will change to:
pH 7.00?[01/Jan/17] PH 4.01?[01/Jan/17] PH 10.0?[01/Jan/17]
ORP+250?[01/Jan/17] The dates shown to the right of the screen are the dates
of the last successful calibration.
7. Select pH7.00. The screen will change to:
PLEASE WAIT Stabilising
000%
The Meter will wait until the readings are stable, then it will send the
calibration command to the Probe, where the calibration takes place. During
calibration, the Calibrating screen is displayed and the progress counter
counts up. If the calibration is successful, the counter will reach 100% and
the following screen will be displayed.
Offset: -1.2mV Calibrating
100% Press [OK]
The top line displays the voltage offset from zero for the pH electrode in
+/-millivolts (mV). If this offset goes beyond +/-25mV at 25°C, the pH
electrode should be serviced.
This value is stored in the Probe’s memory and can be recalled at any time.
See section 10.7. Calibration Data Storage and Retrieval. Press OK then ESC
repeatedly to return to normal reading mode.
Remove the Probe from the calibration bottle, rinse thoroughly in de-ionised
water, shake off any excess and dry the outer sleeve with a soft cloth.
© 2024 Aquaread® Ltd.
www.aquaread.com
Page 36 of 141
Aquaprobe® AP-700, AP-800 & AP-2000 Instruction Manual
10401-00870 Rev W
13.5.2.
Calibrating the Second Point
The pH electrode can now be calibrated at either pH 4.01 or pH 10.00. If you intend to calibrate at both pH 4.01 and pH 10.00, both points must be calibrated in the same session, i.e. without turning the power off.
If the power is removed after calibrating just one additional point (pH 4.01 for example), the probe will automatically calculate and save a calibration value for the uncalibrated third point in order to maintain the electrode’s linearity.
To calibrate the second point, fill a calibration bottle with fresh pH 4.01 or
pH 10.00 solution and drop the Probe in all the way. Follow the procedure
detailed above, but at step 6, select either pH4.01 or pH10.0, dependent upon
the solution you are using. Wait while the Meter stabilises and calibrates.
When the `Calibrating 100%’ screen is displayed, the calibration report will
display the slope for the pH electrode in millivolts (mV) per pH unit. If this
slope
goes below 45mV/pH at 25°C, the pH electrode should be serviced. Press OK then
press the ESC key repeatedly to get back to the main display.
Remove the Probe from the calibration bottle, rinse thoroughly in fresh water, shake off any excess and dry the outer sleeve with a soft cloth.
13.5.3.
Calibrating the Third Point
Without switching the Aquameter® off or disconnecting the Probe, fill a calibration bottle with
fresh pH 4.01 or pH 10.00 solution and drop the Probe in all the way. Follow the procedure
detailed above, but at step 6, select either pH4.01 or pH10.0 dependent upon the solution
you are using. Wait while the Meter stabilises and calibrates. When the `Calibrating 100%’
screen is displayed, the calibration report will display the slope for the pH electrode in
millivolts (mV) per pH unit. If this slope goes below 45mV/pH at 25°C, the pH electrode should be serviced. Press OK then press the ESC key repeatedly to get back to the main
display.
Remove the Probe from the calibration bottle, rinse thoroughly in fresh water, shake off any excess and dry the outer sleeve with a soft cloth. Dampen the sponge in the storage cap with storage solution and fit it to the pH/ORP electrode. pH calibration is now complete.
13.6. Errors During Calibration
If a problem occurs during calibration, an error message will be displayed.
Refer to Calibration Error Messages in section 10 for error message handling.
13.7. pH Electrode Efficiency
If the pH electrode becomes worn or clogged, its efficiency and response time
can be reduced. The efficiency of the pH electrode is constantly monitored and
in the event of the efficiency dropping below 85%, `ERROR 01′ will be flashed
on the bottom line of the display. If this occurs, or if the pH reading
response becomes slow, recondition the electrode as described below.
13.8. Servicing the pH Electrode
1. Remove the pH or combined pH/ORP electrode from the Probe body (see
Electrode Removal and Replacement).
2. Rinse with methyl alcohol. 3. Replace the electrode. 4. Re-calibrate.
© 2024 Aquaread® Ltd.
www.aquaread.com
Page 37 of 141
Aquaprobe® AP-700, AP-800 & AP-2000 Instruction Manual
10401-00870 Rev W
Never place the entire Aquaprobe® in methyl alcohol, as this will cause irreparable damage to the DO/EC electrode. Damaged caused in this way is not covered by the warranty.
If the methyl alcohol rinse does not restore the electrode, perform the following actions:
1. Remove the electrode from the body again. 2. Soak in 0.1M HCl for 5 minutes. 3. Rinse in de-ionised water. 4. Soak in 0.1M NaOH for 5 minutes. 5. Rinse in de-ionised water. 6. Soak in pH4.01 buffer for 10 minutes.
If the above procedure still does not restore performance, replace the electrode.
13.9. Calibrating ORP
ORP electrodes should be calibrated at least once a month to ensure optimum
accuracy. Full calibration involves calibrating at a single point, either
+250mV (at 25C) using a +250mV ORP calibration standard such as Reagecon RS250
Redox Standard, or +229mV (at 25C) using a +229mV ORP calibration standard
such as Zobell Solution.
For best results, calibrate as close to 25C as possible. The probe will automatically compensate for temperature variation in the calibration solution during calibration.
To calibrate the ORP electrode follow these steps:
1. Fill a calibration bottle with fresh calibration solution, remove the storage cap from the pH/ORP electrode, wash the Probe in distilled water, then drop the Probe in all the way.
2. Switch the Aquameter® on and wait until the temperature and ORP measurements are completely stable.
3. Ensure the temperature of the solution is between 5ºC and 40ºC (41ºF 104ºF).
4. Press the MENU key then select Calibration. The following screen will be displayed.
Calibration RapidCal
DO 100% Full Cal
5. Select Full Cal. The screen will change to:
© 2024 Aquaread® Ltd.
Calibration pH/REDOX(ORP)
DO/EC Aux Electrodes
www.aquaread.com
Page 38 of 141
Aquaprobe® AP-700, AP-800 & AP-2000 Instruction Manual
6. Select pH/REDOX(ORP). The screen will change to:
10401-00870 Rev W
pH 7.00?[01/Jan/17] PH 4.01?[01/Jan/17] PH 10.0?[01/Jan/17] ORP+250?[01/Jan/17] 7. Move the arrow to the bottom line. If you are using 250mV calibration solution press the OK key to continue. If you are using 229mV calibration solution, move the cursor to the right, then use the up/down arrow keys to select +229. When the correct solution has been selected, press OK. The screen will change to:
PLEASE WAIT Stabilising
000%
The Meter will wait until the readings are stable, then it will send the calibration command to the Probe, where the calibration takes place. During calibration, the Calibrating screen is displayed and the progress counter counts up. If the calibration is successful, the counter will reach 100% and the following screen will be displayed.
Offset: 5.5mV Calibrating
100% Press [OK] The Calibration Report on the top line displays the voltage
offset between the ORP electrode output and the value of the calibration
solution at the calibration temperature in +/-millivolts (mV). During normal
operation this offset will be subtracted from the ORP electrode output to give
a corrected ORP display.
This value is stored in the Probe’s memory and can be recalled at any time.
See section 10.7. Calibration Data Storage and Retrieval. Press OK then ESC
repeatedly to return to normal reading mode.
Remove the Probe from the calibration bottle, rinse thoroughly in fresh water,
shake off any excess and dry the outer sleeve with a soft cloth. Dampen the
sponge in the storage cap with storage solution and fit it to the pH/ORP
electrode. ORP calibration is now complete.
© 2024 Aquaread® Ltd.
www.aquaread.com
Page 39 of 141
Aquaprobe® AP-700, AP-800 & AP-2000 Instruction Manual
10401-00870 Rev W
13.10. Converting ORP Readings to the Hydrogen Scale
Electrochemical measurements are ultimately referred to the so-called hydrogen
scale, the convention for which is that the electrochemical potential of a
hydrogen electrode in contact with hydrogen gas at one atmosphere partial
pressure and a solution containing hydrogen ions at unit activity is zero at
all temperatures.
The ORP reference electrode used in Aquaread® combination electrodes is a 3MPK1 silver chloride type, and exhibits potentials on the hydrogen scale of:
Temperature 5C 10C 15C 20C 25C 30C 35C 40C
Potential 221 mV 217 mV 214 mV 210 mV 207 mV 203 mV 200 mV 196 mV
Thus, to refer an ORP potential value measured with the Aquaprobe® to the hydrogen scale, the appropriate value above should be added to the measured value.
© 2024 Aquaread® Ltd.
www.aquaread.com
Page 40 of 141
Aquaprobe® AP-700, AP-800 & AP-2000 Instruction Manual
10401-00870 Rev W
DO Electrode Calibration and Maintenance
14.1. Recognising the DO Electrode Type All AP-2000 Aquaprobes are fitted with Optical DO electrodes. The AP-700 and AP-800 may be fitted with either an Optical DO or Galvanic DO electrode. Refer to the photograph on the right to identify your DO electrode type. If your Aquaprobe® features an Optical DO electrode, skip to section 14.6..
14.2. Galvanic DO Electrode The Galvanic DO electrode consists of an
electrolyte filled DO sensor cap with a clear, gaspermeable membrane stretched
across a silver electrode.
14.3. Precautions During Use In order to achieve accurate Dissolved Oxygen
readings with the Galvanic DO electrode, the Probe needs to be either placed
in flowing water, or needs to be stirred or raised and lowered continuously to
ensure a minimum flow rate of 0.3m/s over the DO Electrode. If there is no
water flow across the Probe, the oxygen in the immediate area of the DO
Electrode will be consumed and the reading will start to fall.
14.4. Calibrating the Galvanic DO Electrode The DO electrode should be
calibrated at the Zero saturation point at least once a month. Before each
day’s use, the 100% saturation point should be checked in moist air and
recalibrated if necessary. For optimum accuracy, calibrate the DO100% point as
near to your sample temperature as possible (within the calibration
temperature limits of 5C – 40C).
If you are going to calibrate both the Zero and 100% points at the same time,
ALWAYS calibrate the Zero point first, then the 100% point.
14.4.1. Calibrating the DO Zero Point
1. Remove the lid from a 150mL bottle of DO Zero calibration solution, remove
the storage cap from the pH electrode if fitted, wash the Probe in distilled
water, then drop the Probe in all the way.
2. Switch the Aquameter® on and wait until the DO reading is completely
stable. 3. Ensure the temperature of the solution is between 5C and 40C (41F –
104F). 4. Press the MENU key then select Calibration. The following screen
will be
displayed. Calibration
RapidCal DO 100% Full Cal
5. Select Full Cal. The screen will change to:
Calibration pH/REDOX(ORP)
DO/EC TURB
6. Select DO/EC. The screen will change to:
© 2024 Aquaread® Ltd.
www.aquaread.com
Page 41 of 141
Aquaprobe® AP-700, AP-800 & AP-2000 Instruction Manual
Calibration DOZero?[01/Jan/17] DO100%?[01/Jan/17] EC2570?[01/Jan/17]
10401-00870 Rev W
The dates shown to the right of the screen are the dates of the last successful calibration. 7. Select DOZero. The screen will change to:
PLEASE WAIT Stabilising
000%
The Meter will wait until the readings are stable, then it will send the calibration command to the Probe, where the calibration takes place. During calibration, the Calibrating screen is displayed and the progress counter counts up. If the calibration is successful, the counter will reach 100% and the following screen will be displayed.
Output:0.4 Calibrating
100% Press [OK]
The Calibration Report on the top line will display the voltage output from
the DO cell in millivolts (mV). Please contact Aquaread for current acceptable
calibration values.
This value is stored in the Probe’s memory and can be recalled at any time.
See section 10.7. Calibration Data Storage and Retrieval. Press OK then ESC
repeatedly to return to normal reading mode.
If a problem occurs during calibration, an error message will be displayed.
Refer to Calibration Error Messages in section 10.4 for error message
handling. If If the ‘OUT OF CAL RANGE’ error persists when calibrating the DO
sensor, replace the DO Electrode Membrane Cap.
Remove the Probe from the calibration bottle, rinse thoroughly in fresh water,
shake off any excess and dry the outer sleeve with a soft cloth.
14.4.2.
Calibrating the DO 100% Saturation Point in Moist Air
1. Wash the probe thoroughly in fresh water, then shake off ensuring there are no droplets adhering to the DO membrane.
© 2024 Aquaread® Ltd.
www.aquaread.com
Page 42 of 141
Aquaprobe® AP-700, AP-800 & AP-2000 Instruction Manual
10401-00870 Rev W
2. Moisten a clean cloth or piece of tissue paper with fresh water and wrap
it around the open end of the probe ensuring all the holes are covered. Place
the probe on a flat surface. Do not hold the probe, the heat from your hands
will warm the probe up and interfere with calibration.
3. Switch the Aquameter® on and wait until the temperature measurement is
completely stable. This is very important.
4. Referring back to the screens shown in item 6 above, select DO100% 5. Wait
while the Aquameter® carries out the calibration procedure. 6. When
calibration is complete, the Calibration Report will be displayed.
The top line will display the voltage output from the DO cell in millivolts (mV). This value should be over 20.0 (at 25°C). If the value returned is less than 20.0, the DO Cap should be replaced.
These values are not stored in memory so should be noted down in a calibration record book for the probe. If a problem occurs during calibration, an error message will be displayed. Refer to Calibration Error Messages in section 10.4 for error message handling.
14.5. Replacing a Galvanic DO Electrode Membrane Cap
The Galvanic DO electrode membrane is a very thin sheet of special plastic,
which is permeable to oxygen. Oxygen molecules pass through this membrane into
the Oxygen sensor. The membrane is extremely delicate and is factory fitted
into the DO Membrane Cap. To ensure optimum performance, the DO Membrane Cap
must be replaced every 1-2 months.
Never touch the plastic membrane as the oils in your skin will block the pores in the membrane and stop it from working correctly.
To replace the DO Membrane Cap, follow these simple steps. 1. Remove the Probe sleeve. 2. Unscrew the DO Cap from the end of the DO electrode by rotating it anti-clockwise. 3. Gently scrape away any soft grey deposits from the zinc shaft of the electrode taking care not to damage the electrode in the process. If the deposits are hard and dry, soak the electrode in DO filling solution until the deposits soften up, then remove them. 4. After removing the deposits, rinse the electrode with DO Electrode Filling Solution. 5. Rinse a new DO Membrane Cap with DO Electrode Filling Solution then tap it out so that it is completely empty. 6. Using the DO Electrode Filling Solution dropper bottle, half fill the DO Membrane Cap. Gently tap the cap to ensure any trapped air bubbles are released. 7. Holding the Probe so that the DO electrode is facing downwards, slowly screw the half-filled DO Membrane Cap back onto the DO electrode then tighten the cap. Do not over-tighten. Finger tight is fine. Some solution will overflow. This is normal. 8. Wash the DO electrode with fresh water then replace the lower Probe sleeve. 9. Wait at least six hours (preferably over-night) to allow any oxygen dissolved in the filling solution to be consumed. 10. Carry out both Zero point and 100% point DO calibration as described earlier.
© 2024 Aquaread® Ltd.
www.aquaread.com
Page 43 of 141
Aquaprobe® AP-700, AP-800 & AP-2000 Instruction Manual
10401-00870 Rev W
NEVER re-install a Galvanic DO Membrane Cap once it has been fully tightened. The membrane will be stretched and will not seal properly over the silver cathode a second time. If the membrane does not create a proper seal over the silver cathode, the DO sensor will not operate correctly and any readings given will be erroneous.
Please note: After a period of use and dependant upon how previous cleaning operations have been performed, the surface of the zinc shaft will become rugged and apparently corroded. This is entirely normal and will not affect the performance of the electrode.
14.6. Optical DO Electrode If your Aquaprobe® is fitted with an Optical DO
electrode, the following sections apply. See Appendix 1. The Tech Behind
Aquaread’s Optical DO Measurement System for more technical details of the
Optical DO measurement system.
Unlike the Galvanic DO electrode, the Optical DO electrode does not consume
oxygen during operation so does not require a constant flow of water or
stirring.
14.7. Calibrating the Optical DO Electrode The DO electrode should be
calibrated at the Zero saturation point at least once a month. Before each
day’s use, the 100% saturation point should be checked in moist air and
recalibrated if necessary. For optimum accuracy, calibrate the DO100% point as
near to your sample temperature as possible (within the calibration
temperature limits of 5C – 40C).
If you are going to calibrate both the Zero and 100% points at the same time,
ALWAYS calibrate the Zero point first, then the 100% point.
14.8. Calibrating the DO Zero Point 1. Remove the lid from a 150mL bottle of
DO Zero calibration solution, remove the storage cap from the pH electrode if
fitted, wash the Probe in distilled water, then drop the Probe in all the way.
2. Switch the Aquameter® on and wait until the DO reading is completely
stable. 3. Ensure the temperature of the solution is between 5C and 40C (41F –
104F). 4. Press the MENU key then select Calibration. The following screen
will be displayed.
Calibration RapidCal
DO 100% Full Cal
5. Select Full Cal. The screen will change to:
Calibration pH/REDOX(ORP)
DO/EC Aux Electrodes
6. Select DO/EC. The screen will change to:
© 2024 Aquaread® Ltd.
www.aquaread.com
Page 44 of 141
Aquaprobe® AP-700, AP-800 & AP-2000 Instruction Manual
10401-00870 Rev W
Calibration DOZero?[01/Jan/17] DO100%?[01/Jan/17] EC2570?[01/Jan/17] The dates
shown to the right of the screen are the dates of the last successful
calibration.
7. Select DOZero. The screen will change to:
P
References
- MilliporeSigma | Life Science Products & Service Solutions
- MilliporeSigma | Life Science Products & Service Solutions
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>