doudouX HUGGIES360 Electric Home Care Bed User Manual

June 1, 2024
DoudouX

doudouX HUGGIES360 Electric Home Care Bed

doudouX HUGGIES360 Electric Home Care Bed

Manual Introduction

Electric home-care beds are essential equipment for the well-being, treatment and recovery of bedridden patients. Electric home-care beds allow patients to remain in more comfortable positions as well as in positions sometimes recommended by doctors, which would be very difficult in a normal bed, where one would always have to be placing or removing supports. Electric home-care beds can be manual or automatic. The automatic home-care bed has its functions controlled by buttons on the side panels of the bed or via remote control. The manual home-care bed has 2 movements triggered by cranks, having the elevation of the headboard, which facilitates the moment of feeding, also avoiding refluxes and facilitating breathing and has elevation of the part of the feet that will cause it to improve the blood circulation of the lower limbs, in view of the long period that the patient will spend lying down. On the other hand, patient safety is very important at this time of recovery, so all of our beds have a side protection rails, preventing the patient from going at night and running risks of falling.

Note:

Please carefully read this manual and strictly follow the steps before installing and using. We require all users to have the knowledge of user’s manual and all the cautions.

  • This manual is applicable to HUGGIES360 electric home-care bed. It contains the installation method, operating instructions and maintenance check points.
  • The graphics in this manual are for users’ reference only. In actual use, please refer to the real object.
  • This user manual includes the description, instruction, maintenance and technical parameters of the product.
  • Any questions, please contact the after-sale service center or distributor.
  • Anyone using or operating this product should read and comply with it.
  • Except for some unexpected situations, users are required to be familiar with the usage and precautions.

Installation Guide

After opening the packing case, please check whether the following parts are complete in the packing case:

Table a:

NO. Name Picture Specification QTY Unit
1 Head board Hook distance: L35.8*H2.2 inches 1 pc
2 Foot board Hook distance: L35.8*H2.2 inches 1 pc
3 Side rail Sixth-aluminum alloy side rail 2 pcs
4 Caster 5 inch covered caster 4 pcs
5 IV pole Φ6.3-7.5inches/off white hook 1 pc
6 Bed platform L75.8*W32.3 inches 1 pc
7 Screw

/

| Refer to Table a.1| 1| set

Table a.1:

NO. Name Picture Specification QTY Unit
1 PP mattress holder PP 2 pcs
2 Hexagon screw M3.15*27.6 inches 4 pcs
3 Flat washer Φ3.15Φ6.3H0.47 inches 4 pcs
4 Spring washer Φ3.15Φ5.5H0.98 inches 4 pcs
5 Nut M3.15 inches 4 pcs
6 Hexagon screw M3.15*6.3 inches 8 pcs
7 Flat washer Φ3.15Φ6.3H0.47 inches 8 pcs
8 Spring washer Φ3.15Φ6.3H0.98 inches 8 pcs

Casters installation

Lay a protective pad and put the bed body upside down;
Remove the screw as shown below, and use a 14-inch hex wrench.

Casters Installation

Side guardrails installation

Place the bed upright and install the guardrail, (The picture may not exactly match the real product, this is only for installation instructions)
Take out the screw as shown below, and use a 14-inch hex wrench.

Side Guardrails Installation
Side Guardrails Installation

PP mattress holder installation

PP Mattress Holder Installation
PP Mattress Holder Installation

Head board & Foot board assembly method

(The picture does not exactly match the real product, only for the installation instructions)

Head Board Assembly
Head Board Assembly

IV pole installation

Female part installation: Press the plastic hook (female part) down on the Φ16mm stainless steel pipe.
Male part installation: Place the plastic hook (male part) directly above the female part in a cross shapes, and then aligns the buckles on the male part with the slots on the female part and press down. After hearing a “click”, check whether the end of the buckle is tight, if it is tight, then complete the installation, if not, continue to press down until the buckle is tight.
Insert the installed IV pole vertically into the IV pole prevision. (The arrow in the picture shows the IV pole prevision)

Pole Installation
Pole Installation

Use Method

  1. Head board
  2. Casters
  3. Guardrail
  4. Bed surface board
  5. Drainage hook
  6. Bed frame
  7. Mattress holder
  8. Foot board
    Use Method
Bed surface size Length inch 75.8
width 32.3
The lowest height of the bed from the ground 15.8
Dimensions Length 83
Width 37.8
Height 15.8~28.0
Tilt range Back board / 0-75±5°
Leg board 0-35±5°
Weight Safe working load Lb. 450

Use method of casters

When there is a patient on the bed, the guardrail should be raised before moving the bed.

Except when the bed must be moved, the caster brake must be stepped on during the rest of the time. Need to depress all four wheel brakes.

Lock the caster: After pushing the bed to the desired position, press the ON button with your foot and lock the caster when you hear a “click”, as shown in the figure below.

Use Method

Unlock the caster: Press the OFF button with your foot and unlock the castor when you hear a “click”.

Use Method

Use method of guardrails

Raising and lowering the guardrail

Raise the guardrail: Hold the red gun of the guardrail and lift it up. When you hear the sound of “click”, it means the guardrail is locked.

Use Method

Lower the guardrail: Hold the guardrail red gun and slowly lay down.

Use Method

Use method of drainage hooks

The drainage hooks as shown in the figure.

Use Method
Use Method

Use method of IV pole

Adjusting the height of the IV pole:

  1. First loosen the outer sleeve counter clock carefully, insert the inner tube of the IV pole into the outer sleeve, and then adjust the length of the extension rod up and down;
  2. Second adjust the required length, and then tighten the outer sleeve clock carefully. The maximum bearing of each hook is 2kg, a total of 8kg. (Height Adjustment :890–1540mm).
    Use Method

Use method of hand controller

  1. Back up button
  2. Back down button
  3. Leg up button
  4. Leg down button
  5. Back and leg linkage up(auto-contour) button
  6. Back and leg linkage down(auto-contour)button
  7. Whole bed up button
  8. Whole down Button
    Use Method

Spare Part Replacement And Maintenance

Use 2pcs 17-inch Allen wrenches to unscrew the motor, take out the bolt, and replace the motor. When the replacement is completed, install the bolts in the order shown in the figure, and use the wrench to tighten the knot.
Spare Part Replacement And Maintenance
Spare Part Replacement And Maintenance

Faults Exclusion

Common failures Probable reasons Exclusion methods
Difficult to move the whole bed
  1. The caster is locked without releasing the brake
  2. The caster is damaged

|

  1. Release the brake,check 2.1
  2. Replace the caster,check1.1

Guardrail easy to loosen| The guardrail is not stuck in place| Lift the guardrail upward again and let go of the hand after hearing a “click”
Loud noise when the bed board moves up and down| The moving parts are rusted and the lubrication is insufficient| Add proper amount of lubricating oil at the joint of moving parts
Part of the bed does not work| The hand controller button fails| Replace the hand controller
The electric bed does not work| Unplugged and no power on| Insert the plug and turn on the power

Bed-maintenance

  1. In order to use the bed safely, safety inspections must be done regularly. Conduct a comprehensive inspection every six months, to ensure that there is no looseness in the connection parts and the function of up&down for the bed can be operated normally.

  2. At the joints of the frequently moving parts of the bed, an appropriate amount of lubricating oil needs to be added during daily inspection; if serious wear is found, it should be replaced immediately to ensure safety, add lubricated parts as show.
    Bed-maintenance
    Bed-maintenance

  3. Avoid hitting the bed and scratching the coating on the bed with sharp objects.

  4. Please avoid inhaling moisture into the bed and mattress, which will cause rust, abnormal noise and bacteria breeding. If water adheres, wipe it off immediately.

  5. When the bed frame ages and it reaches a certain servicing time, which the metal part of the bed and the plastic parts of the headboard or footboard, guardrail can be recycle.

Bed-cleaning

  1. When cleaning, please wring out the cloth soaked in neutral detergent diluted with water and wipe it, then wring out the cloth soaked in water to wipe off the remaining detergent ingredients, and finally use a dry wipe dry with a cloth.
  2. Do not use volatile substances (thinner, volatile agent, gasoline, etc.), which may cause chemical reactions and damage the bed.
  3. When using a disinfectant to clean, be sure to dilute it according to its specified concentration before use. Depending on the composition of the disinfectant, it may corrode metal parts, resin parts, etc., causing undesirable phenomena such as discoloration and deformation. Therefore, the content of the disinfectant is recommended as follows:
    0.05~0.2% ammonium chloride disinfectant
    0.05~0.2% Chlorinated phenyl disinfectant
    0.05% diclofenac ethane solution
    0.05~0.2% sodium hypochlorite disinfectant
    Do not use smoking sterilizers, autoclaves, and do not use methyl (phenol) to clean the headboard and foot of the bed. It may cause corrosion, discoloration and deterioration.
    Note : When cleaning the bed or changing the bedclothes, please pay attention to the corners, edges and screws of the frame to prevent scratches.

Mattress-cleaning And Maintenance

  1. When using it, avoid scratching the fabric with sharp-angled tools or knives, cover it with sheets or cleaning pads, and clean it frequently to protect it from dryness.
  2. Clean the mattress regularly with a vacuum cleaner, Mattress cover can be washed in 100℃.
  3. Expose the mattress often to keep the sleep fresh and comfortable, but be careful not to expose the mattress to too long.
  4. If the fabric is accidentally stained, please refer to the corresponding recommended cleaning methods in the instructions, do not use corrosive chemicals to clean the fabric.
  5. If accidentally the chewing gum or other jellies get on the fabric, do not wipe it forcibly. Use an ice pack to allow it harden and gently peel it off.
  6. If the thread is found to be loose due to wear and tear, do not pull it off by hand, use scissors to cut it together.
  • If your order does not have a mattress, please refer to the contract configuration.

Caster-detection And Maintenance

  1. Regularly check the casters to avoid hitting or overloading, which will cause damage to the casters.
  2. Grease the casters or wheels on a regularly, and the frequency of addition should be determined according to the specific actual use conditions. In general, lubricating grease is added every six months.
  3. After cleaning the caster equipment, pay attention to adding grease. The selection of lubricating grease should be based on the actual use environment, special high temperature and special low temperature environment, and the corresponding lubricating grease should be selected.
  4. It is forbidden to force the casters to brake when braking, which will cause damage to the brake system.
  5. The wear of the tire surface of the wheel can be detected by visual inspection. Some thread spun yarn and other accumulations may be wrapped around the wheel, remove the bolts and nuts of the wheel, clean up the debris, and check whether the bearing of the wheel is damaged. If the parts are not damaged, can be reassembled and used.
  6. Judging whether the caster is replaced according to the time used, wear and tear, you can replace the wheels and make sure that the axles bolt and nuts are well put together and connected. If possible, use a lock washers or locknuts to help you with this process.
  7. For the casters equipped with brakes, it is necessary to regularly check whether the brakes are working properly. Check the brakes once a day or before each use. For a home-care bed equipped with multiple brake casters, only one brake caster can be locked at a time which will allow you to try and push the bed also check whether one of the brake performance of each caster is good. If the function of the brake fails due to wear or damage to the wheels, immediately replace the wheels and then retest the brake system.
  8. If the mechanism of the brake system of the caster is damaged, and the brake needs to be repaired or replaced, please contact our after-sales service department or authorized dealer provide by our company. Every time the brakes are replaced, the brake performance of the casters must be retested.

Guardrail-maintenance

  1. Please check whether the side rails are installed correctly to prevent the lying patient from shifting and rolling down.
  2. Please always check the screws of the total position of the guardrail and the wire to prevent loosening and falling off.
  3. Please use the red plastic parts of the guardrail switch correctly, and do not use brute force prevent it from damaging the guardrail.

Spare Part List

No. Name Model No Picture Specification
1 A900/PP/Head foot board S0FJBT010019 Hook distance: 35.8*2.2 inches
2 Aluminum alloy side rail BKLLGH105931 Left side
3 Aluminum alloy side rail BKLLGH105932 Right side
4 IV pole S0FJSY010003 Φ16-19/off-white
5 5 inch caster LGPLU101TS014 φ24
6 Caster square set GPSL101TS297 1.96*1.96 inches
7 HJ motor 100 LGPDJ101TS575 ![](https://manuals.plus/wp-

content/uploads/2024/04/doudouX-HUGGIES360-Electric-Home-Care-Bed-User- Manual-49.png)| HJA58/ Stroke 3.93 inches/ Installation distance 10.8 inches
8| HJ motor 150| LGPDJ101TS574| | HJA58/ Stroke 5.9 inches / Installation distance 12.8 inches
9| HJ motor 200| LGPDJ101TS573| | HJA58/ Stroke 7.9 inches / Installation distance 14.8 inches
10| HJ control box| LGPDJ101TS576|

/

| HJT13E Switch power supply/100- 240V/50/60Hz/white
11| HJ hand controller| LGPDJ101TS577| | HJH18 remote control
12| HJ power line| LGPDJ101TS578| | Straight line 3m/white
13| HJ signal line| LGPDJ101TS579| | W971 signal line/white
14| Basket (Installation distance 560mm)| LS1FJF0102008| | 23.621.33.15 inches

(optional)

15| Drainage hook| GPSL101TS437| | 2.2*2.4 inches
16| Name card| GPSL101TS184| | /

Note : Due to product updates, parts of the original product may not be available for life.
(The company ensures that the sold products can provide the parts that need to be replaced within 6 years after the end of their production)

Packaging, Transport, Storage

  1. The packaging of electric bed is carried out according to the contract or product standard.

  2. Avoid rushing out, violent vibration, and prevent from sunlight and rain during the transportation of the electric bed.

  3. The electric bed should be stored:
    a) Ambient temperature: +5℃ ~ +40℃.
    b) Relative humidity: ≤ 85%.
    c) Atmospheric pressure: 70 kPa ~ 106 kPa.
    d) Power voltage: AC 100-220±22V;
    e)Power frequency: 50±5 Hz

  4. The equipment is with internal power supply, and the protection type is Type B.
    A) Working voltage: DC 24V.
    B) Input power: 500VA.
    C) Fuse model and rating: F3AL 250V, 5×20mm
    D) Working schedule: intermittent loading and continuous operation
    E) Ingress protection :IPX4
    F) Non-AP /APG devices

Note! Check the file that comes with the bed
IPX4 Ingress protection
Class II equipment
Symbol Protective grounding

Contact Information

Office address: 3230 FM 1463 Rd., #5108, Katy, TX, 77494
Real-time customer support: 281-616-1796 (only text, respond in 1h)
Contact email: doudoullc2020@gmail.com (respond in 1h)

Symbols

Logo

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

DoudouX User Manuals

Related Manuals