TR-RH2W 2 Wires Humidity Transmitter Instruction Manual
- May 15, 2024
- TR-RH2W
Table of Contents
TR-RH2W 2 Wires Humidity Transmitter
Product Information
Specifications:
- Humidity Measurement Range: 10% to 95% RH
- Signal Output: 4 – 20 mA
- Humidity Sensor Type: High-precision capacitance type sensor
- Power Supply: DC 9V to 30V
- Wire Configuration: 2 wires for both signal and power supply
- Accuracy:
- 70% R.H.: <1%
- < 70% R.H.: <5%
- Load Impedance Requirement: Up to 500 ohms
- Operating Temperature: 0 to 50°C (32 to 122°F)
- Operating Humidity: Less than 95% RH
- Size: 200 x 23 x 19 mm (7.9 x 0.9 x 0.7 inch)
- Weight: 75 g (0.16 LB)
- Cable Length: Approximately 1.2 m
- Wire Connection: Red wire for Power + and Signal, Black wire for Power –
Product Usage Instructions
Front Panel Description:
- Humidity Sensing Head
- Probe Main Body
- Signal and Power Supply Wires
- Probe Holder
Operation Procedures:
- Using the two screws, fix the Probe Holder on the wall.
- Mount the Probe Main Body to the Probe Holder.
- Use the two wires as signal and power supply wires.
Wires Installation:
Humidity Transmitter TR-RH2W:
- Black Wire: Power Supply –
- Red Wire: Power Supply +, 9V to 30VDC
Calibration Consideration:
Each transmitter is fully calibrated before shipment; zero and span
adjustments are not necessary for measurement.
Frequently Asked Questions (FAQ):
-
Q: Do I need to calibrate the transmitter before use?
A: No, each transmitter is fully calibrated before shipment, so calibration is not required before measurement. -
Q: What is the humidity measurement range of the transmitter?
A: The humidity measurement range is from 10% to 95% RH.
FEATURES
- Two wires 4 to 20 mA humidity transmitter, easy operation.
- 2 wires, both for signal and power supply.
- High precision capacitance type humidity sensor.
- Protection for the reverse wires connecting.
- Use the ” two wires, 4 to 20 mA humidity transmitter ” can cooperate with the optional controller/Indicator or the datalogger system.
SPECIFICATIONS
General Specification
Humidity measurement range | 10% to 95% RH |
---|---|
Signal output | 4 – 20 mA |
10 %RH = 5.6 mA
50 %RH = 12 mA
95 %RH = 19.2 mA
Humidity sensor| High precision capacitance type humidity sensor.
Power Supply| DC 9V to 30 V
- Protected against reverse polarity
Wire numbers| 2 wires, both for signal and power supply.
Accuracy| ≧ 70 %R.H. :
± (3% reading + 1% R.H.)
<70 %R.H. :
± 3%R.H.
Load impedance requirement | Load up to 500 ohms. |
---|
Operating
Temperature
| 0 ℃ to 50 ℃
( 32 ℉ to 122 ℉ ).
Operating Humidity| Less than less than 95% RH.
Size| 200 x 23 x 19 mm
( 7.9 x 0.9 x 0.7 inch ).
Weight Cable length| 75 g ( 0.16 LB ).
Approx. 1.2 m.
Wire Connection| 2 wires :
Red……………….. Power + and signal
Black……………… Power –
Standard accessories| Instruction manual……………… 1 PC
Probe holder………………………. 1 PC
Screws ( for fixing the probe
holder )…………………. 2 PCs
Electrical Specification ( 23±5 ℃)
Accuracy| ≧ 70 %R.H.:
± (3% reading + 1% R.H.)
<70 %R.H.:
---|---
± 3%R.H.
Checklist of input (% RH) and the output (4 – 20 mA)
Input ( %RH ) | Output ( mA ) |
---|---|
0% RH | 4 mA DC |
10% RH | 5.6 mA DC |
33% RH | 9.28 mA DC |
75% RH | 16 mA DC |
95% RH | 19.2 mA DC |
100% RH | 20 mA DC |
FRONT PANEL DESCRIPTION
- 3-1 Humidity sensing head
- 3-2 Probe main body
- 3-3 Signal and power supply wires
- 3-4 Probe holder
OPERATION PROCEDURES
- Using the two screws to fix the ” Probe holder ” ( 3-4, Fig. 1 ) on the wall Mount the ” Probe main body ” ( 3-2, Fig. 1 ) to the ” Probe holder ” ( 3-4, Fig. 1 ).
- Use the two wires ( 3-3, Fig. 1 ) as signal and power supply wires.
The general wire installation will be as follows:
Calibration (zero/span adjustment) consideration:
Each transmitter had been fully calibrated before shipment, zero and span
adjustment are not necessary before measurement.
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>