LORDERAN BJPX-SDW10-20 Stackable Incubated Refrigerated Shaker User Manual
- May 15, 2024
- LORDERAN
Table of Contents
LORDERAN BJPX-SDW10-20 Stackable Incubated Refrigerated Shaker
Preface
Thanks for choosing LORDERAN products!
For your safe and convenient use and reasonable maintenance of this product,
please read the instruction manual carefully before use, and keep it in a safe
place for reference.
For any damage to the instrument caused by the user not following the
“instrument use environment” stated in this manual, or personal failure to
operate per the requirements of the “safety reminder”, LORDERAN has no
obligation and responsibility for this.
The user must do the following three points in use:
- Always use protective equipment correctly (including clothes, gloves, goggles, etc.);
- Always adopt good hygiene habits and operate in strict accordance with product instructions;
- Everyone is obligated to take responsibility for their safety.
Notes: Because LORDERAN products are updated quickly, if the functions described in this manual are different from the functions of the product you purchased, please refer to the actual function.
Safety notes
- Please read this manual carefully when using this product for the first time.
- Stacking horizontal constant temperature shaker can only be operated by trained and authorized personnel.
- The maintenance of the equipment can only be done by the company’s technical engineer or the company’s authorized agent.
- If the operator encounters a situation not mentioned in this manual, please contact the company’s technical engineer or the company’s authorized agent to inquire about the correct handling method.
- Please use the accessories provided by LORDERAN. If the user wants to use other accessories, LORDERAN will not be responsible for the adverse consequences. However, users can apply to Brocade to verify whether the accessories meet LORDERAN requirements.
- Inspection and maintenance must be carried out on the stacked horizontal constant temperature shaker at specified intervals.
- The stacked horizontal constant temperature shaker heats the box through the back to achieve the constant temperature of the box. To ensure the normal operation of the machine and ventilation and heat dissipation, the back left and right sides of the box shall be at least 30cm away from the wall, and the air inlet and outlet shall not be blocked by obstacles.
- In case of machine failure or power failure, the temperature in the working chamber of the superimposed horizontal constant temperature shaker will fluctuate to some extent. If it cannot be repaired in a short time, please take out the culture medium and transfer it to other places that meet the requirements for the preservation temperature of the culture medium for storage to avoid damaging the culture medium.
- The places marked with symbols in the equipment need to refer to this manual when in use, to clarify the potential nature of the hazard and the countermeasures that must be taken.
- Please use the equipment under the methods specified in this manual. If you do not use the equipment under the methods specified by the manufacturer, the protection provided by the equipment may be damaged.
Overview
Structural diagram of superimposed horizontal constant temperature shaker Due to product improvements and different models, the actual product may be different from the sketch, please refer to the actual product.
Working principle
The superimposed horizontal constant temperature shaker is a high-performance product with multi-unit superposition, multi-function combination, multi- amplitude gyration and multi-temperature zone control, which is specially developed for small laboratories. It can be placed on a single platform or two or three platforms on top of each other, to realize the loading configuration of multiple 500ml triangular flasks when the floor area is less than 0.6 m2. One device can meet the experimental requirements of dynamic culture and static culture at the same time, It can greatly meet the experimental requirements of laboratory personnel for miniaturization, large loading and flexibility of products, and is widely used in bacterial culture, fermentation, hybridization, biochemical reaction, enzyme and cell tissue research with high requirements for temperature and oscillation frequency. It has extensive and important applications in medicine, biology, molecular science, pharmacy, food, environmental protection and other research fields Scope and application: Cell Culture, Extraction Experiment, Sample Mixing, Staining and Decolorization, Hybridization Experiment, Solubility Experiment, Plasmid Purification, Bacterial Suspension Preparation, Elution Process.
Main technical parameters
The stacking horizontal constant temperature shaker is composed of an instrument cabinet, storage box, oscillating device and temperature control system.
Product Model
|
BJPX-SDW10
|
BJPX-SDW20
---|---|---
Temperature Control
Range ℃
|
RT+5~60
|
4~60
Display Mode| 5-Inch Touch Screen
Control Mode| P.I.D
Temperature Resolution
Accuracy ℃
|
0.1
Temperature Fluctuation
Degree ℃
|
≤±0.5
Temperature
Uniformity ℃
| ≤±1 ( 37℃ )
Oscillation Mode| Gyratory
Drive Mode| Single Shaft Balanced Drive Type
Speed Range (r/min)| 30~300
Speed Accuracy ( r/min)| ±1
Swing plate Amplitude
(mm)
|
26
Timing Range (min)| 0~999
Rocker Size (mm)| *720490
Number Of Rocker Plates| 1**
Maximum Single Configuration
| *250ml 34 /500ml 24 /1000ml 15 /2000ml 8 /3000ml 6 Pieces**
(Note: If 3000ml is configured, 1. It is necessary to install a wide-mouth flask with a height of ≤ 310mm; 2. When selecting a universal spring clamp, it cannot be installed)
Volume ( L )| 190
Working Room Size
( mm )
| 870590398 ( Unit 1 )
External Dimensions
( mm )
| 1150780640 ( Unit 2 )
Net Weight ( Kg )| 270| 300
Power (W)| 1500W| 2200W
Power Supply| AC200~240V 50/60HZ
Other| Lighting Lamp (1 Piece)
---|---
Sterilization| Ultraviolet Lamp (2 Piece)
Ambient Temperature
Requirement ℃
|
15~35
Maximum Stacking
Quantity/Set
|
3
Data Storage| USB
Studio Materials| High Quality 304 Mirror Stainless Steel
Door Opening Mode
|
Up/Down Swing Door
Note: “220V~” means that the equipment requires 220V AC.
Prepare Notes before Operation
Transportation and handling requirements
- The stacked horizontal constant temperature shaker shall not be transported upside down to avoid shaking and collision during transportation and damage to electrical components inside the chamber In addition, wooden brackets shall be installed at the bottom of the equipment during long-distance transportation, anti-collision corner linings shall be added around the equipment, direct sunlight protection and rain protection devices shall be added, and foot supports shall be locked;
- Lift from the bottom during short-distance transportation, the inclined surface should not be greater than 45 degrees, handle with care, and do not grab the door body or the lining as a stressed part;
- On flat ground, you can directly push the superposition-type horizontal constant temperature shaker to move. When pushing the superposition type horizontal constant temperature shaker, you should release the foot support from the locked state, and pay attention to the casters so as not to damage the power cord. After the movement is completed, you will lock the foot support again to prevent the superposition-type horizontal constant temperature shaker from moving at will
Storage environment requirements
- Indoor storage;
- Storage environment temperature: -4℃~30℃;
- Relative humidity of storage environment: ≤85%;
- Do not place it in a place where the temperature is too high and it is easy to splash;
- The placing surface must be solid, horizontal, incombustible and able to bear the weight of the superimposed horizontal constant temperature shaker;
Operation environment requirements
- Operating environment temperature: 5℃~35℃;
- The relative humidity of the operating environment: ≤80%;
- Rated power supply voltage: 220VAC;
- Rated voltage frequency: 50Hz;
- The installation surface must be solid, horizontal, incombustible and able to bear the weight of the superimposed horizontal constant temperature shaker during operation;
- When in use, it should be placed in a place away from direct sunlight and away from heat sources;
- It is required to leave more than 30cm space around the stacked horizontal constant temperature shaker for ventilation and heat dissipation;
- Do not place it in places where the temperature is too high or splashing water;
- There is no strong vibration or corrosive gas around during use;
Safety notes
- Do not grab the door body or lining as a stressed component;
- Remove all packaging components (including the protective foam in the box and the base of the box) before use;
- Please check the random attachments and materials according to the packing list before use;
- The product should be cleaned once before use;
- Power supply voltage: This equipment uses a 220V/50Hz AC power supply. If the voltage is lower than 198V or higher than 242V, a suitable automatic voltage stabilizer must be installed for use;
- When using this superposition-type horizontal constant temperature shaker, the power supply shall be equipped with a low-voltage air circuit breaker and leakage protection device;
- Dedicated independent sockets must be used and reliably grounded to ensure good grounding after the equipment is connected. Do not extend the length of the power cord arbitrarily. If you really need to extend it, you must use a copper core wire of 2.5mm² or more. The cross-sectional area of the copper core wire in the wall connected to the power socket must also be guaranteed to be above 4mm²;
- It is strictly prohibited to put flammable, explosive dangerous goods, highly corrosive acids, alkalis, etc. into the stacked horizontal constant temperature shaker;
- Do not connect the neutral wire (N end) and the ground wire (E end) of the socket together, otherwise, the shell of the platelet constant temperature incubator will be electrified and an electric shock may occur;
- The power cord cannot be used in bundles, placed under heavy objects, or close to heat sources such as compressors;
- Do not place the equipment where it is difficult to operate the disconnect device;
Note: The stacked horizontal constant temperature shaker just connected to the power supply should not be put into the culture medium immediately. After the temperature of the incubator is stabilized, the culture medium should be put into the incubator for a culture test.
Installation and Debugging
Remove packaging materials and tape
Remove all shipping packaging materials and tape, open the door and ventilate
the equipment. If the shell panel is dirty, please use neutral detergent to
remove the dirt and clean the residual neutral detergent with clean water
(Undiluted detergent will destroy the composition of the plastic, please refer
to the detergent’s manual to dilute the detergent). After cleaning, wipe it
with a damp cloth, and then wipe the panel with a dry cloth.
Adjust the level foot and fixed foot cup
To adjust the foot support, first use an adjustable wrench to screw down and
adjust the nut. After landing horizontally, screw up and tighten the nut to
contact the lower part of the box for locking.
Ground
Warning
Please use a power outlet with a ground wire to prevent electric shock. If the
power socket is not grounded, qualified engineering and technical personnel
must install the grounding wire. Do not ground the equipment through gas
pipes, power supply pipes, telephone lines or lightning rods. This type of
grounding may cause electric shock due to incomplete loops.
For first use
- After removing the package of the stacked horizontal constant temperature shaker, place the stacked horizontal constant temperature shaker (lock the foot support) according to the requirements of the use environment
- The box has been cleaned when leaving the factory, but it is still recommended to use medical alcohol to wipe the stacked horizontal constant temperature shaker and dry it;
- Connect the superposition horizontal constant temperature shaker to the power supply, click the key on the lower part of the display screen on the right side, and the main interface will light up for operation (all parameters have been set for direct operation). The oscillating device starts to operate, and the box temperature tends to be stable in about 30 minutes, indicating that the system is working normally and the commissioning is over;
- For initial use, it is recommended to put the culture medium into the box after the temperature in the box reaches the use temperature;
- To save electricity, try to reduce the number of times and time of opening the door (multiple operations may also damage the door seal and affect the sealing effect of the door body).
Safe operation and preventive measures
-
This manual contains important safety regulations, please read them carefully and follow them strictly. All the items and procedures introduced here are intended for users to use this equipment correctly and safely. If the precautions described here are followed, the user and anyone else will be protected from possible injury.
-
Danger (may cause serious property loss or personal injury)
-
This product must be grounded reliably and far away from electromagnetic interference sources (never use the neutral wire or the neutral wire as the ground wire).
-
Please use a power supply that conforms to the electrical parameters indicated on the nameplate of this equipment.
-
Do not insert metal objects such as nails or wires into any holes and gaps of the device or any exhaust ports for internal air circulation, otherwise, accidental contact between the above objects and moving parts may cause electric shock or injury.
-
The product is not allowed to arbitrarily unplug or plug in the power plug without turning off the power switch during operation.
-
Do not damage or damage the power plug or power cord. If you unplug the plug from the power socket, you should hold the power plug firmly instead of pulling the power plug wire. If the plug wire is loose, do not use the power plug. It is not allowed to connect or cut the power cord at will, otherwise it may cause fire or electric shock.
-
Users are not allowed to disassemble, repair or modify the equipment by themselves. If any of the above operations are carried out by unauthorized personnel, it may cause fire or personal injury due to improper operation.
-
Do not store volatile or flammable materials in this equipment. Otherwise, it may cause an explosion or fire.
-
No objects should block the circulation holes of the air duct during use, or make the circulation of the air duct not smooth;
-
Warning (may cause property damage or personal injury)
-
The user must fully read and understand the instruction manual of this product before proceeding.
-
The 304 stainless steel inner tank is not acid-resistant. Please pay attention to anti-corrosion measures and do not use the acidic medium in the box.
-
When performing the following operations, the power cord of this product must be unplugged:
-
Replace the fuse;
-
The product fails to be inspected and repaired;
-
The product will not be used for a long time; d. Move the product.
-
Please use a power socket with a ground wire to prevent electric shock. If the power socket is not grounded, it must be installed by qualified engineering and technical personnel.
-
Attention (may cause the product to not work normally or affect the service life)
-
When the product is being transported, care should be taken to avoid damaging the vulnerable parts such as the instrument on the panel;
-
The equipment should be installed on solid ground to keep it level. If the ground is not solid enough or the installation location is not suitable, the equipment may tip over or overturn and cause personal injury.
-
The product should be wiped off the condensed water inside the cavity after each storage to avoid corrosion of the cavity and affect its service life.
-
Do not open/close the door of the product box by gravity, otherwise, it will easily cause the door to fall off and damage the product;
-
Do not apply extra pressure to the glass door of the box door or scratch the glass surface with sharp objects, otherwise, it will easily cause the glass to break or produce scratches;
-
Do not place containers or heavy objects filled with water on the product, avoid water splashing on the product and causing short circuits or electric shock hazards, or heavy objects slipping and damage;
-
This product must not be used outdoors.
-
Installation position
To make the equipment operate normally and obtain the best performance, please install the equipment in a place that meets the following conditions: -
A place that will not be directly affected by direct sunlight or airflow from the air conditioner.
-
A place with clean air and sufficient ventilation (do not install in a small, closed room).
-
A place far away from the heat source.
-
A place with a solid and level ground.
-
A place where no flammable or corrosive gas exists.
-
A place is not prone to high humidity.
Warning
Do not use this device outdoors. If this equipment is splashed by rain, it may
cause electric leakage or electric shock. Do not install this equipment in a
humid place or a place where water may be splashed, otherwise, it may cause
electric leakage or electric shock due to reduced insulation.
Warning
Do not install the device in a place where flammable or volatile gas exists,
otherwise, it may cause an explosion or fire. Do not install the equipment in
a location with acid or corrosive gas, otherwise it may cause leakage or
electric shock due to corrosion.
Operation
Menu interface
After the user logs in, click〖Menu〗on any interface to enter the directory
interface. In the directory interface, the user can enter history data】, 【Mode
Set】, 【Parameter Set】and【System Setting】 interface.
Note: Given the special storage requirements after the platelet thawing
is completed, to avoid the system operating parameters disorder caused by the
wrong operation during use by the customer, the instrument will set all the
operating parameters before leaving the factory, without the customer’s
secondary parameter settings. Therefore, customers only need to perform simple
operation operations (the three interfaces of 【Mode Set】, 【User Settings】 and
【System Settings】 do not require customers to perform touch operations), If
there are special requirements for use or problems with the instrument itself,
you need to set the system operating parameters and modes, and you can contact
professional after-sales personnel to authorize one-to-one video guidance
operations.
Main interface
- In【main】interface, users can view or set common data and carry out regular operations.
Icon Description
Key Description
Running state
- The two operating states of the system are displayed in the upper left corner of the [Main Interface], namely [Running] and [Stop running].
Running
- Click the [Run] button and the system directly enters the [Running] state, at which time the machine will directly enter the running state.
Stop running
After the running timing ends, the system closes all outputs and enters the
state of【Stop running】.
System time setting
When there is an error in the system time, you can click the system real-time
time in the [Main Interface] to correct the system time, as shown in the
figure below:
Alarm function
When an alarm occurs in the system, the buzzer will give a warning, and the
e【alarm】in the monitoring interface will turn red and blink. Click to enter
the alarm list interface and check the specific alarm contents.
Temperature alarm
The temperature alarm includes 【over-temperature alarm】, 【lower-temperature
alarm】and 【temperature overflow】, as shown in the following table:
Alarm type | Instructions |
---|
Over temperature alarm
| When “measured temperature > set value + over temperature alarm value”, there will be a temperature deviation alarm and the buzzer will beep;
Lower temperature warning
| When the “measured temperature < set value + under temperature alarm value”, there is a temperature deviation alarm, set to 0 means there is no temperature deviation alarm, and the buzzer beeps;
Temperature overflow
| When the sensor temperature fault occurs, a temperature overflow alarm will be generated, and the buzzer will continuously sing;
Other alarms
Alarm type | Instructions |
---|
Motor Alarm
| The motor fault alarm includes communication faults, power module faults,
stalling, Hall errors, bus under voltage, and bus over voltage. When there is
an alarm, the motor will stop running, the alarm light will be on, and the
buzzer will beep continuously.
Door timeout alarm| When the opening time of the door exceeds the delay time
of the opening alarm, the alarm light will be on and the buzzer will beep.
Operation example
For example: set temperature 22.0℃, set speed 60RPM, set time 1H:30M. The
setting operation flow is shown in the figure below:
Note: After leaving the factory, all parameters have been set except for the [set time] option, and there is no need for the user to perform secondary settings.
No. | Content | Instructions |
---|---|---|
1 | Temperature |
setting
| Click then can set the temperature value
2| Speed setting| Click then can set the motor speed value
3| Time setting| Click the time text box (hours: minutes) and set it to
0:0 means it runs all the time
---|---|---
Curve interface
Real-time temperature curve
Interface instruction
Users can select the display of temperature curves and speed curves in the
real-time temperature curve interface. The ordinate range is 10.0℃ and 50RPM
above and below the set value (when the set value changes, the ordinate
changes and the real-time is cleared.), The horizontal coordinate range is
fixed with the data in the last 2 hours (the refresh rate is adjusted
according to the actual situation).
History curve
Interface instruction
Users can click the【History data】in the【Menu】and enter the【History
data】interface. The automatic temperature measurement, setting value, speed
measurement, setting value, operation and alarm status historical data saved
by the system can be viewed.
The user clicks the [historical data] button in the [directory interface], then clicks the [historical curve] button or slides left to enter the [data curve] interface. [Data curve] Each page of the interface displays a 6-hour historical data curve, temperature curve and speed curve that can be selected, can zoom in and out vertically, move up and down the curve, and can move left and right (curve page, time axis time changes accordingly );
History Data
Interface instruction
Users can click the【History data】in the【Menu】and enter the【History data】interface. The automatic temperature measurement, setting value, speed measurement, setting value, operation and alarm status historical data saved by the system can be viewed.
State Description
When you re-enter the the【Data list】interface, update the last export time,
and when a USB stick is inserted, the USB stick connection status will be
updated to “connected”
Key Description
Export data
If the user needs to export the history data with the U disk, he/she can
click【Export】to pop up the dialogue box confirming whether to export or not.
The user should confirm that the U disk has been inserted to avoid system
errors. Click OK to complete the data guide operation, and update the last
export time after the export is completed. Such as the right:
The exported file is saved in the U disk root, with the current real-time time
as the file name, the file format is.DAT and the data format is a hexadecimal
value.(up to 150000 historical data can be stored, and the maximum export time
is about 3 minutes and 30 seconds).
Note: After the data export is completed, connect the USB flash disk to
the computer and use the data analysis software (data analysis software
v1.16.1) to export the excel data file. You can also directly use the software
to analyze and view the data. Contact the company for after-sales processing
for software acquisition.
Delete the data
If the user feels that there is too much data exported to the U disk, he can
click the 【delete】 to pop up the confirm whether to delete dialogue box.
Note: after deletion, the next U disk export data will be cleared to 0, please
handle this with caution.
Lighting Function
Click the light button on the main interface, and the light will turn on,
press it again, the light will go out, and the light will automatically turn
on after opening the door. If you need to operate after closing the door, it
is recommended to turn off the light manually, otherwise, it will affect the
overall temperature of the studio.
Sterilization Function
Click on the Sterilization button on the main interface, and the Sterilization
Timing dialog box appears, enter the timing time (the recommended disinfection
time is 1h), click OK, and the UV lamp will light up. If you need to pause the
disinfection, you can press it again and click OK. At this time, the UV lamp
is off. It is also recommended to open the door first to reduce the humidity
when the UV function is turned on. (When the machine is running, the
compressor is cooled for a long time, and the humidity in the box may be
high), so that the dry degree in the box can reach the normal indoor humidity
value, otherwise the high humidity may cause the UV lamp to short-circuit,
causing the machine to power off and cause accessories damaged, affecting
normal use.
Safety notes
-
Since the superimposed horizontal constant temperature shaker has a refrigeration system, the distance between the back of the machine and the wall should be more than 30cm, and the distance between the two sides of the box should also be more than 30cm to ensure the normal heat dissipation of the refrigeration system and the air circulation around the instrument..
-
When the equipment is transported in a short distance, the inclination angle shall not be greater than 45° to avoid damage to the equipment’s refrigeration system.
-
After the equipment is transported and placed in place, it should be left for (1-2) days before starting up, which is conducive to the normal operation of the equipment refrigeration system and prolongs its service life.
-
The equipment must be connected to a well-grounded power outlet and placed flat.
-
It is strictly prohibited to use corrosive liquid for disinfection on the superimposed horizontal constant temperature shaker.
-
Please turn off and cut off the power supply when moving, repairing and maintaining the stacked horizontal constant temperature shaker.
-
Use environment requirements:
-
Avoid locations with high temperatures and high humidity
Avoid places with high temperature and high humidity, because the number of bacteria in the air in such places is far more than the number of bacteria in the normal ambient air. If necessary, install air conditioners in the room to avoid operating the instrument in a high temperature and high humidity environment. -
Avoid places with excessive ventilation and places where many people pass by Avoid locations near the door, air conditioner, fan, etc., because the breeze in such locations may make it easier for bacteria to enter the incubator.
-
Installed in a sterile room
To achieve a better preservation effect, please place the equipment in a sterile room as much as possible. -
Use a clean medical flask/culture dish
The pollution is mainly caused by the medical flasks/Petri dishes stored in the cavity. Therefore, please be sure to purchase the medical flasks/Petri dishes produced by manufacturers that are qualified for market sales, qualified for national inspection, and disinfected normally.
-
-
Pay special attention
Before putting it in the culture medium, make the superposition horizontal constant temperature shaker run and reach the corresponding operating conditions of temperature, oscillation amplitude and frequency, then put in the culture medium for culture, mixing and other test operations!
Troubleshooting
The abnormal conditions of the superposition-type horizontal constant temperature shaker are caused by improper use. Before commissioning maintenance, please check and eliminate by yourself according to the following table. Before troubleshooting, please confirm whether the power supply is properly connected, whether the power cord is damaged, and whether the fuse is in good condition.
Fault | Checking parts | Methods |
---|
The button touch is not working
|
Control board
| Check whether the power cord is energized is
normally
Check if the display is damaged
Check whether the connection between the display screen and the control board
is normal
Check whether the control board is working normally
The device is not powered
| Power supply| Check whether the power cord is connected properly
Power line| Check whether the power cord is obvious
damage
Fuse| Check whether the fuse is in good condition
Switching power supply| Check whether the output of the switching power supply
is normal
Control board| Check whether the control board is working
properly
No cooling
| The compressor does not start| Overheating protection, the power supply is not started, the compressor motor coil is
damaged
Compressor starts without
cooling
| Lack of refrigerant
| Heating does not work,
The circulation fan does not rotate
| Check the heating circuit Check the circuit and replace the fan
---|---|---
The evaporator is too icy to exchange heat and cold| Warming up the ice to
reduce the number of door openings and the possibility of air leakage
The fan is not working
| Micro switch| Check whether the micro switch is damaged
Fan| Check whether the fan is damaged
Line| Check whether the fan circuit is intact
Control board| Check whether the control board is working normally
Abnormal temperature
|
Compressor
| Insufficient cooling capacity, leakage of refrigerant, check whether the
compressor and compressor circuit are damaged
The use environment does| Improve environmental conditions
Heating tube| Check whether the heating tube and the heating tube circuit are
damaged
Fan| Change the fan
Sensor
| Built-in temperature tester to check whether the temperature in the box is
consistent with the displayed temperature
The glass door is not closed tightly|
Sealing strip
| Check whether the whole glass door is loose Check whether the door sealing strip is warped or aging
Big noise (more than 70dB) or there are abnormal sounds in operation
| The fan is broken or friction between the fan and the air duct plate| Change
the fan or increase the distance of the air duct
The compressor (forced exhaust type) fan is broken| Sundries should be
removed, or the fan should be replaced
The box is not stable| Place the pad flat
The compressor is not firmly fixed| Check whether the bolts are loose and form
resonance
The operation of the above electrical components must be carried out by a
qualified electrician under safe conditions (cut off the power supply), and
other components are not allowed to be disassembled by themselves. If it is
disassembled without the consent of the technical engineer or agent dealer,
LORDERAN will not be responsible for the adverse consequences; When the
equipment has a failure other than the above, please notify the maintenance
department of our company immediately. For your safety, please do not repair
the equipment by yourself; If you need to order parts, you can find our
technical service department. Please indicate the model and factory number of
the superimposed horizontal constant temperature shaker you purchased.
Note: If the user cannot judge the fault, please contact the agent
engineer as soon as possible, and do not handle it by yourself, otherwise the
warranty will not be given.
Examples of simple parts replacement, such as replacing the fuse (fuse): The
fuse is located at the rear of the left side of the instrument. When
replacing, first turn off the power supply and unplug the plug, use a cross
screwdriver to press counterclockwise and unscrew the fuse socket, remove the
fuse in the fuse socket and replace it with a new fuse, and then press
clockwise and unscrew the fuse socket; The live wire fuse is located at the
lower right side of the incubator. Use a slotted screwdriver to open the fuse
socket and replace it with a new fuse, and then press it back (rated value and
characteristics of the power socket zero wire fuse: F10AL250V, rated value and
characteristics of the power socket live wire fuse: F10AL250V).
Except for the insurance tube, the rest of the parts cannot be replaced by the
user. If you need to replace other parts, please contact our company’s
professional after-sales engineers or agent dealers for replacement.
The final interpretation right belongs to the production and manufacturing
company.
Maintenance
Frequency | Operation |
---|---|
Every day | Clean the inner chamber, swing plate and |
flask clamp
Every week| Cleaning the door seal
Every 1-3 years| Replace the door sealing strip and Z-belt
Wearing parts| Ultraviolet lamp (including lamp holder), lighting lamp
---|---
Fuse
- Clean the chamber and sealing strip in the working chamber Use cotton cloth or towel soaked in distilled water to clean the entire inner cavity and sealing strip, Wipe the contaminated inner cavity and the surface of the sealing strip with cotton cloth soaked in medical alcohol (75% alcohol concentration). After wiping all the foreign objects clean, use a clean dry cotton cloth or towel to wipe the moisture on the inner cavity and the sealing strip, and let it stand for half an hour before normal use.
- Please use protective equipment (including clothes, gloves, goggles, etc.) when cleaning with alcohol;
- Please use the designated reagents for cleaning. Do not use cleaning agents or disinfectants that chemically react with equipment parts or materials contained in the equipment to cause danger. If you have any questions about the compatibility of the disinfectant or cleaning agent with the equipment parts or the materials contained in the equipment, please contact a technical engineer or an agent dealer;
- If dangerous substances leak on the surface of the equipment or enter the inside of the equipment, use medical alcohol for thorough cleaning, and then use a clean dry cotton cloth or towel to absorb the residual moisture before continuing to use.
- Clean the external surface and glass door
- After cleaning the contaminated surface with medical alcohol or thinner, wipe it with a soft cotton cloth or towel.
- The cycle of comprehensive maintenance
- It is recommended that the overall maintenance cycle is one week or 100 working hours.
- Maintenance method
- Daily or weekly maintenance
- Disinfection and cleaning of the operation area (refer to the first description);
- Clean the external surface and glass door around the operating area (refer to the second description); Check whether the various functions of the equipment are abnormal;
- Record this maintenance;
- Monthly maintenance
- Clean the exterior surface and glass door (refer to the first description); 75% medical alcohol must be used to wipe the inner and outer surfaces of the box, the perspective window, the swing plate and the flask clamp of the equipment. After wiping, use cotton cloth soaked in distilled water for secondary cleaning. After cleaning, use a dry cotton cloth to absorb the residual water and ventilate to dry.
- Check whether the various functions of the equipment are abnormal; Record this maintenance.
- Annual maintenance
- Check that the left and right corners of the front glass door are consistent and parallel;
- Check whether there is fluorine leakage in the compressor pipeline;
- Check whether there is water leakage or blockage at the leakage pipe and interface of the stacked horizontal constant temperature shaker;
Record this maintenance
After moving the box each time, let it stand for 24~48 hours before it can be
used normally.
When moving the box every time, try to keep the box parallel to the ground,
and the inclination angle should not be greater than 45°.
Note: The above-mentioned vulnerable parts such as UV lamps and lighting
lamps are consumables, and their service life is not covered by the warranty.
Please check the manual for details.
Circuit diagram
Label Description
- Company logo
- Fuse label
- Ground label
- UV lamp safe use label
Warranty
- The warranty period is twelve months from the date of purchase (excluding the wearing parts indicated in this document).
- If the instrument and equipment during the warranty period become invalid or damaged due to improper use by the user, the company will not undertake the warranty obligation.
- One year later, in the spirit of serving users, we will try our best to provide users with convenience.
- The equipment service life is 6 years, and the production date is.
- Provide equipment drawings and some necessary technical data to maintenance organizations and maintenance personnel trained and approved by the company.
Appendix A
Packing list
No. | Name | Qty. | Remarks |
---|---|---|---|
1 | Stacking horizontal shaker | Host | 1 |
2 | Instructions | File | 1 |
3 | Warranty Card | File | 1 |
4 | Certificate | File | 1 |
5 | Packing List | File | 1 |
6 | 10AFuse Tube | Parts | 2 |
7 | Flask Holder ( 500ml ) | Parts | 24 |
8 | Power Cord | Parts | 1 |
The items listed in this list are consistent with the actual items in the box.