VARIOS QUICK START
Preparation
Confirm that all ordered components are intact. You will need Microsoft®
Windows® computer (Windows 10 or newer), LABROS SoilView software, and LABROS
SoilView-Analysis software to start with. Additional tools are identified in
the VARIOS User Manual (meter.ly/varios-manual).
When the VARIOS is used in combination with the LABROS Balance, place the
Balance on a horizontal, vibration-free, solid surface where the temperature
remains fairly stable (away from the air conditioner and heater vents,
windows, etc.)
IMPORTANT: Temperature fluctuations or vibrations will strongly affect
the accuracy of the measured results.
Remove the plate of the LABROS Balance, and place the VARIOS on top of the
scale. Read the full VARIOS User Manual at (meter.ly/varios-
manual). All products have a 30-day
satisfaction guarantee.
Setup
VARIOS Dryout
Continuous measurement of the thermal conductivity and sample weight loss for
determining water content.
Connect the VARIOS to the LABROS Balance (VARIOS ready) by inserting the
VARIOS connector cable into the back of the LABROS Balance (VARIOS ready).
NOTE: Please use the VARIOS Power Adapter (ordered separately) when using
LABROS Balances without VARIOS ready label.
Download the LABROS SoilView Installation package from the support webpage
(meter.ly/varios-support).
Run the LABROS SoilView.exe file and follow the installation instructions.
VARIOS Standalone
Measurement of the thermal conductivity without simultaneous weight recording.
Measurement–VARIOS Dryout
-
Prepare Soil Sample
Fill a 250 mL soil sampling ring, equipped with two bore holes.
Saturate the soil sample.
-
Install Sensor
Insert the Sensor Needle completely into the soil using the Installation Tool (order separately) to ensure the sensor is installed straight without air gaps.
Wait 15 min after installing the sensor to reach temperature equilibrium between needle and soil.
-
Take a Reading
Install and open LABROS SoilView and SoilViewAnalysis software.
Connect devices.
Configure a measurement by selecting a sample name, an output folder, and a measurement profile.
Select the Start button.
-
Evaluate Reading
Open the generated *.varios file.
Enter the soil sample ring weight and the sample dry soil weight, if applicable.
Fit the data with a suitable fitting model.
Export data.