Viatomtech SA-10BW Continuous Pulse Oximeter Fingertip User Manual

June 16, 2024
Viatomtech

Viatomtech SA-10BW Continuous Pulse Oximeter Fingertip User Manual

Download the APP software

You can view the measurements and record list in the APP.
Scan the below QR code to download the APP software for iOS and Android system.
Please pair the APP with your oximeter according to the number displayed in the lower right corner of the screen after the oximeter is turned on.

Instructions for Safe Operation

M It’s not a medical device. This device is for Sports and Aviation use

only and not intended for medical use.

  • M The Oximeter is not intended for medical use.
  • M The Oximeter is not an apnea monitor and should not be used for arrhythmia analysis.
  • M Do not self-diagnose or self-medicate on the basis of the measurements. Always consult your doctor.
  • M Make sure that there is no visible damage that may affect user’s safety or measurement performance with regard to sensors and clips. It is recommended that the device should be inspected minimally before each use. If there is obvious damage, stop using the device.
  • M Special attention should be paid while the oximeter is used constantly under the ambient temperature over 37°C, burning hurt may occur because of over-heating of the sensor at this situation.
  • M Necessary maintenance must be performed only by qualified service technicians. Users are not permitted to service this device.
  • M The oximeter must not be used with devices and accessories not specified in User Manual.

Warnings and Cautions

  • Explosive hazard— DO NOT use the Oximeter in environment with inflammable gas such as some ignitable anesthetic agents.
  • DO NOT use the Oximeter while the user is under MRI or CT scanning. This device is NOT MRI Compatible.
  • Discomfort or pain may Appear if using the Oximeter continuously on the same location for a long time, especially for user with poor microcirculation. It is recommended that the Oximeter should not be Applied to the same location for longer than 2 hours. If any abnormal condition is found, please change the position of Oximeter.
  • When using the device for continuous prolonged monitoring, the position to which the oxygen probe (sensor) is attached should be checked every 2 hours when the ambient temperature exceeds 37°C, and it is advisable to change the position of the wearing device or move it every 2 hours in the event of skin changes. More frequent examinations may be required for certain patients, such as neonates, patients with perfusion disorders or skin sensitivities.
  • It is not recommended to use the device continuously when the ambient temperature (including the patient’s body temperature) exceeds 41°C, as prolonged and continuous use may increase the risk of unanticipated skin changes such as sensitization, reddening, blistering, or pressure necrosis.
  • The light (the infrared light is invisible) emitted from the device is harmful to the eyes. Do not stare at the light.
  • The Oximeter is not a treatment device.
  • Local laws and regulations must be followed when disposing of the device.
  • Keep the Oximeter away from dust, vibration, corrosive substances, explosive materials, high temperature and moisture.
  • The device should be kept out of the reach of children.
  • If the oximeter gets wet, please stop using it and do not resume operation until it is dry and checked for correct operation. When it is carried from a cold environment to a warm and humid environment, please do not use it immediately. Allow at least 15 minutes for Oximeter to reach ambient temperature.
  • DO NOT operate the button on the front panel with sharp materials or sharp point.
  • DO NOT use high temperature or high-pressure steam disinfection on the Oximeter. Refer to Chapter 8 for instructions regarding cleaning and disinfection.
  • Pay attention to the effects of lint, dust, light (including sunlight), etc.
  • Please keep the cable away from children. It can cause strangulation.

FCC Rules

This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.

Overview

Intended Use

This Pulse Oximeter is intended for measuring the pulse rate and oxygen level (SpO2) through a user’s finger.

Views

Features

  • Display oxygen level, PR, and Plethysmogram
  • Auto power On/Off
  • Setting menu
  • Pulse beep
  • Wireless function

Charging

Charge the battery before using.
Connect the device to computer USB or USB charging adapter with USB cable.
Note: The device cannot be used during charging, and if choosing a third party charging adaptor (Class II), select one that complies with IEC60601-1 or IEC60950-1.
: Fully charged.
: The filled part represents the remaining power. If the filled part moves from left to right, the device is charging.  : Low battery. Please charge the device
Note: Please use the accessories that are original or Approved by our company.

POWER ON/OFF

POWER ON:

Wear the device, it will turn on automatically.

POWER OFF:

Take the device off.

  • It will turn off automatically after 2
  • On the menu interface, if there is no key operation for about 30 seconds, the device will automatically exit the menu and then shut
  • On the recording and playback screen, if there is no key operation for 6 seconds, the device will automatically shut

Start/Stop Measuring

  1. Open the clip and put finger inside the clip (make sure the finger is in full contact with the deep inner side of the clip), and then release theclip.
  2. Wait for 2 seconds, the oximeter will power on andstart to measure. Finger Placement
  3. The display screen showsthe measurement.
  4. Get the finger out, and the device will automatically power

Attentions for measuring:

Do not shake the finger and relax during measurement.
Do not put wet finger directly into sensor.
Avoid placing the device on the same limb which is wrapped with a cuff for blood pressure measurement or during venous infusion.
Do not let anything block the emitting light from device,i.e. do not use finger nail polish/paints.
Existence of high intensive light sources, such as fluorescence light, ruby lamb, infrared heater or strong sunshine, etc. may cause inaccuracy of measurement result. Please put an opaque cover on the sensor or change the measuring site if necessary.
Vigorous exercise and electrosurgical device interference may affect the measuring accuracy.
Nail polish may affect the measuring accuracy, and too long fingernail may cause failure of measurement or inaccurate result.
If the first reading Appears with poor waveform (irregular or not smooth), then the reading is unlikely true, the more stable value is expected by waiting for a while, or a restart is needed when necessary.

Screen

Indications and Icons

Icon : indicates the wireless connection is set up between the mobile device and oximeter.

Status of Definition
Flashing The oximeter is disconnected from the mobile devices.
On The connection between the oximeter and mobile devices is established.

During measuring, long pressing Display key can enter the setup menu screen.

  1. Shortly press Display Key to choose the setting item;
  2. Long press Display Key to active the setting item, then shortly press it to modify the setting parameter;
  3. Long press Display Key to confirm the modification and exit from this setting
  4. Move the setting item to “Save exit menu”, and long press Display Key to exit from the setup
  • “Reminder”: The device supports reminders triggered by user-defined oxygen level or PR It is on by default.
  • “Beep” : Pulse beep option. If it is set to on, every pulse beat makes a
  • “Display”: The display screen is always on by default. You can set the display to automatically turn off after 5 minutes, 3 minutes, or 1 minute. Wake the screen by pressing the display
  • “Restore default”: Shortly press Display Key to choose “Restore default” and long press Display Key to reset all parameters to their default values.

Technical Specifications

Classification

The type of protection against electric shock|

Internally powered equipment

The degree of protection against electric shock|

Type BF Applied parts

Electro-magnetic compatibility| Group I, Class B
Environment
 | Operating| Storage
Temperature| 5 – 40°C| -20 – 55°C
Relative humidity (non-condensing)| 30% – 80%

(non-condensing)

| 10% – 93%

(non-condensing)

Atmospheric pressure| 700 – 1060hPa| 700 – 1060hPa
Degree of dust & water resistance| IP22
Physical
Dimension| 64mm38mm28mm
Weight| About 37 g
Display| OLED
Wireless| Bluetooth 4.2 BLE
Power and supply
Input| DC 5V ±10%
Battery| Rechargeable Lithium-polymer
Battery life| ≤15 hours for typical use
Charge time| About 3 hours
Oxygen level
Oxygen level range| Measuring range: 0% – 100%
Oxygen level Accuracy (Arms)| ±2% (70% – 100%); ±3% (50% – 69%);

No definition (0% – 49%)

Pulse Rate range| 30bpm – 250 bpm
Pulse Rate accuracy| ±2 bpm or ±2%, whichever is greater
Oxygen level Low limit setting range| 85% – 99%

Default setting: 90%

Pulse Rate low limit setting range| 30bpm – 60 bpm Default setting: 50bpm
Pulse Rate high limit setting range| 100bpm – 240 bpm Default setting: 120bpm

Low perfusion performance

| The accuracy of oxygen level and PR measurement still meet the precision described above when the modulation amplitude is as low as 0.6%.
Sensor| Dual-wavelength LED sensor with wavelength
Wavelength| Red light: 663 nm; Infrared light: 890 nm
Maximal average optical output power|

≤2mW

Ambient light interference| The difference between the oxygen level value measured in the

Maintenance and Cleaning

Maintenance

The expected service life (not a warranty) of this device is 5 years. In order to ensure its long service life, please pay attention to the maintenance.

  • Please change the batteries when the low-voltage indicator
  • Please clean the surface of the device before using, with 75% alcohol wipes, then let it air dry or wipe it dry. Do not allow liquid to enter the device.
  • If the Oximeter has not been used for more than 7 days, please charge the Oximeter before
  • The Oximeter is calibrated in the factory before sale, so there is no need to calibrate it during its life cycle. Any oxygen level simulators should not be used to validate the accuracy of the Oximeter, they can only be used as functional testers to verify its

Caution:

  • High-pressure sterilization cannot be used on the
  • Do not immerse the device in liquid.
  • It is recommended that the device should be kept in a dry environment. Humidity may reduce the life of the device, or even damage

Cleaning and Disinfecting Instruction

  • Surface-clean sensor with a soft cloth damped with a solution such as 75% isopropyl alcohol, if low-level disinfection is required, use a mild bleach
  • Then surface-clean with a cloth damped ONLY with clean water and dry with a clean, soft

Caution:

  • Do not sterilize by irradiation steam, or ethylene
  • Do not use the Oximeter if it is

Symbols

FCC Statement

FCC Warning:
FCC ID: 2ADXK-1659
Any Changes or modifications not expressly Approved by the party responsible for compliance could void the user’s authority to operate the equipment.
This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions:

including interference that may cause undesired operation.
Note: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in  particular installation.   If   this   equipment   does   cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on,

Appendix    EMC

The equipment meets the  requirements of IEC 60601-1-2:2014.
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range Applies.
NOTE 2: These guidelines may not Apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
a: Field strengths from fixed transmitters, such as base stations for radio (cellular / cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, and electromagnetic site survey should be considered. If the measured field strength in the location in which The Pulse Oximeter is used exceeds the Applicable RF compliance level above, The Pulse Oximeter should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating The Pulse Oximeter.
b: Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.

Table 4

Recommended separation distances between portable and

mobile RF communication the equipment


The Pulse Oximeter is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of The Pulse Oximeter can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Pulse Oximeter as recommended below, according to the maximum output power of the communications

equipment.

Rated maximum output power of transmitter W(Watts)| Separation distance according to frequency of

transmitter M(Meters)

150kHz to 80MHz

d=1.2 P

| 80MHz to 800MHz

d=1.2 P

| 80MHz to 2,5GHz

d=2.3 P

0,01| N/A| 0.12| 0.23
0,1| N/A| 0.38| 0.73
1| N/A| 1.2| 2.3
10| N/A| 3.8| 7.3
100| N/A| 12| 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance in metres (m) can be determined using the equation Applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.

NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range Applies.

NOTE 2: These guidelines may not Apply in all situations. Electromagnetic propagation is affected by absorption and

reflection from structures, objects and people.

This manual is written by our company and all rights reserved. Without our company’s prior written consent, no part of this manual may be reproduced or copied in any form or method.

Illustration

All illustrations provided in this manual are for reference only, and the settings or data in the illustrations may not be exactly the same as the actual display you see on the product.

Shenzhen Viatom Technology Co., Ltd.
4E, Building 3, Tingwei Industrial Park No. 6 Liufang Road, Block 67 Xin’an Street, Baoan District, Shenzhen, 518101, Guangdong, China

Read More About This Manual & Download PDF:

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals