RigExpert AA-3000 ZOOM Antenna Analyzers User Manual

June 16, 2024
RigExpert

AA-3000 ZOOM Antenna Analyzers

Product Information

Specifications

  • Product Name: RigExpert AA-3000 ZOOM Antenna and Cable
    Analyzer

  • Measuring Range: 100 kHz to 3000 MHz

  • Functions: SWR (standing wave ratio) measurement, return loss
    measurement, cable loss measurement, graphical measurements with
    built-in ZOOM capability, Time Domain Reflectometer mode for fault
    location, Bluetooth Low Energy module for wireless connection

  • Battery: Requires three AA sized batteries (alkaline, Ni-MH or
    LiPO accumulators)

Introduction

Thank you for purchasing a RigExpert AA-3000 ZOOM Antenna and
Cable Analyzer! This analyzer is designed to measure various
parameters of cable and antenna systems, including SWR, return
loss, and cable loss. It operates in the frequency range of 100 kHz
to 3000 MHz and features a built-in ZOOM capability for graphical
measurements. Additionally, it has a Time Domain Reflectometer mode
for locating faults in the feedline system. The analyzer also
includes a Bluetooth Low Energy module for wireless connection with
laptops, tablets, or smartphones.

Operating the AA-3000 ZOOM

First Time Use

To use the analyzer for the first time:

  1. Insert three AA sized batteries (alkaline, Ni-MH or LiPO
    accumulators) into the battery compartment, observing the correct
    polarity.

  2. Press the Power key located at the bottom-right corner of the
    keypad to turn on the analyzer.

  3. The screen will display an initial message showing the firmware
    version and instrument serial number.

  4. After the initial message, the Main menu will appear on the
    screen.

  5. The analyzer will automatically turn off if not in use for an
    extended period of time.

Main Menu

The Main menu is the starting point for accessing different
tasks. To navigate the Main menu:

  1. Use the Cursor up and Cursor down keys to scroll through the
    menu options.

  2. Press OK to select an item.

At the top-left corner of the screen, a battery indicator is
displayed. When the analyzer is connected to a computer, the
battery indicator is replaced with a USB icon.

Multifunctional Keys

The analyzer’s keypad has keys with multiple functions:

  • Numbers (1) are used to enter frequency and other numerical
    parameters.

  • Main functions (2) provide quick access to common tasks.

  • Alternative functions (3) are executed when the F (Functional)
    key is held. Alternative functions are marked with yellow for
    easier identification.

To view a list of active hot keys, press the Help key (1).

Connecting to Your Antenna

To connect your antenna to the analyzer:

  1. Plug the cable into the analyzer’s antenna connector.
  2. Tighten the rotating sleeve of the connector. Ensure that the
    rest of the connector and the cable remain stationary.

FAQ

Q: Where can I find the latest manuals and software

updates?

A: You can find the latest manuals and software updates on the
RigExpert website at http://rigexpert.com.

Q: What is the measuring range of the AA-3000 ZOOM?

A: The AA-3000 ZOOM operates in the frequency range of 100 kHz
to 3000 MHz.

Q: What types of batteries can I use with the analyzer?

A: The analyzer requires three AA sized batteries. You can use
alkaline, Ni-MH, or LiPO accumulators.

AA-3000 ZOOM
RigExpert ®
User’s manual

For latest manuals and software updates, please visit

Головна

Introduction Operating the AA -3000ZOOM
First time use Main menu Multifunctional keys Connecting to your antenna SWR chart Chart ZOOM Data screen Frequency and range entry Return loss chart R X chart Smith chart Memory operation SWR meter SWR2Air Display all parameters MultiSWR mode Multiband mode HOLD functions Frequency generator My Bands
Applications
Antennas Coaxial lines Measurement of other elements
Annexes
Annex 1: Specifications Annex 2: Precautions Annex 3: Tools menu

Table of contents
5 6 6 6 7 7 8 8 9 9 9 10 10 11 11 12 13 13 14 14 15 16 16 17 24 27 27 28 29

Annex 4: Setup menu

34

Annex 5: TDR mode

35

Annex 6: Calibration

39

Annex 7: Dummy loads

41

DECLARATION OF CONFORMITY

42

Introduction

Thank you for purchasing a RigExpert AA-3000 ZOOM Antenna and Cable Analyzer! We did our best to make it powerful yet easy to use.
The analyzer is designed for measuring SWR (standing wave ratio), return loss, cable loss, as well as other parameters of cable and antenna systems in the range of 100 kHz to 3000 MHz. A built-in ZOOM capability makes graphical measurements especially An integrated Time Domain Reflectometer mode can be used to locate a fault within the feedline system. The analyzer is equipped with a Bluetooth Low Energy module for a wireless connection with your laptop, tablet or smartphone.
The following tasks are easily accomplished by using this analyzer: · Rapid check-out of an antenna · Tuning an antenna to resonance · Comparing characteristics of an antenna before and after specific event (rain, hurricane, etc.) · Making coaxial stubs or measuring their parameters · Cable testing and fault location, measuring cable loss and characteristic impedance · Measuring capacitance or inductance of reactive loads
RigExpert AA-3000 ZOOM

1. Antenna connector 2. Color display 3. Keypad 4. USB connector
5

Operating the AA-3000 ZOOM

First time use
Please insert three AA sized batteries (alkaline, Ni-MH or LiPO accumulators) into the battery compartment of the analyzer, watching the polarity.

Press the

(Power) key located at the bottom-right corner of the keypad to turn on

the analyzer. After displaying the initial message (showing a firmware version and a serial

number of the instrument), a Main menu appears on the screen.

The analyzer will be turned off automatically if not in use for too long.

Main menu

The Main menu acts as a starting point from where different tasks may be launched.

Use (Cursor up) and

(Cursor

down) keys to scroll through the menu,

then press

(OK) to select an item.

For your convenience, a battery indicator

is shown at the top-left corner of the

screen. This indicator is replaced with a

USB icon when the analyzer is connected

to your computer.

You may use hot keys for the quick access to certain tasks. For instance, press the 4 (SWR chart) button to open the SWR chart screen immediately.

6

User’s manual

Multifunctional keys
Most keys on the analyzer’s keypad perform several functions.
For instance, numbers (1) are used to enter frequency and other numerical parameters. Main functions (2) provide quick access to most common tasks. Alternative functions (3) are executed if the user holds the F (Functional) key. For the convenience, alternative functions are marked with yellow.
You may press the 1 (Help) key to open a help screen listing all active hot keys.
Connecting to your antenna
Plug the cable to your analyzer’s antenna connector, and then tighten the rotating sleeve. The rest of the connector, as well as the cable, should remain stationary.
If you twist other parts of the connector when tightening or loosening, damage may easily occur. Twisting is not allowed by design of the N-connector.

RigExpert AA-3000 ZOOM

7

SWR chart

Once your antenna is connected to the analyzer, it is time to measure its characteristics. Press the 4 (SWR chart) key to open the SWR chart screen, then press
(OK) to start a new measurement.

A few moments later, the result will be displayed on the analyzer’s screen.

Press the F +

key combination to

run a continuous sweep.

A small triangle at the bottom of the chart corresponds to a point at which the SWR reaches its minimum.

Chart ZOOM

Use the arrow keys to increase or decrease the center frequency or the scanning range.

Watch the chart zooming in or out, or changing its position. Use the F (Functional key)

and

(Cursor up) or (Cursor down) key combination to zoom the vertical scale

Do not forget to press the

(OK) key for the new measurement to start.

Press F (Functional key) and 0 to quickly choose a radio amateur band.

8

User’s manual

Data screen
The data screen is available in all chart modes. Press the 0 (Data) key to display various parameters of a load at cursor.

Frequency and range entry
To enter the center frequency or the sweep range, press the 0 (Frequency, Range) key.

Use cursor keys to navigate, or the 0 to 9 keys to enter values, press the (OK)

key to apply. Hold F key for band selection, use

navigation keys for choosing

bandplan, amateur regional bandplan and custom bandplans are available (if they were

defined), use keys to apply band settings.

Return loss chart
The return loss (RL) chart, which is very similar to the SWR chart, is activated by pressing the F (Functional key) and 4 (RL chart) key combination in the Main menu.

RigExpert AA-3000 ZOOM

9

R,X chart
Press the 5 (R,X chart) key in the Main menu to access the R,X chart mode.
Positive values of reactance (X) correspond to inductive load, while negative values correspond to capacitive load.
The chart will display R and X for series or parallel models of a load. Press F (Functional key) and 1 to switch between these models.
The marker at the bottom of the screen shows a resonant frequency closest to the center of the scan.
Smith chart
The 2 (Smith chart) key opens a screen where the reflection coefficient is plotted on the Smith chart.
For a list of hot keys, press the 1 (Help) key, as usual.

A small marker is used to indicate the center frequency.

10

User’s manual

Memory operation
Device has 250 unified slots for storing your records. Record operations are linked to relevant data screens – SWR, RX, RL, Smith charts mark their records with Z’ slot identificator, TDR chart marks withT’ identificator.
Use 6 , 9 , or F and 9 respectively for Load, Save or Edit Name operations in relevant data screens.

SWR meter, SWR2Air.

This mode allows to measure SWR at

one frequency. Mode provides analog

meter or historical trend view and has

SWR2Air possibility. Change frequency

with

, 3 buttons. Press

y to start or stop the measurement,

F + starts measurement with SWR2Air

function.

SWR2Air allows to manifest measured values by speaker or radio(air), using the same

measured antenna. Measuring is coded in Morse code, values in range 1.0-4.9 come

as two digit word without dot, values 5.0 and above are represented as long dash.

Portable, FM -able radio is typically used to pick up radio channel. Change transmit

frequency by pressing F + 3 , other SWR2Air related parameters you can find

in Setup menu. Use F and

to switch between meter and trend view.

RigExpert AA-3000 ZOOM

11

Display all parameters
To display various parameters of a load on a single screen, press the 8 (All) key.

Do not be confused by negative values of L or C. This can be useful for experienced users.

This screen displays values for series as well as parallel models of impedance of a load.

· In the series model, impedance is expressed as resistance and reactance connected in series:

· In the parallel model, impedance is expressed as resistance and reactance connected in parallel:

12

User’s manual

MultiSWR mode
Press the F (Functional key) and 3 (Multi) key combination to see the SWR at up to five different frequencies. This mode may be useful for tuning multiband antennas.

Use (Up) and (Down) cursor keys to select a frequency to be set or changed, then press the 3 (Frequency) key to enter a new value. Do not forget to press the (OK) key to start the measurement.
Multiband mode
In Tools menu you can find another useful mode for tuning multiband antennas. Mode provides up to 3 overlapped charts of selected bands. In-band values are shown as curve, minimal SWR of band and frequency at this point provided by text. Measuring is done in continous “in loop” manner, providing conviently “alive” picture.

Use

buttons to choose group,

press 3 and select the band for

curve. Use the button to start and

stop measuring.

RigExpert AA-3000 ZOOM

13

HOLD functions
RL and Smith charts. You can keep some measuring result on screen next to fresh measuring result. There are two ways of using HOLD.
Using F + 1 keys allows you to make current curve as reference curve (you can use even RECORD curve). Reference curve will be retained until next arbitrary user action. Using F + 6 enables HOLD last operation. New measurement will be shown with previous curve. Text mark “ref” or “last” indicates current mode of HOLD functionality.
Frequency generator
You can provide on RF connector of analizer square-shaped (or close to it, depends on exact frequency) unmodulated signal. It will be persistent and stable, unlike in any measuring mode. All device’s range is accesible with resolution of 1 Hz. Use button for enabling and dissabling signal at output. “Square wave” icon indicates active output.

14

User’s manual

My Bands
For advanced usage of analizer it is sometimes benefitial to have custom, more flexible bands instead of rigid amateur regional bands. This functionality gives you possibility to manage 10 custom bandplans, which one can contain up to 100 user defined bands. By default, bandplans are named as Plan A-J, with bands named A00-A99 (for Plan A), every band has range as 0, making band in such way unconfigured/dissabled. Bandplans that have 0 configured bands considered unconfigured. Unconfigured bandplans and bands are removed from access list (Frequency entry, Multiband).
Configured bands will appear in access list accordingly to their default number (exactly as they ordered there), regardless of their given name or central frequency. Considering that, we can recommend to place most frequently used bands at start of list, you can also leave “gaps”, having in mind, to insert later some bands with expected place.
Please, explore built-in help for other details and keys functionality.

RigExpert AA-3000 ZOOM

15

Applications

Antennas

Checking the antenna

It is a good idea to check an antenna before connecting it to the receiving or transmit-

ting equipment. The SWR chart mode is good for this purpose. Also you can use Multé

tool located in Tools. Multé will check your rig against all HAM bands defined in your

region and rate all bands where SWR lower than 3.0 in range of 1…5 stars. Select band

for detailed evaluation and press

to run SWR chart for this band.

The picture on the left shows the result for one band real antenna.

The next screen shot shows result for 50 Ohm calibrated load, making in such way ideal all-band antenna (in mathematical terms only).

16

User’s manual

Adjusting the antenna
When the measurement diagnoses that the antenna is off the desired frequency, the analyzer can help in adjusting it. Physical dimensions of a simple antenna (such as a dipole) can be adjusted knowing the actual resonant frequency and the desired one. Other types of antennas may contain more than one element to adjust (including coils, filters, etc.), so this method will not work. Instead, you may use the SWR mode, the All parameters mode or the Smith chart mode to continuously see the results while adjusting various parameters of the antenna.
For multi-band antennas, use the MultiSWR mode. You can easily see how changing one of the adjustment elements (trimming capacitor, coil, or physical length of an aerial) affects SWR at up to five different frequencies.
Coaxial lines
Open- and short-circuited cables
The pictures on the right show R and X charts for a piece of cable with open- and short-circuited far end. A resonant frequency is a point at which X (reactance) equals to zero:
· In the open-circuited case, resonant frequencies correspond to (left to right) 1/4, 3/4, 5/4, etc. of the wavelength in this cable;
· For the short-circuited cable, these points are located at 1/2, 1, 3/2, etc. of the wavelength.

RigExpert AA-3000 ZOOM

17

Cable length measurement
Resonant frequencies of a cable depend on its length as well as on the velocity factor. A velocity factor is a parameter which characterizes the slowdown of the speed of the wave in the cable compared to vacuum. The speed of wave (or light) in vacuum is known as the electromagnetic constant: c=299,792,458 meters (or 983,571,056 feet) per second.
Each type of cable has different velocity factor: for instance, for RG-58 it is 0.66. Notice that this parameter may vary depending on the manufacturing process and materials the cable is made of.
To measure the physical length of a cable,
1. Locate a resonant frequency by using the R,X chart.
Example:
The 1/4-wave resonant frequency of a piece of open-circuited RG-58 cable is 4370 kHz.

2. Knowing the electromagnetic constant and the velocity factor of the particular type of cable, find the speed of electromagnetic wave in this cable.

299,792,458 0.66 = 197,863,022 meters per second – or –
983,571,056
0.66 = 649,156,897 feet per second

18

User’s manual

3. Calculate the physical length of the cable by dividing the above speed by the resonant frequency (in Hz) and multiplying the result by the number which corresponds to the location of this resonant frequency (1/4, 1/2, 3/4, 1, 5/4, etc.)

197,863,022 /4,370,000 x (1/4) = 11.32 meters – or –
649,156,897/4,370,000 x (1/4) = 37.14 feet

Velocity factor measurement
1. Locate a resonant frequency as described above.

Example:
5 meters (16.4 feet) of open-circuited cable. Resonant frequency is 9400 kHz at the 1/4-wave point.

2. Calculate the speed of electromagnetic wave in this cable. Divide the length by 1/4, 1/2, 3/4, etc. (depending on the location of the resonant frequency), then multiply by the resonant frequency (in Hz).

5 / (1/4) x 9,400,000 = 188,000,000 meters per second – or 16.4 / (1/4) x 9,400,000 =
616,640,000 feet per second

3. Finally, find the velocity factor. Just divide the above speed by the electromagnetic constant.

188,000,000 /299,792,458 = 0.63 – or –
616,640,000 /983,571,056 = 0.63

RigExpert AA-3000 ZOOM

19

Cable fault location
To locate the position of a probable fault in a cable, just use the same method as when measuring its length. Watch the behavior of the reactive component (X) near the zero frequency: · If the value of X is moving from ­ to 0, the cable is open-circuited:
· If the value of X is moving from 0 to +â, the cable is short-circuited:

20

User’s manual

Making 1/4-, 1/2- and other coaxial stubs

Pieces of cable of certain electrical length are often used as components of baluns (balancing units), transmission line transformers or delay lines. To make a stub of the predetermined electrical length,
1. Calculate the physical length. Divide the electromagnetic constant by the required frequency (in Hz). Multiply the result by the velocity factor of the cable, then multiply by the desired ratio (in respect to A).

Example: 1/4-» stub for 28.2 MHz, cable is RG-58 (velocity factor is 0.66)
299,792,458 /28,200,000 066 (1/4) = 1.75 meters
– or 983,571,056/28,200,000 066 (1/4) = 5.75 feet

2. Cut a piece of cable slightly longer than this value. Connect it to the analyzer. The cable must be open-circuited at the far end for 1/4-A, 3/4-A, etc. stubs, and short-circuited for 1/2-A, A, 3/2-A, etc. ones.
3. Switch the analyzer to the All parameters measurement mode. Set the frequency the stub is designed for.
4. Cut little pieces (1/10 to 1/5 of the margin) from the far end of the cable until the X value falls to zero (or changes its sign). Do not forget to restore the open-circuit, if needed.

A piece of 1.85 m (6.07 ft) was cut. The margin is 10 cm (0.33 ft). The cable is open-circuited at the far end.
28,200 kHz was set.
11 cm (0.36 ft) were cut off.

RigExpert AA-3000 ZOOM

21

Measuring the characteristic impedance

The characteristic impedance is one of the main parameters of any coaxial cable. Usually, its value is printed on the cable by the manufacturer. However, in certain cases the exact value of the characteristic impedance is unknown or is in question.

To measure the characteristic impedance of a cable,

1. Connect a non-inductive resistor to the far end of the cable. The exact value of this resistor is not important. However, it is recommended to use 50 to 100 Ohm resistors.

Example 1:50-Ohm cable with 75 Ohm resistor at the far end.
Example 2: Unknown cable with 50 Ohm resistor at the far end.

2. Enter the R,X chart mode and make measurement in a reasonably large frequency range (for instance, 0 to 20 MHz).

Example 1: 50-Ohm cable
22

Example 2: Unknown cable
User’s manual

3. Changing the display range and performing additional scans, find a frequency where R (resistance) reaches its maximum, and another frequency with minimum. At these points, X (reactance) will cross the zero line.
4. Switch to the Data at cursor screen by pressing the 0 (Data) key and find values of R at previously found frequencies.

Example 1: 4.46 MHz – min., 8.92 MHz – max.
Example 2: 3.47 MHz – max., 716 MHz – min.
Example 1: 36.2 Ohm – min., 71.6 Ohm- max.
Example 2: 114.0 Ohm- max, 50.0 Ohm – min.

5. Calculate the square root of the product of these two values.

Example 1: square root of (36.2 71.6) = 50.9 Ohm
Example 2: square root of (114.0
50.0) = 75.5 Ohm

The Tools menu (see page 29) contains several automated tools for coaxial line calculations.

RigExpert AA-3000 ZOOM

23

Measurement of other elements
Although RigExpert AA-3000 ZOOM is designed for use with antennas and antennafeeder paths, it may be successfully used to measure parameters of other RF elements.
Capacitors and inductors
The analyzer can measure capacitance from a few pF to about 0.1 µF as well as inductance from a few nH to about 100 µH. Since measuring of capacitance and inductance is not a main purpose of RigExpert analyzers, the user will have to gain some experience in such measurements.
Be sure to place the capacitor or the inductor as close as possible to the RF connector of the analyzer.
1. Ente the R,X chart mode and select a reasonably large scanning range. Perform a scan.

Example 1: Unknown capacitor
24

Example 2: Unknown inductor
User’s manual

2. By using left and right arrow keys, scroll to the frequency where X is -25…-100 Ohm for capacitors or 25…100 Ohm for inductors. Change the scanning range and perform additional scans, if needed.
3. Switch to the Data at cursor screen by pressing the 0 key and read the value of capacitance or inductance.

Example 1: Unknown capacitor

Example 2: Unknown inductor

Transformers
RigExpert analyzers can also be used for checking RF transformers. Connect a 50 Ohm resistor to the secondary coil (for 1:1 transformers) and use SWR chart, R,X chart or Smith chart modes to check the frequency response of the transformer. Similarly, use resistors with other values for non-1:1 transformers.

RigExpert AA-3000 ZOOM

25

Traps
A trap is usually a resonant L-C network used in multi-band antennas. By using a simple one-turn wire coil, a resonant frequency of a trap may be measured.
Example: A coaxial trap constructed of 5 turns of TV cable (coil diameter is 6 cm) was measured.
A one-turn coil (about 10 cm in diameter) connected to the analyzer was placed, co-axially a few centimeters away from the measured trap. The SWR chart shows a visible dip near 10.3 MHz, which is a resonant frequency of the trap.

26

User’s manual

Frequency range: 0.1 to 3000 MHz Frequency entry: 1 kHz resolution Output impedance: 50 Ohm Directivity: 43db(calibrated)

Annex 1
Specifications

SWR measurement range: · 1 to 100 in numerical modes · 1 to 10 in chart modes
SWR system impedance: -selectable 25, 50, 75, 100, 150, 200, 300, 450, 600
Display modes: · SWR at single or multiple frequencies · SWR, return loss, R, X, Z, L, C
at single frequency · SWR chart · R, X chart · Smith chart · Return loss chart · TDR chart (Time Domain Reflectometery) · Cable tools (stub tuner, length
& velocity factor, cable loss and characteristic impedance measurement, LC meter, Multe) · Factory OSL: Yes · User OSL: 10 profiles available · Records: 250 unified memory slots

Power: · Three 1.5V alkaline batteries, type AA · Three 1.2V Ni-MH batteries, type AA · Three 3.7V LiPo batteries, 14200 · Max. 4 hours of continuous
measurement, max. 2 days in standby mode when fully charged batteries are used · When the analyzer is connected to a PC or a DC adapter with USB socket, it takes power from these sources
Interface: · 800×480 color Blanview® display · 6×3 keys on the water-proof keypad · Multilingual menus and help screens · USB connection to a personal computer · Bluetooth Specification v.4.2, LE
Dimensions: 98 × 230 × 53 mm (3.9 × 9.0 × 2.0 in)
Operating temperature: 0…40 °C (32…104 °F)

RF output: · Connector type: N · Output signal shape: square · Output power: -10 dBm (at 50 Ohm load)

Weight: 442 g (15.6 oz) Warranty: 2 years
Made in Ukraine.

RigExpert AA-3000 ZOOM

27

Annex 2
Precautions

Never connect the analyzer to your antenna in thunderstorms. Lightning strikes as well as static discharge may kill the operator.
Never leave the analyzer connected to your antenna after you finished operating it. Occasional lightning strikes or nearby transmitters may permanently damage it.
Never inject RF signal or DC voltage into the antenna connector of the analyzer. Do not connect it to your antenna if you have active transmitters nearby.
Avoid static discharge while connecting a cable to the analyzer. It is recommended to ground the cable before connecting it.

Do not leave the analyzer in active measurement mode when you are not actually using it. This may cause interference to nearby receivers.
If using a personal computer, first connect the cable to the antenna connector of the analyzer, then plug the analyzer to the computer USB port. This will protect the analyzer from static discharges.

28

User’s manual

Annex 3
Tools menu
For the quick access to the the Tools menu, press the F + 8 key combination.
Stub tuner
The Stub tuner mode is designed to help making or checking 1/4- or 1/2- coaxial stubs.
Connect either open or short circuited cable to the analyzer and press (OK) to start.
The analyzer will immediately show resonant frequencies for both quarterwave and halfwave stubs.
Longer cables have lower resonant frequency.

RigExpert AA-3000 ZOOM

29

Length & VF

Knowing the velocity factor, a physical

length of a cable can be easily calculated.

Press the

(Up) button and edit the

value of the velocity factor, then press

(OK) to start measurement.

To find the velocity factor of an unknown

cable, press the

(Down) key and

enter the physical length, then press

(OK).

The velocity factor depends on a type of your transmission line. For instance, RG-58 cable with polyethylene insulator has VF=0.66.

30

User’s manual

Cable loss

All measure range is divided for 4 subranges with convient frequency step. Measure is done as two stage operation. Change of subrange only possible at 1 stage and require full cycle measurement.
Use 3 key to set required subrange, then follow on-screen instructions to perform two stage measurement.

Once the analyzer finishes the measure-

ment, you will see the Loss versus

frequency chart. Use

(Left) and

(Right) cursor keys to change

frequency and watch the loss value in

decibels at the bottom of the analyzer’s

screen.

RigExpert AA-3000 ZOOM

31

Cable impedance

To measure the characteristic impedance, use a piece of an open circuited cable: a half of a meter (or a foot) or longer should be fine. Press (OK) to start.

Next, the far end of the cable should be

short circuited. Press

(OK) to

continue.

There are several reasons why the resulting chart does not look smooth, so we need to use (Left) and (Right) cursor keys to find the location where the impedance is stable. The result is shown at the bottom left corner of the screen.
Use F + (Up) and F + (Down) key combinations to change the scale, if needed.

32

User’s manual

LC Meter
Use 3 to choose one of six modes (Inductance L, Q, tan and Capacitance C, Q, tan ), press to start measure, evaluate result, tune range to most reliable values and restart measure, if required.

Be aware of many factors that will impact result of measure -stray capacitance and inductance, nonuniformities of element connection, more complex behaviour of component at higher frequency.
Avoid pick values at series and parallel resonances (reactance change sign) and adjacent areas, avoid fast changing slopes. Pick values from more stable, persistent areas of chart. In most cases measures at lower frequncies are more reliable.

Self tests

There are several built-in self tests in the

AA-3000 ZOOM analyzer, which can be

run by the user to make sure the analyzer

is working properly. Press

to start

the test and then follow instructions

given on the display.

RigExpert AA-3000 ZOOM

33

Annex 4
Setup menu

For the quick access to the Setup menu, press the F + 3 key combination.

There are several settings in the Setup menu:
· Language ­ select a language for the analyzer’s menus · Palette ­ choose a color scheme · Battery ­ select a power consumption scheme · Battery type – select a type of batteries · Sound volume ­ select sound volume · Bluetooth ­ controls bluetooth module power · Sys. imp. ­ select system impedance (25, 50, 75, 100, 150, 200, 300, 450, 600 Ohm)
which affects SWR and return loss readings. · Units ­ select metric (meters) or imperial (feet) units · Bands ­ select region for highlighting of radio amateur bands · My Bands – sub menu for tuning custom bands · Cable vel. factor ­ choose a velocity factor of the coaxial cable for the TDR mode · Freq. corr. ­ frequency correction of the analyzer’s oscillator · WPM – setup the Words Per Minute rate (SWR2Air) · SWR2Air – choose the mode of SWR2Air · Reset settings ­ reset the analyzer to factory defaults · Clear saved charts ­ clear all memory slots

34

User’s manual

Annex 5
TDR mode

Theory

Time domain reflectometers (TDR) are electronic instruments used for locating faults in transmission lines.

A short electrical pulse is sent over the line, and then a reflected pulse is observed. By knowing the delay between two pulses, the speed of light and the cable velocity factor, the DTF (distanceto-fault) is calculated. The amplitude and the shape of the reflected pulse give the operator idea about the nature of the fault.

Voltage

delay

Pulse sent

Time
Echo received

Instead of a short pulse, a “step” function may be sent over the cable.

Voltage

delay

RigExpert AA-3000 ZOOM

Time

Step

Echo

35

Unlike many other commercially-available reflectometers, RigExpert AA-3000 ZOOM does not send pulses into the cable. Instead, another technique is used. First, R and X (the real and the imaginary part of the impedance) are measured over the whole frequency range (up to 3000 MHz). Then, the IFFT (Inverse Fast Fourier Transform) is applied to the data. As a result, impulse response and step response are calculated. This method is often called a “Frequency Domain Reflectometry”, but the “TDR” term is used in this document since all calculations are made internally and the user can only see the final result.

reflection coefficient

Impulse Response

frequency

IFFT

distance

reflection coefficient

Step Response

frequency

distance

The vertical axis of the resulting chart displays the reflection coefficient: =-1 for short load, 0 for matched impedance load (ZLoad=Z0), or +1 for open load. By knowing the cable velocity factor, the horizontal axis is shown in the units of length.
Single or multiple discontinuities can be displayed on these charts. While the Impulse Response chart is suitable for measuring distance, the Step Response chart helps in finding the cause of a fault.
See the examples of typical Step Response charts on the next page.

36

User’s manual

reflection coefficient
reflection coefficient
reflection coefficient
reflection coefficient
reflection coefficient

Open circuit
distance
Z Load > Z 0
distance
Capacitive discontinuity
distance
Capacitive termination
distance
Lossy cable distance

reflection coefficient
reflection coefficient
reflection coefficient
reflection coefficient
reflection coefficient

Short circuit
distance
Z Load > Z 0
distance
Inductive discontinuity
distance
Inductive termination
distance
Cable with Z>Z 0
distance

RigExpert AA-3000 ZOOM

37

Practice
Use 3 for quick setup popup, choose smallest range approriate for your cable length (shorter distances have better length resolution), tune velocity factor fast. Change of velocity factor corrects range distances.
The result of measure can be viewed in three modes: impulse response, step response, impedance response. To switch view press F + 3 , use 0 for numerical representation all data at marker. Three pictures there show real measure of short open stub in different view modes. Explore built-in help for other buttons functionality.

38

User’s manual

Annex 5
TDR mode
RigExpert AA-3000 ZOOM is designed for high performance and uses factory openshort-load (OSL) calibration. To make usage more flexible for expirienced users there are 10 user OSL profiles, that can be applied above factory OSL.
The standards used for calibration should be of high quality. This requirement is especially important for high frequencies (100 MHz and upper). Three different calibration standards should be used: an “open”, a “short” and a “load” (usually, a 50-Ohm resistor). A place where these standards are connected during calibration is called a reference plane. If the calibration is done at the far end of a transmission line, parameters of this line will be subtracted from measurement results and the analyzer will display “true” parameters of a load.

Reference plane

“open” “short” “load”

Cable
Reference plane

“open” “short” “load”

RigExpert AA-3000 ZOOM

39

Use F + 2 in main menu to go in

calibration profiles managment screen.

Select profile and press

to make

profile active, combination of F +

allows to edit existing profile name.

Key

starts calibration sequence for

selected profile. Follow instructions on

screen to perform 3 step calibration.

You may connect calibration standards to the far end of a cable, so the cable
will be “nulled”.

To apply calibration, press the F + 2 key combination in any measurement mode. The small “CALn” mark will appear on screen, where n is number of active profile(0..9)

40

User’s manual

Amphenol 202109-10 1 Watt terminator plug. Low SWR.

Annex 7
Dummy loads

50 Ohm dummy loads are not all equal. For calibration (see page 32), please use low-power RF terminators which provide low SWR over the wide frequency range. High-power terminators, with cable or not, are suitable neither for calibration purposes (page 32), nor for analyzer self-tests (page 30).
Bird 8201 500 Watt oil-cooled termination High SWR

RigExpert AA-3000 ZOOM

41

EC DECLARATION OF CONFORMITY In accordance with EN ISO 17050-1: 2004

We, Rig Expert Ukraine Ltd. of 2 Solomenska Ploscha, Kyiv, 03035, Ukraine Declare under our sole responsibility that the product:

Equipment Brand name Model number

Antenna and cable analyser RigExpert AA-30000 ZOOM

to which this declaration relates, is in conformity with the following standards and / or other normative documents:

Reference No.: Title:

Edition / Date:

IEC 61000-4-2 IEC 61000-4-3

Testing and measurement techniques Electrostatic discharge immunity test
Testing and measurement techniques – Radiated, radio-frequency, electromagnetic field immunity test

2009 2006

We hereby declare that the above named product is in conformity to all the applicable essential requirements of directive 2004/108/EC (the EMC Directive). The technical documentation relevant to the above equipment will be held at:
SEDAM Communications Limited Old Mill Cottage, Shillington Rd, Gravenhurst, MK45 4JE, The United Kingdom

/Ashot Andrieiev/ Director

June 23,2020

42

User’s manual

RigExpert AA-3000 ZOOM

43

For private households: Information on Disposal for Users of WEEE This symbol on the product(s) and / or accompanying documents means that used electrical and electronic equipment (WEEE) should not be mixed with general household waste. For proper treatment, recovery and recycling, please take this product(s) to designated collection points where it will be accepted free of charge. Alternatively, in some countries, you may be able to return your products to your local retailer upon purchase of an equivalent new product. Disposing of this product correctly will help save valuable resources and prevent any potential negative effects on human health and the environment, which could otherwise arise from inappropriate waste handling. Please contact your local authority for further details of your nearest designated collection point . Penalties may be applicable for incorrect disposal of this waste, in accordance with your national legislation. For professional users in the European Union If you wish to discard electrical and electronic equipment (EEE), please contact your dealer or supplier for further information. For disposal in countries outside of the European Union This symbol is only valid in the European Union (EU). If you wish to discard this product, please contact your local authorities or dealer and ask for the correct method of disposal.

Головна

Copyright © 2020-2022 Rig Expert Ukraine Ltd. “RigExpert” is a registered trademark of Rig Expert Ukraine Ltd.
Doc. date: 11-May-2023

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals