TEKBOX TBCDNE-M3 Coupling Decoupling Network Emission
Product Information
Specifications
- Compliance: TBCDNE-M3
- Maximum supply voltage: Not specified
- Maximum current: Not specified
- Frequency range: 30 MHz to 300 MHz
- Common mode impedance: Not specified
- Phase response: Not specified
- Differential mode impedance: Not specified
- Voltage Division Factor: Not specified
- AE to EUT decoupling: Not specified
- AE to RF output decoupling: Not specified
- RF output connector: Not specified
- EUT / AE connectors: Not specified
- Housing material: Not specified
- Dimensions: Not specified
- Weight: Not specified
- Included: Not specified
- Optional accessories: Not specified
Product Usage Instructions
Measurement Setups
Setup for emission measurement of an EUT with one cable;frequency range 30 MHz to 300 MHz
This setup involves connecting one cable to the EUT (Equipment Under Test) and using the CDNE for emission measurement within the frequency range of 30 MHz to 300 MHz.
Setup for emission measurement of an EUT with two cables connected to adjacent surfaces of the EUT; frequency range 30 MHz to 300 MHz
This setup requires connecting two cables to adjacent surfaces of the EUT and using the CDNE for emission measurement within the frequency range of 30 MHz to 300 MHz.
Setup for emission measurement of an EUT with two cables connected on the same surface of the EUT; frequency range 30 MHz to 300 MHz
In this setup, two cables are connected on the same surface of the EUT. The CDNE is used for emission measurement within the frequency range of 30 MHz to 300 MHz.
Example of arrangement of a luminaire during the radiated (OATS, SAC or FAR) disturbance measurement
This example shows the recommended arrangement of a luminaire during the radiated disturbance measurement using methods such as Open Area Test Site (OATS), Semi-Anechoic Chamber (SAC), or Fully Anechoic Room (FAR).
Example of arrangement of an internal module during the radiated (OATS, SAC or FAR) disturbance measurement
This example illustrates the suggested arrangement of an internal module during the radiated disturbance measurement using methods such as Open Area Test Site (OATS), Semi-Anechoic Chamber (SAC), or Fully Anechoic Room (FAR).
Example of arrangement of an external module during the radiated (OATS, SAC
or FAR) disturbance measurement
This example demonstrates the recommended arrangement of an external module
during the radiated disturbance measurement using methods such as Open Area
Test Site (OATS), Semi-Anechoic Chamber (SAC), or Fully Anechoic Room (FAR).
FAQ
Q: What is the compliance rating of the CDNE?
A: The CDNE is compliant with TBCDNE-M3 standards.
Q: What is the frequency range of the CDNE?
A: The CDNE operates within the frequency range of 30 MHz to 300 MHz.
Q: Are the specifications for maximum supply voltage and maximum current
available?
A: Unfortunately, the user manual does not specify the maximum supply
voltage and maximum current for the CDNE.
TBCDNE-M3
Coupling Decoupling Network Emission (CDNE)
Introduction
CDNEs are a type of Coupling Decoupling Networks used for an alternative
technique of measuring radiated emission of lighting equipment in accordance
with the CISPR 15 (EN55015) standard.
It avoids the need for emission measurements on Open Area Test Sites. This
method was established in CISPR 15 edition 9 as an acceptable alternative to a
traditional OATS test configuration that required antennas and calibrated test
sites. It is specified for the frequency range from 30 MHz to 300 MHz. The
CDNE method and the associated limits up to 300 MHz can be only applied for
EUTs with clock frequencies below or equal to 30 MHz. In such a case, the
product is deemed to comply with the requirements between 300 MHz and 1000
MHz.
When compared to CDNs, CDNEs are substantially more stringently specified,
with reduced common mode (CM) impedance tolerance and additional parameters
for CM phase tolerance and differential mode impedance equal to 100 Ω. A
minimum 20 dB for longitudinal conversion loss shall prevent symmetrical
voltage influencing measurement results.
CISPR 15 edition 9 also specifies that the mains supply cable of the EUT
should be terminated with a CDNE positioned on the reference-ground plane for
the OATS, SAC or FAR measurement method.
Unlike CDNs, CDNEs are not appropriate for immunity measurements The Tekbox
TBCDNE-M3 is designed for measuring EUTs with PE, Neutral, and Line
connections. Tekbox offers the TBCDNE-M2 model for measuring EUTs that do not
have a PE (Earth) connection.
Specifications
- Compliance: CISPR 15 (IEC / EN 55015) Edition 9, CISPR 16-1-2, CISPR 16-2-1
- Maximum supply voltage: 300V AC, 600V DC
- Maximum current: 16A
- Frequency range: 30 MHz – 300 MHz
- Common mode impedance: 150 Ω +10 / – 20 Ω (EUT port)
- Phase response: 0° ± 25° (EUT port)
- Differential mode impedance: 100 Ω ± 20 Ω (EUT port)
- Voltage Division Factor: 20 dB ± 1.5 dB (EUT to RF output)
- AE to EUT decoupling: > 30 dB
- AE to RF output decoupling: > 30 dB
- RF output connector: N – female
- EUT / AE connectors: 4 mm banana safety jacks, 4 mm slots in base plate for GND connection
- Housing material: powder coated aluminium, stainless steel base plate
- Dimensions: 300 x 150 x 150 mm
- Weight: ca. 2.5 kg
- Included: NIST traceable calibration data
- Optional accessories: Adapter panel with shorting bar and N-connector 50 Ω to 150 Ω adapter panel for VDF measurement/calibration OSL adapters for common mode impedance measurement/calibration Balun and OSL adapters for differential mode impedance measurement/calibration
Informative schematic
CDNE-M3, informative schematic
Typ. Voltage Division Factor
Typ. EUT common mode impedance
Typ. EUT differential mode impedance
Typ. EUT common mode phase response
Typ. Decoupling, AE to EUT
Typ. Decoupling, AE to RF
Measurement Setups
When the CDNE method is used and all clock frequencies of the EUT are below or equal to 30 MHz, then the product is deemed to comply with the requirements between 300 MHz and 1 000 MHz if the emissions comply with the limits in the 30 MHz to 300 MHz frequency range as specified in CISPR 15 (IEC / EN 55015). If the CDNE test fails, then any of the other methods and corresponding limits specified in CISPR 15 can still be applied For the CDNE method, the largest dimensions of the EUT are 3 m x 1 m x 1 m (l x w x h). The CDNE restrictions apply to the EUT only, and not the wiring or the total dimension of the system under test. When using the CDNE method to show compliance to 1 GHz, the test report shall include a statement (from the manufacturer) that the clock frequency is below 30 MHz.
Setup for emission measurement of an EUT with one cable; frequency range 30 MHz to 300 MHz
Setup for emission measurement of an EUT with two cables connected to adjacent surfaces of the EUT; frequency range 30 MHz to 300 MHz
Setup for emission measurement of an EUT with two cables connected on the same surface of the EUT; frequency range 30 MHz to 300 MHz
Example of arrangement of a luminaire during the radiated (OATS, SAC or FAR) disturbance measurement
Example of arrangement of an internal module during the radiated (OATS, SAC or FAR) disturbance measurement
Example of arrangement of an external module during the radiated (OATS, SAC or FAR) disturbance measurement
References
- CISPR 16-1-2 Edition 2.0 2014-03
- CISPR 16-2-1 CISPR 16-2-1 Edition 3.0 2014-02
- CISPR 15 Edition 9 (IEC / EN 55015:2019+A11:2020)
Ordering Information
Part Number | Description |
---|---|
TBCDNE-M3 | M3 CDNE |
TBCDN-M3-AP | Adapter panel with shorting bar for M3 CDNE |
TBCDNE-M3-CAL1 | 50 Ω to 150 Ω adapter panel for VDF measurement/calibration |
TBCDNE-M3-CAL2 | OSL adapters for common mode impedance measurement/calibration |
TBCDNE-MX-CAL3 | Balun and OSL adapters for differential mode impedance |
measurement/calibration
History
Version | Date | Author | Changes |
---|---|---|---|
V1.0 | 3.10.2023 | Mayerhofer | Creation of the document |
V1.1 | 2.11.2023 | Mayerhofer | Update of chapter 1 |