PHANTOM-Ceiling Minor Procedure Light Instruction Manual
- June 14, 2024
- Phantom
Table of Contents
PHANTOM
INSTALLATION MANUAL
INTRODUCTION
Please read this manual carefully before proceeding to correctly install the
product, so as to protect “the Service Personnel” and “the Operator” from any
injury. The manual contains the drawings relating to each installation task.
This appliance is a Class 1 medical device pursuant to AS/NZ standards
3200-2-41 (IEC 60601-241).
CONFORMITY
The manufacturer declares that this Product is in compliance with AS/NZ
standards 3200-2-41 (IEC 60601-2-41).
VALIDITY OF MANUAL
This installation manual is valid for the following models:
- Phantom – wall mount
- Phantom – ceiling mount
CUSTOMER SERVICE
The Planet Lighting after-sales service is at your disposal for any further
details you might require regarding the Product, its packaging, its transport
and its installation and for any requests for technical and electrical
diagrams:
Planet Lighting
12 Tamarind Drive
Bellingen NSW 2454
Australia
Ph: +61 2 6659 5600
F: +61 2 6655 2121
Email: info@planetlighting.com
COPYRIGHT
The reproduction and translation, including partial, of any of this manual is
forbidden without written permission from Planet Lighting.
TRANSLATIONS
The original language of this manual is English. For all other translations,
reference must be made to the original manual language.
1.1 GLOSSARY OF TERMS
TERM | DEFINITION |
---|---|
PRODUCT | The EM (Electro-Medical) EQUIPMENT to which this manual refers is the |
Planet Model ‘Phantom’ examination/Minor procedure light. For ease of
description, in this manual this EM EQUIPMENT will be called “Product”.
OPERATOR| Person handling the equipment (e.g., professional health personnel,
nonexpert person assisting the patient).
RESPONSIBLE ORGANIZATION| Entity accountable for the use and maintenance of an
EM equipment or EM system (e.g., a hospital, an individual doctor or a non-
expert person). Preparation and training are included in use.
SERVICE PERSONNEL| Individuals or entities accountable to the responsible
organisation that installs, assembles, maintains or repairs the equipment. In
certain circumstances, the safety of such persons depends on their knowledge
and training and ability to take appropriate precautions when gaining access
to hazardous parts of The Product assembly. By way of example only, the
following professional figures are deemed as SERVICE PERSONNEL:
• Construction Engineer, Draftsman, Building firm duly registered in the
professional Register, (for the masonry works)
• Electrical Engineer, or other Electro-technical expert qualified to work as
an electrician (for the electrical works)
For the installation phase, as regards assembly operations only, a qualified
person is deemed whosoever has attended a course organised by Planet Lighting
or, alternatively, whosoever has carefully read the manual.
1.2 WARNINGS AND SAFETY NOTICES
WARNING| This manual is an integral part of the Product as indicated by AS/NZ
standards 3200-2-41 (IEC 60601-2-41 and subsequent amendments and supplements.
Read and keep this manual close to the Product.
---|---
WARNING| The Product is not suitable for use in premises where explosion risks
exist. The Product is not suitable for use wherever there are inflammable
mixes of
anaesthetics with air, oxygen or N2 0 (laughing gas).
WARNING| Planet Lighting disclaims all liability for any injury to persons or
damage to things caused by the Product having been installed by persons who
are not
“SERVICE PERSONNEL”.
WARNING| The RESPONSIBLE ORGANIZATION is entirely responsible for Product
installation activities; no costs or responsibilities relating to the
installation and/
or commissioning of the Product may therefore be traced back and/or in any
case attributed to Planet Lighting.
WARNING| The ceiling or wall masonry works for Products to be installed on
ceilings or walls, and the electrical works for supplying power to the Product
shall be carried out in a professional manner by SERVICE PERSONNEL to ensure
these are sturdy and safe.
WARNING| The electrical system in the premises must conform to the applicable
localstandards, and any national regulations. A master switch must be
installed with fuse or thermal magnetic circuit breaker to be able to
interrupt power to the Product. DANGER – Electric shock risk To avoid any
risk of electric shocks, the roduct must only be connected to mains supplies
with earth protection.
GENERAL INFORMATION
This installation manual describes how to install, maintain and operate the
Planet Model ‘Phantom’ Examination/Minor procedure light (referred to
throughout this document as the ‘Product”). Prior to installation carefully
read through all the instructions enclosed in this manual.
The Phantom Light Complies with AS/NZ standards 3200-2-41 (IEC 60601-2-41)
2.1 PRODUCT DEFINITION
The Product is a state of the art Examination/Minor procedure light which uses
cutting-edge Solid State Lighting technology. Designed and manufactured in
Australia by Planet Lighting (established 1911) an R&D driven, exporting
designer and manufacturer with over 10 years experience in the design,
manufacture and supply of Solid State Lighting devices worldwide.
The Planet Phantom provides an improved working environment, significantly
reducing ongoing maintenance, with low operational costs and offers a
“greener” alternative to similar products.
The Planet model Phantom allows the accurate positioning of the lamp head
above any work surface. It is designed to be mounted on a wall or ceiling and
be pulled down from above when required, remaining there until repositioned.
At Planet we take great pride in the design and production of our lights and
we are sure you will derive satisfaction from owning and using our products.
When installed and maintained in accordance with the instructions contained in
this service leaflet, this product will provide many years of troublefree
operation.
Should you have any enquiry, comment or special lighting need we would like to
hear from you. We back our products with pride, service and a five year
warranty.
2.1 OPERATOR QUALIFICATION
The following qualification requirements are based on the definitions given
under the ‘Glossary of Terms ‘ section on page 3. The qualification
requirements for various tasks relating to the Product are as follows :
Installation:
SERVICE PERSONNEL
Use: OPERATOR
Cleaning: OPERATOR
Routine maintenance: SERVICE PERSONNEL
Special maintenance: SERVICE PERSONNEL
Disposal: RESPONSIBLE ORGANISATION and SERVICE PERSONNEL
2.2 ENVIRONMENTAL CONSIDERATIONS
2.2.1 PACKAGING
Cardboard boxes containing Product. Dispose of these in compliance with
national directives applicable for waste disposal.
2.2.2 TRANSPORT
Product transport is done by land, sea or air according to the following
characteristics:
Temperature (°C): -15 / +60 Humidity: 10 / 75 %
Atmospheric pressure (hPa): 500 / 1060
2.2.3 STORAGE
The packaged Product must be stored (warehoused) in dry premises having the
following characteristics:
Temperature (°C): -15 / +60 Humidity: 10 / 75 % Atmospheric pressure (hPa):
500 / 1060
2.2.4 PLACE OF INSTALLATION
The premises where the product is started up must have the following
characteristics: Temperature (°C): +10 / +40 Humidity: 30 / 75 % Atmospheric
pressure (hPa): 700 / 1060
2.2.5 RECYCLING
Design of the Product takes into account the materials used in the manufacture
of components. This is to ensure when it reaches its end of life, the disposal
method with the lowest or least negative impact, on human health and the
environment is selected. Unlike the sources of traditional light, LED’s do not
contain mercury, toxic gases, filaments or fragile parts. Due to the fixture
being over 95% recyclable and there being no Bulbs/Globe changing during the
LED life of 50,000+ hours, your environmental footprint will be significantly
reduced.
When your Product reaches the end of its life, contact Planet Lighting or our
local agents to learn about recycling options. The Phantom light must be
disposed of properly according to local laws and, regulations.
2.3 GRAPHIC SIGNS AND SYMBOLS USED IN THE INSTALLATION MAN- UAL
The following safety measures must be put in place during Product
installation, use and servicing.
To emphasize their importance, a number of safety precautions are repeated
throughout the manual.
Follow the safety precautions before using or repairing the Product. Carefully
abiding by the safety precautions improves the ability to use the Product
safely and correctly and helps prevent incorrect maintenance which could be
hazardous and cause damage. The safety measures are ver 09.08.2022-01
approximate and not exhaustive; the Operator, the Responsible Organization and
the Service Personnel must develop their capacities to upgrade and integrate
them.
Indications such as DANGER, WARNING and CAUTION, preceded by the symbol !
indicate the level of risk to which the SERVICE PERSONNEL, the RESPONSIBLE
ORGANIZATION and the PRODUCT could be exposed.
DANGER: indicates an immediately hazardous situation which could result
in death or serious injuries.
WARNING: indicates a potentially hazardous situation that could result in
death or serious injuries.
CAUTION: indicates a potentially hazardous situation which could result
in moderate or light injuries and Product damage.
This triangular symbol together with the explanation alongside (shown above)
indicates the type of hazard to be dealt with.
2.4 GRAPHIC SYMBOLS USED ON PACKAGING
The following table lists the common symbols used by Planet Lighting on our
packaging.
DESIGNATION | SYMBOL | DESCRIPTION |
---|---|---|
THIS SIDE UP | Package must always be transported, handled and stored with |
the arrows always pointing upwards.
FRAGILE| | The cargo is easily broken or damaged. Cargoes marked with this
symbol should be handled carefully and should never be tipped over or slung.
TW0 PERSON LIFT| | Cargo with this symbol weighs in excess of 20kg. For
personal safety these items should only be lifted mechanically, or by two or
more people.
2.5 GRAPHIC SYMBOLS USED ON THE PRODUCT
Below are the symbols to be found on the Product:
DESIGNATION | SYMBOL | DESCRIPTION |
---|---|---|
POSITIVE | +VE | Connect positive wire here |
NEGATIVE | -VE | Connect negative wire here. |
ACTIVE/LIVE | L | Connect active/live wire here |
NEUTRAL | N | Connect neutral wire here |
EARTH | **** | Connect earth wire here |
DATE OF MANUFAC- TURE | Date of manufacture (Month/Year) | |
MANUFACTURER | Manufacturers address | |
READ OPERATING MANUAL | Compulsory to read the operating manual |
2.6 WARRANTY AND LIABILITIES
Planet Lighting disclaims all liability with regard to unreliable Product
operation in the following cases:
- Installation, authorised modifications and repairs have not been performed by SERVICE PERSONNEL
- The Product has not been used for its intended purpose and in conformity with the operating instructions (see operation manual).
- The premises have not been approved for healthcare activities
- The premises are not built in conformity with the law and applicable regulations
- The electrical system in the premises is not in compliance with appropriate requirements
2.7 STRUCTURAL CHANGES OR VARIATIONS
No arbitrary structural changes or variations to the Product are admitted. Any
modifications must have the prior written authorisation of Planet Lighting. In
case of the Product having been tampered with, the warranty shall be
invalidated and the manufacturer disclaims all liability for any injuries or
damage caused to the OPERATOR, the RESPONSIBLE ORGANIZATION and the SERVICE
PERSONNEL.
PLANNING
3.1 PREPARING THE PREMISES STRUCTURALLY
WARNING – SAFE MASONRY WORKS
The masonry works for preparing the ceiling to install the Product must be
sturdy and safe and completed in a professional manner, according to
applicable building regulations.
By way of example only, the professional persons charged with completing the
masonry works could be: Construction Engineer, Structural engineer, or a
Builder, duly registered in the respective professional register.
DANGER – WRONG WALL PERFORATION
In case of wrong perforation of the Product supporting wall (e.g., the
breakage of a reinforcedconcrete ceiling/wall iron) always inform the building
manager as this could affect the stability of the building.
WARNING – CEILING AND WALL STRUCTURAL CAPACITY
The ceiling must be able to withstand a weight of at least 300 kg/m2 and have
a thickness of at least 250 mm. For the wall version, the wall must be a
supporting wall and be made of solid brick. Installation on walls of
perforated bricks and plasterboard is only allowed with the fitting of another
plate on the opposite side of the wall (sandwich closing).
Following are the structural loadings and approx product weights. The listed
weights and loads are static only, and as such do not include any dynamic
loads or externally applied loads such as operator handling. For similar
detail, you may also refer to Fig 2.0 for Ceiling, and Fig 15.0 for Wall
mount- ed versions:
Structural capacity requirements – Ceiling Mounted Total Mass – structural
ceiling fitting only: 10.6kg Mass – arm only:
6.3kg Torque (at mounting plate): 63Nm
Structural capacity requirements – Wall Mounted
Total Mass – structural fitting only: 10.22kg
Mass – arm only: 6.3kg
Torque (at mounting plate): 69Nm
The Product installation premises must conform to local building standards.
After making sure the premises used for medical purposes are in conformity
with the above requirements, proceed to mechanically anchor the ceiling and
wall plate, assessing the type of building and making all consequent
adaptations.
THE SERVICE PERSONNEL have all technical, civil and legal responsibilities
relating to correctly and suitably performing Product anchoring and
installation operations in an adequate manner.
3.2 CORRECTLY WIRING UP THE PREMISES
DANGER – SAFE WIRING INSTALLATIONS
The premises used for medical purposes must be safely wired up in a
professional manner by SERVICE PERSONNEL to power the Product.
DANGER – ELECTRICAL ENVIRONMENT IN COMPLIANCE WITH THE LAW
Before installing the Product, the SERVICE PERSONNEL must make sure the
following conditions exist:
- The wiring system of the environment (premises) in which installation is made must conform with regulations for premises used for medical purposes and with applicable national laws and/or regulations.
- The electrical system must have a certificate of conformity issued by whosoever installed it.
- The earth system must be certified as required by applicable regulations.
INSTALLATION
The Planet Model Phantom arm is designed to be mounted above the immediate
work area. Standard mounting fixtures for Planet Model Phantom are: Ceiling
Mount & Wall Mount (specialised mountings are available on application). For
mounting, please use the relevant Ceiling or Wall Mount Template document
provided on our website – see PHANTOM MOUNTING TEMPLATES.
NOTE: Before using these templates, ensure the scale and measurements
have printed accurately to avoid any installation errors caused by incorrect
printer settings. In all cases the suggested mounting height (measured at
swivel head) is two metres from floor level, as seen in Fig 1.0.
4.1 CEILING MOUNT
WARNING
The qualified electrician who is responsible for the installation of the
Phantom must include an isolation switch( all -pole disconnection) in the
fixed wiring of the Phantom and a cable tie near the terminal block of the
primary(240V) wiring (Phase and Neutral) to secure the cables.
STRUCTURAL LOADS – CEILING MOUNT:
The structural integrity of the mounting surface is the responsibility of the
installer. Structural loads are as follows:
4.1.1 MOUNT CEILING PLATE – AND LEVEL
Use the Phantom Ceiling Template (refer 4.0 Installation) to mark out the hole
locations for the ceiling plate.
-
Installer must ensure sufficient structural integrity of the mounting surface (Refer to 3.1).
-
Fix the ceiling plate to the structural ceiling using suitable fixings (to be selected by the installer).
-
It is very important that the installer ensures the mounting plate and collar are level. This will ensure that the centre axis of the mounting collar is plumb, and the light is balanced.
NOTE: If ceiling tube is the pre-welded type, please go to section 4.1.5
4.1.2 DETERMINE THE CEILING TUBE LENGTH 4.1.3 CUT CEILING TUBE TO LENGTH
WARNING
Do not cut tube end with countersunk holes -
Cut ceiling tube to length, based on the table in Fig. 4.0
-
Measure, and drill 10mm through-hole for the locking bolt, as shown in Fig. 5.0.
4.1.4 PRE CEILING FIT OUT -
Assemble ceiling tube to ceiling plate, with the locking bolt and nut, as shown in Fig. 6.0
-
Reticulate 240V power to the power supply mounting location in accordance with local standards and regulations. Attach the mounting plate earth, and reticulate power to the mounting location, as shown in Fig 7.0.
4.1.5 RETICULATE POWER TO CEILING LEVEL -
Drill cable access hole in the ceiling tube at the sub-ceiling level, as shown in Fig. 8.0. Take care to remove power cords from the drilling area, before drilling.
-
Complete the reticulation of 24V power to the power supply mounting location as shown in Fig 9.0.
4.1.6 POSITION POWER SUPPLY AND TIGHTEN ONTO THE CEILING TUBE -
Slide power supply into mounting position, as close to the ceiling as possible, as shown in Fig. 10.0.
-
Using the u-bolt, nuts and washers supplied, attach the power supply assembly to the ceiling tube, as shown in Fig. 10.0.
4.1.7 CONNECT POWER -
Connect the 24V power feed, and earth for the light fitting into the terminal, shown in Fig. 11.0
-
Connect the 240V AC input, and earth into the terminal, as shown in Fig. 11.0.
4.1.8 ATTACH CEILING COVER -
Move ceiling cover into place, and connect the 24V DC output for the LED indicator light.
-
Push the ceiling cover up against the ceiling and tighten the grub screw to lock it in place, as shown in Fig.13.0.
4.1.9 ATTACH LIGHT FITTING TO CEILING TUBE -
Lift Phantom arm into place, and engage the 24V plugs as shown in Fig. 14.0.
-
Insert the arm bearing into the ceiling tube (taking care to align the threaded holes on the bearing sheath with the countersunk holes on the ceiling tube), as shown in Fig. 14.0.
-
Insert the countersunk-head set screws as shown in Fig. 14.0.
-
Optional – remove or replace the lamp head – twist the conical end of the arm and pull the head and s-bend section, as shown in Fig.14.0.
MAINTENANCE ONLY -
Unscrew the end cap for access to the spring balance adjustment with a 4mm Allen Key, as shown in Fig. 14.0.
-
Adjust the spring balance: using a 4mm allen key, adjust the screw inside the knuckle. Clockwise/Tightening the screw will cause heavier engagement of the spring – this will improve ‘head droop’. Anticlockwise/loosening the spring balance will fix a ‘rising head’ issue, as shown in Fig. 14.0.
-
Friction adjustment: Clockwise/tightening this screw will apply more friction to the arm mechanisms. The arm will require more force to move, and will not move as far with the momentum of an individual movement. This should be set to the staff preference, as shown in Fig 14.0.
4.2 WALL MOUNT
STRUCTURAL LOADS – WALL MOUNT:
The structural integrity of the mounting surface is the responsibility of the
installer. Structural loads are as follows:
4.2.1 MARK UP AND MOUNT WALL UNIT
-
Use the Phantom Wall Mount Template (refer 4.0 Installation). Measure up 1907mm from the floor, and align the bottom of the wall mount template with this mark. Position and fix the template into position, ensuring that both edges are level to the vertical and horizontal, as shown in Fig. 16.0. Use the template to mark hole positions for cable access and mounting.
-
Installer to select suitable fixings, and fix the wall unit to the structurally sound wall, ensuring that the unit is level on all edges Fig. 17.0.
-
Reticulate the 240v power to the power supply location, and connect to the fitting at the terminal block as shown in Fig. 17.0.
4.2.2 MOUNT FITTING -
Lift Phantom arm into place, and engage the 24V plugs as shown in Fig. 18.0.
-
Insert the arm bearing into the wall unit (taking care to align the threaded holes on the arm side bearing sheath with the corresponding holes on the bearing sheath on the wall unit, as shown in Fig. 18.0.
-
Insert and tighten the two grub screws to complete the fixing of the arm to the wall unit as shown in Fig.18.0.
-
Replace the wall unit cover as shown in Fig. 18.0.
-
Fix in place with the countersunk screws provided, as shown in Fig. 18.0.
4.2.3 MAINTENANCE ONLY -
Unscrew the end cap for access to the spring balance adjustment using a 4mm Allen Key, as shown in Fig. 18.0.
-
Adjust the spring balance: using a 4mm allen key, adjust the screw inside the knuckle. Clockwise/Tightening the screw will cause heavier engagement of the spring – this will improve ‘head droop’. Anticlockwise/loosening the spring balance will fix a ‘rising head’ issue, as shown in Fig 18.0.
-
Friction adjustment: Clockwise/tightening this screw will apply more friction to the arm mechanisms, as shown in Fig. 18.0. The arm will require more force to move, however will not continue to move with movement. This should be set to the staff preference.
OPERATING INSTRUCTIONS
5.1 GENERAL OPERATION
The general operation sequence is as follows:
- Switch light on
- Adjust dimming as required
- Using the handle, position the head of the fitting and direct the light at the target location
- Perform task
- Reposition for other tasks if required
- Turn light off after use
WARNING: Take care that you DO NOT direct the lighting beam into your eyes (or another’s eyes) during operation. Looking directly into this light could cause permanent eye damage.
5.1.1 SWITCH AND DIMMING OPERATION
The Products lighting functions are controlled via the use of the membrane
switch located on the head of the fitting, on the opposite side to the handle.
The controls (shown below) offer a simple on/ off switch and an additional
up/down dimming control. The dimming control allows the user to set the
brightness level of all LED’s in the head together.
MAINTENANCE
6.1 MAINTENANCE TASK SCHEDULE
The following periodic maintenance task are recommended for both the Wall, and
Ceiling mounted Products:
TASK | DESCRIPTION | FREQUENCY |
---|---|---|
sterilise handle | Remove handle, and place in an autoclave (or similar | |
sterilisation machine) for sterilisation. See 6.1.1 for more details. | Once |
per use
clean/disinfect Product| Using a suitable disinfectant, wipe down all exterior
surfaces of the fitting. See 6.1.2 for more details.| 1 per day
Adjust balance and friction| Adjust the spring tension and friction settings
as per instructions in 4.2.3 items 9-11.| 1 per 3-12 months (site specific)
6.1.1 STERILISE HANDLE
To sterilise the handle:
- Remove the handle from the fitting (Fig.19).
- Sterilise or autoclave (as per your site procedure).
- Once handle is sterilised, re-attach the handle to the light fitting (Fig.19).
CAUTION: Heat treated sterilisation processes will result in the handle
being quite hot, and may result in burns. Ensure that these handles have
cooled to room temperature before re-attaching them to the lights.
6.1.2 CLEANING AND DISINFECTING
WARNING: Turn off the light and allow to cool before commencing any
cleaning.
Please use the following procedures and precautions during the cleaning
process:
- Let the light cool down. Only disinfect and clean the light and body (including any cables) when it is turned off and cold.
- Apply cleaning solution to the cleaning cloth. Make sure that excess water/cleaning agent has been removed from cleaning cloth before commencing.
- Use a damp cloth. DO NOT use a wet cloth.
- Use alcohol/detergent having medium alkaline level but no active chlorine.
CAUTION: Do not use abrasive products, petrol, paint thinners, alkaline
detergent, acidic cleaning agents or aldehydes. Make sure the proportions of
the detergents/cleaning agents are accurate based on manufacturers
instructions.
WARNING: The disinfectants/cleaning agents may contain substances harmful
to health: use disinfectants only recommended by your health authorities and
follow the manufacturer’s user instruction carefully.
CAUTION: DO NOT spray cleaning solutions directly onto the Product as it
may affect product life and performance. Spray onto a cloth and wipe the
product.( See Fig. 20)
Customer Service can be contacted on the phone number below Mon – Friday 9 am – 5 pm
Made in Australia Exported to the World
Tamarind Drive Bellingen NSW 2454 Australia
Phone + 61 (0)2 6659 5600
Fax + 61 (0)2 66552121
info@planetlighting.com
www.planetlighting.com