TITANOX s r l M35581EN Dry Heat Sterilizing Unit Instruction Manual
- June 13, 2024
- TITANOX s r l
Table of Contents
- INTENDED USE
- SYMBOLS AND SAFETY WARNINGS
- ENVIROMENTAL CONDITIONS
- INSTALLATION
- SAFETY WARNINGS
- TIME – PROGRAMMED RUNNING
- CONTINUOS – CYCLE RUNNING
- ORDINARY MAINTENANCE
- PERIODIC MONTHLY MAINTENANCE
- PERIODIC ANNUAL MAINTENANCE
- PERMITTED USES AND FORBIDDEN USES
- DETACHABLE COMPONENT PARTS
- SELLING OFF
- WARRANTY CONDITIONS
- USEFUL LIFE
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
TITANOX s r l M35581EN Dry Heat Sterilizing Unit Instruction Manual
INTENDED USE
The device is designed to be used in sanitary environments for the purpose of
sterilizing non-heat sensitive medical and surgical instruments. The process
of disinfection is based exclusively on the thermal destruction of the
microorganisms present upon the instruments at a typical temperature of about
180º C.
Effectively reaching a sterile state for the instruments inserted into the
sterilizers depends upon multiple factors:
- The level of initial contamination of the instruments (total bacterial load);
- The type of contaminating microorganisms;
- The permeability of all the contaminated instrument parts by the hot air produced by the sterilizer.
For more information see the paragraph ATTENTION ahead.
The device must be used only and exclusively by healthcare professionals
(doctors, paramedics, nurses or similar) who are familiar with the principles
of controlling sterilization processes in general and dry heat sterilization
processes in particular.
Index
- The regulatory int. temp. 50210° C (±6°C)v
- Prog. timer 2 hours or manual (±3°)
- Thermometer int. temp. 50200°C (±10°C)
- Connection plug with fuses
- Warning light “timer” or “manual” connected
- Warning light “resistances”
- Air-inlets for air circulation
- Closure with key
- Silicone gasket
- Security thermostat 232°C (±8°C)
SYMBOLS AND SAFETY WARNINGS
ATTENTION: TEMPERATURE LEVEE
READ INSTRUCTIONS
PROTECTION **BINDING CLAMP
**
CAUTION: DISCONNECT VOLTAGE BEFORE OPENING
Models | A3-213-400 | A3-214-535 | A3-215-670 |
---|---|---|---|
Loaded Chamber | 3 kg | 4 kg | 5 kg |
External sizes:
Width mm| 570| 705| 835
Height mm| 400| 475| 555
Depth mm| 345| 450| 565
Internal sizes:
Width mm| 405| 535| 670
Height mm| 210| 345| 415
Depth mm| 255| 320| 420
Weights:| | |
Net weight| 13 kg| 22 kg| 35 kg
Gross weight| 15 kg| 25 kg| 41 kg
Electrical Characteristics:
Nominal tension (Voltage)| 230 V| 230 V| 230 V
Nominal power (Watt)| 450 W| 950 W| 1900 W
Nominal frequency (Hz)| 50/60 Hz| 50/60 Hz| 50/60 Hz
Net’s fuses (mm 5×20)| F5A-250 V| F5A-250 V| F10A-250V
The device is in conformity to electrical safety norms provided for by the normative institutes and supplied with bipolar plug which assures perfect electrical grounding.
NON-COMPLIANCE WITH INSTRUCTIONS DESCRIBED IN THIS BOOKLET SHALL FREE COMPANY TITANOX S.R.L. FROM ANY LIABILITY.
SYMBOLS | MEANING |
---|---|
Warning! |
| See the annexed documentation
| Use heat insulating gloves
| Caution! Hot surface (max. 210 °C)
| It points the Manufacturer’s name
| This symbol, together with the
year, points the production date
| **** Keep away from sunlight
| Keep in a cool, dry place
| Medical Device
| RAEE symbol to handle electrical and electronical devices wastes
Available operation:
TIME-PROGRAMMED: timer programmable up to 2 hours
CONTINUUOUS CYCLE: manual (without timer)
ENVIROMENTAL CONDITIONS
- Ambient temperature from 5 to 40° C.
- Relative humidity max. 80% for temperatures up to 31° C with linear decrease up to 50% at the temperature of 40° C condensing included.
- Atmospheric Pressure from 500 to 1060 hPa.
- Voltage supply variation not higher than 10%.
- Value of transitory over-voltage in conformity to the installation category (which provides for 2500V).
INSTALLATION
The device has been calibrated and tested at factory and as such does not
require any further calibration or adjustments before installation and start-
up.
Unpack the device and follow the next advices:
- Position the device on a levelled flat, hard, and smooth surface, established with non inflammable material.
- Allow space not lower than 10 cm from walls and surrounding furniture.
- Do not install the device nearby water sinks or similar to avoid contacts with water or substances which could cause electrical short-circuit to the system.
- Install the device in a well aerated location, not near windows or external doors which could cause an unnatural circulation of the air inside the device and therefore compromise its correctly running.
- Do not install the device nearby heat sources or near other electrical devices.
- Install the device in such a way that the power cable is never twisted or bent, but it should connect to the socket free and unhindered. Avoid to positioning the power cable nearby heat sources or other devices that shall cause damage on it in a long run.
- In the event the device is placed on a trolley, always check that the lower part of the device is not enclosed or hindered, as to allow continuous and sufficient ventilation.
Once the device is correctly installed and power cable connected, it is ready
for use.
SAFETY WARNINGS
- The device is designed for use in internal locations.
- The device is not designed for use in presence of gasses or explosive vapours.
- Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation.
- No water or other liquids should be poured into device neither on its base.
- Before any cleaning or maintenance actions, the power cable must always first be removed.
- Make sure that the electrical system has electrical grounding and that it’s in conformity to the safety norms in the country of installation.
- Do not remove any label or plate, in case of need, ask for more.
- Request only original spare parts cause in case of substitution with no original ones, there could be an increase of electromagnetic emissions or a decreased of electromagnetic immunity of this equipment. This may vary the performance and safety of the equipment.
- Portable RF communications equipment should be used no closer than 30 cm to any part of the sterilizer, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
- Emissions class and group: Class A Group 1; Insulation class: 1; IP20
- Do not open the door of the device until the thermometer of the internal temperature do not indicate a temperature lower than 30°C.
- Medical devices, before being placed in the sterilizer, should be washed and dried. The residual water steam may create deposits on the probes and resistances by altering the precision.
TIME – PROGRAMMED RUNNING
- Insert the supplied connection socket of the mains cable in the plug of the device (Pos. 4) and the feeder- plug in the wall mains socket, after checking the voltage value.
- Turn the thermoregulator knob (Pos. 1) and set the desired temperature. This operation shall always be carried out when device is cold.
- Place materials to be sterilized inside the device.
- Lock up the door (Pos. 8).
- Keep the air-inlets half-opened (Pos. 7) to allow a better circulation of the air inside the device and to distribute the heat even to the corners and hidden spaces. Close the above air-inlets only at the end of the sterilization cycle in order to seal the device and to keep the material inside in temperature for a long time
- Set, by turning the timer knob (Pos. 2) anti-clockwise, the sterilization time adding necessary time in order to bring the device and the material inside to the desired temperature. The necessary time to reach the temperature of 180°C is about 30 minutes in case of empty chamber sterilizer, and about 60 minutes with Loaded Chamber well arranged on the trays. This operation switch on the warning light (Pos. 5) indicating that the device is working and the warning light “resistance” (Pos. 6) indicating that the resistances are turned on in the circuit. The graduated thermometer (Pos. 3) allows to check the progressive increase of the internal temperature. When the internal temperature has reached the programmed value, the warning light “resistance” (Pos. 6) is turned off for then, after brief time, again turned-on indicating the re-connection of the resistances in the circuit. This operation allows to keep inside the device a constant temperature up to the end of the sterilization cycle. After programmed time has elapsed, the device turns itself off automatically.
CONTINUOS – CYCLE RUNNING
The device has also a continuous cycle running. For use this kind of running proceed up to point 5 of the procedure for time-programmed running described above. Turn, at this moment, the timer knob (Pos. 2) anticlockwise. The indefinite-time manual continuous-cycle is inserted with a click. To turn off the device, turn back the timer know to position 0.
ORDINARY MAINTENANCE
Before initiating any maintenance operation, make sure that:
- the device is not connected to the power source.
- the device is at ambient temperature.
Keep the internal parts and the grid perfectly clean. Even though they may
change colour and become brownish, never use abrasive or inflammable products
for cleaning them.
Keep the external parts perfectly clean that they must be always specular to
prevent corrosion and dust.
PERIODIC MONTHLY MAINTENANCE
After removing the plug from the power socket, check that the fuses are not
oxidised specially when the device is not used for a long time or kept in a
humid ambient.
The power socket should not change colour or oxidise. If that happens, replace
it immediately.
The power cable must be integral, and it should not show cuts, abrasions or
bending.
The resistances and the internal electrical system do not require any
maintenance.
PERIODIC ANNUAL MAINTENANCE
It is recommended to carry out at least annually the verification of the leakage currents and the verification of the earth conductor with an electrical safety tester in accordance with the requirements of EN 60601-1 or EN 62353. Even if it is advisable for the user to provide a method of physical validation of the device upon installation and routine checks (for example using biological indicators during each cycle), it is still recommended to check the status of the devices at least annually, to control systems of the sterilizer, in particular the temperature control system, to verify that the expected performance is maintained.
PERMITTED USES AND FORBIDDEN USES
The device must be used to sterilize metal materials whose point of melting is
higher than 300°C (surgical instruments, plates, and metal screw).
Inside the sterilizer place only metal containers without plastic parts and
without textile material.
It is forbidden to place in the device items whose temperature of melting is
lower than 300°C or it is not known to the operator.
DETACHABLE COMPONENT PARTS
- 2 Internal shelves
- 1 Main cable
SELLING OFF
The sterilizing unit is made of various materials with mechanical, electro-
mechanical and electronic parts.
The selling off has to be made according with the regulations in force in the
utilizing Nation.
ATTENTION
Laboratory tests following a cycle of 120 minutes at 180ºC on Bacillus
subtilis var niger ATCC 9372 spores demonstrated the efficiency of the
sterilizer.
Titanox does not guarantee and cannot ensure the effective achievement of the
sterile state of the instruments placed in the sterilizer, according to the
definition of sterile medical instruments foreseen by the EN 556 Norms.
The user of the sterilizer therefore has the responsibility to conduct all of
the confirmation procedures of the sterilization process and the necessary
verifications in order to ensure the effective completion of every single
sterilization cycle.
For this reason, in order to ensure a correct routine control of the
sterilization cycle, it is recommended to use biological indicators in the
load to be sterilized, as prescribed by ISO 20857. In any case, it is the
user’s responsibility to use a method that allows to unambiguously identify
successfully sterilized and non sterilized devices.
WARRANTY CONDITIONS
- The device is warranted for a period of one year from date of purchase.
- Warranty covers the substitution or repairs free-of-charge of components with manufacturing defects.
- The device will be repaired only at our factory. Charges, risks arising from the transport of the device shall be on purchaser’s account.
- In the event of repairs at purchaser’s home, purchaser shall be charged fixed call costs covering partial reimbursement of travel and professional visit by personnel.
- Warranty coverage excludes internal lighting, damages caused by carelessness of purchaser, incorrect and improper uses and installations not conforming to warnings, indicated in these booklet instructions or in any case results from phenomenon unrelated to the normal working of the device.
- The warranty expires when the device is tampered with or repaired by unauthorised personnel.
- It is excluded the substitution and the extension of the warranty following a breakdown.
- It is excluded any compensation for damages direct or indirect of any nature to persons or objects arising from use or suspension of use of the device.
- The warranty expires immediately if the relative certificate shows alterations, erasing, or it is not issued or convalidated by us. The certificate must accompany the device or handed to maintenance personnel for homerepairs.
The manufacturing company Titanox S.r.l. is responsible for the safety, reliability, and performance of the device if:
- the assembly, the additions, the re-setting, the modification, or repairs are carried out by personnel of Titanox S.r.l.
- the electrical system to which it is connected conforms to safety norms in country of installation.
- the device is used in conformity to instructions of use and maintenance.
This liability expires immediately when the device is tampered with or repaired by unauthorised personal.
For any further requirements of spare parts, repairs or checks, it’s
necessary contact directly the manufacturer: TITANOX S.r.l. – Via Canove 2/A –
Loc. Canove de’ Biazzi – 26038 Torre de’ Picenardi (CR) – Italia – Tel. (0039)
0375 394065 – Fax (0039) 0375 394067 communicating the registration number of
the device to repair.
All serious accidents concerning the medical device supplied by us must be reported to the manufacturer and Competent Authority of the member State where your registered office is located.
USEFUL LIFE
The useful life of the device is established in years 5 from the commissioning. For this period Titanox S.r.l. guarantees the availability of spare parts and safe operation if the environmental and use conditions defined in the instructions for use are respected by the user.
Symbols
TITANOX s.r.l.
FABBRICA ARTICOLI MEDICO SANITARI MEDICAL SANITARY ITEM FACTORY
26038 Torre dè Picenardi (CR) – ITALY – Via Canove, 2/A – Canove dè Biazzi
Tel. (0039) 0375 394065 (r.a.) –
Fax (0039) 0375 394067
Home page: http://www.titanox.com
E-Mail info@titanox.com