FKG R-Motion Glider Instructions

June 12, 2024
R-Motion

R-Motion logo Instructions for use
Endodontic instruments
Description
General information

FKG R-Motion Glider

see www.fkg.ch for additional information

R-Motion® Glider|
---|---
R-Motion® Glider C|
R-Motion® 20|
R-Motion® 25|
R-Motion® 30|
R-Motion® 40|
R-Motion® 50|

ISO diameter (1) identification Depth marks (in millimetres)
Available on 21 / 25 / 31 mm instruments
Example on 25 / 31 mm instrumentsFKG R-Motion Glider -
Example

Parameters for use :

R-Motion® instruments must only be used in reciprocating (counterclockwise) mode.
Indications for use :
R-Motion® endodontic instruments are intended for use in medical or hospital facilities, by qualified health professionals.
R-Motion® Glider and R-Motion® Glider C instruments are designed for catheterisation of root canals in reciprocating mode.
R-Motion® 20/25/30/40/50 instruments are designed for shaping and cleaning root canals in reciprocating mode.

Contraindications :

Nickel-Titanium instruments must not be used on individuals with a known allergic sensitivity to these metals.
Adverse reactions: None known.

Notice and precautions :

  • Take several radiographs from different angles to determine the anatomy of the root canals (length, width and curvature).
  • Irrigate the canal thoroughly and frequently after the R-Motion® instrument is removed.
  • Regularly inspect the R-Motion® instrument during use, and discard if it shows any signs of wear (for example : straightening).
  • If the R-Motion® instrument is not progressing easily, withdraw it from the canal, clean it, check its cutting edges, then irrigate the canal and repeat the process with an ISO 010 hand file.
  • Single use instruments must not be retreated for subsequent use. This could compromise their integrity.
  • Always inspect the instrument(s) before use, and discard (it) them if there are any visible defect(s).
  • When an instrument reaches the end of its life, please dispose of it in accordance with the applicable laws and regulations.

Protocol for use

  1. Create straight-line coronal and radicular access.

  2. Use an ISO 010 hand file to explore the canal.

  3. Determine the working length using a radiograph and/or an apex locator.

  4. Use the mechanised R-Motion® Glider instrument, applying gentle strokes of 2-3  mm to establish a glide path until the working length is reached. After 3  strokes, remove and clean the instrument, then irrigate ; recapitulate with an ISO  010 hand file. Use of the R-Motion® Glider C is recommended for canals which require an instrument with more rigidity, and/or if the R-Motion® Glider is no longer progressing and can not reach the working length.

  5. Select the R-Motion® single-file shaping instrument which is best suited to the anatomy of the canal being treated. In most cases, and depending on your approach to apical preparation, the R-Motion®  25 or R-Motion®  30 will be appropriate. If the radiographs do not allow you to clearly identify the anatomy of the canal, it can be gauged using hand files.

  6. Start the shaping procedure, applying your usual irrigation/disinfection protocol.

  7. Use gentle inward-outward movements of 2-3 mm, applying very light apical pressure and allowing the instrument to progress passively along the canal. After 3  strokes, remove and clean the instrument, then irrigate ; recapitulate with an ISO 010 hand file.

  8. Repeat the previous step until the instrument has reached its working length. If necessary, a brushing motion may be applied to remove coronal interferences and/or ensure uniform shaping of irregular canals. If the instrument is no longer progressing, make one or more passes with the R-Motion® 20 instrument, then resume using the R-Motion® 25 or the R-Motion® 30. Once the working length has been reached, remove the instrument to prevent over-enlargement of the foramen.
    Once all the canals have been shaped, proceed with the cleaning of the canals. For an optimal result, the use of XP-endo® Finisher is recommended.
    Once the cleaning of the canals is completed, proceed to the next phase of treatment :

  9. Obturate with gutta-percha and sealer. TotalFill® BC Points™ and TotalFill® BC Sealer™ are recommended.

Symbols

| Sterilized product (irradiation)
---|---
| Single use (max. 4 canals)
| Reciprocating motion (counterclockwise)
| Warning: refer to the user manual
| Batch number
| Expiry date
| Do not use if package is damaged
| Nickel-Titanium instrument
| Reference number
| Manufacturer

R-Motion logo1639
FKG Dentaire Sàrl
Le Crêt-du-Locle 4, 2322 Le Crêt-du-Locle, Switzerland
Tel. +41 32 924 22 44,
info@fkg.ch,
www.fkg.ch
FKG REF. 99.7AA.10.21A.XX – n°112 – 2021/10