Preparation and operation

WARNING

  • Risk of suffocation if full-face masks without an exhalation system are used!
  • If full-face masks without an exhalation system are used, the CO2 concentration may rise to critical values and put the patient at risk.
  • Use full-face masks with an external exhalation system if there is no integrated exhalation system.
  • Follow the Instructions for Use for the exhalation system
  •  Connect the mask to the breathing tube (see Instructions for Use for the breathing mask).
  • The proper position and arrangement of the mask on the face of the patient is critical for uniform use of the device.

Start therapy

Requirement

  • The device is set up and connected Set up and connect the device
  1. If the display is dark: press any key briefly. The device switches to standby.
  2. Briefly press the On/off key or
  • If the autoSTART function is activated: breathe into the mask.
  • Current therapy pressure appears in the display. Therapy starts.
  • For more information on auto START Menu settings

End therapy/switch off the device

  • Briefly press the Or/off key Or
  • If the autoSTART function is activated: remove the mask.
  • The device displaces the therapy hours for the current day and then switches to standby.
  • To save energy, you can disconnect the plug from the socket during the day.

Set humidifier

  • Requirement
  • Humidifier is connected and filled with water (see Instructions for Use for humidifier).
  • The humidifier symbol can be seen in the display.
  1. Start therapy Start therapy.
    • The humidifier switches on automatically. The humidifier symbol goes green
  2. To increase the humidifier stage: press the key.
  3. To decrease the humidifier stage: press the key.
  4. To switch off the humidifier: press the key until appears in the display.
  • The humidifier stage suitable for you depends on room temperature and humidity. If you have dry airways in the morning, the heating output is set too low.
  • If condensation has formed in the breathing tube in the morning, the heating output is set too high.
  • When the water level in the humidifier is too low, the device switches off the humidifier automatically.
  • If the humidifier symbol is flashing, you need to fill the humidifier with water (see Instructions for Use for a humidifier).

$Performing the mask test

  • Requirement: Therapy is running.
  1. Press menu buttonPrisma-SmartPlus-APAP-Therapy-Set-FIG-8
  2. To start the mask test : Press
    • The remaining time and the mask test pressure are displayed.
  3. If necessary: Press or button to change the mask test pressure.
  4. Check mask for leaks.
    • Proper mask location: Green checkmark.
    • Mediocre mask location: Leakage display turns orange.
    • Poor mask location: Leakage displays flashes.
  5. If necessary: Adjust mask.
  6. Wait until the device has completed the mask test. or