TAT-5000S-RS232-QR IFU Mindray Accurate Temperature with Gentle Forehead
Scan
User Manual
TAT-5000S-RS232-QR IFU Mindray Accurate Temperature with Gentle Forehead
Scan
TAT-5000S-RS232-QR IFU
Mindray
The Temporal Scanner is an infrared thermometer designed for accurate,
completely noninvasive temperature assessment by scanning the temporal artery
(TA).
Temperature is measured by gently stroking the Temporal Scanner across the
forehead, and includes a momentary touch of the probe to the neck area behind
the ear lobe, to account for any cooling of the forehead as a result of
diaphoresis. The patented arterial heat balance technology (AHB™)
automatically measures the temperature of the skin surface over the artery and
the ambient temperature. It samples these readings some 1000 times a second,
ultimately recording the highest temperature measured (peak) during the course
of the measurement. The Temporal Scanner emits nothing – it only senses the
natural thermal radiation emitted from the skin.
It has been clinically proven in premier university hospitals to be more
accurate than ear thermometry, and better tolerated than rectal thermometry,
and is supported by more than 50 peer-reviewed published studies covering all
ages from premature infants to geriatrics in all clinical care areas. It is a
superior method for patient and clinician alike.
A 40-page compendium on Temporal Artery Temperature Assessment is available at
www.exergen.com/medical/PDFs/tempassess.pdf,
and a complete list of peer-reviewed published clinical studies is available
at www.exergen.com/c. Complete multilanguage
information on clinical use, instruction manuals, and training is available at
www.exergen.com/s, which includes links to a
specialized clinical site http://www.exergen.com/tathermometry/index.htm.
The link to www.exergen.com/s appears on the front
label of the instrument as a scannable “QR” symbol for easy linking to the
site.
Important Safety Instructions
READ ALL INSTRUCTIONS BEFORE USING
Intended Use: The Exergen Temporal Scanner is a handheld infrared thermometer
used by medical professionals for the intermittent measurement of human body
temperature of people of all ages, by scanning the forehead skin over the
temporal artery. Intended users are physicians, nurses, and nursing assistants
at all levels who normally provide patient care. The thermometer provides a
peak temperature reading from plural readings during the step of scanning.
Electronic circuitry processes the measured peak temperature to provide a
temperature display based on a model of heat balance relative to a detected
arterial temperature, the electronic circuitry computing an internal
temperature of the body as a function of ambient temperature (Ta) and sensed
surface temperature. Training materials that are supplementary to this
instruction manual are available at
www.exergen.com/s, and recommended for first time
users.
TAT-5000S Series thermometers are used by medical professionals in clinical
environments. Such medical professionals include physicians, nurses, nurses’
aides, patient care technicians, and others who are trained to take the
temperature of patients. Clinical environments include areas where medical
professionals are providing medical services for patients, including
hospitals, outpatient clinics, primary care offices, and other settings where
temperature is taken as part of patient care. Clinical environments include
Emergency Medical Services environments.
Additionally, the TAT-5000S series thermometers are not for use aboard
aircraft or near High Frequency Surgical Equipment or Radio Frequency shielded
rooms, such as MRI (Magnetic Resonance Imaging) areas.
When using the product basic safety precautions should always be followed,
including the following:
- Use this product only for its intended use as described in this manual.
- Do not take temperature over scar tissue, open sores or abrasions.
- The operating environmental temperature range for this product is 60 to 104°F (15.5 to 40°C).
- Always store this thermometer in a clean, dry place where it will not become excessively cold (-4°F/-20°C), or hot (122°F/50°C) or humid (max RH 93% non-condensing, at 50 to 106 kPa).
- The thermometer is not shockproof. Do not drop it or expose it to electrical shocks.
- Do not Autoclave. Please note cleaning and sterilizing procedures in this manual.
- Do not use this thermometer if it is not working properly, if it has been exposed to temperature extremes, damaged, been subject to electrical shocks or immersed in water.
- There are no parts that you can service yourself except for the battery, which you should replace when low by following the instructions in this manual. For service, repair, or adjustments, return your thermometer to Exergen. Warning: no modification of this equipment is allowed.
- Never drop or insert any object into any opening, unless stated in this manual.
- If your thermometer is not used regularly, remove the battery to prevent possible damage due to chemical leakage.
- Follow the battery manufacturer’s recommendations or your hospital policy for the disposal of used batteries.
- Not suitable for use in the presence of flammable anesthetic mixtures.
- Communication cables for the TAT-5000S that are field replaceable are specific to the model and patient monitor. Only compatible cables may be used, to maintain compliance of the TAT-5000S thermometers with requirements for Emissions and Immunity.
- If the device fails to operate as described above, see the FAQ section of this manual. Additionally, ensure that you are not in the presence of electromagnetic disturbances.
- If you have any additional questions regarding use or care of the thermometer, please see www.exergen.com or call customer service at 617-923-9900.
WARNING: Use of this equipment adjacent to or stacked with other
equipment (other than TAT-5000S compatible patient monitors) should be avoided
because it could result in improper operation. If such use is necessary, this
equipment and the other equipment should be observed to verify that they are
operating normally.
WARNING: Use of accessories, transducers and cables other than those
specified or provided by the manufacturer of this equipment could result in
increased electromagnetic emissions or decreased electromagnetic immunity of
this equipment and result in improper operation.
WARNING: Portable RF communications equipment (including peripherals such
as antenna cables and external antennas) should be used no closer than 30 cm
(12 inches) to any part of the TAT-5000S thermometer, including cables
specified by the manufacturer. Otherwise, degradation of the performance of
this equipment could result.
SAVE THESE INSTRUCTIONS.
Introduction to Temporal Artery Thermometry
Temporal artery thermometry (TAT) is a completely new method of temperature
assessment, using infrared technology to detect the heat naturally emitting
from the skin surface. In addition, and of key importance, this method
incorporates a patented arterial heat balance system to automatically account
for the effects of ambient temperature on the skin.
This method of temperature assessment has been shown to improve results and
reduce costs by non-invasively measuring body temperature with a degree of
clinical accuracy unachievable with any other thermometry method.
Before Using, Familiarize Yourself with the Instrument
- To Scan: Depress the red button. The instrument will continually scan for the highest temperature (peak) as long as the button is depressed.
- Clicking: Each fast click indicates a rise to a higher temperature, similar to a radar detector. Slow clicking indicates that the instrument is still scanning, but not finding any higher temperature.
- To Retain or Lock Reading: The reading will remain on the display for 30 seconds after button is released. If measuring room temperature, the temperature will remain on the display for only 5 seconds.
- To Restart: Depress the button to restart. It is not necessary to wait until the display is clear, the thermometer will immediately begin a new scan each time the button is depressed.
Alternate sites when temporal artery or behind ear are unavailable:
- Femoral artery: slowly slide the probe across groin.
- Lateral thoracic artery: slowly scan side-to-side in the area ~midway between the axilla and the nipple.
Let the instrument acclimatize for at least 10 minutes in the area in which it will be used.
2-Step Infant Temperature Measurement
Step 1
Place probe flush on center of forehead and depress button. Keeping button
depressed, slowly slide probe mid-line across forehead to the hair line.
Step 2
Release button remove from head and read.
How to improve the accuracy of your measurements on infants
The preferred site is the temporal artery area. Unless visibly diaphoretic,
one measurement here is typically all that is required| If the temporal artery
is covered, then the area behind the ear, if exposed, can be an alternate
site.
Measure straight across the forehead and not down side of face.
At mid-line, the temporal artery is about 2 mm below the surface, but can go
deeply below the surface on the side of the face.| Brush the hair aside if
covering the area to be measured. Measurement site must be exposed.
3-Step Adult Temperature Measurement
Step 1 Slide across forehead.
Place probe flush on center of forehead and depress button. Keeping button
depressed slowly slide probe mid-line across forehead to the hair line.|
Step 2 Slide behind ear.
Keeping button depressed, lift probe from forehead, touch behind ear halfway
down the mastoid process and slide down to the soft depression behind the
earlobe.| Step 3 Release button and read.
How to improve the accuracy of your measurements on adults
Measure only the up-side on a patient in a lateral position. The down-side
will be insulated preventing the heat from dissipating, resulting in falsely
high readings.| Think of a sweatband.
Measure straight across the forehead and not down the side of the face. At
mid-line, the temporal artery is about 2 mm below the surface, but can go
deeply below the surface on the side of the face.| Measure exposed skin.
rush the hair and bangs aside if covering the area to be measured.
Minimum measuring time: 2 seconds.
Minimum time between successive measurements: 30 seconds
FAQs
How does the temperature from a temporal scanner relate to core temperature?
Temporal artery temperature is considered a core temperature because it has been demonstrated as accurate as the temperature measured by a pulmonary artery and esophageal catheter, and as accurate as a rectal temperature on a stable patient. Rule of thumb: Rectal temperature is about 1°F (0.5°C) higher than an oral temperature and 2°F (1°C) higher than an axillary temperature. It will be easy to remember if you think of core temperature as a rectal temperature, and apply the same protocol you would use for a rectal temperature. If your thermometer is marked Arterial/Oral and has a serial number beginning with “O” (standard model start with “A”), it is programmed to compute the normal average cooling effect at the mouth, and automatically reduces the higher arterial temperature by that amount. This calibration allows the hospital to maintain existing protocols for fever workups based on oral temperature, and results in a reading consistent with the 98.6°F (37°C) mean normal oral temperature, in the range of 96.6 - 99.5°F (35.9 - 37.5°C) you now see.
What should I do if I get an abnormally high or low reading, how do I confirm my reading?
• Repeat the reading with the same Temporal Scanner; a correct reading will be reproducible. • Repeat the reading with another Temporal Scanner. Two Temporal Scanners with the same reading will confirm the reading. • Sequential readings on the same patient in rapid succession will cool the skin; it is best to wait about 30 seconds for the skin to recover from the cold probe.
Possible causes of abnormal readings.
Type of abnormal Temperature | Possible cause | Helpful hint |
---|---|---|
Abnormally low Temperature | Dirty Lens | Clean lens of scanner every two |
weeks.
Releasing the button before finished| Release the button after finished
measuring.
Measuring when an ice pack or wet compress is on the forehead| Remove ice pack
or wet compress, wait 2 minutes, and re-take temperature.
Measuring a completely diaphoretic patient| Complete diaphoresis includes
diaphoresis of area behind the ear and suggests that the temperature is
rapidly dropping. Use an alternative method of temperature measurement in
these cases until the patient is dry and the temporal artery measurement can
be repeated.
Improperly scanning down the side of the face| Scan straight across forehead.
The temporal artery is closest to skin in that area.
Abnormally high temperature| Anything covering the area to be measured
would insulate and prevent heat from dissipating, resulting in false high
readings.| Confirm measurement site has not recently been in contact with heat
insulators such as hats, blankets, and hair.
Scan the area not covered or wait about 30 seconds for the previously covered
area to equilibrate to the environment.
DISPLAY DIAGNOSTICS CHART
The following chart summarizes the conditions that may occur while the
Temporal Scanner is in use, and the associated indications:
Condition | Display | Range |
---|---|---|
High Target | HI | >110 OF (43 0C) |
Low Target | LO | <61 of (16 0C) |
High Ambient | HI A | >104 OF (40 0C) |
Low Ambient | LOA | <60 OF (16 0C) |
Low Battery | bAtt | |
No or Very Low Battery | blank display | |
Processing Error | Err | Restart. Return to Exergen for repair if error message |
persists.
Scanning (Normal Operation)| |
Care and Maintenance
-
Handling: The Temporal Scanner is designed and built to industrial durability standards in order to provide long and trouble-free service. However, it is also a high precision optical instrument, and should be accorded the same degree of care in handling as you would provide other precision optical instruments, such as cameras or otoscopes.
-
Cleaning the case: The Temporal Scanner case can be wiped down using a cloth dampened with 70% isopropyl alcohol. The industrial grade housing and design of the electronic components allow for completely safe cleaning with 70% isopropyl alcohol but should not be immersed in fluid or autoclaved.
-
Cleaning the sensor lens: With normal use, the only maintenance required is to keep the lens on the end of the probe clean. It is made of special mirror-like, silicon infrared transmitting material. However, dirt, greasy films or moisture on the lens will interfere with the passage of infrared heat and affect the accuracy of the instrument. Regularly clean the lens with a cotton swab dipped in alcohol in accordance with the instruction label on the instrument (see below). Use only light force for cleaning, to avoid damaging the lens. Water can be used to remove any residual film left by the alcohol. Do not use bleach or other cleaning solutions on the sensor lens Sterilization: Sterilization is not recommended for cabled versions of the Temporal Scanner.
DO NOT SUBMERSE THE THERMOMETER IN ANY CLEANING SOLUTION. -
Calibration: Factory calibration data is installed via a computer which communicates with the Temporal Scanner’s microprocessor. The instrument automatically self-calibrates each time it is turned on using this data, and will never require recalibration. If readings are not correct, the instrument should be returned for repair.
-
Battery: A standard alkaline 9V battery provides approximately 15,000 readings. ** To replace, insert the end of a bent paper clip into the pinhole on the side of the unit to release the battery compartment door. Disconnect the old battery and replace with a new one in the same location. Replace the cover. Use only high quality alkaline batteries.
Approximate number of readings when scanning for 5 seconds and reading the
temperature display for 3 seconds before turning thermometer off.
Instructions for Fahrenheit or Celsius Conversion
The Temporal Scanner can be used in either °F or °C. To convert from one
scale to the other, the only tools necessary are a paper clip and the tip of a
small screwdriver.
For °F/°C Conversion:**
- Insert the end of a bent paper clip into the pinhole on the side to release and remove the cover. Remove the battery from the compartment.
- Locate the switch, and with the tip of a screwdriver, slide left or right to the opposite position.
- Remove the screwdriver.
- Replace cover.
Cable Replacement (TAT-5000S-RS232-QR only)
- Disconnect the scanner cable from the monitor’s communication port.
- Insert the end of a bent paper clip into the pinhole on the side to release and remove the cover. Remove the battery from the compartment.
- Locate the cable release tab, and with the tip of a screwdriver in the small round depression in the tab, push the tab down.
- Pull the cable out.
- Replace with new cable – push the cable in until it clicks.
- Put the battery back into the compartment and replace the cover.
- Reconnect the scanner cable to the monitor’s communication port.
Repair
If repair is required, contact Exergen at (617) 923-9900 or
rma@exergen.com for a Return Materials Authorization
(RMA) Number.
-
Mark the RMA number on the outside of your package and packing slips.
-
Include a description of the fault if possible.
-
Send the instrument freight/postage prepaid to:
Exergen Corporation
400 Pleasant Street
Watertown, MA 02472
USA -
Include the address the instrument should be returned to. The instrument will be returned freight/postage prepaid.
Part Numbers
Exergen p/n | Description |
---|---|
124237-AF-MR | TAT-5000S-RS232-QR, Arterial, Deg F |
134203 | Disposable Caps, box of 1000 |
129462 | Sheaths , box of 250 |
Guidance and Manufacturer’s Declaration-Electromagnetic Emissions
The infrared forehead thermometer model TAT-5000S series is intended for use
in the electromagnetic environment specified below. The user of the TAT-5000S
series should assure that it is used in such an environment.
Emissions test| Compliance| Electromagnetic environment-
guidance
RF emissions CISPR 11| Group 1| The TAT-5000S series thermometer uses no RF
energy therefore any emissions are unlikely to cause any interference in
nearby electronic equipment
RF emissions CISPR 11| Class B| The TAT-5000S series thermometer is suitable
for use by a healthcare professional in a typical health care environment.
Harmonic emissions| Not applicable
Voltage fluctuations| Not applicable
Guidance and manufacturer’s declaration-electromagnetic immunity
The TAT-5000S series thermometer is intended for use in the electromagnetic
environment specified below. The user of the TAT- 5000S series should assure
that it is used in such an environment.
Immunity test| IEC 60601 test level| Compliance level|
Electromagnetic environment-guidance
Conducted RF
IEC 61000-4-6| 3Vrms
150 kHz to 80 MHz| 3Vrms| Portable and mobile RF communications equipment
should be used no closer to any part of the TAT-5000S series including cables
if applicable, than the recommended separation distance calculated from the
equation applicable to the frequency of the transmitter.
Recommended separation distance
d=1,2P1/2
d=1,2P1/2 80 MHz to 800MHz
d=1,2*P1/2 800MHz to 2,5 GHz
Where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer and d is the recommended separation
distance in meters(m).
Field strength from the fixer RF transmitters, as determined by an
electromagnetic site survey, a. should be less than the compliance level in
each frequency range and b. interference may occur in the vicinity of
equipment with the following symbol:
Radiated RF
IEC 61000- 4-3| 10V/m
80 MHz to 2,7 GHz| 10V/m
Note 1 At 80MHz and 800MHz, the higher range applies.
Note 2 These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects,
and people.
a. Field strengths from fixed transmitter, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and
FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the
measured field strengths in the location in which the TAT-5000S series
thermometer is used exceeds the applicable RF compliance level above, the TAT-
5000S series thermometer should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such
as reorienting or relocating the TAT-5000S.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less
than 3V/m.
c. Portable and mobile RF communications equipment can affect performance.
d. ESD compliance has been verified by testing. UL Report Number R11965326
-TAT5000S-EMC.
Immunity test| IEC 60601 test level| Compliance level|
Electromagnetic environment-guidance
---|---|---|---
Electrostatic discharge (ESD) IEC61000-4-2| 8kV contact 15kV air| 8kV contact
15kV air| Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should be at least 30%.
Electrical fast transient/ burst
IEC 61000-4-4
| 2kV for power supply lines 1kV for input output lines| Not applicable| Mains
power quality should be that of a typical health care environment.
Surge IEC 61000-4-5| 1kV line(s) to line(s) 2kVline(s) to earth| Not
applicable| Mains power quality should be that of a typical health care
environment.
Interruptions and voltage variations on power supply Input lines IEC
61000-4-11| <5% UT (>95% dip in UT) for 0,5 cycle
** 40% UT (60% dip in UT) for 5 cycles
70% UT (30% dip in UT) for 25 cycles
< 5% UT (>95% dip in UT) for5 sec.| Not applicable| Mains power is not
applicable. The TAT-5000S series is powered by battery and battery only.
Power frequency (50/60 Hz) magnetic field IEC 61000-4-8| 30A/m| 30A/m| Power
frequency magnetic fields should be at the level characteristic of a typical
location in a typical health care environment.
Note UT is the a.c. mains voltage prior to the application of the test
level
Recommended separation distances between portable and mobile RF
communication equipment and the TAT-5000S Series**
The TAT-5000S series forehead thermometer is intended for use in an
electromagnetic environment in which radiated RF disturbances are controlled
or the user of the TAT-5000S series thermometer can help prevent
electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the TAT-
5000S series thermometer as recommended below, according to the maximum output
power of the communications equipment.
Rated maximum output power of transmitter (W)| Separation distance
according to frequency of transmitter m
150 KHz to 80 MHz d=1,2 P 1/2| 80 MHz to 800 MHz d=1,2 P 1/2| 800
MHz to 2,5 GHz d=2,3 P 1/2
0,01| 0,12| 0,12| 0,23
0,1| 0,38| 0,38| 0,73
1| 1,2| 1,2| 2,3
10| 3,8| 3,8| 7,3
100| 12| 12| 23
For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in meters (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
Note 1 At 80 MHz and 800 MHz the separation distance for the higher
frequency range applies.
Note 2 These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects
and people.
Specification| TAT-5000S-RS232
---|---
Clinical Accuracy| ± 0.2°F or 0.1°C Per ASTM E1112
Temperature Range| 61 to 110°F (16 to 43°C)
Arterial Heat Balance Range for Body Temperature | 94 to 110°F (34.5 to 43°C)
Operating Environment| 60 to 104°F (16 to 40°C)
Resolution| 0.1° F or C
Response Time| ~0.04 seconds
Time Displayed On Screen| 30 seconds
Size| Instrument : 7.9″ X 1.8″ X 1.6″ Cable : 32″ retracted
Weight| 0.7 lb
EMI and RFI Protection| Stainless steel enclosure on upper part inside of
casing
Storage Conditions| -4 to 122°F (-20 to 50°C)
Display Type and Size| Large bright LED’s
Construction Method| · Industrial duty impact resistant casing
· Chemically resistant casing and lens
· Hermetically sealed sensing system
· Stainless steel probe
Warranty| Thermometer: Lifetime; Cable: 5 years
*Automatically applied when temperature is within normal body temperature range, otherwise reads surface temperature.| * Approximate number of readings when scanning for 5 seconds and reading the temperature display for 3 seconds before turning thermometer off.
16 C rounded up from 15.5 C.| Laboratory accuracy outside physiological
range is +/-0.3 °C (0.5 °F).
****| Symbol for Manufacturer| | Do not throw this device away in the trash,
contact Exergen Corp. for disposal and recycling instructions.
---|---|---|---
| Attention, Consult Accompanying Documents|
IPX0
| Ordinary Equipment
| Consult Instructions for Use|
| “On” (only for part of Equipment)
| Degree of Protection Against Electrical Shock Defibrillation-Proof Type BF
Applied Part, Battery Operated| | MEDICAL ELECTRICAL EQUIPMENT
ANSI/AAMI/ES60601-1: 005/(R)2012
3rd Edition including Amendment 1;
CAN/CSA-C22.2 No. 60601.1: 2014; IEC 60601-1-6; ISO 80601-2-56:
Particular Requirements For Basic Safety and Essential
Performance of Clinical Thermometers For Body Temperature Measurement
The CLINICAL THERMOMETER is an ADJUSTED MODE CLINICAL THERMOMETER. Correction
method is proprietary. Laboratory testing protocol for laboratory accuracy
available upon request.
EMERGO EUROPE
Westervoortsedijk 60 6827 AT Arnhem
The Netherlands
EXERGEN CORPORATION • 400 PLEASANT STREET • WATERTOWN, MA 02472, USA • PH
(617) 923-9900 www.exergen.com
Verification Testing
All Exergen infrared thermometers are designed to permanently maintain their
accuracy and normally recalibration is not required unless the thermometer has
been physically damaged or experiences component failure. In the unlikely
event recalibration might be required, the thermometer must be returned to
Exergen for the procedure. However, calibration can be verified in the lab or
clinical units quite easily using a device known as a portable blackbody. A
portable blackbody is a reference heat generator (Figure 1),
which is a self-contained device providing a stable reference target
temperature in the clinical temperature range. The device is then used to
verify the calibration of any Exergen thermometer in question, or for quality
checks done on a routine basis. The verifier operates with either a 9-volt
power supply plugged directly into any 120 VAC wall receptacle, allowing
extended use in the laboratory, or it can be completely powered by a 9-volt
battery for portable use on the nursing floors. There are two ways to use the
portable blackbody to verify the calibration accuracy of the thermometer in
question, either (1) with a certified master reference infrared thermometer,
or (2) by using two identical thermometers as a reference against the one in
question.
Using the Portable Blackbody
- Turn on the verifier device, using either a 9-volt battery or the power supply. Make sure the red LED is illuminated.
- Allow device ~5 minutes for warm-up and stabilization time.
- Allow certified master or the two reference thermometers and the instrument to be tested to acclimate in the same ambient temperature for at least 10 minutes.
- For all instruments, make sure the lens at the tip of the probe is clean. To clean, use an alcohol prep or a swab dipped in alcohol, followed by a damp wipe with water to remove any residue.
- Alternately insert the reference instrument(s) and the instrument being verified into the aperture opening, comparing the readings.
Portable Blackbody Calibration Verifier
- Power On LED
- ON/OFF Switch
- Battery Compartment
- Power Supply Jack
Using a Certified Master Reference Thermometer in a Portable Blackbody to
Verify Calibration Figure 1
Accuracy Limits: Comparison between the reference instrument(s) and the
instrument being verified should be within ±0.4 ºF (0.2 ºC) for acceptable
limits. If not, repeat the process. In the event they still differ by more
than the acceptable limits, call Exergen for repair or replacement of the
failed instrument.
Verifier Specifications:
Power Source | 9-volt battery, or 9-volt power supply. |
---|---|
Battery Life | Approximately 1 hour continuous use. |
Low Voltage Indicator | Red LED shuts off when battery voltage drops below |
lower threshold -5 volts.
Temperature Range| 97-104 °F (36-40 °C)
Cleaning| Wipe down with alcohol. Do not immerse.
SERIAL PROTOCOL CONFIDENTIAL
TAT devices using this protocol only transmit and do not receive data The serial port settings an 9600 baud. eight bits. no parity. one stop bit. and no handshaking. The data packet is sent after the display temperature for the unit is determined. which is very shortly after the button is released. AU data is transmitted most significant byte first. The bytes of data in a 32 bit floating point format are transmitted MO first, followed by MI. M2, and lastly EXP. For reference, here are a few examples of floating point representations of hers:
In this example, the firmware version number of the device is 7953 (Ox fl I). the device is in Fahrenheit mode and there were no errors the displayed temperature a 72.8 degrees (0x0248), the serial number is fifty-six, and the reference measurement site is arterial (0x41).
Recommended guidelines for implementations
Verifying valid packets
‘Ib ensure the correctness of a received packet, software should enforce the
singular correct values for the gart byte, the protocol version, the eheektum,
and the final saline (bytes I. 4. 31. and 32). Furthermore, there are limits
on the permissible values for the displayed temperature and the temperature
reference site (bytes 5. 7. and 8). It is suggested that these bytes also be
verified to have permissible valum, as indicated in the description of thaw
bytes in the data packet format table. Enforcing particular values for the
firmware version number is no! recommended, so that any future TAT devices
manufactured with updated firmware can interface with the software. so long as
the protocol version has not changed. For those extra-paranoid users, it is
possible that a portion of the reserved bytes (bytes 13-30) may be unused and
stuffed with roma ant values, for a given version of firmware. Those bytes are
subject to change without notice in any future revisions of the firmware. Than
bytes should only he enforced in conjunction with particular values of
firmware version, and after consultation with Exergen. Any implementations
attempting to verify these reamed bytes should be mindful of correct operation
with future firmware versions. and not try to verify the reserved bytes if the
firmware version number differs from the expected known salm(s).
Timeout mechanism
Enforcing a minimum allowable time between received serial bytes in a packet
is encouraged. The TAT device will send serial bytes in quick succession.
therefore any large delay between received bytes can he interpreted as an
error. and the bytes received prior to the delay discarded. One transmission
of a packet of data has started, the delay between the end of one byte’s stop
bits, and the beginning of the start hit of the next byte of data should be
weU below one millisecond, until all 32 bytes of the packet have been sent.
Recovery In the event of detected errors
Software neut. to be able to recover from a break condition and any framing
errors on the serial line. In the event of a detected error while verifying
the received packet. it is suggested that the software drop the first received
byte and attempt to verify the mat 32 bytes received as a valid packet. This
will ensue that any glitches on the serial line (due to startup transients.
electrostatic discharges. plugging or unplugging devices. or whatnot) do not
CHUM’ the software to reject any valid packets sent subsequently to said
glitches.
Pinout
Ground| QR Plug End of Cable| Cable| Hirose Connector End of
Cable
---|---|---|---
Signal| sleeve| Black| Pin 1
RS232 Signal| pin| Red| Pin 4
Disposable Covers:
Disposable (Resposable) covers, meaning they can be used once and discarded, or reused on the same patient, are available for all levels of cross- contamination protection should they be preferred for certain patient populations, and are still very cost effective. These options include disposable caps and full instrument sheaths, the sheaths being mainly used for isolation patients.
-
Using the Disposable (Resposable) Caps:1. Apply cap by pushing onto the probe head with fingers.
2. Remove cap by pushing edge forward with thumb.
3. Caps may be reused on the same patient. -
Using the Full Instrument Sheaths: 1. Insert instrument into sheath bottom end first.
• If instrument is on a cable, insert probe end first and twist sheath at neck with fingers to assure film is smooth over probe lens.
2. Wrap additional film around probe neck.
• Film should be smooth over probe lens.
3. Slide additional film under fingers while using.
Disposable (Resposable) Covers can be discarded in normal trash.The operator is responsible for checking the compatibility of the thermometer, probe cover, and monitoring equipment. Incompatible components can result in degraded performance.