AIRBOX P01554 Variflow User Manual
- October 30, 2023
- AIRBOX
Table of Contents
AIRBOX P01554 Variflow
Product Information
- Flow Range: Adjustable 5 -17 l/min
- Battery Pack: Battery for Variflow/V8/Lowflow – Lead acid 12V 9Ah Recharge
- Charge Time: 6-8 hours
- Duration: 2+ Hours @ 16 l/min, 8+ Hours @ 8 l/min
- Display: OLED, elapsed run time and battery indicator
- Material: HF ABS
- Waterproof: IP65
- Dimensions: H 38cm x W24cm x D12cm
- Mast extends: (Up to 2 meters)
- Weight: 4.6kg
- CE Marked
Product Usage Instructions
Setting up and using the Variflow pump
Before using the VARIFLOW pump, ensure it is fully charged. A battery level
indicator shows the available power when the unit is first switched on.
Charging the battery
When the charger is plugged into the charging socket, a red LED will be
illuminated. The LED will change to green when the battery is fully charged.
It takes approximately 6-8 hours to completely charge the battery.
Removing the waterproof covers
IMPORTANT : Before switching the unit on, ensure the gold airoutlet
valve and sampling head are not covered.
Specifications
(tested with 8.0µm MCE filter)
- Flow Range: Adjustable 5 -17 l/min
- Battery Pack: Battery for Variflow/V8/Lowflow – Lead acid 12V 9Ah Recharge
- Charge Time: 6-8 hours
- Duration: 2+ Hours @ 16 l/min, 8+ Hours @ 8 l/min
- Display: OLED, elapsed run time and battery indicator
- Material: HF ABS
- Waterproof: IP65
- Dimensions: H38cm x W24cm x D12cm
- Mast extends: (Up to 2 meters)
- Weight: 4.6kg
CE Marked
Switching on and setting the flow rate
- Before switching on the VARIFLOW connect a flow meter to the air sampling head.
- Press start/pause to switch on the unit. The display will show, in this order, the serial number, the current firmware version and the battery status.
- If start/pause is pressed the display shows ‘Flow not set’.
- Press either arrow to open and start Calibration Mode and to Set Flow.
- The pump is set at aprox 8 l/min as a default, to fine tune this:-
- Press start/pause to start calibration. NOTE: Run for 45 seconds to allow to stabilise. Press up/down arrows to adjust flow rate. (A single press advances the display by one increment, holding while pressing advances the display by 10 increments). Check the flow rate against the flow meter and adjust acording to flow requirements. (Any increment changes will be saved for the post test flow check, see 2.2 & 2.3).
- To confirm and set the flow rate press stop.
IMPORTANT: If pump is shutdown between setting the flow and starting the test, flow will have to be re-calibrated by repeating the steps above.
Setting test mode
- Press start/pause the display will show ‘Set test mode’.
- Press arrows to select ‘Free Running’ or ‘Timed test’.
If Free Running is selected
- Press start/pause the pump runs and the timer starts counting, (you can pause during the test by pressing start/pause)
- Press stop during test and the display shows ‘Test Stop? Are you sure?’ If you are not sure, press start/pause to continue test, if you are sure:-
- Press stop display shows ‘Test completed – Check Flow’.
- Press stop display shows ‘Connect flow meter’ (now reconnect your flow meter to check the flow rate).
- Press start/pause, display shows ‘Flow Check’ (any increment changes made when setting the flow rate (see 2) will be shown).
- NOTE: Wait 45 seconds for flow rate to stabilise.
- Press stop display shows ‘Flow Check Complete’.
- Press stop to shutdown.
If Timed Test is selected
- Press start/pause the display shows ‘Enter Run Time’
- Press arrows to adjust time:-
- Press start/pause to move between (Hrs) 00 (Mins) 00 (Sec) 00.
- Press start/pause the pump runs and the timer starts. When the time has elapsed the display shows ‘Test Completed – Check Flow’.
- Press stop display shows ‘Test completed – Check Flow’.
- Press stop display shows ‘Connect flow meter’ (now reconnect your flow meter to check the flow rate).
- Press start/pause display shows ‘Flow Check’ (any increment changes made when setting the flow rate (see 2) will be shown.
- NOTE : Wait 45 seconds for flow rate to stabilise.
- Press stop display shows ‘Flow Check Complete’
- Press stop to shutdown.
- Note: If the pump switches itself off during either Free Running or Timed Test (i.e if the battery dies or the sampling head gets blocked) on restart it will either show:-
- ‘Test Incomplete, Restart Test?’, Press start/pause to resume test and see where the previous test got to.
- Test Completed Check Flow’, in which case continue as instructions above.
- ‘Low battery’, in which case either replace battery (within 10 minutes of disconnecting old one) or charge pump directly from the mains for a short time to complete the flow check.
Maintenance
Replacing the battery
The VARIFLOW pump has been specifically designed to make replacing the battery
a quick and easy task. Simply undo the two screws on the front of the battery
housing, pull off the terminal connectors, remove the battery and replace.
- NOTE: This process is made easier if pump is lying on it’s back.
- IMPORTANT: The terminals on the battery are colour coded red and black to match the red and black wires of the unit. When fitting a new battery make sure that these match up correctly.
Replacing the mast
The mast is secured on the unit in a threaded housing, so to replace just
unscrew and remove. We would also recommend removing the mast when cleaning
the unit (3.3).
Cleaning the unit
The casing of the VARIFLOW pump is made from a hard wearing ABS structural
foam, making it incredibly rugged and also very easy to clean. Tested to IP65
standards (‘Protected against low pressure water jets from any direction.’)
the unit can be cleaned under a shower hose.
NOTE: Do NOT Submerge the product fully in water.
Servicing & Warranty
The Variflow pump is covered by our exclusive warranty and in the event that
the unit fails we will undertake the repair free of charge for up to 2 years
from the date of purchase. The warranty covers the costs of materials and
labour (with the shipping costs to be covered by the customer) and is valid
under the following conditions:-
- a) Please complete and return the warranty registration card.
- b) The defect must not have been caused by improper handling or misuse.
- c) Return the unit to us in the decontamination bag provided by Airbox. We will not assume responsibility for transport damage under any circumstances.
- d) A short description of the defect must be included with the returned unit.
A full warranty statement is available on our website or from our customer service team.
Troubleshooting
- Pump will not start, red LED illuminated
- Battery voltage may be too low, check the battery level and recharge the battery. If the pump still doesn’t start, contact the Airbox service department.
- Pump will not start, no display
- Flat battery, recharge or replace the battery.
- Flow rate doesn’t reach required level
- Ensure any yellow waterproof safety caps have been removed.
- Check tube for kinks and that the sampling head is not blocked.
- If flow rate is still insufficient, contact the Airbox service department.
- Pump stops while running, red LED illuminated
- Check tube for kinks and that the sampling head is not blocked.
- Check the battery level and recharge if required. Replace the battery if necessary. If the flow rate is still insufficient, contact the Airbox service department.
- Pump fails to run for specified duration
- Check the battery level and recharge if required. Replace the battery if necessary.
Parts
Product Code | Name | Description |
---|---|---|
P01554 | VARIFLOW | VARIFLOW Air Sampling Pump (with UK/ EU charger) |
P01554/KIT | VARIFLOW | VARIFLOW Air Sampling Pump (with UK/ EU charger, |
sampling head & spare mast)
P01839| Sampling head| AIRBOX SAMPLING HEAD + 2x O-rings +
2x Filter support pads
P029347LF| Charger for Variflow 1Ah| Charger for P029347LF – 12V 1Ah Lead acid
SLA – (Choose lead)
PT27259| Lead acid battery- Variflow| Battery 12V NP sealed- lead acid 8.5Ah,
(min) Rechargeable.
PT29482LF| Variflow mast| Telescopic mast for VARIFLOW -2m
PT28297LFKIT| Cap & Plug Kit (x10)| Sampling head cap & plug kit – (pack of
10)
Warranty Registration
Don’t forget to register your new Airbox product by filling in the
registration form provided and returning it to us at the address below.
Alternatively you can do this online at:
www.airboxsp.com
Airbox Sampling Products , Technology Centre,
Easting Close, Worthing, West Sussex BN14 8HQ
Thank you for choosing this Airbox product. To ensure that you get the best
possible use from it, please read this manual carefully.
Airbox Sampling Products is part of the FermionX Group
Technical support
For all technical support or customer service enquiries
please contact us:
telephone: +44(0)1903 524 600
email: sales@airboxsp.com