TEKBOX TBL50100-1 Line Impedance Stabilisation Network Instruction Manual

June 10, 2024
TEKBOX

TEKBOX-logo

TEKBOX TBL50100-1 Line Impedance Stabilisation Network

Product Information

The product is an AC LISN (Line Impedance Stabilization Network)that is used in electromagnetic interference testing. It is designed to measure conducted emissions from electrical and electronic devices that are powered by AC mains. The LISN provides a standardized impedance to the device under test (DUT) and separates the common-mode and differential-mode noise components. The product has a maximum source voltage rating of 250V.

Product Usage Instructions

Warning
The LISN involves dealing with potentially lethal voltages and high ground leakage currents. Therefore, the LISN shall only be operated by qualified staff. The LISN housing is connected to the negative/ground SOURCE and EUT pin of the terminal block and the ground of the RF connector. Inadvertently connecting the positive voltage to the ground pin can expose you to the risk of electric shock.

Application
The product is used to measure conducted emissions from electrical and electronic devices that are powered by AC mains. The LISN provides a standardized impedance to the device under test (DUT) and separates the common-mode and differential-mode noise components.

Operation
The LISN should only be operated by qualified staff. Connect the LISN to the DUT as per the wiring variant required for the specific device. The LISN provides a standardized impedance to the DUT and separates the common-mode and differential-mode noise components, which can then be measured using appropriate equipment.

Wiring Variants
The product supports various wiring variants depending on the type of device being tested. Refer to Picture 9, Picture 10, Picture 11, and Picture 12 for measurement setups for DC supplied  equipment, single-phase AC supplied equipment, triple-phase AC supplied equipment in Delta configuration, and triple-phase AC supplied equipment in Star configuration, respectively.

Calibration Data
The LISN calibration data is provided in Table 1, which includes voltage division ratio EUT terminals to RF connector for various frequencies.

Optional Accessories
The product comes with optional accessories that include a BNC adapter for TBL50100 or TBL05100 EUT/SOURCE socket and a 50 Ohm BNC termination.

Note: For more detailed information on the product and its usage, refer to the user manual.

Introduction

The TBL50100-1 50µH is a single channel 50 Ω // 50 µH LISN for the frequency range 150 kHz – 30 MHz according to CISPR 16-1-2.
In addition to conducted noise measurements on AC or DC supplied equipment, the LISN can be used to establish a defined impedance for a variety of other standardized tests.
The LISN is inserted into the supply line of the EUT (Device Under Test). Conducted noise, which is present at the supply terminals of the EUT can be measured at the BNC connector using a spectrum analyzer or a measurement receiver. The SOURCE (supply) terminal and the EUT terminal are decoupled by a 50 µH inductor.
DC or single-phase measurements typically require a pair of TBL50100-1. In conjunction with the Tekbox LISN-Mate common mode noise and differential mode noise can be measured separately. EUTs with 3-phase supply require three or four pieces of TBL50100-1, depending on whether the product is connected in delta or star configuration.
The product includes a BNC-male to N-male RG232 cable, mating connectors for SOURCE and EUT and an individual calibration protocol.

TBL50100-1-Line-Impedance-Stabilisation-Network-1

Parameters

  • LISN type: V-AMN, 50 Ω // 50 µH; no 250 µH pre-filter
  • Specified frequency range: 150 kHz – 30 MHz
  • Characterized frequency range: 9 kHz -100 MHz
  • DC Resistance: < 20 mΩ
  • Maximum current: 70 A continuous, 100A for 10 minutes; see heat up characteristics plot
  • Operating voltage range: 0 – 250V DC; 0 – 250V AC (50/60 Hz)
  • High current plug / screw terminals; male: Phoenix Contact 1762741, female: Phoenix Contact 1762592
  • RF connector: BNC female
  • Dimensions: 350 mm x 230 mm x 230 mm;
  • Weight: 5.3 kg

Conformity
The TBL50100-1 is compliant with CISPR 16-1-2.
In line with the CISPR 16-1-1 standard compliant design and setup specification, the TBL50100-1 LISN exhibits high ground leakage currents and non-standard isolation and consequently cannot meet the safety requirements of EN 61010-1.
Furthermore CISPR 16-1-2 specifies high capacitance values for the capacitors from Line / + to Ground. These capacitance values are not commercially available in Y2 rating and are implemented using AC rated capacitors with suitable voltage rating.
In order to prevent the danger of lethal electric shock, the operator is responsible for ensuring protective measures in line with IEC 60364-4-41 and IEC 60364-5-54 and to follow all safety related information of this manual.

Safety
To ensure safe operations, the user must adhere to all safety-related information in this manual. All metal parts of the housing are connected to the earth pins of the Source and EUT sockets. Before connecting any other conductor to the LISN, connect protective Earth to the base panel.
Protective Earth must not be disconnected from the base panel unless all other wiring has been disconnected first. Special care must be taken to avoid connecting AC – line or VDC+ to the EARTH pin of the SOURCE connector by mistake.
When operating the LISN with alternating current, use an isolation transformer.
While the source connector is under voltage, do not connect or disconnect it.
The TBL50100-1 shall be operated by qualified laboratory staff only.

Warning
Spectrum Analyzer / Measurement Receiver protection:
The TBL50100-1 LISN does not contain any protective elements in the RF path. Use an external attenuator and / or limiter, if your EUT may produce harmful transients or high RF noise levels, in order to protect the spectrum analyser / measurement receiver input. If using without transient limiter and testing AC supplied products, it is highly recommended to protect the input of the connected analyzer with TBHPF1-9kHz or TBHPF1-150 kHz highpass filters.
Safety:
Operating an AC LISN involves dealing with potentially lethal voltages and high ground leakage currents. The LISN shall only be operated by qualified staff.

Read this manual carefully and be sure to understand the operation of the LISN. Make sure that the conducted noise test equipment is set up correctly and that the necessary earth connections are reliably bonded to avoid the risk of lethal electric shocks. The ground panel must be securely connected to the protective earth conductor available on site, before making any other electrical connection. This connection must not be separated, until the supply voltage is disconnected from the LISN setup.
When connecting the LISN to AC mains voltage, always use an isolation transformer.
Take precautions, such as validating the signal amplitude at the RF output prior to connecting a spectrum analyzer or using attenuators and/or limiters to prevent damage to your test receiver or spectrum analyzer.
Do not carry out any modifications or manipulations of the TBL50100-1.
Avoid touching the housing, when operating the LISN at maximum current over extended time. The housing temperature may rise up to 60°C. Turn off the EUT after measurements to avoid unnecessary dissipation.
The LISN housing is connected to the negative / ground SOURCE and EUT pin of the terminal block and the ground of the RF connector. Inadvertently connecting the positive voltage to the ground pin will expose you to the risk of electric shock. The maximum source voltage rating with respect to component ratings is 250V.

Principle schematic

TBL50100-1-Line-Impedance-Stabilisation-Network-3

Impedance

TBL50100-1-Line-Impedance-Stabilisation-Network-4

Phase

TBL50100-1-Line-Impedance-Stabilisation-Network-5

Isolation

TBL50100-1-Line-Impedance-Stabilisation-Network-6

Thermal characteristics

TBL50100-1-Line-Impedance-Stabilisation-Network-7

Calibration data according to CISPR 16 -1-2 Annex A8

TBL50100-1-Line-Impedance-Stabilisation-Network-8

Picture 6: Calibration set up according to CISPR 16-1-2 Annex A.8.

TBL50100-1-Line-Impedance-Stabilisation-Network-9

Picture 7: Voltage division ratio EUT terminals to RF connector

Frequency [MHz]| Voltage Division Ratio EUT port to RF port [dB]| Frequency [MHz]| Voltage Division Ratio EUT port to RF port [dB]| Frequency [MHz]| Voltage Division Ratio EUT port to RF port [dB]
---|---|---|---|---|---
0,009| -6,85| 3| -0,04| 24| -0,14
0,01| -5,77| 4| -0,04| 25| -0,11
0,015| -1,63| 5| -0,04| 26| -0,08
0,02| -0,53| 6| -0,05| 27| -0,07
0,025| -0,21| 7| -0,06| 28| -0,06
0,05| -0,04| 8| -0,06| 29| -0,05
0,075| -0,04| 9| -0,06| 30| -0,06
0,1| -0,04| 10| -0,08| 35| -0,21
0,125| -0,04| 11| -0,07| 40| -0,12
0,15| -0,04| 12| -0,07| 45| -0,19
0,175| -0,04| 13| -0,07| 50| -0,24
0,2| -0,04| 14| -0,07| 55| -0,23
0,3| -0,04| 15| -0,06| 60| -0,29
0,4| -0,04| 16| -0,06| 65| -0,30
0,5| -0,04| 17| -0,06| 70| -0,15
0,6| -0,04| 18| -0,07| 75| -0,89
0,7| -0,04| 19| -0,08| 80| -0,34
0,8| -0,04| 20| -0,11| 85| -0,37
0,9| -0,04| 21| -0,14| 90| -0,32
1| -0,04| 22| -0,16| 95| -0,35
2| -0,04| 23| -0,16| 100| -0,28

Table 1 – LISN calibration data, voltage division ratio, typical data

Application

The abbreviation LISN stands for Line Impedance Stabilisation Network.

  • It is a low pass filter typically placed between a power source and the supply terminals of an equipment under test (EUT).
  • It has a feed-through path to supply the EUT with power
  • It provides a well-defined RF-impedance to the EUT
  • It couples electrical emissions generated by the EUT to a 50 Ω RF port, which can be connected to a spectrum analyser or measurement receiver
  • It suppresses electrical noise from the supply side towards the EUT
  • It suppresses electrical noise from EUT side towards the supply

Picture 8: Basic diagram of a conducted emission measurement setup with a LISN
Note that the above basic diagram is simplified. Typically, a standard conformant setup needs two TBL50100-1. One LISN is inserted in the positive supply line and the other LISN is inserted in the negative supply line. Conducted emission measurements have to be carried out on both supply lines. While measuring the emissions on one of the supply lines, the RF output of the other LISN must be terminated with 50 Ohm. Similarly, when testing an AC supplied product, one LISN must be inserted in the phase line and the other LISN in the neutral line. Products with 3 phase AC supply can be tested using three (delta configuration) or four (star configuration) TBL50100-1.

TBL50100-1-Line-Impedance-Stabilisation-Network-10

Picture 8: Basic diagram of a conducted emission measurement setup with a LISN

Operation
The TBL50100-1 LISN does not contain a 250µH pre-filter. It contains a high pass to reduce the residual 50 Hz amplitude at the RF output, but no transient limiter and no attenuator.
To protect the spectrum analyzer / measurement receiver input, it is highly recommended to use an external filter/attenuator/transient limiter such as the TBFL1 or similar. External attenuators may also be necessary, depending on the behaviour of the EUT.
Furthermore, the TBL50100-1 LISN has an 8µF capacitor to ground. In AC applications, the capacitor draws a significant amount of blind current, which would trip the mains ground fault switch. Hence, the TBL0550-1 must be supplied through an isolation transformer.
Separate measurement of differential and common mode noise can be carried out using two TBL50100-1 in combination with the Tekbox LISN Mate TBLM1.

Wiring Variants

TBL50100-1-Line-Impedance-Stabilisation-Network-11

Picture 9: Measurement setup for DC supplied equipment

TBL50100-1-Line-Impedance-Stabilisation-Network-12

Picture 10: Measurement setup for single phase AC supplied equipment

TBL50100-1-Line-Impedance-Stabilisation-Network-13

Picture 11: Measurement setup for triple phase AC supplied equipment in Delta configuration

TBL50100-1-Line-Impedance-Stabilisation-Network-14

Picture 12: Measurement setup for triple phase AC supplied equipment in Star configuration

TBL50100-1-Line-Impedance-Stabilisation-Network-15

Picture 13: Measurement setup example for separate measurement of common mode noise and differential mode noise, using a LISN-Mate, the same principle can be applied to any wiring variant

Ordering Information

Part Number Description
TBL50100-1 50µH LISN, 2pcs. female terminal blocks Phoenix Contact 1762592, 1

pc. 75 cm BNC-male to N-male RG223 cable, calibration protocol

Optional Accessories:

Part Number Description
TBLA-PH2 BNC adapter for TBL50100 or TBL05100 EUT/SOURCE socket
TBT-BNCM-50 50 Ohm BNC termination

History

Version Date Author Changes
V1.0 12.4.2023 Mayerhofer Creation of the document

TekBox Digital Solutions Vietnam Pte. Ltd. www.tekbox.com Factory 4, F4, Lot I-3B-1, Saigon Hi- Tech Park, Tan Phu Ward, District 9, Ho Chi Minh City, Vietnam

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

TEKBOX User Manuals

Related Manuals