greenTEG AG gSKIN Heat Flux Sensors for R and D Applications
Product Information
The CONTENT Sensor is designed to measure heat flux, which describes the transfer of thermal power per surface unit. The sensor has an electrical connection that generates a voltage signal based on the temperature difference between its two sides. The positive side of the sensor generates a positive voltage signal assuming a heat flux from top (hot side) to the bottom (cold side). The sensor is sensitive to temperature changes and must be handled with care to avoid damage.
The sensor comes with a datasheet specifying its electrical resistance range, which includes the resistance of cables. A resistance below 0.5 ohms indicates a short circuit, while a resistance higher than the specified value indicates a physical wearout of the sensor and/or its cables.
Product Usage Instructions
Mounting the Sensor
Before mounting the sensor, perform a functionality test as described in
Section 3. Once the sensor’s functionality has been verified, prepare the
mounting surface by ensuring that it is flat, dry, and free of dust and
grease. Clean the sensor surface with ethanol or isopropanol and avoid using
acids or bases for cleaning the sensor. Mount the sensor using a mounting
substance as described in Section 4.
Calculating Heat Flux
The heat flux is calculated using the formula: Q = U / S [W/m2], where Q is
the heat flux, U is the voltage signal generated by the sensor, and S is the
surface area of the sensor.
Product Functionality Test
Checking the Electrical Resistance of the Sensor
Perform an electrical resistance test using a standard multimeter via a
two-probe resistance measurement. The resistance measurement must be done
without any applied temperature gradient. The resistance for each sensor will
be in the range specified in its respective datasheet. A resistance below 0.5
ohms indicates a short circuit, while a resistance higher than the specified
value indicates physical wear out of the sensor and/or its cables. In both
cases, the sensor is not functional and must be replaced.
Checking Sensor Behavior with a Temperature Difference
Connect the sensor to a voltmeter with a resolution preferably in the 0.1mV
range. Place the sensor on a metallic surface at room temperature. When
touching the sensor with a warm finger on the upper surface, you should get a
signal in the mV range. A sensor signal below 0.1 mV indicates a short
circuit. If the signal randomly fluctuates between a positive and negative
signal, or the voltages are in the +/- 1 V range, you may have an open
circuit. Check the connection of your electrical probes. If the signal shows
any of these features, the sensor is not functional and must be replaced.
Contact the vendor in this case.
Mounting Substance
Use a mounting substance to attach the sensor to a flat, dry, and dust-free
surface. Follow the detailed description of the sensor mounting provided in
Section 4.
Preface
gSKIN® Heat Flux Sensors are sensors of high quality. To take advantage of
their outstanding performance, specific precautions must be taken during
storage, assembly, and packaging. Therefore, please read the following
instructions carefully. Exposure to volatile organic compounds requires
special care. High concentration and long-term exposure to specific chemicals
must be avoided. Critical conditions are known to occur in a manufacturing
environment and/or during storage. Therefore, the handling and the choice of
housing and packaging materials are crucial. Ambient conditions, however, do
not affect the performance of the gSKIN® sensors. Upon receiving gSKIN® Heat
Flux Sensors check the packing list, verify that all the equipment is present
and inspect for any signs of damage.
Applicability
This document is applicable to all gSKIN® Heat Flux Sensors for R&D
applications supplied by greenTEG AG.
SHORT USER GUIDE
About the gSKIN® Heat Flux Sensors
The gSKIN® Heat Flux Sensors measure thermal power passing through the sensor
surface. The high sensitivity of the modules coupled with its thin design and
low thermal resistance ensure precise measurements with minimal influence on
the thermal flow.
Picture showing the layers (Package, Contacts, Thermopiles, Substrate, Interconnects) of gSKIN® Heat Flux Sensor (thermopiles are highly integrated in the sensor substrate, which leads to high sensitivity sensor modules).
Measurement preparation
- Procure a mounting substance (i.e. tape, paste, or glue) to mount the sensor in your setup.
- Procure a read-out device (i.e. gSKIN® DLOG Data Logger, voltmeter, or 3rd party read-out device)
Mounting the sensor
Before mounting the sensor, the sensor must be tested as described in Section
3. Once the sensor’s functionality has been verified, the mounting surface has
to be prepared.
Ensure that the mounting surface is flat, dry, and free of dust and grease.
Clean the sensor surface with ethanol or isopropanol. Do not use acids or
bases for cleaning the sensor. Mount the sensor using a mounting substance. A
detailed description of the sensor mounting is given in Section 4.
Figure 2: Schematic diagram of mounting and functionality of a gSKIN® Heat
Flux Sensor.
How to calculate heat flux
The heat flux φ describes the transfer of thermal power per surface unit and
is calculated using the following formula:
- φ = U / S [W/m2]
where
- U is the sensor output voltage, in μV
- S is the temperature-corrected sensitivity of the sensor, in μV/(W/m2)
INTRODUCTION
gSKIN® HEAT FLUX SENSOR INTRODUCTION
The positive side of the gSKIN® Sensor
Figure 3: This placement shows the positive side of the sensor (i.e. assuming a heat flux from the top (hot side) to the bottom (cold side), this sensor will generate a positive voltage signal). Alternatively, the correct sensor orientation can be determined by placing the flex print in such a way, that the serial number is not visible.
The positive and negative side of the gSKIN® Heat Flux Sensor
Heat energy always flows from the hot side to the cold side of a system. The
gSKIN® Heat Flux Sensor should be mounted with the positive side of the sensor
in the direction of the expected positive heat flux. However, the gSKIN® Heat
Flux Sensors can be used bi-directionally: If the direction of the heat flux
is reversed, the sign of the sensor voltage output changes (i.e. from positive
to negative or vice versa).
Note
It is very important, that applying any force to the flex print (i.e. tearing)
needs to be avoided. This destroys the electrical connection to the sensor and
therefore results in breaking the sensor.
FUNCTIONALITY TEST
All gSKIN® Heat Flux Sensors adhere to high manufacturing standards. Before shipping, the performance of each gSKIN® Heat Flux Sensor is individually checked. However, external factors (e.g. transportation, prior use), may affect the functionality of the sensor module. Before permanent installation, the sensor functionality must be tested. This section describes the necessary steps to perform the functionality test.
Checking the electrical resistance of the sensor
Electrical resistance testing is done using a standard multimeter via a two-
probe resistance measurement. The resistance measurement must be done without
any applied temperature gradient (e.g. with the sensor hanging in the air
holding it at the cables). The resistance for each sensor will be in the range
specified in its respective datasheet. These values include the resistance of
cables. Resistance below 0.5 ohms indicates a short circuit, while a
resistance higher than the value stated in the datasheet indicates physical
wear out of the sensor and/or its cables. In both cases, the sensor is not
functional and must be replaced.
Checking sensor behavior with a temperature difference
Connect the sensor to a voltmeter (resolution preferably in the 0.1mV range).
Place the sensor on a metallic surface at room temperature. When touching the
sensor with a warm finger on the upper surface, you should get a signal in the
mV range. A sensor signal below 0.1 mV indicates a short circuit. Check
whether the resistance of the sensor is > 0.5 ohm as described in Section 3.1.
If the signal randomly fluctuates between a positive and negative signal, or the voltages are in the +/- 1 V range, you may have an open circuit. Check the connection of your electrical probes. If the signal shows one of the three described features above, the sensor is not functional and has to be replaced. In this case, please contact the vendor.
INSTALLATION OF A gSKIN HEAT FLUX SENSOR
This section describes two general mounting options of the gSKIN® Heat Flux Sensors. The various application notes available on greenTEG’s webpage, describe additional task-specific mounting options. All application notes are available at www.greenTEG.com.
Mounting substance
In order to obtain meaningful measurement data, the gSKIN® Heat Flux Sensor
has to be mounted with adequate mounting substances. Adequate mounting
substances feature high thermal conductivity and low thickness. Three types of
mounting substances are suitable: adhesive tape, thermally conductive paste,
and thermally conductive glue.
The mounting substance should be chosen based on the measurement setup.
Greentech offers each of these mounting substances through its webshop
www.shop.greenTEG.com.
Adhesive tape
Adhesive tape should be used for simple tasks, where quick setup is crucial
and thermal coupling is of secondary importance. Clean the surface to be
measured and apply the tape to the backside of the sensor. Mount the sensor
onto the surface by applying gentle pressure to establish adhesion. You can
add the thermal paste (see next section) for improving thermal coupling to the
surface.
Thermally conductive paste
The thermally conductive paste is recommended in applications where pressure
is used to fix the gSKIN® Heat Flux Sensor in the measurement setup. It
generates a very strong thermal coupling as the paste adapts to surface
inhomogeneities. Clean the surface to be measured and spread a thin layer of
paste onto the backside of the sensor. Then press the sensor onto the surface.
You may need to hold the sensor in place with tape across the electric cables.
Thermally conductive glue
Thermally conductive glue is suitable for applications where additional
mechanical stability is required. Similar to the paste, it generates a strong
thermal coupling and adapts to surface inhomogeneities. Clean the surface to
be measured and spread a thin layer of thermal glue onto the backside of the
sensor. Then press the sensor onto the surface and follow the curing
instructions of the glue.
Removal of the mounting substance
To remove the different mounting substances, refer to the respective
manufacturer’s instruction manual. If no instructions are available, contact
the supplier. Isopropanol and ethanol can be used as cleaning agents whereas
acids and bases must be avoided to avoid damage to the sensors. Rub the
surface gently with a soaked tissue to remove residues of the mounting
substance.
Note
To remove thermally conductive tape put isopropanol or alcohol on the sensor
and wait 60 seconds and then slowly try to remove the sensor (at no point tear
at the contacts).
In case you use thermally conductive glue the sensor cannot be removed
anymore.
Mounting
The sensor responds to all three types of heat transfer: conduction, convection and radiation. gSKIN® Heat Flux Sensors are fully calibrated for measuring conductive heat flux. The conductive calibration ensures highly precise measurements for the following two measurement scenarios.
At the interface between a solid surface and gas
Figure 4: gSKIN® Heat Flux Sensor mounted on a solid surface
Mounting instructions:
- Select a representative area of the surface you want to study.
- Ensure that the area of interest is flat, dry, and free of dust and grease. Clean the sensor surface with ethanol or isopropanol. Do not use acids or bases for cleaning the sensor.
- Apply the sensor using any of the above-described mounting substances. When mounting the sensor, make sure no air is trapped between the surface and the sensor. Air gaps are thermally insulating and heavily distort the measurement results.
- Mount the sensor with the positive side of the sensor in the direction of the expected positive heat flux (as described in Section 2)
- Do not apply more than 100 N per cm2 of compressive force to the sensor at any time.
- In order to ensure meaningful results, we recommend making the exposed sensor surface similar to the finish of the surface to be measured. For example, if the surface to be measured is covered with white paint, you will get maximum accuracy by painting the sensor surface with the same paint.
Between two solid materials
Figure 5: gSKIN® Heat Flux Sensor mounted between two solid materials
Mounting instructions:
- Ensure that both solid bodies have a contact area at least as large as the gSKIN® Heat Flux Sensor.
- Ensure that the two solid planes are perfectly parallel to each other and that the contact surfaces are flat, dry and free of dust and grease. Clean the sensor surface with ethanol or isopropanol. Do not use acids or bases for cleaning the sensor.
- Mount the sensor with the positive side of the sensor in the direction of the expected positive heat flux.
- Sandwich the sensor between the two contact areas using any of the above-described mounting substances. It is highly recommended to use thermally conductive paste or glue to increase the quality of the thermal contacts between the surfaces and the sensor.
- Do not use too much thermal paste or glue as it increases the risk of thermal short-cuts between the two contact surfaces. Furthermore, ensure that no air is trapped between the surface and the sensor. Air gaps are thermally insulating and heavily distort measurement results.
- A clamping force of 10N – 100N per cm2 is recommended in order to optimize the thermal contact. The maximal value of 100N per cm2 should not be exceeded at any time.
Mounting material
Materials | Preparation | Installation |
---|---|---|
Metal | Smooth surface : |
Degrease by Organic Solvent
Rough surface :
Make the surface smooth
| Use adhesive tape
Use thermally conductive glue Use the thermally conductive paste
Concrete| Make the surface smooth| Embed to the material
Use thermally conductive glue Use thermally conductive paste
Glass, tiles| Degrease by organic solvent| Use adhesive tape
Wood| Place the sensor according to
the direction of grain
| Use thermally conductive glue
Use thermally conductive paste
Glass wool| Make the surface as smooth as possible| Embed the material
Use thermally conductive glue Use thermally conductive paste
Perlite board| Make the surface as smooth as
possible
| Use thermally conductive glue
Use thermally conductive paste
SENSOR READ-OUT
The gSKIN® Heat Flux Sensor’s output is an analog voltage signal. Depending on the measurement task, the voltage signal can be in the μV to mV range. To read-out the sensor signal, three options are available: the gSKIN® DLOG Data Logger, a voltmeter, or a 3rd party read-out device. The following section describes each option separately.
DLOG Data Logger
The gSKIN® DLOG Data Loggers are specifically developed for reliable and
straightforward heat flux measurements in combination with the gSKIN® Heat
Flux Sensors. The gSKIN® DLOG Data Loggers work as a complete solution with
included software. The gSKIN® DLOGs can be set to measure either the analog
voltage signal (in V) or heat flux signal (in W/m2). Please follow the
Instruction Manual, which is available for all gSKIN® DLOG Data Loggers.
Optionally, the gSKIN® DLOG Data Loggers can be equipped with 2 temperature sensors. Detailed information is available at greenTEG’s webshop www.shop.greenTEG.com.
Applicability
The gSKIN® DLOG Data Loggers are compatible with all gSKIN® Heat Flux Sensors
with a plug.
Readout Software
The gSKIN® DLOG Data Loggers are compatible with Heat Flux Sensor Software
(download link).
Voltmeter as a read-out device
Voltmeters are used for simple measurement tasks and/or for sensor
functionality tests. In order to read the output voltage of the sensor with
high accuracy, you need a voltmeter with high resolution. The resolution of
the heat flux measurement is limited by the voltmeter resolution and noise.
Table 1 demonstrates the relevance of voltmeter resolution.
Heat flux resolution [W/m2]
- Voltmeter resolution = 1mV 143
- Voltmeter resolution = 1μV 0.14
Table 1: Heat flux resolution of a gSKIN®-XP 26 9C Heat Flux Sensor with a sensitivity of 7.00μV/(W/m2).
The voltmeter resolution is the most critical feature when choosing the optimal device. Due to the low electrical resistance of the sensor, there are no special requirements regarding the input resistance of the voltmeter. The gSKIN® Heat Flux Sensors can be used bi-directionally: If the direction of the heat flux is reversed, the sign of the sensor voltage output changes (i.e. from positive to negative). Since the sensitivity of the sensor does not depend on the direction of the heat flux, the measurement of the reversed heat flux has the same accuracy. However, on some voltmeters, the measurement of negative voltages may not be possible or may be less accurate than the measurement of positive voltages. Further information about the positive and negative sides of the sensors can be found in Section 2.
Applicability
Voltmeters are compatible with all gSKIN® Heat Flux Sensors without a plug.
3rd party read-out device
A data logger is highly recommended for the measurement of time-dependent
variations of the sensor signal. For the choice of a suitable device, the same
considerations as for the voltmeter apply.
Applicability
3rd party read-out devices are compatible with all gSKIN® Heat Flux Sensors
without a plug.
Surface-temperature measurement read-outa
The surface temperature is measured with an NTP thermistor and can be read out
with an ohm-meter which has at least a range between 0.5 and 200kOhm. Simply
connect the green and black colored wire to the ohm-meter and read out the
resistance. With the lookup table the resistance can be transformed into the
corresponding temperature.
Temperature (°C) | Resistance (kOhm) |
---|---|
-10 | 42.889 |
-5 | 34.196 |
0 | 27.445 |
5 | 22.165 |
10 | 18.010 |
15 | 14.720 |
20 | 12.099 |
25 | 10.000 |
30 | 8.309 |
35 | 6.939 |
40 | 5.824 |
Figure 6: Small section of the look-up table. The entire table can be found onlinec
This lookup table can be approximated very well 8max 0,2° in different temperature ranges using the following equations (Steinhart Hart equations):
- 10 – 50°C: [°]=3354ln() + 8.952−273.15 where R is the measured resistance in kOhms.
- below 10°C: [°]=3171ln() + 8.296−273.15 where R is the measured resistance in kOhms.
- 50°C – 80°C: [°]=3466ln() + 9.300−273.15 where R is the measured resistance in kOhms.
DATA ANALYSIS
This section contains the basic analysis methods needed to interpret data from the gSKIN® Heat Flux Sensors. All data necessary for these can be found in the following documents:
- Calibration certificate
- This certificate is delivered with every Heat Flux Sensor for R&D Applications. It contains the sensor-specific sensitivity at calibration temperature (So) and the correction factor (Sc).
- Datasheet
- The datasheet provides an overview for all technical parameters of the gSKIN® Heat Flux Sensor. It also states the sensor area, which is necessary for calculating thermal power.
Temperature corrected sensitivity
The sensitivity of the gSKIN® Heat Flux Sensors depends on the temperature at
which they are used. This can be neglected in case your measurement takes
place at room temperature and uncertainties of +/-5% are accepted. Use the
following instructions to achieve the highest precision, especially when
operating at high temperatures (>30ºC) and low temperatures (<10ºC).
The temperature-corrected sensitivity of the sensor is calculated using the following formula:
- S= So + (Ts – To) * Sc
[μV/(W/m2)]
where
- So is the sensitivity at calibration temperature, in μV/(W/m2)
- Sc is the linear correction factor, in (μV/(W/m2))/°C
- To is the calibration temperature, in °C
- Ts is the mean sensor temperature level, in °C
Values So, Sc, and To are sensor-specific calibration values and are provided together with each gSKIN® Heat Flux Sensor purchase in the calibration certificate.
If Ts is not measured, it can be approximated by the following formula:
- Ts = (Th + Tc) / 2
[°C]
where Th and Tc are the respective temperatures of the hot and the cold reservoir of the system, in °C.
If the sensor is mounted onto a hot surface exposed to air (i.e. gas), Ts is better approximated by
- Ts = Th
[°C]
Example for calculating a temperature-corrected sensitivity
The sensor is mounted on a warm surface and is exposed to air. The surface has
a temperature of 40°C, which is a good approximation for Ts (see last
section). Furthermore, the following specifications are given by the
calibration certificate:
- So = 9.41 μV/(W/m2)
- Sc = 0.0049 (μV/(W/m2))/°C
- To = 22.5°C
The temperature-corrected sensitivity is calculated as follows:
- S= So + (Ts – To) * Sc
- = 9.41 μV/(W/m2) + (40°C – 22.5°C) * 0.0049 (μV/(W/m2))/°C
- = 9.41 μV/(W/m2) + 0.0858 μV/(W/m2)
- = 9.4958 μV/(W/m2)
The higher temperature induced a change in sensitivity of 0.9%.
Heat flux measurement
The heat flux φ describes the transfer of thermal power per surface unit and
is calculated using the following formula:
- φ = U / S
[W/m2]
where
- U is the sensor output voltage, in μV
- S is the temperature-corrected sensitivity of the sensor, in μV/(W/m2)
Example for a heat flux measurement
- Measured voltage U = 320 μV
- Temperature-corrected sensitivity S = 10.02 μV/(W/m2)
- Calculated heat flux φ = 320 μV / 10.02 μV/(W/m2) = 31.94 W/m2
Thermal power measurement
The thermal power Φ describes the amount of heat energy that passes through
the sensor area per second. The unit of thermal power is W and is calculated
using the following formula:
- Φ = A U / S = A φ
[W]
where
- A is the sensor area, in m2
- U is the sensor output voltage, in μV
- S is the temperature-corrected sensitivity of the sensor, in μV/(W/m2)
Example for a thermal power measurement
- Measured voltage U = 320μV
- Temperature-corrected sensitivity S = 10.02μV/(W/m2)
- Sensor area (datasheet) A = 10mm * 10mm
- Calculated thermal power Φ = 320μV 10mm 10mm / 10.02μV/(W/m2) = 3.19mW
MAINTENANCE OF THE SENSOR
Removing a sensor from a measurement setup
If the gSKIN® Heat Flux Sensor has been mounted using thermally conductive
tape or paste, it can be easily removed without destroying the sensor.
greenTEG’s thermally conductive tape and thermally conductive paste can be
removed following the instructions mentioned in Section 4.1.
Cleaning the sensor
Cleaning is only necessary before mounting the sensor. Clean the sensor
surface with ethanol or isopropanol. Once the sensor is mounted, no further
cleaning is necessary.
Storage
Store unused gSKIN® Heat Flux Sensors at ambient temperature in a clean and
dry place. No further care is required.
ADDITIONAL CONSIDERATIONS
Electromagnetic interference
Due to the low electrical resistance of the sensor and the aluminum coating,
the output signal is resistant to electromagnetic interference. In most cases,
no countermeasures are necessary. If electromagnetic interference is a problem
within an application, typical countermeasures (e.g. shielded cables,
grounding) have to be taken.
Troubleshooting electrical problems
In case of electrical problems, check all connections and cables. Check for
loose connections and/or short circuits in the leads. In some cases, corroded
cables are the issue. If the problem cannot be located in the leads/cables,
the sensor may be broken and has to be replaced.
Application in temperatures outside of the calibration temperature range
The calibration temperature range of the gSKIN® Heat Flux Sensors is stated in
the respective data sheets. Within this temperature range, greenTEG guarantees
a relative error of +/-3%. Outside of this range, the relative error may
exceed this value.
Influences of radiative heat flux
Electromagnetic radiation from deep ultraviolet wavelengths (DUV) to mid-
infrared (MIR) may interfere with your measurement (e.g. when measuring
windows).To achieve the highest precision make sure to block off this
radiation.
Use in fluids
The use in fluids is not recommended. However, the sensor may be exposed to
moisture or clean neutral water at temperatures less than 50° C (323.15 K) for
a short time by properly insulating all electrical parts. This may be done by
sealing with lacquer, silicone rubber or similar materials. In these
environments, it is recommended to mount the sensor with thermally conductive
glue. However, long-term exposure to wet ambient conditions is not recommended
as this may corrode the metallic leads. In any case, do not expose the sensor
to strong acids or bases.
General problems
For additional problems and considerations that are not listed in this
instruction manual please refer to the FAQ Heat Flux Measurements (download
link).
LIST OF SYMBOLS
Name | Symbol | Unit |
---|---|---|
Heat Flux Density | φ | W/m2 |
Thermal power | Φ | W |
Sensor output voltage (measured) | U | V |
Temperature corrected sensitivity | S | µV/(W/m2) |
Sensitivity at calibration temperature | So | µV/(W/m2) |
Correction factor | Sc | (µV/(W/m2))/°C |
Thermal resistivity | k | Km/W |
Absolute thermal resistance | K | K/W |
Heat transfer coefficient | H | W/(m2K) |
Thermal conductivity, Heat conductivity | λ | W/(mK) |
Electrical resistance | R | Ohm |
Temperature of the hot side | Th | °C |
Temperature of the cold side | Tc | °C |
Calibration temperature | To | °C |
Sensor temperature | T | °C |
Sensor area | A | m2 |
Sensor thickness | d | µm |
Disclaimer
The above restrictions, recommendations, materials, etc. do not cover all
possible cases and items. This document is not to be considered to be complete
and it is subject to change without prior notice.
Revision History
Date | Revision | Changes |
---|---|---|
11. November 2013 | 0.1 (preliminary) | Initial revision |
23. September 2014 | 1.0 | Updated figures, minor corrections |
20. January 2015 | 1.5 | Updated product portfolio |
1. April 2015 | 1.7 | Minor revisions |
30. October 2015 | 1.8 | Minor revisions |
12. February 2018 | 2 | Information on XO-Temp added |
23. March 2022 | 2.1 | Updated sensor polarity convention |
CONTACT INFORMATION
- Hofwisenstr. 50A 8153 Rümlang, Switzerland
- T: +41 44 515 09 15
- info@greenTEG.com
- greenTEG.com.
References
- Heat Flux & Temperature Sensing Solutions - greenteg
- Heat Flux & Temperature Sensing Solutions - greenteg
- Heat Flux & Temperature Sensing Solutions - greenteg
- murata.com/en-us/products/productdata/8796842262558/NTHCG271.txt?1437970080000
- NCP15XH103D03RC|NTC for Temperature Sensor|Temperature Sensors|Sensors|NTC for Temperature Sensor|NTC Thermistors|Thermistors|Murata Manufacturing Co., Ltd.