Linshang LS163 Transmission Meter User Manual

June 9, 2024
Linshang

Linshang LS163 Transmission Meter

Linshang LS163 Transmission Meter product

Product Introduction

The LS163/LS163A Transmission Meter is a portable instrument used for measuring the visible light transmittance, IR rejection
(transmission) value, and UV rejection (transmission) value of solar film, stick-film glass, and other materials.
The transmission meter is a portable light transmission measuring instrument, it’s able to simultaneously measure the visible light transmittance, IR rejection (transmission) value and UV rejection (transmission) value of solar film, stick-film glass and other materials.

Parameters

The product has two different models: LS163 and LS163A. The IR peak wavelength of LS163 is 940nm, and IR peak wavelength of LS163A is 1400nm.

  1. UV Peak wavelength: 365nm
  2.  IR Peak wavelength: 940nm/ 1400nm
  3. Visible light wavelength: 380-760nm, conform to the CIE photopic vision function standard
  4. Accuracy: ±2% (Colorless and transparent material)
  5.  Resolution: 0.1%
  6. Test Slot: 6mm 58mm (WD)
  7. Meter Size: 146mm 73mm 24mm (LWH)
  8.  Display: 128*128 Dot matrix LCD
  9. Weight: 187g (Include Batteries)
  10. Power supply: 4*AAA alkaline battery

Operation

Power On/Off
  • To power on the meter, short press the button. The test sample cannot be placed in the test slot before powering on otherwise the self-calibration cannot be completed.
  • The serial number and version number will be displayed on the interface, and the meter will enter into the measurement interface after 3 seconds.
  • To power off the meter, long press the button for 3 seconds.
  • The meter will automatically power off after 3 minutes without any operation.

Power on self-test: Short press the “ ” button to power on the meter. The test sample can’t be placed in the test slot before powering on otherwise the self-calibration can’t be completed. The serial number, version number will be displayed on the interface and the meter will enter into the measurement interface after 3s. The meter displays the corresponding measurement interface according to the setting parameters.
Manual shutdown: Long press the “ ” button for 3s to power off the meter.
Auto shutdown: The meter will automatically power off after 3 minutes without any operation.

Parameter Setting
  • In the off state, long press the button for 3 seconds to enter into the parameter setting mode.
  • Short press the button to select Chinese or English, long press for 2 seconds to confirm the option and enter the next setting item.
  • Short press the button to select the parameters, and long press the button for 2 seconds to confirm the parameters.
  • The meter enters into the measurement interface. Optional parameters are UVR/IRR/VLT (default option), UVT/IRT/VLT, and
    VLT.

In the off state, long press the “ ****” button for 3s and enter into the parameter setting mode.

Language
Short press the button to select Chinese or English, long press for 2s to confirm the option and enter the next setting item. Display Item
Short press the “ ” button to select the parameters and long press the “ ” button for 2s to confirm the parameters. Now the meter enters into the measurement interface. Optional parameters are as following:

  • UVR/IRR/VLT (default option): UVR (UV rejection), IRR (IR rejection), VLT (Visible light transmittance) are displayed on the interface.
  • UVT/IRT/VLT: UVT (UV transmission), IRT (IR transmission), VLT (Visible light transmittance) are displayed on the interface.
  •   VLT: Only VLT (Visible light transmittance) is displayed on the interface.

UVR/IRR/VLT
UVT/IRT/VLT
VLT

Operation
  • When testing the sample in the measurement state, short press the button to hold the data. The HOLD symbol will be displayed on the lower left corner on the interface.
  • To return to the measurement state, short press the button again.
  • In the HOLD state, if the backlight is off, short press the button to light up the backlight.
  • If the backlight is on, it will cancel the HOLD state and start a new measurement.

When we test the sample in the measurement state, short press the “ ” button to hold the data, the “HOLD” symbol will be displayed on the lower left corner on the interface. Short press the “ ” button again, the meter will return to the measurement state. In the “HOLD” state, if the backlight is off, short press the “ ” button to light up the backlight. If the backlight is on, it will cancel the “HOLD” state and start a new measurement.

Display Interface Autorotation

The meter display interface will rotate adaptively, which is convenient for reading measurement data.  The interface shown as below: Linshang LS163
Transmission Meter 06

 Backlight Auto Lit

In the measurement state, if the backlight is off, the backlight will be automatically on when the test sample is placed in the slot and be off again without any operation after 30 seconds.

 Low Power Alarm

When the battery power is low, the empty battery symbol will display as screen and flash, indicating that the battery needs to be replaced.

Features

  1.  Test slot is 6mm which can directly test the window glass.
  2. Adopt LCD dot matrix with backlight which can be used without lights.
  3. The display interface will rotate adaptively which is convenient for reading measurement data.
  4. . Equipped with real-time dynamic self-calibration function, which is automatically self-calibrated after powering on.
  5. Three display interfaces can be selected. (UVR/IRR/VLT, UVT/IRT/VLT and VLT)
  6. Ultra-low power consumption, 4 AAA alkaline batteries can maintain continuous measurement for more than 100 hours
  7. UV rejection meter, IR rejection meter, light transmittance meter. Three functions in ONE device.
  8.  Simple operation, putting the testing sample in the test slot, UV rejection, IR rejection, visible light transmittance values of the sample will be displayed simultaneously.
  9. Applicable to solar film, stick-film glass and other materials performance exhibition, production, quality inspection and other occasions.

Precaution

  1.  When the meter can’t complete power on self-calibration, the interface will prompt “E30”. Possible causes:
    • The ambient light is too strong.
    •  Test sample is placed in the test slot, just take away the test sample and turn on the meter again.
    • Dust enters the test hole and blown off the dust with a dust blower.
    •  The meter has hardware failures and needs to be returned to the factory for maintenance.
  2.  When the meter is not in use, turn off the power.
  3.  Avoid to contact with corrosive materials and be away from high temperature and high humidity environment.
  4. After using for a long time, due to the luminous efficiency of the LED light source (the temperature of the LED light source increases, the efficiency decreases), the data of UVR, IRR and VLT may not be displayed as “0.0, 0.0, 100” when there is no test sample or VLT, UVT ,IRT may not be displayed as “100, 100, 100” . At this time, turn off the meter and then restart it , the meter will self-test and self-calibrate, the measurement accuracy and normal use of the meter will not be affected.

Packing List

No. Description Quantity Unit
1 Transmission Meter 1 set
2 User Manual 1 pcs
3 AAA Battery 4 pcs
4 Certificate / Warranty Card 1 pcs
5 Drawstring Bag 1 pcs

Service

  1. The meter has one-year warranty. If the meter works abnormally, please send the whole meter to the company for maintenance.
  2. Provide users with spare parts and lifelong maintenance services.
  3.  Provide the users with the meter inspection service for free.
  4. Free technical support for long term.

Manufacturer: Shenzhen Linshang Technology Co., Ltd.
Website: www.linshangtech.com
Service hotline: 086-0755-86263411
Email: sales21@linshangtech.com

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals