HEATHROW SCIENTIFIC 120209 Vortexer Mixer Instruction Manual
- June 10, 2024
- HEATHROW SCIENTIFIC
Table of Contents
- Safety Precautions
- Specifications
- EC Declaration of Conformity
- Modes of Operation
- Loading Tubes
- Vortexer head removal
- Vortexer elastic cord positioning
- Care and Handling
- Disposal Responsibilities
- Product Return Information
- Warranty
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
HEATHROW SCIENTIFIC 120209 Vortexer Mixer
To immediately shut off power to Vortexer device, pull out the AC cord from
the lower central back area of the device
Thank you for purchasing the Heathrow Scientific Vortexer. Your
Vortexer mixing instrument is designed for general laboratory use
Safety Precautions
Your Vortexer instrument is designed for general laboratory use. This
instruction manual cannot address all safety hazards. It is the responsibility
of the user to consult and observe all health and safety precautions and to
assess the instrument’s suitability to perform the task.
Caution!
Before starting the Vortexer for the first time, please read this instruction
manual carefully. This manual contains important information, safeguards, and
operating instructions.
- Never use this product in any manner inconsistent with these instructions
- This product is intended to be used indoors only.
- Always ensure that the disconnecting plug (power plug) is accessible.
- This product is intended for vortexing and mixing aqueous solutions in approved test tubes compatible to the speed and power of this unit. Note on Vortexing: use only approved containers with screw-tight or lockable lids.
- The product is not intended to vortex very dense materials. Do not use materials with a density greater than 1.2g/cm.
- Observe general laboratory safety precautions and regulations when using this product.
- Do not attempt to operate with the mixing head attachment removed. Vortexer may become destabilized in operation.
- The mixing head attachment must always be securely fastened when in operation: Always make certain that the head attachment is seated completely onto the top shaft before operating.
- Do not use if the Vortexer is damaged.
- Do not put hands onto or under the mixing head attachment unless the Vortexer is completely stopped.
- Do not move unit while the Vortexer is in operation.
- Do not immerse the product in water. Injury due to shock or fire may result.
- The mixing head attachment must be loaded symmetrically. Operating this unit with an unbalanced load will cause severe vibration and damage, which is not covered by warranty.
- Do not fill tubes while inserted onto the Vortexer mixing head. Spilled liquid may harm the unit or the user.
- Close tube lids/caps and confirm that they are completely secure before starting the Vortexer instrument. Open tube lids can be torn off during vortexing and damage the unit. Open lids can cause fluid to be dispersed inside the vortex instrument
- Do not insert non-standard tubes or other foreign articles into the Vortexer head.
- Follow all precautions from the Material Safety Data Sheet (MSDS) for any reagent you use with this instrument.
- Do not use solvent or flammable liquids near this or other electrical equipment.
- Do not vortex materials that in combination or alone are flammable, explosive, or corrosive or may cause any other hazard.
- Do not operate in a hazardous or flammable environment
- Do not use any solvent on the unit that may degrade the plastic of the Vortexer main housing or head.
- Ensure the mixing head is protected from corrosion and mechanical damage. The mixing head must be cleaned with a pH-neutral cleaning liquid.
- Always work in a manner which endangers neither the user nor any other person.
- Should the instrument fail to work to specification, immediately stop vortexing. Contact the authorized dealer or the manufacturer if trouble persists.
- Do not attempt to make contact with the mixing head attachment while unit is running. Doing so may cause the unit to fail and will void the warranty.
- Only use original manufacturer’s mixing head attachment, AC adapters, and any other spare parts.
- If unit is malfunctioning or needs to be returned, please contact either the authorized dealer or Heathrow Scientific® LLC .
- Improper use of the Vortexer instrument will void the warranty. If there is a failure during the warranty period, contact Heathrow Scientific® LLC for warranty service.
- Do not attempt to use units that have not been correctly installed or repaired.
- Do not attempt to disassemble or modify this product. Any attempt to do so will void the warranty.
- Store at room temperature in a dry area. Do not expose to sunlight, moisture, or extreme temperatures for prolonged periods of time.
- Before using any cleaning or decontamination methods please contact Heathrow Scientific® LLC to ensure that the proposed method will not damage the equipment. A pH-neutral cleaning liquid must be used.
- Vortexer must be unplugged from electrical source before any cleaning or maintenance is performed.
To immediately shut off power to Vortexer device, pull out the AC cord from the lower central back area of the device
Attention: If main electrical power is interrupted, Vortexer device will
immediately begin operating if 3-position switch has not been set to OFF
position when power is restored.
If power is interrupted: 1) Unplug cord from back of device. 2) Set Vortexer
switch to OFF
Warning!
- Always check for device movement when running in continuous mode. If movement of device is noted, adjust speed accordingly. Different loading of mixing head may result in reaching a resonance frequency whereby the device will move on the work surface/ table top.
- If the unit is not used as recommended by the manufacturer, the overall safety will be impaired.
- Check MSDS, wear required Personal Protective Equipment, and observe all applicable local and national regulations before dispensing and disposing of hazardous samples.
- The laboratory procedure should ensure that no person or hazardous substances are present within a 12″ (30cm) zone around the Vortexer instrument when operating.
- This product does not contain bio-seals as per IEC/EN/CSA 61010-2-20 and cannot provide any level of containment in case of a spill or release of toxic, radioactive, or pathogenic micro-organisms thus these materials are not recommended to be used in this product.
Specifications
Product Description/Features: This device is a benchtop laboratory
Vortexer with a variable speed control (OFF, and up to approx. 3000 rpm),
capable of vortexing microplates and 6 different tube sizes with the
convenience of a single mixing head attachment. This Vortexer has enough
flexibility to perform the standard tasks that a vortex mixer is required to
perform, plus it has the ability to hold a variety of tubes, containers, and
microplates all within one mixing head design.
Supplied Equipment and Spare Parts:
- 1 ea. Vortexer (classified as Installation Category 2) HS120209, US/North America: 120VAC (110V 120V, 50-60 HZ, 55 Watts, 1.0 Amps) HS120318, Japan (PSE cord): 100VAC (100V 110V, 50-60 Hz, 66 Watts, 1.3 Amps) HS120214, United Kingdom: 230V (220V 240V, 50-60 Hz, 44 Watts, .4 Amps) HS120212, EU: 230VAC (220V 240V, 50-60 Hz, 44 Watts, .4 Amps) HS120210, Australian: 230VAC (220V 240V, 50-60 Hz, 44 Watts, .4 Amps)
- 1 ea. AC power cord (WARNING only use power cords supplied by Heathrow Scientific): HS120329: N. America HS120488: Japan PSE rated HS120331: UK HS120332: CE HS120333: Australia
- 1 ea. mixing head, HS120327
- 3 additional elastic retention cords, HS120422
- 3 additional central pads for intermittent mixing area, HS120328
**Technical Features
**
Speed range: From 0 to approximately 3000 rpm – shaking to vortexing
Operation:
1) 3-position slider power switch:
Top position: Intermittent operation Center position: OFF Bottom position:
Continuously ON
2) Infinite position rotary speed control for speeds from Off up to
approximately 3000 rpm
Dimensions: 6.75″ W x 7.5″ D x 6″ Height of main unit. 7.75″ Height with
Mixing Head attached. (171mm W x 190mm D x 197mm H)
Weight: 8.66 lbs (3.93 kg) total Vortexer weight with Mixing Head
attached
Orbital Diameter: 3.7mm
Electrical: See HS part numbers listed above.
Note: Product has a Thermal Protection Device.
Environmental:
For Indoor Use Only Altitude: up to 2,000 m Operating Temperature: 2° C to 40°
C (35.6° F to 104° F), Max. Humidity 95%Max. Relative Humidity 80% non-
condensing
This mark is the confirmation that the unit conforms to the EU guidelines and has been tested according to the following EU Declaration of Conformity.
Conforms to:
ANSI/UL Std. 61010-1 ANSI/UL Std. 61010-2-051
IEC/EN 61010-1
IEC/EN 61010-2-051
CAN/CSA C22.2 No. 61010-1 CAN/CSA C22.2 No. 61010-2-051
This Product is in Compliance with: RoHS 3, PSE Certified Cord (Japan), WEEE
EC Declaration of Conformity
Manufacturer : Heathrow Scientific® LLC Address: 620 Lakeview Parkway,
Vernon Hills, IL 60061 USA European Contact: Emergo, Europe
Address: Princessegracht 20, 2514AP, The Hague, The Netherlands
Model: Vortexer, HS120209, HS120210, HS120212, HS120214, HS120318
This unit has been constructed and conforms to the following:
Safety Standards:
IEC/EN61010-1:2010, ed. 3.1, 2017
IEC/EN61010-2-051, ed. 3, 2016
UL Std. 61010-1, ed. 3, 2016
UL Std. 61010-2-051, ed. 3 (R2016)
CAN/CSA C22.2 No. 61010-1-12 (R2017)
CAN/CSA C22.6 No. 61010-2-051 (R2016)
2014/35/EU (Low Voltage Directive)
EMC Standards:
EN61326-1: 2013 (IEC 61326-1:2012)
FCC Part 15, Subpart B : 2015 ICES-003, Issue 6
2012/19/EU (WEEE)
2015/863/EU (RoHS)
Signed
:
Gary Kamees
Vice President Product Development/Manufacturing
Vernon Hills, IL
January 2020
Note: Changes or modifications to the product not expressly approved by
the manufacturer could void the user’s authority to operate the equipment.
Note: This product has been tested and found to comply with the limits of
a Class 1A digital device pursuant to Part 15 of the FCC rules. These limits
are designed to provide reasonable protection against harmful interference
when the product is operating in a commercial environment. The product
generates, uses, and can radiate radio frequency energy and if not installed
and used in accordance with the user’s manual may cause harmful interference
to radio communications. Operation of the product in a residential area is
likely to caused interference in which case the user will be required to
correct the interference.
Your Vortexer is supplied with a pop-on mixing head attachment and an AC power cord located in the top corrugated box insert (1) of the product packaging. After you fully unpack your Vortexer from the package and remove the protective plastic bag, install the mixing head attachment (check that it is right side up).
IMPORTANT: YOU MUST PULL OUT INNER FRONT SUPPORT CARD (3) BEFORE
REMOVING YOUR HEATHROW SCIENTIFIC® VORTEXER. USE TWO HANDS TO GRASP THE UNIT
WHEN LIFTING OUT OF THE BOX. (4)
TO INSTALL THE MIXING HEAD ATTACHMENT YOU MUST PRESS DOWN FIRMLY WITH BOTH
HANDS UNTIL YOU FEEL THE MIXING HEAD ATTACHMENT SNAP FULLY INTO PLACE.
This will ensure that the mixing head attachment will not come off the
Vortexer during operation
Note: On units equipped with a frequency switch (located above the power cord connection on rear) set switch to either 50Hz or 60Hz to match the local power company supply. Turn the speed rotary dial turned all the way to the left- (in the OFF position). Plug the device into a properly grounded outlet.
**** To immediately shut off power to Vortexer device, pull out the AC cord from the lower central back area of the device
Modes of Operation
WARNING: If the Vortexer is used in a manner not specified by Heathrow Scientific, the protection provided by the equipment may be impaired.
Duty Cycle for ALL VERSIONS: Maximum run time of 20 minutes followed by 60 minute OFF time when in Continuous Operation mode. 100V-110V Vortexer: Set power input frequency switch on rear of unit to either 50Hz or 60Hz to match local power company supply. 220V-240V Vortexer: Set power input frequency switch on rear of unit to either 50Hz or 60Hz to match local power company supply.
The rotary speed control dial, located in the lower center of the front panel,
has infinite adjustability (OFF, and 0 to approx 3000 rpm). Vortexing speed
increases to a maximum of approx. 3,000rpm as the knob is turned clockwise.
Before turning on the Vortexer, set the speed dial to OFF position (rotate
counter-clockwise until the knob stops and a “click” is felt). To begin
operation, move the power switch to either the Intermittent TOUCH or ON
position and turn the speed control to the desired setting. If the power
switch is in the ON position, the mixing head will begin moving immediately.
If the power switch is in the Intermittent TOUCH position, operation will
begin when the rotary speed control dial is rotated clockwise to the maximum
speed and an object (test tube) is placed in the center of the mixing head,
held at the top of the test tube, and moderate pressure is applied to push
down the mixing head. Mixing will cease when the object is removed and
pressure is released.
Intermittent/touch mode:
Place the 3-position switch in the Top Intermittent TOUCH position and rotate
the speed control dial clockwise to maximum speed. When pressure is released
from the mixing head, operation will stop. Intermittent operation will begin
when a test tube is pressed down in the center area of the mixing head, while
the tube is held firmly at the top.
Hands free/continuous on mode:
Place the 3-position switch in the ON (bottom) position and set the rotary
speed control as described above. Hands Free mode allows mixing or vortexing
of the tubes or microplate and will continue to run until you stop the mixer
by placing the 3-position switch to the OFF position.
Note: To prevent overheating of the Vortexer during continuous use, there
is an internal Temperature Protection Device that will temporarily shut off
the device. Once the device cools down, it will be operational again. Please
set switch to OFF position whenever it is not in operation.
Attention : It is recommended to NOT leave the Vortexer unattended while
in operation.
OFF mode:
Always place the 3-position switch to the OFF position (center) when not in
use!
Loading Tubes
The Vortexer mixing head is designed to hold several different tube sizes for
hands-free mixing/vortexing.
Caution: Always load onto mixing head so that the tubes are balanced
B. Loading of
Standard Assay tray (maximum of 1)
C. Loading of 50ml tubes (maximum of 2)
Note: Two retention cords should be used to secure 50 mL tubes
D. Loading of 15 mL tubes (maximum of 2)
E. Loading of 5 mL tubes (maximum of 2)
F. Loading of 1.5/2 mL tubes (maximum of 4)
G. Loading of .5 mL tubes(maximum of 6)
H. Loading of .2 mL tubes (2 tube strips – maximum of 24)
Vortexer head removal
In the event of a cleaning being needed for the Vortexer, please follow the sequence of removal for the mixing head.
Turn the power OFF by placing the 3-position switch into the OFF (center) position.
Grasp the installed mixing head attachment with two hands and pull straight up
until it releases from the shaft. A technique of lifting the attachment with
your fingertips while using your knuckles as a fulcrum will give you the
leverage to remove the installed attachment component smoothly. Once Vortexer
is cleaned, reinstall the mixing head per instructions.
Vortexer elastic cord positioning
Care and Handling
Your Vortexer should be given the care normally required for any electrical appliance. Avoid wetting or unnecessary exposure to fumes. The finish can be washed with water (after unplugging) and mild soap or detergents, using a soft cloth or sponge. Avoid exposing any electrical components of the Vortexer instrument to water, cleaning agents, or other liquid! Keep the unit clean by immediately blotting any spills. Replacement parts are available through most laboratory equipment distributors or directly from Heathrow Scientific, LLC.
CAUTION! Unplug from power before cleaning. Do not immerse. Avoid getting any liquid into the Vortexer housing.
Disposal Responsibilities
The equipment you purchased may contain hazardous substances that could impact the environment. Per regulations on electronic devices in the European Community, you must use the appropriate disposal systems to avoid exposure of these substances to the environment. The disposal systems will reuse or recycle hazardous materials from your equipment responsibly.
The crossed-out wheeled bin symbol invites you to use those systems.
If you need more information on the collection, reuse and recycling systems, please contact your local or regional waste administration. You can also contact us for more information on the environmental performance of our products.
Product Return Information
Important:
Transporting hazardous materials without a permit is a violation of federal
law.
Heathrow Scientific® LLC will not accept any product return that is not
appropriately cleaned and decontaminated. In the unlikely event of damage to
the unit necessitates return, contact Heathrow Scientific® LLC and obtain
return authorization before sending your product for service.
Warranty
IN NO EVENT WILL HEATHROW SCIENTIFIC® LLC’S OBLIGATION UNDER THIS WARRANTY
EXCEED THE PRICE OF THE PRODUCT.
Limited Warranty Heathrow Scientific® LLC warrants that Vortexer will be
free from defects in workmanship and material for 5 years from the date of
purchase. If you believe that there is a defect in the product, you must,
during the warranty period, notify Heathrow Scientific® LLC, provide proof of
purchase, and return the product to Heathrow Scientific® LLC with a Return
Authorization form. To obtain a Return Authorization form, please call
1-847-816-5070. If Heathrow Scientific®
LLC is properly notified and, after inspection, confirms that there is a
defect and the warranty period has not expired, Heathrow Scientific® LLC will
repair, modify, or replace the product, at its sole option, at no charge.
OTHER THAN THIS LIMITED WARRANTY, HEATHROW SCIENTIFIC® LLC MAKES NO
WARRANTIES, EXPRESS OR IMPLIED, WITH RESPECT TO THE QUALITY OR PERFORMANCE OF
THE PRODUCT, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF
MERCHANTABILITY AND FITNESS FOR PARTICULAR PURPOSE WHICHARE HEREBY DISCLAIMED
AND EXCLUDED. HEATHROW SCIENTIFIC® LLC WILL IN NO EVENT BE LIABLE FOR ANY LOSS
OF USE, LOSS OF PROFITS, CONSEQUENTIAL, SPECIAL, EXEMPLARY OR PUNITIVE
DAMAGES.
THIS WARRANTY DOES NOT COVER:
ANY DEFECT OR DAMAGE CAUSED BY IMPROPER OR UNREASONABLE USE OF THE PRODUCT.
(HEATHROW SCIENTIFIC® LLC VORTEXER IS DESIGNED FOR USE ONLY BY TRAINED
LABORATORY TECHNICIANS. USE BY ANYONE ELSE WILL VOID THIS WARRANTY.)
ANY PRODUCT THAT HAS BEEN, IN HEATHROW SCIENTIFIC® LLC’S SOLE JUDGEMENT,
TAMPERED WITH, ALTERED, OR REPAIRED BY ANYONE OTHER THAN HEATHROW SCIENTIFIC®
LLC.
ANY PRODUCT THAT IS INOPERATIVE BECAUSE OF: (a) WEAR OCCASIONED BY USE, (b)
NEGLIGENCE, (c) ACCIDENT, (d) INCORRECT MAINTENANCE, OR (e) USE UNDER ABNORMAL
CONDITIONS OF TEMPERATURE, DIRT OR CORROSION, OR USE WITH ABRASIVE OR
CORROSIVE MATERIALS.
ACCESSORY PARTS, SUCH AS RUBBER AND PLASTIC PARTS THAT ARE DAMAGED BY LIQUIDS
OR MISUSE
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>