Enbio PRO Autoclave Enbio S LED Light User Manual
- June 10, 2024
- enbio
Table of Contents
- Enbio PRO Autoclave Enbio S LED Light
- Product Information
- Enbio S Sterilization Parameters
- Enbio PRO Sterilization Parameters
- Precautions, Requirements, Recommendations
- Product Usage Instructions
- Scope of Delivery and Unpacking
- Installation and Preparation for Operation
- Preparation and Loading of Instruments
- Operating the Device
- Maintenance and Upkeep
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
Enbio PRO Autoclave Enbio S LED Light
Product Information
The Enbio S and Enbio PRO are small class B steam sterilizers
that comply with the ISO 13060 standard. They are classified as
class IIb medical devices according to Annex IX of the Medical
Devices Directive 93/42/EEC.
The devices are suitable for use in primary health care
practices, dental practices, and treatment rooms. However, they are
not designed to sterilize liquids, pharmaceutical products,
biomedical waste or materials incompatible with steam
sterilization. Processing such loads may result in improper
sterilization and/or damage to the autoclave and pose a risk to
user/patient safety.
Enbio S Sterilization Parameters
- Operating Temperature: 121°C
- Operating Time: 15-20 minutes
- Pre-Vacuum: Yes
- Drying Time: 1-20 minutes (adjustable)
Enbio PRO Sterilization Parameters
-
Operating Temperature: 121°C, 134°C
-
Operating Time: 6-35 minutes (depending on selected
program) -
Pre-Vacuum: Yes
-
Drying Time: 1-20 minutes (adjustable)
Precautions, Requirements, Recommendations
Before using the device, it is important to carefully read and
follow all instructions provided in the user manual. Here are some
general precautions and recommendations:
-
Only use the device for its intended purpose.
-
Do not attempt to modify or repair the device yourself.
-
Ensure that the device is properly grounded and connected to a
suitable power supply. -
Use only distilled or deionized water for the device.
-
Do not overload the device beyond its specified capacity.
-
Regularly inspect and clean the device according to the
instructions provided in the user manual.
Product Usage Instructions
Scope of Delivery and Unpacking
When you receive your Enbio S/Enbio PRO sterilizer, follow these
steps for unpacking:
-
Remove all packaging materials from the device.
-
Inspect the device for any visible signs of damage that may
have occurred during shipping. If damage is found, do not use the
device and contact Enbio immediately. -
Ensure that all parts and accessories are included as described
in the user manual.
Installation and Preparation for Operation
Follow these steps to install and prepare your Enbio S/Enbio PRO
sterilizer for use:
-
Install the HEPA filter.
-
Connect the device to a suitable feed water source using the
WATER IN connection. -
Connect the device’s wastewater outlet to a suitable wastewater
tank using the WATER OUT connection. -
If you are using the Enbio Magic Filter, connect it to the
device’s feed water and wastewater outlets. -
Power on the device and follow the instructions provided in the
user manual to set it up for use.
Preparation and Loading of Instruments
Before sterilizing instruments, follow these steps:
-
Ensure that all instruments are properly cleaned and packaged
according to the instructions provided in the user manual. -
Place the packaged instruments on the sterilizer tray according
to the instructions provided in the user manual.
Operating the Device
To operate the Enbio S/Enbio PRO sterilizer, follow these
steps:
- Select the desired program from the device’s menu.
- Load the sterilizer tray into the device.
- Close and lock the device’s door.
- Follow the instructions provided in the user manual to start
the sterilization cycle.
Maintenance and Upkeep
To ensure proper maintenance and upkeep of the Enbio S/Enbio PRO
sterilizer, follow these steps:
-
Regularly clean and inspect the device according to the
instructions provided in the user manual. -
Replace any consumable parts (such as the HEPA filter) as
needed.
2022.04.27 The latest version of the manual is available at www.enbio.com
Enbio S / Enbio PRO User Manual EN
Enbio S
General view
561 mm
Rear view
HEPA filter seat
WATER IN feed water inlet
Condensate outlet WATER OUT
Enbio PRO
General view
561 mm
Rear view
HEPA filter seat
WATER IN feed water inlet
Condensate outlet WATER OUT
Drawer front
LCD display
Locking knob
252 mm
Power socket USB socket Main switch
Drawer front
LCD display
Locking knob
270 mm
USB socket Power socket Main switch
202 mm
162 mm
Table of Contents
1. Introduction
2
1.1 Purpose and scope
2
1.2 Applicable legislation
2
1.3 General description of the device
2
1.4 Purpose of the device
3
1.5 Enbio S sterilization parameters
4
1.6 Enbio PRO sterilization parameters
6
1.7 Symbols used on the device
8
1.8 Precautions, requirements, recommendations
8
2. Scope of delivery and unpacking
10
2.1 Unpacking the device
10
2.2 Sterilizer accessories
11
3. Installation and preparation for operation
12
3.HEPA filter installation
12
3.2 Feed water quality
13
3.3 Water connection with feed and wastewater tank
14
3.3.1 Connecting the supply water WATER IN
14
3.3.2 Connecting the wastewater WATER OUT
14
3.4. Water recycling connection (Enbio Magic Filter)
16
3.4.1 WATER IN feed water connection with Enbio Magic FIlter
16
3.4.2 WATER OUT wastewater connection with Enbio Magic FIlter
16
3.5. Setting up the device
19
3.6. Connecting the device to the power supply
19
4. Preparation and loading of instruments
20
4.1 Preparation of packages
20
4.2 Arranging packs on sterilizer tray (sterilizing packaged loads)
21
4.3 Arranging the packages on the sterilizer tray (sterilization of loads without packaging)
21
5. Operating the device
22
5.1. First run
22
5.2 Program selection
22
5.3 Test programs
26
5.4 Info menu
31
5.4.1 LED lighting
32
5.4.2 Counters
32
5.5 Restart
35
6. Maintenance and upkeep
37
6.1 Consumable parts
39
7. Data archiving
41
8. Enbio Data Viewer software
42
8.1 Installing the software
42
8.2 Program structure and main functionalities
45
9. Warning messages and error codes
48
9.1 Warning messages
48
9.2 Information messages
48
9.3 Error codes
49
10. Complaint procedure
53
11. Warranty conditions
53
12. Technical data
54
13. EC declaration of conformity
56
1
1. Introduction
1.1 Purpose and scope
The purpose of these operating instructions is to provide information on the
ENBIO S and ENBIO PRO sterilizer. In particular, information concerning:
· Intended use · Correct installation and settings · Correct use and operation
· Safe and reliable operation · Regular and proper maintenance and servicing ·
Handling situations
1.2 Applicable legislation
The ENBIO S and ENBIO PRO sterilizers are designed and manufactured to meet
the following requirements: · EN 13060 standard “Small Steam Sterilizers” and
related documents · Directive 93/42 / EEC on medical devices · Directive
2012/19 / EU on waste electrical and electronic equipment (WEEE) · Law on
Medical Devices
1.3. General description of the device
ENBIO S and ENBIO PRO are small steam sterilizers designed to sterilize
medical devices with steam. They have a hermetically sealed, heated chamber.
The sterilized load is placed inside the chamber on a special perforated tray.
After closing the chamber, the user selects and starts the appropriate
sterilization program via a touch screen. Sterilization proper begins after
the pre-vacuum phase. The steam generator generates steam and introduces it
into the chamber. This steam transfers its energy to the sterilized load.
Inside the chamber, the correct temperature and pressure, depending on the
sterilization cycle selected, are maintained for a specific period of time.
After this time, the chamber is emptied of steam and the drying cycle begins.
When the sterilization process is complete, the device displays a summary
information and the result of the process to the user.
1.4. Purpose of the device
The ENBIO S and ENBIO PRO devices are small class B steam sterilizers in
accordance with the EN 13060 standard; they are also classified as class IIb
medical devices according to Annex IX of the Medical Devices Directive 93/42 /
EEC.
The devices are suitable for sterilizing loads for which steam sterilization
is specified. The following medical devices can be sterilized in the ENBIO S
and ENBIO PRO: solid loads, small porous items, small porous loads, full
porous loads, simple hollow items, items with narrow clearance, multiple
packages that can be unpackaged or packaged (single- or multi-layer). The
134°C FAST process is dedicated exclusively to solid, non-porous, simple
instruments and dental instruments (e.g. scissors, handles, pliers, chisels,
probes, etc.) only unpackaged, non-textile.
The ENBIO S and ENBIO PRO are not designed to sterilize liquids,
pharmaceutical products, biomedical waste or materials incompatible with steam
sterilization. Processing such loads may result in improper sterilization
and/or damage to the autoclave and pose a risk to user/ patient safety.
The ENBIO S and ENBIO PRO devices can be used in primary health care
practices, in dental practices, and in treatment rooms. The sterilizers, which
comply with the guidelines of the ISO 13060 standard, are suitable for
operation in the environment of other electrical medical devices.
The Enbio PRO sterilizer has a PRION program dedicated to one of the
decontamination steps for items suspected of having or may have come into
contact with transmissible prion protein (e.g. CreutzfeldtJakob disease, BSE,
etc.). Detailed information and recommendations for infection control of
transmissible spongiform encephalopathies are provided in the document “WHO
Infection Control Guidelines for Transmissible Spongiform Encephalopathies –
Report of a WHO consultation. (Geneva, Switzerland, 2326 March 1999)”. It is
the responsibility of the user of this device to follow these guidelines and
any local/ national guidelines.
Non-medical use: The ENBIO S and ENBIO PRO devices can also be used for non-
medical applications, e.g. in beauty and wellness salons, veterinary
practices, cosmetic surgeries, tattoo studios, piercing studios and
hairdressing salons.
The devices are intended for professional use only by properly trained staff.
2
3
1.5. Enbio S sterilization parameters
The table below, which presents the characteristics of the individual
sterilization programs in the Enbio S device, should be read in detail. It is
the user’s responsibility to select the program appropriate to the type of
load and not to exceed the maximum weight of the sterilized load. Failure to
comply with the following rules may endanger the health of patients and the
operation of the device.
Enbio S
Program name Description of the load
134°C
solid loads, small porous objects, small porous loads, full porous loads, simple dimpled objects, narrow clearance objects, multiple packages that may be unpackaged or packaged (single or multiple layers)
Sterilization temperature
134°C
Sterilization Drying time time
4 minutes
3 minutes
Max. load weight
Number of Total process pre-vacuums time (for a load of
specified weight)*
500 grams 3
load of 100 grams 13 minutes
Max. water consumption
115 ml
Sterilization process class
B
121°C 134°C FAST **
solid loads, small porous objects, small porous loads, full porous loads,
simple dimpled objects, narrow clearance objects, multiple packages that may
be unpackaged or packaged (single or multiple layers)
solid, non-porous, simple dental instruments and appliances (e.g. scissors,
handles, pliers, chisels, probes, etc.) only unpackaged, non-textile
121°C 134°C
15 minutes 5 minutes
500 grams 3
3.5 minutes no drying
500 grams 1
load of 100 grams 110 ml
B
26 minutes
load of 100 grams 105 ml
S
7 minutes
- Ambient temperature can have an effect on increasing the total process time. * The total duration of the first process may be longer due to the need for the appliance chamber to warm up. After sterilization at 134°C FAST the loads are wet. It is recommended to leave the drawer ajar in order to evaporate the remaining water and to lower the temperature of the instruments. After sterilization in the 134°C FAST program, the instruments are intended for immediate use.
4
5
1.6. Enbio PRO sterilization parameters
The table below shows the characteristics of each sterilization program in
Enbio PRO. It is the user’s responsibility to select a program suitable for
the type of load and not to exceed the maximum weight of the sterilized load.
Failure to observe the following precautions may compromise patient health and
device operation.
Enbio PRO
Program name
Description of the load
Sterilization temperature
134°C
solid loads, small porous objects, small porous loads, full porous loads, simple dimpled objects, narrow clearance objects, multiple packages that may be unpackaged or packaged (single or multiple layers)
134°C
Sterilization time
4 minutes
Drying time
Max. load weight
4 minutes 800 grams
Number of Total process pre-vacuums time (for a load of
specified weight)*
3
load of 200 grams
18 minutes
Max. water consumption
190 ml
Sterilization process class
B
121°C
134 ° C FAST **
solid loads, small porous objects, small porous loads, full porous loads,
simple dimpled objects, narrow clearance objects, multiple packages that may
be unpackaged or packaged (single or multiple layers)
solid, non-porous, simple dental instruments and appliances (e.g. scissors,
handles, pliers, chisels, probes, etc.) only unpackaged, non-textile
121°C 134°C
15 minutes
5 minutes 800 grams 3
3.5 minutes
no drying 800 grams 1
load of 200 grams 180 ml
B
31 minutes
load of 100 grams 140 ml
S
10 minutes
134 ° C PRION solid loads, small porous objects, small porous loads, full porous loads, simple dimpled objects, narrow clearance objects, multiple packages that may be unpackaged or packaged (single or multiple layers)
134°C
18 minutes
5 minutes 800 grams 3
load of 800 grams 230 ml
B
45 minutes
- Ambient temperature can have an effect on increasing the total process time. * The total duration of the first process may be longer due to the need for the appliance chamber to warm up. After sterilization at 134°C FAST the loads are wet. It is recommended to leave the drawer ajar in order to evaporate the remaining water and to lower the temperature of the instruments. After sterilization in the 134°C FAST program, the instruments are intended for immediate use.
6
7
1.7. Symbols used on the device
This symbol is located on the front of the appliance, on top of the drawer
front, and advises extreme caution due to the high temperature of the
compartment and its immediate surroundings. This symbol is located on the
unit’s rating plate and identifies its individual serial number. This symbol
appears on the unit’s rating plate and identifies the Notified Body involved
in the conformity assessment process for Directive 93/42/EEC. This symbol is
located on the unit’s rating plate and identifies the year the unit was
manufactured. This symbol is located on the equipment rating plate and
identifies the equipment manufacturer.
This symbol can be found on the unit’s rating plate and requires the user to
read and follow the information in this manual.
This symbol is located on the unit’s rating plate and identifies the
Authorized Representative.
1.8. Precautions, requirements, recommendations
· The user is responsible for the correct installation, proper use and
maintenance of the unit in accordance with the guidelines in this document. If
necessary, contact the service department or your device supplier.
· The sterilizer must not be used in the presence of flammable gases or
explosive vapors in the ambient air.
· At the end of each sterilization cycle, the load is hot. Instruments or
packages should be removed from the chamber using appropriate protective
gloves or other equipment to prevent burns.
· Do not remove the rating plate or other markings from the unit – otherwise
the warranty will be void. · Do not disassemble the unit’s housing or other
parts – otherwise the warranty will be void. · Follow the guidelines for
preparing instruments for sterilization. · Flooding the device with water or
other liquids may cause a short circuit and threaten the user’s safety. · Turn
off the unit and disconnect the power supply before inspecting, performing
maintenance, or
servicing. · Inspection and servicing of the device may be carried out only by
trained service technicians using
original spare parts. · After sterilization at 134°C FAST the loads are wet.
It is recommended to leave the drawer ajar to
evaporate the remaining water and to lower the temperature of the instruments.
· ** After sterilization in the 134 ° C FAST program, the instruments are
intended for immediate use.
8
· The correct and safe operation of the device is based on a thorough reading
and observance of this document, on the installation and use of the device in
accordance with the descriptions given therein and on the observance of all
safety conditions. Any other use, not complying with this manual, may lead to
dangerous accidents, for which the manufacturer will not be held responsible.
· Access to the device must be restricted to unauthorized persons and the
operating personnel must be trained. Operating personnel are understood to be
persons who, as a result of their training, experience and knowledge of the
relevant standards, documentation and regulations relating to safety and
working conditions, are authorized to carry out sterilization and who can
recognize possible dangers and avoid them.
· This document must be supplied with the appliance; it contains detailed
information on assembly and installation as well as commissioning, use, repair
and maintenance. If the appliance is operated as intended, this manual
contains sufficient information necessary for qualified personnel.
· This document should always be kept near the device and easily accessible. ·
The manufacturer reserves the right to make changes that do not affect the
safety of operation and
maintenance of the device, without notice to users. · The manufacturer shall
not be liable for damages during the period of waiting for service,
inspections
and warranty repairs, or any other damage to the Customer’s property other
than the device, in particular for errors resulting from improper installation
or incorrect operation of the device. · Failure to follow the instructions in
this document may endanger the safety of the device user and patients. · The
user must follow all unpacking, installation and operation guidelines –
otherwise the warranty is void. · In particular, the user must ensure the
availability of water of adequate quality – under the pain of losing the
warranty. · It is unacceptable to use any liquids, solutions, chemicals in the
sterilization process – the device can be powered only with water of
appropriate quality. The use of water of incorrect quality or solutions other
than water will void the warranty. · It is the user’s responsibility to
perform timely maintenance and inspections of the device – under penalty of
losing the warranty. · It is the user’s responsibility to select a program
appropriate to the type of load and not to exceed the maximum weight of the
load to be sterilized.
9
2. Scope of delivery and unpacking the device
2.1. Unpacking the device
If the sterilizer has been transported or stored at a temperature and humidity
different from that in the place of installation, wait at least 60 minutes
after delivery. If the device is moved from a cold to a warm place, it may
contain humidity which, by adversely affecting the electrical components of
the device, may damage it when it is turned on.
Carefully unpack the device from its packaging.
The packaging and its contents should be checked for external damage. In the
event of damage, immediately contact the seller or the transport company to
write a damage report. Do not use the device.
It is recommended to leave the carton for possible transportation of the
autoclave.
Used electrical and electronic equipment must not be placed, thrown away or
stored with other waste. Used equipment should be delivered to a local
collection point for used electrical equipment, which is registered with the
respective environmental protection office and conducts selective waste
collection.
2.2. Equipment of the sterilizer
Before starting the installation, it is recommended to check the contents of
the package. The following items should be included in the package 1. Enbio S
/ Enbio PRO sterilizer 2. HEPA filter (inside the chamber of the device). 3.
Power cord. 4. Water and condensate connection hoses (blue and red) with
plastic plugs. 5. Enbio Magic Filter water filter 6. USB flash drive. 7. User
manual 8. Device validation report 9. TÜV certificate 10. Enbio Medical
Sterilization sticker 11. Warranty 12. Enbio Data Viewer software * –
electronic version on pendrive
1
2
3
4
5
6
7
8
9
10
11
12
10
11
3. Installation and preparation for operation
Please read these operating instructions carefully before using the ENBIO S or
ENBIO PRO. During operation, observe the guidelines in these instructions as
well as the applicable local health and safety rules. The ENBIO S and ENBIO
PRO units are designed for self-assembly by the user and do not require
specialized installation on site. The user is responsible for proper
installation of the unit on site in accordance with these instructions.
3.1. HEPA filter installation.
To protect the component during transport, a HEPA filter is not installed in
the device. It was placed inside the chamber, on a tray. Take it out of the
chamber, take it out of the bag and by yourself, carefully screw it into the
slot provided for this purpose (see fig.). The filter should be screwed in by
hand (approx. 7 turns) until resistance is felt – further screwing in will
damage the filter and immobilize the sterilizer.
3.2. Feed water quality
It is unacceptable to use any liquids, solutions, chemicals, additives to the
feed water – the device may be powered only with demineralized or distilled
water of appropriate quality. No chemicals or additives must be added to the
sterilization water, even if they are specifically intended for use in steam
generators, steam production or for use as additives in sterilization,
disinfection, cleaning or corrosion protection. The use of incorrect quality
water or solutions other than water reduces the effectiveness of sterilization
and leads to damage to the device and loss of warranty. The total mineral
content of the sterilization water must be less than 10 ppm or, in the case of
conductivity measurement, it must be less than 15 S/cm – therefore tap water
cannot be used as feed water for the device. The table below shows recommended
water hardness and conductivity parameters for Enbio S and Enbio PRO
sterilizers.
Permissible parameters of water used for sterilization · hardness <0.02 mmol /
l · conductivity (at 20 ° C) <15 S / cm · pH – from 5 to 7.5 · chemical
additives – none
The use of water with an impurity content in excess of the above levels will
damage the device and will void the warranty.
The water in the supply tank should be replaced at least every three months
(due to increased conductivity due to prolonged contact with air). If the tank
is dirty, also change it to a new one. The tank should be closed with the
attached stopper so that water does not deteriorate its chemical composition
as a result of contact with air.
The guarantee granted by the manufacturer becomes void if the autoclave has
been used with water of a quality inconsistent with the recommended
If the autoclave is operated in a water recycling system using the Enbio Magic
Filter, the filter must be replaced with a new one every 6 months or less when
the resin has completely discolored to an amber color. As long as the blue
color is visible, the resin has still water-purifying properties, i.e. water
hardness <0.02 mmol / l and conductivity <15 S / cm. In order to ensure the
correct parameters of the water supplying the device, it is recommended to
check the water tank at least quarterly. In the event of contamination, the
tank should be emptied, cleaned and filled with new demineralized water.
Failure to comply with these obligations may disrupt the sterilization process
and will void the warranty on the sterilizer.
12
13
3.3. Water connection with feed and wastewater tank
To connect the Enbio autoclave, you need 2 water tanks (feed and waste) of
identical capacity, or a feed water tank and drainage access to the sewage
system.
3.3.1 Connecting the WATER IN feed water
· Connect the blue connection hose to the blue feedwater connection, which is
marked on the rear panel of the WATER IN device. The tubing should be inserted
into the connector, to the depth of the black line (placed on the tubing).
· The other end of the blue tube should be immersed in the tank with the feed
water. The device is equipped with a water suction pump, so it is not
necessary to place the water tank above or on the same level as the device.
· To secure and immobilize the water supply hose to the machine, use the plug
supplied with the hose and insert the plug into the opening of the water
supply tank. An unsecured tubing may jump out of the water and cause
sterilization errors.
· To prevent the hose from coiling in the water tank, install the included
weight with a rubber ring at 2 cm from the end of the hose (fig.)
· The minimum water load in the tank is 300 ml. · Remember and check that the
blue tube is always immersed in the water. · Check the water level in the tank
on a regular basis, depending on the frequency of the processes
3.3.2 WATER OUT Wastewater connection
· Connect the red wastewater hose to the orange wastewater connector that is
marked on the rear panel of the WATER OUT device. The tubing should be
inserted into the connector, to the depth of the black line (placed on the
tubing).
· Wastewater should be discharged directly to the water and sewage system or
to a special wastewater tank. If a tank is used, the other end of the red tube
should be placed in the wastewater tank. To secure and immobilize the water
drain hose from the machine, use the plug provided with the hose and insert
the plug into the opening of the wastewater tank. The unsecured tubing may
jump out of the tank and flood the room.
· Make sure that the red tube is never immersed in water, otherwise the water
will not drain properly, causing sterilization errors.
· The wastewater tank or drainage outlet must always be located below the
unit. · When using wastewater tanks, we recommend that you use tanks with the
same capacity as the feed
water tank. Their simultaneous replacement will protect against the
possibility of wastewater overflow. · The water level in the tank should be
checked on a regular basis, depending on the frequency of the
processes.
14
Correct connection of the hoses to the feed and wastewater connections
Correct installation of the weight with the securing ring
Correct positioning of the hoses in the feed and wastewater tanks.
15
20 mm
3.4. Connection with water recycling (Enbio Magic Filter)
Enbio Magic Filter is a filter for recycling water used in the sterilizer. The
ion-exchange resin contained in Enbio Magic Filter allows you to purify
wastewater to the level of deionized water in accordance with the requirements
of EN 13060 – that is, water which is recommended and approved for use with
Enbio autoclaves. You only need 1 water tank to use Enbio autoclave with
filter.
3.4.1 Connection of feedwater WATER IN with Enbio Magic Filter
· Connect the shorter part of the blue Enbio Magic Filter connection hose to
the blue water supply connection, which is marked on the rear panel of the
WATER IN device. The tubing should be inserted into the connector, to the
depth of the black line (placed on the tubing).
· Submerge the longer part of the blue Enbio Magic Filter hose into the feed
water tank. The unit is equipped with a water intake pump, so there is no need
to place the water tank above or at the same level as the unit
· To ensure proper operation, Enbio Magic Filter must always be placed in a
vertical position · To prevent the hose from coiling in the water tank,
install the included weight with a rubber ring at 2
cm from the end of the hose (fig.) · To secure and immobilize the water supply
hose to the machine, use the plug supplied with the Enbio
Magic Filter and insert the plug into the opening of the water tank. An
unsecured tubing may jump out of the water and cause sterilization errors. ·
The minimum water load in the tank is 300 ml. · Remember and check that the
blue tube is always immersed in the water. · Check the water level in the tank
on a regular basis, depending on the frequency of the processes
3.4.2 WATER OUT wastewater connection with Enbio Magic Filter
· Connect the red wastewater hose to the orange wastewater connector that is
marked on the rear panel of the WATER OUT device. The tubing should be
inserted into the connector, to the depth of the black line (placed on the
tubing).
· Wastewater should be drained into a water tank. To secure and immobilize the
water drain hose from the unit, use the plug supplied with Enbio Magic Filter
and insert the plug into the opening of the water tank. The unsecured tubing
may jump out of the tank and flood the room.
· Make sure that the red tube is never immersed in water, otherwise the water
will not drain properly, causing sterilization errors.
· The water tank must always be located below the unit.
16
Correct connection of the hoses to the feed and wastewater connections
Correct installation of the weight with the securing ring
Correct positioning of the hoses in the water tank
17
1 cm
5 cm
3.5. Setting up the device
· The unit should be placed on a flat, level surface. Do not use the unit if
it is at an angle. · Only demineralized or distilled water should be used in
the devices. The use of incorrectly specified
water (see “3.2. Supply water quality”) reduces the sterilization efficiency
and leads to loss of warranty due to damage to the device. · Leave a space of
5 cm at the rear of the appliance and at least 1 cm at the sides of the
appliance from walls or other elements for proper ventilation and fresh air
access. · The unit should be located to provide easy access to the main switch
located on the rear panel of the unit. · Do not place the unit near a sink
with water or other places where the unit could be flooded – danger of short
circuit. · The device should be installed in a well-ventilated room, far from
heat sources and rooms where mixtures of gases or liquids or other dangerous
factors may arise.
The following conditions must be ensured for safe and efficient operation of
the device: · ambient temperature from + 5 ° C to + 25 ° C, · relative
humidity from 0 to 90%, · storage temperature from -20 ° C to + 60 ° C, · 0 to
90% relative humidity.
3.6. Connecting the device to the power supply
Connect the appliance only to power sources that are grounded and equipped
with residual current protection, and that have the same voltage rating as the
appliance (see “12. Technical Data”). It is unacceptable to use extension
cords to connect the device to the power source. Connecting the device to an
incorrect power source may damage it and void the warranty.
18
19
4. Preparation and loading of instruments
The devices are suitable for sterilization of loads for which steam
sterilization is stipulated. The instruments must be sterilized only if they
are clean and dry. Therefore they must be washed and disinfected before being
loaded onto the tray in accordance with applicable regulations. Residues of
used chemicals or other solid particles may render the sterilization process
impossible or even damage the device. In addition, sterilization of
instruments that have not been previously cleaned and disinfected is a
biological hazard and can lead to damage (both to the instruments and to the
sterilizer). For instruments that must be lubricated, use steam sterilization
lubricants. Excess lubricant must also always be removed · The 134 ° C FAST
program is intended only for sterilization of unwrapped instruments (see “1.5
Enbio S
Sterilization Parameters” and / or “1.6 Enbio PRO Sterilization Parameters”).
· After sterilization with 134°C FAST the instruments are moist – it is
recommended to leave the drawer
open for a few minutes to evaporate excess moisture. · After sterilization of
unpackaged loads, they are intended for immediate use.
4.1. Preparation of packages
It is recommended to use sterilization packaging that meets the requirements
of EN ISO 11607-1: 2020 and EN 868-2: 2017. Appropriate packaging should: ·
ensure good penetration of the sterilizing agent into the inside of the
package, · provide resistance to damage during the sterilization process, ·
ensure tight and durable sealing of contents and safe removal for use, ·
provide a barrier to microorganisms and unwanted substances and contaminants,
· Use disposable sterilization sleeves (intended for steam sterilization) as
recommended by their
manufacturer · sleeves should only be filled to about 3/4 full to allow for
proper sealing and minimize the risk of
damage to the package · a distance of at least 30 mm should be maintained
between the seals and the sterilized load. 30 mm · the sharp edges of the load
should be protected to avoid damaging the packaging · the packaging material
must not be placed too loosely or be tightly stretched so as not to affect
pressure variations during sterilization · a label with information about the
contents of the package, the code of the packer, date of sterilization
and use-by date as well as parameters of the sterilization process should be
placed on the package
4.2. Arranging packages on the sterilizer tray (sterilization of packed items)
· packets on the tray should be placed so that the paper side touches the
paper side, because penetration of the sterilizing agent and air exchange can
only take place through the paper
· the packages should be placed on the tray in such a way as to eliminate the
contact of the package with the door seal and the phase of the sterilization
chamber – failure to comply with this rule may cause the chamber to become
unsealed and the sterilization cycle incorrectly carried out
· the edges of the packs must not protrude beyond the sterilizer tray, as this
would cause the chamber to leak and the sterilization cycle to fail
· if the sterilizer chamber is heavily loaded, the first packs must face the
foil side towards the bottom of the tray (this ensures faster and more
efficient drying of the packs).
4.3. Arranging tools on the sterilizer tray (sterilization of loads without
packaging)
· in the case of sterilization without packages – the instruments must be
placed in such a way that they do not have direct contact with each other, no
element of them falls into the openings of the tray, and does not rest against
the edge of the tray or protrude beyond its outline
· Failure to follow the above recommendations may cause permanent and
irreversible damage to the sterilization chamber phase, which will result in
the lack of tightness of the sterilizer and loss of warranty
Do not exceed the maximum load weight: 500 g for Enbio S devices and 800 g for
Enbio PRO devices. In order to verify the correctness of the performed
sterilization, it is recommended to place a sterilization chemical indicator
in the chamber for each process, which discoloration during properly conducted
sterilization.
Example of a packed load.
Example of an unpacked load.
20
21
5. Operating the device
5.1. First run
Turn on the device using the main switch located on the rear panel of the
device. Make sure that the supply and wastewater hoses are connected
correctly, and that there is water in the supply tank and the wastewater tank
is empty. Place the load on the tray, slide the drawer in and lock it, i.e.
turn the knob clockwise. The sterilizer provides audible information about
changes and the completion of the process.
5.2 Program selection
Depending on the type of load to be sterilized, the user must select the
appropriate program – in accordance with the instructions of the sterilizer
manufacturer (see sections: “1.4 Intended use of the device”, “1.5 Enbio S
sterilization parameters” and “1.6 Enbio PRO sterilization parameters”) and
the load manufacturer’s recommendations. · Carrying out sterilization
processes in the ENBIO S / ENBIO PRO device does not affect the
biocompatibility of the materials. · All device components that come into
direct contact with the sterilized load do not cause toxicity,
sensitization or irritation.
Process selection screen Enbio S
22
Process selection screen Enbio PRO
After switching on the device, the start screen appears on the display. To go
to the next screen, press the screen once with your finger (anywhere).
This screen only appears on first run, before the first process is performed.
Each time the device is switched on again, the welcome screen appears in the
display. To go to the next screen, press the screen once with your finger
(anywhere).
On the next screen, we can go to sterilization, test the device or open the information menu. The following sterilization programs can be selected from the main menu: 121 ° C, 134 ° C, 134 ° C FAST and 134 ° C PRION (Enbio PRO only).
Test programs
Sterilization programs
Settings menu
Selection of sterilization programs When the chamber is opened, the DOOR OPEN
symbol flashes.
23
When the chamber is closed by turning the locking knob clockwise, the START
symbol appears on the display, which informs that the chamber is properly
closed.
Now we can select the program by pressing the appropriate symbol of the
temperature in which we want to sterilize 121 ° C, 134 ° C, 134 ° C FAST or
134 ° C PRION, which will highlight the selected program. The selected program
is started by pressing the START symbol.
If the USB memory is not inserted in the device, the USB disk symbol is not
displayed in the lower right corner of the screen.
A message appears that USB memory is missing. The program data will not be
saved. We can continue working without saving data to the USB memory stick by
pressing the YES box, or abort the work by selecting the NO box to place the
stick in the port and start the program steps from the beginning. It is
recommended to use a USB drive with each of the processes. The data stored on
it will allow you to print reports from selected processes.
24
If work continues or the START field has been selected, a symbolic pressure diagram of the entire process is displayed on the screen with the current program progress highlighted against a background and information on the next process steps in the upper left corner of the screen. During the program run, the screen displays the temperature of the selected sterilization program or the current temperature in the process chamber in the lower left corner, the current pressure in the chamber in the lower right corner, and the time remaining until the end of the process in the upper right corner of the screen. This is an estimated time, which may be increased due to the weight and type of the load.
Program stage
Temperature of the selected process
Approximate time to the end of the process
Program flow diagram
The current temperature in the chamber
STOP button
Current pressure in the chamber
During the course of the program, in place of the START field, the STOP field
is displayed, which allows the user to stop the process at any time. In the
upper left corner of the screen, the names of the individual stages of the
program are displayed consecutively, e.g. chamber lock, heating of the working
chamber
If the process was successful, the display alternately shows information
screens about the end of the process and the sterility of the load and about
the possibility of opening the chamber of the device. By pressing the FINISH
field, we go to the welcome screen.
WARNING! After the process, the chamber, tray and load are hot. Be especially
careful and use protective gloves to remove the load or wait until it cools
down. In the 134C FAST program, instruments are hot and wet after
sterilization.
25
5.3 Test programs
The devices are equipped with special test programs to verify the correct operation of the sterilizer.
Enbio S / Enbio PRO Process temperature Number of pre-vacuums Sterilization time Drying time Total process time
Bowie & Dick / Helix 134°C 3 3.5 min 3 min 15 min
Vacuum test 1 16 min
By pressing the Test field you go to the test programs menu. From this level, we can choose between the Vacuum test and Helix / B&D test programs. We select the appropriate program by pressing the required field on the display.
When the device working chamber is closed, the word DOOR OPEN changes to START and by pressing this field the selected test program is started.
If a USB flash drive has not been inserted into the unit, the USB symbol is not displayed in the lower right corner of the screen and a message indicating no USB flash drive is displayed. The test program data will not be saved. We can continue without writing data to the USB flash drive by pressing the YES box or abort by selecting the NO box to place the USB flash drive in the port and start the program from the beginning.
26
Vacuum test
The vacuum test should only be performed on a cold device before starting
work. During the test, the device verifies: · vacuum pump capacity. ·
tightness of the pneumatic system.
After selecting a vacuum test program and starting it with the button, the
vacuum test program run screen is displayed, with Information about loss of
pressure in the working chamber and the duration of the test.
[Vacuum test PASSED] When the test program ran successfully.
[Vacuum test FAILED] When the test program did not run successfully.
After pressing the CONTINUE field, the welcome screen is displayed.
During the vacuum test the sterilizer chamber must be completely dry and cold.
If not, the vacuum test may not be reliable even if the sterilizer is fully
operational. When the test is complete a message indicating the result will
appear on the display. If the result is negative check, clean or replace the
seal, clean the front edge of the chamber and repeat the test. If the Vacuum
test fails again, contact the supplier or manufacturer.
27
Bowie & Dick test
The Bowie & Dick test, also known as the steam penetration test, simulates a
small, highly porous load. To perform the test it is necessary to have a
special test packet and placed inside the chamber. The package is not an
accessory of the device, the user should purchase it on his own. This test
evaluates the device’s performance in sterilizing loads of porous objects: ·
Pre-vacuum performance and steam penetration. · Temperature and pressure of
saturated steam reached for a certain period of time.
How to run the test: · The test must be carried out with an empty chamber in
accordance with EN 13060. · Place the Bowie-Dick test package in the chamber
in the center of the tray.
After selecting the Helix / B & D test program and starting it with the START
button, the program sequence screen is displayed. Information about the
process parameters is displayed.
The Helix / B & D test program can be stopped at any time by pressing the STOP
field, which is associated with incorrect termination of the test.
When the test program is complete, alternating screens are displayed: [READY]
Test complete / [READY] Chamber is safe to open. The process chamber of the
sterilizer can be safely opened.
The control test should be removed. WARNING! The package will be hot. To
correctly interpret the test, refer to the instructions provided by the test
manufacturer.
Open the package and remove the indicator chemical from the center of the
package.
Positive result The chemical indicator turned a uniform dark color over the entire surface.
Negative result A bright field remained in the middle of the test because of the remaining air inside the tested device.
Any color change, uneven coloring of the test, indicates the presence of air during the test cycle due to a malfunction of the sterilizer. If the test result is abnormal, check the expiration date of the test pack and repeat the test.
After opening the process chamber, the welcome screen is displayed.
28
29
Helix test
The Helix test corresponds to the sterilization of instruments with A-holes in
accordance with EN 13060. It consists of a 1500 mm long tube open on one side
and a closed test capsule on the other side. The indicator strip is inside the
test capsule.
Helix test kit
This test is used to evaluate the efficiency of the device in the
sterilization of hollow and porous loads, in particular: · Pre-vacuum
performance and speed and uniformity of steam penetration. · Saturated steam
temperatures and pressures reached for a certain period of time.
How to run the test: · The test must be carried out with an empty chamber in
accordance with EN 13060 · Follow the test manufacturer’s guidelines · Place
the test in the center of the tray in the chamber. · At the end of the cycle,
open the sterilizer and remove the test.
WARNING! The test kit will be hot. To correctly interpret the test, refer to
the instructions provided by the test manufacturer. Open the capsule and
remove the test strip.
Positive result all areas of the indicator strip have turned dark
Negative result Part of the indicator strip did not turn dark due to the presence of air inside the capsule.
Insufficient color change of the indicator strip fields indicates the presence
of air during the test cycle due to a malfunction of the sterilizer. If the
test result is incorrect, check the use-by date of the test pack, additionally
check that the tubing is not clogged and obstructed, and repeat the test.
30
5.4 Information menu
The info menu is accessible by pressing the Info. button. This level displays
information about the unit type, serial number, number of processes performed
and available USB memory space for saving process data, as well as the
COUNTERS service menu – process counters for filter changes. It is also
possible to change the date and time. To set the date or time touch the digits
on the display. When a certain field is selected, it starts flashing and the
arrows for changing the value up or down are displayed. This is how to set the
date and time correctly. Pressing the number again confirms it and you can
move on to changing the settings of the next parameter. In the same way, you
can select the language by clicking on its abbreviation.
The button labeled B turns off and on the blue backlight in the depth of the
screen.
Clicking the LED button launches the backlight control menu on the sides of
the device.
Clicking a button button turns off the button sounds.
Autoclave serial number
Process count
Setting the date and time
Key sounds deactivation Language choice
LED illumination settings (Enbio S only)
Counter settings
Disable shaded background
31
5.4.1 LED lighting
LED lighting has two modes: · Free mode, where the user (by moving the
sliders) freely sets the colors, intensity and
brightness of the light to their preference. · Continuous mode that indicates
the stages of the entire sterilization process with colors LED lighting is
only available in ENBIO S.
Decorative illumination
Illumination during a process step
Decorative illumination color change
If the process count exceeds 1450, the device will inform the operator or user
of this via a warning screen and display this value on the counter screen:
If the process number in exceeds 1500, the device informs the operator or user
to perform a mandatory periodical service.
Decorative illumination color change
Turning the LED illumination on and off
5.4.2 Counters
Enbio S and Enbio PRO sterilizers record the number of processes performed.
This lets you know when you need to replace consumable parts and when a
service inspection is due. ENBIO sterilizers count down to the required
overhaul 12 months or 1500 processes from the time the first process was
performed, whichever occurs first.
Counter section with the counters reset to zero. The number of processes
performed is on the left, while on the right is the number at which the
component should be replaced or a service inspection should be performed.
After replacing the filter, the user can reset the values by pressing the
RESET button. The value for the service check can only be reset by an
authorized service technician.
During the first process (Vacuum, Helix, FAST, 134, 121), the unit will record
the current date as the date of the first process (bottom line on the counter
screen). The device will count processes and days since the first startup.
20 days before the service due date, the device will inform the operator or
user of this via a warning screen and display this value on the counter
screen:
After 12 months of the first process, the device will inform the operator or
user about the necessity to perform the service.
32
33
The counter can only be reset by Enbio or an authorized external service by selecting the RESET button on the counter screen and entering a unique service code.
Displaying the counter values in yellow or red does not block the operation of the unit. However, exceeding the required replacement time may have a significant impact on the operation of the unit and the sterility of the load. For replacement of individual components, please contact the manufacturer or supplier.
No.
Name
Replacement frequency (cycles)
1
HEPA filter
1000
Yellow Early replacement call (cycles)
980
Red Latest replacement after (cycles)
1000
5.5 Restart
The restart of the process is forced when the user stops the process by
pressing the STOP field, in the case of a power outage or an error during the
process, for example lack of supply water. If the STOP field is selected, the
following messages are displayed alternately: · the user has stopped the
process · equalizing the pressure in the working chamber · process is
incorrect, which means that the load is not sterile.
When the pressure in the working chamber is equalized, the following messages
appear alternately on the display. Now you can open the device freely. After
opening the chamber, the screen appears.
By selecting the field, we can return to the welcome screen. In the event of an error, we must additionally enter the 4-digit security code 0000. Entering this code is tantamount to the operator’s declaration that he is aware that the sterilization process has not been carried out properly and that the batch is non-sterile.
34
35
If an incorrect code is entered, a message will appear on the display.
Enter the code again. The arrow enables erasing incorrectly entered digits.
After entering the correct code, a welcome screen will appear on the display
of the device.
6. Maintenance and upkeep
Cleaning the tray
Keeping the tray clean helps to ensure proper operation of the appliance. The
tray and its technical condition are a good indicator of using the correct
water. A stony, brown tray indicates the use of poor quality water. It is
recommended to clean the inside of the tray on a weekly basis with a mild
detergent that does not add chlorine and does not react with aluminum. After
cleaning, the tray requires thorough rinsing with water. In order to clean the
tray properly, it should be removed from the front of the device.
Enbio S Lift the tray up gently and pull away from the front face. The mounting pins have notches into which the drawer fits. Before reinstalling the tray into the unit, drain the tray and slide it over the front face pins and press down gently to lock.
Enbio PRO Unscrew the 3 screws securing the tray to the front face. Remove the tray from the autoclave. Dry the tray before putting it back into the device.
Cleaning the process chamber
Keeping the chamber clean helps to ensure proper operation of the device. It
is recommended to clean the inside of the process chamber once a week with a
mild detergent without the addition of chlorine. After cleaning, the chamber
should be wiped dry with a soft cloth. Cleaning should be performed on a cool,
cooled chamber
Cleaning of external surfaces
Cleaning the external parts of the device should be done with a soft cloth
moistened with water and a mild detergent (no chlorine added and not reacting
with plastics, varnish coatings, aluminum). Strong detergents should not be
used. The use of mild detergents to maintain the device does not affect the
risk of toxic components coming into contact with device components.
36
37
Cleaning the gasket
It is recommended to clean the gasket each time after 100 processes. Use warm,
clean water and microfiber to clean the gasket (microfiber with silver
particles is allowed). Blunt and sharp cleaning tools are not allowed.
Cleaning with chemicals is not allowed. Carry out cleaning on a cooled down
device, after opening the drawer. Be careful not to bend the drawer. After
cleaning, leave the device open until the gasket is dry. During this time,
protect the device against damage.
Replacement of consumables
The sterilizer is equipped with a high performance sterilization system. A
message on the screen informs the user when each component should be replaced.
If the sterilizer is in regular operation, alternating replacement screens
will appear after the welcome screen has been pressed. The replacement screens
are described in detail in section “9. Warning Messages and Error Codes”.
Cleaning the water container
In order to ensure the correct parameters of the water supplying the device,
it is recommended to check the water tank at least quarterly. If contamination
is found, the tank should be emptied, cleaned and refilled with new water.
In order to ensure an efficient sterilization process and the correct
functioning of the device, it is recommended to replace consumables in a
timely manner.
Replacing the Enbio Magic Filter
The filter should be replaced with a new one once every 6 months or sooner
when the resin has completely discolored to an amber color. Failure to do so
may disrupt the sterilization process and will void the sterilizer warranty.
Dispose of the used filter according to local guidelines. The filter and all
its parts are not recyclable.
New filter
Filter half used up
Need to replace
6.1 Consumables
The table below contains items subject to periodic replacement and items subject to natural wear. Spare parts must be ordered directly from the manufacturer. The use of other spare parts voids the warranty and does not guarantee the correct functioning of the device.
Name HEPA filter Enbio S / PRO Enbio S front kit with gasket Enbio S tray Enbio PRO front kit with gasket Tray set 3 pieces Enbio PRO Wastewater set (plug + red hose 1.5 m) Feed water set (plug + blue hose 1.5 m) Enbio Magic Filter
Part no 1-8-27720A5 1-8-14672D2 1-8-14738B2 1-8-12433A45 1-8-1097954A1 1-8-42988A1 1-8-42992A1 1-8-1116687A5
38
39
To ensure proper functioning of the Enbio S / Enbio PRO steriliser, it is recommended that the wearing parts be replaced according to the schedule below. And periodic inspection of the individual components of the sterilizer in accordance with the following guidelines.
Name HEPA bacteriological filter Connection / drain hose
Plugs for water / condensate containers Front drawer with seal
Enbio Magic Filter
Replacement frequency Every 1000 cycles or every 12 months If damage is
observed or once a year
If damage is observed Replacement at mandatory service after 1500 cycles/365
days Replace the filter once every 6 months or sooner when the resin has
completely discolored to an amber color
Element to be controlled Front gasket
Bacteriological filter Connection / drain hose
Container stoppers Enbio Magic Filter
Frequency of inspections
weekly or in the event of incorrect operation – performed by the user
every week – performed by the user
weekly or in the event of incorrect operation – performed by the user
weekly – performed by the user
weekly – performed by the user
7. Data archiving
The course of each performed sterilization is automatically saved on a data
carrier (USB stick). · The memory slot is located on the rear panel of the
device. · It is recommended to periodically archive the data on another medium
e.g. computer, laptop. · The USB flash drive should not be removed from the
slot during the process.
USB socket USB flash drive
40
41
8. Enbio Data Viewer software
EnbioDataViewer software allows you to view and archive sterilization programs
on your computer and print them out. Minimum hardware requirements to install
the software: · Operating system – Windows min. Windows 7 or higher · Free
disk space – min. 100 MB · Minimum processor requirements – min. 1 GHz ·
Minimum operating memory – min. 512 MB Ram · Screen resolution – min. 1200×720
or higher
8.1 Installing the software
The software is delivered with the device and is on a portable drive – USB
flash drive or the latest version can be downloaded from the manufacturer’s
website http://enbio.com/service/
To install the software, double click on the software installation file. After
this operation an installation window will be displayed regarding language
selection.
After confirming, you must accept the license terms for the software being
installed.
42
You will then be prompted to place a shortcut to the software on your computer
desktop.
After clicking “Next>”, the folder where the program should be installed is
selected.
After making your selection, click “Next”. By clicking the Install button, the
Enbio Data Viewer software will be installed.
43
After the installation, the following message appears.
8.2 Program structure and main functionalities
The main window consists of three main areas
We can now run the software or finish the installation without running the
software by clicking the Finish button.
If you select the option with immediate start, the main program window is
displayed.
Tree of all processes that have been synchronized with the flash drive have been sorted by execution dates
Temperature and pressure graph with main autoclave and process data (date and number).
The function buttons, eg “PDF report”, which will enable printing the protocol from the process, are marked in dark blue.
Data on the duration and achieved parameters of each process step.
The most important sterilization parameters.
Possibility to save notes for each process.
44
45
Drop-down menu:
By clicking on the File window, we have access to the following options: ·
loading the saved process flow from a flash drive or from another location ·
print a saved process · create a report to a PDF file · exporting data to a
file in order to send a package with data to the manufacturer in case of
problems · data export to CSV format · closing the program
By clicking on the drop-down menu in Help, we have access to the following options: about the program and its version
By clicking on the Tools window, we have access to the following options: ·
synchronization of all files with saved processes after selecting memory
location on
pendrive (this process, depending on the number of sterilizations and tests
performed can take up to several seconds) · searching for any saved process
from the database · adding your own logo to PDF reports
46
Search
The program allows you to search for processes by: · date range · process
number · process type · process outcome (success/fail)
Process no. 03052
Date 08-03-22
Program 134 Fast (S)
Start 12:07:41
Stop 12:15:02
S/N ST01-CH-21-10835
Firmware 33.3
Stage
0 Locking door 1 Heating up 2 Vacuum I 3 Steam Pulse I 4 Vacuum II 5 Steam
Pulse II 6 Vacuum III 7 Pressurizing 8 Sterilizing 9 Depressurizing 10 Drying
11 Equalizing 12 Ulocking door
Time [mm:ss]
Step [mm:ss]
p [bar]
t [C]
00:01 01:15 -0.858 111.8
01:17 02:18 06:00
01:00 03:41 01:16
2.198 2.210 -0.396
136.2 136.3 94.6
07:17 07:20
00:02
-0.152 94.5 -0.096 94.5
t calc. [C] 111.8
136.2 136.6 94.6 94.5 94.5
Sterilization Complete
STERILIZATION TEMPERATURE, Time Min Temperature Max Temperature Min Pressure Max Pressure
134
03:41 136,1 136,6 2,17 2,25
Label
Print 11-04-2022
PDF report
The program allows you to generate a report on each process performed by the
autoclave. It contains all the necessary process data and the sterilization
result.
47
9. Warning messages and error codes
In the event of a device malfunction, the display shows the relevant
information, warning and error codes.
9.1 Warning messages
The warning messages refer to the replacement of individual consumables. The
item to be replaced is highlighted in red, and the screens are displayed
alternately.
Filter replacement screen
9.2 Information messages
Mandatory inspection screen
Screen about overpressure or underpressure resulting from the natural processes of cooling the chamber. It can occur immediately after starting the machine.
Message resulting from interruption of the process after the sterilization stage – during drying.
48
9.3 Error codes
The table below contains error codes that may appear when working with the
ENBIO S / ENBIO PRO sterilizer
No Error code
Description
Recommendations
1 “Chamber over temperature”
Maximum chamber temperature exceeded
Contact with the service
2 “Steam gen. Over temperature”
Steam generator temperature too · Too high weight of sterilized instruments –
high
repeat the process with less instruments
(max. 0.5 kg S, 0.8 kg PRO)
· Contact with the service
3 “Process over temperature”
Process temperature too high
Contact with the service
4 “Overpressure error”
Pressure error
Contact with the service
5 “Sterilization pressure too Drying pressure too low low”
· Check that there is water in the bottle with the blue hose
· Correct the position of the blue hose so that the end is completely
submerged in water. Add a sinker to eliminate the problem in the future
· Check that the water supply hose (blue) is not damaged (After correcting the
position / replacing the hose or refilling with water, restart the machine)
· Contact with the service
6 “Sterilization temp. Too low”
Sterilization temperature too low
· Check the water level in the bottle with the blue hose
· Check that the red tube is not pointing up along its entire length, creating
the socalled air trap
· Contact with the service
7 “Too high pressure during Drying pressure too high drying”
· Make sure the red drain hose is not immersed in water. The hose must not be
kinked, the liquid must flow down by gravity
· Check that the weight of the sterilized instruments is not too high
· Contact with the service
8 “Too many steam pulses / Too many steam pulses. No feed · Check the water connection to the “water
no water”
water.
in” connector
· Check the distilled water level in the feed
water tank (blue plug)
· Check that the weight of the load does
not exceed the allowable weight.
· Contact with the service
9 “Drainage error”
Clogged drain
· Check the level of the wastewater and the connection of the hoses
· Check the level of the used water in the bottle with the red cap. If the
bottle is full, discard the used water
· Check that the red hose is not kinked and that it is pointing downwards
along its entire length
· Check that there is no debris in the outlet opening (inside the chamber)
· Contact with the service
49
10 “Chamber heating error” Chamber heating error
11 “Steam generator heating error”
12 “Prevacuum fail / check outlet”
Steam generator error Vacuum pump / drain fault
13 “Power failure”
Momentary voltage loss during operation
14 “Pressure during standby” Overpressure while waiting
15 “Locking door error” 16 “Unlocking door error”
Door lock error Door unlock error
17 “Valve V3 / HEPA filter error”
V3 valve / HEPA filter error
18 “Pressure sensor error”
Pressure sensor error
19 “USB disc error / Change Writing error on pendrive –
disc”
damage to medium
20 Min. Chamber temperature Chamber temperature too low during the process
21 Chamber temperature sensor failure
Chamber temperature sensor failure
22 Steam gen. Temp. Sensor Steam generator temperature
failure
sensor failure
23 Process temp. Sensor failure
Process temperature sensor failure
24 Autoclave has too low temperature
Autoclave temperature too low / temperature sensor error
31 “Internal flash error”
Internal memory error
· Mains voltage too low – consult an electrician for the location where the
autoclave is to be installed
· Contact with the service · Repeat the process · Contact with the service ·
Check the level of used water in the bottle
with the red cap and pour out the excess · Check that the red hose is not
submerged
or kinked · Check that the autoclave setting provides
free airflow for cooling the unit · The red hose must point downwards
along its entire length, no section may point upwards · Clean the chamber door
seal · Contact with the service · Restart the device and make sure it is
properly plugged into the outlet · Consult an electrician competent for the
site where the autoclave is installed to check the installation · Restart your
device · Contact with the service Contact with the service · Turn off the
autoclave and turn it on again, start the process and stop it after a few
seconds. There must be no overpressure in the chamber, i.e. the following
information must be displayed: “READY / Chamber is safe to open” · Contact
with the service · Replace the HEPA filter · Contact with the service Contact
with the service Copy the contents from your current flash drive – buy and
install a new one Contact with the service
Contact with the service
Contact with the service
Contact with the service
· Leave the device switched off for 3 hours at room temperature
· Contact with the service Contact with the service
Messages “Aborted by user”
“Vacuum test failed” “No USB memory” “Equalizing pressure”
“Overpressure during standby”
Process interrupted by the user. Non-sterile cartridge if interrupted during
or before the sterilization process.
Vacuum test error
This message appears when the user terminates the process. This does not mean
that there is a malfunction. Start a new process.
Contact with the service
No USB memory stick
Pressure when stationary. Equalization of pressure to atmospheric.
Hypertension in standby mode
Check the USB port and mount the memory. Contact with the service.
· The message occurs in certain cases as a result of natural processes.
· If the message appears frequently, contact the service center.
The reason for this error is that the hot sterilizer is left with the chamber
closed (e.g. overnight). As the sterilizer cools down a vacuum is created in
the chamber which causes a startup error. Wait until the device has equalized
the pressure automatically – the message will disappear automatically
50
51
Here are some examples of error codes: Alternating screens: pressure
equalization, please wait.
A QR code is displayed in error message screens. By scanning this code with a
mobile phone with the option of reading QR codes, the user will be redirected
to a website containing recommendations on the possibility of eliminating the
error.
10. Complaints procedure
In order to report a problem with the device, complete the complaint form on
the manufacturer’s website www.enbio.com or contact the hotline. In the event
of transport damage, a claim should be sent along with a bill of lading and a
purchase document as well as photos documenting the damage. If you have any
questions, please contact us by e-mail at: support@enbio.com
WARNING! The complaint process will be started when the service department
receives a correctly completed complaint application.
When returning the device for service, clean the chamber and the tray of the
device, carry out the decontamination process and secure it properly during
transport. The device should preferably be returned in its original packaging.
In the absence of appropriate packaging, please contact the service or
supplier. If you need to transport the device: · Disconnect the demineralized
water and condensate hoses · Allow the working chamber to cool. · Use original
or suitable packaging with protective inserts Damage caused during transport
to the service due to improper protection of the device is the responsibility
of the sender.
11. Warranty conditions
Enbio S / Enbio PRO sterilizers are covered by a standard 12- or 24-month
warranty. Please contact the supplier for details.
52
53
12. Technical data
Device parameters Power supply Installed power Maximum current consumption Working pressure Maximum pressure Maximum process temperature
Enbio S 220-240V AC 50 / 60Hz 2.25 kW max. 10 A 2,1 bar 2,3 bar max 137°C
Process chamber volume
2,7 l
Weight
15 kg
Process chamber dimensions (LxWxH) 292 x 192 x 45 mm
External dimensions of the device (LxWxH)
561 x 252 x 162 mm
Degree of protection
IP20
Noise level
38dB(A)
Archiving process data
Pendrive
Enbio PRO 220-240V AC 50 / 60Hz 3.25 kW max. 15 A 2,1 bar 2,3 bar max 138°C
5,3 l 20 kg 300 x 200 x 90 mm 561 x 270 x 202 mm
IP20 40dB(A) Pendrive
Environmental conditions Working temperature range
Relative humidity Storage temperature range Relative humidity Ambient pressure
range
from + 5°C to + 25°C
0-90% -20°C to + 60°C 0-90% 900-1100 hPa
Enbio Group AG Eichengasse 3 4702 Oensingen Switzerland
54
Test connector – for use by authorized service only. If found to be used by the user, it will void the warranty.
Rating plate located on the bottom of the device.
Enbio S
Manufacturer Enbio Group AG Eichengasse 3 4702 Oensingen Switzerland
Enbio Technology sp. z o.o. ul. Sportowa 8C 81-300 Gdynia, Poland
Enbio STM
Class B
RoHS PATENTED
Power supply 220-240V AC 10A 50/60Hz 2,25kW max
2022 ST01-CH-XX-XXXXX
Enbio PRO
Enbio PROTM
Class B
Manufacturer Enbio Group AG Eichengasse 3 4702 Oensingen Switzerland
Enbio Technology sp. z o.o. ul. Sportowa 8C 81-300 Gdynia, Poland
RoHS PATENTED
Power supply 220-240V AC 15A 50/60Hz 3,25kW max
2022 ST02-CH-XX-XXXXX
55
13. EC declaration of conformity
EC Declaration of Conformity
Company: Enbio Technology Sp. z o. o., ul. Sportowa 8 C, 81-300 Gdynia, Poland
declares with sole responsibility, that medical devices: STEAM STERILIZERS,
models: · Enbio S · Enbio PRO · Enbio XS
complies with provisions of Council Directive 93/42/EEC of 14 June 1993
concerning medical devices (as amended). The device has been classified in
Class IIb in accordance with rule 15 of Annex IX of the above mentioned
Directive.
Conformity assessment has been carried out in accordance with Annex II without
point 4 of the above Regulations.
The conformity assessment has been conducted by the Notified Body No. 2274 TUV
Nord Polska Sp. z o.o. Mickiewicza 29 Street 40-085 Katowice Poland
Gdynia 14.01.2021
Marek Krajczyski President of the Board
Sebastian Magrian Member of the Board
Version: 05_01.2021
56
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>