AIR TECHNIQUES ULR01.1A1 ScanX Smart Reader User Manual
- June 8, 2024
- AIR TECHNIQUES
Table of Contents
ULR01.1A1 ScanX Smart Reader
User Manual ScanX Smart Reader
ULR01.1A1
EN-US Installation and Operating Instructions
ULR01.1A1 ScanX Smart Reader
Important information
1 About this document
These installation and operating instructions are an integral part of the
unit.
Air Techniques shall not be held liable and offers no guarantees of the safe
and smooth operation of this unit if you fail to comply with notes and
instructions contained in these Installation and Operating Instructions.
The German version of the installation and operating instructions is the
original manual. All other languages are translations of the original manual.
These Installation and Operating
Instructions apply to order number: 2162100010 (ULR01.1A1).
1.1 Warnings and symbols
Warnings
The warning notes in this document highlight possible injury to persons or
damage to machinery. They are marked with the following warning symbols:
General warning symbol
The warnings are structured as follows:
SIGNAL WORD
Description of type and source of danger
Here you will find the possible consequences of ignoring the warning
❯ Follow these measures to avoid the danger.
The signal word differentiates between different levels of danger:
– DANGER
Direct danger of severe injury or death
– WARNING
Possible danger of severe injury or death
– CAUTION
Risk of minor injuries
– NOTICE
Risk of extensive material/property damage
Miscellaneous symbols
These symbols are used in the document and on or in the unit:
| Note, e.g. specific instructions regarding the efficient use of the
unit.
---|---
| Part number
| Serial number
| Model number
| CE mark
| Manufacturer
| Distributor
| Dispose of correctly in accordance with EU Directive 2012/19/EU (WEEE).
| Take note of the accompanying electronic documents.
| Please note the operating instructions.
| Refer to Operating Instructions.
| Wear hand protection.
| Do not reuse
| DC current
1.2 Copyright information
All electronic drawings, processes, names, soft- ware, and appliances
mentioned here are protected under copyright.
Printing or copying these Installation and Operating Instructions, including
excerpts thereof, may only be carried out with the written approval of Air
Techniques.
Safety
The unit has been developed and designed appropriately such that hazards are
largely excluded if the unit is used in accordance with its
Intended use.
Therefore, please note the following.
Despite this, the following residual risks can remain:
– Personal injury due to incorrect use/misuse
– Personal injury due to mechanical effects
– Personal injury due to electric shock
– Personal injury due to radiation
– Personal injury due to fire
– Personal injury due to thermal effects on the skin
– Personal injury due to lack of hygiene, e.g. infection
2.1 Unauthorized modification
Pursuant to Part 15.21 of the FCC rules, any changes or modifications to this
equipment not expressly approved by the manufacturer may cause, harmful
interference and void the FCC authorization to operate this equipment.
2.2 General safety information
❯ Comply with the guidelines, laws, rules and regulations applicable at the
site of operation when you use this unit.
❯ Prior to each use, check the function and proper condition of the device.
❯ Do not convert or modify the unit.
❯ Comply with the Installation and Operating Instructions.
❯ Make the Installation and Operating Instructions always available to the
operator in the vicinity of the device.
2.3 Specialist personnel
Operation
Persons operating the unit must ensure safe and correct handling based on
their training and knowledge.
❯ Instruct or have every user instructed in handling the unit.
Installation and repairs
❯ All installation, resetting, alteration, expansion, and repair work must be
carried out either by Air Techniques personnel or by a suitably qualified
person approved by Air
Techniques.
2.4 Protection from electric shock
The device is designed for use in a basic electromagnetic environment with
connection to the public power supply network in accordance with IEC 61326‑1
(EN 61326‑1).
❯ Replace damaged cables or plugs immediately.
2.5 FCC note
This device complies with Part 15 of the FCC Rules. Operation is subject to
the following two conditions:
❯ This device may not cause interference.
❯ This device must accept any interference, including interference that may
cause undesired operation.
Note: This equipment has been tested and found to comply with the limits for a
Class B digital device, pursuant to part 15 of the FCC Rules.
These limits are designed to provide reasonable protection against harmful
interference in a residential installation. This equipment generates, uses and
can radiate radio frequency energy and, if not installed and used in
accordance with the instructions, may cause harmful interference to radio
communications.
However, there is no guarantee that interference will not occur in a
particular installation. If this equipment does cause harmful interference to
radio or television reception, which can be determined by turning the
equipment off and on, the user is encouraged to try to correct the
interference by one or more of the following measures:
❯ Reorient or relocate the receiving antenna.
❯ Increase the separation between the equipment and receiver.
❯ Connect the equipment into an outlet on a circuit different from that to
which the receiver is connected.
❯ Consult the dealer or an experienced radio/TV technician for help.
2.6 ISED Statement
This device contains licence-exempt transmitter(s)/receiver(s) that comply
with Innovation, Science and Economic Development Canada’s licence-exempt
RSS(s). Operation is subject to the following two conditions:
- This device may not cause interference.
- This device must accept any interference, including interference that may cause undesired operation of the device.
Canada ICES-003 compliance statement This Class B digital apparatus complies
with Canadian ICES 003.
2.7 Only use genuine parts
❯ Only use accessories and optional items specified or approved by Air
Techniques.
❯ Only use original working parts and spare parts.
Air Techniques accepts no liability for damage resulting from the use of
nonapproved accessories, optional items or any parts other than original spare
and wear parts.
The use of non-approved accessories, optional items or non-genuine wear parts
/ replacement parts (e. g. mains cable) can adversely affect the electrical
safety and EMC.
2.8 Transport
Only the original packaging ensures optimum protection for the unit during
transport.
If necessary, the original packaging for this unit can be ordered from Air
Techniques.
Air Techniques cannot be held responsible for any damage resulting from
transport in unsuitable packaging, even during the warranty period.
❯ Only transport the unit in its original packaging.
❯ Keep all packaging away from children.
❯ Do not expose the unit to any strong vibrations or shocks.
2.9 Disposal
Disposal of the units, electronic circuitry and PSPs must be accomplished only
at the appropriate facilities for recovery and recycling.
Make sure to dispose of such items in accordance with current federal,
national, state and local government rules and regulations.
2.10 Protection from cybersecurity threats
The unit is to be connected to a computer that can be connected to the
Internet. Therefore, the system needs to be protected from threats from the
Internet.
❯ Use antivirus software and update it regularly.
Look for evidence of possible virus infection and, if applicable, check with
the antivirus software and remove the virus.
❯ Perform regular data backups.
❯ Provide access to units only to trustworthy users, e.g. by means of user
name and password.
❯ Make sure that only trustworthy contents are downloaded. Install
manufacturer-authenticated software and firmware updates only.
Overview
- ScanX Smart Reader
- USB cable
3.1 Scope of delivery
The following items are included in the scope of delivery (possible variations
due to country-specific requirements and/or import regulations):
ScanX Smart Reader . . . . . . . . . . . . . .. G8809
– ScanX Smart Reader basic device
– USB cable (2 m)
3.2 Accessories
The following items are required for operation of the device, depending on the
application:
USB cable (2 m) . . . . . . . . . . . . . . . . . . . . G8808
Technical data
Electrical data of the device
Nominal voltage (supplied via USB)| V DC| 5
Max. current consumption| mA| 200
Max. power consumption| W| < 1
Power draw in standby| W| 0.16
General technical data
Dimensions (W x H x D)| mm| 90 X 13 X 90
in| 3.54 X 0.51 X 3.54
Weight| kg| approx. 0.13
lb| approx. 0.29
Ambient conditions during operation
Temperature| °C| +10 to +35
°F| +50 to +95
Relative humidity| %| 20 – 80
Air pressure| hPa| 750 – 1060
Elevation above sea level| m| < 2000
ft| < 6562
Ambient conditions during storage and transport
Temperature| °C| -20 to + 60
°F| -4 to +140
Relative humidity| %| Oct-95
Air pressure| hPa| 750 – 1060
USB Connector
USB technology| | USB 2.0 or higher
Data Rate| | Full Speed
Support for high-performance devices| mA| 500
Connection, device side| | USB type C
Technical data for the RFID module
Frequency| MHz| 13.56
Supported standard| | ISO/IEC 15693
Modulation| | ASK
Max. power| mW| 400
4.1 Model identification plate
The type plate is located on the underside of the device.REF Order number
SN Serial number
4.2 Conformity assessment
This device has been subjected to conformity acceptance testing in accordance
with the current relevant guidelines of the European Union.
This equipment conforms to all relevant requirements.
4.3 Simplified declaration of conformity
The manufacturer hereby declares that the device satisfies the requirements
of, among others, Directive 2014/53/EU.
The full text of the EU declaration of conformity can be viewed online in the
Download Center:https://qr.duerrdental.com/conformity
Function
- Status LED
- Scanner
The Smart Reader is an RFID reader that connects to a computer via a USB
cable. It reads RFID chips and transmits the read data to software on the
connected computer.
With the Smart Reader and SmartScan, for example, a phosphor storage plate is
assigned to a specific patient via the imaging software.
This allows the phosphor storage plate to be read without a scan job on all
devices in the network that support SmartScan.
5.1 Status LED
Depending on the software used, the status LED displays different colors.
Refer to the relevant descriptions for the status LED in the software used.
5.2 Connections
The connections are located on the rear of the unit.
- USB port (type C)
Requirements
6.1 Installation/setup room
The room chosen for set up should fulfil the following requirements:
– Closed, dry, well-ventilated room
– Should not be a room made for another purpose (e.g., boiler room or wet
cell)
6.2 System requirements
For the system requirements of the computer systems, visit the download area
at www.airtechniques.com (document no. E7201).
Installation
7.1 Setting up the unit
Portable and mobile HF communication appliances can interfere with the
effectiveness of electrical medical devices.
❯ Do not stack the unit next to or together with other appliances.
❯ If, however, this unit is operated next to other units or stacked with other
units, monitor the unit carefully in the configuration selected in order to
ensure normal operation.
❯ Place the unit on a firm, horizontal surface.
7.2 Connecting the unit
Combining devices safely
❯ When connecting the unit to other devices, such as a computer system: make
sure that e.g. the computer, monitor, or printer that is/are connected is/are
compliant with at least the standard IEC 60950‑1 or IEC 62368‑1.
Connecting the unit via the USB port
❯ Plug the supplied USB cable into the USB port of the device.❯ Connect the USB cable to the computer.
Commissioning and first start-up
NOTICE
Short circuit due to the build-up of condensation
❯ Do not switch on the unit until it has warmed up to room temperature and is
dry.
The unit supports the following imaging programs:
– VisionX made by Air Techniques
Always use the current version of the imaging program in the commissioning of
the device. Check the version of the enclosed imaging program versus the
versions available at www.airtechniques.com.
The device is a so-called ‘plug & play’ device, so it can be used straight
away after connecting it.
Operation
The device scans RFID chips from different sources. Whether or not the scanned
RFID chip can be used depends on whether it is supported and on the devices
and
software used.
9.1 SmartScan
Use SmartScan to assign a phosphor storage plate to a particular patient via
the imaging software. As soon as a phosphor storage plate is assigned to a
patient in the imaging
software, all devices that support SmartScan will change status to “ready for
image acquisition”. Afterwards any phosphor storage plates that were
previously linked to a patient can be scanned on any of these devices in an
arbitrary order. The images are then automatically assigned to this patient by
the imaging software.
SmartScan works with the following imaging programs:
– VisionX made by Air Techniques
Cleaning and disinfection
When cleaning and disinfecting the unit and its accessories, comply with
national directives, standards and specifications for medical products as well
as the specific specifications for dental practices or clinics.
NOTICE
The use of unsuitable agents and methods can damage the unit and accessories
❯ Only use the disinfection and cleaning agents specified or approved by Air
Techniques and the EPA.
❯ Comply with the operating instructions of the disinfectants and cleaning
agents.
** Wear hand protection.
10.1 Surface of the unit
The surface of the unit must be cleaned and disinfected if it is contaminated
or soiled.
NOTICE
Liquid can cause damage to the unit**
❯ Do not spray the unit with cleaning agents or disinfectants.
❯ Make sure that liquid penetrates into the unit.
❯ Remove any soiling with a soft, lint-free cloth that has been dampened with
cold tap water.
❯ Disinfect the surfaces with a disinfectant wipe.
Alternatively, use a spray disinfectant on a soft, lint-free cloth. Comply
with the operating instructions of the disinfectant.
Maintenance
The unit is maintenance-free.
Tips for operators and service technicians
Any repairs above and beyond routine maintenance may only be done by suitably
qualified personnel or by one of our service technicians.
12.1 Software error
Error | Possible cause | Remed y |
---|---|---|
Imaging software fails to recognize the unit | Cable between unit and computer | |
not correctly connected | ❯ Check the connecting cable. | |
Connecting cable between device and computer defective | ❯ Check the connecting |
cable and use a new cable if necessary.
Computer does not detect any connection to the unit| ❯ Check the connecting
cable.
❯ Use a different USB port.
Hardware error| ❯ Contact technician.
12.2 Fault on the unit
Error | Possible cause | Remedy |
---|---|---|
Unit does not switch on | Device not connected correctly, or connecting cable | |
or USB port defective | ) Check the connecting cable and use a new cable if |
necessary.
) Use a different USB port.
Hardware error| ) Contact technician.
Device does not recognize the phosphor storage plate| Incorrect phosphor
storage plate used| ) Only use phosphor storage plates that are supported by
the device.
RFID tag damaged or detached|
) Check the phosphor storage plate for damage and replace it if required.
Device not responding even though the status LED is lit up| Fault in the
device or in the soft- ware| ) Depending on the software used, the status LED
displays different colors. Refer to the relevant descriptions for the status
LED in the software used.
Hersteller / Manufacturer:
DÜRR DENTAL SE
Höpfigheimer Str. 17
74321 Bietigheim-Bissingen
Germany
Fon: +49 7142 705-0
www.duerrdental.com
info@duerrdental.com
Manufactured for / Distributed by:
Air Techniques, Inc.
1295 Walt Whitman Road
Melville, New York 11747-3062, USA
Phone: 800-247-8324
Fax: 888-247-8481
www.airtechniques.com
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>